Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Similar documents
Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Fast methods for the determination of ibuprofen in drug products

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Basic Principles for Purification Using Supercritical Fluid Chromatography

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Application Note. Authors. Abstract. Pharmaceuticals

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Volume 6, Issue 2, January February 2011; Article-015

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

Modernizing the USP Monograph for Acetaminophen

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Multi-Channel SFC System for Fast Chiral Method Development and Optimization

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

Expediting Achiral SFC Method Development Using a Multi-Channel SFC System with MS Detection

Journal of Chemical and Pharmaceutical Research, 2013, 5(5): Research Article

Chapter 4: Verification of compendial methods

Routine MS Detection for USP Chromatographic Methods

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

LC Technical Information

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Analysis of PEGylated Proteins with Agilent AdvanceBio SEC Columns

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Int. J. Pharm. Sci. Rev. Res., 30(2), January February 2015; Article No. 09, Pages: 63-68

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Separation of Large and Small Peptides by Supercritical Fluid Chromatography and Detection by Mass Spectrometry

BRIEFING. (EXC: K. Moore.) RTS C Propylparaben C 10 H 12 O Benzoic acid, 4 hydroxy, propyl ester; Propyl p hydroxybenzoate [ ].

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Revision Bulletin 27 Jan Feb 2017 Non-Botanical Dietary Supplements Compliance

Reverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

Agilent LC & LC-MS Cannabis Testing

The new Agilent 1290 Infinity Quaternary LC

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

Validated HPLC Methods

Revision Bulletin 29 Dec Jan 2018 Non-Botanical Dietary Supplements Compliance

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

LC and LC/MS Column Selection Flow Chart

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Figure 1. Structures for Vitamin B2 and Vitamin B1.

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS.

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Application Note. Edgar Naegele. Abstract

Dilution(*) Chromatography

Successfully Scaling and Transferring HPLC and UPLC Methods

with diode array detection

Appendix II- Bioanalytical Method Development and Validation

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Transcription:

Agilent Application Solution Transfer of a USP method for tolazamide from normal phase HPLC to SFC using the Agilent 126 Infinity Hybrid SFC/UHPLC System Improving peak shape and providing wider UV selectivity Application Note Pharmaceutical QA/QC 4 Normal phase method 19.44 3 2 13.233 Authors Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India 1 1 8 6 SFC method 5 1 15 2 25 3.963 4.848 Vinayak AK Amrita School of Biotechnology Amrita Vishwa Vidyapeetham Kollam, India 4 2 Abstract 1 2 3 4 Normal phase liquid chromatography methods often have long run times and involve environmentally toxic/costly solvents. Supercritical fluid chromatography (SFC) methods on the other hand, are faster, inexpensive, and eco-friendly. SFC involves the use of low viscosity supercritical carbon dioxide, which can be operated at flow rates up to three times higher than liquid chromatography (LC) without losing separation efficiency, thereby leading to faster analysis. This Application Note describes a method to transfer a United States Pharmacopeia (USP) normal phase HPLC assay method to SFC. The Agilent 126 Infinity Hybrid SFC/UHPLC System was used to perform both normal phase as well as the SFC methods. The results show that the SFC method meets the system suitability criteria and displays narrower and taller peaks. In addition, the SFC runs were 4 faster and created 19 lower solvent expenses than the normal phase method. Robustness tests demonstrate excellent results for routine analysis with SFC.

Introduction is an oral, blood glucose lowering drug of the sulfonylurea class. The USP assay method for tolazamide uses a normal phase method that includes chloroform as a sample diluent, while hexane is used as the mobile phase. Chloroform is a known carcinogen, potentially toxic to analysts and expensive to dispose. Hexane is included in the list of chemicals on the US Toxic Release Inventory (TRI). SFC is considered a green technology with a low risk of flammability, due to the use of carbon dioxide (CO 2 ) as a major component of the mobile phase. Recently, SFC has demonstrated that it can replace many achiral LC methods. Especially, compared to normal phase methods, SFC methods offer faster separation without losing efficiency, and faster column re-equilibration 1. This Application Note shows the development of a SFC method to replace a normal phase tolazamide assay in which tolbutamide is used as an internal standard (Figure 1). This method uses methanol as a sample diluent instead of chloroform. The Agilent 126 Infinity SFC/UHPLC Hybrid System was used to perform both the normal phase as well as the SFC method on the single instrument 2. With this unique hybrid solution, there is no need to invest in two individual systems, thus excluding system-to-system variability and saving significant cost and laboratory space. A O O S Figure 1 Molecular structures of (A) and (B). Experimental Instruments An Agilent 126 Infinity Hybrid SFC/UHPLC System (G439A) consisting of the following modules was used: Aurora SFC Fusion A5 module Agilent 126 Infinity Degasser Agilent 126 Infinity SFC Binary Pump Agilent 126 Infinity SFC Autosampler Agilent 126 Infinity Thermostatted Column Compartment Agilent 126 Infinity Diode array Detector In addition the following components were needed: A gilent 126 Infinity Binary Pump (G1312B) Agilent 129 Infinity Universal Valve Drive (G117A) Agilent 2-position/1-port valve kit 6 bar (G4232A) Agilent 126 SFC/UHPLC Hybrid Capillary Kit (G436A) Software Agilent ChemStation B.4.3 H O H B O O S H Reagents and materials All solvents used were HPLC grade. Purified water was used from a Milli-Q water purification system (Millipore, USA). Methanol super gradients were purchased from Lab-Scan. HPLC grade, hexane, tetrahydrofuran, glacial acetic acid, and chloroform were purchased from Sigma-Aldrich (India). and tolbutamide (Vetranal, analytical reagent >99%) were also purchased from Sigma-Aldrich (India). O H 2

Chromatographic parameters The chromatographic parameters for SFC chromatography using the 126 Infinity Hybrid SFC/UHPLC System are shown in Table 1. During a normal phase run, the SFC flow rate and backpressure regulator (BPR) were maintained at a low value of 1 ml/min and 9 bar respectively. These settings help to keep the system under pressure while switching to SFC after normal phase runs. Preparation of standards Preparation of water-saturated hexane: Add 1:1 ratio of pure hexane and H 2 O in a separating funnel mix well. After phase separation, draw out the water from the bottom and collect the remaining water saturated hexane Normal phase internal standard solution: Dissolve a suitable quantity of tolbutamide in alcohol-free chloroform to obtain a solution with a known concentration of approximately 1.5 mg/ml. Normal phase standard solution: Dissolve an accurately weighed quantity of USP RS in normal phase internal standard solution to obtain a solution with a known concentration of about 3 mg/ml. SFC internal standard solution: was accurately weighed out, and dissolved in 1% methanol to obtain 1.5 mg/ml. Parameters Normal phase method SFC method Column: Agilent ZORBAX Rx-SIL 4.6 25 mm, 5 µm (p/n 88975-91) Agilent ZORBAX Rx-SIL 4.6 25 mm, 5 µm (p/n 88975-91) Thermostatted Column Compartment solvent preheating: 25 C 4 C Thermostatted Column Compartment solvent post conditioning: 25 C 4 C Detection: Flow cell: 254/4 nm (Ref 36/1 nm) 1 Hz acquisition rate 1 mm path length, 13-µL volume high pressure flow cell Injection volume: 5 µl* 5 µl Injector program: Yes Yes BPR: 9 bar 15 bar SFC flow rate: 1 ml/min 3 ml/min Normal phase flow rate: 1.5 ml/min ml/min SFC run: - Time % B 4.5 5. 5.1 5.5 Normal phase run: 1% A isocratic - 254/4 nm (Ref 36/1 nm) 1 Hz acquisition rate 1 mm path length, 13-µL volume high pressure flow cell Run time: 3 minutes 5.5 minutes Mobile phase: Hexane, water-saturated hexane, tetrahydrofuran, methanol, and glacial acetic acid (475:475:2:15:9). A: Supercritical fluid CO 2 B: Methanol *The injector volume was decreased from 1 µl to 5 µl to fit the 5-µL fixed loop. Table 1 Chromatographic parameters used in the Agilent 126 Infinity Hybrid SFC/UHPLC System. 12 12 SFC standard solution: Dissolve an accurately weighed tolazamide in SFC internal standard solution to obtain a concentration of approximately 3 mg/ml. 3

Procedure The Agilent 126 Infinity Binary Pump of the hybrid system used the normal phase pump seal (p/n 95-142). The pump was equilibrated with isopropyl alcohol prior to normal phase solvents. The 126 Infinity SFC/UHPLC Hybrid System was operated in normal phase mode by switching the 2-position/1-port valve. The normal phase runs were performed using the normal phase standard solution to determine the USP system suitability parameters. The 2-position/1-port valve was then switched to SFC mode for the SFC runs, to determine the system suitability parameters. In the SFC mode, robustness studies were also performed using SFC standard solution. To evaluate the robustness of the method, six method parameters were evaluated: Flow rate ± 2% Results and discussion Separation and detection The system suitability mixture was used to optimize the separation conditions. The separation was initially performed at initial SFC conditions (TCC temperature of 35 C, backpressure regulator at 15 bar, Agilent ZORBAX Rx-Sil column and flow rate of 3. ml/min). The methanol percentage was varied from 2% B (methanol) isocratic by decreasing it systematically to 5% isocratic in different runs. However, the separation was found to be good in gradient runs where the mobile phase was changed from % B to 12% B in 4.5 minutes. 4 Normal phase method Flow rate optimization followed mobile phase optimization. The flow rate was changed from 1.5 ml/min to 3.5 ml/min in increments of.2 ml/min where area/rt of the peaks was recorded. The ideal flow rate was determined to be 3 ml/min. The TCC temperature was also varied from 25 C to 45 C where the ideal temperature was found to be 4 C. Figure 2 shows the chromatogram of the SFC method performed at the final optimal condition overlaid with the USP normal phase method. The detector was set at 254 nm, as suggested in the USP method. An additional peak appears at 2.9 minutes in the SFC method (Figure 2). This peak originates from the super gradient methanol used to dilute the sample. 19.44 Column temperature ± 2.5% Injector volume ± 3% Absorption wavelength ± 1 nm Modifier concentration ± 1% Backpressure ± 2 bar 3 2 1 13.233 For each robustness parameter, a SFC standard preparation of 1 ppm solution of tolazamide and tolbutamide was injected, and six replicates were used to calculate area, RT. The resolution of tolazamide was compared to tolbutamide. The original method (without robustness parameters) was also performed in six replicates. The percentage deviation (% accuracy) of area/retention time (RT) was calculated from the original method. 1 8 6 4 2 SFC method 5 1 15 2 25 3.963 4.848 min 1 2 3 4 5 min Figure 2 Separation of standard solution of tolazamide and tolbutamide using an Agilent ZORBAX Rx- SIL 4.6 25 mm, 5 µm column. 4

The system suitability test was performed using both methods. The SFC method provided similar relative retention time (RRT) values and resolution (Table 3). The area precision for four replicate injections was performed as per USP method. The results are better in the SFC method than the USP normal phase method. An added benefit of SFC is being able to run the sample at a faster flow rate and use methanol as diluent. SFC chromatographic runs produced sharper and narrower peaks compared to normal phase methods. The tolazamide peak width at half height was.4 minutes, 7 lower compared to the normal phase method (Table 4). Another advantage of the SFC method versus the USP normal phase method is a significant cost reduction (Table 5). A 4 decrease in analysis time and a 19 decrease in cost were achieved with the SFC method for every analytical run. Assuming analysis time to be US $ 8/hr, the overhead cost would decrease to US $ 2/hr. System suitability Parameter USP method USP normal phase method SFC method Relative retention time 1. 1. 1..6.7.8 Resolution NLT 2 13 13 Std injection (n=4) () RSD area NMT 2% 2% 1% Table 3 USP tolazamide system suitability acceptable limits compared with USP normal phase method and SFC method. NLT = not less than, NMT = not more than. Peak properties Parameter Normal phase method SFC method Peak width at half height.3.2.4.4 Peak height 43 21 12 66 Peak symmetry 1.7 1.3 1. 1. Peak tailing.7.9 1. 1. Table 4 Comparison of peak properties obtained from normal phase and SFC method. USP normal phase method SFC method Savings Analysis time per sample (min) 2 5.5 3.6 x Cost per sample (US $) 248 13.2 19 x Table 5 Savings in analysis time and solvent cost per 1 analysis using SFC. 5

Increased sensitivity in SFC through lower UV cut off The solvents/modifiers used in the normal phase methods have various UV cut off regions. Table 6 shows the UV cut off of normal phase solvents used in the tolazamide method. Chloroform displaying the highest cut off of 245 nm, restricts the usage of wavelength at a lower region., with a UV maximum at 228 nm (Figure 3), cannot be used with chloroform. The typical SFC modifier and sample diluent methanol displays a UV cut off of 25 nm, thereby making it ideal for the analysis of tolazamide. A 28 increase of sensitivity was observed when tolazamide was analyzed using 228 nm (see table in Figure 4). UV cut off Solvents Common normal phase solvents Common SFC method Hexane 195 - THF 212 - Chloroform 245 - Methanol 25 25 Super critical fluid CO 2-19 Table 6 UV cut off values of solvents used in normal phase and SFC methods 3. Absorbance 3 25 2 15 1 228 nm (UV max for tolazamide) 254 nm (USP) 5 22 24 26 28 3 32 34 36 38 Wavelength nm Figure 3 The UV spectra of tolazamide showing 228 nm as UV maximum and 254 nm wavelength recommended as per USP. Compound name 254 nm 228 nm Area 254 nm S/N 228 nm S/N 322 984 25 5934 199 636 15 4377 3 25 3.98 4.856 2 15 1 228 nm 5 254 nm 3 3.25 3.5 3.75 4 4.25 4.5 4.75 5 5.25 min Figure 4 Chromatogram of tolazamide and tolbutamide performed at 228 nm (red) and 254 nm (blue). Insert: peak areas and signal-to-noise ratio (S/N) obtained using two different UV acquisition wavelengths. 6

Robustness To test the robustness of the method, the SFC standard solution containing 3 mg/ml of tolazamide and 1.5 mg/ml of tolbutamide was used. Six critical method parameters (flow rate, column temperature, injector volume, absorption wavelength, modifier concentration, and backpressure) were varied individually (Table 7) 4. The peak areas from the six replicate injections were compared. The allowed deviations for the area and retention time were set to ± 5% and ± 3% respectively. The red numbers indicate where the result exceeded the allowed deviation. Changes in flow rate and TCC temperature do not vary the method. The injection volume of 15 µl is taken in order to overfill the fixed injection loop of 5 µl three times. A deviation in injection volume of ± 3% from 15 µl does not affect the method. It is recommended to use 15 µl to overfill the 5 µl injection loop. The absorption wavelength in robustness studies was chosen to be 254 nm as per USP recommendation. However, for better sensitivity, 228 nm can also be used. A backpressure regulator change of 2 bar affected only the internal standard area. The resolution of tolazamide was not found to be changing at any of the robustness testing methods. Robustness results indicate that the method is reliable for normal usage, where, to a great extent, the performance remains unaffected by deliberate changes of the method parameters. However, some parameters, such as the backpressure are critical, and must be carefully controlled. Parameters Changes % Area % RT Resolution % Area % RT Flow: 3 ml/min ± 2% High: 3.6 ml/min.5 1. 12.8 4. 1.2 Low: 2.94 ml/min -3.1-1.1 12.6 -.8-1.3 TCC: 4 C ± 2.5% High: 41 C -.9 -.6 12.8.9 -.5 Low: 39 C -1.3.3 12.7.5.2 Injector: 15 µl ± 3% High: 15.5 µl -1.2 -.3 12.8.4.2 Low: 14.5 µl -1.4 -.3 12.8. -.4 Wavelength: 254 ± 1 nm 255 nm -4.3 -.4 12.9-3.2 -.4 253 nm 3.8. 12.9 3.3 -.1 Modifier concentration: High: 12.2% B.5.4 12.9-1.1.1 12% B ± 1% Low: 11.8% B.2-1. 12.8-1.8-1 Backpressure: 15 ± 2 bar High : 152-1.4. 12.8-6.7 -.2 Low : 148-1.9 -.7 12.9-6.7 -.9 Table 7 SFC robustness test results. The red values in the table indicate that the deviations exceeding the allowed limits of 5% for area and 3% for retention time. 7

Conclusion The 126 Infinity Hybrid SFC/UHPLC System was used to develop a novel SFC tolazamide assay method. This method was then compared to the original USP normal phase method. While meeting the system suitability requirements, the new SFC method was 4 faster and 19 less expensive than the normal phase method. The peaks obtained by SFC method were narrower and taller compared to USP method. The low UV cut off of SFC modifiers and sample diluents allow for greater flexibility to choose appropriate UV regions with the SFC method, which can lead to increased sensitivity as shown in this example. Robustness test results were excellent for the SFC method. Additionally, the SFC method does not require the purchase and disposal of expensive environmentally hazardous chemicals. Hence, the newly developed tolazamide SFC method provides a fast, cost effective, safe, and sensitive solution compared to normal phase method. References 1. Colin F. Poole, Progress in Packed Column Supercritical Fluid Chromatography: Materials and Methods, J. Biochem. Biophys. Methods, 43: 3-23, 2. 2. Martin Vollmer, Markus Becker, Agilent 126 Infinity Hybrid SFC/UHPLC System Agilent Technical Overview. Pub No. 599-9514EN, 212. 3. C.M.White, R.K.Houck, Supercritical fluid chromatography and some of its applications: A review, J. High Resolution Chromatograph & Chromatography Com, 9: 4-17, 1986. 4. Bieke Dejaegher, Y van Vander Heyden, Ruggedness and Robustness Testing, J. Chrom A., 1158: 138-157, 27. www.agilent.com/chem/sfc Agilent Technologies, Inc., 212 Published in the USA, June 1, 212 5991-395EN