Long Path Industrial OCT High-precision Measurement and Refractive Index Estimation

Similar documents
Interference- Michelson Interferometer. Interference lecture by Dr. T.Vishwam

Measurements in Optics for Civil Engineers

Michelson Interferometer. crucial role in Einstein s development of the Special Theory of Relativity.

Fiber optic Sensors for Measurement ofrefractive Index of Liquids by Using Fresnel Ratio Meter

1 N star coupler as a distributed fiber-optic strain sensor in a white-light interferometer

Scattering properties of the retina and the choroids determined from OCT-A-scans

Gratings in Electrooptic Polymer Devices

10. OPTICAL COHERENCE TOMOGRAPHY

POLARISATION-SENSITIVE OPTICAL COHERENCE TOMOGRAPHY FOR MATERIAL CHARACTERISATION AND TESTING

arxiv: v1 [quant-ph] 18 Sep 2017

Advanced Optical Coherence Tomography techniques: novel and fast imaging tools for non-destructive testing

Physics 208 Exam 1 Oct. 3, 2007

iprom Optical Interferometry Prof. Dr. -Ing. Rainer Tutsch Institut für Produktionsmesstechnik IPROM Technische Universität Braunschweig

Experiment and Simulation Study on A New Structure of Full Optical Fiber Current Sensor

Near-Infrared Spectroscopy of Nitride Heterostructures EMILY FINAN ADVISOR: DR. OANA MALIS PURDUE UNIVERSITY REU PROGRAM AUGUST 2, 2012

Ambiguity of optical coherence tomography measurements due to rough surface scattering

Double-Slit Interference

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

Simultaneous Temperature and Strain Sensing for Cryogenic Applications Using Dual-Wavelength Fiber Bragg Gratings

Some Topics in Optics

PRINCIPLES OF PHYSICAL OPTICS

Optics, Light and Lasers

A Guide to Experiments in Quantum Optics

Resolution improvement in optical coherence tomography with segmented spectrum management

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a).

High-Resolution Imagers

GRATING CLASSIFICATION

Modern Physics Laboratory MP2 Blackbody Radiation

Low-coherence heterodyne photon correlation spectroscopy

Evaluation of Second Order Nonlinearity in Periodically Poled

The IPIE Adaptive Optical System Application For LEO Observations

Simple and Low Cost Polarimeter System

Supplemental material for Bound electron nonlinearity beyond the ionization threshold

Michelson Interferometer

A Portable Optical DSPI System for Residual Stresses Measurement by Hole Drilling Using the Integral Method in Terms of Displacement

CHM 424 EXAM 2 - COVER PAGE FALL

Lecture 19 Optical MEMS (1)

Rapid yet accurate measurement of mass diffusion coefficients by phase shifting interferometer

Single Emitter Detection with Fluorescence and Extinction Spectroscopy

Single-beam optical fiber trap

Astronomy 203 practice final examination

Optics.

Instructor: Welcome to. Phys 774: Principles of Spectroscopy. Fall How can we produce EM waves? Spectrum of Electromagnetic Radiation and Light

Experiment 4. Michelson Interferometer. 4.1 Introduction. References

LECTURE 23: LIGHT. Propagation of Light Huygen s Principle

Recent progress in SR interferometer

Introduction. Procedure. In this experiment, you'll use the interferometer to EQUIPMENT NEEDED: Lens 18mm FL. Component holder.

An alternative method to specify the degree of resonator stability

The 10th International Workshop on Accelerator Alignment, KEK, Tsukuba, February 2008 ALIGNMENT OF S-LSR

More Optical Telescopes

Designs of One-Element Refracting System for Gaussian and Annular-Gaussian Beams Transformations

EUV Reflectivity measurements on Acktar Sample Magic Black

Low Coherence Vibration Insensitive Fizeau Interferometer

Visualizing the bi-directional electron transfer in a Schottky junction consisted of single CdS nanoparticles and a planar gold film

Density Field Measurement by Digital Laser Speckle Photography

LECTURE 23: LIGHT. Propagation of Light Huygen s Principle

Blackbody Radiation EX-9920 ScienceWorkshop Page 1 of 8. Blackbody Radiation

The Michelson Interferometer

Comparison of different composite Nd:YAG rods thermal properties under diode pumping

Liquid refractometer based on immersion diffractometry

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

Concave mirrors. Which of the following ray tracings is correct? A: only 1 B: only 2 C: only 3 D: all E: 2& 3

Visualization of Convective Structures by Methods of the Hilbert Optics and Phase-Shift Interferometry

Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown and Twiss Photon Antibunching Setup

Quantum Optical Coherence Tomography

LECTURE 11 ELECTROMAGNETIC WAVES & POLARIZATION. Instructor: Kazumi Tolich

1 cm b. 4.4 mm c. 2.2 cm d. 4.4 cm v

A tunable corner-pumped Nd:YAG/YAG composite slab CW laser

LECTURE 11 ELECTROMAGNETIC WAVES & POLARIZATION. Instructor: Kazumi Tolich

Ultra-narrow-band tunable laserline notch filter

AST 101 Intro to Astronomy: Stars & Galaxies

Novel metrology for measuring spectral purity of KrF lasers for deep UV lithography

Hiromitsu TOMIZAWA XFEL Division /SPring-8

Wavelength switchable flat-top all-fiber comb filter based on a double-loop Mach-Zehnder interferometer

Chapter 6 Lecture. The Cosmic Perspective. Telescopes Portals of Discovery Pearson Education, Inc.

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Resonantly Pumped Er:YAG and Er:YAP Lasers

Chapter 6 Lecture. The Cosmic Perspective Seventh Edition. Telescopes Portals of Discovery Pearson Education, Inc.

Jitter measurement by electro-optical sampling

A NEW AUTOMATED MEASURING INSTRUMENT FOR MINUTE PHOTOELASTICITY

Detection of Single Photon Emission by Hanbury-Brown Twiss Interferometry

Chapter 6 Telescopes: Portals of Discovery. Agenda. How does your eye form an image? Refraction. Example: Refraction at Sunset

Propagation losses in optical fibers

Experiment 3 1. The Michelson Interferometer and the He- Ne Laser Physics 2150 Experiment No. 3 University of Colorado

Speed of Light in Glass

Light - electromagnetic radiation

Plasma Formation and Self-focusing in Continuum Generation

PS210 - Optical Techniques. Section VI

OPTI 511L Fall A. Demonstrate frequency doubling of a YAG laser (1064 nm -> 532 nm).

n(λ) = c/v(λ). Figure 1: Dispersion curves for some common optical glass types.

Introduction to Microeletromechanical Systems (MEMS) Lecture 9 Topics. MEMS Overview

Calibrating the Thermal Camera

A novel scheme for measuring the relative phase difference between S and P polarization in optically denser medium

Experimental study of nonlinear laser-beam Thomson scattering

Refractive Index Measurement by Gain- or Loss-Induced Resonance

Wataru Ohnishi a) Student Member, Hiroshi Fujimoto Senior Member Koichi Sakata Member, Kazuhiro Suzuki Non-member Kazuaki Saiki Member.

Chapter 5. Telescopes. Copyright (c) The McGraw-Hill Companies, Inc. Permission required for reproduction or display.

Fizeau s Interference Experiment with Moving Water Contradicts the Special Theory of Relativity.

For more sample papers visit :

Sensing Rotation with Light: From Fiber Optic Gyroscope to Exceptional Points

Transcription:

Long Path Industrial OCT High-precision Measurement and Refractive Index Estimation Tatsuo Shiina Graduate School of Advanced Integration Science, Chiba University, 1-33 Yayoi-cho, Inage-ku, Chiba, Japan Keywords: Abstract: OCT (Optical Coherence Tomography), Industry, Long Path, High Accuracy, Refractive Index, Group Index. Long-path optical coherence tomography was developed for industrial use. The system is compact and easy variable to change the measurement speed and range. In this study, its precision and long-path were designed as 1μm and 100mm, respectively. Refractive index of water was analysed by changing the temperature. The results well coincided with the theoretical curve of group index of refraction. 1 INTRODUCTION Recently, needs of high-precision optical measurement devices are increased in the industrial field due to the technology development of outer shape measurement and in-vivo measurement of materials. 3 dimensional outer shape measurement based on optical measurement technology is essential to the industrial field. Recently non-contact optical probe takes the place of the contact type on the 3 dimensional measurement. Many of principles for it are proposed and commercialized. The traditional high-precision measurement technology is optical interference technology in industrial field. These technology installs laser and white-light source into it. Laser interferometer, laser displacement meter, and white-light interferometer are commercialized. In these high-precision optical measurement devices, long path measurement is included. Combinational lens such as telescopic lens is essential to evaluate and analyse their lenses matching to optimize their performance. In the case of crystal growth and material compounding operation, the feedbacks from the interior condition sensing to the temperature and concentration controls are important. On the other hand, the long path measurement on the laser and white-light interferometers utilizes linear stage, and they are lack of repeatability. Furthermore, these apparatuses are large and expensive. They have restriction to use. The optical coherence tomography : OCT, which is developed in medical field, is recently adapted to the industrial use. The OCT technology is the low coherent interferometer and obtains the cross-sectional image by non-invasive and nondestructive measurement, Mainly it is used in ophthalmology.(danielson 1991, Huang 1991, Brezinski 1999) The combination of super luminescent diode : SLD and optical fiber interferometer adds the flexibility of measurement to the device and also compactness. In this study, a portable OCT scanner has been developed for industrial use.(shiina 2003, 2009, 2014) In this report, to improve the repeatability on the long path measurement, we state the development of long path industrial OCT, which has the rotational optical path change mechanism. It can repeat the measurement, and also change the measurement range by adjusting the rotational radius of mechanism. We evaluate the accuracy and applied it to the refractive index measurement. We aimed the concrete accuracy of 1μm within the measurement range of 100mm to verify the measurement result with the five-figure accuracy. 2 LONG PATH INDUSTRIAL OCT The low coherence interferometer changes its reference path length, and interferes it with the sample path, where the same optical path length inside the specimen. Due to the optical path change, interior information of the specimen is visualized. 345 Shiina, T. Long Path Industrial OCT - High-precision Measurement and Refractive Index Estimation. DOI: 10.5220/0005842903430348 In Proceedings of the 4th International Conference on Photonics, Optics and Laser Technology (PHOTOPTICS 2016), pages 345-350 ISBN: 978-989-758-174-8 Copyright c 2016 by SCITEPRESS Science and Technology Publications, Lda. All rights reserved

OSENS 2016 - Special Session on Optical Sensors Therefore, it is important to scan precisely the optical path change. Our long path industrial OCT utilizes the rotational optical path change mechanism. The rotation radius and speed decides the measurement range and scan rate, respectively. This scanning mechanism consists of a rotating corner reflector and a fixed mirror. The optical path change is represented by equation (1). l All = l 1 + l 2 + (l 1 l 4 ) 2s = 2l 1 + l 2 (1 sin 2θ) 2s (r + s)(1 cosθ) l 1 = (r + s)sinθ tan(π /4+θ) l l 2 = 3 cos(π /4+θ) (r + s)(1 cosθ) l 3 = 2s+ sin(π /4+θ) l 4 = l 2 sin(2θ) (1) Figure 1 shows the geometrical arrangement of the mechanism with the optical path of l 1 l 4. θ is rotation angle [deg], r is rotation radius, s is the offset length from the optical axis. The fixed mirror reflects the thrown beam to the same path. The optical path change becomes the approximately linear motion. The optical path change of the rotation radius of 10mm is shown in Fig.2. The actual motion has the distortion from the linear motion. It becomes an ogive. The distortion is about 1 2% within the rotation angle of +/-20 deg. The long path industrial OCT has a rotation disk of 60mm radius, of which measurement range reaches 100mm.(Fig.3) A servo motor is installed. To stabilize the rotation, the rotation disk is balanced its weight. The motor has a rotary encoder to monitor the rotation jitter.the optical setup of the long path industrial OCT is illustrated in Fig.4. Optical Path Change [mm] Reflector Angle [deg] Figure 2: Optical path length. Figure 3: Optical path change mechanism. Figure 1: Optical path change by rotating reflector. Figure 4: Structure of long path industrial OCT. 346

Long Path Industrial OCT - High-precision Measurement and Refractive Index Estimation The interferometer consists of an optical fiber coupler. SLD beam (Anritsu Co. Ltd) is divided by the coupler, one goes to the reference path and the other goes to the measurement path, which has the optical probe to focus it to the specimen. Both of reflected beam are combined and cause the interference within the same coupler, and detected by the photodiode. 3 ACCURACY EVALUATION The SLD source of 800nm-band is installed into the long path industrial OCT. As the rotation radius of the reflector is 60mm, the measurement range reaches 100mm. Here it is restricted to 80mm by the reflector size. The rotation speed is 200rpm. The interference signal is detected as the Gaussian envelope through amplifiers and filter circuits. At first, the system was evaluated its accuracy. The sample is a mirror on the linear stage. By alternating the measurement length, the interference position was obtained. The linear stage (Mitsutoyo Co. Ltd) is 1m long and its accuracy was 5μm. Figure 5 shows the experimental result. The Z-phase signal of the rotary encoder is utilized as trigger and becomes the zero position through the measurement range of 80mm. The optical path change becomes approximately linear change. In the measurement, T1, T2, and T3 [ms] were obtained. T1 and T2 are rise-up and fall-down times of A-phase signal of the rotary encoder when the peak position as time T3 of the interference-envelop signal appeared. The rotation motor controls its rotation with 6- pole coils, and its rotation speed slightly fluctuates. Therefore, T1 T3 has a little bit fluctuations. It should be compensated. How to compensate it is explained in Fig.6. As T1 average of 10 times measurement is standard, T3 value is corrected, which is called T1 correct. As T2 average of 10 times measurement is standard, T3 value corrected, which is called T2 correct. Converting the measured interference signal time to the position, its average time is taken the place of the optical path length with equation (1). Each value of Fig.5 is corrected T3 average on each path length. The approximation line is slightly s-curve (ogive). In this study, 3 rd approximate curve is adapted on this corrected result. Standard variation of 10 times measurement at each path length is summarized in Fig.7. To search the peak position of the interference signal, moving average and center search program, which pursued due to the interference signal height, are used. T2 correct on the center search program minimized the standard variation as 1.43μm. The goal value of 1μm was not accomplished, while we decided it the extensional accuracy to progress the experiment. Optical Path [mm] Figure 5: Experimental result and 3 rd approximation line. RMS Deviation A-Phase signal Interference signal T1 correct T2 correct T1 Average T2 Average Figure 6: Jitter correction for reflector rotation. Moving Average Center Program Optical Path [mm] Figure 7: Jitter correction result at each measurement points. 347

OSENS 2016 - Special Session on Optical Sensors 4 REFRACTIVE INDEX MEASUREMET 4.1 Apparatus Thermometer Thermo Camera As the application of the long path industrial OCT, the refractive index measurement was conducted. The experimental set up and water tank measurement part are shown in Fig.8 and Fig.9, respectively. The measurement target is 5cm x 5cm water tank (small tank). 15cm x 15cm water tank (large tank) has a cooler terminal, and control the water temperature including the inner small tank. In the measurement, water temperature is lowered, and the refractive index, which depend on the temperature, was calculated by measuring the optical path change between the inside glass walls of the small tank. To stabilize the controlled temperature inside the small tank, a stirrer rotates the large tank water slowly. The temperature distribution of the small tank was monitored by a thermo camera. The OCT measurement probe was set to enter the small tank within the measurement range. The interference signals of the small tank were obtained at four positions from its glass walls (each side of the walls). Figure 10 shows the interference signals. The refractive index was calculated by the optical path length between the inner water-sides of the small tank walls. The temperature was controlled from the 25 to 2 degrees at the step of 0.5 degrees. T2 correct with the center search program was adapted into the refractive index calculation to fix the maximum interference signal and to compensate the rotation jitter. As the concrete calculation, 3 rd approximation curve on Fig.5 was utilized, that is, the refractive index was estimated by changing the 3 rd approximation curve to the linear equation. Here, the 3 rd approximation curve is represented as follows, y = 0.00010456x 3 + 0.0035485x 2 (2) + 2.0526x +13.845 while the linear change equation obtained by the whole measurement range is expressed as follows. y = 2.0636 +14.023 (3) To change the equation (2) to the equation (3) is conducted by calculating the difference between them, which is shown in Fig.11. The optical path length of 12mm became the center and the distortion of ogive from the linear line was balanced. The difference was changed as the distortion value to the linear one. SLD Figure 8: Water refractive index measurement by longpath OCT. OCT Probe Mirror Stirrer Figure 9: Measurement of water refractive index. Inner Signal1 Outer Signal1 Inner Signal1 Water Tank Figure 10: Interference Signals on long-path OCT. 4.2 Group Index Estimation Thermo Controller Inner Signal2 Inner Signal2 Cooler Outer Signal2 The refractive index depends on material density, temperature, and incident wavelength. Absolute refractive index equation shown as equation (4) is a regression formula due to the above parameters based on Lorentz-Lorentz equation. 348

Long Path Industrial OCT - High-precision Measurement and Refractive Index Estimation n 2 +1 n 2 + 2 1 D = a 0 + a 1 D + a 2 T + a 3 λ 2 T + a (4) 4 λ + a 5 2 λ 2 2 λ + a 6 UV λ 2 2 λ + a 7 D2 IR D = D / D 0, T = T / T 0, λ = λ / λ 0 where n is the absolute refractive index of pure water, D is density scale represented by the ratio between the pure water density D and the standard density D 0 [kg/m 3 ], T is temperature scale represented by the ratio between the pure water temperature T [K] and the standard temperature T 0 (=273.15K). λ is wavelength scale represented by the ratio between the wavelength in vacuum λ and the standard wavelength λ 0 (=0.589μm). a 0 a 7 are optimized coefficients and λ UV and λ IR are UV / IR resonances. [7] The OCT light source (here, SLD light source) has wide spectrum. It disperses in a material, and difference of speed (group index) due to the refractive index occurs. That is, the refractive index estimated by the OCT system becomes group index of refraction n g. It is expressed as equation (5). n g = n(λ) λ dn(λ) (5) dλ Figure 12 shows the absolute index calculated by the equation (4) and the group index calculated by the equation (5) against the wavelength of 859.681nm, which is same as the experiment. The experimental results were compared with this group index. The estimated experimental result is shown in Fig.13. The measurement was conducted by lowering the temperature from the room temperature to 2 degrees. In the figure, the results are represented as average and the center search program with 10 times measurement. Both of the estimations well matched with the theoretical value of group index. The maximum errors from the theoretical curve of the average and the program were 0.00070 and 0.00057, respectively. Both of the estimations get the five-figured accuracy. The maximum error occurred on the longest path length. It is caused by the 3 rd approximation curve we used. RMS Deviation [μm] Figure 11: debiation between 3rd approximation and linear motion. Refractive Index Figure 12: Absolute refractive index and calculated group refractive index. Group Refractive Index Figure 13: Group refractive index of experiment and theoretical values. 5 SUMMARY Angle [deg] Temperature [deg] Temperature [deg] Group Normal Average Center Program In this study, we set our goal to the high-precision of 1μm in the measurement range of long path length of 100mm. It means five-figured accuracy. As a result, the measurement accuracy achieved 1.43μm as standard variation. The main reason not to get less than 1μm accuracy is indetermination of the linear stage we used (5μm). The rotation motor has the rotation jitter of 0.00036 degrees as standard variation. It is equal to 0.74μm of the fluctuation of 349

OSENS 2016 - Special Session on Optical Sensors the optical path difference. It means that the experimental result has room for improvement. At the next step, we prepare the high-precision linear stage to verify the high accuracy of our long path system. In this report, the refractive index of pure water was estimated successfully with the five-figured accuracy. This application is not only to the pure water, but also to the mixed liquid solution to confirm the mixing ratio and the concentration. By scanning the OCT probe against the optical axis, the concentration distribution and its fluctuation of the liquid solution and temperature distribution can be evaluated. The measurement range can be enlarged by magnifying the rotation radius. The rotation disk, however, is difficult to rotate stably. In this study, plural reference paths make it possible to expand the measurement range with the same accuracy with this experiment.(harvey 1998) We seek the improvement of the measurement range and its new applications. REFERENCES Danielson B. L. and Boisrobert C. Y. 1991, Absolute optical ranging using low coherence interferometry, App. Opt. Vol. 30, No.21 pp.2975-2979. Huang D., Swanson E. A., Lin C. P., Schuman J. S., Stinson W. G., Chang W., Hee M. R., Flotte T., Gregory K., Puliafito C. A., J. G. Fujimoto, 1991, Optical Coherence Tomography, Science, Vol. 254, pp.1178-1181. Brezinski M. E. and Fujimoto J. G., 1999, Optical Coherence Tomography: High-Resolution Imaging in Nontransparent Tissue, IEEE J. Quant. Electron., Vol. 5, No. 4, pp.1185-1192. Shiina T., Moritani Y., Ito M., and Okamura Y., 2003, "Long optical path scanning mechanism for optical coherence tomography", Applied Optics, Vol.42, No. 19, pp.3795-3799. Shiina T., Miyazaki H., and Honda T., 2009, Factory built-in type simplified OCT system for industrial application, IEEJ C Vol.129, No.7, pp.1276-1281, (Japanese). Shiina T., 2014, Optical Coherence Tomography for industrial application Handbook of Optical Metrology 2 nd Edition, CRC Press. Harvey A. H., Gallagher J. S., and J. M. H. L., 1998, Revised Formulation for the Refractive Index of Water and Steam as a Function of Wavelength, Temperature and Density, J. Phys. Chem. Ref. Data, Vol.27, pp.761-774. Shiina T., PCT/JP2010/070844. 350