Verification for Non-Automatic Weighing Instruments (NAWI)

Similar documents
9.1 Compatibility of modules for nonautomatic weighing instruments (NAWIs) Common requirements

After publication, this Indian Standard will supersede the following Indian Standards:

OIML R 76-2 RECOMMENDATION. Edition 2007 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Non-automatic weighing instruments

Non-automatic weighing instrument. Automatic weighing instrument. Weighing instrument

INTERNATIONAL OIML R 61-2 RECOMMENDATION

OIML R 51-2 RECOMMENDATION. Edition 2006 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Automatic catchweighing instruments

Load Cell Approval Field and Laboratory Evaluation Manual

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

+ + ( indicator ) + (display, printer, memory)

OIML R 50-3 RECOMMENDATION. Edition 2014 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

A combinatorial technique for weighbridge verification

THE DETERMINATION OF GRAVITY ZONE FOR WEIGHING INSTRUMENTS IN SLOVENIA

OIML R RECOMMENDATION. Edition 2006 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Instruments for measuring the areas of leathers

ORGANISATION INTERNATIONALE DE MÉTROLOGIE LÉGALE INTERNATIONAL RECOMMENDATION

Guidelines on the Calibration of Non-Automatic Weighing Instruments

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

Accuracy Calibration Certificate

ACCEPTANCE PROCEDURE FOR ROMET ROTARY METER BODIES AND/OR MECHANICAL NON-CONVERTING AND CONVERTING MODULES

Measuring Instruments Directive (MID) MID/EN14154 Short Overview

INTERNATIONAL JOURNAL OF ADVANCED RESEARCH IN ENGINEERING AND TECHNOLOGY (IJARET)

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

INTERNATIONAL OIML R 15 RECOMMENDATION

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

Understanding Weighing A Grocer s Guide to Metrology

Type examination and testing of 825 non-automatic weighing indicator according to EN Performed for Cardinal Detecto Scale Manufacturing Company

Data sheet FIXED CHIP RESISTORS; RECTANGULAR TYPE & WIDE TERMINATION. AEC-Q200 qualified. RoHS COMPLIANCE ITEM Halogen and Antimony Free

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

E C t y p e - e x a m i n a t i o n C e r t i f i c a t e

Data sheet FIXED THICK FILM CHIP RESISTORS; RECTANGULAR TYPE AND HIGH POWER ANTI SURGE. AEC-Q200 qualified

Non-destructive testing of steel forgings- Part 1: Magnetic particle inspection (BS EN :1999)

Standard burettes for verification officers

Data sheet FIXED THICK FILM CHIP RESISTORS; RECTANGULAR TYPE ANDULTRAHIGH VOLTAGE. AEC-Q200 qualified. RoHS COMPLIANCE ITEM Halogen and Antimony Free

Operating Instructions for Digital Manometer. Model: MAN-SD

DECREE No. 44. of the Ministry of Economy of the Slovak Republic. dated 19 February 1999,

PRELIMINARY DRAFT INDIAN STANDARD METHOD FOR CALIBRATION AND CLASSIFICATION OF TORQUE MEASURING DEVICES INCLUDING TORQUE WRENCH TESTER

3.0 INDICATION AND REGISTRATION REPRESENTATION Already exist in Pub 14

E/ECE/324/Rev.2/Add.127/Amend.7 E/ECE/TRANS/505/Rev.2/Add.127/Amend.7

Determination of particle size distribution Single particle light interaction methods. Part 4:

ISO defines the Rockwell superficial hardness value, HRN, as:

Data sheet FIXED THICK FILM CHIP RESISTORS; RECTANGULAR TYPE RMC1/32,1/20,1/16S,1/16,1/10,1/8,1/4,1/2,1. AEC-Q200 qualified (Without RMC1/32)

E v a l u a t i o n C e r t i f i c a t e

ME6504 METROLOGY AND MEASUREMENTS

White Paper. Moisture Analyzer Routine Testing

Issued August DATA SHEET. 3VT5 MCCB up to 1600 A

Electrical Double Layer Energy Storage Capacitors Up to 3 V Operating Voltage

Carbon Film Resistors, Standard

Ruggedized Electrical Double Layer Energy Storage Capacitors

Chap 4. Software Reliability

Calibration of Volumetric Glassware. Prepared by Allan Fraser May 2016 APPLICATION Note 1

Manual Railway Industry Substance List. Version: March 2011

DeFelsko PosiTector Dew Point Meter with Anemometer (DPMA)

DROP WEIGHT TEST WORK

Level Sensor PanCake PR 6251 and Mounting Kit PR 6051

Technical Procedure for General Laboratory Equipment

LAB 15. Traceability: Volumetric Apparatus. Edition 3 February 2019 Draft for Consultation. LAB 15 / Issue: 3 Page 1 of 12

CAPACITORS, FIXED, TUBULAR, POROUS TANTALUM CATHODE AND ANODE, GELLED ELECTROLYTE, HERMETICALLY SEALED BASED ON TYPE ST79

Precision and Accuracy Assessing your Calorimeter

FIS Specifications for Flex Poles (Edition May 2008) Original Text: German

Buckling Analysis Of Straight Helical Compression Springs Made Of ASTM A229 Gr-II, ASTM A 313 Materials (Type 304 & 316).

Precision and Accuracy Assessing your Calorimeter

Calibration of Analog Scales Based on. a Metrological Comparison Between. the SIM Guide and OIML R 76-1

National Type Evaluation Program (NTEP) Weighing Sector Meeting Agenda

Woodford County Erosion Prevention Plan and Permit. Application #

Control Standard for Handling Chemical Substances in Products, Parts and Materials

Revision of GRE/2017/05

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS

Test report CVN Project number Page 2 of 2

MATHEMATICAL SKILLS MOMENTS OF FORCE (RATIOS) AND EQUILIBRIUM AND ASSOCIATED EXAMINATION QUESTIONS.

3VT4 Molded Case Circuit Breakers up to 1000 A

RESISTOR, FIXED, SURFACE MOUNT, THIN FILM, NON-HERMETICALLY SEALED BASED ON TYPE TNPS. ESCC Detail Specification No. 4001/029

A compression load cell, with strain gauges, tested as a part of a weighing instrument. Manufacturer : Molen or Precia Molen.

Commenting template CEN/TC 102 Date : Document: pren 13060

WELDED STEEL PIPE NIPPLE PRICE SHEET

So far we discussed random number generators that need to have the maximum length period.

Solid Tantalum SMD Capacitors TANTAMOUNT, Hi-Rel COTS, Low ESR, Metal Case

WELDED STEEL PIPE NIPPLE PRICE SHEET

INDIANA DEPARTMENT OF TRANSPORTATION OFFICE OF MATERIALS MANAGEMENT. VERIFYING THERMOMETERS ITM No T

ISCC Audit Procedure Classified chemicals

STATIC & DYNAMIC CHARACTERISTICS OF MEASUREMENT SYSTEM

AGGREGATED RESULTS ONGAS PRICES FOR 2010 FOR CATEGORIES OF CUSTOMERS IN GAUTENG, FREE STATE, KWAZULU-NATAL AND MPUMALANGA PROVINCES.

4. How to prove a problem is NPC

TEST REPORT IEC and/or EN Photobiological safety of lamps and lamp systems

How Small Can a Launch Vehicle Be?

LED Retrofit. Fast-track proposal and feedback. W. Schlager Karlsruhe, 23-July-2015

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 1

Article for the 28 th Sensing Forum

CONCENTRATED VOLUMETRIC & VOLUMETRIC SOLUTIONS

Chip tantalum capacitors

DEVELOPMENT, CONSTRUCTION, DELIVERY AND WARRANTY OF A DEVICE ABLE TO DETERMINE BOUNCE PARAMETERS OF A TABLE TENNIS BALL ON A TABLE SURFACE

Chapter 4. Inequalities

EU DIRECTIVES PREPACKAGES AND IMPLEMENTATIONS IN CROATIAN NATIONAL LEGISLATION 76/211/EEC 2007/45/EC 75/107/EEC

Issued August DATA SHEET. 3VT3 MCCB up to 630 A

Solid-Electrolyte TANTALEX Capacitors, Hermetically Sealed, Axial-Lead, CECC Approved

Infrared Thermometer Calibration A Complete Guide

T e c h n i c a l L e x i c o n

Precision Thin Film Chip Resistor Array

610,- KERN ADB. Analytical balances. Quick-Finder Analytical balances. Large glass draught shield with 3 sliding doors

Calibration Traceability Guidelines

NEW VSL Conditioning plate!

Transcription:

Verification for Non-Automatic Weighing Instruments (NAWI) Every instrument should be accurate and correct to be permitted for use in any commercial transaction, particularly at the Point Of Sale. An instrument may be accurate but may not be correct. To be accurate, the instrument must meet the applicable tolerance limit () and some metrological requirements. It should pass some specific performance tests. To be correct, an instrument, should not only be accurate, it should meet some specific legal requirements also - type (design) or model approval, appropriate class for usage, descriptive markings, proper maintenance etc. 1. Definitions. 1.1 Non-Automatic Weighing Instrument (NAWI) Instrument that requires the intervention of an operator during the weighing process to decide that the weighing result is acceptable. 1.2 Verification scale interval, e Value, expressed in units of mass, used for the classification and verification of an instrument. 1.3 Actual scale interval, d Value, expressed in units of mass of: the difference between the values corresponding to two consecutive scale marks, for analog indication; or the difference between two consecutive indicated values, for digital indication. 1.4 Number of verification scale intervals, n Quotient of the maximum capacity and the verification scale interval: n = Max / e 1.5 Maximum capacity (Max) Maximum weighing capacity, not taking into account the additive tare capacity. 1.6 Minimum capacity (Min) Value of the load below which the weighing results may be subject to an excessive relative error 1.7 Maximum permissible error () Maximum difference, positive or negative, allowed by regulation between the indication of an instrument and the corresponding true value, as determined by reference standard masses or standard weights, with the instrument being at zero at no-load, in the reference position 1

2. Classification of NAWI All NAWI's are classified into four categories: Class I: Special accuracy - ultra micro, micro, semi micro, macro Class II: High accuracy - precision balances Class III: Medium accuracy - NAWI's for trade use Class IIII: Ordinary accuracy - For low accuracy trade use 2.1 Identification of NAWI classes Identification of NAWI classes is based on the following table. Table 1 Class e value in g Verification Scale Interval n = max/e min capacity (g) I 0.001 or above 50 000 100e II 0.001 to 0.05 0.1 or above 100 to 100 000 5 000 to 100 000 20e 50e III 0.1 to 2 5 or above 100 to 10 000 500 to 10 000 20e 20e IV 5 or above 100 to 1000 10e Check if declared, e value matches with the accuracy class Find n = max/e and check whether it matches with the accuracy class Check whether min value matches with the accuracy class Some instruments do not declare minimum capacity. Calculate the minimum capacity from the table and check whether the instrument can measure below that quantity 3. Metrological requirements An instrument must be allowed to warm up before commencing test. All NAWIs should pass the test under normal conditions. Once an instrument fails in a test, it is no use carrying out any further test. Accuracy classes Verification scale interval, e Actual scale interval, d Sealing/stamping mechanism 2

4. Testing Procedures of NAWI other than Weighbridge. 4.1 Visual examination All NAWI shall have a permanently fixed/engrooved descriptive plate with marking: Manufacture`s mark and model Accuracy class I, II, III, IIII Verification scale interval, e scale interval, d (when d < e) maximum capacity (max) minimum capacity (min) NOTE: Scale Class I, II and III, shall be Verified by Standard mass of class E, F and M respectively. 4.2 Determination of maximum permissible error () In every class of digital scale has which corresponds to verification intervals of the scale (e). See the graphs below. Class I 1.5e 1e 0.5e Class II 0 50000 200000 >200000 n 1.5e 1e 0.5e 0 5000 20000 100000 n 3

Class III 1.5e 1e 0.5e 0 500 2000 10000 n Class IIII 1.5e 1e 0.5e 0 50 200 1000 n The above graphs are simplified in Table 2 below: Table 2: for initial verification. Class I Class II Class III Class IIII ± Load in e min to 50 000 min to 5 000 min to 500 min to 50 0.5e Load in e 50 001 to 200 000 5 001 to 20 000 501 to 2 000 51 to 200 1e Load in e above 200 000 20 001 to 100 000 2 001 to 10 000 201 to 1 000 1.5e NOTE: For subsequent verification, is twice of initial From the minimum and maximum value obtained for particular scale, this will show where the scale falls between boundaries of, eg. 0.5e, 1e, 1.5e. 4

Example: Class III, Max Capacity 8000g, e=1g Minimum Capacity = n min e Maximum Capacity = n max e When Minimum capacity is not given, refer to Table 1 above for class III which is 20e. n min = Minimum number of verification scale intervals n max = Maximum number of verification scale intervals n min = Min. Capacity e.. n min = 20e = 20 intervals. e n max = Max. Capacity e = 8000 1 = 8000 intervals From the graph of class III, our weighing instrument 8000g capacity lies between three boundaries as follows. (i) 20 500 (boundary i) (ii) 501 2000 (boundary ii) (iii)2001 8000 (boundary iii) So the for these 3 boundaries are as follows. Boundary i: = 0.5e = 0.5(1) = ± 0.5g Boundary ii: = 1e = 1(1) = ± 1g Boundary iii: = 1.5e = 1.5(1) = ± 1.5g Convert the number of scale intervals into Mass (grams) by n e... n 1 e = 20 1 = 20g n 2 e = 500 1 = 500g n 3 e = 2000 1 = 2000g n 4 e = 8000 1 = 8000g 5

Graph Class III 1.5 e 1 e 0.5e i ii iii 0 20 500 2000 8000 10000 n 0.5g 1 g 1.5 g 20 g 500 g 2000 g 8000 g Range intervals and s shown in table 3, Table 3 Range (g) 20g to 500g ±0.5 >500g to 2000g ±1.0 >2000g to 8000g ±1.5 4.3 Weighing Performance Test On carrying out performance test, the instrument must be allowed to warm up, and then loaded with a known mass not exceeding maximum capacity at least three times. The Instrument is tested by loading the mass in the following sequence (i) Loading in ascending order up to maximum capacity (ii) Unloading in descending order to zero. 4.3.1 Procedures: (a) Select at least five different standard mass : Minimum capacity of instrument. 20% of the maximum capacity of the instrument. 40% of the maximum capacity of the instrument. 60% of the maximum capacity of the instrument. 80% of the maximum capacity of the instrument. 100% of the maximum capacity of the instrument. NOTE: of the instrument must be known according the class of weighing instrument and standard mass used. 6

(b) Determine E o (error calculated at/or near zero) by putting the smallest mass that can be detected. (c) Put the selected mass as stipulated in part (a) above in turn on the instrument and note the indications (The same procedure repeated backwards). On every indication (I) observed above, add smallest mass (ΔL) that will change the readings. From the above procedure error is calculated by the following formula. E = I + 1/2 e ΔL L Now error corrected. E c = E E 0 with E 0 = error calculated at or near zero All results should be recorded in the following table, Arrow indications and means forward and backwards respectively. Table 4 Load, L Indication, I Add Load ΔL Error, E Corrected error, E c In every test if Ec mpe, the instrument Passed Remarks: 4. Eccentricity Test The aim of the test, regardless position of load on instrument, the instrument must indicate actual value of load within. Procedures Determination of Load receptor. Check whether the load receptor of instrument has : Four or less points of support. More than four points of supports. In most cases weighing scales other than Weighbridges have four or less points of supports. Instruments with four or less points of support a load corresponding to one-third of the maximum capacity shall be applied as follow: 7

Ο Loading Position When commencing test observe corresponding of ⅓ of the maximum capacity for particular instrument. (i) Set instrument at zero (ii) Apply ⅓ of maximum capacity in all marked load position in-turn. (iii)record the load (L) and indication (I) (iv) In all position add and record smallest mass (ΔL) that will be detected. (v) In every test, Calculate E by using formula E = I + ½ e ΔL L and Ec = E E0 (vi) Record the results above in Table 5 Table 5 location Load, L Indication, I Add Load ΔL Error, E Corrected error, E c 1 2 3 4 If Ec, the instrument Passed The indications for different positions of a load shall meet the s, when the instrument is tested. 5. Discrimination This test is done to check if the instrument is capable of sensing a small change in load. The actual scale interval for a class I or II digital instrument may be d and not e. As the majority of instruments to be verified will be class III or IIII where e = d, e is used in the procedure. If the instrument to be tested has d e then e becomes d in this instance for the whole procedure. Procedure Select three different standard mass as stipulated below: (i) Minimum capacity of the instrument (ii) Half capacity of the instrument (iii) Maximum capacity of instrument Start putting the selected standard mass on the instrument and note the indication (I). On every indication (I 1 ) observed above add 1.4d and note indication (I 2 ) where d is the actual scale interval which is the difference between the values corresponding to two consecutive indicated values. 8

In all three tests compute I 2 I 1 Results are recorded in Table 6 Table 6 Load, L Indication, I 1 Extra load = 1.4d Indication, I 2 I 2 -I 1 Finally, in every tests if I 2 I 1 d the instrument Passed 6. Repeatability This test is designed to check whether the instrument gives consistent reading for the same load when the load is placed on the same position on the receptor for a number of times. Procedures Repeatability test is done at half load of the maximum capacity of the instrument. NOTE: of the instrument must be known at the particular load. Start putting selected half load on the instrument for about 10 times and add smallest mass (ΔL) that will change the indications Calculate E by using the formula. E = I + ½e L ΔL Record results calculated above in Table 7. Table 7 Load Indication Add. Load ΔL E Finally in every tests: If E and E max E min the instrument Passed. 9

TEST FORM FOR NAWI (ELECTRONIC) OTHER THAN WEIGHBRIDGES Owner Name and Address. General Examination: Type Value of e Serial Number Model. Value of d.. Last Verification Sealing Mechanism (Provide/not Provide) Maximum Capacity Minimum Capacity.. Visual Examination: (a) (b) (c) Passed Failed 1. WEIGHING PERFOMANCE TEST. Table Load, L Indication, I Add Load ΔL Error, E Corrected error, E c Check if Ec the instrument Passed Passed Failed Remarks.. 10

2. ECCENTRICITY TEST Table location Load, L Indication, I Add Load ΔL Error, E Corrected error, E c 1 2 3 4 Check if Ec the instrument Passed Passed Failed Remarks.. 3. DISCRIMINATION TEST Table Load, L Indication, I 1 Extra load = 1.4d Indication, I 2 I 2 -I 1 Finally, in every tests if I 2 I 1 d the instrument Passed Passed Failed Remarks.. 11

Table 4. REPEATABILITY Load Indication Add. Load ΔL E Finally in every tests: If E and E max E min the instrument Passed. Passed Failed Remarks.. Final Remarks: Name of Inspector: Name of Owner/Agent:.. Signature:...... Signature: Date: Date:.. 12