TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS

Similar documents
Abstract. 1. Introduction

Study on a thermal diffusivity standard for the laser flash method measurements 1

New analysis for determination of hemispherical total emissivity by feedback-controlled pulse-heating technique

UNCERTAINTY IN THE TEMPERATURE OF SILICON WAFERS MEASURED BY RADIATION THERMOMETRY BASED UPON A POLARIZATION TECHNIQUE

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

Potential for improved radiation thermometry measurement uncertainty through implementing a primary scale in an industrial laboratory

Guide to the Realization of the ITS-90

Novel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007

Introduction to metrology

Infrared Calibration Development at Fluke Corporation Hart Scientific Division

EVOLUTIONS IN THE DEVELOPMENT OF A MEASUREMENT REFERENCE STANDARD FOR THE TEMPERATURE RANGE ABOVE 962 ºC

Toward Reliable Industrial Radiation Thermometry

Temperature Measurements

Section 7. Temperature Measurement

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

Carbonized Electrospun Nanofiber Sheets for Thermophones

Sapphire Fiber-Optic Temperature Sensor Based on Black-Body Radiation Law

Performance of the MCP-PMT for the Belle II TOP counter

InAs photodiode for low temperature sensing

Calibration of temperature sensors within Length Standard Section of NMIJ

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

The International Temperature Scale of 1990

A TES Bolometer for THz FT-Spectroscopy

Automated volume measurement for weihts using acoustic volumeter

= (fundamental constants c 0, h, k ). (1) k

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

Chapter 1. Blackbody Radiation. Theory

Chapter 3. Design for manufacture of Fabry-Perot cavity sensors

VTS Process Photodiodes

DETERMINING SIZE OF SOURCE FOR HANDHELD INFRARED THERMOMETERS THEORY AND PRACTICE

Photoelectric Effect

AN INHERENTLY LINEAR TRANSDUCER USING THERMAL SIGMA-DELTA MODULATOR

Herschel PACS ICC. The bandwidth of the PACS photometric system. Issue: 1.1

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS

Methods of total spectral radiant flux realization at VNIIOFI

EXPERIMENT NO. 4. Thermal Radiation: the Stefan-Boltzmann Law

The in-flight calibration system for the airborne imager GLORIA

Introduction to Blackbody Sources

On the effects of temperature dependence of spectral emissivity in industrial radiation thermometry

RADIOMETRIC COMPARISON BETWEEN A NATIONAL LABORATORY AND AN INDUSTRIAL LABORATORY

3-DIMENSIONAL SPECTROSCOPIC-TOMOGRAPHY OF BIOLOGICAL MEMBRANE BY THE IMAGING-TYPE 2-D FOURIER SPECTROSCOPY

TEST-LOCATION SPECIFICATION BY MEANS OF HARDNESS MAPPING ON VICKERS BLOCK SURFACE

Photoelectric Effect

IR spectral characterization of customer blackbody sources: first calibration results

Sensor Technology Summary

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

PLANCK S CONSTANT IN THE LIGHT OF AN INCANDESCENT LAMP

Application of the Multi-current Transient Hot-Wire Technique for Absolute Measurements of the Thermal Conductivity of Glycols

Approximation of ITS-90 with High Temperature Thermometers up to 1085 C

1 THE CONCEPT OF TEMPERATURE

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

DEVICE FOR MEASUREMENT OF THERMAL EMISSIVITY AT CRYOGENIC TEMPERATURES

Tohru IUCHI*, Tomoyuki TSURUKAWAYA** and Akira TAZOE***

Supplementary Figures

Care should be taken to give an appropriate number of significant figures in the final answers to calculations.

Radiation Thermometry Review Protocol

The Stefan-Boltzmann Law

Measurement method for the proficiency testing program

Infrared Temperature Calibration 101 Using the right tool means better work and more productivity

C101-E116. Talk Letter. Vol.4 March 2010

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

Sensors - April A Certified-Emissivity Blackbody for Calibrating Infrared Thermometers

Uncertainty of temperature measurements by infrared thermography for metal cutting applications

SUPPLEMENTARY INFORMATION

Resistance Thermometry based Picowatt-Resolution Heat-Flow Calorimeter

Burner Tubing Specification for the Turbulent Ethylene Non-Premixed Jet Flame

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Recent improvement of the LHD Thomson scattering system

Sensors and Actuators Sensors Physics

PHYSICS ADVANCED HIGHER. Unit 3 Electromagnetism Homework

Satellite Type Estination from Ground-based Photometric Observation

REFERENCE SPECTROMETRY FOR CALIBRATION OF OPTICAL EARTH OBSERVATION SYSTEMS

Thermal Image Resolution on Angular Emissivity Measurements using Infrared Thermography

Correction for window transmission losses in radiation thermometry using high temperature fixed points

Thermal Radiation: The Stefan-Boltzmann Law

Calibration of capacitance diaphragm gauges with 1333 Pa full scale by direct comparison to resonant silicon gauge and static expansion system

Model TS318-11C55 Thermopile Sensor

INFRAMET. 2.1 Basic laws

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 55, NO. 4, AUGUST /$ IEEE

Standards for Radiation Thermometry

Operator's Manual. optris BR 400. Calibration source

INFLUENCE OF SURFACE EMISSIVITY AND OF LOW EMISSIVITY SHIELDS ON THE THERMAL PROPERTIES OF LOW DENSITY INSULATING MATERIALS

Maximum heat transfer along a sapphire suspension fiber for a cryogenic interferometric gravitational wave detector

Extraction of incident irradiance from LWIR hyperspectral imagery

Temperature Sensors & Measurement

ABB temperature measurement Radiation thermometry. Measurement made easy. Process temperature measurement practice--non-contacting

The Curve of Growth of the Equivalent Width

Xenics Presentation. A 25 minute crash course in Thermography

Development of a Bolometer Detector System for the NIST High Accuracy Infrared Spectrophotometer

CCT-K5: Comparison of local realizations of the ITS-90 between the silver point and 1700 C using vacuum tungsten strip lamps as transfer standards

Partial Energy Level Diagrams

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

Measuring Laser Diode Optical Power with an Integrating Sphere

First step: make the detector, a current-to-voltage converter using an op amp and a photodiode

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators

Application of high-precision temperature-controlled FBG f ilter and light source self-calibration technique in the BOTDR sensor system

Energy Resolution Enhancement of Retarding Type Electron Analyzer for Photoelectron Thermometry

Method to Remove the Effect of Atmosphere and Ambient Radiation on Colorimetric Temperature Measurement

Transcription:

XIX IMEKO World Congress Fundamental and Applied Metrology September 6, 29, Lisbon, Portugal TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS Fumihiro Sakuma NMIJ, National Institute of Advanced Industrial Science and Technology, Tsukuba, Japan, f-sakuma@aist.go.jp Abstract The ambient or instrument temperature affects the outputs of a standard radiation thermometer through various parts such as a detector, a feedback resistance and an interference filter. Usually the output was corrected by one coefficient assuming the linear dependence. If the filter wavelength changes, the dependence is different at each target temperature. We measured the temperature coefficients of Topcon radiation thermometers at different target temperatures and found the target temperature dependence as well as the feed back resistance dependence. Therefore a high stability radiation thermometer requires controlling the detector, filter and amplifier temperatures. Keywords: ambient temperature effect, radiation thermometer, wavelength shift. INTRODUCTION Topcon radiation thermometers are used in the traceability system of radiation thermometers in Japan,.65 µm from 96 o C to 28 o C and.9 µm from 4 o C to 2 o C. Stability is one of the important factors for the uncertainty of radiation thermometers. We reported the long term stability of.65 µm Topcon radiation thermometers and found that the output change was mainly caused by the shift of the center wavelength of the interference filter []. We also reported that the range ratios of most Topcon radiation thermometers were stable within ±.%/year [2]. This paper describes the effect of ambient temperature on the thermometer output. The ambient or instrument temperature change causes the change in response through many factors. The response of the silicon detector has a large temperature coefficient at the longer wavelength than µm. The amplifier gain changes by the change of the feedback resistance. The filter changes its transmittance and center wavelength. It has been very difficult to measure the shift of the filter center wavelength therefore its effect was assumed small, the temperature coefficient was measured at one target temperature and the value was applied for the whole temperature range. We studied the temperature coefficients about the target temperature dependence as well as the detector dependence, the wavelength dependence, the filter dependence and their stability. 2. EXPERIMENTAL 2.. Radiation thermometers The characterization of.9 µm and.65 µm radiation thermometers are described in references 3 and 4 [3,4]. Eight thermometers, R to R8 were used for.65 µm, six, S to S6 for.9 µm and two, T and T2 for µm. Most.65 µm thermometers used S336-5BK of the Hamamatsu Photonics as a detector, and DIF-BPF-2 of the Optical Coating of Japan (OCJ) or hard coating one of the Barr Associates as an interference filter. Thermometer R5 had a different detector S226-5BK and a different filter TO-399. The band width was about 5 nm. The early made.9 µm thermometers S and S6 had S874-5K of the Hamamatsu Photonics as a detector. They were changed to S336-5BK. Other thermometers had S336-5BK except for S2 thermometer which had S2386-5K as a detector. All.9 µm thermometers had DIF-BPF-4 of OCJ as a filter with a band width of about 8 nm. All µm thermometers had S336-5BK as a detector and DIF-BPF-4 of the OCJ as a filter with a band width of about 9 nm. For monitoring the instrument temperature, Topcon radiation thermometers used an IC temperature sensor, µpc66 placed close to the detector. The output of the sensor was multiplied ten times and monitored by a voltmeter, K2 of the Keithley. The output signal of the silicon photodiode was amplified by an operational amplifier, OPABM with a feed back resistance from kω, kω, MΩ, MΩ and MΩ. The amplified output was monitored by a voltmeter, 3458A of the Agilent. 2.2. Measurement of temperature coefficient Four kinds of fixed-point blackbodies were used, copper (84.62 o C), silver (96.78 o C), aluminium (66.323 o C) and zinc (49.527 o C) [5]. The temperature coefficients of.65 µm thermometers were measured only at copper and silver. Usually the coefficients of.9 µm thermometers were measured any of the first three fixed points and the zinc point coefficient was measured especially for this research.

In measuring the temperature coefficient, the laboratory temperature had been set at 2 C from the previous evening. Next morning two freezing curves of the fixed point were measured. Then the laboratory temperature was increased to 25 C. After waiting for three hours two freezing curves were measured again. 3. MEASURED RESULTS 3.. Filter wavelength.65 µm Fig. shows an example of the temperature coefficient measurement of the.65 µm radiation thermometer R. The laboratory temperature was 2 C and 25 C and the instrument temperature was about 22.4 C and 27 C, respectively. The output signal of the thermometer decreased from 44.58 mv to 44.53 mv. From the slope of the linear fitting, the temperature coefficient of the thermometer was obtained as -.27 %/ C. Fig. 2 shows temperature coefficients of eight.65 µm radiation thermometers from 994 to 28. In the legend after Rn (n= to 8), A denotes that the detector is S2386-5BK and B denotes that the filter is a hard coating type of the Barr Associates, respectively. R: The temperature coefficient of R using a Barr filter was measured six times from 2 and the average and the standard deviation were -.27 %/ C and.4 %/ C, respectively. R2: The temperature coefficient of R2 using an OCJ filter was measured four times from 999 and the average and the standard deviation were +.5 %/ C and.2 %/ C, respectively. The filter of R2 was changed to a Barr filter in early 26. The temperature coefficient was -.2 %/ C in 26 and 28. R3: The temperature coefficient of R3 using an OCJ filter was +.6 %/ C in 2 and +.8 %/ C in 2 then it increased to +.42 %/ C in 23 and +.88 %/ C in 25. This change corresponded to the filter center wavelength shift to longer wavelength,.5 nm from 2 to 23 and 4 nm from 23 to 25 []. The filter of R3 was changed to a Barr filter in early 26. The temperature coefficient became negative, -. %/ C in 26 and 28. R4: The temperature coefficient of R4 using a Topcon filter was measured three times from 994 and the average and the standard deviation were -.45 %/ C and. %/ C, respectively. R5: The temperature coefficient of R5 using a Barr filter was -.8 %/ C in 999 and after changing the detector and the amplifier, it was -.2 %/ C. R6: The temperature coefficient of R6 using an OCJ filter was +. %/ C in 994. After changing the filter to a Barr filter, it became -.2 %/ C in 998 and 28. R7, R8: The temperature coefficients of R7 and R8 using OCJ filters, were +.2 %/ C. Output Signal (V).4459.4458.4457.4456.4455.4454.4453.4452 22 24 26 28 R Linear fit Fig.. Instrument temperature dependence of the output signal of the.65 µm radiation thermometer R at the copper point...5 -.5 -. 994 995 996 997 998 999 2 2 22 23 Time (year A.D.) Fig. 2. Temperature coefficients of.65 µm radiation thermometers. B in the legend box means a Barr filter. 24 25 26 27 28 29 R B R2 R2 B R3 R3 B R4 A R5 B R6 R6 B R7 R8 3.2. Filter wavelength.9 µm Temperature coefficient stability Fig. 3 shows temperature coefficients of six.9 µm radiation thermometers from 983 to 28 AD. S: The temperature coefficient of S with an S874-5K detector and a.88 µm filter was +.23 %/ C. The detector and filter were changed and the coefficient became +.2 %/ C and was stable from 994 to 28. S2: The temperature coefficient of S2 with an S2386 detector was measured seven times from 999 and the average and the standard deviation were +.6 %/ C and.4 %/ C, respectively. S3: The temperature coefficient of S3 with an S336-5BK detector was measured five times from 999 and the average and the standard deviation were +.29 %/ C and.6 %/ C, respectively. S4: The temperature coefficient of S4 with an S336-5BK detector was +.25 %/ C. S5: The temperature coefficient of S5 with an S336-5BK detector was +. %/ C in 999 and in 26.

.25.9 Temperature Coefficient (%/oc).2.5..5 S S2 S3 S4 S5 S6.8.7.6.5.4.3.2...26 nm/oc 2 29 28 27 26 25 24 23 22 2 2 999 998 997 996 995 994 993 992 99 99 989 988 987 986 985 984 983 Time (year A.D.) Fig. 3. Temperature coefficients of.9 µm radiation thermometers S6: The temperature coefficient of S6 with an S874-5K detector was +. %/ C. After changing the detector to S336-5BK the coefficient became smaller,.4 %/ C in 989 and.3 %/ C in 24. Dependence on target temperature Fig. 4 shows the target temperature dependence of the temperature coefficient of S at the four fixed points. Gain was used at the Zn and Al points, gain at the Ag point, and gain. at the Cu point. The solid line shows the temperature coefficient when the center wavelength shifts.26 nm/ C to longer wavelength. This was calculated by the following equation dv V c2 dλ = ( 5 + ) λt λ, () derived from the Wien s equation. Here dv is the change of the output V, c 2 is the second radiation constant, λ is the center wavelength, dλ is the wavelength shift and T is the target temperature. The error bar shows the standard deviation of the coefficient calculated from the four measured data as in Fig.. The standard deviation at 42 C was large because the.9 µm thermometer output was small. At other points the error bars were too small to see. Fig. 5 shows the target temperature dependence of the temperature coefficient of S2 at the four fixed points. Different gains were used at the Al, Ag and Cu points to study the gain dependence of the coefficient. The solid line shows the temperature coefficient when the center wavelength shifts.26 nm/ C to longer wavelength. The target temperature dependence of the temperature coefficients of five.9 µm radiation thermometers were measured at the four fixed points. Most data were measured at the Ag and Cu points. Most measured coefficients are matched as the center wavelength shifts to longer wavelength between.5 nm/ C and.35 nm/ C. 4 5 6 7 8 9 2 Fig. 4 Target temperature dependence of the temperature coefficient of S at the four fixed points. Different gains to. were used. The line shows the effect of wavelength shift by.26 nm/ o C..6.5.4.3.2. 4 5 6 7 8 9 2...5 nm/oc Fig. 5 Target temperature dependence of temperature coefficient of S2 at the four fixed points. Different gains to. were used. The line shows the effect of wavelength shift by.5 nm/ o C. 3.3. Filter wavelength µm The temperature coefficients of two µm radiation thermometers were measured. Figure 6 shows the temperature coefficient of T at the four fixed points. The coefficient was much larger than.9 µm and was from +.33 %/ C to +.4 %/ C. The standard deviation at the Zn point was much smaller than that of.9 µm. The measured data can be explained by the sum of two effects. One is a center wavelength shift of the filter by.65 nm/ C to longer wavelength side (dashed line) and the other is a constant shift of.28 %/ C due to the detector (dotted line). The sum is shown by a solid line. The temperature coefficient of T2 was from +.23 %/ C to +.3 %/ C. Figure 7 shows the target temperature dependence of the temperature coefficient of T2 at four fixed points. The standard deviation was about.2 %/ C so the deviation cannot be seen in the figure. The solid line and dotted line show the effect of wavelength shift by.57 nm/ C and constant offset.82 %/ C, respectively and the

.45.4.35.3.25.2.5..5 T.65 nm/oc.28 %/oc Sum Simulation 4 5 6 7 8 9 2 Fig. 6 Target temperature dependence of temperature coefficient of T at the four fixed points (diamond). The dashed line and dotted line show the effect of wavelength shift by.65 nm/ o C and constant offset.28 %/ o C, respectively and the thick solid line shows the sum of the two. A square shows the simulation result based on the spectral responsivity and detector temperature coefficient..3.25.2.5..5 4 5 6 7 8 9 2 T2 2//6 T2 23/9/4.57 nm/oc.82 %/oc Sum Simulation Fig. 7 Target temperature dependence of temperature coefficient of T2 at the four fixed points. Diamonds and triangles show data of different date. The solid line and dashed line show the effect of wavelength shift by.57 nm/ o C and constant offset of.8 %/ o C, respectively and the dotted line shows the sum of the two. A square shows a simulation result based on the spectral responsivity and detector temperature coefficient. solid line shows the sum of the two. The tendency is the same as T but the coefficient was a little less. 4. DISCUSSION The detector, the filter and the amplifier contributed the temperature coefficient of radiation thermometers. 4.. Detector dependence In case of Topcon radiation thermometers four kinds of detectors were used. They were S874-5K, S226-5BK, S2386-5K and S336-5BK. Old S874 and S226 had small temperature dependence at.65 µm and large temperature dependence of.3 %/ C at.9 µm and.7 %/ C at µm. S2386 and S336 had no temperature dependence from.6 µm to.9 µm and about. %/ C at µm and increased rapidly in longer wavelength [6]. The main factor of the temperature coefficients of µm thermometers is the detector. In Fig. 3 all six.9 µm thermometers used OCJ filters. In 983 the detectors of S and S6 were S874 and had large temperature coefficients. After changing the detectors to S2386 or S336 the temperature coefficients were about +.2 %/ C. Simulation of µm thermometer The output V of a radiation thermometer with a relative spectral responsivity R( measuring a blackbody of temperature T is calculated as follows V ( T ) = a R( L( λ, T ) dλ. (2) Here a is a constant and L is the Planck s function at the wavelength λ. Using the temperature coefficient α at λ, the instrument temperature t a dependence of the output is shown as follows. dv dta dr( α ( =, (3) R( dt a = a α( R( L( λ, T ) dλ. (4) The measured spectral responsivity data of T and S2 are shown in Fig. 8 as a thick solid line and a dotted line, respectively. The temperature coefficient shown as a thin solid line started to increase from 94 nm. The responsivity of S2 was influenced negligibly in this case. The integrands of eq. (2) at the copper point and zinc point are shown in Fig. 9 as a thick solid line and a dotted line, respectively. Because of the characteristics of the Spectral Responsity (a.u.).8.6.4.2 T S2 Temperature Coefficient (%/oc) 8 85 9 95 5 Wavellength (nm) Fig. 8 Spectral responsivity of T (thick solid line) and S2 (dotted line). Temperature coefficient of S336 (thin solid line) [5]..8.6.4.2

Spectral Responsivity and integrand (a.u.).9.8.7.6.5.4.3.2. T TCu TZn 95 97 99 3 5 7 Wavellength (nm) Fig. 9 Spectral responsivity of T (thin solid line) and normalized integrands at the copper point (thick solid line) and the zinc point (dotted line). Planck s equation, the weight of the integrand shifts to the longer wavelength as the temperature decreases. The simulated temperature coefficients of T and T2 are shown as squares in Figs. 6 and 7. The simulated coefficients were similar in tendency to and less than the measured ones, about 7 % for T and 9 % for T2, respectively. The target temperature dependences of the simulation were also smaller than the measured data. The difference might come from the detector temperature coefficient data because only typical values were shown in the manufacturer s data sheet. 4.2. Filter dependence.65 µm: In Fig.2 the temperature coefficients of.65 µm thermometers with Barr filters were negative and about -.2%/ C while those with OCJ filters were positive and about +.2%/ C. When the filter was changed from the OCJ to the Barr for R2, R3 and R6, the coefficient changed from positive to negative. This fact clearly shows that the filter contributes the temperature coefficients. The difference of the temperature coefficients between the Barr filters and the OCJ filters was about.4 %/ C which corresponded to.2 nm/ C at the copper point. This value is the same order of the temperature coefficients of interference filters in a manufacturer s catalogue [7]..9 µm: In Figs. 4 and 5 the temperature coefficients of.9 µm thermometers S and S2 had clear target temperature dependence. The temperature coefficient increased as the target temperature decreased. This dependence is very well explained by the center wavelength shift of.26 nm/ C for S and.5 nm/ C for S2 to the longer wavelength shown as a solid line in the Figs. 4 and 5. If the main contributions of the temperature coefficients of five.9 µm thermometers were filter center wavelength shift, the shifts were from.5 nm/ C to.35 nm/ C. µm: The temperature coefficients of µm thermometers T and T2 increased as the target temperature decreased. The amount of increase was estimated as.65 nm/ C for T and.57 nm/ C for T2 in shift of the center wavelength with a offset of.28 %/ C for T and.82 %/ C for T2. The main wavelength shift and offset comes from the detector. More detailed knowledge is necessary to discuss the filter change of µm thermometers. 4.3. Thermometer S2 gain dependence The target temperature dependence of the temperature coefficient of S2 was about.5 nm/ C. However the coefficient at the Cu point was higher by.5 %/ C. S3 had similar dependence. We checked the gain dependence of the coefficients in Fig. 4 and found that gain dependence exists and that the dependence was less than.5 %/ C 4.4. Stability Figs. 2 and 3 show the stability of temperature coefficients. Unless the detector, filter and amplifier were changed, the temperature coefficients were usually stable for years; nine years for R, seven years for R2, twelve years for R4, nine years for R6, twelve years for S, six years for S2, nine years for S3 and seven years for S5. Therefore the coefficients are usually stable in time. When the filter center wavelength shifted in time as R3, the coefficient also changed. 4.5. Correction error Usually the measured temperature coefficient was assumed to be a constant and independent of the target temperature. If the coefficient was mainly contributed from the shift of the filter center wavelength, the coefficient had the target temperature dependence. Here we consider the error in the mistreatment of the instrument temperature correction for.65 µm and.9 µm thermometers..65 µm: The solid line in Fig. shows the relative output change dv/v of a.65 µm radiation thermometer due to the center wavelength shift of.2 nm. The output change was.7 % at 27 C and.35 % at the copper point. If the temperature coefficient is corrected with the value at the copper point,.35 %, the error in the correction dt is calculated by the following equation, 2 dv λt dt = (.35% ). (5) V c The error is shown as a dotted line in Fig. 6 and is about. C at 27 C..9 µm: If the relative output change dv/v of a.9 µm radiation thermometer was due to the center wavelength shift of.3 nm, the output change was.7 % at 2 C,.22 % at the copper point and.62% at 4 C as shown in Fig.. The temperature error in correcting with.22 % is +.5 C at 2 C and -. C at 4 C. 2

Relative Output Change 4.5E-4 4.E-4 3.5E-4 3.E-4 2.5E-4 2.E-4.5E-4.E-4 5.E-5 dv/v dt Cu.E+ -.2 9 2 5 8 2 24 27 3 Fig. Relative output change due to.2 nm change in center wavelength of a.65 mm radiation thermometer (solid line). Open circle shows the copper point. Broken line shows the temperature difference when the temperature coefficient is corrected with the value at the copper point (circle),.35 %. Relative Output Change 8.E-4 7.E-4 6.E-4 5.E-4 4.E-4 3.E-4 2.E-4.E-4.E+ -.2 4 8 2 6 2.6.5.4.3.2.. -. o C) Temperature Diffference (.6.4.2..8.6.4.2. o C) Temperature Difference ( dv/v dt Cu Fig. Relative output change due to.3 nm change in center wavelength of a.65 mm radiation thermometer (solid line). Open circle shows the copper point. Broken line shows the temperature difference when the temperature coefficient is corrected with the value at the copper point (circle),.23 %. 5. CONCLUSIONS The temperature coefficients of Topcon radiation thermometers were measured by calibrating the fixed-point blackbodies while changing the ambient temperature. The measured coefficients were from -.5 %/ C to.2 %/ C, from. %/ C to.3 %/ C and about.3 %/ C for.65 µm,.9 µm and µm, respectively. The temperature coefficients of.65 µm thermometers using OCJ filters were about.2 %/ C while those using Barr Associates filters were about -.2 %/ C. Usually the temperature coefficients are stable in years unless the filter, detector and amplifier are changed. If the center wavelength shifted in time, the coefficient changed. The target temperature dependences of the temperature coefficients of.9 µm radiation thermometers were measured at the copper, silver, aluminum and zinc points. The coefficients were larger at lower target temperatures and the result was well described by the shift of the filter center wavelength. The estimated shift was from.5 nm/ C to.35 nm/ C. Also a small temperature coefficient due to the gain difference was observed and the difference was within.5 %/ C. The target temperature dependences of µm thermometers were also observed. This effect was caused by the large detector temperature coefficient which increased rapidly in a longer wavelength. When the filter center wavelength shifts in the ambient temperature change, the correction of the instrument temperature should be different from other change, such as the detector response or the feedback resistance. The effect of the wrong correction on the temperature error was estimated. The best way of suppressing the instrument temperature effect is to control the instrument temperature including the detector, the filter and the amplifier. Recently we developed with the Chino a new standard radiation thermometer with the instrument temperature controlled at 3 C [8]. ACKNOWLEDGMENTS The author would like to thank Ms. Laina Ma and Ms. Yunfen Ou of AIST for their assistance in the fixed-point measurements. REFERENCES [] F. Sakuma and L. Ma, Long-Term Stability of.65 µm Radiation Thermometers at NMIJ, Int J Thermophys 29 7/5 (28) [2] F. Sakuma and L. Ma, Measurement and stability of range ratios of standard radiation thermometers Transa. Soc. Instrum. Contr. Engin. Vol.44 No. 2 29/28 (28) in Japanese. [3] F. Sakuma and L. Ma, Calibration and Characterization of the Transfer Standard Radiation Thermometer for an APMP Intercomparison in Temperature, Its Measurement and Control in Science and Industry, Vol. 7 edited by D.C. Ripple AIP 589/594 (23) [4] F. Sakuma and S. Hattori Establishing a Practical Temperature Standard by Using a Narrow-band Radiation Thermometer with a Silicon Detector in Temperature, Its Measurement and Control in Science and Industry, edited by J.F. Schooley, AIP New York Vol. 5, 42/427 (982) [5] F. Sakuma and S. Hattori A Practical-Type Fixed-Point Blackbody in Temperature, Its Measurement and Control in Science and Industry, edited by J.F. Schooley, AIP New York Vol. 5, 535/539 (982) [6] Hamamatsu Photonics, Photodiode Catalogue No.KPDJ8 Jul. 2T P9 Fig.8 (2) [7] Andover Corporation Quality Optical Filters Catalogue 28 29 p.4. [8] F. Sakuma, L. Ma, T. Suzuki, T. Kobayashi, A. Nakanishi, Calibration and Characterization of Chino 9 nm Silicon Narrow-Band Radiation Thermometer Proc. SICE Annual Conf. 24 in Sapporo.