SIZE-SELECTED AEROSOL FILTER ANALYSIS USING ELECTROSTATIC CLASSIFICATION AND LIBS

Similar documents
CARBON NANOTUBE ANALYSIS USING THE TSI LIBS DESKTOP ANALYZER

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS

MEASURING NANOPARTICLE EXPOSURE

NANOSCAN SMPS SPECTROMETER COMPARED TO THE TSI SMPS SPECTROMETER

MACROIMS MACROION MOBILITY SPECTROMETER MOLECULAR WEIGHT AND PARTICLE SIZE DETERMINATION OF POLYVINYLPYRROLIDONE (PVP)

TSI AEROTRAK REMOTE 4-20 ma SENSORS LIFE SCIENCE

Supplementary information

Estimation of Mass with the Model 3321 APS Spectrometer

Heat Flow Calculations Made Using the Models 8386 and 8386A VELOCICALC Plus Air Velocity Meters Application Note ITI-124

Supplementary Material

ELECTROSTATIC CLASSIFIER MODEL 3082

Rapid Measurements of Aerosol Size Distributions Using a Fast Integrated Mobility Spectrometer (FIMS)

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

AMMS-100 Atmospheric Metal Monitoring System

Jasco V-670 absorption spectrometer

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

AN1542 APPLICATION NOTE

The new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch. Symposium Stockholm Mai

Analysis Repeatability of Trace and Major Elements in a Water Sample

Application of the EQP and EQS mass spectrometers to time resolved studies in ion beam and plasma processes

PD300 PD300-1W PD300-3W PD300-TP

Fast, Effective XPS Point Analysis of Metal Components

ICP-MS. High Resolution ICP-MS.

SOP SMPS BaltimorePM supersite Version /29/01 Page 1 of 13 SMPS

HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY

Using Alba with the FemtoFiber laser by Toptica for 2-photon quantitative imaging

Laser Dissociation of Protonated PAHs

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

MULTI-WAVELENGTH OPTICAL CALIBRATION OF THERMAL/OPTICAL ANALYZER AND

Fluorescence tracer technique for simultaneous temperature and equivalence ratio measurements in Diesel jets

Quantitative Analysis of Carbon Content in Bituminous Coal by Laser-Induced Breakdown Spectroscopy Using UV Laser Radiation

Simultaneous Velocity and Concentration Measurements of a Turbulent Jet Mixing Flow

Particle Size Distribution Measurements with the Novel 1 nm-smps

Multi-Instrument Manager Tool for Data Acquisition and Merging of Optical and Electrical Mobility Size Distributions

Imagent for fnirs and EROS measurements

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

Laser heating of noble gas droplet sprays: EUV source efficiency considerations

for XPS surface analysis

Introduction to Fourier Transform Infrared Spectroscopy

Mobile Atmospheric Chemistry Laboratory

Supplementary Material for. Reactive Uptake of Ammonia by Secondary Organic Aerosols: Implications for Air Quality

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Supplementary Materials for

VIBRATING ORIFICE AEROSOL GENERATOR MODEL 3450

Developments & Limitations in GSR Analysis

Meet Our Uncle: 12 Stability Applications on One Platform

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Effect of Sampling Depth on the Analyte Response in Laser Ablation Inductively Coupled Plasma Mass Spectrometry

Projects in the Remote Sensing of Aerosols with focus on Air Quality

For the rapid, sensitive and accurate detection of Caspase 5 inhibition by various compounds

Caspase 7 Inhibitor Drug Detection Kit

Q-300 TCT Application Note - A-Q-300TCT-BallGridArrays-001a-EN

Differential Spectral Imaging of the CN Violet Band in Laser-Induced Plasmas on TNT Simulant Molecules

OES - Optical Emission Spectrometer 2000

Caspase 3 Inhibitor Drug Detection Kit

Discriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS)

ab Cervix cancer tissue array of 102 cases (1.5mm)

Introduction to Fourier Transform Infrared Spectroscopy

Vibrational imaging of tablets by epi-detected stimulated Raman scattering microscopy

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Graphene photodetectors with ultra-broadband and high responsivity at room temperature

Audit Report. Group photo taken during the audit day July 4, Pha Din Station. General Conclusion:

Intercomparison of Mobility Particle Size Spectrometers

Meet Stunner: The one-shot protein concentration and sizing combo

3D-Printed Fluidic Devices Enable Quantitative Evaluation of Blood Components in Modified Storage Solutions for Use in Transfusion Medicine

Performance Characterization of A New Cam System M.J. Koskelo 1, J.C. Rodgers 2, D.C. Nelson 2, A.R. McFarland 3 and C.A. Ortiz 3

Calibration checks of particle counter using primary and other techniques at the laboratory level

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Laser desorption/ionization on the layer of graphene nanoparticles coupled with mass spectrometry for characterization of polymer

Total NADP and NADPH Assay Kit (Colorimetric)

ab Caspase 3, Caspase 8 and Caspase 9 Multiplex Activity Assay Kit (Fluorometric)

Microsoft Excel Directions

RamanStation 400: a Versatile Platform for SERS Analysis

Differential Mobility Particle Sizer (Aerosol measurements)

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Plasma Spectroscopy in ISTTOK

MICHAEL J. WITTE 1 AND CHRISTIAN G. PARIGGER 1,a

White Rose Research Online URL for this paper: Version: Accepted Version

Positive and Nondestructive Identification of Acrylic-Based Coatings

Obsolete Product(s) - Obsolete Product(s)

SOME ASPECTS OF PYROMETRY OF "WHITE" OBJECTS V.E.

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Lab 5: Calculating an equilibrium constant

Experiment 4 Radiation in the Visible Spectrum

Strategies to compare far-uv Circular Dichroism spectra of similar proteins using Chirascan -plus ACD

New Instruments from GRIMM

Intercomparison of Mobility Particle Size Spectrometers

Laser Ablation for Chemical Analysis: 50 Years. Rick Russo Laser Damage Boulder, CA September 25, 2012

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

PRECISION AGRICULTURE APPLICATIONS OF AN ON-THE-GO SOIL REFLECTANCE SENSOR ABSTRACT

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY

Instrumental Limitations for High Absorbance Measurements in Noise

Application Note. Understanding Performance Specifications for Low Background Alpha Beta Counters. FOM What Is It and Is It Useful?

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Keysight Technologies Measuring Polarization Dependent Loss of Passive Optical Components

Ultrafast Dynamics and Single Particle Spectroscopy of Au-CdSe Nanorods

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a).

Transcription:

SIZE-SELECTED AEROSOL FILTER ANALYSIS USING ELECTROSTATIC CLASSIFICATION AND LIBS APPLICATION NOTE LIBS-023 Quantitative elemental composition analysis is possible using size-selected aerosol samples collected on a filter. This is accomplished by combining size-selection using a TSI Model 3080 Electrostatic Classifier with elemental analysis using the TSI Model 3464-01 LIBS Desktop Analyzer. A method is described for loading filters, collecting spectra, and determining calibrations for individual elements in a multi-component mixture. This technique offers a fast and robust way of conducting filter analysis to predict element loadings. Sampling Setup The sampling setup is shown in Figure 1. A metal solution is aerosolized, and the particles are dried using a diffusion drier. Size-selected metal particles (D m = 200 nm) are collected onto a 47 mm polycarbonate filter. Mass loading is calculated based on the filter area, mass concentration, volume flow and exposure time. Figure 1: Sampling Setup TSI and TSI logo are registered trademarks of TSI Incorporated. SMPS is a trademark of TSI Incorporated.

Sample Preparation A sample solution containing ~20 mg/l of eight different metals (High Purity Standards, Charleston, SC) was prepared and aerosolized. Resulting aerosol mass concentrations and filter loading (per metal) ranged from 18 to 32 g/m 3 and 20 to 370 ng/cm 2 for exposure times that ranged from 0.75 to 18.5 hours, respectively. The metal concentrations are given in Table 1. Table 1: Metal Concentrations in Multi-Component Sample Solution (µmol/l) Li Na K V Mo Ag Ba Pb 3200 980 580 440 230 210 160 110 Aerosol Size Distribution Figure 2 shows the mass-weighted size distribution of the 200 nm classified metal particle aerosol measured with a TSI Model 3034 SMPS spectrometer. 327 scans covering more than 16 hours were averaged. The calculated total mass concentration was 30.2 µg/m 3. The bimodal distribution was likely caused by a) some doubly-charged dimmers being classified as 200 nm, and b) coagulation of particles in the mixing chamber. The intensity of the larger peak is more pronounced in the mass weighted graph. Figure 2: Mass-Weighted Size Distribution of 200 nm Classified Metal Particles. Graph Shows Average ± 1 Standard Deviation of 327 Scans. Insert Shows Number-Weighted Size Distribution -2-

Sample Analysis The filter samples were analyzed with a Model 3464-01 LIBS Desktop Analyzer, using a 50 mj, 1064 nm Nd:YAG laser and a 4 channel broadband spectrometer. Instrument settings were as follows: Laser output: 60% (~30 mj/pulse) Spot size: 100 µm Sample shots: 1 (shot per spot) Layout: 15 x 15 grid (225 spots total) Spacing: 1.5 mm x 1.5 mm Acquisition delay: 1 µs Data Analysis and Results Acquired spectra were analyzed for metal content at different loadings in order to develop a calibration for each metal type. One or several peaks for each metal were selected, integrated after baseline subtraction, and normalized against a carbon background peak. The selected peaks for the different elements are shown in Table 2. Table 2: Element Peaks Chosen for Calibration Element Wavelength (nm) Element Wavelength (nm) Element Wavelength (nm) Li 610.3542 V 292.4017 Ag 328.068 Li 670.7775 V 310.2292 Ag 338.2887 Na 588.995 V 311.0709 Ba 455.4033 Na 589.5924 Mo 277.5402 Ba 614.1713 K 766.4899 Mo 281.6158 Ba 649.6898 K 769.8964 Mo 390.2953 Pb 220.3534 C 247.8561 For each metal, calibration curves were developed for normalized peak area versus surface loading. Figure 3 shows a calibration curve for silver. Linear calibrations for all eight elements were found. Correlation coefficients for each metal are given in Table 3. Figure 3: Calibration Curve for Silver, using the 338.29nm Line -3-

Potential sources of error in the y-axis surface loading values include: SMPS spectrometer measurement error Error in surface area calculation Error in volume flow measurement Error in exposure time measurement Error in particle density determination Estimates for these errors are as follows: SMPS measurement: Estimated ±10% One aspect of this error is the limited size range of the SMPS spectrometer, i.e. the mass weighted size distribution extends past 500 nm. To evaluate this error, two log-normal functions were fitted to the data. The total mass was calculated from the fit and compared with the measured values. The fit was then extended to ~1 µm, and the extra mass was calculated. The error between calculated and measured mass values was between 0.1 and 0.3%, and the extra mass between 4.3 and 12.5%. Surface area: Volume flow: Exposure time: Particle density: The diameter can easily be measured within ±1 mm, with an exposed diameter of 43 mm, this equals ±5% in area (rounded) The 4140 flow meter is specified at 2% error. As a worst case, exposure time was controlled with 15 seconds. This equals a total error of <0.6%. Not a source of error in this work because the same particle type was used for all measurements. Assuming an SMPS measurement error of 10%, the total error ranges from 12 to 17%. The y-error bars shown in Figure 3 indicate these error calculations. Table 3: Correlation Coefficients r 2 for Predicted Values Versus Measured Values Line Value Line Value Line Value Li 610 0.9247 V 292 0.9176 Ag 328 0.9499 Li 670 0.9234 V 310 0.9255 Ag 338 0.9595 Na 588 0.9427 V 311 0.9248 Ba 455 0.9375 Na 589 0.8918 Mo 277 0.9343 Ba 614 0.9361 Na comb. 0.9274 Mo 281 0.9369 Ba 650 0.9377 K 766 0.9478 Mo 390 0.9245 Pb 220 0.9430 K 769 0.9446 The calibrations were used to predict surface loadings using the leave-one-out method of cross validation. Calibrations were built using 13 of the 14 data points for each metal and then the calibration was applied to the normalized peak area of the omitted point. Typical prediction results are shown in Figure 4 which compares the known surface loading to the predicted loadings for Mo and Ag, resulting in average prediction errors of 28% and 21%, respectively, and limits of detection around 20 ng/cm 2 for each metal. Similar results were obtained for the other metals. The average slope of the correlations was calculated as 0.937 ± 0.013, indicating an under-prediction of ~6%. -4-

Figure 4: Plots of Predicted Values (y-axis) Versus Measured Values (x-axis) of Surface Loading Ambient Measurements Based on a detection limit of 20 ng/cm 2, detection limits for ambient air can be calculated. Volume flow and sample time are variable, and the use of a 47 mm filter is assumed for these calculations. Table 4 below shows the detection limits (in ng/m 3 ) for flows of 1, 5 and 16.7 l/min, and sampling times between 1 and 24 hours. Higher flows and longer sampling times will lower the required concentration in the aerosol. Table 4: Ambient Detection Limits (in ng/m 3 ) Sample Flow Sample Time (hr) m 3 /hr l/min 1 4 12 24 0.06 1 4841 1210 403 202 0.3 5 968 242 81 40 1 16. 7 290 73 24 12-5-

Conclusions Size-segregated aerosol material collected on polycarbonate filters was quantified for metal concentrations using the combination of aerosol classification and LIBS. Simple linear calibration models were fairly accurate in predicting concentrations. Better results may be obtained using more sophisticated models. Even in this straightforward method, limits of detection as low as 20 ng/cm 2 could be achieved. LIBS analysis provided an easy method to perform quantitative elemental analysis with little required sample preparation and quick data acquisition and analysis. The light element capability of LIBS allowed for straightforward normalization using carbon peak data. For further information about this application, please contact: Dr. Phillip Tan Sr. Product Manager phillip.tan@tsi.com Dr. Steve Buckley Dir. Market Development steve.buckley@tsi.com TSI Incorporated Visit our website www.tsi.com for more information. USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 4 91 11 87 64 Germany Tel: +49 241 523030 India Tel: +91 80 67877200 China Tel: +86 10 8251 6588 Singapore Tel: +65 6595 6388 LIBS-023 Rev C 2013 TSI Incorporated Printed in U.S.A.