Preparation of High Purity Plutonium Oxide for Radiochemistry Instrument Calibration Standards and Working Standards

Similar documents
Plutonium 239 Equivalency Calculations

&OAE- 9b I 0 IS& - - I Determination of Plutonium in Urine: Evaluation of Electrothermal Vaporization Inductively Coupled Plasma Mass Spectroscopy.

Determinktion of Iodine to Compliment Mass Spectrometric Measurements

The Radiological Hazard of Plutonium Isotopes and Specific Plutonium Mixtures

Test Method: CPSC-CH-E

PROJECT PROGRESS REPORT (03/lfi?lfibr-~/15/1998):

Use of Gamma Rays from the Decay of 13.8-sec "Be to Calibrate a Germanium Gamma Ray Detector for Measurements up to 8 MeV

Measurement of Low Levels of Alpha in 99M0Product Solutions

K.M. Hoffmann and D. C. Seely

A Few-Group Delayed Neutron Model Based on a Consistent Set of Decay Constants. Joann M. Campbell Gregory D. Spriggs

ION EXCHANGE SEPARATION OF PLUTONIUM AND GALLIUM (1) Resource and Inventory Requirements, (2) Waste, Emissions, and Effluent, and (3) Facility Size

TECHNETIUM-99 IN SOIL

(4) How do you develop an optimal signal detection technique from the knowledge of

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials

ASTER CONF-~W OSTI MAY DISTRIBUTION OF THIS

AlliedSianal. Measurement U ncertai n ty of Silicone Fluid Leakage Testing for Rolamites. W. E. Holland. Federal Manufacturing & Technologies

8STATISTICAL ANALYSIS OF HIGH EXPLOSIVE DETONATION DATA. Beckman, Fernandez, Ramsay, and Wendelberger DRAFT 5/10/98 1.

AEC Research and Development REPORT

Gross Alpha-Gross Beta Analysis in Water by Liquid Scintillation Counting (LSC)

Plasma Response Control Using Advanced Feedback Techniques

Alex Dombos Michigan State University Nuclear and Particle Physics

D. L. Strenge PNL-SA A GENERAL ALGORITHM FOR RADIOACTIVE DECAY WITH BRANCHING AND LOSS FROM A MEDIUM DISCLAIMER. July 1995

APPLICATION SINGLE ION ACTIVITY COEFFICIENTS TO DETERMINE SOLVENT EXTRACTION MECHANISM FOR COMPONENTS OF NUCLEAR WASTE

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

The Concentration of 236 Pu Daughters in Plutonium for Application to MOX Production from Plutonium from Dismantled US Nuclear Weapons

12/16/95-3/15/96 PERIOD MULTI-PARAMETER ON-LINE COAL BULK ANALYSIS. 2, 1. Thermal Neutron Flux in Coal: New Coal Container Geometry

Preparation of Pure Plutonium Metal Standards for Nondestructive Assay

Inside Wall Temperature Measurements of DSTs Using an Infrared Temperature Sensor

Abstract of paper proposed for the American Nuclear Society 1997 Winter Meeting Albuquerque, New Mexico November 16-20, 1997

C o f l F - 9m307-- SPREADSHEET APPLICATION TO CLASSIFY RADIOACTIVE MATERIAL FOR SHIPMENTa. A. N. Brown

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Evaluation of the Total-Evaporation Method for Mass Spectrometric Analysis of Uranium and Plutonium Safeguards Samples: Report of Phase I

BASAL CAMBRIAN BASELINE GEOLOGICAL CHARACTERIZATION COMPLETED

SUMNER ASSOCIATES. 100 Cienega, Suite D Santa Fe, New Mexico Telephone: FAX CEIVED DEC 16 W STI

R. A. Bartsch, Z. Y. Zhang, S. Elshani, and W. Zhao

BWXT Y-12 Y-12. A BWXT/Bechtel Enterprise SMALL, PORTABLE, LIGHTWEIGHT DT NEUTRON GENERATOR FOR USE WITH NMIS

The role of LSD spikes in safeguarding nuclear reprocessing plants R Wellum, Y Aregbe, A Verbruggen, S Richter

ACCURATE QUANTIFICATION OF RADIOACTIVE MATERIALS BY X-RAY FLUORESCENCE: GALLIUM IN PLUTONIUM METAL

Removal of Uranium from Plutonium Solutions by Anion Exchange

Meraj Rahimi, JAI Co. Dale Lancaster, TRW Environmental Safety Systems

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

Tell uric prof i 1 es across the Darrough Known Geothermal Resource Area, Nevada. Harold Kaufniann. Open-file Report No.

PROGRESS REPORT JULY TO SEPTEMBER

Causes of Color in Minerals and Gemstones

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Cable Tracking System Proposal. Nick Friedman Experimenta1 Facilities Division. June 25,1993. Advanced Photon Source Argonne National Laboratory

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

CQNl_" RESPONSE TO 100% INTERNAL QUANTUM EFFICIENCY SILICON PHOTODIODES TO LOW ENERGY ELECTRONS AND IONS

PROCEEDINGS THIRD WORKSHOP GEOTHERMAL RESERVOIR ENGINEERING. December 14-15,1977

Neutron Detector Gamma Insensitivity Criteria

Los Alamos. OF nm LA-"'- Title: HYBRID KED/XRF MEASUREMENT OF MINOR ACTINIDES IN REPROCESSING PLANTS. S. T. Hsue and M. L.

PRECISION STUDIES OF NUCLEI

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

From Single Bone Analysis to Total Skeleton Content

The Graphite Isotope Ratio Method (GIRM): A Plutonium Production Verification Tool

PROJECT PROGRESS REPORT (06/16/1998-9/15/1998):

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

A Statistical Approach to Reporting Uncertainty

N. Tsoupas, E. Rodger, J. Claus, H.W. Foelsche, and P. Wanderer Brookhaven National Laboratory Associated Universities, Inc. Upton, New York 11973

Preparation and Certification of IRMM1027k, Large-Sized Dried (LSD) spike

Validation of detection efficiency of alpha particles in commercial liquid scintillation counters

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

DETECTING AND QUANTIFYING LEWISITE DEGRADATION PRODUCTS I N ENVIRONMENTAL SAMPLES USING ARSENIC SPECIATION*

ISO INTERNATIONAL STANDARD. Measurement of radioactivity in the environment Soil Part 5: Measurement of strontium 90

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

ISO Water quality Determination of carbon 14 activity Liquid scintillation counting method

Multi-Scale Chemical Process Modeling with Bayesian Nonparametric Regression

DO NOT CIRCULATE LOS ALAMOS SCIENTIFIC LABORATORY LA-8399 PERMANENT RETENTION REQUIREDBY CONTRACT. Separation of Trace Uranium from

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

ions which are abundant in waste water. Since zeolites are

ELECTRODEPOSITION OF ACTINIDES

Sequential Isotopic Determination of Plutonium, Thorium, Americium, Uranium, and Strontium in Air-Filter Sample

UNIT OPERATIONS AND CHEMICAL PROCESSES. Alan M. Krichinsky

Certificate of Analysis

A Two-Dimensional Point-Kernel Model for Dose Calculations in a Glovebox Array. D. E. Kornreich and D. E. Dooley

PROCEDURE FOR PLUTONIUM DETERMINATION USING Pu(V1) SPECTRA. Report/Distribution to the mxblic

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)

LQSAlamos National Laboratory

On the HalfiLife of LABORATORY ERNEST ORLANDO LAWRENCE BERKELEYNATIONAL. and Particle Astrophysics Nuclear Science and Physics Divisions

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Method NumberUSTUR 150

GA A26686 FAST ION EFFECTS DURING TEST BLANKET MODULE SIMULATION EXPERIMENTS IN DIII-D

MAGNETICALLY CONTROLLED DEPOSITION OF METALS USING GAS PLASMA. Quarterly Progress Report April - June 1997

Analysis of Shane Telescope Aberration and After Collimation

STP-TS THERMOPHYSICAL PROPERTIES OF WORKING GASES USED IN WORKING GAS TURBINE APPLICATIONS

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Three-Dimensional Silicon Photonic Crystals

Survey, Alignment and Beam Stability at the Advanced Light Source

Revised material in Section 4, Analytical Chemistry

ASTER. Dose Rate Visualization of Radioisotope Thermoelectric. Generators. RECElVED BEC OF THIS DOCUMEMT IS UNLlMIi L, Hanford Company

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS

The photoneutron yield predictions by PICA and comparison with the measurements

The temperature dependence of the C1+ propylene rate coefficient and the Arrhenius fit are shown in Figure 1.

Development of a High Intensity EBIT for Basic and Applied Science

NOTE: This method does not include all of the. specifications (e.g., equipment and supplies) and procedures

Eichrom Technologies, Inc. Analytical Procedures Rev. 1.5 February 10, 2005 Page 1 of 9

TECHNETIUM-99 IN WATER

Transcription:

. LAUR 97-1 Preparation of High Purity Plutonium Oxide for Radiochemistry Instrument Calibration Standards and Working Standards Amy S. Wong and Nelson D. Stalnaker CST-8, MS G740 Los Alamos National Laboratory Los Alamos, NM 87545, USA April 1997 A manuscript to be submitted for publication in AppZied Radiation and Isotopes journal. International Committee for Radionuclide Metrology, 1997 National Institute of Standards and Technology, Gaithersburg, Maryland Htl MSTRIBUTION OF THiS DOCUMENT IS UNtlMITEO

DISCLAIMER This report was p r e p a d as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, make any warranty, express or implied, or assumes any legal Liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed,or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof.

Portions of this document may be illegible in electronic image produck Images are pmduced from the best available original docuult!i&

Preparation of High Purity Plutonium Oxide for Radiochemistry Instrument Calibration Standards and Working Standards Amy S. Wong* and Nelson D. Stalnaker CST-8, MS G740, Los Alamos National Laboratory, Los Alamos, NM 87545, USA Abstract Due to the lack of suitable high level National Institute of Standards and Technology (NIST) traceable plutonium solution standards fiom the NIST or commercial vendors, the CST-8 Radiochemistry team at Los Alamos National Laboratory (LANL) has prepared instrument calibration standards and working standards fiom a well-characterized plutonium oxide. All the aliquoting steps were performed gravimetrically. When a 241Am standardized solution obtained from a commercial vendor was compared to these calibration solutions, the results agreed to within 0.04% for the total alpha activity. The aliquots of the plutonium standard solutions and dilutions were sealed in glass ampules for long term storage. Introduction In any analytical measurement, a good calibration standard is necessary to obtain reliable results. This calibration standard should be traceable to the National Institute of Standards and Technology (NIST) or a certified laboratory or agency (e.g., the New Brunswick Laboratory, NBL). At Los Alamos National Laboratory (LANL), the CST-8 Radiochemistry team performs radiochemical analysis for plutonium and americium at levels of less than 1 nci per ml to 10 mci per ml. There are no suitable standardized solutions from the NIST or commercial vendors for these analyses because of the high activity levels and radionuclide type. To meet the requirements for appropriate standard solutions, the Radiochemistry Team prepared standards from a well-characterized plutonium oxide (PuO,). First the PuO, was assayed and analyzed for impurities, then 241Am and the plutonium isotopic content was determined by mass spectrometry. * To whom all correspondence should be addressed. Phone: 505-667-5096; Fax: 505-665-4737; Email: Wong@lanl.gov 1

Materials and Methods Radiochemistry calibration standard solutions were prepared from a well-characterized plutonium oxide (PuO,) sample (Batch ID: PEOR3258-STD1). This batch was originally prepared by the LANL Plutonium Facility specifically for the preparation of working reference materials. The PuO, powder was blended, calcined (between 900 C and 960 C), pulverized, and screened to generate a stable (low surface area) material. The Radiochemistry team set aside 3 g of the PuO, powder fiom this batch after the analytical characterizations. Characterization of Plutonium Oxide. Five samples were taken fiom the batch of 120 g of PuO,. The samplings represented different vertical and horizontal regions of the PuO, batch in the bottle. Plutonium assay (by coulometric titration), Pu isotopic content (by mass spectrometry), and 241Am analysis (by mass spectrometry and radiochemistry) were done on these five samples of PuO,. Analyses were done on three aliquots of samples to determine particle size distribution, metallic impurities (by DC Arc Emission and ICP-MS), and anion composition (by ion chromatography) of PuO, sample. Alpha spectroscopy was used to determine 244Cm and 252Cf levels in the PuO, sample. Analyses on all radionuclides and most impurities were either traceable to the NBL or NIST via their certified reference materials (CRMs) or standard reference materials (SRMs). Preparation of Primary Standard and VeriJication Sample. Two 100 mg cuts of PuO, were taken from the 3 g oxide sample reserved for the Radiochemistry standard. The aliquots were dissolved by the Pu Assay team which has extensive experience at dissolving PuO,. Each aliquot was weighed to the nearest 0.01 mg. The dissolution method was ASTM C1168-90 Standard Practice for Preparation and Dissolution of Plutonium Materials for Analyses. This method employs a sealed reflux tube dissolution using a mixture of HC1, NO,, and HF under 2

pressure to ensure complete dissolution. These solutions were quantitatively transferred to two weight calibrated 100-mL volumetric flasks. One of the flasks was labeled as Primary PuO, Standard, and the other was labeled as Verification Sample. The solutions were diluted to volume with 4 N HC1. The weight of the solutions was recorded. In these PuO, solutions, the expected concentration was 1.001 mg Pu0,lg solution. The alpha specific activity was 68.15 pci/g solution for the Primary Standard and was 68.13,uCi/g solution for the verification sample. Preparation of Diluted Standard Solutions. The Primary Standard and Verification Sample solutions were too concentrated for preparation of gross alpha standard. Consequently, dilutions and aliquoting were required. Special disposal plastic pyconometers made from plastic transfer pipettes were used to transfer and weight the plutonium solutions. All the dilution and aliquoting steps were performed gravimetrically. Five aliquots of primary standard dilutions and three aliquots of verification sample dilutions were prepared. Figure I shows the schematic for preparing the primary standard and verification sample. The solution aliquot weights ranged from 93 to 160 mg. The aliquots were diluted in weight calibrated 100-mL volumetric flasks. The dilution and weighing steps were performed quickly to minimize any loss due to evaporation. These diluted plutonium solutions contained approximately 0.9 to 1.5 pg (-60 to 104 nci) PuO, per gram of solution. 3

Figure I. Schematic on Preparation of Radiochemistry Calibration Standards. I Primary Standard Oxide (PEOR3258-STD-1) 100 mg of PuO, 100 mg PuO, Primary Standard Solution (100 ml) Verification Sample (100 ml) Y Y I V Five 100-mL Diluted Standard Solutions (-100 mg of Primary Standard Solution in Each Flask) Four - Glass Plate calibration Standards (Label as # Gl-1, G1-2, G1-3, G1-4) Seven Liquid Scintillation Calibration Standards (Label as # Ll-1, L1-2, L1-3, L1-4, etc.) Diluted Standard # 1 - Four Glass plate Calibration Standards (Label as # G2-1, G2-2, G2-3, G2-4) Seven - Liquid Scintillation Calibration Standards (Label as # L2-1, L2-2, L2-3, L2-4, etc.) Diluted Standard # 2 Four - Glass Plate Calibration Standards (Label as # G3-1, G3-2, G3-3, G3-4) Seven -Liquid Scintillation Calibration Standards (Label as # L3-1, L3-2, L3-3, L3-4, etc.) Diluted Standard # 3 Four - Glass Plate Calibration Standards (Label as # G4-I, G4-2, G4-3, G4-4) Seven Liquid Scintillation Calibration Standards (Label as # L4-1, L4-2, L4-3, L4-4, etc.) Diluted Standard # 4 Diluted Standard # 5 v Three 100-mL Diluted Standard Solutions (-100 mg of Verification Sample Solution 1 Each Flask) - Four - Glass Plate Calibration Standards (Label as # G5-1, G5-2, G5-3, G5-4) Seven -Liquid Scintillation Calibration Standards \(Label as # L5-1, L5-2, L5-3, L5-4, etc.) Ver # 1 Five- Liquid Scintillation Samples (Label as # VLI-I, VLl-2) Ver # 2 Five- Liquid Scintillation Samples (Label as # VL2-1, VL2-2) Ver # 3 Five- Liquid Scintillation Samples (Label as # VL3-1, VL3-2) 4

, Verijication Measurements of Plutonium Oxide Solutions. The diluted primary standard and verification sample solutions were compared by liquid scintillation (LS) counting. The counting efficiency of a LS counter for alpha particles is around 100 f 1 %. Five LS samples were prepared from each dilution. Approximately 0.1 to 0.5 g diluted solution was added to the plastic LS vial. Then 10 ml biodegradable LS cocktail (Packard Ultima Gold XR) was added to the vials. In addition, two more LS samples were prepared fiom each of the primary standard dilutions using glass scintillation vials. All LS vials were counted for 10 minutes each on the Packard Tri-Carb 2200CA LS counter. Each vial was counted twice on two separated days. Preparation of Commercial 241Am Standurds. The 241Amstandardized solution (AMP10040, batch 95/241/49) was purchased from the Amersham Corp. (Arlington Heights, IL). This standardized solution is directly traceable to the National Physical Laboratory (NPL in the U.K.). A total of 68.25 pci (as of January 6, 1987) of 241Amactivity was transferred quantitatively to a calibrated 50-mL volumetric flask. The standard certificate gave the total uncertainty o f f 0.80 % at 99.7% confidence (3 0). Five aliquots (90 to 180 nci) of diluted Amersham 241Amstandard solution were prepared for the LS counting. The commercial 241Am LS samples were used to verify the alpha counting efficiency of the LS counter, which was established by the Radiochemistry primary standard and was verified by the verification sample previously. Standardized Solutions Preservation and Storage. The concentrations of plutonium primary standard and verification sample, plutonium diluted solutions, and 241Amdiluted solutions were carefully checked and recorded in the Radiochemistry Laboratory Standard Notebook. Aliquots of these standard solutions were sealed in glass ampules. Each ampule 5

contained - 5 to 12 g of solution. The weight of solution and the amount of radioactivity in each glass ampule were recorded. Then the exterior of each glass ampule was checked for contamination. Each labeled ampule was placed in a heat sealed plastic bag and stored in an upright position in a tray. Results and Discussion Characterization of Plutonium Oxide. The results of analytical characterizations of the PuO, samples are reported in the Reference 1 and summarized in Table I. Plutonium assay and Pu isotopic distribution and 241Amcontent are the key components for calculating the alpha specific activity of the radiochemistry standards. Table I. Plutonium Assay, Isotopic Distribution, 241AmAnalysis, 244Cmand 252Cflevel Component (Five aliquots, duplicate analysis performed on each aliquot) 238pu(4 239Pu 240Pu 241Pu 242Pu Weight percent at 95% CL@) 0.000145 f 0.000008 0.937614 & 0.000013 0.059445 k 0.000005 0.002237 k 0.000008 0.000559 k 0.000004 766.7 f 2.2 pg/g PuO, by mass spectrometry 762.4 f 2.6 pg/g PuO, by radiochemistry < 0.5 pg/g PuO, by alpha spectrometry < 0.1 pg/g PuO, by alpha spectrometry Coulometric titration at 95% CL@) 2 4 1 b 244Cm 252Cf Plutonium Assay (Five aliquots, duplicate analysis performed on each aliquot) g Pu / g PUO, 0.87827 f 0.00072 (a) 238Puwas determined based on alpha spectrometry and mass spectrometry methods. (b) The uncertainty of these values was estimated based on long term and short term measurement uncertainty at 95% CL. 6

Alpha Specific Activity of the Plutonium Sample. The alpha specific activity is calculated using the plutonium assay, the Pu and Am isotopic composition of the PuO,, the half-life of each isotope, and the % alpha emission for each The calculated alpha specific activity for the radiochemistry standard (PEOR3258-STD1) was 77.1 140 mci/g of plutonium or 67.73 mci/g of PuO, (87.827 % Pu) on July 15, 1995. The uncertainty of PuO, specific activity which included the uncertainties of Pu and Am isotopic specific activity and isotopic abundance was rt 0.36 % at 95 % confidence level. Standards Comparison. The plutonium and americium standards were compared by determining the alpha counting efficiency of a Packard LS counting using each solution. The alpha counting efficiency of the LS counter was determined from the average counts per minute in the alpha region of the LS spectrum. The LS counting efficiency data for primary standard dilutions, diluted verification samples, and 241Amsolutions on the Packard LS counter are summarized in Table 11. The counting efficiency for alpha particles determined by the primary dilution standards was 100.20 f 0.12 %. The uncertainty of measurements is expressed as f t * o / h. There were no difference in counting efficiency between plastic and glass scintillation vials. The glass vials did give a broader alpha particle peak. From the data set, the third and fourth dilutions (L3-x, L4-x) were clearly affected by possible transcription error during the initial aliquoting step for dilution. However, the overall precision of all measurements was good. The counting efficiency determined by the verification sample dilutions was 100.26 f 0.05 %. There was an excellent agreement (within 0.06 YO)between the counting efficiencies obtained from the primary standard and the verification sample. 7

The 241Amstandard solution purchased from Amersham was used to verify the plutonium measurements. The LS counter alpha counting efficiency obtained by the 241Amactivity was 100.18 k 0.21 %. The difference of counting efficiency obtained fiom the Radiochemistry plutonium standard and the commercial source was only 0.04 %. The overall counting efficiency of the Packard LS counter was 100.21 _+ 0.40 %. The uncertainty o f f 0.40 % included the uncertainty of PuO, alpha activity and measurements at 95 % CL. This good agreement between all the results suggests that the Radiochemistry PuO, standard solutions are indeed wellcharacterized and prepared. The solutions prepared fiom the PuO, standard can be used to prepare high quality instrument calibration standards and working standards. Table 11. Liquid Scintillation Data on Radiochemistry PuO, Standards and 241AmSolutions Primary Standard") Counting Efficiency Dilution AVG STD 0.05 PR-Dil#1 100.13 PR-Di1#2 100.08 0.07 PR-Di1#3 99.54 0.05 PR-Dil#4 101.07 0.07 PR-Dil#5 100.15 0.12 Overall AVG= 100.20 f 0.12 %(4)( Verfication Sample(2) Counting Efficiency Dilution AVG STD VR-Dil #I 100.31 0.08 VR-Dil#2 100.13 0.15 VR-Dil#3 100.34 0.05 Overall AVG = 100.26 f 0.05 %(4) 241Am(3) Counting Eff., % Ll Am-96 100.44 L2Am-96 100.11 L3Am-96 100.09 L4Am-96 100.25 L5Am-96 99.99 AVG= 100.18 f 0.21 %(') Amersham LS Sample (1) For each primary dilution, seven LS samples were prepared. Each LS sample was counted twice. (2) For each verification dilution, five LS samples were prepared. Each LS sample was counted twice. (3) For 241Amstandardized solution, five LS samples were prepared. Each LS sample was counted twice. (4) The uncertainty of measurements is expressed as t * o h. _+ Uncertainty E~tirnate~,~,~. The details of all the measurement uncertainty estimates are given in Reference 5. The uncertainties in the measurement results are estimated based on the random and systematic errors of the measurements, and they are expressed at the 95 % confidence level. 8

Conclusion Using a well-characterized PuO, sample, standards of high quality for calibrating radiochemistry instrumentation were prepared. The overall uncertainty in the alpha specific activity (pci/g solution) of primary standard solution and verification sample solution is _+ 0.35% at 95 % CL. With accurately known alpha specific activity and through decay corrections, these standardized solutions will be used as the calibration standards for the Radiochemistry team. Acknowledgment The authors thank all the members of CST Analytical Chemistry Groups who performed the characterizations of PuO, materials. Special thanks to Bob Marshall, the project leader of Working Reference Materials Program, who provided the radiochemistry the PuO, materials; and to the Pu Assay team, who provided the dissolution of PuO, samples; and to Larry Bruckner and Larry Ticknor, the statisticians of ESA- 1, who helped us to determine the uncertainty of the PuO, characterizations. References 1. Sample analysis report on CST # 200016529, LANL CST Analytical Chemistry, October 1995. 2. Firestone, R. B., Shirley, V. S. editors (1996) Table oflsotopes, 8* Edition, John Wiley & Sons, New York. 3. Browne, E. and Firestone, R. B. (1986), Table of Radioactive Isotopes, Shirley, V. S. editor, John Wiley & Sons, New York. 4. Wong, A. S. and Stalnaker,N. D. (1997) Preparation of High Purity Plutonium Oxide for Radiochemistry Instrument Calibration Standards and Working Standards, a Los Alamos Technical Report, in preparation. 5. Taylor, B. N. and Kuyatt, C. E. (1993), Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results, NIST Technical Note 1297, National Institute of Standards and Technology, Gaithersburg, MD. 6. Certification of StandardizedRadioactive Sources (1968), ICRU Report 12, International Commission on Radiation Units and Measurements, Washington, DC. 9