Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies

Similar documents
Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Accurate Determination of TiO 2 Nanoparticles in Complex Matrices using the Agilent 8900 ICP-QQQ

Agilent 8900 Triple Quadrupole ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Agilent 8900 Triple Quadrupole ICP-MS LEAVE INTERFERENCES BEHIND WITH MS/MS

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Single particle analysis using the Agilent 7700x ICP-MS

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Determination of Nanoparticle Sizeand Number Concentration using the npquant Evaluation Module for Qtegra ISDS. The world leader in serving science

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

Characterization of nanoparticles in aqueous samples by ICP-MS

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2.

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology

Agilent ICP-MS Journal December 2016 Issue 67

Sample Analysis Design PART II

Simultaneous quantification of peptides and phosphopeptides by caplc-icp- MS using the Agilent 8800/8900 Triple Quadrupole ICP-MS

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Introduction to the New Agilent 7900

Single Nanoparticle Analysis of Asphaltene Solutions using ICP-QQQ

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Nanoparticle Analysis: An Upcoming Challenge for Environmental Analysis

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Multi-residue analysis of pesticides by GC-HRMS

ICP-MS. plasma 17OO. High Resolution Multi-Collector ICP-MS.

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Sample Analysis Design PART III

Agilent 8800 Triple Quadrupole ICP-MS: Understanding oxygen reaction mode in ICP-MS/MS

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Fergus Keenan Shona McSheehy Julian Wills

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

ICP-Mass Spectrometer

Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Accurate Sizing and Precise Counting of 10 nm Gold Nanoparticles using the Enhanced Sensitivity of the NexION 2000 ICP-MS

A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Concentration Measurements by Single Particle ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Very recently Roussis and Proulx clearly demonstrated

Powerful Scan Modes of QTRAP System Technology

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Agilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012

Agilent Technologies at TIAFT 2013

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Defining quality standards for the analysis of solid samples

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Quantitative characterization of silica nanoparticles by asymmetric flow FFF MALS-ICP-QQQ using the Agilent 8800 ICP-QQQ

Sources of Errors in Trace Element and Speciation Analysis

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Secondary Ion Mass Spectroscopy (SIMS)

Autenticity control, provenance testing and fraud detection using mass spectrometry

ATOMIC SPECTROSCOPY APPLICATIONS IN THE CONTRACT ENVIRONMENTAL LABORATORY. Primer July 2015 Edition

Sample Analysis Design Internal Standard

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Advantages of ICP-QQQ for nuclear waste characterisation

Analytical Approaches to Arsenic

Detection of engineered cerium oxide nanoparticles in soils

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

GC-APCI IMS of Diesel

Formation of iron oxide nanoparticles for the photooxidation of water: Alteration of finite size effects from ferrihydrite to hematite

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Finnigan LCQ Advantage MAX

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

ABI 3200 Q TRAP LC/MS/MS System

Analysis for Arsenic and Hexavalent Chromium:

Information Dependent Acquisition (IDA) 1

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Low Level Detection of Priority Inorganic Pollutants in Water. Anthony Neumann, MSc Darren Crichton, BSc

TOMAHAQ Method Construction

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Agilent All Ions MS/MS

Identifying Disinfection Byproducts in Treated Water

Transcription:

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software Alain Desprez Agilent Technologies 1

The Total Solution Choice of Mass Spectrometers 7800 quadrupole ICP-MS cost-effective, 3 ms dwell time 7900 quadrupole ICP-MS ultra-sensitive, ultra-fast 0.1 ms dwell time 8900 ICP-QQQ highest sensitivity and lowest background, MS/MS for interference-free determination of SiO 2 and TiO 2 nanoparticles Fully automated acquisition and data analysis configuration within MassHunter 4.3 software guided by Method Wizard Supports single particle mode with 1 or 2 elements and field-flow fractionation (FFF) coupling 7800 ICP-MS 7900 ICP-MS 8900 ICP-QQQ 2

Signal intensity (cps) Single Particle-ICP-MS (spicp-ms) Uses normal sample intro system and TRA data collection method If sample is sufficiently dilute, each nanoparticle gives a distinct transient signal Particle concentration, particle size (diameter), size distribution and composition are obtained Signal from one nanoparticle event Nanoparticle sample time (sec) 3

Method Wizard now supports nano analysis by both snp and FFF modes 4

Method Wizard now supports nano analysis by both snp and FFF modes Most parameters are automatically entered after providing a couple of simple inputs such as sample tube ID and Reference Material Name 5

Just Complete the Sample List, Add it to the Queue And Review the Results 6

100µs integration time and no settling time with the Agilent 8900 and 7900 captures all the data Zoom in 7

Raw Data to Right Results in Seconds Composite Sample 60nm (NIST 8013) 50 ng/l + 30nm (NIST 8012) 5 ng/l Observed concentration (particles/l) Observed concentration (ng/l) Prepared total concentration (ng/l) Recovery 4.78 x10 7 57.6 55 105% Reference particle size by TEM (nm) Observed particle size (nm) Prepared particle number ratio (%) Observed particle number ratio (%) 56.0±0.5 58 55.6 57.8 27.6±2.1 28 44.4 42.2 8

Determining Really Small Nanoparticles What are the limitations and how can ICP-MS/MS solve them? There are only 2 things that limit the minimum size nanoparticle that can be detected in single particle mode. 1. Sensitivity 2. Background (raw instrument background, spectral interferences, dissolved ionic background) Everything else is secondary. Nanoparticle size (diameter) is a function of the cube root of the mass. If the diameter of a particle is reduced by half, the mass and therefore the signal is reduced by 8X. HIGH SENSITIVITY IS THE CRITICAL FACTOR FOR DETECTING SMALL PARTICLES. Background is also critical because nanoparticle peaks are detected as x times the signal to noise. Small peaks can be lost in the noise. Agilent application note 5991-6596 9

Challenge of Si or Ti Measurement by Conventional (Single) Quadrupole ICP-MS Interferences Si is interfered by C, N and O overlaps (ubiquitous in most aqueous/organic samples) Ti is interfered by polyatomic overlaps in P, S, Si and C matrices These interferences make the analysis of SiO 2 or TiO 2 NPs in real samples difficult 48 Ti (74% abundance) cannot be measured reliably by ICP-QMS, due to the 48 Ca isobaric interference Less interfered 49 Ti (5.4% abundance) has less sensitivity, so detection of smaller particles is compromised ICP-MS/MS can solve these problems 10

ICP-MS/MS Technique for Si (On-Mass Method) ICP-MS/MS uses two quadrupoles separated by a collision/reaction cell H 2 gas N 2 H + COH + 14 N 2 + 12 C 16 O + 28 Si + Si+ Q1 28 amu Q2 28 amu N 2+ + H 2 N 2 H + + H CO + + H 2 COH + + H Si + + H 2 No Reaction 11

SiO 2 Nanoparticle Reference Material Results 50 nm Nominal size (nm) Median size (nm) Most Freq. Size (nm) Mean Size (nm) TEM Diameter (1) (nm) BED (2) (nm) 50 49 50 49 46.3 ± 3.1 23 60 59 58 60 57.8 ± 3.5 22 60 nm 100 99 100 100 97.0 ± 4.8 25 200 199 204 199 198.5 ± 10.5 25 100 nm (1) values supplied by nanocomposix (2) Background equivalent diameter ~ size detection limit most likely limited by dissolved Si. 200 nm 12

ICP-MS/MS Technique for Ti (Mass-Shift Method) SO + + O 2 No Reaction SiO + + O 2 No Reaction POH + PO + + H (collision) CO 2+ + O 2 CO 2 + O + 2 Ca + + O 2 CaO + + O CaO + + H 2 CaOH + + H Ti + + O 2 TiO + + O O 2 + H 2 gas SO +, SiO +, PO +, CO 2 CaOH + 32 S 16 O + 30 Si 18 O + 31 P 16 OH + 12 C 18 O + 2 48 Ca + 48 Ti + Q1 m/z 48 Q2 m/z 64 48 Ti 16 O + 13

TiO 2 analysis - Interference Removal by O 2 -H 2 Mode Determining interferences in ionic matrix solutions Quantitative result as Ti (ppb) Mode Sensitivity (cps/ppb) BEC (ppb) P 100 ppm S 100 ppm Ca 50 ppm Si 50 ppm Ethanol 0.1% Matrix mixture No gas 155000 0.016 1.7 6.0 225 0.39 0.14 261 O 2 + H 2 79000 0.001 0.010 0.001 0.18 0.054 0.001 0.023 No gas : Single Quad mode, Q2=48 O 2 + H 2 : MS/MS mode, Q1=48, Q2=64 Advantage of O 2 + H 2 mode Can use the most abundant isotope of Ti (mass 48, 73.7% relative abundance) for highest sensitivity Can remove almost all interferences including 48 Ca 14

TiO 2 NPs in Matrix Samples O 2 -H 2 MS/MS mode Signal Distribution Size Distribution Sunscreen in DI water Mean size : 77 nm BED : 16 nm Sunscreen in tap water Mean size : 79 nm BED : 10 nm Sunscreen in matrix mixture (100 ppm S & P, 50 ppm Ca & Si, 0.1% ethanol) Mean size : 84 nm BED : 22 nm 15

Conclusions The sp-icp-ms/ms method with MS/MS for SiO 2 and TiO 2 nanoparticle analysis: Provides high sensitivity for Si and Ti Effectively eliminates polyatomic ions which interfere with Si or Ti analysis thereby significantly lowering the minimum detectable particle size even in complex matrices Permits the use of major isotopes ( 28 Si, 48 Ti), even in the presence of polyatomic and isobaric ( 48 Ca) overlaps Provides quick and accurate results for SiO 2 and TiO 2 particles smaller than 100 nm 16