A Method for the Low-Level (ng/g) Determination of Perfluorooctanoate in Carpet by LC MS MS Using Matrix Extracted Standards

Similar documents
Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Supporting Information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Fast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

THE IMPLEMENTATION OF A SCREENING WORKFLOW FOR ION MOBILITY QUADRUOPOLE TIME-OF-FLIGHT MASS SPECTROMETRIC ANALYSIS OF PFOS ISOMERS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Impact factor: 3.958/ICV: 4.10 ISSN:

Fast and Accurate Quantitation of Perfluorinated Sources from Textiles using Gas Chromatography-Triple Quadrupole Mass Spectrometry

Sulfotepp impurities in Chlorpyrifos EC formulations

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Development and validation a RP-HPLC method: Application for the quantitative determination of quetiapine fumarate from marketed bulk tablets

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

DEVELOPMENT OF PFCs HIGH SENSITIVITY ANALYSIS METHOD APPLIED RETENTION GAP TECHNIQUE WITH UPLC/MS/MS Tatsuya Ezaki 1, Nobutake Sato 1, Jun Yonekubo 1,

Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass Spectrometry

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

CHAPTER 19 PARACETAMOL + IBUPROFEN

Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

Appendix II- Bioanalytical Method Development and Validation

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

SUPPORTING INFORMATION

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

EVOLUTE WCX Columns for Solid Phase Extraction of Quaternary Amines and Strongly Basic Compounds from Aqueous Samples

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

International Journal of Pharmacy and Pharmaceutical Sciences Vol 2, Issue 1, 2010

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Received: ; Accepted:

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

8 Laboratory of Environmental Pollution Control and Remediation Technology, Sun

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

CHAPTER - 3 ANALYTICAL PROFILE. 3.1 Estimation of Drug in Pharmaceutical Formulation Estimation of Drugs

Determination of Polyfluoroalkyl Substances (PFAS) in Food Products

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Supporting Information. Sweetened Swimming Pools and Hot Tubs

Environmental Mass Spectrometry DR. RALPH N. MEAD

Dual Use of a Chemical Auxiliary: Molecularly Imprinted Polymers for the Selective Recovery of Products from Biocatalytic Reaction Mixtures

Analytical Method Development and Validation of Lafutidine in Tablet dosage form by RP-HPLC

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

H 3 CO H 3 CO S CH 3

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

11 APPENDICES Appendix A. Analytical Method Dicamba on Filter Paper

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Development of Validated Analytical Method of Mefenamic Acid in an Emulgel (Topical Formulation)

Volume 6, Issue 2, January February 2011; Article-015

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

DEVELOPMENT AND VALIDATION OF NEW RP-HPLC METHOD FOR THE DETERMINATION OF AFLOQUALONE IN HUMAN PLASMA AND FORMULATION

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Supporting Information. Detection and Occurrence of Chlorinated By-products of Bisphenol A, Nonylphenol and

Method Development and Validation for the Estimation of Darunavir in Rat Plasma by RP-HPLC

Determination of trace anions in concentrated hydrofluoric acid

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEY WORDS. Simultaneous high-throughput screening of melamine (MEL) and cyanuric acid (CYA).

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

TECHNICAL BRODIFACOUM

Transcription:

A Method for the Low-Level (ng/g) Determination of Perfluorooctanoate in Carpet by LC MS MS Using Matrix Extracted Standards Karen L Empereur 1, *, Marilyn Stadalius 1, Yongdong Zhu 2, Bashir A Mansoori 2, Tsuguhide Isemura 3, Mary A. Kaiser 4, Wolfgang Knaup 5, and Masahiro Noguchi 6 1 Critical Path Services, 3521 Silverside Rd, Wilmington, DE 19810; 2 Quest Pharmaceutical Services, 3 Innovation Way, Suite 240, Newark, DE 19711; 3 Asahi Glass Research Center, 1150 Hazawa-cho, Kanagawa-ku, Yokohama-shi Kanagawa 221-8755, Japan; 4 E.I. du Pont de Nemours & Co., Wilmington, DE, 19880-0402; 5 Clariant GmbH, Werk Gendorf R & D Fluortelomers, Burgkirchen 84504, Germany; 6 Daikin Industries, LTD, 1-1 Nishi Hitotsuya, Settsu-Shi, Osaka 566-8585, Japan Abstract Fluorotelomer-based acrylic polymers are applied to the surface of carpet to impart oil, stain, and water repellence properties. Concerns that fluorotelomer-based polymers are a possible source of low level exposure to humans, coupled with their widespread use have prompted the need to develop a method to detect and measure perfluorooctanoate (PFO) in carpet. A liquid chromatography tandem mass spectrometry method for the determination of PFO in carpet using a dual labeled 13 C- perfluoroctanoic acid ( 13 C-PFOA) internal standard is successfully developed and validated. Levels of PFO are determined using a gradient, reversed-phase high-performance liquid chromatography (HPLC) method with acetic acid acidified water methanol, separated on a 50 mm Phenomenex Synergi Polar RP column. Ions monitored are 413 (parent) and 369 (daughter) for PFO and 415 (parent) and 370 (daughter) for dual labeled 13 C-PFOA internal standard. Accuracy and precision over three days for 5 to 900 ng/g PFO in carpet ranged from 2.4% to 7.6% and 3.7% to 14.1%, respectively. Overall extraction efficiency for samples (n = 30) fortified with 13 C-PFOA at 20 ng/g and perfluorooctanoic acid (PFOA) at 5, 50, and 500 ng/g is 98.9% ± 8.1%. Specificity of the method was evaluated with two different carpet samples. Introduction Recent studies indicate potential exposure of the general human population to perfluorooctanoic acid (PFOA) at very low levels (1 7). Other reports showed those low levels of PFOA and other fluorinated compounds could be found in wildlife and in the environment (8,9). Hence, interest in understanding more about the source and mode of exposure of these materials in humans and in the environment has increased. Fluorotelomer-based acrylic polymers are applied to the surface of carpet fibers to impart oil and water repellency properties (10,11). Small quantities of PFOA may be generated as an unintended by-product in the production of the polymers (12 16). * Author to whom correspondence should be addressed. Although fluorotelomer-based polymers are not made using PFOA nor is PFOA added during the manufacture or use of telomer products, questions have arisen as to the possibility of trace level PFOA impurities in telomers and the potential for telomers to transform into PFOA (17). A method to determine extractable perfluorooctanoic acid (PFOA) in water, sweat simulant, saliva simulant, and methanol from textile and carpet samples by liquid chromatograpy tandem mass spectrometry (LC MS MS) has been developed (18). Low level PFOA determinations are particularly challenging as carpet fiber and backing is a complex matrix. Most carpet today is tufted. Tufted carpet consists of face yarn, primary backing fabric, a bonding compound such as latex, polyvinylchloride, or polyurethane, and often a secondary backing fabric. With so many components in carpet samples, the MS MS signal can either be enhanced or attenuated, depending upon what other compounds might be co-eluting with the analyte. Also background PFOA exceeding the limit of quantitation (LOQ) of 5 ng/g can be observed. Fortification of untreated carpet samples was used to determine if any major problems are evident in recovery of the analyte from the matrix. Ideally an isotopically enriched surrogate might be added during the manufacturing process, but is not practical. PFOA contamination can originate from solvents, labware, lab contamination, and even from the components of standard HPLC instrumentation. Extraction and injection solvents as well as labware must be evaluated for contamination by routinely running reagent blanks; contamination from work areas must be minimized by washing these areas extensively with methanol, isopropanol, or other appropriate solvent. Experimental Reagents Pure analytical perfluorooctanoic acid (CAS # 335-67-1, Catalog # 001319, chemical purity 97%) was purchased from Oakwood Products, Inc (West Columbia, SC) and perfluoro- 632 Reproduction (photocopying) of editorial content of this journal is prohibited without publisher s permission.

octanoic acid (1,2-di- 13 C) PFOA, 100% chemical purity was obtained from DuPont Haskell Laboratory (Newark, DE). Methanol (HPLC grade) and acetic acid were purchased from Fisher Scientific. High purity water was prepared using an NEU- ION system (Baltimore, MD). It is necessary to check the methanol for the presence of contaminants by LC MS MS before use since certain lots have been found to be unsuitable for use. Carpet Untreated tufted carpet was provided by E.I. du Pont de Nemours and Company. Table I. HPLC Gradient Time (min) % A % B Flow (ml/min) 0.0 70 30 0.3 1.0 70 30 0.3 3.0 0 100 0.3 5.0 0 100 0.3 5.1 70 30 0.3 Table II. MS Parameters Ionization Declustering Collision Compound mode potential energy Transition PFO ESI 30 20 412.7 368.8 13 C-PF (IS) ESI 30 20 415.0 369.9 Table III. PFO Recovery in Carpet Over Three Days Interday Intraday Sample PFO average Interday average Intraday description (ng/g)* recovery (%) S.D. recovery (%) S.D. Day 1 5.00 97.8 6.79 Day 2 5.00 106 3.82 97.8 6.79 Day 3 5.00 120 24.4 Day 1 15.0 101 11.3 Day 2 15.0 95.7 1.41 101 11.3 Day 3 15.0 114 0.94 Day 1 150 104 0.94 Day 2 150 108 1.89 104 0.94 Day 3 150 99.0 1.41 Day 1 900 101 3.85 Day 2 900 104 4.40 101 3.85 Day 3 900 102 5.58 * n = 2 for each day at each concentration. Equipment A Shimadzu HPLC equipped with a binary gradient pump and autosampler was used. An Applied Biosystems, Inc. triplequadrupole mass spectrometer (API 3000) with Sciex Turbo Ion Spray Liquid Introduction Interface (MDS Sciex, Concord, Ontario, Canada) was used as the detector. Negative ions were monitored in MRM mode; 413 (parent) to 369 (daughter) for PFOA and 415 (parent) to 370 (daughter) for dual 13 C-PFOA. System control, data acquisition, and processing were done by Analyst software (Applied Biosystems, Inc.). The results were calculated by 1/x 2 weighted linear regression analysis using Watson LIMS software (v6.4.0.04 InnaPhase Corporation). The reversed-phase gradient separation was accomplished by injecting 3 µl samples onto a Phenomenex Synergi Polar RP, 2 mm i.d. 50 mm, 4 µm, column (Torrance, CA) maintained at ambient temperature. The mobile phase comprised of A: water acetic acid (100:0.1, v:v) and B: methanol acetic acid (100:0.1, v:v) using the following gradient (Table I). The sample solvent was methanol water (50:50, v:v). The MS MS parameters were optimized to transmit the parent ions, fragment them and monitor the daughter ions. Ions monitored were: 413 (parent) and 369 (daughter) for PFOA; 415 (parent) and 370 (daughter) for dual 13 C-PFOA. Negative ions were monitored in MRM mode (Table II). Extraction procedure Samples of approximately 10 grams of carpet with backing (approximately 9 9 cm piece) were cut, weighed and placed into glass beakers. The pieces of carpet were spiked (if necessary) with 1.0 ml aliquots of the calibration standard solutions or the fortification solutions. The carpet pieces were also spiked with 1.0 ml of internal standard solution. The spiked carpet pieces were allowed to sit for ~15 min to allow the solvent to dry. The spiked carpet was extracted by adding 200 ml of methanol to the carpet in the beakers. The beakers were covered and allowed to shake on an Eberbach reciprocal shaker for ~15 min; after shaking, the beakers were placed in an ultrasonic water bath for ~30 min. An aliquot of 5.0 ml of the methanol extract from each beaker was transferred into separate glass test tubes. The 5 ml extracts were then evaporated with nitrogen to dryness. The residue was reconstituted with 2.0 ml methanol water (50:50, v:v). The Table IV. Interday Accuracy and Precision of PFO in Carpet Over Three Days PFOA Fortification Levels in Carpet Sample description 5.00 ng/g 15.0 ng/g 150 ng/g 900 ng/g Day 1 5.13 16.4 156 886 4.65 14.0 158 935 Day 2 5.14 14.5 165 908 5.41 14.2 161 964 Day 3 6.85 17.1 150 955 5.12 17.3 147 884 3 Day Mean 5.38 15.6 156 922 S.D. 0.759 1.52 6.74 34.5 % Coefficient 14.1 9.7 4.3 3.7 of variance % Difference 7.6 4.0 4.0 2.4 n 6 6 6 6 633

reconstituted samples were centrifuged before transferring to a glass microvial for LC MS MS analysis. Standards and fortification solutions A stock standard solution of PFOA was prepared at a concentration of 1123.6 µg/ml by dissolving 11.584 mg of the standard (corrected for purity) in methanol. A stock internal standard solution of 13 C-PFOA was prepared at a concentration of 1110.6 µg/ml by dissolving 11.106 mg of the standard (corrected for purity) in methanol. Calibration standard solutions from 50 to 10,000 ng/ml were prepared in methanol water (50:50, v:v). A 200 ng/ml 13 C-PFOA internal standard solution was prepared by dilution of the stock solution with methanol water (50:50, v:v). Fortification solutions of 9,000, 1,500, 150, and 50 ng/ml PFOA were prepared by diluting the stock solution with methanol water (50:50, v:v). Figure 1. Representative chromatogram of the PFO peak (MRM 415.0 to 368.8) for lowest extracted standard corresponding to 5 ng/ml in carpet (0.625 ng/ml). Standards for analysis were prepared by spiking seven (10 g) untreated carpet pieces on each day of analysis with 1.0 ml of each of the seven calibration standards, respectively. The carpets were also spiked with 1.0 ml of the internal standard solution. The concentration levels of PFOA on the carpet ranged from 5.0 (LOQ) to 1000 ng/g. The concentration level of the internal standard was 20 ng/g. The carpet pieces were processed through the extraction procedure, and the resultant solutions used as calibration standards. The fortified carpet samples were prepared on each of 3 successive days by spiking duplicate (10 g) untreated carpet samples with the 1.0 ml of each of the fortification solutions to give concentrations on the carpet at the proposed LOQ (5 ng/g), at 3 the proposed LOQ (15 ng/g), at 30 the proposed LOQ (150 ng/g) and at 180 the proposed LOQ (900 ng/g). The carpets were also spiked with 1.0 ml of the internal standard solution. The carpet pieces were processed through the extraction procedure and the resultant solutions were used for determining recovery, precision and accuracy. Matrix-only samples (methanol extract of carpet samples where the PFOA was added to the methanol after extraction) were prepared for extraction efficiency and matrix effect assessments. Untreated carpet samples (10 g) were extracted with 200 ml of methanol. A 5.0-mL aliquot of supernatant was spiked with 25 µl of standard solutions to give concentrations equivalent to 50, 500, or 5000 ng/ml. A 25-µL aliquot of internal standard solution was added as well. Samples were evaporated with nitrogen to dryness, the dried sample reconstituted with 2.0 ml methanol water (50:50, v:v) and centrifuged before transfer to a glass microvial for LC MS MS analysis. Neat samples (without carpet matrix) were prepared for extraction efficiency and matrix effect assessment. A 5.0-mL aliquot of methanol was added to each glass test tube. Respective tubes were spiked with 25 µl of solutions at concentration of 50, 500, or 5000 ng/ml, then each tube was spiked with 25 µl of internal standard. Samples were evaporated with nitrogen to dryness, the volume reconstituted with 2.0 ml methanol water (50:50, v:v) and samples centrifuged before transfer to a glass microvial for LC MS MS analysis. Results and Discussion Figure 2. Representative chromatogram of the mobile phase blank (MRM 415.0 to 368.8). Chromatographic performance Quantitation of PFO, having a retention time of ~3.0 min, was accomplished by tur- 634

bospray LC MS MS analysis. Excellent sensitivity was accomplished at the lowest standard concentration of 5.0 ng/g (which corresponds to a final concentration of 0.625 ng/ml) in carpet extract (Figure 1). To maintain good peak shape and consistent retention, standards and samples were injected in a solvent system with 50% water, and the injection volume did not exceed 3 µl. Total analyses time, including washing, re-equilibration, and injection steps was approximately 10 min. In this method, carryover and contamination from injection solvents were evaluated by routinely running mobile phase blanks (Figure 2). Linearity The linearity of PFO for each analysis set was determined using standard curves in carpet extracts obtained from the peak area ratio between the native analyte and its dual labeled 13 C analog (internal standard) at a minimum of seven concentrations, including the LOQ. Untreated carpet was found to contain small amount of PFOA, the source of which is uncertain. In order eliminate the contribution of the background amount the mean response of the three blank plus internal standard samples was subtracted from responses of all samples. Results showed a linear fit from 5.00 to 1,000 ng/g for PFO. Correlation coefficients of r 2 0.994 were readily achievable. Representative linear regression equation with 1/x 2 weighting (Figure 3) Matrix effect The matrix effect study evaluated the suppression or enhancement of the analyte and internal response by the matrix. The matrix effect was determined in carpet fibers with backing at 5, 50, and 500 ng/g PFO and at 20 ng/g internal standard. The neat solutions served as the reference samples. Matrix suppression, determined by percent difference from the solvent peak areas, ranged from 7.6% to 12.0%. To ensure that samples could be diluted with blank matrix without affecting the calculated concentration, a carpet sample fortified at 200 ng/g was diluted by a factor of ten and five replicates were analyzed. The difference between the mean concentration of the diluted replicates and the nominal concentration was 7.7%. Specificity Specificity was evaluated by extracting and analyzing two lots of carpet fortified with PFOA and internal standard (10 ng/g). Analyte values (11.477 ± 0.256 ng/g) were within the acceptable range. Recovery, precision, and accuracy The method recovery, precision, and accuracy were assessed by measuring the concentrations found in duplicate carpet samples fortified at four levels over three days (Table III). Interday recovery at all concentration levels ranged from 95.7% to 120%. Interday precision, calculated using the coefficient of variation (CV), ranged from 3.7% to 14.1%. Interday accuracy, determined by the percent difference from the nominal concentration, ranged from 2.4% to 7.6% (Table IV). Extraction efficiency The efficiency of the liquid liquid extraction process was determined comparing carpet extract solutions fortified with 5, 50, and 500 ng/g PFO and at 20 ng/g internal standard to reference samples spiked with an equivalent amount of analyte. The results for recovery of PFO ranged from 96.3% to 110.7% and recovery of internal standard ranged from 96.3% to 98.0%. Conclusion Figure 3. Representative curve for PFO corresponding to concentrations of 5.00 to 1000 ng/g in carpet extract. Linear regression equation with 1/x 2 weighting: y = 0.0402x + 0.0124. Reproducible good peak shape and good gradient retention in carpet extracts are readily obtained with this LC MS MS methodology without significant sample clean-up. Excellent sensitivity of this method at 5 ng/g is achieved by the use of a gradient separation and by drying and reconstituting samples in smaller volumes of solvent. Because PFO contamination can originate from work areas, solvents, labware, and HPLC components, low level determinations are particularly challenging as background PFO can exceed the LOQ of 5 ng/g. The method was validated by extracting duplicate carpet samples fortified at the proposed LOQ (5 ng/g), at 3 the proposed LOQ (15 ng/g), at 10 the proposed LOQ (50 ng/g) and at 180 the proposed LOQ (900 ng/g) on three successive days. The efficiency of the extraction procedure was validated by comparing peak areas of five replicates at the LOQ (5 ng/g), at 10 the proposed LOQ (50 ng/g), and at 100 the proposed LOQ (500 ng/g) with extracts fortified at equivalent concentrations. Based on the data, accuracy, precision, repeata- 635

bility, recovery and specificity of the method were established. Because PFO was detected in untreated samples, interday precision, ranging from 3.7% to 14.1% and interday accuracy ranging from 2.4% to 7.6% are well within the acceptance criteria for the method. Overall extraction efficiency ranging from 92.4% to 110.7% with no clean-up is quite impressive at these low levels. Acknowlegments The authors thank Barbara S. Larsen of E.I. du Pont de Nemours and Company, Wilmington, DE, USA for providing high quality technical input during the course of this work and the authors also acknowledge the Telomer Research Program, for without their support and funding, this work would not have been conducted. References 1. K.J. Hansen, L.A. Clemen, M.E. Ellefson, and H.O. Johnson. Compound-specific, quantitative characterization of organic fluorochemicals in biological matrices. Environ. Sci. Technol. 35: 766 70 (2001). 2. K. Kannan, S. Corsolini, J. Falandysz, G. Fillmann, K.S. Kumar, B.G. Loganathan, M.A. Mohd, J. Olivero, N. van Wouwe, J.H. Yang, and K.M. Aldous. Perfluorooctanesulfonate and related fluorochemicals in human blood from several countries. Environ. Sci. Technol. 38: 4489 95 (2004). 3. G.W. Olsen, K.J. Hanen, L.A. Stevenson, J.M. Burris, and J.H. Mandel. Human donor liver and serum concentrations of perfluorooctanesulfonate and other perfluorochemicals. Environ. Sci. Technol. 37: 888 91 (2003). 4. C. Sottani and C. Minoia. Quantitative determination of perfluorooctanoic acid ammonium salt in human serum by high-performance liquid chromatography with atmospheric pressure chemical ionization tandem mass spectrometry. Rapid Commun. Mass Spectrom. 16: 650 54 (2002). 5. Z. Kuklenyik, J.A. Reich, J.S. Tully, L.L. Needham, and A.M. Calafat. Automated solid-phase extraction and measurement of perfluorinated organic acids and amides in human serum and milk. Environ. Sci. Technol. 38: 3698 3704 (2004). 6. C. Kubwabo, N. Vais, and F.M. Benoit. A pilot study on the determination of perfluorooctanesulfonate and other perfluorinated compounds in blood of canadians. J. Environ. Monit. 6: 540 45 (2004). 7. M.M. Schultz, D.F. Barofsky, and J.A. Field. Fluorinated alkyl surfactants. Environmental Engineering Science. 20(5): 487 501 (2003). 8. J.P. Giesy and K. Kannan. Global distribution of perfluorooctane sulfonate in wildlife. Environ. Sci. Technol. 35: 1339 42 (2001). 9. J.W. Martin, M.M. Smithwick, B.M. Braune, P.F. Hoekstra, D.C.G. Muir, and S.A. Mabury. Identification of long-chain perfluorinated acids in biota from the Canadian arctic. Environ. Sci. Technol. 38: 373 80 (2004). 10. Organofluorine Chemistry: Principles and Commercial Applications. R.E. Banks, B.E. Smart, and J.C. Tatlow, Eds. First Edition. Plenum Press, New York, 1994. 11. E. Kissa. Fluorinated Surfactants and Repellents. 2nd Ed. Marcel Dekker, New York, 2001. 12. B. Huang and F. Wu. A new method of the synthesis of polyfluoralkyl carboxylic acids from polyfluoroalkyl iodides. Youji Huaxue. 13: 403 404 (1993). 13. C. Dapremont-Avignon, P. Calas, A. Commeyras, and C. Amatore. Synthesis of perfluoroalkyl carboxylic acids by reaction of perfluoroalkyl iodides with electrogenerated superoxide ion. J. Fluorine Chem. 51: 357 59 (1991). 14. S. Benefice-Malouet, H. Blancou, P. Calas, and A. Commeyras. Synthese d acides perfluoroalcane carboxylique et sulfinique par reduction electrochimique d iodures de perfluoroalkyle sur cathode en fibres de carbone dans le solvant N,N-dimethylformamide. Application a la synthese de perfluoro α,ω diacides. J. Fluorine Chem. 39: 125 40 (1988). 15. B. Huang, A. Haas, and M.A. Lieb. A new method for the preparation of perfluorocarboxylic acids. J. Fluorine Chem. 36: 49 62 (1987). 16. C. Hu and Z. Xu. A new method for the synthesis of perfluorocarboxylic acids from perfluoroalkyl iodides. Huaxue Xuebao. 48: 936 38 (1990). 17. M.F. Dominiak. PFOS/PFOA: EPA Update on Perfluorinated Compounds, Synthetic Organic Chemical Manufacturers Association, Presentation, March 2004. 18. M.P. Mawn, R.G. McKay, T.W. Ryan, B.Szostek, C.R. Powley, and R.C. Buck. Determination of extractable perfluorooctanoic acid (PFOA) in water, sweat simulant, saliva simulant, and methanol from textile and carpet samples by LC/MS/MS. Analyst. 130: 670 78 (2005). Manuscript received March 16, 2007; revision received July 10, 2007. 636