Scattering properties of the retina and the choroids determined from OCT-A-scans

Similar documents
CHIRP OPTICAL COHERENCE TOMOGRAPHY

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a).

10. OPTICAL COHERENCE TOMOGRAPHY

Long Path Industrial OCT High-precision Measurement and Refractive Index Estimation

Ambiguity of optical coherence tomography measurements due to rough surface scattering

In Vivo Total Retinal Blood Flow Measurement by Fourier. domain Doppler Optical Coherence Tomography

POLARISATION-SENSITIVE OPTICAL COHERENCE TOMOGRAPHY FOR MATERIAL CHARACTERISATION AND TESTING

Imaging thermally damaged tissue by polarization sensitive optical coherence tomography

PRINCIPLES OF PHYSICAL OPTICS

Resolution improvement in optical coherence tomography with segmented spectrum management

Degree of polarization in laser speckles from turbid media: Implications in tissue optics

Physical Optics. Lecture 7: Coherence Herbert Gross.

Laser Speckle and Applications in Optics

Metrology and Sensing

Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium

Light Propagation in Free Space

Michelson Interferometer. crucial role in Einstein s development of the Special Theory of Relativity.

Michelson Interferometer

Advanced Optical Coherence Tomography techniques: novel and fast imaging tools for non-destructive testing

Review Article Nanoparticle Imaging with Polarization Interferometric Nonlinear Confocal Microscope

UvA-DARE (Digital Academic Repository)

Optics.

Analysis of the signal fall-off in spectral domain optical coherence tomography systems

Metrology and Sensing

Chemistry Instrumental Analysis Lecture 15. Chem 4631

UvA-DARE (Digital Academic Repository)

Imaging carious human dental tissue with optical coherence tomography

Optical properties of ocular tissues in the near infrared region

Thickness and Birefringence of Healthy Retinal Nerve Fiber Layer Tissue Measured with Polarization-Sensitive Optical Coherence Tomography

Quantitative Biomedical Optics

Flow measurement without phase information in optical coherence tomography images

Effect of source spectral shape on task-based assessment of detection and resolution in optical coherence tomography

Edward S. Rogers Sr. Department of Electrical and Computer Engineering. ECE426F Optical Engineering. Final Exam. Dec. 17, 2003.

High-Resolution. Transmission. Electron Microscopy

OPTI 511L Fall A. Demonstrate frequency doubling of a YAG laser (1064 nm -> 532 nm).

Optical Coherence Tomography: Modeling and Applications

Coherence and width of spectral lines with Michelson interferometer

Analysis of second-harmonic generation microscopy under refractive index mismatch

Blood flow measurement and slow flow detection in retinal vessels with Joint Spectral and Time domain method in ultrahigh speed OCT

Single Emitter Detection with Fluorescence and Extinction Spectroscopy

Quantitative retinal blood flow measurement with. three-dimensional vessel geometry determination. using ultrahigh-resolution Doppler optical

Imaging of vibrating objects using speckle subtraction

Measurement of Corneal Thickness by Loser Doppler Interferometry

High-Resolution Imagers

Constructive vs. destructive interference; Coherent vs. incoherent interference

Lecture 9: Introduction to Diffraction of Light

Effects of retinal scattering in the ocular double-pass process

Lichtausbreitung in streuenden Medien: Prinzip und Anwendungsbeispiele

Recent progress in SR interferometer

Electricity & Optics

Unit 4 Parent Guide: Waves. What is a wave?

EUV Reflectivity measurements on Acktar Sample Magic Black

Lecture 19 Optical MEMS (1)

Low Coherence Vibration Insensitive Fizeau Interferometer

Correction of Errors in Polarization Based Dynamic Phase Shifting Interferometers

Astronomy 203 practice final examination

Metrology and Sensing

Separation of absorption and scattering profiles in spectroscopic optical coherence tomography using a least-squares algorithm

A novel laser guide star: Projected Pupil Plane Pattern

SIMG Optics for Imaging Solutions to Final Exam

Modeling Focused Beam Propagation in scattering media. Janaka Ranasinghesagara, Ph.D.

Quantification of optical Doppler broadening and optical path lengths of multiply scattered light by phase modulated low coherence interferometry

INTERFEROMETRIC METHOD FOR THE STUDY OF SPATIAL PHASE MODULATION INDUCED BY LIGHT IN DYE-DOPED DNA COMPLEXES

Supplementary Figure 1. Characterization of the single-photon quantum light source based on spontaneous parametric down-conversion (SPDC).

Lecture 11: Introduction to diffraction of light

Frequency-resolved interferometer measurements of polarized wave propagation in scattering media

Scattering of Poincaré beams: polarization speckles

Digital Holographic Measurement of Nanometric Optical Excitation on Soft Matter by Optical Pressure and Photothermal Interactions

Particle-Wave Duality and Which-Way Information

Fast simulation of finite-beam optical coherence tomography of inhomogeneous turbid media

Where are the Fringes? (in a real system) Div. of Amplitude - Wedged Plates. Fringe Localisation Double Slit. Fringe Localisation Grating

Metrology and Sensing

School. Team Number. Optics

SOFT X-RAYS AND EXTREME ULTRAVIOLET RADIATION

PHY410 Optics Exam #3

Optics, Optoelectronics and Photonics

Part 1 - Basic Interferometers for Optical Testing

Imaging through random media using low-coherence optical heterodyning

Multiple and dependent scattering effects in Doppler optical coherence tomography Kalkman, J.; Bykov, A.V.; Faber, D.J.; van Leeuwen, T.G.

PS210 - Optical Techniques. Section VI

Probing the orbital angular momentum of light with a multipoint interferometer

Concave mirrors. Which of the following ray tracings is correct? A: only 1 B: only 2 C: only 3 D: all E: 2& 3

Optics. n n. sin c. sin

Density Field Measurement by Digital Laser Speckle Photography

The Michelson Interferometer

Dielectric Meta-Reflectarray for Broadband Linear Polarization Conversion and Optical Vortex Generation

arxiv: v1 [quant-ph] 18 Sep 2017

Unbalanced lensless ghost imaging with thermal light

Development of measurement technique to evaluate thermal conductivity of thermoelectric Bi 2 Te 3 submicron thin films by photothermal radiometry

Optical coherence tomography (OCT) is a ranging technique

Gratings in Electrooptic Polymer Devices

Phase-sensitive swept-source interferometry for absolute ranging with application to measurements of group refractive index and thickness

Tutorial on Biomedical Optics

Measurement of the laser cut front geometry

Low-coherence heterodyne photon correlation spectroscopy

ABSTRACT. Introduction

For more information, please contact

TA/TI survey. Phy Phy

Optics, Light and Lasers

Noise adaptive wavelet thresholding for speckle noise removal in optical coherence tomography

Transcription:

International Ophthalmology 23: 291 295, 2001. J.R. Sampaolesi (ed.), Laser Scanning: Update 1, 109 113. 2001 Kluwer Academic Publishers. Printed in the Netherlands. 291 Scattering properties of the retina and the choroids determined from OCT-A-scans H. Hammer, D. Schweitzer, E. Thamm, A. Kolb & J. Strobel University of Jena, Department of Ophthalmology, Jena, Germany Key words: OCT, tissue scattering, retina, choroid Abstract Goal: To determine the coefficient and the anisotropy of scattering as well as the refractive indices in the retina and in the choroid noninvasively in vivo. Methods: The power of coherent reflected light versus fundus depth is recorded in OCT-A-scans. The ratio of refractive indices is derived from the height of the reflection peaks. Provided that the absorption coefficient is known from fundus reflectometry, the scattering coefficient and anisotropy are calculated from the offset and the slope of the signal behind the reflection peaks on the basis of a single backscattering model. Results: We found scattering coefficients of 12/mm (retina) and 27.5/mm (choroid) as well as anisotropy values of 0.97 (retina) and 0.90 (choroid). Discussion: The OCT is usually employed for the measurement of intraocular distances. The new technique described here gives the unique opportunity to determine further interesting parameters of single ocular layers. The values given above are in good agreement with in vitro results. Introduction Recently, optical coherence tomography (OCT) is widely used to observe hidden morphologic structures in the depth of biological tissues. In the ophthalmology, OCT is used to obtain tomographic images of the ocular fundus. This enables the measurement of the thicknesses of retinal structures. However, coherent reflected light carries more information characterising the optical properties of tissue. The goal of this paper is to investigate the feasibility of OCT to determine the refractive index, the extinction coefficient, and the scattering anisotropy of tissues using a single backscattering model. This will be demonstrated at the example of an OCT-scan from a human ocular fundus in vivo. Methods OCT is an interferometric technique. Let us consider, e.g., a Michelson interferometer 7with a scattering sample in one arm and a mirror in the other one (reference arm). In order to avoid the superposition of many interfering waves backscattered from the sample, a light source with a high spatial but a short temporal coherence is used. This results in interference only if the optical path difference in both arms of the interferometer is not larger than the coherence length of the source [1]. If L + L is the distance of the reference mirror and L that of the surface of the sample (z = 0) from the beam splitter (50/50), and if there is a reflection R(z) in the depth z in the sample, at the detector the power [2, 3]. P(z) = P 0 [(1 + R(z))/4 +1/2 R(z) γ(2( L nz)/c) = cos(4π( L nz)/λ)] (1) is measured, where P 0 is the power of the light source, n is the refractive index of the sample, c is the velocity of the light, λ is the wavelength, and γ is the temporal coherence function. As involuntary eye movements make it difficult to observe the interference between waves reflected from the eye and from a stationary mirror, Fercher et al. [4, 5, 6] used a dual beam interferometer: The light of the source is divided by an interferometer so that [ 109 ]

292 there are two wavefronts with a path difference equal to the optical length of the eye incident on the eye. Thus, interference between the wave reflected at the cornea and that reflected at the ocular fundus is observed. In order to avoid losses of the signal resulting from the interference of a plane wave reflected at the fundus with a curved wave reflected at the cornea, a diffractive lens focusing 40% of the incident light at the cornea and 60% at the fundus is used. For this optical arrangement, Equation (1) has to be modified: P(z)/P 0 = 0.4R c + 0.6R(z) + 2 0.4R c 0.6R(z) γ(2( L nz)/c) cos(4π( L nz)/λ). (2) Here, R c is the reflection at the cornea and R(z) is the reflection at the fundus in the depth z. The interference term (third summand of Equation (2)) is separated from the incoherent power by shifting the reference mirror of the interferometer with a constant speed v =dl/dt and filtering the detected signal with the Doppler frequency f D = v/λ. If R(z) results from the scattering at cells and cell compartments for which the distance is shorter than the coherence length of the light source, the measured power results from a coherent summation over all scattered waves. To calculate this, the local distribution of the scatterers as well as the amplitude and the phase of their scattering function have to be known. Since this is difficult in biological tissues, a simple single scattering approach [3, 7] is used instead, making the following assumptions: The scatterers are randomly distributed in the sample and their distance is large compared to the wavelength. Only those photons contribute to the coherent signal, which reach the aperture of the detector after a single scattering event. The Stokes-vektor remained unchanged in this backscattering. In compliance, with these requirements we get for the mean power with the frequency f D P(z) /P 0 = 2 0.4 0.6 R c CF = (1 (z f)/f) 2 + ((z f)πr 2 /λf 2 ) 2, (4) where r is the radius of the laser beam in the pupil and f is the focal length of the eye. Thus, R and the extinction coefficient µ t were calculated solving the non-linear optimization problem P(z) 2 0.4 0.6 R c z R exp( 2µ t z )/CF γ(2n(z z )/c) 2 dz 2 Min (5) by the use of Powell s conjugate direction set method [9]. Here, P(z)is the measured coherent power versus depth normalized to the power of the incident beam. R can be determined by the integration of the scattering phase function over the aperture angle of the observation. Assuming the Henyey Greenstein phase function [10], we get R = 1 π+ϕ 4 π (1 g 2 )/(1 + g 2 2g cos θ) 2/3 dθ, π (6) from which equation the anisotropy coefficient g of scattering can be calculated if the aperture angle ϕ = arctan(pupil radius/axial length of the eye) is assumed. In OCT-scans of the ocular fundus (Figure 1), peaks resulting from specular reflection at the inner limiting membrane (ILM, border between vitreous and retina) and the retinal pigment epithelium (RPE) are observed. From the height of these peaks, the refractive indices of single layers may be determined using Equation (2) and Fresnel s reflection formula. For small angles of incidence of the laser beam onto cornea and retina, this results in P P 0 = 2 0.4 0.6 (n air n cornea )/(n air + n cornea ) R (exp( 2µ t z )/CF ) γ(2n(z z )/c) 2 dz, (3) (n vitreous n retina )/(n vitreous + n retina ). (7) where R is that fraction of light, which is scattered into the aperture of the detector [8]. The term CF is a correction of the power recorded from a scatterer in the depth z out of the focal plane in a confocal arrangement and is given by Yadlowsky [8] with Results The coherent reflection at a human ocular fundus in vivo is given in Figure 1. This scan was recorded by [ 110 ]

293 Figure 1. Measured OCT-scan through retina, RPE, and choroid as well as the calculated coherent power according to Equation (3) using the parameters given in Table 1. Baumgartner et al. [11] using a dual beam interferometer. The light source was a superluminescent diode (C86142E, EG&G Optoelectronics, Canada) emitting at λ = 855 nm with a coherence length in air of 13 µm. The power of the beam in the pupillary plane was about 200 µw at a cross section of 0.6 0.8 mm 2. The coherence function γ was determined from the reflex of a glass plate. Unfortunately, the data of Baumgartner et al. [11] were not given in power units. Therefore, it was impossible to calculate refractive indices. Equation (7) was employed for scaling the reflectance data using the refractive indices of Gullstrand s eye model [12]: n cornea = 1.376, n vitreous = 1.336, and n retina =1.35. R and the extinction coefficient µ t were determined solving the optimization problem Equation (5) and the scattering anisotropy g was obtained from Equation (6) for the retina (34.33 34.57 mm from the cornea) and for the choroids (34.71 35.01 mm from the cornea). In the calculation for the choroid, the incident power was corrected for the extinction e µ t z of the retina. The reflected coherent power according to the single backscattering model Equation (3) is shown in Figure 1 and all data are given in Table 1. To obtain an error measure (Table 1), we calculated the variation of one parameter ( R or µ t ) needed to obtain a value of the RMS deviation between measurement and model twice as high as the minimum, while the other parameter was kept constant at it s optimal value. Discussion The results in Figure 1 and Table 1 show that the single backscattering model, first employed in the interpretation of OCT-scans by Schmitt et al. [3], is applicable to scans from the ocular fundus too. However, there is one complication: Schmitt et al. [3] assumed the coherence length of the light source to be small compared to the mean free pathlength of the photons in the sample 1/µ t. This allowed them to replace the coherence function γ by a Dirac delta function and, therefore, to avoid the solution of the integral in Equation (3). As can be seen from the values of µ t in Table 1 and from the coherence length (13 µm), this is not possible in our case. Therefore, we had to solve the integral numerically. Comparing the data in Table 1 with in vitro measurements from bovine retina and porcine choroids [13], we found complete agreement of the scattering anisotropy. However, the value of the extinction coef- [ 111 ]

294 Table 1. Extinction coefficient and scattering anisotropy of retina and choroids µ t µ t R R g g Retina 12 mm 1 +120 mm 1 0.0000407 +0.000061 0.97 0.037 10.8 mm 1 0.0000407 +0.031 Choroids 27.5 mm 1 +154 mm 1 0.000146 +0.000292 0.90 0.159 22 mm 1 0.000146 +0.104 ficient, which should be comparable to the scattering coefficient since absorption is nearly negligible at 855 nm, is about as double as high in the in vitro measurement than measured by OCT. This may be due to morphological differences between human and bovine or porcine tissues or, more likely, to rapid post mortem changes enhancing the opacity of the tissues. On the other hand, the differences between in vivo (OCT) and in vitro measurements are much smaller than the huge errors of the OCT-measurement resulting from the noise of the OCT signal. The signal to noise ratio is principally limited by the photon shot noise. Additionally, the thermal noise of the detector and the electronic noise of the amplifier have to be considered. These broadband noise may be reduced by filtering the signal with the Doppler frequency f D, whereas the bandwidth of the filter has to be at least 2v/ if v is the speed of the reference mirror and a depth resolution shall be achieved [14]. Furthermore, the coherent signal is affected by speckle. This could be reduced by an incoherent averaging over measurements at different wavelengths or by a local averaging which, however, reduces the resolution. Therefore, Schmitt et al. [15] used a detector array in order to reduce the speckle by averaging over measurements under slightly different angles. A further enhancement of the signal to noise ratio is possible if the observed depth plane z is always confocal to the detector. This requires a synchronization of the shift of the reference mirror with that of the focus. A solution of this problem is published by Lexer et al. [16]. The theory of single backscattering assumes complete loss of coherence after the second scattering event. The influence of multiple scattered light was estimated by Schmitt et al. [7] calculating the effective detector cross sections for single and multiple scattered light. For a specimen with optical properties similar to that reported here, they found a remarkable contribution of multiple scattered light for depths greater than 0.3 mm. Since the thicknesses of the retina and the choroid are smaller, the single scattering model can be regarded as sufficient for the description of the coherent signal. Despite all problems mentioned above, the measurement of the extinction coefficient and of the anisotropy of scattering has shown to be possible. To get reliable values for the extinction coefficient, the signal to noise ratio of the measurement has to be improved substantially. The determination of the refractive indices should be possible too if the coherent reflection is measured in power units and is normalized to the incident power. Acknowledgement The authors would like to thank Prof. Fercher and his team for providing the data of the OCT-scan shown in Figure 1. References 1. Huang D, Swanson EA, Lin CP, Schuman JS, Stinson WG, Chang W, Hee MR, Flott T, Gregory CA, Pulafito CA, Fujimoto JG. Optical coherence tomography. Science 1991: 254: 1178 1181. 2. Danielson BL, Whittenburg CD. Guided wave reflectometry with micrometer resolution. Appl Opt 1987; 26: 2836 2842. 3. Schmitt JM, Knüttel A, Bonner RF. Measurement of optical properties of biological tissues by low-coherence reflectometry. Appl Opt 1993; 32: 6032 6042. 4. Fercher A F, Mengedoht K, and Werner W (1988) Eye length measurement by interferometry with partially coherent light. Optics Letters 13: 186-189 5. Hitzenberger CK. Optical measurement of the axial eye length by laser Doppler interferometry. Invest Ophthalmol Vis Sci 1991; 32: 616 624. 6. Fercher AF. Optical coherence tomography. J Biomed Opt 1996; 1: 157 173. 7. Schmitt JM, Knüttel A, Grandjbakhche A, Bonner RF Optical characterization of dense tissue using low-coherence interferometry. Proc SPIE 1993; 1889: 197 211. 8. Yadlowsky MJ, Schmitt JM, Bonner RF. Multiple scattering in optical coherence microscopy. Appl Opt 1995; 34: 5699 5707. [ 112 ]

295 9. Brent RP. Algorithms for Minimization without Derivatives. Prentice-Hall, Englewood Cliffs, 1973. 10. Henyey LG, Greenstein JL. Diffuse radiation in the galaxy. Astrophys J 1941; 93: 70 83. 11. Baumgartner A, Hitzenberger CK, Sattmann H, Drexler, W, Fercher F. Signal and resolution enhancement in dual beam optical coherence tomography of the human eye. J Biomed Opt 1998; 3: 45 54. 12. Naumann GOH. Pathologie des Auges, Bd. 1, Springer- Verlag, Berlin, Heidelberg, New York, 1997. 13. Hammer M, Roggan A, Schweitzer D, Müller G. Optical properties of ocular fundus tissues an in vitro study using the double-integrating-spere technique and inverse Monte Carlo simulation. Phys Med Biol 1995; 40: 963 978. 14. Yadlowsky MJ, Schmitt JM, Bonner RF. Contrast and resolution in the optical-coherence microscopy of dense biological tissue. SPIE Proc 1995; 2387: 193 203. 15. Schmitt JM. Array detection for speckle reduction in optical coherence microscopy. Phys Med Biol 1997; 42: 1427 1439. 16. Lexer F, Fercher AF, Sattmann H, Drexler W, Molebny S. Dynamic coherent focus for transversal resolution enhancement of OCT. Proc SPIE 1998; 3251: 85 90. [ 113 ]