Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Similar documents
Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Determination of Volatile Organic Compounds in Air

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Optimizing Standard Preparation

Maximizing Production While Minimizing Costs

Solid Phase Micro Extraction of Flavor Compounds in Beer

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Optimizing. Abstract: is standardd. procedures. altered to

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Maximizing Sample Throughput In Purge And Trap Analysis

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Solid Phase Micro Extraction of Flavor Compounds in Beer

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Methanol Extraction of high level soil samples by USEPA Method 8260C

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Ed George and Anaïs Viven Varian, Inc.

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

2017 EPA Method Update Rule and EPA Method 624.1

EPA TO-17 Volatile Organic Compounds

Target Compound Results Summary

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Analytical Trap Comparison for USEPA Method 8260C

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

CALA Directory of Laboratories

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

Kendrick mass = IUPAC mass x (14/ ) (1) Kendrick mass defect = nominal Kendrick mass exact Kendrick mass (2)

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

STANDARD OPERATING PROCEDURES

In-Tube Extraction Sample Preparation for Gas Chromatography

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Meeting NJ Low Level TO-15 Air Testing Method Requirements

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Thermal Desorption Technical Support

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Associates. Project No.: October 25, 2017

Study of Residual Solvents in Various Matrices by Static Headspace

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Volatile Organic Compounds in Water PBM

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Readiness thru Health

Copies: Dave Favero, RACER File

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Table 1 Soil and Groundwater Sampling Summary Table

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Headspace Technology for GC and GC/MS: Features, benefits & applications

CALA Directory of Laboratories

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Transcription:

JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling Anne Jurek Introduction: #134 Static headspace sampling has always been a viable option for the detection of Volatile Organic Compounds (VOCs) and is readily used in Europe and Canada for testing water samples. In order to detect low level contamination of water, it is essential for the static headspace sampling and GC/MS analysis parameters to be enhanced. This poster will investigate the optimization and detection of VOCs in water using static headspace analysis. Discussion: The USEPA methods for extracting VOCs from environmental samples require purge and trap sampling. Due to this constraint, environmental laboratories in the United States do not use static headspace analysis. However, in Europe and Canada, it is common to use static headspace sampling for the measurement of VOCs. Purge and trap sampling involves purging the VOCs out of the matrix and trapping the analytes onto an analytical trap, the trap is then desorbed to the GC/MS. This process has a number of pros and cons. On the positive side, purge and trap is more sensitive. It is also the recommended sampling technique for USEPA Methods. Furthermore, the advent of autosampling systems has simplified sample preparation. However, purge and trap does have some negatives, including active sites, worries about foaming samples, and trap degradation. Static headspace sampling, on the other hand, is much simpler than purge and trap sampling. For this sampling technique, the sample is brought to equilibrium and a portion of the headspace is transferred to the GC/MS for separation and analysis. The simplicity of this technique is a definite pro. Moreover, this sampling process does not develop active sites, has no need for an analytical trap and the linear calibration range can be much higher than that of purge and trap sampling. Conversely, samples need to be manually prepped thus, losing their sample integrity. Additionally, the detection limits are higher for a number of compounds. Finally, some of the analytes do not partition into the headspace well enough and need method optimization. In recent years, GC/MS systems have become much more sensitive. The advent of SIM/Scan acquisition techniques has made low level detection a much simpler proposition. This analysis will focus on the headspace sampling and analysis of over 50 volatile organic compounds. Experimental: The sampling system used for this analysis was the EST Analytical FLEX autosampler fitted with a 2.5ml headspace syringe. An Agilent 7890 GC and 5975 MS were used for separation and analysis. The GC was configured with a Restek Rxi 624 Sil MS 30m x 250mm x 1.4µm column and a SKY 2mm x 6.5 x 78.5 splitless inlet liner. The MS was run in SIM/Scan mode. Refer to Tables 1 and 2 for the analysis and sampling parameters.

GC/MS Inlet Inlet Temp. Inlet Head Pressure Split Liner Column Oven Temp. Program Column Flow Rate Gas Total Flow Source Temp. Quad Temp. MS Transfer Line Temp. Solvent Delay Acquisition Mode Scan Range SIM Ions: 50, 52, 62, 64, 66, 85, 87, 94, 96 SIM Ions: 61, 63, 96, 101, 103, 153 SIM Ions: 49, 61, 84, 86, 96 SIM Ions: 63, 64 SIM Ions: 52, 61, 62, 75, 77, 78, 83, 85, 96, 97, 98, 110, 117, 119, 128, 130. 168 SIM Ions: 41, 63, 76, 83, 85, 88, 93, 95, 112, 114, 130, 174 SIM Ions: 75, 77 SIM Ions: 91, 92 SIM Ions: 76, 78, 83, 85, 97, 107, 109, 127, 129, 164 SIM Ions: 52, 82, 91, 106, 112, 114, 117, 131, 133 SIM Ions: 78, 91, 104, 106, 173, 175 SIM Ions: 77, 83, 85, 91, 105, 120, 126, 156 SIM Ions: 105, 111, 119, 120, 134, 146, 150, 152 SIM Ions: 91, 111, 134, 146 SIM Ions: 75, 155 SIM Ions: 102, 128, 180, 182, 190, 225 Agilent 7890/5975 Split/Splitless 200ºC 12.153 psi 5:1 Restek SKY Liner Splitless, 2mm x 6.5 x78.5 Rxi-624 Sil MS 30m x 0.25mm I.D. 1.4µm film thickness 45ºC hold for 2.0 min, ramp 15ºC/min to 220ºC hold for 1.33 min, 15 min run time 1.0ml/min. Helium 9ml/min. 230ºC 150ºC 180ºC 0.7 min SIM/Scan m/z 35-265 0.70 to 2.12 min 2.13 to 2.62 min 2.63 to 3.25 min 3.26 to 3.69 min 3.70 to 4.84 min 4.85 to 5.86 min 5.87 to 6.19 min 6.20 to 6.55 min 6.56 to 7.45 min 7.46 to 8.09 min 8.10 to 8.72 min 8.73 to 9.40 min 9.41 to 10.17 min 10.18 to 10.80 min 10.81 to 11.61 min 11.62 to 15 min Table 1: GC/MS Experimental Parameters

Autosampler Method Type Incubation Temp. Incubation Time Agitation Speed Agitation Delay Agitation Duration Syringe Temperature Syringe Needle Depth Sample Depth Speed Sample Volume Sample Fill Rate Sample Fill Delay Needle Depth Speed Needle Depth Injection Rate Injection Volume Pre-Injection Delay Post-Injection Delay Injection Start Input General Sample Incubate Agitate Sample Fill Injection FLEX Headspace 60ºC 20min 80% 0.0min 19min 60ºC 90% 20% 1000µl 10% Off 30% 90% 40% 1000µl Off Off Start Sweep Needle Needle Temperature 150ºC Syringe Pumps 5 Syringe Pump Volume 1200µl Syringe Pump Speed 20% Table 2: Flex Autosampler Experimental Parameters The 8260 standards were acquired from Restek. Next, several midpoint standards were made in order to determine the optimum experimental conditions. Ultimately, it was found that ten milliliters of standard added to two grams of sodium chloride provided the optimum analyte response. The most effective sampling and analysis conditions are listed in the previous two tables. After the experimental conditions were established, a linear curve was run from 0.5 to 200ppb. Then, seven replicate low level standards were run in order to determine method detection limits. Furthermore, a second set of replicate samples were run at the mid-level of the curve in order to ascertain the precision and accuracy of the sampling and analysis. Experimental results are listed in Table 3 and SIM and Scan chromatograms of the curve midpoint can be found in Figures 1 and 2.

Compound Curve Ave. %RSD %Recovery MDL %RSD Curve RF 50ppb 50ppb Dichlorodifluoromethane 7.70 0.397.32 0.81 101.56 Chloromethane 11.17 0.356 0.39 10.03 109.17 Vinyl Chloride 11.57 0.423 0.26 8.45 114.23 Bromomethane 13.94 0.149 0.33 5.33 95.62 Chloroethane 10.15 0.269 0.23 8.41 102.60 Trichlorofluoromethane 9.26 0.423 0.28 9.94 97.13 1,1-Dichloroethene 6.44 0.356 0.17 8.49 101.55 Methylene Chloride 8.22 0.216 0.22 5.76 100.97 cis-1,2-dichloroethene 8.29 0.314 0.13 5.42 103.09 1,1-Dichloroethane 10.22 0.667 0.21 7.45 105.25 trans-1,2-dichloroethene 9.50 0.373 0.19 5.93 99.73 2,2-Dichloropropane 11.14 0.414 0.23 5.42 96.03 Bromochloromethane 6.32 0.103 0.16 3.34 98.29 Chloroform 7.36 0.537 0.17 6.14 102.71 1,1,1-Trichloroethane 8.63 0.650 0.25 7.59 99.53 Carbon Tetrachloride 7.61 0.544 0.26 8.73 100.03 1,1-Dichloropropene 9.15 0.618 0.28 7.80 101.56 Benzene 8.10 1.486 0.19 5.97 104.32 1,2-Dichloroethane 10.17 0.253 0.18 8.52 98.46 Trichloroethene 6.02 0.369 0.20 6.57 99.12 1,2-Dichloropropane 7.21 0.238 0.10 5.24 100.52 Dibromomethane 6.69 0.057 0.27 4.43 102.68 Bromodichloromethane 5.66 0.254 0.13 3.43 100.66 cis-1,3-dichloropropene 6.08 0.273 0.15 3.92 100.12 Toluene 5.11 0.712 0.12 5.16 104.50 1,1,2-Trichloroethane 5.69 0.118 0.12 2.40 96.71 Tetrachloroethene 6.11 0.482 0.11 7.50 100.22 1,3-Dichloropropane 5.61 0.215 0.09 3.58 102.15 Dibromochloromethane 5.25 0.125 0.06 2.49 98.30 1,2-Dibromoethane 7.66 0.091 0.23 1.61 96.97 Chlorobenzene 4.33 0.891 0.09 4.26 100.53 1,1,1,2-Tetrachloroethane 5.80 0.290 0.10 4.60 99.10 Ethylbenzene 6.79 1.911 0.15 5.52 105.21 Xylene (m+p) 8.92 1.421 0.10 4.90 109.24 Styrene 11.04 0.885 0.10 3.47 110.98 Xylene (o) 8.65 1.341 0.08 4.22 108.33 Bromoform 11.09 0.089 0.34 3.80 93.60 Bromobenzene 7.58 1.176 0.10 5.34 99.46 1,1,2,2-Tetrachloroethane 6.71 0.326 0.10 5.10 94.44 n-propylbenzene 8.91 4.871 0.19 5.79 105.37 2-Chlorotoluene 5.14 0.800 0.13 5.33 102.81 4-Chlorotoluene 5.11 0.808 0.15 3.69 103.71 1,3,5-Trimethylbenzene 10.07 3.039 0.16 5.12 106.84 tert-butylbenzene 8.16 2.741 0.13 8.02 103.68 sec-butylbenzene 8.99 0.832 0.16 9.16 104.84 1,2,4-Trimethylbenzene 9.87 2.907 0.21 3.94 107.61 1,3-Dichlorobenzene 5.29 1.380 0.14 3.46 99.16 1,4-Dichlorobenzene 6.09 1.330 0.09 2.52 98.15 1,2,-Dichlorobenzene 5.00 1.121 0.08 3.21 98.93 n-butylbenzene 8.74 3.237 0.19 5.63 106.36 1,2-Dibromo-3- chloropropane 10.91.063 0.40 6.30 94.22 1,2,4-Trichlorobenzene 7.38 0.606 0.21 2.86 99.11 Naphthalene 6.75 0.955 0.13 2.94 95.02 Hexachlorobutadiene 12.36 0.543 0.25 10.71 97.42 1,2,3-Trichlorobenzene 12.38 0.486 0.09 3.56 97.42 Average 8.07 0.828 0.18 5.62 101.37 Table 3: Experimental Results Summary

Figure 1: 50ppb Chromatogram in SIM Figure 2: 50ppb Chromatogram in Scan

Conclusions: The FLEX Autosampler in conjunction with the Agilent GC/MS proved to be an outstanding system for USEPA Method 8260 VOC analysis by static headspace sampling. The system was able to accurately determine VOCs down to 0.5ppb using the SIM/Scan acquisition mode. The curve %RSD results showed the linearity of the curve to meet the USEPA Method 8260 requirement of 15% or better. The method detection limits of all the compounds tested also passed method requirements. Lastly, the precision and accuracy of the FLEX Autosampler was excellent, with the average precision at less than 6% RSD and the average %recovery at just over 101%. References: 1. Volatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS); United States Environmental Protection Agency Method 8260B, Revision 2, December 1996. Headquarters JSB International Tramstraat 15 5611 CM Eindhoven T +31 (0) 40 251 47 53 F +31 (0) 40 251 47 58 With courtesy of Zoex Europe Tramstraat 15 5611 CM Eindhoven T +31 (0) 40 257 39 72 F +31 (0) 40 251 47 58 Sales and Service Netherlands Apolloweg 2B 8239 DA Lelystad T +31 (0) 320 87 00 18 F +31 (0) 320 87 00 19 Belgium Grensstraat 7 Box 3 1831 Diegem T +32 (0) 2 721 92 11 F +32 (0) 2 720 76 22 Germany Max-Planck-Strasse 4 D-47475 Kamp-Lintfort T +49 (0) 28 42 9280 799 F +49 (0) 28 42 9732 638 UK & Ireland Cedar Court, Grove Park Business Est. White Waltham, Maidenhead Berks, SL6 3LW T +44 (0) 16 288 220 48 F +44 (0) 70 394 006 78 info@go-jsb.com www.go-jsb.com