CuSO 4.5H 2 O Analysis with a Hyphenated TGA 8000 and Hiden Analytical HPR-20 EGA Mass Spectrometer

Similar documents
HYPHENATED TECHNOLOGY GUIDE

HYPHENATED TECHNOLOGY GUIDE

It is the purpose of this paper to demonstrate that the detection limit is directly related to the sample size, for both TG and TG/FTIR.

TGA Thermal Gravimetric Analysis

Thermogravimetric Analysis Advanced Techniques for Better Materials Characterisation

CATALYSIS & THERMAL ANALYSIS

Technical know-how in thermal analysis measurement

Hiden CATLAB Systems Microreactor for Catalysis Studies & Thermal Analysis

Hiden Isochema. Gravimetric Gas & Vapor Sorption Analyzers. Hiden Isochema IGA Series. Advancing Sorption Analysis

DETERMINATION OF THE DECOMPOSITION PRODUCTS OF CALCIUM OXALATE USING THERMAL GRAVIMETRY AND INFRARED SPECTROMETRY

Evolved gas analysis by simultaneous thermogravimetric differential thermal analysis-fourier transformation infrared spectroscopy (TG-DTA-FTIR)

Collected Applications Thermal Analysis EVOLVED GAS ANALYSIS

Hiden SIMS Secondary Ion Mass Spectrometers. Analysers for surface, elemental and molecular analysis

DSC AND TG/DTA AS PROBLEM-SOLVING TOOLS: CHARACTERIZATION OF PHARMACEUTICAL COMPOUNDS

1. Prepare the MALDI sample plate by spotting an angiotensin standard and the test sample(s).

THERMO- GRAVIMETRIC ANALYZER TGA 1000

DYNAMIC VAPOR SORPTION ANALYZER FULLY AUTOMATED COMPACT BENCH-TOP DVS ANALYZER, FOR FAST AND ACCURATE SORPTION MEASUREMENTS.

DSC AS PROBLEM-SOLVING TOOL: BETTER INTERPRETATION OF Tg USING CYCLIC DSC

Mercury, total-in-sediment, atomic absorption spectrophotometry, nameless, direct

Minimum (-31) every 10 s -35 (-31)

Technical Data Sheet: Thermokinetics 3

High Pressure DSC Differential Scanning Calorimeter

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

costech instruments ECS 4010 Elemental Combustion System CHNS-O

Chemistry (14-16) SSER Ltd. Exothermic & Endothermic Reactions

Hydra-C Application Note: 1076

Lab I. 2D Motion. 1 Introduction. 2 Theory. 2.1 scalars and vectors LAB I. 2D MOTION 15

Lab I. 2D Motion. 1 Introduction. 2 Theory. 2.1 scalars and vectors LAB I. 2D MOTION 15

Lab 4 Motion in One-Dimension Part 2: Position, Velocity and Acceleration Graphically and Statistically (pre-requisite Lab3)

Supplementary Information


EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.

Thermal Analysis Mass Spectrometer Coupling. Evolved Gas Analysis Method, Techniques and Applications

Determination of the Chemical Formula

Alka Seltzer Lab: Reaction rate - Factors that affect the rate of chemical reactions

Complete Materials Deformulation Using TGA-IR

Fourier Transform Infrared Spectroscopy (Perkin Elmer - Spectrum One)

Pressure & temperature sampling options for Hiden gas analysis systems

Nutshells of Thermal Analysis. Heat it up! Burn it! Thermal Analysis

Rapid, Efficient Phase Pure Synthesis of Ca 2 AlNO 3 Layered Double Hydroxide

Supporting Information

Uncover the insight with Hyphenation. Gerlinde Wita October 2015

Innovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer

Relative Sensitivity RS Measurements of Gases

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Application of the EQP and EQS mass spectrometers to time resolved studies in ion beam and plasma processes

MassHunter Software Overview

DVS INTRINSIC. The easy-to-use solution to complex water sorption challenges.

GAS CHROMATOGRAPHY (GC)

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

AUTOMATED FORMULATION SYSTEMS

1. (a) The diagrams below show the arrangement of atoms or molecules in five different substances A, B, C, D and E.

Petrochemical. Transformer Oil Gas Analysis - TOGA

What Is The Fastest Route to High Quality Data? AxION DSA: Quality data without chromatography or sample preparation. AxION 2 TOF MS and AxION DSA

CREATE A MOTION. Purpose Demonstrate your understanding of motion graphs by creating real-world motion(s) that match a graph(s) you ve been given.

XEMIS NEXT GENERATION GRAVIMETRIC SORPTION ANALYZER

novaa 800 D Atomic Absorption Spectrometer

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Protocol Particle size distribution by centrifugal sedimentation (CPS)

SYNTHESIS OF CARBON NANOPARTICLES. 4.0 Production and Characterization of Carbon Nanoballs and other Nanoparticles

Thermal Methods of Analysis Theory, General Techniques and Applications. Prof. Tarek A. Fayed

ST EDWARD S OXFORD. Lower Sixth Entrance Assessment. November Chemistry. 1 Hour. Candidates name:... St Edward's School 1

NCFS Sample Preparation and Instrumental Parameters

Structural and spectroscopic characterization of potassium fluoroborohydrides

INTRO TO THE MOLE -PART 1

Stabilizing vitamin D 3 by conformationally selective co-crystallization

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

Defining quality standards for the analysis of solid samples

THE UNITED REPUBLIC OF TANZANIA MINISTRY OF EDUCATION AND CULTURE FORM TWO SECONDARY EDUCATION EXAMINATIONS INSTRUCTIONS

U.S. EPA Method 8270 for multicomponent analyte determination

Microkinetics of H 2 S Removal by Zinc Oxide in the Presence of Moist Gas Atmosphere

Supporting Information:

Production of Mesoporous Carbon from Waste Tire

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

Applications of Newton's Laws

1.4 Enthalpy. What is chemical energy?

Supporting Information

Year 10 Science Chemistry Examination November 2011 Part A Multiple Choice

General Physics I Lab. M1 The Atwood Machine

Thermal stability of azurite and malachite in relation to the formation of mediaeval glass and glazes

Chem Page IX - 1 LAB MANUAL Differential Scanning Calorimetry 09_dsc131.docx EXPERIMENT IX

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

Total analysis with DSC, TMA and TGA-EGA

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Determination of Caffeine by HPLC

or a chemical change in several experimental trials.

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Scholarship Examination

Catalysis Letters, 79 (1-4): Kluwer Academic Publishers-Plenum Publishers.

Chapter 31 Gas Chromatography. Carrier Gas System

System Identification of RTD Dynamics (Tom Co, 11/26/2006; 11/25/2007, 11/15/2009)

Quantum Magnetic Levitation. By: Austin Mullins and Zach Steinberg

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Determination of Physical Properties of Organic Compounds

Determining the Concentration of a Solution: Beer s Law

Electronic Supplementary Information. Precursor Salt Assisted Syntheses of High-Index Faceted Concave Hexagon and Nanorod like Polyoxometalates

Transcription:

CuSO 4.5H 2 O Analysis with a Hyphenated TGA 8000 and Hiden Analytical HPR-20 EGA Mass Spectrometer Phil Robinson, Ruston Services Ltd, UK Jim Melling, Hiden Analytical Ltd, UK Dennie Wezendonk, Utrecht University, Netherlands Background As part of the commissioning of a new hyphenation system, a sample of copper sulphate pentahydrate was analysed using the PerkinElmer TGA 8000 hyphenated with a Hiden HPR-20 EGA mass spectrometer to provide a series of weight losses with simple inorganic gases. For simplicity, a small sample mass of the copper sulphate (~5mg) was used, and a heating rate of 40 C/min as a proving test for the evolved gas hyphenation connection. Sample Presentation and Analysis A small sample of around 5.4mg of CuSO4.5H2O was loaded into the TGA ceramic sample pan following gentle grinding in an agate mortar and pestle to produce a fine powder. By reducing the particle size in this way, the gas path length for interactions between the solid phase and the gas phase is minimised, which produces better resolution in both the TGA weight loss curve and the mass spectrometer gas evolution peaks. The sample was distributed evenly across the inside of the TGA ceramic pan to provide a thin layer, which further contributes towards maximising resolution between adjacent weight loss events. The following temperature program was used for this testing, which was carried out under an argon purge with a total flow rate of 80mL/min (balance purge flow of 50mL/min and sample purge flow of 30mL/min). 1. Isothermal at 30 C for 1 minute 2. Trigger the MS to start collecting data at start of segment 3. Heat from 30 C to 900 C at 40 C/min 4. Isothermal at 900 C for 0.1 minute 5. Trigger MS to stop collecting data at start of segment In some applications the isothermal segment shown in step 1 can be used to complete purging with argon after loading the sample. For example, if carbon monoxide (CO) was expected as an evolved species, any N2 in the system would interfere with the detection of CO at ppm levels below 1000ppm. In this situation, using argon would solve the detection issue but would require effective purging to remove any nitrogen resulting from the lab air around the sample during loading. This purging could be achieved by allowing perhaps 3-4 minutes at the starting temperature after closing the TGA following sample loading. 1

Equipment The following equipment was used PerkinElmer TGA 8000 with 48 position Autosampler. This unit includes mass-flow control for both the balance purge and the sample purge, with the flow rates to each specified in the test method used in Pyris Software for Windows. Hiden Analytical HPR-20 EGA quadrupole mass spectrometer with the EGAsoft time based data collection software. The interface between the TGA and the MS is designed to have minimal dead-space, which gives good resolution between events and eliminates tailing between weight loss events. The heated transfer line also eliminates cold spots to prevent condensation of evolved organic fragments passing from the TGA. Both systems were controlled from the same PC, which also controlled the trigger box to signal start the mass spectrometer data collection when the TGA data collection started. Treatment of Results Displaying the imported data on the same graph as the sample data allows a clear correlation of species evolved in relation to the events observed in the TGA. Therefore, after data collection by the MS software, data for each mass species of interest is exported as ASCII text data, which can then be imported directly into the PerkinElmer Pyris TGA software for display alongside the sample data. After importing and display of the mass data with the TGA weight loss profile has been completed, the mass data can be to be saved in native Pyris data file format which can then be readable by Pyris without any further treatment if required for subsequent re-display. 2

Results Curve #1 TGA Data for CuSO 4.5H 2 O Decomposition The overall change in this copper sulphate material can be characterised by the following equation:- 2CuSO 4.5H 2 O 2CuO + 2SO 2 + O 2 + 10 H 2 O The weight loses are characterised as follows: 50 C to 330 C Loss of five waters of crystallisation 550 C to 870 C Loss of SO 2 and O 2 Notice the shape of the first derivative curve which, as is shown below, almost exactly duplicates the shapes of the total evolved gas profile seen at the mass spectrometer. This is an important feature since it demonstrates that all the gases evolved have reached the mass spectrometer, although with the fixed gas evolved species here is less useful. 3

Curve #2 TGA Data With Total Evolved Gas Species As can be seen in the plot above, the shape of the total evolved species curve follows very closely with the TGA first derivative curve. The actual size of the individual EGA peaks will depend on the response of the MS to each of the evolved species, but nevertheless, much can be learned from the shapes of curves, including whether there has been any loss in the gas transfer process between the TGA and MS. There is a small delay perhaps 1 C - 2 C between the peaks on the TGA derivative and the MS peaks, although some of the difference is as a result of the frequency with which the data from the MS is collected with respect to the heating rate. In this example, data was collected at approximately 8.5 second intervals, which is equivalent to about 5.7 C between each data point at 40 C/min. 4

Curve #3 TGA Data With Individual Evolved Gas Species This display shows the individual ion species detected by the mass spectrometer alongside the TGA weight and TGA derivative weight loss curves. For a simple inorganic decomposition, these species are relatively straightforward to predict from the empirical equation. This display also demonstrates the ease with which the decomposition fragments from the empirical equation can be confirmed. 5