NITROGEN, TOTAL, HR, Test N Tube

Similar documents
NITROGEN, TOTAL, Test N Tube * Method 10071

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

NITROGEN, AMMONIA, High Range, Test N Tube

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Mercaptoacetic Acid Method*

Method to 0.50 mg/l NH 3 N Powder Pillows

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Method to 0.50 mg/l NH 3 N Powder Pillows

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

DR/4000 PROCEDURE. IRON, Total

Nitrogen, Total Inorganic

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

NITRITE, Low Range (0 to mg/l NO 2- -N)

DPD Test N Tube Method *

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

SILICA, Low Range (0 to 1.60 mg/l)

Chloramine (Mono) and Nitrogen, Free Ammonia

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Oxygen Demand, Chemical

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

CHLORINE, TOTAL (0 to 4.00 mg/l)

Oxygen Demand, Chemical

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Chlorine, Free and Total, High Range

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Chlorine, Free and Total, High Range

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

p-dimethylaminobenzaldehyde Method*

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Quaternary Ammonium Compounds

Chlorine, Free and Total, Low Range

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

DR/4000 PROCEDURE. HARDNESS, Total

TPH (Total Petroleum Hydrocarbons)

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Manganese III Digestion Method * (with chloride removal)

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

MERCURY (0.1 to 2.5 µg/l) Method 10065

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Sample Preparation LCW 909. Total Kjeldahl Nitrogen, TKN NB: NEW! LCW 909 GB

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Trihalomethane Formation Potential (THMFP)

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Ammonia Gas Sensing Combination Electrode

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

Ammonia Gas Sensing Combination Electrode

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Polychlorinated Biphenyls (PCB) in Soil

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Nitrogen total LR VARIO TT mg/l N b) Persulphate Digestion

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

RAPID LIQUID TESTING. Hach Company, 1996 All rights reserved. Printed in the U.S.A dld/dp ed

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University

RAPID KJELDAHL BENCHNOTES

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

SILICA, Low Range SiLR

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Cyanide, colorimetric, pyridine-pyrazolone

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

The Synthesis and Analysis of Aspirin

Determination of an Equilibrium Constant

The Fluorometric Determination of Acetylsalicylic Acid in an Aspirin Tablet

SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation)

Determining the Concentration of a Solution: Beer s Law

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Transcription:

Method 10072 NITROGEN, TOTAL, HR, Test N Tube 10 to 150 mg/l N TNT Persulfate Digestion Method For water and wastewater Digestion is required for determining total nitrogen. The digestion procedure is included in the method. 1. Turn on the COD Reactor. Heat to 103 106 C (best temperature is 105 C). Place the plastic shield in front of the reactor. Note: Ensure safety devices are in place to protect the analyst should splattering and leakage occur. Note: For proof of accuracy, run a 125 mg/l NH 3 -N standard through digestion and analysis. 2. Prepare a reagent blank: Using a funnel, add the contents of one Total Nitrogen Persulfate Reagent Powder Pillow to one HR Total Nitrogen Hydroxide Digestion Vial. Note: Wipe off any reagent that gets on the lid or the tube threads. 3. Add 0.5 ml of organic-free water. Cap and shake vigorously for about 30 seconds. Process this reagent blank exactly the same as the sample, including digestion and color finish. Proceed to step 6. Note: Alternate water must be free of all nitrogen-containing species. Note: The persulfate reagent may not dissolve completely after shaking. Note: One reagent blank is sufficient for each set of samples using the same lots of reagents. Note: The reagent blank is stable for as long as seven days when stored in the dark; see Blanks for Colorimetric Measurement following this procedure. 4. Prepare the sample: Using a funnel, add the contents of one Total Nitrogen Persulfate Reagent Powder Pillow to a HR Total Nitrogen Hydroxide Digestion Vial. Note: Wipe off any reagent that gets on the lid or the tube threads.

5. Add 0.5 ml of sample to the vial. Cap and shake vigorously for about 30 seconds. Note: The persulfate reagent may not dissolve completely after shaking. 6. Place the vials in the COD Reactor. Heat for 30 minutes. 7. Using finger cots or gloves, remove the hot vials from the reactor and allow to cool to room temperature. Note: It is very important to remove the vials from the COD Reactor after exactly 30 minutes. 8. Enter the stored program number for Test N Tube HR Total Nitrogen. Press: 3 9 5 READ/ENTER The display will show: Dial nm to 410 Note: DR/2000s with software version 3.0 and greater will not display Dial nm to message if the wavelength is already set correctly. The display will show the message in step 9. Proceed with step 10. Note: For DR/2000s without this stored program, see Instrument Setup following these steps. 2

410 nm 9. Rotate the wavelength dial until the display shows: 410 nm Note: For DR/3000s, set the wavelength to 410 nm and press CLEAR. 10. Press: READ/ENTER The display will show: mg/l N HR TN TNT 11. Add the contents of one Total Nitrogen Reagent A Powder Pillow to the vial containing the digested blank or sample. Cap vial and shake for 15 seconds. Press: SHIFT TIMER after shaking. A three-minute reaction period will begin. Note: For DR/3000s, press 3 TIMER. 12. After the timer beeps, add one Total Nitrogen Reagent B Powder Pillow to the vial. Cap the vial and shake for 15 seconds. Press: SHIFT TIMER after shaking. A two-minute reaction period will begin. Note: The reagent will not completely dissolve. The solution will begin to turn yellow. Note: For DR/3000s, press 2TIMER. 13. After the timer beeps, remove the cap from a Total Nitrogen Reagent C vial. Add 2 ml of digested, treated sample (or reagent blank) to the vial. 14. Cap and invert slowly 10 times to mix. The vial will be warm. Note: Proper mixing is important for complete recovery. Hold the vial vertical with the cap up. Invert the vial and wait for all of the solution to flow to the cap end. Pause. Return the vial to the upright position and wait for all of the solution to flow to the vial bottom. This is one inversion (10 inversions = 30 seconds) 15. Press: SHIFT TIMER A five-minute reaction period will begin. Do not invert the vial again. Note: The yellow color will intensify. Note: For DR/3000s, press 5TIMER. 16. Insert the COD Vial Adapter into the cell holder with the marker to the right. Note: For DR/3000s, the groove in the adapter faces the front of the instrument. 3

17. When the timer beeps, wipe the outside of the Total Nitrogen Reagent C vial containing the reagent blank. Place the vial in the adapter with the Hach logo facing the front of the instrument. Place the cover on the adapter. Note: Wipe with a damp towel, followed by a dry one, to remove fingerprints or other marks. 18. Press: ZERO The display will show: then: WAIT 0 mg/l N HR TN TNT Note: For DR/3000s, press MANUAL PROGRAM. Then press ZERO. Press ZERO again if the display does not show 0.000. Enter the concentration factor (200.29) and press CONC FACTOR. Then press 0 CONC. 19. Wipe the Total Nitrogen Reagent C vial containing the sample. Place the vial into the cell holder with the Hach logo facing the front of the instrument. Place the cover on the adapter. Note: Multiple samples may be read after zeroing on one reagent blank. 20. Press: READ/ENTER The display will show: WAIT then the result in mg/l total nitrogen will be displayed. Note: If the test overranges, repeat the digestion and colorimetric measurement with diluted sample. The entire digestion must be repeated for accurate results. Note: For DR/3000s, pressing READ is not necessary. The instrument will automatically display the results. Sampling and Storage Collect samples in clean plastic or glass bottles. Best results are obtained with immediate analysis. Preserve the sample by reducing the ph to 2 or less with concentrated sulfuric acid (at least 2 ml per liter). Store at 4 C (39 F) or less. Preserved samples may be stored up to 28 days. Warm samples to room temperature and neutralize with 5 N sodium hydroxide before analysis. Correct the test result for volume additions; see Correcting for Volume Additions in Section I of the Procedures Manual. Accuracy Check This method generally yields 95-100% recovery on organic nitrogen standards. For proof of accuracy Hach offers a set of three Primary Standards for Kjeldahl Nitrogen. 1. Prepare one or more of the following three solutions. Each preparation is for an equivalent 120 mg/l N standard. Use water that is free of all organic and nitrogen-containing species. a. Weigh 1.6208 g of Ammonium p-toluenesulfonate (PTSA). Dissolve in a 1000-mL volumetric flask with deionized water. Add deionized water to the 1000-mL mark. 4

b. Weigh 2.1179 g of Glycine p-toluenesulfonate. Dissolve in a 1000-mL volumetric flask with deionized water. Add deionized water to the 1000-mL mark. c. Weigh 2.5295 g of Nicotinic p-toluenesulfonate. Dissolve in a 1000-mL volumetric flask with deionized water. Add deionized water to the 1000-mL mark. 2. Analyze each of these solutions using the test procedure above. Calculate the percent recovery for each using this formula: The percent recovery should be: Hach analysts have found Ammonia-PTSA to be the most difficult to digest. Other compounds may yield different percent recoveries. Standard Solution Method For proof of accuracy, substitute 0.5 ml of a 125-mg/L ammonia nitrogen standard solution for the sample in the procedure. To prepare a 125-mg/L ammonia nitrogen standard, use a 25-mL Class A pipet to add 25.00 ml of a 1000-mg/L Ammonia Nitrogen Standard (see OPTIONAL REAGENTS on page 14) to a 200-mL Class A volumetric flask. Dilute to the line with organic-free water. Standard Additions Method Compound measured concentration % recovery = --------------------------------------------------------------- 100 120 a. Fill three 25-mL graduated mixing cylinders with 25 ml of sample. b. Snap the neck off an Ammonia Nitrogen Voluette Ampule Standard Solution, 1000 mg/l as NH 3 -N. c. Use the TenSette Pipet to add 0.1, 0.2 and 0.3 ml of standard, respectively, to the three mixing cylinders. d. Stopper each cylinder and mix thoroughly. Lowest Expected % Recovery Ammonia-PTSA 95% Glycine-PTSA 95% Nicotinic-PTSA 95% e. Add 0.5 ml of each prepared solution, respectively, to three HR Total Nitrogen Hydroxide Digestion Vials. f. Analyze each standard addition sample as described in the procedure. The nitrogen concentration should increase by about 4 mg/l N for each 0.1 ml of standard added. g. If these increases do not occur, see Standard Additions in Section I of the Procedures Manual for troubleshooting information. 5

Blanks for Colorimetric Measurement The reagent blank may be used repeatedly for measurements using the same lots of reagents. Store the reagent blank in the dark at room temperature (18 25 C) for a maximum of seven days. If a small amount of white floc appears prior to the end of one week, discard the reagent blank and prepare a new one. Method Performance DR/2000: In a single laboratory, using a standard solution of 125 mg/l NH 3 -N and two lots of reagent with a DR/2000 Spectrophotometer, a single operator obtained a standard deviation of less than 3 mg/l N. DR/3000: In a single laboratory, using a standard solution of 125 mg/l NH 3 -N and two lots of reagent with a DR/3000 Spectrophotometer, a single operator obtained a standard deviation of less than 3 mg/l N. Interferences The substances in the following table have been tested and found not to interfere up to the indicated levels (in mg/l): Substance Maximum Level Tested (mg/l) Barium 10.4 Calcium 1200 Chromium (3+) 2 Iron 8 Lead 26.4 ppb Magnesium 2000 Organic Carbon 600 ph 13 ph units Phosphorus 400 Silica 600 Silver 3.6 Tin 6 Interfering substances that resulted in a concentration change of ±10%: Substance Bromide Chloride Level and Effect >240 ppm; positive interference 3000 ppm; positive interference The large amounts of nitrogen-free organic compounds in some samples may decrease digestion efficiency by consuming some of the persulfate reagent. Samples known to contain high levels of organics should be diluted and re-run to verify digestion efficiency. Summary of Method An alkaline persulfate digestion converts all forms of nitrogen to nitrate. Sodium metabisulfite is added after digestion to eliminate halogen oxide interferences. Nitrate then reacts with chromotropic acid under strongly acidic conditions to form a yellow complex with an absorbance maximum at 410 nm. 6

Instrument Setup DR/2000 with 1.261 and 1.27 software Enter the calibration as an operator-programmed calibration. Follow the steps in the Operation section of the DR/2000 Instrument Manual. Store the method as follows: nm = 410 Decimal = 0000. Units = mg/l Symbol = N HR TN TNT Timer 1 = 3:00 Timer 2 = 2:00 Timer 3 = 5:00 Enter the calibration with 0.000 absorbance values for the zero and #1 standard. To do this, do not place anything in the sample cell compartment. Begin by storing standards 0 and 1 as the concentrations shown in the table below. Accept 0.000 Abs as the absorbance value for all standards. Store the calibration values by pressing: SHIFT READ/ENTER Next, edit the absorbance values for the standards to the values given below. Follow the steps given in the Operations section of the DR/2000 Instrument Manual. Std. Conc. Abs. # 0 0 0.000 # 1 150 0.740 The method is now stored as an operator-programmed method number between 950 and 999. Record the method number for future reference when using this method. 7

DR/2000 with software versions 2.0 and 2.2 Enter the calibration as an update to Hach stored programs. 1. Press: 2. Press: 3. Press: 4. Within 3 seconds, press: The display will show: ENTER nm Note: If you make an error, press SHIFT CLEAR and re-enter the number. When the number is correct, press READ/ENTER. 5. Press: The display will show: DECIMAL? 00.00 6. Use the arrow keys to correctly position the decimal point. Press the down arrow key twice. The display will show: DECIMAL? 0000. 7. Press: READ/ENTER. The display will show: UNITS? 8. Use the arrow keys to select the appropriate unit of measure. Press the down arrow key once. The display will show: mg/l 9. Press READ/ENTER when the correct unit of measure is displayed. The display will show: SYMBOL? 10. Construct the correct symbol display: N HR TN TNT a. Select a letter by scrolling to the correct symbol with the arrow keys. b. To make a letter upper case, press the SHIFT key. c. To enter subscript numbers, enter the digit on the numeric keypad. d. To enter superscript numbers, enter the digit on the numeric keypad number then make it a superscript by pressing SHIFT. e. The space is the character displayed after one press of the down arrow. f. Accept each symbol by pressing READ/ENTER. 8

g. To end symbol entry, press READ/ENTER a second time after accepting the last character. 11. When the instrument is out of symbol entry mode, the display will show: TIMER? 12. This method has 3 timed steps, so press SHIFT TIMER. The display will show: MM:SS TIME 1? 13. Enter a timer value of 3 minutes. Press: 14. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 2? 15. Enter a timer value of 2 minutes. Press: 16. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 3? 17. Enter a timer value of 5 minutes. Press: 18. Press READ/ENTER to accept the timer entry. The display will show: MM:SS TIMER 4? 19. Press READ/ENTER to complete the timer entry. The display will show: #1 Data 20. Enter the following 12 numbers as shown. Complete each number entry by pressing the READ/ENTER key. Display Number Entry #1 Data 0 #2 Data 3340 #3 Data 3341 #4 Data 3085 #5 Data 3340 # 6 Data 3341 #7 Data 3085 #8 Data 65535 #9 Data 65535 #10 Data 21845 #11 Data 544 Checksum 23617 The final number is a check value which determines if the data sequence was correctly entered. If an error was made during number entry, the display will return to the prompt for #1 Data and the entire sequence must be re-entered. If all numbers are correctly entered, the display will return to the method prompt and is ready for use. 9

DR/2000 with software version 3.0 and above 1. Turn the instrument on. Press SHIFT METHOD to enter the configuration mode. The display will show: MOMENTARY or CONSTANT ON 2. Press the up arrow key twice to select HACH UPDATE. Press READ/ENTER. The display will show: ENTER #: 3. Press: The display will show: P395 ENTER nm Note: If you make an error, press SHIFT CLEAR and reenter the number. When the number is correct, press READ/ENTER. 4. Press: 5. Press READ/ENTER. The display will show: P395 Decimal? 00.00 6. Use the arrow keys to correctly position the decimal point. Press the DOWN ARROW key twice. The display will show: DECIMAL? 0000. 7. Press: READ/ENTER. The display will show: P395 UNITS? 8. Use the arrow keys to select the appropriate unit of measure. Press the down arrow key twice. The display will show: P395 mg/l 9. Press READ/ENTER when the correct unit of measure is displayed. This display will show: P395 mg/l 10. Construct the display to read the correct symbol. The symbol must be entered exactly as shown including dashes and spaces between characters: N HR TN TNT 10

To enter the display symbol, press the following number sequences, following each with an enter key to place the character then enter again to accept and move to position. a. To enter N press b. To enter press (enters a space character) c. To enter H press d. To enter R press e. To enter press (enters a space character) f. To enter T press g. To enter N press h. To enter press (enters a space character) i. To enter T press j. To enter N press k. To enter T press Alternately, you can scroll to each of the proper characters with the arrow keys by using the following procedures: a. Select letters and numbers by scrolling to the correct character with the arrow keys. b. To make a letter uppercase, press the SHIFT key. c. To make a number a subscript, press SHIFT until the symbol is correct. d. The space is the character displayed after one press of the down arrow key. e. Make sure to enter the display line exactly as shown, including the spaces. Do not enter trailing spaces. f. Accept each symbol by pressing READ/ENTER. g. When the last character of the symbol is accepted, display entry mode automatically concludes. 11. When the instrument is out of symbol entry mode, the display will read: P395 TIMER 11

12. This method has three timed steps, so press SHIFT TIMER. The display will show: MM:SS TIME 1? 13. Enter a timer value of 3 minutes. Press: 14. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 2? 15. Enter a timer value of 2 minutes. Press: 16. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 3? 17. Enter a timer value of 5 minutes. Press: 18. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 4? 19. Press READ/ENTER to complete the timer entry. The display will show: #0 STANDARD 20. Press READ/ENTER to display the zero data pair. The display will show: 0.000 Abs 0000. mg/l 21. Press READ/ENTER. The display will show: #1 STANDARD 22. Press READ/ENTER. The display will prompt for entry of the first concentration point: # 1 0000. mg/l 23. Enter concentration point #1 from the table below by pressing 0150 so that the display shows: # 1 0150. mg/l 24. Press READ/ENTER. The display will prompt for entry of the first absorbance point: # 1 0.000 Abs 25. Enter the absorbance point #1 by pressing 0740. The display will show: # 1 0.740 Abs 26. Press READ/ENTER. The display will show the first data pair: 0.740 Abs 0150. mg/l 12

27. Press READ/ENTER to accept the first data pair. The display will show: #2 STANDARD 28. The data pair values from the table below are now entered. 29. When the last point pair is entered the display will show: # 2 STANDARD 30. Press SHIFT READ/ENTER to complete data point entry. The display will show: #: Standard Concentration Absorbance 31. Enter the validation number: 3032 so that the display shows: #: 3032 #0 [ 0] mg/l [0.000] Abs #1 [ 150] mg/l [0.740] Abs 32. Press READ/ENTER. The display will show: COMPLETED then: P395 mg/l N HR TN TNT Note: If the display shows INCORRECT #, then prompts for the validation number, you may have made an error during data entry. Make sure that the validation number is correct. If so, then the error occurred during some other portion of the method entry. You must press METH and respond to the ABORT? message by pressing READ/ENTER, then re-enter the method. The instrument is now ready for use with method 395. REQUIRED REAGENTS Test N Tube HR Total Nitrogen Reagent Set (50 vials)...27141-00 Includes: (1) 26718-46, (1) 26719-46, (1) 26720-46, (50) Hydroxide digestion vials,* (50) Acid Solution vials* Quantity Required Description Per Test Unit Cat. No. HR Total Nitrogen Hydroxide Digestion Vials... 1 vial... 50/pkg... * Total Nitrogen Persulfate Reagent Powder Pillows... 1 pillow... 50/pkg...26718-46 Total Nitrogen Reagent A, Bisulfite Powder Pillows... 1 pillow... 50/pkg...26719-46 Total Nitrogen Reagent B, Indicator Powder Pillows... 1 pillow... 50/pkg...26720-46 Total Nitrogen Reagent C Vials, Acid Solution... 1 vial... 50/pkg... * * These Items are not sold separately. Please order the complete set (Cat. No. 27141-00) as a replacement. 13

REQUIRED APPARATUS Quantity Required Description Per Test Unit Cat. No. COD Reactor, 115/230V, North American Plug... 1...each... 45600-00 COD Reactor, 230V, European Plug... 1...each... 45600-02 COD Vial Adapter, DR2000/3000... 1...each... 44799-00 Funnel, Micro... 1...each... 25843-35 Pipet, TenSette, 0.1-1.0 ml... 1...each... 19700-01 Pipet Tips for 19700-01... 2...50/pkg... 21856-96 Safety Shield, for COD Reactor... 1...each... 23810-00 Test N Tube Cooling Rack... 1...each... 18641-00 OPTIONAL REAGENTS Nitrogen, Ammonia, 1000 mg/l NH 3 -N... 1 liter... 23541-53 Nitrogen, Ammonia, Voluette Ampule, 1000 mg/l NH 3 -N, 10 ml...16/pkg... 23541-10 Primary Standards for Kjeldahl Nitrogen... set of 3... 22778-00 Sodium Hydroxide Standard Solution, 5.0 N... 59 ml MDB... 2450-26 Sulfuric Acid, ACS... 500 ml... 979-49 Water, Organic-free... 500 ml... 26415-49 OPTIONAL APPARATUS Ampule Breaker Kit...each... 21968-00 Balance, analytical, 115 Vac...each... 26103-00 Balance, analytical, 230 Vac...each... 26103-02 Cots, finger...2/pkg... 14647-02 Cylinder, graduated mixing, 25 ml... 3...each... 20886-40 Flask, Volumetric, Class A, 200 ml...each... 14574-45 Flask, Volumetric, Class A, 1000 ml... 3...each... 14574-53 ph Indicator Paper, 1.0 to 11.0 ph units...5 rolls/pkg... 391-33 Pipet, Volumetric, Class A, 25 ml... 2...each... 14515-40 14

Hach Company, 1998. All rights reserved. Printed in the U.S.A. 8158 jb/dh 10/98 1ed