Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Similar documents
The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Automated and accurate component detection using reference mass spectra

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Monitoring Protein PEGylation with Ion Exchange Chromatography

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

of mass spectrometry

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

HILIC Method Development in a Few Simple Steps

Determination of trace anions in concentrated hydrofluoric acid

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Elemental Analysis: CHNS/O characterization of polymers and plastics

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Drugs of Abuse by Direct Sample Probe Using Single Quadrupole GC/MS

ab Alkaline Phosphatase Assay Kit (Luminometric)

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Dionex IonPac AS28-Fast-4µm column

1. Intended use. 2. Introduction. 3. Assay description. 4. Limitations. Instruction For Use Gyrolab Generic PK Kit

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Benzoylecgonine in Urine by SAMHSA GC/MS

ab83360 Ammonia Assay Kit

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Certified Reference Material - Certificate of Analysis

ENVIRONMENTAL analysis

Determination of Volatile Substances Proof of Food Adulteration

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Enhancement of Linearity and Response in Charged Aerosol Detection

Introduction to Fourier Transform Infrared Spectroscopy

Non-Interfering Protein Assay Kit Cat. No

Determination of Elemental Sulfur in Explosives and. Explosive Residues by Gas Chromatography-Mass Spectrometry. Kimberly S.

ab Caspase 3, Caspase 8 and Caspase 9 Multiplex Activity Assay Kit (Fluorometric)

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Complete Materials Deformulation Using TGA-IR

Fast methods for the determination of ibuprofen in drug products

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

FORMIC ACID in URINE by GC/MS-headspace - Code GC74010

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry

ab83360 Ammonia Assay Kit

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Transcription:

Mass Spectral Studies of Polypropylene Chromatographic Well Plates Brian King, Thermo Fisher Scientific, Runcorn, Cheshire, UK Detlev Lennartz, Thermo Fisher Scientific, Langerwehe, Germany White Paper 2115 Key Words GC Extraction, Extractables, Silicone Mats, Sensitivity, Well Plates, Polypropylene Introduction In a previous independent study, commercially available deep well microplates were shown to harbor significant levels of extractable compounds. It has been proposed, that this is from the use of lower grade polypropylene polymers and/or deforming reagents used during the molding process itself. This study has been undertaken to identify low extraction 96-well deep well plates that are suitable for use with high sensitivity chromatographic methods. Mass spectral data shows that persistent contamination from a range of compounds found in the raw polymer master batch is evident in many of the micro plates tested. Background The effect of extractables identified in this report is complex and depends on the exact application for which the plate is designed. However, it is clear that long chain hydrocarbon contaminants will certainly cause extraneous signals in any spectroscopic method including UV detection in HPLC. These unwanted small molecules can easily mask drugs of abuse or their metabolites and cause ion suppression, which leads to inaccurate quantification. Scientists using deep well micro plates must ensure regular screening of the plates in use if they are to avoid such unwanted contamination or re-contamination of (expensively) cleaned up samples during storage in well plates. Testing can be carried out in-house, but this study shows that plates can be pre-tested for their suitability for applications involving injection into chromatographic instruments with mass spectral detection. Thermo Scientific WebSeal system for chromatographers.

2 Method A new, unused plate was selected from each batch. Testing for polymer leachate and extractable contamination was performed by incubating (overnight) an appropriate volume of HPLC grade methanol in 12 wells in each sample plate. The plates were sealed and left to stand overnight. After overnight incubation, 1 µl aliquot of each well sample was subjected to analysis on a GC-MS system using splitless injection at 25 C. Separation was performed on a capillary column using the appropriate temperature gradient. Detection was by positive ion EI-MS (Electron Ionization-Mass Spectrometry). GC/MS Instrument: Software: Column: Carrier Gas: Flow Rate: Oven Program: Thermo Scientific ISQ Series GC-MS with Thermo Scientific Triplus AS Autosampler Thermo Scientific Xcalibur Thermo Scientific TraceGOLD TG-5MS, 3 m.25 mm.25 μm Helium 1.2 ml/min C, hold for.5 min; 15 C/min to 15 C, hold for 1 min; 1 C/min to 29 C, hold for 5 min Inlet Temperature: 25 C; Split flow: 5 ml/min Injection Volume: 1 μl splitless MS Transfer Line: 29 C MS Ion Source: 23 C MS Detection: Positive EI; Full scan 5 to 65 Results A number of candidate plates showed minimal extraction peaks as shown in Figures 1 and 2. A number of these plates have been submitted to stringent certification to allow assurance for their usage in testing MS analysis. Where a number of extraction peaks were determined antioxidants (stabilizers) were most likely to be found in low levels. Examples are shown in Figure 3 where a single peak extract was identified and Figure 4 where isolation and identification of discrete peaks are determined from the MS spectra. Where multiple peaks are determined the plate has a range of extractants including antioxidants and cannot be recommended for chromatographic use. An example shown in Figure 5. External contamination sources should not be ignored and the use of silicones and solvents with the plates can cause leaching and a transfer of silicone oligomers as shown in Figure 6. Discrete peaks for mold release can be seen in Figure 7 for a rejected plate format.

3 Contaminant Groups The extracted contaminants were compared against an NIST database to identify the chemicals involved. These chemicals were shown to be in a group which acts as plastics enhancers and mold flow agents to assist ease of injection molding. It is likely that the above chemicals are incorporated in the polymer mix at an early stage to enhance ease of molding and fast cycle time of the mold to produce the relevant products. Table 1: Contaminants and their commercial use. Contaminants Eucrylamide Dodecanoic Acid Hexadecanoic Acid, Methyl Ester Octadecanoic Acid, Methyl Ester 9-Octadecanoic Acid Bis 3, 5, Dimethyl 4-hydroxy Benzenepropionic Acid Ester Commercial Use Anti-caking agent, lubricant Mold release agent Softener, accelerator activator Plastics lubricant Plastics lubricant, release agent Antioxidant Conclusion It is possible to select microplates manufactured with high purity polypropylene for use as a certified chromatography injection or storage plate. Some plates were determined to have limited extraction components and may be used for routine HPLC and other chromatographic techniques (with water samples or limited polar solvent content) if the extraction products are avoided. The use of methanol in this study is significant. As a polar solvent it is very good at dissolving small molecules and has been shown to extract most non-polymer bound compounds from polypropylene microplates. It is also a very suitable solvent for GC-MS. Contamination of plates from external sources were also found, which, while isolated, did show that transport, storage and packaging must also be considered when selecting a suitable chromatography plate. Recommendations It is recommended that deep well microplates used for long-term storage of samples in organic solvents are regularly assessed to check for extractable contamination. Consequently, it is important that this phenomenon be understood before your stored compounds or samples are further analyzed. The introduction of a certified range of microtiter and deep well plates has been carried out to aid the user when utilizing MS and sensitive chromatographic analysis. These plates show minimal extraction with methanolic solutions as the raw material is a medical grade polypropylene with good solvent compatibility and minimal additives. WebSeal plates (deep well plates with round and square wells with 1 and 2 ml volume) and PTFE covered silicone mats were used to produce the results in Figures 1 and 2. A number of plates with low levels of additive extracts have been chromatographically assessed and have been found to give low background suitable for routine HPLC applications. This study also shows that external contamination of plates should be avoided. The use of silicone fluids may give a characteristic profile in total ion current (TIC) mode due to the presence of Si (n) oligomers. Finally, some plates exhibit significant contamination from plastic extraction of mold release and lubrication products and should not be used due to potential interference and suppression in this mode.

4 Blank WebSeal Test Plate 9 27.2 27.48 25.59 24.49 23.85 23.36 7.87E6 Blank1 9 27.3 27.82 26.48 25.59 24.94 24.21 23.7 7.22E6 MTP96d 7 22.98 22.68 5 22.26 21.98 21.67 21.38 3 2.75 2 2.23 19.52 1 18.84 5.26 6.96 17.47 8.73 9.52 1.79 13.13 15.23 16.37 6 8 1 12 14 16 18 2 22 24 26 28 3 7 22.95 5 22.67 22.41 22.12 21.91 3 21.56 21.16 2 2.67 19.96 1 19.21 5.24 5.46 17.95 7.3 8.73 9.1 1.78 12.74 14.5 15.7 16.85 6 8 1 12 14 16 18 2 22 24 26 28 3 27.11 27.12 9 9 73.11 7 5 73.11 3 2 77.8 133.1 193.11 135.14 281.14 1 119.1 163.8 221.2 267.13 283.18 341.18 1.16 429.24 461.18 489.28 535.27 563.33 625.17 5 15 2 25 3 35 45 5 55 65 7 5 119.1 3 193.11 77.9 133.1 2 281.16 96.8 1 149.14 267.13 283.19 221.2 341.17 1.19 429.27 461.2 489.24 535.26 563.36 9.31 639.21 5 15 2 25 3 35 45 5 55 65 Figure 1: Comparison of blank methanol sample injection and a test plate with minimal extraction components. Figure 2: Certification traces used for characterization of low background polypropylene plates. No significant additional peaks are seen in the sampled plate, compared to an injection blank. This has a low background suitable for MS analysis with minimal suppression from potential extraction compounds. 2 2 TIC Blank Certified WebSeal PP Plate 1.E7 acnblk_3 1.E7 platesamples _p1_16 1. 1.5 11. 11.5 12. 12.5 13. 13.5 14. 14.5 15. 15.5 16. 16.5 17. 17.5 18. 18.5 19. 19.5 2. 2 GC Spectra Blank 8.37E4 platesamples_p1_17#25 9-5 RT: 1.-2. AV: 2942 T: {,} + c EI Full ms [5.-65.] 2 Certified WebSeal PP Plate 7.71E4 platesamples_p1_16#25 9-5 RT: 1.-2. AV: 2942 T: {,} + c EI Full ms [5.-65.] 5 15 2 25 3 35 45 5 55 65

5 Figure 3: Example of a chromatography tested plate with low levels of extractants which may be seen at low levels. 9 7 5 23.67 23.51 23.26 23.1 22.86 22.1 24.96 24.55 24.26 1.77E6 2252a 3 2 1 22. 14.11 21.66 21.14 2.78 2.9 5.2 8.73 19.51 6.95 1.77 17.68 18.62 7.3 9.15 11.51 12.75 14.77 16.61 1.62 6 8 1 12 14 16 18 2 22 24 9 73.11 27.14 7 5 3 2 1 193.12 133.11 96.1 281.18 119.1 177.1 221.23 267.16 341.17 295.21 1.2 429.26 475.27 53.32 549.33 566.31 625.33 5 15 2 25 3 35 45 5 55 65 Figure 4: Match of extractant example from a lower quality polypropylene plate. Match was 95.71% when reprocessed. Compound is a component of a Phenolic Primary Antioxidant for processing and long-term thermal stabilization of polyolefins. 95 9 85 75 7 146.97 277.48 65 292.23 55 5 45 57.5 219.12 35 3 25 2 15 1 5.1 86.97 115.11 161.1 22.81 234.97 259.26 3.71 534.21 328.24 346.3 429.6 458.86 477.99 516.69 553.92 578.77 5.18 641.96 5 15 2 25 3 35 45 5 55 65

6 Figure 5: Example of methanol extraction from a poor quality polypropylene plate with antioxidant extract showing at 13.31 min with high extraction background from C16 C18 fatty acids. 9 7 5 3 2 1 5.15 6. 6.83 7.36 8.2 9.3 1.23 1.41 11.42 13.8 13.31 15.23 12.94 14.21 14.73 12. 16.54 19.85 1.26E8 19.67 MeOH_Soa 19.38 k4465_2 19.21 19.7 18.62 18.57 18.1 17.78 5 6 7 8 9 1 11 12 13 14 15 16 17 18 19 2 2 57.9 71.11 97.13 111.15 125.16 139.17 177.13 14.1 253.1 281.14 327.7 5.14 355.19 429.26 553.49 479.31 582. 628.61 1.15E6 MeOH_Soak4465_2#44 2-4852 RT: 5.-2. AV: 4411 T: {,} + c EI Full ms [5.-65.] 2.3E7 MeOH_Soak4465_2#28 85 RT: 13.31 AV: 1 T: {,} + c EI Full ms [5.-65.] 2 78.5 135.8 178.8 28.13 238.28 328.99 389.61 416.42 4.2 554.32 1.64 628.83 5 15 2 25 3 35 45 5 55 Figure 6: Example of external contamination with silicone fluid from processing of plates. This can also be observed if unprotected silicone sealing mats have been exposed to solvents during storage on the plates. 95 9 85 75 6.86 8.61 1.61E8 PlateSampl es_p1_12 7 65 55 5 45 1.73 12.75 14.52 16.1 17.54 35 3 25 2 3.96 4.4 7.63 18.83 2.2 21.14 15 1 5 4.28 5.23 1.46 13.5 6.43 7.96 13.13 19.37 9.55 1.96 15.86 18.47 11.97 16.64 22.18 23.17 24.3 25.65 25.85 3 4 5 6 7 8 9 1 11 12 13 14 15 16 17 18 19 2 21 22 23 24 25 26 27

7 Figure 7: Contaminated plate extraction with example of multiple component contamination. 95 9 85 75 7 65 13.6 1.24E9 PlateSampl es_p1_13 White Paper 2115 55 5 45 7.63 1.46 35 3 25 13.13 2 15 7.96 1 5 6.36 6.52 4.4 5.98 8.16 3.13 6.66 11.15 4.28 5.71 8.65 12.7 1.33 13.27 14.1 15.86 16.44 17.77 19.56 2.82 21.69 22.62 23.13 24.21 25.3 25.61 3 4 5 6 7 8 9 1 11 12 13 14 15 16 17 18 19 2 21 22 23 24 25 26 27 Products Used in This Paper WebSeal Well Plates, Non Coated, Non Sterile, Chromatography Certified Description Well Format Total Volume (µl) Working Volume Range µl/well Catalog Number Pack of 96-Well Microplate, Round Well U-Shape, 7 mm Diameter 5 9 1-P211 5 U-Shape, 7 mm Diameter 5 9 1-P21 5 96-Well Microplate, Square Well V-Shape, 8 mm Diameter 2 5 19 1-P212 5 V-Shape, 8 mm Diameter 2 5 19 1-P22 5 WebSeal Mats, Non Sterile Description Color Material Well Design Pre-slit Catalog Number Pack of WebSeal Mat Blue Silicone/PTFE 96 Round Well Flat Base, Step, 7 mm Diameter No 1-M111 5 Blue Silicone/PTFE 96 Round Well Flat Base, Step, 7 mm Diameter Yes 1-M112 5 Blue Silicone/PTFE 96 Square Well Flat Base, Step, 8 mm Diameter, Large Penetration Area No 1-M12 5 Blue Silicone/PTFE 96 Square Well Flat Base, Step, 8 mm Diameter, Large Penetration Area Yes 1-M122 5 For more information, visit our website at www.thermoscientific.com/chromatography 215 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. USA and Canada +1 332 3331 Australia 13 735 292 (free call domestic) China 81 5118 (free call domestic) 65 5118 France +33 ()1 92 48 34 Germany +49 () 2423 9431 2 or 21 India +91 22 6742 9494 +91 27 1766 2352 WP2115-EN 815M Japan 12 753 67 (free call domestic) 12 753 671 (fax) Korea +82 2 342 8 United Kingdom +44 () 1928 534 11 New Zealand 933 966 (free call domestic) Singapore +65 6289 119 All Other Enquiries +44 () 1928 534 5 Technical Support For advice and support, please visit our website: www.thermoscientific.com/chromexpert