Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators

Similar documents
Neutron Fluence and Energy Spectra Around the Varian Clinac 2lOOC/23OOC Medical Accelerator

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

The neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room

Neutron fluence measurements of the Siemens Oncor linear accelerator utilizing gold foil activation

Secondary neutron spectra from modern Varian, Siemens, and Elekta linacs with multileaf collimators

Estimation of neutron and gamma radiation doses inside the concrete shield wall for 10 and 15 MV medical linear accelerators

Peter J. Biggs Ph.D., Massachusetts General lhospital, Harvard Medical School, Boston, MA 02114

Progress in Nuclear Science and Technology, Volume 6,

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom

Chapter 4. QUANTIFYING THE HAZARD II: DATA & ANALYSIS. The dose equivalents for spheres in air with 10 cm radius centred at a point in the

The Radiological Physics Center s standard dataset for small field size output factors

Quality-Assurance Check of Collimator and Phantom- Scatter Factors

Neutron Sources in the Varian Clinac 2100~2300C Medical Accelerator Calculated by the EGS4 Code

NEUTRON RADIATION FROM MEDICAL ELECTRON ACCELERATORS*

A comparison of methods for monitoring photon beam energy constancy

Radiation protection issues in proton therapy

Units S H I E L D I N H = D Q. H: Dose equivalent (Sv) D: Dose (Gy) Q: Quality Factor. 1Sv = 1 J/Kg. 1Gy = 1 J/Kg

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

IRPA 2 nd European Congress, Paris, May Session: RP in Medicine - Workers. Activation Products In Medical Linear Accelerators

NEUTRON SPECTROMETRY WITH BUBBLE DETECTORS

DESIGN OF A TREATMENT ROOM FOR AN 18-MV LINAC

In our recent paper [1], applicability of the self-activation of an NaI scintillator has been studied for the photo-neutron monitoring around

Peter J. Biggs Ph.D. Massachusetts General Hospital Harvard Medical School Boston, MA 02114

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

) for Varian TrueBeam high-dose-rate therapy beams

Radiation exposure of personnel during IORT: radiation protection aspects.

The Influence of Brass Compensator Thickness and Field Size on Neutron Contamination Spectrum in 18MV Elekta SL 75/25 Medical Linear Accelerator

A study on the cost of concrete shielding in a standard radiotherapy facility room

I. INTRODUCTION EXPERIMENTAL

BEAMnrc: a code to simulate radiotherapy external beam sources

ASSESSING LEAKAGE WORKLOADS OF MEDICAL LINEAR ACCELERATORS FOR IMRT AND TBI TECHNIQUES

Comparison of Primary Doses Obtained in Three 6 MV Photon Beams Using a Small Attenuator

Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling.

THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR*

International Journal of Scientific & Engineering Research, Volume 7, Issue 2, February-2016 ISSN

Michael Dunn Nuclear Data Group Leader Nuclear Science & Technology Division Medical Physics Working Group Meeting October 26, 2005

NEUTRONS FROM MEDICAL EJ,ECTRON ACCELERATORS '

Measurement of induced radioactivity in air and water for medical accelerators

Use of a radioactive check device for redundancy check of ionization chambers

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Response Characteristics of a Large-Area Ion Chamber with Various Radiotherapy Beams Makan Farrokhkish, B.Sc.

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

The effect of dose calculation uncertainty on the evaluation of radiotherapy plans

COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES *

Work supported by Department of Energy contract DE-AC03-76SF00515

Radiation Shielding for Megavoltage Photon Therapy Machines

Monte Carlo simulation of the photoneutron field in linac radiotherapy treatments with different collimation systems

Low energy linear acceleratordriven subcritical assembly

Chapter 10 Acceptance Tests and Commissioning Measurements

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515

Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.

Secondary Particles Produced by Hadron Therapy

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Today, I will present the first of two lectures on neutron interactions.

QUESTION 8 GIVEN. Electron beam kinetic energy = 20 MeV Peak current = 1 A Beam pulse length = 1 microsecond Beam pulse frequency = 10 Hz

PHYS 5020 Computation and Image Processing

8/3/2016. Chia-Ho, Hua St. Jude Children's Research Hospital. Kevin Teo The Hospital of the University of Pennsylvania

Radiation Shielding of a 230 MeV Proton Cyclotron For Cancer Therapy

Objectives. Tolerance Limits and Action Levels for IMRT QA. The Overall Process of IMRT Planning and Delivery. Chain of IMRT Process

Simulation Modeling in Dosimetry

Thermal and resonance neutrons generated by various electron and X-ray therapeutic beams from medical linacs installed in polish oncological centers

Radiation shielding for gamma stereotactic radiosurgery units

SPS Chapter Research Award Interim Report

arxiv: v2 [physics.med-ph] 29 May 2015

Dosimetric Properties of the Field Sizes of 12MV Photon Beams: A Monte Carlo Study

Magnetic removal of electron contamination in radiotherapy x-ray beams

New irradiation zones at the CERN-PS

Hospital Cyclotrons: Radiation Safety Aspects. Matthew Griffiths

Neutron Interactions Part I. Rebecca M. Howell, Ph.D. Radiation Physics Y2.5321

Absorption spectra variations of EBT radiochromic film from radiation exposure

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron

Electron therapy Class 2: Review questions

Using Prompt gamma ray emission to address uncertainties in proton therapy

Commissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100 linear accelerators

Applications of MCBEND

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER

Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit

Available online at ScienceDirect. Physics Procedia 69 (2015 )

1 Introduction. A Monte Carlo study

Evaluation of an electron Monte Carlo dose calculation algorithm for electron beams

Introduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose

NEUTOR: Neutrons Monitor for. Project financed by: Radiotherapy

Physics of Radioactive Decay. Purpose. Return to our patient

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Monitor Unit Calculations for Photon and Electrons. AAMD Meeting Raleigh, NC October 3, John P. Gibbons Chief of Clinical Physics

Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor

Physics. Sunday, March 4, :30 a.m. 10:00 a.m.

Shielding Design Considerations for Proton Therapy Facilities

ROC7080: RADIATION THERAPY PHYSICS LABORATORY LAB A : MECHANICAL MEASUREMENTS LINAC QA

ELECTRON INTERACTIONS

BD-100: THE CHALK RIVER NUCLEAR LABORATORIES NEUTRON BUBBLE DETECTOR*

Compton Camera. Compton Camera

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ

Secondary Radiation and Shielding Design for Particle Therapy Facilities

Turkey s first photonuclear reactions performed at Akdeniz University: Matter, Anti-matter, Pure Energy and Alchemy

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Neutron source strength measurements for Varian, Siemens, Elekta, and General Electric linear accelerators David S. Followill,* Marilyn S. Stovall, Stephen F. Kry, and Geoffrey S. Ibbott Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030 Received 20 December 2002; accepted for publication 14 March 2003 The shielding calculations for high energy 10 MV linear accelerators must include the photoneutron production within the head of the accelerator. Procedures have been described to calculate the treatment room door shielding based on the neutron source strength Q value for a specific accelerator and energy combination. Unfortunately, there is currently little data in the literature stating the neutron source strengths for the most widely used linear accelerators. In this study, the neutron fluence for 36 linear accelerators, including models from Varian, Siemens, Elekta/Philips, and General Electric, was measured using gold-foil activation. Several of the models and energy combinations had multiple measurements. The neutron fluence measured in the patient plane was independent of the surface area of the room, suggesting that neutron fluence is more dependent on the direct neutron fluence from the head of the accelerator than from room scatter. Neutron source strength, Q, was determined from the measured neutron fluences. As expected, Q increased with increasing photon energy. The Q values ranged from 0.02 for a 10 MV beam to 1.44( 10 12 ) neutrons per photon Gy for a 25 MV beam. The most comprehensive set of neutron source strength values, Q, for the current accelerators in clinical use are presented for use in calculating room shielding. 2003 American College of Medical Physics. DOI: 10.1120/1.1571671 PACS number s : 87.53. j, 87.52. g Key words: neutron source strength, shielding, photoneutron INTRODUCTION In most radiotherapy facilities, the older low-energy linear accelerators linacs are being replaced with new dual-energy linacs with photon beams 10 MV; as well, new treatment rooms are being built to accommodate the new dual-energy linacs. Thus, the number of linacs with high-energy photon beams 10 MV is increasing. These new linacs have the capacity to produce photoneutrons in the target, flattening filters and collimating devices if operated at energies above 10 MeV. The neutron component in treatment rooms where photon energies 15 MV are produced is significant and, as such, extra shielding is required. Measurements by Palta et al. 1 have shown a six-fold increase in the neutron dose around two Siemens Mevatron 77 accelerators Siemens Medical Solutions, Concord, CA as the photon energy was increased from 15 to 18 MV. The data required to perform the shielding calculations for the neutrons, especially the shielding in the treatment door, are lacking. National Council on Radiation Protection and Measurements NCRP Reports 51 and 79 addressed the shielding requirements for the higher x-ray energies and photoneutrons, respectively; however, they were published over 16 years ago and have been superceded by new research and concepts. 2,3 McGinley 4 described a method to calculate the amount of shielding needed in the treatment door to account for photoneutron production using what he referred to as the neutron source strength for a specific linac. McGinley provided a table 189 1526-9914Õ2003Õ4 3 Õ189Õ6Õ$17.00 2003 Am. Coll. Med. Phys. 189

190 Followill et al.: Neutron source strength measurements... 190 of neutron source strengths; however, the table lacked values for the variety of newer linac models produced by current manufacturers. Both NCRP Report 79 and McGinley s research described a method to determine the total neutron fluence n/cm 2 per unit x-ray dose at isocenter produced by several different linacs. 3,4 The total neutron fluence is given by the sum of the direct, scatter, and thermal neutron fluences: total dir sc th. A Monte Carlo analysis by McCall et al. 5,6 of the effects of a concrete room on the neutron spectrum indicated that each of the neutron fluence components could be determined using the following empirical relationships: total dir sc th aq/4 d 2 5.4aQ/S 1.26Q/S. Equation 2 can be broken down into the following three components: dir aq/4 d 2, sc 5.4aQ/S, th 1.26Q/S. The quantity a is the transmission factor for neutrons that penetrate the linac head shielding. The transmission factor a has a value of 1.0 for lead and 0.85 for tungsten. The quantity S is the treatment-room surface area in square centimeters, and Q is the neutron source strength in neutrons from the head of the treatment unit per x-ray dose Gy delivered at isocenter. The quantity d is the distance cm from the target to the point where the direct fluence is evaluated. From the relationship in Eq. 2 the total neutron fluence, total, can be determined. The total neutron fluence is then used to estimate the capture gamma dose at the maze entrance per x-ray dose delivered at isocenter. 6 Once the capture gamma dose is known at the maze entrance as well as the scattered and transmitted x-ray dose, the lead thickness of the maze door shielding can be calculated. Neutron source strength values Q values have been published for only a limited number of linac models and x-ray energies. The neutron source strengths for many of the current and newer linacs are not available in the literature. 4,7 12 The work presented here reports the Q values for 36 modern linacs representing 12 different combinations of linac models and x-ray energies. MATERIALS AND METHODS Estimates of total neutron fluence in and around a radiation therapy treatment room require that measurements of fast and thermal neutrons be made. For this work, the fast neutron fluence in a number of treatment rooms was measured using the gold-foil activation technique outlined previously in the literature. 1,13 Neutron measurements were made for Varian models 2100C, 2300CD, and 2500 Varian Associates, Palo Alto, CA, Siemens models Primus, KD, MD, and MD2 Siemens Medical Solutions, Concord, CA. Elekta models SL20 and SL25 Elekta Oncology Systems, Norcross, GA, and General Electric model Saturn 43 General Electric Medical Systems, Buc, France. Briefly, the gold foils 2 cm diameter 0.025 mm thick were weighed to subsequently account for the different foil sizes used and were then placed inside neutron polyethylene moderators Reactor Experiments, Sunnyvale, CA. The moderators were used to thermalize the fast neutrons such that the gold foils were exposed only to thermal neutrons. Gold foils have a high cross section of interaction for thermal neutron energies and are subsequently activated by the neutrons. The moderators also had a cadmium covering, which absorbs incident thermal neutrons so that only thermalized fast neutrons reached the gold foils. The moderators were placed on the treatment couch at various locations in the patient plane, as shown in Fig. 1. The foils were located in a plane perpendicular to the central axis of the beam located 100 cm from the target. The center of the foil was considered the point of measurement. 1 2 3

191 Followill et al.: Neutron source strength measurements... 191 FIG. 1. Location of the gold foils and moderators points A E for the neutron fluence measurements around the linear accelerators. The gold foils were exposed to the neutron fluences generated by x-ray beams of energies ranging from 10 to 25 MV. The accelerators used and the photon energies available from each are listed in Table I. The neutron fluence measured from an accelerator varies with the collimator setting. The secondary collimators were set to a 20-cm 20-cm opening for most measurements, which was used to represent a clinically relevant jaw setting. It has been shown however, that as the jaw setting is reduced to a 0-cm 0-cm opening, the neutron production increases. 1 The collimator setting for the intensity modulated radiation therapy setups was approximately 4 cm 4 cm. A total of 30 photon Gy at the depth of maximum dose was delivered to isocenter for all beams. The activated gold foils were allowed to decay for approximately one day before they were counted on an Eberline model BC-4 Beta Counter Eberline Instrument Corp., Santa Fe, NM. The day of decay allowed trace amounts of activated elements such as sodium, to decay so that only the 198 Au emissions were detected. The measured induced activity in each gold foil was a function of the fast neutron fluence rate that was incident on the moderator. The time of irradiation, time elapsed between irradiation, and foil counting, and the count time duration was recorded. Following a procedure described by Palta et al. 1 and in the American Association of Physicists in Medicine Task Group Report 19, 13 the count rate per gram for each activated gold foil was related to the fast neutron fluence through a neutron fluence per count rate calibration factor determined for the Eberline Beta counter. The calibration factor was determined by sending several gold foils of known mass to the National Institute of Standards and Technology to be irradiated in a known neutron fluence with an uncertainty of 2% at the one standard deviation level. These calibration foils were subsequently counted on the Eberline counter to determine the count rate to neutron fluence relationship. The calibration factor used was 3.515 10 6 (ncm 2 gram per counts s 1 ). The total neutron fluence excluding the thermal component was then determined at each point of measurement points A E. The measured count rate for each foil and the neutron fluence calibration factor determined for the counting system were used to calculate the fast neutron fluence at each point of measurement. The total neutron fluence is comprised of three components, thermal, scattered and direct, as indicated in Eq. 1. Solving Eq. 2 for Q, excluding the thermal component, yields:

192 Followill et al.: Neutron source strength measurements... 192 TABLE I. Neutron source strength Q values for various linac model and energy combinations. Manufacturer Model Nominal MV Q values ( 10 12 neutrons per Gy Std. Deviation ( 10 12 ) No. of Linac s Published Q values ( 10 12 ) Varian 1800 10 0.06 Varian 1800 15 0.76 Varian 1800 18 1.22 Varian 2100C 18 0.96 0.11 17 Varian 2100C b 18 0.87 1 Varian 2300CD 18 0.95 0.03 2 Varian 2500 24 0.77 1 Siemens MD2 10 0.08 1 Siemens MD 15 0.20 0.02 2 Siemens KD 18 0.88 0.10 2 Siemens KD 20 0.92 Siemens Primus a 10 0.02 1 Siemens Primus a 15 0.12 1 Siemens Primus b 15 0.21 1 Siemens Primus 15 1 0.20 Elekta SL-20 17 0.69 Elekta SL-20 18 0.46 1 Elekta SL-25 18 0.46 1 Elekta SL-25 22 2.37 Elekta SL-25 25 1.44 0.31 3 GE Saturne 41 12 0.24 GE Saturne 41 15 0.47 GE Saturne 43 18 1.32 1 1.50 GE Saturne 43 25 2.40 a With MiMIC device in place. b With the multileaf collimator set to a 3.8-cm 2 3.8-cm 2 field. Q dir sca 1 5.4 2 4 d S. The measured fast neutron fluence for each point of measurement and Eq. 4 were used to determine the neutron source strength for a linac and energy combination. Each Q value determined for a linac and energy combination was an average value for all of the measurement points A E see Fig. 1. Measurements at isocenter would have resulted in an incorrect elevated neutron source strength value for the linac due to extra photoneutron production in the moderator. None of the measurements of neutron fluence at the isocenter were included in this analysis. RESULTS AND DISCUSSION The results of the neutron fluence measurements as a function of room surface area can be seen in Fig. 2. Depending on the location of the measurement point, the magnitude of the neutron fluences differ. However, there is little to no dependence on room surface area. Scatter and thermal neutron fluence components at any point within a treatment vault depend on the room surface area. The data in Fig. 2 suggest that the measured neutron fluence in the patient plane is primarily from the direct neutron fluence originating from the head of the machine and very little contribution is from scattered fast neutrons originating from the rooms walls. Table I lists the neutron source strengths Q determined for the linac models and energies measured. The Q values in 10 12 n per photon Gy) range from 0.02 for a Siemens Primus 10 MV beam with MiMIC device attached to 1.44 for an Elekta SL25 25 MV x-ray beam. Table I adds another 13 Q values to the data already published. 4,11 Many of the Q values in Table I are average 4

193 Followill et al.: Neutron source strength measurements... 193 FIG. 2. The measured neutron fluence (n/cm 2 10 6 per photon Gy) produced by the 18 MV x-ray beam from 15 Clinac 2100C linear accelerators at six locations in the patient plane as a function of room surface area. values for measurements on more than one linac of a specific model and energy. In addition to the Q values for the new linac and energy combinations, five linac/energy combinations already had Q values listed in the literature. 4,11 These five sets of redundant data are in good agreement considering the inherent uncertainty 20% in neutron fluence measurements. Apparent exceptions to the good agreement of the results are the relationship between the Q values for the Elekta SL25 22 and 25 MV beams, which are 2.37 10 12 and 1.44 10 12, respectively, and the Philips SL20 17 and 18 MV beam s Q values that are 0.69 10 12 and 0.46 10 12, respectively. There is also a difference between the 24 MV beam from a Clinac 2500 and the 18 MV beam from a Clinac 2100C. The Q value for the 2100C 18 MV beam is greater than that for the 24 MV beam and this is best explained by differences in head shielding. In general, for a given linac model, the number of neutrons produced is expected to increase with energy. There are differences in the photon energies, but the change in Q is in the opposite direction than would normally be expected. There may also be differences in the head shielding which would explain some of the differences noted. The Q value for the SL25 25 MV beam presented in this work represents the average of three linacs and has a standard deviation of 0.31 10 12, whereas the data published by McGinley comes from measurements of one linac which gives one more assurance that the Q value from the three accelerators is correct. 4,10 The Philips SL20 data (Q 0.46) measured in this work agrees with the measurements for the SL25 (Q 0.46) at the same nominal x-ray energy. The difference between the 17 and 18 MV beams noted in Table I is most likely attributed to differences in x-ray energy and uncertainty in the neutr on measurements. Another contribution to the observed differences in Q values between accelerators might be due to differences in the head design and shielding. Most current accelerators of the same manufacturer have similar head designs, but subtle changes are made during the manufacturing process and the head design information is considered proprietary by the manufacturers. The subtle changes to the head design probably do not change the neutron production significantly. Some accelerators such as the Clinac 2100C series did undergo a change in head design when the MLC was added. 11 The Q values for the Clinacs with and without MLCs were indistinguishable, indicating no effect on the neutron production even with a substantial change to the machine head. The Q values are not significantly different between different models of accelerators from the same manufacturer and energy, however they do differ between accelerators from different manufacturers. CONCLUSION The data in Table I are a compilation of all of the published Q values to date. These data will enable physicists to perform the appropriate shielding calculations for most linac/energy combinations following published guidelines for teletherapy shielding. 4

194 Followill et al.: Neutron source strength measurements... 194 *Author to whom correspondence should be addressed; fax: 713 794-1364; email address: dfollowi@mdanderson.org Email address: mstovall@mdanderson.org Email address: sfkrv@mdanderson.org Email address: gibbott@mdanderson.org 1 J. R. Palta, K. R. Hogstrom, and C. Tannanonta, Neutron leakage measurements from a medical linear accelerator, Med. Phys. 11, 498 501 1984. 2 Radiation Protection Design Guidelines for 0.1 100 MeV Particle Accelerator Facilities, NCRP Report No. 51 National Council on Radiation Protection and Measurements, Bethesda, MD, 1977. 3 Neutron Contamination from Medical Accelerators, NCRP Report No. 79 National Council on Radiation Protection and Measurements, Bethesda, MD, 1984. 4 P. H. McGinley, Shielding Techniques for Radiation Oncology Facilities Medical Physics Publishing, Madison, WI, 1998. 5 R. C. McCall, T. M. Jenkins, and R. A. Shore, Transport of acceleration produced neutrons in a concrete room, IEEE Trans. Nucl. Sci. NS-26, 1593 1602 1979. 6 R. C. McCall, P. H. McGinley, and K. E. Huffman, Room scattered neutrons, Med. Phys. 26, 206 207 1999. 7 R. C. McCall, Neutron yield of medical accelerators, SLAC-Pub-4480, Stanford, CA 1987. 8 P. H. McGinley and J. C. Landry, Neutron contamination of x-ray beams produced by the Varian Clinac 1800, Phys. Med. Biol. 33, 975 980 1989. 9 P. H. McGinley, Photoneutron dose levels produced by the Siemens KD medical accelerator, Emory Univ. J. Med. 2, 40 46 1988. 10 P. H. McGinley, S. Ghavidel, and J. C. Landry, A study of photoneutron dose levels produced by the Philips SL medical accelerators, Radiat. Prot. Manage. 10, 45 50 1993. 11 J. P. Lin, T. C. Chu, S. Y. Lin, and M. T. Liu, The measurement of photoneutrons in the vicinity of a Siemens Primus linear accelerator, Appl. Radiat. Isot. 55, 315 321 2001. 12 J. O. Fenn and P. H. McGinley, Stray photoneutron fields produced by the GE Saturne accelerator, Radiat. Prot. Manage. 12, 39 45 1995. 13 American Association of Physicists in Medicine, Neutron measurements around high energy x-ray radiotherapy machines, AAPM Report No. 19, New York 1986.