High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Similar documents
Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Elemental analysis of river sediment using the Agilent 4200 MP-AES

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Samples compliant with the WATER

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

ICP-OES Application Note Number 35

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Application Note. Abstract. Authors

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Elemental Scientific. Spike Recovery. Elemental Scientific

Single particle analysis using the Agilent 7700x ICP-MS

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Elemental Scientific. Stability. Elemental Scientific

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Elemental Scientific. Spike Recovery. Elemental Scientific

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Trace Analyses in Metal Matrices Using the ELAN DRC II

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Analysis Repeatability of Trace and Major Elements in a Water Sample

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Agilent ICP-MS Journal September 2009 Issue 39

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Application Note. Abstract. Authors. Environmental

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Not Just for Paint Chips Anymore. High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Agilent 8900 Triple Quadrupole ICP-MS

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

On-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Low Level Phosphorus: A Method Comparison Study

Transcription:

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 3380 146th Place SE, Suite 300, Bellevue Washington, 98007 USA Takeo Kuwabara and Tetsushi Sakai Agilent Technologies Tokyo Analytical Division 9-1 Takakura-cho, Hachioji Tokyo 192-0033 Japan Abstract Agilent has further improved the sample throughput capabilities of its 7500cx ICP-MS with Octopole Reaction System (ORS) using a newly configured Integrated Sample Introduction System Discrete Sampling (ISIS-DS) accessory, and helium collision mode. Employing this new methodology, a complete suite of 30 or more elements can be analyzed in compliance with USEPA criteria (spectrum mode, three replicates, and sub-ppb MDLs) in approximately 75 seconds, sample to sample, with excellent removal of polyatomic interferences. Performance data showing stability, interference control, accuracy, precision, and washout are presented. The new system is applicable to labs requiring extremely high sample throughput and with its low sample consumption of ~2.2 ml/sample, for applications where sample volume is limited.

Introduction The Agilent Integrated Sample Introduction System (ISIS) has always permitted the highest sample throughput, using either high-speed uptake with constant flow nebulization or discrete sampling using a six-port valve and sample loop with timeresolved acquisition. In keeping with Agilent's continual focus on product enhancement, ISIS Discrete Sampling (ISIS- DS) has been reconfigured to further improve productivity. The resulting enhanced ISIS-DS sampling mode takes advantage of the ability of the 7500cx ICP-MS to analyze environmental samples using a single collision cell mode (He mode). This new mode of operation permits USEPA-compliant analysis (spectrum mode, three replicates, and sub-ppb MDLs) of a complete suite of 30 or more elements in approximately 75 seconds, sample to sample. ISIS Configuration Figure 1 shows the ISIS configuration used. It is a typical discrete sampling configuration with a couple of important modifications. Pump 1 (P1) is the large ISIS sample uptake peristaltic pump. Pump 2 (P2) is the standard 7500 nebulizer pump. The ISIS uptake pump (P1), which is located downstream of the valve, draws the sample from the autosampler into the sample loop. As a result, the sample loaded in the sample loop is never exposed to peristaltic pump tubing, thereby eliminating a common source of contamination and carryover. This high-speed, high-capacity peristaltic pump is capable of rinsing and filling the sample loop in approximately 10 seconds when using the Cetac ASX-520 autosampler with the wide-bore 0.8 mm id probe. The other modification is the addition of the tee joint between the valve and nebulizer to allow the use of online internal standard addition. By minimizing both the length and diameter of the tubing between the valve and nebulizer, the time from rotation of the valve (sample injection) to the realization of a constant analyte signal is less than 15 seconds. A 300-µL loop is sufficient to allow more than 30 seconds of continuous spectrum mode acquisition. Larger loop sizes can be used to achieve any duration of acquisition required. After acquisition has completed, the valve returns to the load position, flushing any remaining sample to waste and rinsing the nebulizer and spray chamber with clean rinse solution. At this point, approximately 15 seconds is required for the signal to return to baseline in preparation for the next analysis (Figure 2). Figure 1. ISIS-DS sampling with online internal standard configuration. Valve in "load" position on left and in "inject" position on right. 2

Figure 2. Analyte and internal standard profile during discrete sampling analysis (log scale for signal intensity). Time 0 autosampler probe enters the sample and loop is loaded. Inject at 10 seconds, 15 seconds acquisition delay, 30 seconds acquisition, 10 seconds rinse, 10 seconds overhead. Total time is approximately 75 seconds. Experimental A sequence of 216 samples was analyzed in 4 hours 29 minutes using the new ISIS-DS sampling configuration depicted in Figure 1 and the acquisition conditions listed in Table 1. The 7500cx was operated in a single cell mode (He collision mode) resulting in both excellent removal of polyatomic interferences and very fast acquisitions since no cell gas switching or stabilization was required 1. The sequence consisted of a single initial calibration at 0.1, 1, 10, and 100 ppb for all elements, followed by repeated (n = 26) analyses of a block of samples consisting of: 50 ppb calibration check (CCV) NIST 1643e water CCB (blank) USEPA Interference Check Solution A (ICS-A) Blank USEPA ICS-AB (spiked with all analytes at 100 ppb to monitor carryover) Blank Blank 1 This is a key benefit over reaction cell ICP-MS instruments that have to operate in multiple cell modes to cover all analytes. While it is possible to use multiple cell modes with discrete sampling, the resulting acquisition time is significantly lengthened, minimizing the benefits in terms of both run time and matrix exposure. If multiple cell modes are employed using the Agilent Octopole Reaction System, the small cell volume and very rapid gas switching reduce the cost in time and matrix exposure. 3

Table 1. ISIS/7500cx ICP-MS Acquisition Conditions for Spectrum Mode Discrete Sampling Analysis Plasma Robust mode 1550 watts Nebulizer Glass concentric (standard) Number of elements (including internal standards) 31 ORS mode Helium - 4 ml/min (single mode) Integration time per point 0.1 seconds (all elements) Points per peak 1 Replicates 3 Total acquisition time (3 replicates) 29 seconds Loop volume 300 µl Loop rinse and fill time 10 seconds Acquisition delay (after valve rotation to inject) 15 seconds Steady state signal time (before valve rotation to fill again) 30 seconds Results Total Run Time and Sample Consumption The resulting run-to-run time was measured at approximately 75 seconds per sample. Total sample consumption was determined by weighing each sample before and after analysis and was calculated to be 2.2 ml per sample per analysis. The method thus lends itself to samples in which the volume available for analysis is limited, and because small amounts are used, waste disposal costs are reduced. The small sample consumption also permits samples to be automatically reanalyzed by intelligent sequencing if needed from a 10-mL autosampler vial while allowing the ASX-520 to be configured for the maximum possible number of samples. Stability Long-term stability was monitored using internal standards. The abstracted internal standard data are illustrated in Figure 3, and show no downward drift, even after repeated (52 total) injections of ICS-A and ICS-AB. Only 6 Li demonstrated matrix suppression greater than 10% in the highest matrix samples, otherwise internal standard recoveries were within ~ ± 10% for the entire sequence. Calibration stability was monitored by measuring a 50-ppb CCV once in each 8-sample block (Figure 4). USEPA limits for CCV recovery are ± 10%. No CCV failures occurred; in fact, nearly all CCV recoveries were within ± 5% for the entire sequence. Figure 3. Internal standard recoveries compared to calibration blank for all 216 samples. 4

Figure 4. CCV recoveries (50 ppb) for entire sequence. USEPA limits for CCV recoveries in both Method 200.8 and 6020 are ± 10% (shown in red). Accuracy Long-term accuracy and precision were also determined through repeated analysis of NIST 1643e Certified Reference Water (n = 26). Results are tabulated in Table 2, showing recoveries within ± 10% or better of certified values and relative standard deviations near 1% for most elements. Be and Se had higher %RSDs due to the short integration times used and slightly lower count rates for these elements in He mode. Longer integration times can be used if higher counts or better precision are required for these elements. When ultimate DLs for Se are required (low ppt), H 2 mode is recommended. Due to the fast switching time of the ORS, Se can be measured in H 2 mode with only ~30 seconds added to the sample to sample run time. 5

Table 2. Precision (%RSD) and Mean Recovery of NIST 1643e Water for 26 Separate Analyses Mean Certified Mean Mass/ measured RSD value recovery element value (µg/l) (%) (µg/l) (%) 9 Be 13.8 2.5 14.0 101.0 23 Na 22689.2 2.0 20740.0 109.4 24 Mg 7300.3 2.1 8037.0 90.8 27 Al 142.3 3.3 141.8 100.4 39 K 1837.8 1.1 2034.0 90.4 43 Ca 32170.1 0.7 32300.0 99.6 51 V 37.8 1.1 37.9 99.8 53 Cr 19.2 1.7 20.4 93.9 55 Mn 38.0 0.9 39.0 97.6 56 Fe 98.1 3.9 98.1 100.0 59 Co 28.8 0.7 27.1 106.4 60 Ni 59.2 0.8 62.4 94.9 63 Cu 23.2 0.8 22.8 101.9 66 Zn 70.0 0.5 78.5 89.2 75 As 54.3 0.9 60.5 89.8 78 Se 10.0 3.4 12.0 83.2 95 Mo 121.7 1.1 121.4 100.3 107 Ag 1.1 1.4 1.1 101.1 111 Cd 6.2 0.8 6.6 94.3 121 Sb 59.5 0.9 58.3 102.0 205 Tl 7.4 0.8 7.4 100.0 208 Pb 19.6 0.9 19.6 99.7 Table 3. Washout Performance (Mean value of 26 ICS-AB spikes [100 ppb], each immediately followed by two consecutive blanks. Percent reduction calculated as 1-([mean Blank]/[mean ICS-AB]) in percentage.) ICS-AB Percent Percent Mass/ spike Blank 1 reduction Blank 2 reduction element Mean Mean Mean Mean Mean 9 Be 94.9315 0.0199 99.979 0.0097 99.990 23 Na 96707.6923 19.6032 99.980 13.5090 99.986 24 Mg 79238.8462 14.2332 99.982 9.8046 99.988 27 Al 75758.0769 11.7913 99.984 7.8004 99.990 39 K 82694.2308 17.6441 99.979 13.2657 99.984 43 Ca 9092.8462 1.4105 99.984 0.9697 99.989 53 Cr 95.7327 0.0419 99.956 0.0441 99.954 55 Mn 94.8977 0.0132 99.986 0.0069 99.993 56 Fe 77021.9231 12.5122 99.984 8.0837 99.990 57 Fe 75266.5385 12.0863 99.984 7.7304 99.990 59 Co 106.8577 0.0140 99.987 0.0092 99.991 60 Ni 101.3692 0.0161 100.016 0.0129 100.013 63 Cu 98.5700 0.0163 99.984 0.0043 99.996 66 Zn 99.9350 0.0055 99.994 0.0011 99.999 75 As 95.8615 0.0290 99.970 0.0171 99.982 78 Se 94.0162 0.0841 99.911 0.0428 99.955 95 Mo 1862.3077 1.4281 99.923 0.6278 99.966 107 Ag 96.8769 0.0181 99.981 0.0098 99.990 111 Cd 104.0538 0.0134 99.987 0.0084 99.992 121 Sb 109.1346 0.2629 99.759 0.1077 99.901 205 Tl 93.4731 0.0339 99.964 0.0131 99.986 208 Pb 92.4704 0.0175 100.019 0.0241 100.026 Washout Washout is always a concern in high sample throughput applications, particularly when analyzing high-matrix, variable samples. In order to evaluate the washout for each element, two sequential blank samples were measured immediately after each spiked ICS-AB sample. The spiked ICS-AB contained 100 ppb of all calibrated elements, plus very high concentrations of Na, Mg, Al, K, and Fe. Memory effects were determined by measuring the blank immediately following the ICS-AB. Any subsequent carryover was measured in the second blank (Table 2). In all cases, greater than 3 orders of magnitude reduction (> 99.9%) was achieved before the first blank, even for the high-concentration matrix elements. The second blank showed nearly no additional reduction, indicating that essentially complete washout was achieved during the configured sample uptake and rinse-out steps of the analysis. Even "sticky" elements, such as Mo, Sb, and Tl, demonstrated the same high degree of washout. This level of washout is comparable to or better than standard peristaltic pumped systems using much longer rinse times. 6

Conclusions The results of this simple experiment illustrate that discrete sampling in spectrum mode (as opposed to time-resolved mode) using the Agilent Integrated Sample Introduction System can achieve extremely high sample throughput for typical environmental analyses using USEPA criteria. These data highlight that this novel method, using ISIS, easily exceeds the demanding USEPA requirements for stability, interference control, accuracy, precision, and washout. The ISIS-DS system offers several advantages over other discrete sampling systems: Full integration into the ICP-MS mainframe, fully integrated software, compatibility with the industry standard ASX 520 autosampler, no vacuum pump and associated pump valve to wear and replace, very low sample consumption (~2.2 ml/sample), and the flexibility to use the ISIS for other supported sample-introduction tasks, such as constant-flow nebulization, autodilution, or hydride generation. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. 7

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2009 Printed in the USA March 11, 2009 5990-3678EN