CERTIFICATE OF ANALYSIS

Similar documents
Certified Reference Material

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS

Certified Reference Material

Certified Reference Material

Certification Report. Certified Reference Material. BAM-M383a, b, c. Pure Copper. January 2014 (revised March 2015)

Bundesanstalt für Materialforschung und -prüfung (BAM) in co-operation with the

Certified reference materials

Federal Institute for Materials Research and Testing. Certified Reference Materials

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Certificate of Analysis

BAM Federal Institute for Materials Research and Testing. in Co-operation with the

CERTIFICATE OF ANALYSIS

Certified Reference Materials

Certified reference materials

Federal Institute for Materials Research and Testing. Certified Reference Materials

Certified Reference Materials

ERM -BB501b Customer information

ANALYSIS CERTIFICATE OF

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

REFERENCE MATERIALS

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

CERTIFICATE OF ANALYSIS SARM 74 FERROCHROME METAL

PT-Providers: Providers: Key Partners of

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

CERTIFICATE OF. Single Analyte Custom Grade Solution. ICP Assay NIST SRM 3120a Lot Number: Calculated NIST SRM Lot Number: See Sec. 4.

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

CERTIFICATE OF ANALYSIS

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3165 Lot Number: EDTA NIST SRM 928 Lot Number: 928

Certificate of Analysis

CERTIFICATE OF. Single Analyte Custom Grade Solution g/ml (measured at 20 ± 4 ºC) Graviametric NIST SRM Lot Number: See Sec. 4.

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# Calculated See Sec. 4.

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Inspection Documents for Type 2.1 Coverage:

BATCH: MATERIAL. Description. water. Quantity QC METAL LL2. is prepared (V/V). Use. be used in the quality. nitric acid is. certificate. intended use.

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)

CERTIFIED REFERENCE MATERIAL RRM Au014

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3103a Lot Number:

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3122 Lot Number:

NPL, India activity related to water as per the need of the country

Certificate of Analysis

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) EDTA NIST SRM 928 Lot Number: 928

METALS IN WATER BATCH: MATERIAL. Description. Cu, Ni and. Quantity 5% (V/V). Use. the mark in. such a way. water. The are given in QC METAL HL2

Hiroshi Sawai TKG Laboratory Center International Standardisation Office

Locations. Branch Fabeckstraße Unter den Eichen Berlin. Branch Adlershof Richard-Willstätter-Straße Berlin

Certificate of Analysis

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples.

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3112a Lot Number: Cr3

Inspection Documents for Type 2.1 Coverage:

Certificate of Analysis

This document is a preview generated by EVS

Certificate of Analysis

Inspection Documents for Type 2.1 Coverage:

Analysis For quality control the reference material is analysed at the same time and in the same manner as other samples.

Inspection Documents for Type 2.1 Coverage:

Single Analyte Mass Spec Solution ± 0.60 µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# Calculated See Sec. 4.

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3162a Lot Number:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction

Single Analyte Custom Grade Solution. 999 ± 3 µg/ml g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3112a Lot Number:

Certificate of Analysis

Measurement Uncertainty Principles and Implementation in QC

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

Inspection Documents for Type 2.1 Coverage:

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

Inspection Documents for Type 2.1 Coverage:

Reference Materials Update

Comparability and Traceability Point of view from a metrological institute

Measurement Uncertainty in Mechanical Testing

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

What is measurement uncertainty?

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis

XUV 773: X-Ray Fluorescence Analysis of Gemstones

A Statistical Approach to Reporting Uncertainty

Ch. 8 Introduction to Optical Atomic Spectroscopy

Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater

OES - Optical Emission Spectrometer 2000

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Measurement uncertainty and legal limits in analytical measurements

Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13. Determination of Major, Minor and Trace Elements in a Clay Sample

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Philipp Koskarti

CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu

Transcription:

in cooperation with the WG Lead of the Committee of Chemists of GDMB CERTIFICATE OF ANALYSIS ERM -EB102a PbCaSn Certified Values Certified value 1) Uncertainty 2) Mass fraction in % Ca 0.0635 0.0022 Sn 1.01 ± 0.05 Al 124 ± 11 Ag 170 ± 6 Bi 73.7 ± 2.6 Cu 1.3 ± 0.4 Tl 30.2 ± 1.5 1) Unweighted mean value of the means of accepted sets of data, each set being obtained in a different laboratory and/or with a different method of measurement. The values are traceable to the SI (Système International d Unités) via calibration using pure metals or substances of known stoichiometry. 2) Estimated expanded uncertainty U with a coverage factor of k=2, corresponding to a level of confidence of approx. 95 %, as defined in the Guide to the expression of uncertainty in measurement, ISO, 1993. This certificate is valid until 10/2059; this validity may be extended as further evidence of stability becomes available. DESCRIPTION OF THE SAMPLE The Reference Material is available in the form of discs (40 mm diameter and 40 mm height). It is intended for establishing and checking the calibration of optical emission and X-ray spectrometers (excluding microanalysis) for the analysis of samples of similar materials. NOTE European Reference Material ERM -EB102a was produced and certified under the responsibility of BAM Federal Institute for Materials Research and Testing in cooperation with the Committee of Chemists of the GDMB, Gesellschaft für Bergbau, Metallurgie, Rohstoff- und Umwelttechnik according to the principles laid down in the technical guidelines of the European Reference Materials co-operation agreement between BAM-LGC- IRMM. Information on these guidelines is available on the Internet (http://www.erm-crm.org). Page 1 of 5

Accepted as an ERM, Berlin, 2009-09-30. BAM Department I Analytical Chemistry; Reference Materials 12200 Berlin, Germany BAM Division I.1 Inorganic Chemical Analysis; Reference Materials 12200 Berlin, Germany Prof. Dr. U. Panne (Head of Department) Dr. N. Jakubowski (Head of Division) Indicative Values 1) Indicative value 2) Uncertainty 3) Sb 4 ± 4 As < 2 In < 2 Te < 1.1 Zn < 0.5 1) Values were not certified, but given as indicative values, when the number of accepted data sets was considered to be too low (< 5), when the uncertainty from the inter-laboratory certification was considerably larger than the expected range or when only an upper limit can be given. 2) Unweighted mean value of the means of accepted sets of data, each set being obtained in a different laboratory and/or with a different method of measurement. The values are traceable to the SI (Système International d Unités) via calibration using pure substances of known stoichiometry. 3) Estimated expanded uncertainty U with a coverage factor of k=2, corresponding to a level of confidence of approx. 95 %, as defined in the Guide to the expression of uncertainty in measurement, ISO, 1993. Additional Material Information 1) 1) 2) Mean or upper limit Std.-dev 2) Fe < 2 Mg < 1 Na 4 ± 1 S < 3 Values were obtained in an interlaboratory comparison with spark emission spectrometry. Standard deviation calculated from the mean of means of the interlaboratory comparison with n = 6 for Fe, n = 4 for Mg, n = 8 for Na and n = 12 for S. Page 2 of 5

MEANS OF ACCEPTED DATA SETS Certified values Indicative values Mass fraction in % Line no. Ca Sn Al Ag Bi Cu Tl Sb As In Te Zn 1 0.0582 --- 111 165 67.0 0.56 28.7 0.62 0.11 0.15 0.08 0.38 2 0.0613 0.937 119 166 70.2 0.62 29.0 0.94 0.63 0.32 0.43 0.41 3 0.0616 0.962 119 167 70.7 0.84 29.1 1.03 0.81 1.00 0.50 0.45 4 0.0617 0.999 120 168 72.1 0.98 29.1 2.13 2.00 1.44 1.05 0.47 5 0.0622 1.011 121 168 72.2 1.00 29.6 5.08 < 0.1 < 1 < 1.5 0.53 6 0.0632 1.013 122 169 72.2 1.03 29.6 6.87 < 0.4 < 3 < 1 0.55 7 0.0633 1.016 126 171 74.7 1.20 31.1 8.87 < 1 < 1 --- 8 0.0654 1.023 126 172 75.7 1.21 32.0 < 1 < 0.5 9 0.0662 1.028 127 172 76.1 1.45 33.3 < 2.5 < 1 10 0.0671 1.031 131 172 76.1 1.45 11 0.0680 1.039 131 175 77.4 1.71 12 1.040 133 178 79.7 1.79 13 1.053 2.08 14 2.09 M : 0.0635 1.01 124 170 74 1.3 30.2 3.5 0.886 0.726 0.51 0.947 s M : 0.0030 0.04 7 4 4 0.5 1.6 3.3 0.799 0.602 0.401 0.2882 s i : 0.0008 0.008 2 2 1.21 0.19 1.1 0.23 0.075 0.055 0.101 0.128 The laboratory mean values have been examined statistically to eliminate outlying values. Each laboratory mean consists of at least 4 but usually 6 single values. Where --- appears in the table it indicates that an outlying value has been omitted (Grubbs 99 %). " < "-values have not been considered in statistical evaluation. M : mean of means of data sets s i : mean of standard deviations of data sets under repeatability conditions s M : standard deviation of means of data sets TECHNICAL REPORT A detailed technical report (in German) describing the analysis procedures and the treatment of the analytical data used to certify ERM -EB102a is available on request. Page 3 of 5

ANALYTICAL METHOD USED FOR CERTIFICATION Line no. Method --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- Ca 1, 8, 9 FAAS 2, 3, 4, 5, 6, 7, 10,11 ICP-OES Sn 2, 4, 5, 6, 7, 8, 9, 10, 11, 13 ICP-OES 3, 12 FAAS Al 1, 2, 4, 5, 6, 7, 9, 11, 12 ICP-OES 3, 10 FAAS 8 FAAS after separation of the matrix Ag 1 FAAS after separation of the matrix 2, 3, 5, 6, 7, 8, 9, 11, 13 ICP-OES 4 ICP-MS 10, 11, 12 FAAS Bi 1, 2, 3, 6, 8, 9, 10, 11, 12 ICP-OES 4, 5 FAAS 7 ICP-MS Tl 1, 2, 7, 8, 9 ICP-OES 3, 4 FAAS 5, 6 ICP-MS Cu 1 ICP-OES after separation of the matrix 2, 3, 5, 6, 8, 10, 11, 12 ICP-OES 4, 9 ICP-MS 7, 13 FAAS 14 ETAAS Sb 1, 8 ICP-MS 2, 3, 4, 5, 6, 9 ICP-OES 7 FAAS As 1 ICP-OES after separation of the matrix 2, 3, 6 ICP-OES 4 Spectrophotometry 5, 7 ICP-MS In 1 ICP-OES after separation of the matrix 2, 3, 4, 6 ICP-OES 5 ICP-MS Te 1 ICP-OES after separation of the matrix 2, 3, 4, 5 ICP-OES 6, 7 ICP-MS Zn 1, 2, 3, 4 ICP-OES 5 ICP-MS 6 ICP-OES after separation of the matrix 8, 9 FAAS Abbreviations: ICP-OES: Inductively coupled plasma optical emission spectrometry ICP-MS: Inductively coupled plasma mass spectrometry FAAS: Flame atomic absorption spectrometry ETAAS: Electrothermal atomic absorption spectrometry Page 4 of 5

INTENDED USE The CRM is intended for establishing and checking the calibration of optical emission and X-ray spectrometers (excluding micro-analysis) for the analysis of samples of similar materials. The minimum sample size for wet chemical analysis is 0.5 g. INSTRUCTIONS FOR USE Before use, the surface of the material must be prepared by milling or turning on a lathe. STORAGE The material should be stored in a dry and clean environment at room temperature. PARTICIPANTS Aurubis AG, Hamburg, Germany BAM Bundesanstalt für Materialforschung und -prüfung, Berlin, Germany BERZELIUS Stolberg GmbH, Stolberg, Germany BSB Recycling GmbH, Braubach, Germany Exide Technologies GmbH, Bad Lauterberg, Germany GfE Fremat GmbH, Freiberg, Germany Johnson Controls Sachsen-Batterien GmbH & Co. KG, Zwickau, Germany Johnson Controls Power Solutions Europe, Guardamar del Segura, Spain Johnson Controls, VB Autobatterie GmbH & Co. KGaA, Hannover, Germany Muldenhütten Recycling und Umwelttechnik GmbH, Freiberg, Germany Nyrstar, Overpelt, Belgium ThyssenKrupp Steel AG, Duisburg, Germany TU Bergakademie Freiberg, Freiberg, Germany WESER-METALL GmbH, Nordenham, Germany Supply of Reference Materials by: BAM Bundesanstalt für Materialforschung und -prüfung Richard-Willstätter-Straße 11, 12489 Berlin, Germany Phone: +49 30 8104 2061 e-mail: sales.crm@bam.de Fax: +49 30 8104 1117 internet: www.bam.de Page 5 of 5