HIGH-PRECISION INTERCOMPARISON AT ISOTRACE

Similar documents
corresponding l Nuclear Physics Laboratory and Quaternary Isotope Laboratory University of Washington, Seattle, Washington INTRODUCTION

MAURICE ARNOLD, EDOUARD BARD, PIERRE MAURICE HELENE VALLADAS and J C DUPLESSY

that our measurement system can not only produce precise dates from very small amounts of materials, but that these dates are also accurate.

RADIOCARBON ANALYSIS OF DIFFEREENT SAMPLES AT BINP AMS

A Survey of Modern Radiocarbon Dating

apparent presence of 1

RADIOCARBON DATING BEYOND 50,000 YEARS BY LIQUID SCINTILLATION COUNTING

THE CARBONATE 14C BACKGROUND AND ITS COMPONENTS AT THE LEIBNIZ AMS FACILITY

SOURCES OF RANDOM ERROR IN THE DEBRECEN RADIOCARBON LABORATORY

ACCELERATOR MASS SPECTROMETRY AT THE LUND PELLETRON ACCELERATOR

DETERMINATION OF RADON BY LIQUID SCINTILLATION a/f PARTICLE SPECTROMETRY: Towards the Resolution of a 14C Dating Problem

RATE s Radiocarbon: Intrinsic or Contamination?

CAPABILITIES OF THE NEW SUERC 5MV AMS FACILITY FOR 14 C DATING

Comparison of Different Methods of Environmental Radioactivity Measurements at the Zagreb Radiocarbon and Tritium Laboratory

RICHARD BURLEIGH, MORVEN LEESE, and MICHAEL TITS Research Laboratory, The British Museum, London WC 1 B 31)G, England

Analysis of Radiocarbon Dating Statistics in Reference to the Voynich Manuscript

D. Kunikita, K. Yoshida, Y. Miyazaki, H. Matsuzakii, H. Kato, K. Suzuki, T. Sato, G. Medvedev, E. Lipnina

Unit A (225 m.y. old) Unit B (how old is it?) Unit C (275 m.y. old)

Measurable 14 C in Fossilized Organic Materials: Confirming the Young Earth Creation-Flood Model

THE FIFTH INTERNATIONAL RADIOCARBON INTERCOMPARISON (VIRI): AN ASSESSMENT OF LABORATORY PERFORMANCE IN STAGE 3

Minze Stuiver and Henry A. Polach. Discussion: Reporting of 14 C Data. Originally published in Radiocarbon 19(3), 1977, p

THE NEW 14 C ANALYSIS LABORATORY IN JENA, GERMANY. A Steinhof 1,2 G Adamiec 3 G Gleixner 1 GJ van Klinken 4 T Wagner 1

3.5. Accelerator Mass Spectroscopy -AMS -

HIGH-PRECISION 14C MEASUREMENT OF GERMAN AND IRISH OAKS TO SHOW THE NATURAL 14C VARIATIONS FROM 7890 TO 5000 BC

FIRST RADIOCARBON MEASUREMENTS AT BINP AMS

A STUDY ON TRAPPING CO 2 USING MOLECULAR SIEVE FOR 14 C AMS SAMPLE PREPARATION

Advances in radiocarbon measurement of water samples

The Norwegian Institute of Technology, Trondheim, Norway

C - AMS. Introduction Identification of 14 C. Remarks for users

DATING POLAR ICE BY 14C ACCELERATOR MASS SPECTROMETRY MICHAEL ANDREE, JUERG BEER, H P LOETSCHER, ERNST MOOR, H.ANS OESCHGER

STUDYING RADIOCARBON PAPERS MAY SOLVE THE ENIGMA OF THE RADIOCARBON DATING OF THE SHROUD OF TURIN.

It was first employed by Purser and others (1977) and Bennett

Seattle, Washington 98195

RADIOCARBON DATING REPRODUCIBILITY AT THE MUSEO DE LA PLATA RADIOCARBON LABORATORY

Ultratrace analysis of radionuclides by AMS

Measuring the Age of things (Astro 202 2/12/08) Nomenclature. Different Elements. Three Types of Nuclear Decay. Carbon 14 Decay.

Accelerator Mass Spectroscopy

R JUERG BEER, HANS OESCHGER, and MICHEL ANDREE Physikalisches Institut, Universitat Bern, CH-3012 BERN, Switzerland

12- is measured in a side Faraday cup at the low-energy end of the accelerator

PUSHING THE LIMITS OF AMS RADIOCARBON DATING WITH IMPROVED BAYESIAN DATA ANALYSIS

University of Groningen. The HVEE C-14 system at Groningen Gottdang, Andreas; Mous, Dirk J.W.; van der Plicht, Johannes. Published in: Radiocarbon

ISOTOPIC FRACTIONATION OF NORWEGIAN MATERIALS FOR RADIOCARBON DATING STEINAR GULLIKSEN. Trondheim, Norway

HIGH-SENSITIVITY RADIOCARBON DATING IN THE 50,000 TO 70,000 BP RANGE WITHOUT ISOTOPIC ENRICHMENT

LABORATORY OF COSMOCHEMISTRY RADIOCARBON MEASUREMENTS I

National Bureau of Standards, Washington, DC, 20234

CLIMATE CHANGE, CATASTROPHE AND THE TIDES OF HISTORY. 1. CLIMATE THE LONG VIEW.

IUAC Pelletron Accelerator

TOWARDS HIGH-PRECISION AMS: PROGRESS AND LIMITATIONS

Take the measurement of a person's height as an example. Assuming that her height has been determined to be 5' 8", how accurate is our result?

AN OVERVIEW OF ALL THREE STAGES OF THE INTERNATIONAL RADIOCARBON INTERCOMPARISON

Lithogeochemistry Using a Portable X-Ray Fluorescence (pxrf) Spectrometer and Preliminary Results From the Eagle Ford Shale

Physics of Aquatic Systems II

DEVELOPMENT OF A COMBUSTION SYSTEM FOR LIQUID OR GAS SAMPLES

LAST TAUPO ERUPTION: EVIDENCE FROM TREE-RING SEQUENCES

SCIENTIFIC DATING IN ARCHAEOLOGY

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

NUMB3RS Activity: How to Get a Date. Episode: Bones of Contention

Advancing Microplastic Laboratory Analysis

SYSTEMATIC INVESTIGATIONS OF 14C MEASUREMENTS AT THE VIENNA ENVIRONMENTAL RESEARCH ACCELERATOR

4/4/17. Dating Methods. Dating in Archaeology. These things date to 2500 B.C. (but there are no completion dates written anywhere on or near them)

Tuesday 10 June 2014 Afternoon

Time: 1 hour 30 minutes

The QL system has been described by Grootes and Stuiver (1979). The QL system consists of 23 single columns combined to 5 completely

THE STATISTICS OF LOW-LEVEL COUNTING USING THE NEW GENERATION OF PACKARD LIQUID SCINTILLATION COUNTERS

Lecture Notes on Inductive Definitions

Physics 2020 Laboratory Manual

Uncertainty, Error, and Precision in Quantitative Measurements an Introduction 4.4 cm Experimental error

Name Date Time to Complete

PROCEEDINGS OF THE 16 TH INTERNATIONAL 14 C CONFERENCE, EDITED BY W. G. MOOK

THE ROLE OF 36(1 AND 14C MEASUREMENTS IN AUSTRALIAN GROUNDWATER STUDIES. example approximating this requirement.

Statistics of Radioactive Decay

Pacific Northwest Rock Lab, Part II. Igneous Rocks. Name Per.

Radioactive Clocks or Radiometric Dating

The Canadian Ceoscience Knowledge Network. - A Collaborative Effort for Unified Access to Ceoscience Data

Time: 1 hour 30 minutes

3.2 Ga detrital uraninite in the Witwatersrand Basin, South. Africa: Evidence of a reducing Archean atmosphere

Sample Homogeneity Testing

GAS PROPORTIONAL VERSUS LIQUID SCINTILLATION COUNTING, RADIOMETRIC VERSUS AMS DATING PALL THEODORSSON

Radiometric Dating and the Age of the Earth

Calibration of the Modular Neutron Array (MoNA)

MEEM Design of Experiments Final Take-home Exam

7.5 to ca./< 4 ka >7.5 & ca./< 4 ka

THE RELIABILITY OF AMS RADIOCARBON DATING OF SHELLS FROM CHINA. Weijian Zhoul M J Headl Fubao Wang2 D J Donahue3

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry

Earth s Changing Surface Chapter 4

Martin David Kamen (August 27, 1913 August 31, 2002) and Sam Ruben (November 5, 1913 September 28, 1943) discovered how to synthesize the isotope

Phys 243 Lab 7: Radioactive Half-life

* * MATHEMATICS (MEI) 4767 Statistics 2 ADVANCED GCE. Monday 25 January 2010 Morning. Duration: 1 hour 30 minutes. Turn over

Your quiz will have some multiple-choice questions, matching, fill-in-the-blank questions, and short answer. To review, study DQs, notes from class,

IMPROVED COLLAGEN EXTRACTION BY MODIFIED LONGIN METHOD

GRADE 7 MATH LEARNING GUIDE. Lesson 26: Solving Linear Equations and Inequalities in One Variable Using

RADIOCARBON DATING OF LIME FRACTIONS AND ORGANIC MATERIAL FROM BUILDINGS

Overview: In this experiment we will study the decay of a radioactive nucleus, Cesium. Figure 1: The Decay Modes of Cesium 137

C 7 L D ARGONNE NATIONAL LABORATORY, ARGONNE, ILL mm.'it-is, isai. defvf-tjds $ G. UB/TIB Hannover <ANL~PHYSICS DIVISION'

Back to the Big Question

Radiocarbon dating. Physical age determinations using natural radionuclides

Vocabulary and Section Summary B

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

X-Ray Emission and Absorption

Transcription:

[RADIOCARBON, VOL 32, No. 3, 1990, P 335-339] HIGH-PRECISION INTERCOMPARISON AT ISOTRACE ROELF P BEUKENS IsoTrace Laboratory, University of Toronto, Toronto, Ontario M5S 1A7, Canada ABSTRACT. I conducted a high-precision comparison at the 0.2% to 0.3% level with samples supplied by the radiocarbon laboratory of the Quaternary Research Center at the University of Washington (QRC). Four samples with ages ranging from modern to > 50,000 BP were dated in a blind test. The absence of cosmic-radiation background in AMS dating is a major advantage for dating samples > 35,000 BP. The reliability of AMS dates > 35,000 BP depends entirely on understanding the contamination processes. By comparing results with laboratories capable of sample enrichment, such as QRC, it is possible to identify and estimate the intrinsic 14C in the background samples as well as the contamination introduced by sample preparation. INTRODUCTION Beukens, Gurfinkel and Lee (1986) tested precision and accuracy at the start of operations of the IsoTrace Laboratory. I demonstrated precision at the 0.25-0.37% level in reproducibility measurements, which showed that the precision, calculated from all known random errors, accounted for all the variance in the data. The intercomparison with samples from the Geological Survey of Canada (GSC) and the Australian National University (ANU) radiocarbon dating facilities showed good agreement at the 1% level, but several discrepancies did exist. Although some of these could be attributed to the way the GSC shell ages were reported (normalized to a base of S13C = O%o), discrepancies for some of the older samples were evident. Subsequently, I observed occasionally similar discrepancies on other samples. To understand the cause of these discrepancies and to test our laboratory at high precision, I initiated a four-part research program: 1. Determination of the machine background 2. Determination of the contamination contributed by sample preparation 3. Intercomparison of old and background samples with the GSC 4. High-precision intercomparison at the 0.2% level. Analysis Procedures. All samples, analyzed as part of this program, contained 200-3OO g of carbon. Prior to analysis, each sample was cleaned for 10 minutes with the primary cesium sputter beam. To avoid cratering during the analysis, I measured 16 spots on a sample. For every spot, the 14C+3 ions were accumulated for 10 seconds while the 12C+3 and 13C+3 currents and their variances were determined before and after the measurement. This process was repeated for all samples, including the standards, 12-20 times until sufficient precision was obtained. I obtained conventional radiocarbon dates, corrected for natural and sputter fractionation to base of &3C = -25%o, by comparison with the averaged results of 2 or 3 NBS I oxalic acid standards, as described previously (Beukens, Gurfinkel & Lee 1986). The measurements were conducted over a suf- ficiently long period to obtain finite results and avoid bias due to Poisson statistics. I reprocessed and remeasured every sample at least once and the results presented here are the weighted averages of these measurements. The background samples, in particular, were measured several times over a period of several months. No time-dependent variations were observed in these measurements, but remeasurements are being performed on a regular basis as a quality check. All results are presented without background or contamination corrections unless otherwise specified. MACHINE BACKGROUND Machine background in Accelerator Mass Spectrometry (AMS) is due to electronic noise, ion source contamination and 12C ions from the 12C' sputter tail, mimicking "C ions (E/q ambiguity). 335

336 Roelf P Beukens At the IsoTrace AMS facility, this last contribution is eliminated by a 45 electrostatic analyzer after the ion source. Electronic noise can be measured separately and was shown to contribute <0.001 pmc to the background. Ion source contamination appears to be unavoidable because at least 90% of all sputtered 14C is deposited somewhere in the source region. It can be minimized, however, by thorough cleaning and a proper ion source design. I measured the contribution of this contamination to the apparent age of an infinitely old sample on blank aluminum sample holders that yielded 0.002 ± 0.001 pmc or an apparent age of 85,000 ± 4000 BP. CONTAMINATION DUE TO SAMPLE PREPARATION I measured this contamination by preparing and analyzing samples that are geologically very old. Most old materials, however, are not free of intrinsic 14C contamination because such samples can easily be contaminated during their long period of interment or during conservation after excavation or retrieval. As it is impossible to predict this contamination, many samples had to be analyzed to determine which were best suited for this purpose. Table 1 shows the current results of this contamination study. Sample preparation at AMS facilities can be divided into two stages. The first stage consists of sample pretreatment and conversion into CO2 either by acid hydrolysis of carbonates or combustion of organics. The second stage consists of the conversion of CO2 into graphite. The pretreatment, combustion and hydrolysis techniques used here, are identical to those employed for any submitted sample and have not changed appreciably over the last five years. The lower 14C levels in these measurements should therefore be attributed entirely to the lower intrinsic 14C contamination of these samples and not to changes in sample preparation or analysis techniques. TABLE 1 Sample-preparation-related background measurements Sample treatment Material 14C content age (pmc) Graphitization CO2 from natural gas 0.077 0.005 540 Acid hydrolysis Marble 0.076 0.009 920 Combustion Wood fragment (Yukon) 0.152 0.025 1310 Fossilized redwood (Axel Heiberg Island) 0.211 0.018 680 At IsoTrace, the graphitization process is different from that of other AMS laboratories. I first convert the CO2 into acetylene using the standard Li-carbide synthesis. The acetylene is then dissociated in an electrical high voltage AC discharge and produces two machine-ready samples at one time (Beukens & Lee 1981). Industrial CO2. produced by the combustion of natural gas, yields an apparent age of 57,630 ± 540 BP or 0.077 ± 0.055 pmc. Thus, this value represents the upper limit for the contamination contribution of the graphitization process. Many carbonate samples were analyzed to test the combined contamination contribution of acid hydrolysis and graphitization. For several years, I obtained results of 0.15 to 0.13 pmc for optical grade calcite (Iceland Spar) and a carbonate aggregate from the upper Amazon region. Recently, a marble has yielded a result of 0.076 ± 0.009 pmc, equivalent to an apparent age of 57,690 ± 920 BP. This upper-limit estimate for the combined acid hydrolysis and graphitization processes is statistically identical to that of the graphitization process alone. I obtained an upper limit of 52,140 ± 1310 BP or 0.15 ± 0.02 pmc for the combined contribution of the combustion and graphitization processes, on the cellulose fraction of a wood

High-Precision Intercornparison at IsoTrace 337 sample submitted to the laboratory. As this was a surface find from an old shoreline of ancient Lake MacKenzie (Yukon, Canada) it is not exactly the most ideal sample and better suited material will surely be found. One sample that I expected to be the ideal blank is a well-preserved redwood sample from early Tertiary fossil forests on Axel Heiberg Island (North West Territories, Canada). The result (Table 1) was very disappointing, however, as the 14C content of this sample is significantly higher than that of the first wood sample, implying that the Axel Heiberg wood sample is contaminated. INTERCOMPARISON WITH THE GEOLOGICAL SURVEY OF CANADA As part of the search for suitable background material, I re-analyzed calcite and anthracite background samples from the GSC, using identical pretreatment procedures. The GSC assumes these samples to be 14C-free and subtracts their count rates from the samples, analyzed in their facility. This is intended to provide a correction for the cosmic radiation and electronic background as well as a correction for sample preparation-induced contamination. However, the IsoTrace re-analysis of these samples (Table 2) clearly shows that these samples are not 14C-free because they yield higher 14C contents than the limits in Table 1. This means that the GSC dates are over-corrected, resulting in ages that are too old. This fact is demonstrated by the re-analysis of the CO2 from a shell and a wood sample for which the GSC quotes ages of > 50,000 BP (Table 3). At the 2a limit, the difference between the uncorrected IsoTrace results and those for the GSC background materials agrees with the 0.2 pmc limit. Therefore, the IsoTrace results, corrected by the background results of Table 1, indicate that these samples are actually 41,000-46,000 BP instead of > 50,000 BP, as quoted by the GSC. It is to be expected that many radiocarbon dating facilities, which have not checked the intrinsic 14C content of their background material, are also quoting ages older than the actual ages. TABLE 2 Re-analysis of the background samples from the Geological Survey of Canada Sample treatment Material 14C content (pmc) age Acid hydrolysis Calcite 0.54 0.04 600 Combustion Anthracite 0.36 0.03 700 TABLE 3 Intercomparison with the Geological Survey of Canada Wood sample GSC measurement pmc IsoTrace measurement ± 0.03 pmc GSC anthracite 0.36 ± 0.03 pmc Difference 0.10 ± 0.04 pmc Shell sample GSC measurement <0.20 pmc IsoTrace measurement 0.65 ± 0.04 pmc GSC calcite 0.54 ± 0.04 pmc Difference 0.11 ± 0.06 pmc

338 Roel f P Beukens HIGH-PRECISION INTERCOMPARISON WITH THE QUATERNARY RESEARCH CENTER OF THE UNIVERSITY OF WASHINGTON The radiocarbon facility at QRC regularly dates samples to a precision of better than 0.2% and has a proven dating capability to > 65,000 BP. Their contamination due to sample preparation has been carefully checked using a 14C enrichment process. Thus, QRC is an ideal partner for intercomparison checks. I divided this intercomparison study into three analytical stages: 1. Comparison of NBS I oxalic acid standards 2. Comparison of pretreated samples, submitted as CO2 3. Comparison of untreated samples. The results of the first two stages are presented below. Comparison of NBS Standards. As all sample ages are obtained by comparing their 14C content to standards, this is the logical place to start. I do not actually prepare NBS oxalic acid standards at IsoTrace. They are prepared at the United States Geological Survey (USGS) as acetylene. This acetylene is then graphitized at the IsoTrace laboratory. I compared these standards with standards prepared from CO2 supplied by QRC. The ratio of the IsoTrace standard to the QRC standard was 0.9992 ± 0.0020, giving the IsoTrace standard a clean bill of health. Comparison of Pretreated Samples. For this stage, Minze Stuiver selected three samples of widely varying ages, which had been analyzed previously to high precision by QRC. The samples were supplied as CO2. To ensure a true blind test, the QRC ages were not made available until after I reported my results. Table 4 shows the uncorrected and background-corrected results of this intercomparison. It is unusual for a decay-counting laboratory to show uncorrected results because background subtraction is integral to their procedures. AMS laboratories frequently do show uncorrected results as the effect of the correction is negligible, in most cases. A background- corrected AMS date is recalculated after subtracting the blank correction from the results for the sample as well as the results for the standards used. I used the measured upper limit for the contamination by the IsoTrace graphitization process of 0.077 ± 0.005 pmc as an estimate for the background correction factor. The agreement at this level of precision is quite satisfactory and the comparison shows no statistically significant offsets. TABLE 4 High-precision intercomparison with the Quaternary Research Center (QRC) QRC results IsoTrace results Lab no. Age of samples analyzed age corrected age QL-11288 4132 ± 18 18 QL-11312 6973 ± 20 16 20 20 QL-1787 > 55,000 16 340 600

High-Precision Intercomparison at IsoTrace 339 CONCLUSION Stage 3, the analysis of the untreated samples, is currently underway. Comparison of these results with the Stage 2 results will allow a better estimate of the upper limit for the contamination by the combustion process. Over the years, many "very old" samples, which had never actually been measured, have been submitted to our laboratory. I think it is prudent to distrust such samples until they have been measured, as most samples, to a certain extent, appear to be contaminated. Finally, it is quite surprising that old samples have never been included in international intercomparisons. I believe that this is a major oversight as many of the participating laboratories claim to date samples older than 35,000 BP. REFERENCES Beukens, RP and Lee, HW 1981 The production of small carbon samples by RF dissociation of acetylene, In Kutschera, W, ed, Symposium on accelerator mass spectrometry, 2nd, Proc. Argonne Natl Lab: 416-425. Beukens, RP, Gurfinkel, DM and Lee, HW 1986 Progress at the Isotrace radiocarbon facility. In Stuiver, M and Kra, RS, eds, Internatl 14C conf,12th, Proc. Radiocarbon 28(2A): 229-236.