Organophosphorus Pesticides in Apple Matrix by GC/MS/FPD using an Agilent J&W DB-35ms Ultra Inert GC Column

Similar documents
Application Note. Abstract. Authors. Introduction. Food Safety

Challenging Pesticide Analysis Using an Agilent J&W DB-35ms Ultra Inert GC Column

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection

Improved GC/MS Analysis of Tomato Pesticides with Agilent Deactivated Silica Tubing

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Highly Sensitive and Rugged GC/MS/MS Tool

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Agilent Bond Elut QuEChERS Food Safety Applications Notebook: Volume 2 INNOVATIVE APPROACHES FOR TODAY S FOOD ANALYSIS CHALLENGES

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Application Note. Abstract. Authors. Environmental

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analysis of Pesticide Residues in Apple using Agilent Bond Elut QuEChERS European Standard EN Kits by LC/MS/MS Detection

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Application Note. Authors. Abstract. Xin Sun Shandong Agriculture University, Taian, , China

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

The Analysis of Organophosphate Pesticides by LC/MS Application

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Application. Gas Chromatography February Introduction

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Author. Abstract. Introduction

Sample Pre-Treatment : Introduction to Commonly Used Techniques for Gas

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

Analysis of Pesticide Residues in Apple Using Agilent Bond Elut QuEChERS AOAC Kits by GC/MS

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

An Optimal Method for the Analysis of Pesticides in a Variety of Matrices

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

Validation Report 18

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Fully automated QuEChERS clean-up and LC/MS-QQQ analysis of pesticides in fruits and vegetables.

STANDARD OPERATING PROCEDURES

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Application Note. Abstract. Authors. Introduction

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

USP<467> residual solvents

Sulfotepp impurities in Chlorpyrifos EC formulations

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

UCT. Selectively removing Chlorophyll without Compromising your Planar Pesticides INNOVATION THROUGH CHEMISTRY E N V I R O

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Application. Gas Chromatography March 1998

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

ENVIRONMENTAL analysis

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Reduce Cost of Pesticide Residue Analysis

Headspace Technology for GC and GC/MS: Features, benefits & applications

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Analysis of BTEX in Natural Water with SPME

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

The Pulsed Flame Photometric Detector for analysis of phosphorus pesticides in wastewater and food commodities

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Application Notebook AGILENT FOOD TESTING SOLUTIONS

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Transcription:

Organophosphorus Pesticides in Apple Matrix by GC/MS/FPD using an Agilent J&W DB-5ms Ultra Inert GC Column Application Note Environmental and Food Safety Author Doris Smith and Ken Lynam Agilent Technologies, Inc. 850 Centerville Road Wilmington, DE 9809-60 USA Abstract The Agilent J&W DB-5ms Ultra Inert (UI) 0 m 0.8 mm 0.8 µm column effectively resolved the analytes of interest producing excellent peak shape for even the more problematic organophosphorus (OP) pesticides. The detection limits for most of the pesticides were 5-5 ng/ml. Recovery studies were performed by spiking with a standard solution to achieve the desired concentrations in an apple matrix; 50, 00 and 750 ng/ml GC/MS/SIM and 50, 00, 50 ng/ml FPD in phosphorus mode. Recoveries were >77% for most of the pesticides by GC/MS/SIM and >75% by GC/FPD. This application note details a quick and effective analytical method for the determination of low ppm and trace level organophosphorus pesticides residues in apple extract. A capillary flow technology (CFT) device was installed post-column to split the effluent between the MSD and FPD, implementing an automated backflush that diminished residual sample carryover and reduced instrument cycle times. This multisignal configuration allowed for full scan, selective ion monitoring (SIM), and flame photometric detection from a single injection. A simplified QuEChERS (Quick, Easy, Cheap, Effective, Rugged, and Safe) method provided sufficient sample matrix cleanup while preserving low level analyte detection.

Introduction Organophosphorous (OP) pesticides are widely used in the agricultural industry for crop protection. Human toxicities for this class of molecules have shown acute as well as chronic effects from pesticide poisoning. OP pesticides affect the nervous system of insects and mammals by inhibiting an enzyme, acetylcholinesterase, important in helping regulate nerve impulses []. Children are considered more susceptible to organophosphate toxicity because their pesticide dose per body weight is larger compared to that of adults []. Children also have lower levels of detoxifying enzymes that deactivate OP pesticides, contributing to their vulnerability to pesticide exposure [,]. Recent studies have shown a correlation between OP pesticides exposure and an increased risk for attention deficit hyperactivity disorder (ADHD) and other neurodevelopmental deficits in children [,5,6,7]. Because the main source of exposure for children is through consumption of food containing OP pesticide residues [,8], analytical testing capable of determining residual pesticides in food samples is critical. The multiresidue determination of pesticides in fruits and vegetables usually involves an organic extraction of the pesticides from the plant matrix, followed by a cleanup procedure to remove co-extractives and other interferences. Anastassiades et al [9] developed a QuEChERS method for the analysis of pesticide residues in produce. This approach simplifies the traditional, labor intensive extraction and cleanup procedure, while providing a fast, robust, and cost effective method suitable for extracting pesticide residues. Chromatographically active compounds such as organophosphorus pesticides can adsorb onto active sites in the sample flow path, particularly at trace levels, compromising an analyte s response. These pesticides tend to show peak tailing through interaction with active sites in a chromatographic system. This makes analysis challenging, particularly in difficult sample matrices. Minimizing activity in the GC column is essential to ensure accurate quantitation. Agilent s J&W DB-5ms Ultra Inert (UI) column minimizes column activity so difficult and active analytes can be consistently analyzed at trace levels. The use of the midpolarity DB-5ms UI phase also offers additional selectivity over a nonpolar phase, which can assist in resolving potentially coeluting peaks, or shift a peak of interest away from matrix interferences. A gas chromatographic system capable of multisignal detection can provide complementary data for identification, confirmation, and quantitation of target analytes from a single injection. This method provides simultaneous detection of organophosphorus pesticides by GC/MS/SIM and FPD in phosphorus mode by splitting the column effluent between the MSD and FPD. The approach chosen here uses a GC/MSD/FPD system to identify and confirm the order of elution for peaks of interest. Once the elution order is established, the chromatographic parameters can easily be transferred to a GC/FPD system. The use of FPD detection without flow splitting is expected to increase sensitivity threefold, further improving lower level detection. The GC/MS system was also equipped with backflush capability, which shortens instrument cycle time by backflushing late-eluting matrix components through the inlet purge valve. Long bakeout times between injections are avoided by using this technique. Backflushing has the additional benefit of increasing the time intervals for source cleaning by effectively clearing deleterious matrix components from the system [0]. Experimental An Agilent 7890 GC with an Agilent 5975C MSD equipped with a flame photometric detector and Agilent 768B automatic liquid sampler were used for this series of experiments. A purged two-way capillary flow technology (CFT) device was used to split the effluent : to the MSD:FPD. The CFT device also allowed for post-column backflush. Table lists the chromatographic conditions used for these analyses. Table lists the flow path consumable supplies used in these experiments. Table. GC/MSD: Sampler: CFT Device: MSD Restrictor: FPD Restrictor: PCM : Inlet: Chromatographic Conditions Agilent 7890 GC/Agilent 5975C Series GC/MSD Agilent 768B automatic liquid sampler, 5.0 µl syringe (Agilent p/n 58-7) Purged -way splitter (Agilent p/n G80B) Split Ratio : MSD:FPD. m 0.5 mm id deactivated fused silica tubing. m 0.5 mm id deactivated fused silica tubing.8 psi constant pressure µl splitless; 50 C, purge flow 60 ml/min at 0.5 min, gas saver on at min 0 ml/min Column: Agilent J&W DB-5ms UI 0 m 0.8 mm 0.8 µm (Agilent p/n-8ui) Carrier: Helium, constant pressure.5 psi at 95 C Oven: 95 C (. min), 5 C/min to 5 C, 5 C/min to 65 C,.5 C/min to 95 C, 0 C/min to 80 C (.75 min) Postrun Backflush: 5 min at 80 C, PCM pressure 70 psi during backflush, psi inlet pressure during backflush MSD: 0 C transfer line, 0 C source, 50 C quad

Table. Flow Path Supplies Vials: Amber crimp top glass vials (Agilent p/n 58-96) Vial Caps: Crimp caps (Agilent p/n 58-0) Vial inserts: 50 µl glass/polymer feet (Agilent p/n 58-887) Syringe: 5 µl (Agilent p/n 58-7) Septum: Advanced Green (Agilent p/n 58-759) Inlet liner: Deactivated dual taper Helix liner (Agilent p/n 588-598) Ferrules: 0. mm id short; 85/5 vespel/graphite (Agilent p/n 58-) PCT fittings: Internal nut (Agilent p/n G855-050) PCT ferrules: SilTite ferrules, 0.5 mm id (Agilent p/n 588-56) 0x magnifier: 0x Magnifier loop (Agilent p/n 0-00) Reagents and Chemicals All reagents and solvents were HPLC or Ultra Resi grade. Acetonitrile (ACN) from Honeywell (Muskegon, MI, USA), toluene from Burdick & Jackson, and acetone from JT Baker was purchased through VWR International (West Chester, PA, USA). The -component custom pesticide standard was prepared by Ultra Scientific (N. Kingstown, RI, USA). Solutions and Standards The OP pesticide stock standard solution (00 µg/ml of organophoshorus pesticides) was diluted in acetone to yield spiking solutions and 0 µg/ml. A surrogate standard, triphenyl phosphate (TPP), was prepared at concentrations of, 5 and 00 µg/ml in toluene. The spiking solutions were used to prepare the calibration curves in the matrix blank extract by appropriate dilution. Sample Preparation An organic apple sample was purchased from a local grocery store. The apple was chopped into small cubes and frozen at 80 C overnight. The samples were then comminuted thoroughly to achieve sample homogeneity. The sample extraction method used the QuEChERS method. Figure illustrates the sample preparation procedure graphically in a flow chart. Samples containing 5 (± 0.) grams of apple were weighed into centrifuge tubes. QC samples were spiked with appropriate amount of spiking solutions to yield QC samples with quantitative concentrations relative to the : split ratio of 50, 00, and 750 ng/ml levels for GC/MS-SIM determination, and 50, 00, and 50 ng/ml by flame photometric detection. Each sample received a 5-mL aliquot of ACN. Two ceramic bars (Agilent p/n 598-9) were added to each sample to aid in sample extraction. The samples were vortexed for minute. An original Agilent Bond Elut QuEChERS extraction salt packet (Agilent p/n 598-5555) containing 6 grams of MgSO and.5 grams sodium chloride was added to each centrifuge tube. The capped tubes were shaken on a Geno/Grinder @500 rpm for minute. The samples were centrifuged at 000 rpm for 5 min. An 8 ml aliquot of the upper layer was transferred to an Agilent Bond Elut QuEChERS General Fruits and Vegetables dispersive SPE 5 ml tube (Agilent p/n 598-5058). The dspe tube was vortexed for minute and then centrifuged at 000 rpm for minutes to complete the sample extraction. The extract from the dspe tube was transferred to a GC vial and analyzed by SIM GC/MS and GC/FPD using the chromatographic conditions listed in Table. Extractions of water and acetonitrile aliquots were prepared in the same manner as the samples and served as reagent blanks. QuEChERS Sample Preparation Workflow Weigh 5 g sample (± 0.g) 50 ml centrifuge tube Add surrogate/is solution, and QC spike solution if necessary. Vortex min Figure. Add 5 ml of ACN and ceramic bars to the sample (Agilent p/n 598-9) Vortex min Add Agilent Bond Elut QuEChERS extraction salt packet (Agilent p/n 598-5555) Shake vigorously for min on Geno/Grinder at 500 rpm Centrifuge at 000 rpm for 5 min Transfer 8 ml of upper ACN layer to AOAC General Fruits and Vegetables dispersive SPE 5 ml tube (Agilent p/n 598-5058) Vortex min, centrifuge at 000 rpm for 5 min Transfer extract to autosampler vial Analyze GC/MS/FPD Flow chart of the Agilent QuEChERS extraction procedure for apple samples.

Discussion of Results The organophosphorus pesticides were resolved on an Agilent J&W DB-5ms UI 0 m 0.8 mm 0.8 µm analysis column in about 0 minutes. The -component pesticide matrix-matched standard shown in Figure shows good peak shapes for the pesticides in both the GC/MS/SIM and FPD chromatograms. Organophosphorus pesticides, particularly the more polar pesticides can be problematic, often yielding broad peak shapes or excessive tailing making reliable quantitation at low levels difficult. The high level of inertness of the DB-5ms UI column results in better peak shape and decreased sample adsorption allowing lower detection limits. Figure depicts the excellent peak shape seen for four of the more polar OP pesticides (oxydemeton-methyl, methamidophos, mevinphos, acephate) with the DB-5ms UI column. The performance of the DB-5ms UI high efficiency column yielded excellent linearity and recovery over the calibration range of this study. The linearity of the column as defined by the r values of the pesticide standard curve was 0.99 for all the pesticides using both detectors. The individual OP pesticide analyte values are shown in Table. The GC/MS/SIM analysis was able to detect down to the 5 0 ng/ml range for most of the pesticides. A higher SIM signal is necessary to quantify the more volatile pesticides below the 0 ng/ml range due mainly to matrix interferences. Because flame photometric detection in phosphorus mode is selective only for analytes containing phosphorus, it is able to detect low levels of OP pesticides in complex matrices without the matrix interferences. The FPD was able to detect the OP pesticides down to 5 ng/ml with the exception of naled, which could only be detected at higher levels (>5 ng/ml). Naled can undergo debromination, which can have an impact on detection, especially at trace levels. The detection levels for the targeted OP pesticides were well below the US maximum residue levels (MRLs) in an apple matrix, except in the case of chlorpyrifos, which has an MRL of 0 ppb for apples and grapes []. Analysis by GC/FPD without flow splitting offers increased sensitivity to monitor the lower levels of detection needed for chlorpyrifos. Separation of OP Pesticides on Agilent J&W DB-5ms UI column 8000 000 0000 6000 000. Oxydemeton-methyl. Methamidophos. Mevinphos. Acephate 5. Naled 6. Diazinon 7. Dimethoate 6 5 7 8. Chlorpyrifos 9. Malathion 0. Methidathion. TPP (surrogate std). Phosmet. Azinphos methyl 8 9 0 MSD (SIM): 600 ng/ml 600000 00000 800000 00000 000000 600000 00000.00 6.00 8.00 0.00.00.00 6.00 8.00 0.00.00.00 6.00 8.00 0.00 Time 0 6 7 8 9 5.00 6.00 8.00 0.00.00.00 6.00 8.00 0.00.00.00 6.00 8.00 0.00 Time FPD (P): 00 ng/ml Figure. GC/MS-SIM and FPD chromatograms of a matrix matched organophosphorus pesticides standard analyzed on an Agilent J&W DB-5ms UI 0 m 0.8 mm 0.8 µm capillary GC column (Agilent p/n -8UI). Chromatographic conditions are listed in Table. The effluent split ratio is MSD:FPD = :.

Excellent Peak Shape for Polar Pesticides at Low Levels on Agilent J&W DB-5ms UI column. Oxydemeton-methyl. Methamidophos. Mevinphos. Acephate MSD (SIM): 5 ng/ml FPD (P): 5 ng/ml.00 5.00 6.00 7.00 8.00 9.00 0.00.00 5.00 6.00 7.00 8.00 9.00 0.00 Figure. Enlarged section of GC/MS/SIM and FPD chromatograms of the more problematic polar pesticides analyzed on an Agilent J&W DB-5ms UI capillary column (Agilent p/n -8UI). Chromatographic conditions are listed in Table. The effluent split ratio is MSD:FPD = :. Table. Correlation Coefficients for the OP Pesticides Calibration Standards Analyzed by GC/MS-SIM and FPD in Phosphorus Mode with a Split Ratio for MSD:FPD = : Excellent Linearity of OP Pesticides on Agilent J&W DB-5ms UI Column MSD FPD Oxydemeton-methyl 0.99 0.997 Methamidophos 0.997 0.997 Mevinphos 0.997 0.999 Acephate 0.997 0.999 Naled 0.99 0.996 Diazinon 0.996 0.997 Dimethoate 0.997 0.999 Chlorpyrifos 0.997 0.998 Malathion 0.995 0.999 Methidathion 0.996 0.999 TPP 0.999 0.997 Phosmet 0.997 0.999 Azinphos methyl 0.995 0.999 r values for 0, 75, 50, 00, 55, 750, 500 ppb MSD Calibration Levels 5, 50, 00, 75, 50, 500 ppb FPD Calibration Levels The extraction process using the QuEChERS method was effective in retaining the OP pesticides in the spiked apple sample and providing sufficient cleanup of the sample matrix for GC/MS analysis. Figure shows the organophosphorus pesticide mix spiked into an apple matrix sample. The matrix was prepared using a QuEChERS sample preparation approach that included extraction/partitioning and dispersive- SPE. A GC/MS/SIM blank matrix trace is shown below the analyte trace to indicate the level of potential matrix interference with the analytes of interest. Peak shapes for the organophosphorus pesticides are still quite sharp and well resolved indicating excellent performance on the DB-5ms UI column in fruit matrix. Recoveries were determined by GC/MS/SIM at the 50, 00, and 750 ng/ml levels, and 50, 00, and 50 ng/ml using the FPD in phosphorus mode. The recoveries for most of the pesticides were greater than 75% with average RSDs below 0%. Recoveries for the individual OP pesticides are listed in Table. Lower recoveries were noted for the more polar pesticides: oxydemeton-methyl, methamidophos, and acephate. One possible explanation is that these polar, highly water soluble pesticides may have been partially lost through incomplete partitioning into the aqueous layer during the extraction step []. 5

GC/MS SIM Chromatogram of Apple Extract Blank Relative to Spiked Sample after Agilent Bond Elut QuEChERS Extraction and Dispersive SPE Abundance 9000 8000 7000 6000 5000 000 000 000 000. Oxydemeton-methyl. Methamidophos. Mevinphos. Acephate 5. Naled 6. Diazinon 7. Dimethoate 5 6 8. Chlorpyrifos 9. Malathion 0. Methidathion. TPP (surrogate std). Phosmet. Azinphos methyl 7 8 9.00 6.00 8.00 0.00.00.00 6.00 8.00 0.00.00.00 6.00 8.00 0.00 Time 0 50 ppb spiked apple extract Blank apple extract Figure. GC/MS/SIM chromatogram of the apple extract blank and a 50 ng/ml spiked apple extract analyzed on an Agilent J&W DB-5ms UI capillary column (Agilent p/n -8UI). Chromatographic conditions are listed in Table. Table. Recovery and Repeatability of OP Pesticides in Spiked Apple Matrix with an Agilent J&W DB-5ms UI Column (Agilent p/n -8UI) (continued) Recovery and Repeatability of OP Pesticides in Spiked Apple Matrix by SIM GC/MS with Agilent J&W DB-5ms UI Column 50 ng/ml fortified QC 00 ng/ml fortified QC 750 ng/ml fortified QC Analysis % Recovery RSD (n=6) % Recovery RSD (n=6) % Recovery RSD (n=6) Oxydemeton methyl 6.6 5.5 6.0 5.0 6.9 8. Methamidophos 68.9 8.5 78.6.9 8.8.9 Mevinphos 88.7. 9..6 97.0. Acephate 77.5 6. 80. 5. 8..6 Naled 9.9 6. 87.6. 80. 5.7 Diazinon 8.5.0 89.0.7 90.9.0 Dimethoate 90.6.9 9.9. 96.6.6 Chlorpyrifos 87.0.7 9.7. 95.6.6 Malathion 9.5.8 9.9. 97..6 Methidathion 89.6. 9.. 99..7 TPP (surrogate std) 00.6.8 0.5. 00..0 Phosmet 85.9 5. 86.8. 95..8 Azinphos methyl 88.6. 8..5 95..9 6

Table. Recovery and Repeatability of OP Pesticides in Spiked Apple Matrix with an Agilent J&W DB-5ms UI Column (Agilent p/n-8ui) Recovery and Repeatability of OP Pesticides in Spiked Apple Matrix by GC/FPD with Agilent J&W DB-5ms UI Column 50 ng/ml fortified QC 00 ng/ml fortified QC 750 ng/ml fortified QC Analysis % Recovery RSD (n=6) % Recovery RSD (n=6) % Recovery RSD (n=6) Oxydemeton methyl 5.8 8.5 60.6 9. 6. 7.7 Methamidophos 6. 9. 75.5 5.9 8.7. Mevinphos 80.5 5.0 90..5 9.5. Acephate 6.. 78.5 7.5 8. 7.9 Naled 97..0 87.6 8.0 78.6 8.7 Diazinon 80.. 86.7.9 90..5 Dimethoate 80.6 7. 9.0.6 9.6.6 Chlorpyrifos 80.8.7 9.5 5.5 96.6. Malathion 8.7. 9.9.5 96.9.8 Methidathion 8.7 7.9 9.9.6 96.. TPP (surrogate std) 99.6. 99.7 6.0 95.9. Phosmet 76.6 5.5 89..6 9..5 Azinphos methyl 79. 7.6 88.5.5 9.8. Conclusions This application note successfully shows a quick and efficient analytical method to monitor low and trace level organophosphorus pesticides residue in apple samples. Splitting the column effluent to both an MSD and FPD facilitated selectivity, identification, and confirmation of OP pesticides from a single injection, thereby increasing laboratory productivity. Using GC/MS in Full Scan mode enabled identification of specific pesticides, while SIM mode offered selectivity and sensitivity for quantitation of the pesticides at trace levels. Confirmation and further specificity was achieved by FPD in phosphorus mode. FPD detection was effective at minimizing matrix interferences enabling lower detection. The Agilent Bond Elut QuEChERS method for general fruits and vegetables was successful at providing enough sample cleanup to minimize matrix interferences while still maintaining low level analyte detection. The simple QuEChERS extraction method allows for faster sample prep facilitating higher sample throughput. Residual sample matrix carryover is removed through use of backflush, which eliminates the need for a bakeout cycle, significantly reducing analytical run times. The Agilent J&W DB-5ms UI capillary column resolves the targeted OP pesticides and provides excellent peak shapes for the polar pesticides allowing for more reliable quantitation at low levels. Detection levels for the OP pesticides were at or below the US maximum residue levels (MRLs) for various fruits. Matrix-matched calibration standards yielded regression coefficients r 0.99 and recoveries from fortification studies were greater than 75% with an average RSD <0% for both GC/MS/SIM and FPD, further demonstrating the effectiveness of using an Agilent J&W DB-5ms UI column for residual pesticide determination. 7

References. L. G. Sultatos, Mammalian Toxicology of Organophosphorus Pesticides. Journal of Toxicology and Environmental Health. 99; ():7 89.. NRC (National Research Council) (99) Pesticides in the Diets of Infants and Children, National Academy Press, Washington, DC.. B. Eskenazi, A. R. Marks, A. Bradman A, et al. Organophosphate Pesticide Exposure and Neurodevelopment in Young Mexican-American Children. Environ Health Perspect. 007; 5:79-798.. C. E. Furlong, N. Holland, R. J. Richter, A. Bradman A, A. Ho, B. Eskenazi, PON Status of Farmworker Mothers and Children as a Predictor of Organophosphate Sensitivity. Pharmacogenet Genomics. Mar 006; 6():8-90. 5. M. F. Brouchard, D. C. Bellinger, R. O. Wright, and M. G. Weisskopf, Attention-Deficit/Hyperactivity Disorder and Urinary Metabolites of Organophosphate Pesticides. Pediatrics 00; 5; e70-e77. 6. A. R. Marks, K. Harley, A. Bradman, K. Kogurt, D. B. Barr, C. Johnson, N. Calderon, and B, Eskenazi, Organophosphate Pesticide Exposure and Attention in Young Mexican-American Children, Environ. Health Perspect. doi:0.89/ehp.00056. 7. M. Bjørling-Poulsen, H. R. Andersen, and P. Grandjean, Potential Developmental Neurotoxicity of Pesticides Used in Europe, Environmental Health 008, 7:50 doi:0.86/76-069x-7-50. 8. C. Lu, K. Toepel, R. Irish, R. A. Fenske, D. B. Barr, and R. Bravo, Organic Diets Significantly Lower Children s Dietary Exposure to Organophosphorus Pesticides, Environ. Health Perspect. 006;, 60-6. 9. M. Anastassiades, S. J. Lehotay, D. Štajnbaher, F. J. Schenck, Fast and Easy Multiresidue Method Employing Acetonitrile extraction/partitioning and Dispersive Solid-Phase Extraction for the Determination of Pesticide Residues in Produce, J. AOAC Int. 00; 86, -. 0. C. J. Meng, Improving Productivity and Extending Column Life with Backflush, Agilent Technologies publication 5989-608EN.. http://www.mrldatabase.com/. F. Schenck, J. Wong, C. Lu C, J. Li, J. R. Holcomb, L. M. Mitchell, Multiresidue Analysis of 0 Organophosphorus Pesticides in Produce at Parts-per- Billion Levels Using a Modified QuEChERS Method and Gas Chromatography with Pulsed Flame Photometric Detection, Journal of AOAC INTERNATIONAL, Vol. 9, No., 009, 56-57. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 0 Printed in the USA January 6, 0 5990-765EN