Application Note GCMS-03 Reliable high-throughput quantification of melamine and related analogs below regulatory limits in food using the EVOQ GC-TQ

Similar documents
Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Document: LABERCA/08MEL-Al.7 7 th revision of 20/10/08. Research and methodological development: B. Veyrand, S. Elaudais, Ph.

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Determination and quantification by GC-MS of melamine (1 mg/kg), ammeline and cyanuric acid (20 mg/kg)

Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

TargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

GC-MS Screen for the Presence of Melamine, Ammeline, Ammelide and Cyanuric Acid

ENVIRONMENTAL analysis

Food Safety. Abstract. Authors

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEY WORDS. Simultaneous high-throughput screening of melamine (MEL) and cyanuric acid (CYA).

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Multi-residue analysis of pesticides by GC-HRMS

Technical Report. Abstract: Pu Wang 1, Huizhong Sun 1, Qinghua Zhang 1, Feifei Tian 2, Lei Cao 2

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Estimation of LOQ for the Analysis of Persistent Organic Pollutants, in particular PCDD/Fs and PCBs

WADA Technical Document TD2003IDCR

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Liquid Chromatography Mass Spectrometry. Shimadzu Corporation, UK; 2 Phytocontrol, France; 3 Shimadzu France, France. Trimesium S + CH 3 ETU

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

At-a-Glance. Verapamil C 27 H 38 N 2 O 4 M+H + =

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Characterization of petroleum samples via thermal analysis coupled to APCI FTMS

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Validation of GC-MS/MS for Detection and Confirmation of Low-Level Dioxins

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Determination of 59 potential Allergens in Perfumes by twin-line fast GCMSMS

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Quantitative Determination of Dioxins in Drinking Water by Isotope Dilution using Triple Quadrupole GC-MS/MS

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

MRMS axelerate rapidly detected micropollutants and plant response metabolites in poplar leaves

GC/MS/MS Dioxin and Dioxin-like PCBs in Feed and Food. Jessica Westland, LSAG Applications Chemist

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Benzoylecgonine in Urine by SAMHSA GC/MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Confirmation and Quantitation of THC in Oral Fluid Using an MRM Method on a GC-Triple Quadrupole MS

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry

Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method)

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Analysis and Quantification of Mycotoxins in Cereals Using MRM HR on the SCIEX X500R QTOF System with SCIEX OS

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Highly Sensitive and Rugged GC/MS/MS Tool

COMPLEX HEROIN COMPARISON INVESTIGATION

Application Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

U.S. EPA Method 8270 for multicomponent analyte determination

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Proven robustness for large-scale metabolomics studies using the Phenomics Workhorse

CALIBRATION OF GC/MS METHOD AND VALIDATION OF THE MODIFIED SAMPLE PREPARATION FOR DETERMINATION OF POLYCYCLIC AROMATIC HYDROCARBONS

Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Choosing Fit-For-Purpose Food Safety Methods

ACETONE IN URINE BY GC/MS Headspace Code GC10110

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Automated and accurate component detection using reference mass spectra

Transcription:

Application Note GCMS03 Reliable highthroughput quantification of melamine and related analogs below regulatory limits in food using the EVOQ GCTQ Introduction is a molecule that presents a high content of nitrogen, 67% by mass. It has many industrial applications in the form of melamine or melamine resins. For example, it can be found in kitchenware, whiteboards, laminate flooring, colorant or fire retardant additives. In recent years, melamine has been fraudulently added to food and feed in order to increase their nitrogen ratios. The main reason for these fraudulent activities is to provide the food item with apparent higher protein content since many protein tests are only based on nitrogen content detection. In 2007, after the death of a large number of animals, melamine was found in wheat gluten destined to petfood manufacturing. Later in 2008, a food scandal occurred in China involving milk and infant formula which was adulterated with melamine, leading to kidney stones and other renal failure among young children. According to a WHO report from 2009 and a report from the Chinese Ministry of Health, 294,000 infants had been affected by melaminecontaminated infant formulae by the end of November 2008. More than 50,000 infants were hospitalized and six deaths have been confirmed. Authors Marc Bourgeois, Bruno Veyrand, Philippe Marchand, Fabrice Monteau, Bruno Le Bizec. LUNAM Université, ONIRIS, Ecole Nationale Vétérinaire, Agroalimentaire et de l Alimentation NantesAtlantique, Laboratoire d Etude des Résidus et Contaminants dans les Aliments (LABERCA), AtlanpôleLa Chantrerie, BP 50707, Nantes F44307, France. Alexandre Collgros, application engineer. Bruker Daltonique France, Chemical & Applied Market Keywords Analogs Instrumentation and Software EVOQ GCTQ

Chemical structure of melamine and its analogs Cyanuric acid Ammelide sonication and centrifugation for about 10 min. at 2000 g. The obtained liquid phase was filtered through a 0.45 µm nylon filter, and 50 µl of filtrate was then spiked with benzoguanamine (external standard). This solution was evaporated, dissolved in 50 µl of MSTFA and then submitted to derivatization by heating 60 C for 45 min. Ammeline GCMS/MS Conditions The samples and calibration solutions were injected with the following GC and MS/MS parameters. For ionization, filament current was set to 80 µa. Argon was used as the collision gas at a pressure of 2.0 mtorr. Chromatographic conditions Benzoguanamine (External standard) Column Gas ZB5 MS 30 m x 0.25 mm x 0.25 µm Helium, 1.0 ml/min constant flow H 2 N N Injector 280 C, Splitless N N Injection volume 2 µl NH 2 Taking into account the available occurrence data, the level of 2.5 mg/kg in foodstuff was chosen as an appropriate level to distinguish between the unavoidable background presence of melamine and unacceptable adulteration. For infant formula, this level was lowered to 1 mg/kg, as described in Commission Regulation (EU) No 594/2012, amending Regulation 1881/2006/EC. These levels take into account the need to ensure a large margin of safety. The development of a specific and sensitive method based on Selected Reaction Monitoring (SRM) with a GCMS/MS system was conducted to ensure unambiguous identification of this compound in foodstuff. With the goal of having a fast and highthroughput method, a rapid sample preparation was used without any SPE cleanup. Since it was proven that the toxicity of melamine increased in the presence of cyanuric acid, it was decided to investigate broader contamination profiles, i.e., including not only melamine but also its analogs using the EVOQ GCTQ. Structures of these compounds are illustrated above. The calibration standards ranged from 0.5 to 10 mg/kg of melamine and analogs. All analyzed food samples (unknown, blank and supplemented) were submitted to solid / liquid extraction, by adding 4 ml of a H 2 O / Acetonitrile mixture (50/50, v/v), and agitated for 30 sec., followed by 30 min. Oven temperature program Rate ( C/min) Instrument Temperature( C) 120 1 10 320 2 Gas chromatography (436 GC, Bruker) coupled to a triple quadrupole mass spectrometer (EVOQ GCTQ, Bruker). Material and reagents Hold (min), ammeline, ammelide, cyanuric acid and benzoguanamine were purchased from BCP Instruments (Irigny, France), C 3 labeled melamine and cyanuric acid from Cluzeau (SainteFoyLaGrande, France). Mixed solutions containing melamine and its three analogs were made in H 2 O / diethylamine (80/20, v/v). Benzoguanamine solution was prepared in pyridine, and C 3 labeled melamine and cyanuric acid solution were prepared in ultrapure water. Diethylamine, acetonitrile, MSTFA and pyridine were purchased from SigmaAldrich (SaintQuentin Fallavier, France).

Sample preparation 100 mg of sample (ground into a fine powder) was used for the analysis. The internal standard mixture was added to the sample at a concentration of 2 mg/kg. A blank food sample (containing no traces of melamine), and a range of various supplemented food samples used for calibration were prepared using the same protocol. MS Conditions Ionisation mode Electron Impact, 70eV Source temperature 250 C Transfer Line Temperature 310 C Chromatogram: blank (top) and spiked (bottom) sample of powdered milk Blank Spiked at 2.0 mg/kg Internal Standard Blank : Acquisition mode Quadrupoles Resolutions SRM Standard on Q1 (i.e. 2.0 amu width) Unit on Q3 (i.e. 0.7 amu width) Blank : The various transitions and collision energies are summarized in the table below. SRM Transitions monitored for each target analyte Compound Transition 1 (CE, V) 351.1>147.0 Transition 2 (CE, V) Spiked at 2.0 mg/kg : Internal Standard 345.1>147.0 345.1>188.0 5 V Ammelide Ammeline 344.1>171.0 V 343.1>171.0 V 344.1>189.0 343.1>189.0 V Spiked at 0.5 mg/kg : 348.1>173.0 20 V Benzoguanamine 342.1>171.0 20 V 331.1>171.0 20 V 342.1>327.1 331.1>316.1 Spiked at 0.5 mg/kg : For acquisition, windows of 3 minutes width were centered at the retention time for each compound, and scan times were set by the software, between 60 and 300 ms, adjusted for peak width (in sec), desired number of data points per peak, and number of transitions simultaneously monitored. Figure 1: Blank (top) and spiked (bottom) sample of powdered milk. Blank was spiked only with melamine, whereas spiked sample contained also melamine at 0.5 mg/kg.

In every sequence, five calibration solutions were injected plus a blank solution containing only external and internal standards on matrix. Quality control sample were also injected to ensure the best level of performance of the system. Samples were then injected for detection and quantification. Calibration curve for Prior to SRM analysis, full scan and productscan analysis were performed on MSTFAderivatized molecules to optimize SRM parameters. Results According to the analytical criteria described in Commission Decision 2002/657/EC, two diagnostic SRM transitions were recorded for each analyte, providing unambiguous identification of compounds. In order to evaluate the instrument s sensitivity for melamine without SPE cleanup step during sample preparation, a blank and a spiked sample were analyzed. Both samples were infant formula (powdered milk) and the spiked sample contained 0.5 mg/kg melamine, as illustrated in Figure 1. The spiked sample at 0.5 mg/kg represents half of the maximum residue limit in infant formulae, and the melamine was easily identified. Molecule R² 0.999492 Ammelide 0.999960 Ammeline 0.999958 0.999804 Figure 2: linearity obtained for each target analyte in spiked infant formula samples. Figure 2 illustrates the linearity achieved for all compounds (coefficient of determinations greater than 0.999). Isotope labeled cyanuric acid was used as the internal standard for ammelide, and labeled melamine was used as the internal standard for ammeline. Figure 3 illustrates the calibration curve for melamine, with r² equal to 0.999804. Linearities of melamine and analogs between 0.5 and 10 mg/kg were achieved on the same infant powdered formula, using isotope dilution. Calibration curves for melamine and its analogs between 0.5 and 10 mg/kg were achieved on the same infant powdered formula using isotope dilution. The limits of quantification are 10 to 20 times lower than the regulation (Maximum Residue Limit of 1.00 mg/kg for infant formula). Cyanuric acid had less sensitivity than the other analytes. Despite the fact that the quantifier for this acid was easily observed at 0.5 mg/kg, the lack of specificity of the transitions lead to a limit of quantification of only 1.0 mg/kg. Figure 3: Example of calibration curve obtained for melamine Chromatogram: spiked powdered milk at 0.5 mg/kg Ammelide Ammeline (spiked at 2.0 mg/kg) (spiked at 2.0 mg/kg) Figure 4: typical diagnostic ion chromatograms (quantifier SRM transition) obtained for a spiked powdered milk sample. melamine and cyanuric acid (internal standards) were spiked at 2.0 mg/kg and native melamine and analogs at 0.5 mg/kg

These limits of quantification represent 10 to 20 times lower than the regulation (Maximum Residue Limit of 1.00 mg/kg for infant formula). Cyanuric acid presented less sensitivity. Despite the fact that the quantifier for this acid was well observed at 0.5 mg/kg, the lack of specificity of the transitions lead to a limit of quantification of only 1.00 mg/kg. Since it was proven that the toxicity of melamine is increased in the presence of cyanuric acid, structural analogs were also analyzed. They were included at the same concentration as melamine, 0.5 mg/kg, and cyanuric acid had his own isotope labeled internal standard, C 3 cyanuric acid. Figure 4 shows the response of all compounds (quantifiers) from a spiked milk sample. Limits of quantification were determined to be between 0.050 and 0.100 mg/kg for melamine, ammelide and ammeline. Conclusion The sensitivity of the EVOQ GCTQ system combined with the specificity of SRM analysis, allowed the efficient detection and quantification of melamine and its analogs even without SPE cleanup, which provides a high throughput workflow. could be quantified approximately 20 times lower than the Maximum Residue Limit for infant formulae. Acknowledgements The authors would like to thank Bruker Daltonique France (service, sales and application teams) for giving the opportunity to evaluate the performances of this GCMS/MS in the field of food safety analysis. References Method used in LABERCA, Nantes, France. http://www.laberca.org/newsletter/en/laberca08melal7. pdf Toxicological and Health Aspects of and Cyanuric Acid. Report of a WHO Expert Meeting in collaboration with FAO. Health Canada, Ottawa, Canada 1 4 December 2008. http://whqlibdoc.who.int/publications/2009/97892497951_ eng.pdf EU COMMISSION DECISION 2002/657/EC N L221/8 of 12 August 2002, Implementing Council Directive 96/23/EC concerning the performance of analytical methods and the interpretation of results. EU COMMISSION DECISION N L273/18 of 14 October 2008 imposing special conditions governing the import of products containing milk or milk products originating in or consigned from China, and repealing Commission Decision 2008/757/EC. EU COMMISSION REGULATION 594/2012 N 176/43 of 5 July 2012 amending Regulation (EC) 1881/2006 as regards the maximum levels of the contaminants ochratoxin A, non dioxinlike PCBs and melamine in foodstuffs.

Bruker Daltonics is continually improving its products and reserves the right to change specifications without notice. Bruker Daltonics 0420, GCMS03, 1837128 For research use only. Not for use in diagnostic procedures. Bruker Daltonik GmbH Bruker Daltonics Inc. Bremen Germany Phone +49 (0)42122050 Fax +49 (0)4212205103 Billerica, MA USA Phone +1 (978) 6633660 Fax +1 (978) 6675993 ms.sales.bdal@bruker.com www.bruker.com