Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Similar documents
Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Thermal Desorption Technical Support

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Maximizing Sample Throughput In Purge And Trap Analysis

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Determination of Volatile Organic Compounds in Air

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Solid Phase Micro Extraction of Flavor Compounds in Beer

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

CIA Advantage-xr. Cryogen-free automated canister and whole-air sampling system

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Target Compound Results Summary

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

EPA TO-17 Volatile Organic Compounds

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

STANDARD OPERATING PROCEDURES

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Application Note 032

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Methanol Extraction of high level soil samples by USEPA Method 8260C

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Ed George and Anaïs Viven Varian, Inc.

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Optimizing. Abstract: is standardd. procedures. altered to

2017 EPA Method Update Rule and EPA Method 624.1

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Application Note 059 Direct Desorption of Car Trim Materials for VOC and SVOC Analysis in Accordance with VDA Method 278

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

Meeting NJ Low Level TO-15 Air Testing Method Requirements

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Thermal Desorption Technical Support

TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Study of Residual Solvents in Various Matrices by Static Headspace

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Date: March 6, 2008 RWDI Reference #: W B Pages (Including Cover):

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Maximizing Production While Minimizing Costs

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analytical Trap Comparison for USEPA Method 8260C

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

Optimizing Standard Preparation

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Author. Abstract. Introduction

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

NIST gas standards containing volatile organic compounds in support of ambient air pollution measurements

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

CALA Directory of Laboratories

J. Li et al. Correspondence to: S. D. Xie

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Indoor Air Sampling and Forensic Analysis at Peter Cooper Village and Stuyvesant Town Properties October 2017

Headspace Technology for GC and GC/MS: Features, benefits & applications

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

TD-100. Automated thermal desorber. Universal TD system for up to 100 RFID-tagged or untagged tubes

Transcription:

Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the excellent performance offered by Markes International s automated cryogen-free thermal desorption (TD) system for the analysis of volatile organic compounds (VOCs) in ambient air, sampled using sorbent tubes. The system used is compliant with Chinese EPA Method HJ 3, and features automated quantitative re collection of split flows, which allows repeat analysis, method development, and results verification. Introduction Volatile organic compounds (VOCs) are important precursors for atmospheric photochemical reactions, including those that generate low-level ozone and particulate matter that give rise to poor air quality. In addition, some of these VOCs are harmful to health in their own right, and these air toxics or hazardous air pollutants (HAPs) are therefore monitored in many industrial and urban environments. They range in volatility from chloromethane (methyl chloride) and acetone to hexachlorobutadiene and dodecene, and include polar as well as nonpolar compounds. Several national and international standard methods have been developed for air toxics and related applications, including US EPA Methods TO-5 (canisters) and TO-7 (sorbent tubes). In China, the Twelfth Five-Year Plan () emphasises the need to deepen the prevention and control of particulate matter pollution, and strengthen the control of volatile organic pollutants and toxic emissions. Control of VOCs has therefore become a priority in recent years, as reflected by the 3 Atmospheric Pollution Control Act and standard method HJ -3, released by the Chinese Ministry of Environmental Protection (MEP), and titled Ambient air Determination of volatile organic compounds Sorbent adsorption and thermal desorption gas chromatography mass spectrometry method.

In response to demand for measurement of air toxics, analytical technologies have been developed that offer an automated platform compliant with sorbent-tube-based methods. In this application note, we describe the sampling of VOCs in ambient air using sorbent tubes, followed by HJ -compliant analysis using an automated, cryogen-free thermal desorption-gas chromatography-mass spectrometry (TD-GC/MS) system. Background of thermal desorption Thermal desorption (TD) is a versatile GC preconcentration technology used to analyze volatile and semivolatile organic compounds (VOCs and SVOCs) in a wide range of sample types. By concentrating organic vapors from a sample into a very small volume of carrier gas (Figure ), TD maximizes sensitivity for trace-level target compounds, helps minimize interferences, and routinely allows analyte detection at the ppb level or below. It also greatly improves sample throughput by allowing full automation of sample preparation, desorption/extraction, preconcentration, and GC injection. The xr range of TD instruments from Markes International enhances these capabilities, offering a wide analyte range (C -C including reactive species), automated re-collection and reanalysis of split portions for method validation and compliance with standard methods, optional internal standard addition for improved confidence in results, and electronic/manual options for control of carrier gas. In this study, we use the TD-xr for fully automated analysis of up to sorbent tube samples. Experimental Standard A -component gas standard in nitrogen (which includes the 3 target compounds in HJ ), carried in pure helium, was loaded onto the sorbent tubes using a Calibration Solution Loading Rig (Markes International). Table. Instrumental parameters. A B Tube desorption and inlet split: Sample tube heated in a flow of carrier gas and analytes swept into an electrically cooled focusing trap, typically held between ambient and 3 C. Trap desorption and outlet split: Focusing trap rapidly heated (up to C/s) in a reverse flow of carrier gas (backflush operation), to transfer the analytes to the GC column. Sample tube Sample tubes and traps can contain multiple sorbents, for analysis of an extended range of analytes. Focusing trap Focusing trap Split/re-collection tube During either stage, the flow of analytes can be split and re-collected onto a clean sorbent tube. Split/re-collection tube To GC Parameter Value Real air sample Pumped sampling: ACTI-VOC (Markes International) Flow rate: 5 ml/min for minutes (total volume L) TD Instrument: TD-xr (Markes International) Tube: Universal (Markes International part no. C3-AXXX-5) Trap: Air toxics (Markes International part no. U-T5ATA-S) Tube dry-purge:. minutes Tube desorb: C ( minutes) Trap low: 5 C Trap heating rate: C/s Trap high: C for 3 minutes TD flow path: C GC Carrier gas: Helium GC column: Agilent J&W DB-, m.8 mm,. µm Mode: Constant-flow,. ml/min Oven ramp: 35 C (3 minutes), then 5 C/min to 9 C ( minutes) MS Ion source: 5 C Quadrupole: 5 C Transfer line: C Full scan range: m/z 35-3 Figure. How two-stage thermal desorption works.

Results and Discussion Chromatography Figure shows the chromatogram obtained for a sample equivalent to L of a ppbv standard. Excellent peak shape was obtained for all analytes, as illustrated for the challenging polar compound trichloromethane (expanded peak). These sharp peaks resulted from the rapid desorption of the focusing trap in the TD instrument, and are maintained even with splitless analysis, ensuring optimum sensitivity for trace-level analytes. Linearity System linearity was assessed by analyzing a variety of VOCs at.5,,, 5, and 9 ppbv (equivalent to a L sampling volume, and consistent with ambient concentrations). All R values were above.99 (Figure 3 and Table ). 8 Peak area,,-trichloro-,,-trifluoroethane (#) Benzene (#) Toluene (#5) Benzyl chloride (#3),-Dichloroethane (#5) Figure 3. Linearities for five key compounds from the standard. R.998.998.99.995.9979 8 Concentration (ppbv, L equivalent) 8 3 Abundance 3* 5.8.9.9 7 9 8 3 8 5 7 9 3 5 7 3 9 3 8 3 33 3 5 7 8 9 3.,-Dichloroethene.,,-Trichloro-,,-trifluoroethane 3. Chloropropene*. Dichloromethane 5.,-Dichloroethane. cis -,-Dichloroethene 7. Trichloromethane 8.,,-Trichloroethane 9. Tetrachloromethane.,-Dichloroethane. Benzene. Trichloroethene 3.,-Dichloropropane. cis-,3-dichloropropene 5. Toluene. trans -,3-Dichloropropene 7.,,-Trichloroethane 8. Tetrachloroethene 9.,-Dibromoethane. Chlorobenzene. Ethylbenzene. m,p -Xylene 3 o-xylene. Styrene 5.,,,-Tetrachloroethane. -Ethyltoluene 7.,3,5-Trimethylbenzene * Chloropropene was not present in the standard mix used; the label indicates where it would be expected to elute. Figure. Chromatogram obtained for a sample equivalent to L of a ppbv standard, with the 3 target compounds listed in HJ labeled. The trichloromethane peak is expanded, showing excellent peak shape. 3 8.,,-Trimethylbenzene 9.,3-Dichlorobenzene 3.,-Dichlorobenzene 3. Benzyl chloride 3.,-Dichlorobenzene 33.,,-Trichlorobenzene 3. Hexachlorobutadiene

Table. Linearity, detection limit, and repeatability data for the target VOCs. Note that chloropropene (#3) was not present in the standard mix used, and that chlorobenzene (#) is obscured by the internal standard chlorobenzene-d 5, so quantitation was not possible. Analyte concentration (ppbv) Quant ion (m/z) Run Run Run 3 Run Run 5 Run Run 7 No. Compound t R (min) R Mean conc. (ppbv) Std. dev. (ppbv) Peak area RSD (%) LOD (ppbv),-dichloroethene.3.995..38.95.93.8.7.5...8.7,,-trichloro-,,- trifluoroethane..998 5.9.7....3.8....3 Dichloromethane 3.3.998 9.33.377.5.8.3.3.39..3 7.5.95 5,-Dichloroethane 3.88.9979 3.79..535.58.98.8.3.85.3 7.5. cis-,-dichloroethene.5.998.38.37..3.387.3.3.38.3 5.9.7 7 Trichloromethane.9.988 85..3.59.53.53.8.7.9.8 5.7.88 8,,-Trichloroethane 5.5.98 97.3.7.95.3.93.7..8. 5..9 9 Tetrachloromethane 5..98 7.3.3.359.35.3.39.3.33.8 5..5,-Dichloroethane 5.3.997 78.53.5.58.58.5.5.97.5.3.3. Benzene 5..998.53.5.588.58.5.53.99.53.3.3.8 Trichloroethene.7.99 3.5.58.578.57.53.59.53.5.3 5..95 3,-Dichloropropane.7.995 3.53.55...3.57.53.578.3.. cis-,3-dichloropropene.97.99 75...9.8.77..7.5.3.9.98 5 Toluene 7.7.99 9.53.55.575.557.5.5.7.58.3.8. trans-,3-dichloropropene 7.5.995 75.7..5.9.5.59.3.77.9..9 7,,-Trichloroethane 7..9959 97.53.53.583.578.57.535.5.58.7.9.85 8 Tetrachloroethene 7.77.99..7.99.8.73.9.7.5.3 5.. 9,-Dibromoethane 8..998 7.8.3.77.7.3.8.3.8..8.7 Ethylbenzene 8.9.99.537.53.58.5.55.5.87.537.3.. m,p-xylene 8.8.979.95.9.7.735.7.58.9.9.8.8.5 3 o-xylene 9.7.988.9.95.5.53.5.83.5.5.3..9 Styrene 9.9.988.89.95.53.53.5.8.5.5.3..98 5,,,-Tetrachloroethane 9.8.995 85.573.559....57.55.585.9.9.9 -Ethyltoluene.3.988.5.57.57.83.73.9.5.59.3 7..7 7,3,5-Trimethylbenzene..9799.5.8.5.9.75.3.5.3.3 7.8.3 8,,-Trimethylbenzene.5.985...8.5..7.383..3 7.. 9,3-Dichlorobenzene.7.989.7.55.739.98.9.5..75.39 5.8. 3,-Dichlorobenzene.78.978.7.3.733.77.9.9..77.37 5.5.7 3 Benzyl chloride.93.995 9.38.35.3.38.385.3.3.38. 7..83 3,-Dichlorobenzene..98.3.35.9.7.59..58.3.3 5.. 33,,-Trichlorobenzene.7.975 8.553.55.7..58.553.57.5.33 5.8. 3 Hexachlorobutadiene.89.959 5...57.7.37.9..8..9.

Repeatability and limits of detection Seven repeat analyses of a.5 ppbv gas standard ( L equivalent) were conducted, and showed excellent consistency of retention times (Figure ). Relative standard deviations (RSDs) of repsonses for all target compounds were below 8% (Table ), and limits of detection (LODs) were below.5 ppbv. Quantitative re-collection of split flows Markes TD instruments have the ability to re-collect samples by directing the split flow (which would otherwise carry excess sample to vent) onto a sorbent tube. This process of re-collection can either be onto the tube from which the sample was originally desorbed, or onto a clean tube. This process can be fully automated, and offers a significant advantage over solvent extraction, because it allows a single sample to be analyzed multiple times. Sample splitting and re-collection makes method validation easier, and more importantly, avoids the need to collect another sample should the analysis unexpectedly fail to proceed correctly. 7 Run 7 5 Run Abundance Run 5 3 Run Run 3 Run Run 3 5 7 8 9 3 Figure. Seven repeated analyses of the same.5 ppbv standard. The baseline of each chromatogram is offset by for clarity. 5

Figure 5 shows a repeat analysis of a gas standard equivalent to L at ppbv with a : split flow, illustrating quantitative re-collection over a wide range of volatilities. Note the consistency of peak position and shape. The accuracy of the analysis was also validated by loading a sorbent tube with ng benzene and ng toluene, and performing re-collection/analysis cycles with a split ratio of.5: after the initial analysis. This showed excellent consistency with the theoretical values (Figure ). 8 Abundance Original sample Re-collected sample Ratio of response (re-collection/original) Benzene Toluene Theoretical decay curve 8 8 Figure 5. Original analysis (black) and repeat analysis (red) of tubes containing analytes equivalent to L of a ppbv standard with a : outlet split. 3 5 7 8 9 Run number Figure. Evaluation of the decay in response for benzene and toluene from a tube repeatedly analyzed with a.5: outlet split.

Real air samples To evaluate the performance of the system for a real-world scenario, L of ambient air was collected onto a sorbent tube in an office, a laboratory, and semirural outside air, followed by analysis under the conditions described earlier. The results clearly show raised levels of VOCs in the office air (Figure 7). 5 3 Abundance 3 3 5 3 7 8 8 7 9 9 Office air 5 Laboratory air Semirural outside air 5 5 3 35. Methanol. -Methylbutane 3. Ethanol. Acetone 5. Isopropanol. -Methylpentane 7. 3-Methylpentane 8. Hexane 9. Ethyl acetate. -Methylhexane. Cyclohexane. Trimethylhexane 3. Heptane. Acetic acid 5. Butan--ol. Toluene 7. Hexanal 8. m- and p -Xylene 9. o-xylene. α-pinene. Cyclohexanone. α-myrcene 3. Limonene. Phenol 5. Menthol. -Phenoxyethanol Figure 7. Analysis of three L air samples collected at three different locations using sorbent tubes, and analyzed by TD-GC/MS. Compounds listed in HJ are marked in bold type. 7

Conclusions This application note demonstrates that Markes International s automated, cryogen-free thermal desorption systems offer excellent results for monitoring air toxics in ambient air in accordance with Chinese EPA Method HJ. A particular feature of this method is the automated quantitative re-collection of split flows offered by the TD instrument, which makes method development and validation of results straightforward. Trademarks ACTI-VOC, CSLR, and TD-xr are trademarks of Markes International. DB- is a trademark of Agilent Technologies. Applications were performed under the stated analytical conditions. Operation under different conditions, or with incompatible sample matrices, may impact the performance shown. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 7 Printed in the USA July 7, 7 599-87EN