New neutron detector based on Micromegas technology for ADS projects

Similar documents
PoS(FNDA2006)010 FAST NEUTRON DETECTORS BASED ON MICROMEGAS TECHNOLOGY. S. Andriamonje * I. Giomataris and J. Pancin

A new detector for neutron beam monitoring

Neutron imaging with a Micromegas detector

DIRECT EXPERIMENTAL TESTS AND COMPARISON BETWEEN SUB-MINIATURE FISSION CHAMBERS AND SPND FOR FIXED IN-CORE INSTRUMENTATION OF LWR

Breakdown limit studies in high-rate gaseous detectors

FEASIBILITY STUDY OF NEW MICROSCOPIC FISSION CHAMBERS DEDICATED FOR ADS

GEM: A new concept for electron amplification in gas detectors

Advances in the Micro-Hole & Strip Plate gaseous detector

Fission Fragment characterization with FALSTAFF at NFS

Neutron capture and fission reactions on. sections, -ratios and prompt -ray emission from fission. 1 Introduction and Motivation

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Simulating the Charge Dispersion Phenomena in Micro Pattern Gas Detectors with a Resistive Anode

arxiv: v3 [physics.ins-det] 19 Sep 2017

RADIOLOGICAL IMPACT OF THE TRIGAACCELERATOR-DRIVEN EXPERIMENT (TRADE)

Light ion recoil detector

PHOTOELECTRON COLLECTION EFFICIENCY AT HIGH PRESSURE FOR A GAMMA DETECTOR ENVISAGING MEDICAL IMAGING

GEM at CERN. Leszek Ropelewski CERN PH-DT2 DT2-ST & TOTEM

Background optimization for a new spherical gas detector for very light WIMP detection

He-3 Neutron Detectors

PoS(EPS-HEP2015)232. Performance of a 1 m 2 Micromegas Detector Using Argon and Neon based Drift Gases

Present and Future of Fission at n_tof

A fast triple GEM detector for high-rate charged-particle triggering

ATLAS New Small Wheel Phase I Upgrade: Detector and Electronics Performance Analysis

Precision Calibration of Large Area Micromegas Detectors Using Cosmic Muons

MUSE-4 BENCHMARK CALCULATIONS USING MCNP-4C AND DIFFERENT NUCLEAR DATA LIBRARIES

Flux and neutron spectrum measurements in fast neutron irradiation experiments

Micro Pixel Chamber with resistive electrodes for spark reduction

Detecting low energy recoils with Micromegas

A novel large-volume Spherical Detector with Proportional Amplification read-out

A Monte Carlo Simulation for Estimating of the Flux in a Novel Neutron Activation System using 252 Cf Source

EEE4106Z Radiation Interactions & Detection

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

New irradiation zones at the CERN-PS

Neutrons for Science (NFS) at SPIRAL-2

Performance of a triple-gem detector for high-rate particle triggering

Study of GEM-like detectors with resistive electrodes for RICH applications

Nuclear waste management

Development of New MicroStrip Gas Chambers for X-ray Applications

ACTAR TPC: an active target and time projection chamber for nuclear physics

Discharge studies in Micromegas detectors in low energy hadron beams

Electron transparency, ion transparency and ion feedback of a 3M GEM

M.Cagnazzo Atominstitut, Vienna University of Technology Stadionallee 2, 1020 Wien, Austria

arxiv: v1 [astro-ph.im] 6 Dec 2010

VERIFICATION OF MONTE CARLO CALCULATIONS OF THE NEUTRON FLUX IN THE CAROUSEL CHANNELS OF THE TRIGA MARK II REACTOR, LJUBLJANA

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

Track Resolution Measurements for a Time Projection Chamber with Gas Electron Multiplier Readout

Design and test of an Accelerator Driven Neutron Activator at the Joint Research Centre of the European Commission

Spatial Resolution of a Micromegas-TPC Using the Charge Dispersion Signal

Total kinetic energy and fragment mass distributions from fission of Th-232 and U-233

arxiv: v1 [astro-ph.im] 7 Nov 2011

A Complete Simulation of a Triple-GEM Detector

Nuclear cross-section measurements at the Manuel Lujan Jr. Neutron Scattering Center. Michal Mocko

R&D on Astroparticles Detectors. (Activity on CSN )

Performance test of triple GEM detector at CERN n_tof facility

MICROMEGAS PERFORMANCE BASED IN ARGON- ISOBUTANE AND ARGON-DEMETHYL-ETHER

Gas-filled Detectors

1. General information and technical data of TAPIRO research reactor

Measurement of the transverse diffusion coefficient of charge in liquid xenon

arxiv: v1 [physics.ins-det] 10 May 2011

Nuclear Instruments and Methods in Physics Research A

Determination of the boron content in polyethylene samples using the reactor Orphée

Measurement of the n_tof beam profile in the second experimental area (EAR2) using a silicon detector

CCD readout of GEM-based neutron detectors

USA HTR NEUTRONIC CHARACTERIZATION OF THE SAFARI-1 MATERIAL TESTING REACTOR

GEM-based photon detector for RICH applications

Estimation of 210 Po Losses from the Solid 209 Bi Target Irradiated in a Thermal Neutron Flux

A complete simulation of a triple-gem detector

arxiv:physics/ v2 [physics.ins-det] 31 Oct 2003

Physics sources of noise in ring imaging Cherenkov detectors

Chapter V: Interactions of neutrons with matter

Geant4 simulation for LHC radiation monitoring

MODELLING OF HTRs WITH MONTE CARLO: FROM A HOMOGENEOUS TO AN EXACT HETEROGENEOUS CORE WITH MICROPARTICLES

X-ray ionization yields and energy spectra in liquid argon

Neutronics Experiments for ITER at JAERI/FNS

A proposal to study gas gain fluctuations in Micromegas detectors

Particle Energy Loss in Matter

CONCEPTUAL STUDY OF NEUTRON IRRADIATOR-DRIVEN BY ELECTRON ACCELERATOR

Chem 481 Lecture Material 4/22/09

Micro-pattern gaseous detectors

Reactivity monitoring using the Area method for the subcritical VENUS-F core within the framework of the FREYA project

ACTAR TPC: an active target and time projection chamber for nuclear physics. 17/09/2015 T. Roger COMEX 5 1

A Time Projection Chamber for Precision ^^^Pu(n,f) Cross Section Measurement

physics/ Sep 1997

Search for low-mass WIMPs with Spherical Detectors : NEWS-LSM and NEWS-SNO

PANDA-?? A New Detector for Dark Matter Search

ACTIVATION ANALYSIS OF DECOMISSIONING OPERATIONS FOR RESEARCH REACTORS

arxiv:physics/ v2 27 Mar 2001

NEUTRONIC ANALYSIS STUDIES OF THE SPALLATION TARGET WINDOW FOR A GAS COOLED ADS CONCEPT.

Results with Micromegas modules at LP-TPC

& SUN ZhiJia 1;2. Abstract

arxiv: v1 [physics.ins-det] 4 Nov 2013

Recent developments on Micro-Pattern Gaseous Detectors

Measurement of the Ionizing Energy Depositions after Fast Neutron Interactions in Silicon

DETECTORS. I. Charged Particle Detectors

NEW DEVELOPMENTS IN GASEOUS DETECTORS. Fabio Sauli. CERN, Geneva, Switzerland

A progress report on the development of the CsI-RICH detector for the ALICE experiment at LHC

arxiv:physics/ v1 3 Aug 2006

Experimental High Energy Physics & Plasma Etching

Particle Energy Loss in Matter

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center

Transcription:

1 New neutron detector based on Micromegas technology for ADS projects Samuel Andriamonje a*, Grégory Andriamonje b, Stephan Aune a, Gilles Ban c, Stéphane Breaud d, Christophe Blandin d, Esther Ferrer a, Benoit Geslot d, Arnaud Giganon a, Ioannis Giomataris a, Christian Jammes d, Yacine Kadi e, Philippe Laborie c, Jean François Lecolley c, Julien Pancin a, Marc Riallot a, Roberto Rosa f, Lucia Sarchiapone e, Jean Claude Steckmeyer c, Joel Tillier c a CEA-Saclay, DSM/DAPNIA, F- 91191 Gif-sur-Yvette, France b IXL - Université Bordeaux 1 - BAT. A31-351 cours de la Libération - F-33405 Talence Cedex, France c CNRS/IN2P3 LPC Caen, 6 Boulevard Maréchal Juin,F-14050 Caen Cedex, France, d CEA/DEN/ Cadarache, 13108 Saint-Paul Lez Durance-France, e CERN CH 1211 Geneva-Switzerland f ENEA-Casaccia, Via Anguillarese, 00060 Roma, Italy Abstract A new neutron detector based on Micromegas technology has been developed for the measurement of the simulated neutron spectrum in the ADS project. After the presentation of simulated neutron spectra obtained in the interaction of 140 MeV protons with the spallation target inside the TRIGA core, a full description of the new detector configuration is given. The advantage of this detector compared to conventional neutron flux detectors and the results obtained with the first prototype at the CELINA 14 MeV neutron source facility at CEA-Cadarache are presented. The future developments of operational Piccolo-Micromegas for fast neutron reactors are also described. Keywords: Neutron; Detector; ADS; TRIGA; Micromegas; MC simulation PACS: 24.10.Lx; 25.85.Ec; 28.41.Ak; 28.41.Rc; 28.50.-k; 29.40.Cs; 28.20.-v; 28.90.+I; 29.25.Rm * Corresponding author. Tel.: +33-1-69-08-55-86; fax: +33-1-69-08-75-84; e-mail: sandriamonje@cea.fr (S. Andriamonje).

2 1. Introduction In the ADS (Accelerator Driven System) project, one of the most important aspects needed for approval of the demonstrator is the experimental verification of the simulation. Of particular interest is the determination of the neutron spectrum (i.e. neutron flux as a function of the neutron energy) for different configurations of the sub-critical device. The neutron flux in ADS consists of neutrons produced via spallation reactions in the target and fissions from the multiplying blanket. Unfortunately the neutron spectra cannot be measured using only one type of detector. To cover the complete energy range of the produced neutrons, a new neutron detector concept based on Micromegas technology has been developed. The requirements for neutron spectrum measurements in the ADS project and the description of the new detector configuration dedicated to the neutron flux (i.e. neutron spectra) measurement in the present ADS project are given. In order to be placed inside an empty rod of a reactor this detector needs to be compact (3.5 cm x 3.5 cm x 3.5 cm) hence the origin of the Piccolo-Micromegas name. The advantage of this detector compared to conventional neutron flux detectors and the results obtained with the first prototype at the CELINA 14 MeV neutron source facility at CEA-Cadarache are finally presented. 2. Simulated neutron spectra in ADS project One example of the possible subcritical configuration based in TRIGA-ADS project [1], is shown in Fig. 1. The central rod and the rods named ring B are reserved for the spallation target and its cooling system. The measurement can be only achieved starting from ring C. The distribution of neutrons in three-dimensional space as a function of energy and time is simulated using the innovative simulation codes (FLUKA and EA-MC [2,3] at CERN, and MCNP-4C [4] at ENEA/Casaccia). The analysis of neutron flux spectra is a necessary step in order to characterize the system. The TRIGA- ADS project is essentially a thermal system, cooled and moderated by water and using a fuel matrix rich in hydrogen. Fig. 1. Schematic view of the Fuel configuration for Sub-critical operation with possible locations of the neutron flux detectors. Despite the high fuel enrichment (20 % 235 U) and the rather small pitch between fuel elements (~ 2.35 mm for ~ 18 mm radius fuel elements), the abundance of this element guarantees the thermalization of the fission neutrons inside the fuel, therefore defining the very characteristic two-humped spectrum (the thermal and the high-energy humps). Moreover, the presence of a water buffer region (empty fuel channels in ring B) causes a prompt energy loss for the spallation neutrons and their diffusion, increasing the efficiency of the source neutrons. Fig. 2 shows the neutron energy spectrum for different regions of the core for k src = 0.97. The target presents a harder neutron flux and the energy spectrum shows a high-energy peak around 2 3 MeV (spallation peak) and the resonances of tantalum (in particular, the ~15,000 barn resonance at 4 ev). The thermal hump (Maxwell distribution) is due to the neutrons reflected into the centre of the core. The fuel regions present similar neutron energy spectra, with decreasing integral value with the distance from the centre of the core. The presence of the spallation source produces a relative energy shift between the high-energy flux (target region) to the thermal distribution (reflector region) permitting, to a limited extent, the analysis of

3 different elements under different spectrum conditions. 10 11 10 10 10 9 10 8 Ring E Ring C Tantalum target 10 7 10-3 10-2 10-1 10 0 10 1 10 2 10 3 10 4 10 5 10 6 10 7 Neutron energy (ev) Fig. 2. Neutron flux spectra at selected locations of the TRIGA- ADS reference configuration. 3. Micromegas as neutron detector Ring B 3.1. Brief description of MICROMEGAS technology The general description of the principle of Micromegas (MICRO-MEsh-GAseous Structure) technology can be found in references [5,6]. We describe here the typical aspect of the Piccolo- Micromegas detector for in-core measurement in nuclear reactor. The principle of the Micromegas neutron detector used in n_tof experiment [7,8] is reported in Fig. 3. As shown in this figure, the amplification occurs between the mesh plane and the microstrip plane. The device operates as a two-stage parallel plate avalanche chamber. Ionization electrons, created by the energy deposition of an incident charged particle in the conversion gap, drift and can be transferred through the cathode micromesh, they are then amplified in the small gap, between anode and cathode, under the action of the electric field which is high in this region. The electron cloud is finally collected by the anode microstrip, while the positive ions are drifting in the opposite direction and are collected on the micromesh. For this experiment, appropriate neutron/charged particle converters have been used (see Fig. 3): (i) 6 Li(n,α) for the neutron having an energy up to 1 MeV, (ii) H(n,n')H and 4 He(n,n') 4 He for higher energy neutron. This method has been extended to the measurement of neutron flux in-core environment. One of the main qualities of the detector is its high resistance to the radiation. Several experiments have shown that the Micromegas detector meets with this requirement. Only 10% variation of the gain has been observed after an irradiation of the detector of 24 mc/mm 2, this is equivalent of ten years irradiation of the detector on LHC (Large Hadron Collider) project [9]. Fig. 4. Schematic view of the Piccolo-Micromegas detector (in horizontal position) for neutron flux measurements. 3.2. Description of Piccolo-Micromegas Fig. 3. The principle of Micromegas for neutron detector used in n_tof experiment [8]. The principle of Piccolo-Micromegas detector is shown in Fig. 4. The different materials, stainless steel and ceramic, have been chosen in order to respect the following conditions: - low activation with high neutron flux, - small perturbation of the neutron flux, - homogeneity of the dilatation coefficients, - low out gassing for a sealed gaseous detector,

4 The chamber body of the detector is a stainless steel cylinder with 35 mm of external diameter, 30 mm of internal diameter and 35 mm of height and closed in the bottom part. The cover is made of stainless steel equipped by a special piece for handling the detector inside the reactor rod and six special connectors stuck or welded by a special technique already used for the micro fission chamber [10]. The drift electrode is composed of three neutron/charged particle converters: 10 B, 235 U, 232 Th and one empty for the recoil nucleus of H of Ar- C 4 H 10 filling gas. According to the neutron interaction cross section: - the 235 U is used for monitoring, since the fission cross section of 235 U from thermal to several MeV neutron energy is well known, - the 10 B for the measurement of epithermal neutron flux (up to several kev) - the 232 Th for the measurement of fast neutrons having an energy larger than the fission threshold (~ 1 MeV) and the H+ recoil for the neutron having an energy greater than the energy threshold of the detector (several kev). The Micromesh is a stainless steel woven with 19 µm wires. 500 lines per inch pitch insure the transparency. The drift gap (1 mm) and the amplification gap (160 µm the distance between the Micromesh and the anode pad) are ensured by an insulating part in ceramic opening by four holes having the same diameter that the deposit sample. 4. Experiment test and first results homemade fast amplifier. An oscilloscope-pc has been used for the data acquisition. The energy resolution of the detector at 5.9 kev 55 Fe X-ray is about 20%. The gains of the detector as a function of the different voltage applied on the different pads have been measured. Fig. 5. a) Gain of the Piccolo-Micromegas detector versus HV- PAD for HV-Mesh = +110 V and the drift cathode in the ground and 1 bar Ar+CO 2, b) Example of the fission fragment pulse. The results show clearly (Fig. 5a) that a very low polarization of the pad is largely sufficient for the observation of the signal from fission fragments (several MeV compared to 5.9 kev energy deposited by the X ray of 55 Fe radioactive source). 4.2. Experiments and results with neutron source The first test of the Piccolo-detector has been performed at the 14 MeV neutron source CELINA facility at CEA-Cadarache. Two types of gas mixtures (cf. Paragraph 4.1) have been used to fill at 1 bar the sealed detector. 4.1. Characteristics of Piccolo-Micromegas detector To determine the main characteristics of the new Micromegas detector, 55 Fe X-ray source and two types of gas mixture (Ar-10%CO 2 and Ar-2%C 4 H 10 ) have been used. The pressure of the sealed detector has been set to one bar. The ionization power of fission fragment compared to recoil nucleus or alpha particle are very different. In order to adapt the dynamic range of the collected charge, a new method has been used. The amplification gap has been taken identical for the four pads (160 µm). The drift electrode is grounded. The mesh and the four reading pads have been polarized positively and individually. The signal from each pad is collected and shaped by a Fig. 6. a) Amplitude histogram of 235 U fission fragments, b) Amplitude histogram of H+ recoils.

5 An example of the pulse obtained from the fission fragment emitted by the interaction of a neutron with 235 U is shown in Fig. 5b. The contribution of the fast electron and the slow ions are shown clearly in the figure. A special routine has been developed with Matlab [11] to analysis and sort the raw data for eventual spark and saturation. Two histograms as a function of the amplitude of the pulse (proportional to the energy deposited by the charged particle) are reported in Fig. 6a and Fig. 6b. The first one corresponds to the 235 U fission fragments and the second one to the recoils of hydrogen. The gain used for the fission fragments is ~20 times lower than the one used for the H + recoils. The resolution of the detector is not good enough to observe the two bumps of the fission fragments distribution. 4.3. Comparison with simulation A full simulation of the Piccolo-Micromegas detector placed inside the CELINA-Cadarache neutron source facility has been performed using FLUKA [2]. The study has been centered on the particular case of 235 U, because of the well-known fission cross section. The model used consisted of a 14 MeV neutron source moderated by graphite with an intensity of about 2x10 9 n/s/4π and samples of 235 U and 232 Th 50 µg each. Fig. 7. Neutron flux distribution (arbitrary units) inside the Piccolo-Micromegas detector The neutron flux distribution inside the Piccolo- Micromegas detector is shown in Fig. 7. The detector seems homogeneously illuminated by the 14 MeV neutron source placed a few tens of cm away, inducing some ~0.08 fissions/s per µg of 235 U. This is in good agreement with the measured integral of the fission fragment spectrum (Fig. 6a). 5. Conclusion A new neutron detector named Piccolo- Micromegas has been developed. The first results presented in this paper show clearly the performance and the ability of this detector to measure simultaneously from thermal to high-energy neutron flux in the nuclear reactor especially in ADS project. The next step is the use of a special material for the connectors and the cable for operation at high temperature up to 300 C. References [1] The TRADE Working group, TRiga Accelerator Driven Experiment (TRADE) Feasibility Report. ENEA Report, March 2002. [2] A. Fasso, et al., Intermediate Energy Nuclear Data: Models and C, Proceedings of a Specialist s Meeting, Issy les Moulineaux (France) 1994 [3] C. Rubbia, et al., Conceptual Design of a Fast Neutron Operated High Power Energy Amplifier, CERN/AT/95-44 (EET) Geneva, September 29, 1995. Y. Kadi, et al., The EA-MC Monte Carlo Code Package, in Proc. of the Fifth International Meeting on Simulating Accelerator Radiation Environment SARE-5: Models and Codes for Spallation Neutron Sources, OECD Headquarters, Paris, France, July 17-18, 2000. [4] J. F. Briesmeister, Editor A MCNP TM, General Monte Carlo N-Particle Transport Code, Version 4C, Los Alamos National Laboratory report LA-13709-M, March 2000). [5] I. Giomataris, et al., A high-granularity position sensitive gaseous detector for high particle-flux environments, Nucl. Instr. and Meth. A 376 (1996) 29. [6] G. Charpak et al., Micromegas a multipurpose gazeous detector, Nucl. Instr. and Meth. A 478 (2002) 26. [7] S. Andriamonje, et al., Experimental studies of a MICROMEGAS neutron detector, Nucl. Instr. Meth. A 481 (2002) 36-45. [8] J. Pancin, et al., Nucl. Instr. and Meth. A 524 (2004) 102-114 [9] G. Puill, et al., IEEE Trans. Nucl. Sci. NS-46 (6) (1999) 1894 [10] G. Bignan and J.C. Guyard (1994), Sub-miniature Fission Chamber with Tight Feedthrough, Patent CEA N 94-14293. [11] http://www.mathtools.net/matlab/