Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Similar documents
VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Separation and Characterization of Indian and Australian Sandalwood Oils

Quantification of Rice Aroma, 2-Acetyl-1-Pyrroline (2-AP), Using TurboMatrix Headspace Trap Coupled with GC/NPD and GC/MS

Author. Abstract. Introduction

U.S. EPA Method 8270 for multicomponent analyte determination

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated

Benzoylecgonine in Urine by SAMHSA GC/MS

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analytical Trap Comparison for USEPA Method 8260C

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Automated Characterization of Compounds in Fire Debris Samples

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Advances In Fire Debris Analysis

Headspace Technology for GC and GC/MS: Features, benefits & applications

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Tar analysis by Solid Phase Adsorption (SPA) associated with Thermal Desorption (TD) and Gas Chromatography (GC) analysis

Method for the determination of dimethyl sulfate

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

Determination of Volatile Substances Proof of Food Adulteration

Methanol Extraction of high level soil samples by USEPA Method 8260C

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Chemistry Instrumental Analysis Lecture 27. Chem 4631

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

ELECTRONICALLY REPRINTED FROM. Volume 34 Number 3 MARCH

2401 Gas (liquid) Chromatography

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Application Note 032

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Automated PIONA Analysis by ASTM D8071 and VUV PIONA+

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

HYPHENATED TECHNOLOGY GUIDE

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Volatile Organic Compounds in Every Day Food

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Detection of Chemical Warfare Agents by Transportable GC/MS

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

Application. Gas Chromatography February Introduction

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Gas Chromatography (GC)

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Drugs of Abuse by Direct Sample Probe Using Single Quadrupole GC/MS

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles

Opiates in Urine by SAMHSA GC/MS

Routine MS Detection for USP Chromatographic Methods

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

HYPHENATED TECHNOLOGY GUIDE

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Analysis of BTEX in Natural Water with SPME

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

TDTS. Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination

Gas Chromatography. Introduction

Analysis of Polyester by a Combination of Curie-Point Heating Techniques and Preparative Liquid Chromatography

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Determination of Volatile Organic Compounds in Air

Transcription:

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene investigation through analytical chemistry. If a fire s ignition is suspicious, the analysis will include gas chromatography (GC) as a means to detect the presence of a liquid accelerant; this analysis may include mass spectrometry (MS) as the initial detector or as confirmatory analysis. Gas chromatography mass spectrometry (GC/MS) provides a means to identify the accelerant through its chromatographic fingerprint; mass spectral data confirms this determination by identifying marker components of this fingerprint and their fragmentation. Author William Goodman PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA GAS CHROMATOGRAPHY / MASS SPECTROMETRY A P P L I C A T I O N N O T E The ASTM classification table (Table 1 - Page 2) breaks liquid accelerants into 6 major classes. This table is a useful tool to assist the analyst in the identification of an accelerant. Each class of accelerants is described across a row of the table; included in this description are characteristic compounds and ion fragments. Relating the detected ion fragments of the analysis to specific compounds will link the experimental data to an ASTM classification of the accelerant 1. Prior to GC/MS analysis, fire-debris samples require sample collection and sample preparation to acquire a representative sample and transform the matrix of this sample into one which is appropriate for GC analysis. Recommendations for samplematrix type, sample collection, storage containers, and sample preparation methods are found throughout literature and within methods developed by the American Society for Testing and Materials (ASTM International). The sample preparation methods found in these sources generally fall into 3 classes: headspace, solvent extraction, and steam distillation. Sample preparation for GC/MS analysis is typically performed by headspace or solvent extraction.

Dynamic headspace with automated thermal desorption (ATD) sample collection is discussed in the article Sample preparation for the chemical analysis of debris in suspect arson cases by Bertsch and Zhang 2. This article presents the advantages and disadvantages of the common types of sample preparation. In this article, dynamic headspace and thermal desorption are described as very sensitive and clean methods. The only disadvantage is the inability to perform multiple analyses on a single sample tube. The goal of this study is to eliminate the use of solvents for sample preparation and to implement sample recollection on the ATD, allowing multiple analyses of a single sample tube. The discussion will include sample matrix, sample collection and storage, sample containers, sample introduction and analysis by GC/MS using the unique solutions provided by the technology of the TurboMatrix 650 ATD and Clarus GC/MS system. Experimental Sample Collection For this study, test samples of lumber were burned with and without an added liquid accelerant gasoline on a clean, concrete surface. The samples were extinguished with water and stored in a 0.5-L glass jar. Sample Preparation In preparation for GC/MS analysis, a dynamic headspace extraction was performed on the debris samples. This procedure consisted of heating the jar of debris to 80 C and equilibrating it at this temperature for 20 minutes. The contents of the jar were purged onto a metal ATD sample-collection tube packed with Tenax TA with a flow of clean, dry air at 50 ml/min. The sample collection time was 2 minutes; during this time the ATD sample tube remained at room temperature (21 C), outside the oven. Tenax TA was used as the adsorbent throughout this study; arson samples often contain a large amount of water Tenax was selected for its hydrophobic properties so that less water would be introduced into the analytical system. Standards were obtained from Restek (Bellefonte, PA). The weathered gasoline kit was used here; this included un-weathered, as well as 25, 50, 75 and 99% weathered standards. 5 μl of the liquid standard, diluted in methanol, was spiked directly onto an ATD tube; the tube was then purged with clean, dry air at 50 ml/min for 5 minutes at room temperature to remove the methanol. Instrumental Analysis The instrumental platform used in this study was the PerkinElmer TurboMatrix 650 ATD coupled to the PerkinElmer Clarus 600 T GC/MS this system offers the laboratory an analytical platform for the analysis of liquid accelerants. The heated zones of the TurboMatrix 650 ATD were set to 250 C to prevent condensation of high-molecularweight compounds. A dual-stage desorption with recollection of the secondary split allows for re-analysis and/or for archival purposes. The pneumatics operated in constant-flow mode, utilizing an auxiliary temperature sensor in the GC oven (Table 2 - Page 3). Table 1: ASTM Accelerant Classification Table ASTM Method E1387-01. Class Dominant Number Class Name Hydrocarbon Range Component Classes Diagnostic Ions (m/z) 1 Light Petroleum Distillates C 4 - C 8 2 Gasoline C 4 - C 12 Naphthalene 128,142,156 3 Medium Petroleum Distillates C 8 - C 12 4 Kerosene C 9 - C 16 5 Heavy Petroleum Distillates C 10 - C 23 Naphthalenes 128,142,156 0 Miscellaneous Variable 0.1 Oxygenated Solvents 0.2 Isoparaffins Alcohols 31,45 0.3 Normal Alkanes Ketones 43,58 0.4 Aromatic Solvents Esters 43,73 0.5 Naphthenic/paraffinic solvents Terpenes 93,136 Adapted from the American Society for Testing and Materials Methods 2

As indicated on Table 3, a 0.32-μm diameter fused-silica transfer line connected the ATD to the Elite-1MS capillary column. The GC oven program began at 35 C and was ramped at 15 C/min to 280 C; the total GC run time was 20 min. The mass spectrometer cycled at 4 scans per second across a range of m/z 30-500; the heated zones of the MS were maintained at 280 C. The instrument was controlled by and the data was reported with TurboMass GC/MS software. Table 2: Detailed Operating Parameters for TurboMatrix 650 ATD Instrument. Thermal Desorption PerkinElmer TurboMatrix Unit: 650 ATD Transfer-Line Temperature: 250 C Valve Temperature: 250 C Tube Temperature: 250 C Trap Temperature (Low): -25 C Trap Temperature (High): 250 C Trap Heating Rate: Ballistic Trap Hold Time: 10 min Desorb Time: 1 min Desorb Flow Time: 1 min GC Cycle Time: 30 min Inlet Split: On Outlet Split: On Injections Per Tube: Multiple Split Mode: Flow Thermal Desorption Mode: 2-Stage Desorb & Re-collect Dry Purge: Off Carrier Mode: Flow Column: Re-collect Flow: 50 ml/min Desorb: Column Setting During Desorb: Transfer-Line Type: Short Trap-Packing Type: Tenax TA Tube Type: Metal Tube-Packing Type: Tenax TA Table 3: Detailed Operating Parameters for GC/MS System. Gas Chromatograph: PerkinElmer Clarus 600 GC Analytical Column: Elite - 1MS (15 m x 0.25 mm x 0.1 μm) Carrier Gas Type: He Oven Program: Temp Hold Time Rate 35 C 1 min 15 C/min 280 C 3 min END Detector Type: Mass Spectrometer Mass Spectrometer: PerkinElmer Clarus 600 MS GC Inlet Line Temp: 280 C Ion Source Temperature: 280 C Function Type: Full Scan Full-Scan Range: m/z 30-500 Full-Scan Time: 0.2 sec InterScan Delay: 0.05 sec Results In preparation for the analysis of an arson sample, a series of weathered-gasoline standards were analyzed. Pictured in Figure 1 is the weathered gasoline series as you can see, the total ion fingerprint of gasoline vs. 99% weathered gasoline is quite different. In Figures 2 through 4 (Page 4), each of the characteristic compound classes identified by the ASTM are displayed in detail. Figure 1. Total ion chromatograms of un-weathered (bottom) and 99% weathered (top) gasoline standards. 3

Figure 2 displays the alkylbenzene or BTEX (benzene, toluene, ethylbenzene, xylene) compounds as you can see, these are the primary components of un-weathered gasoline. Resulting from their low molecular weight and high volatility, the alkylbenzenes are first to evaporate and comprise a very low percentage of 99% weathered gasoline. Conversely, the naphthalenes (Figure 3) comprise a small percentage of un-weathered gasoline, are much less volatile, and represent the majority of the 99% weathered gasoline fingerprint. The alkanes in gasoline (Figure 4) span a wide volatility range this results in high mass discrimination in the highly-weathered standards. Figure 4. The comparison of the extracted ion chromatogram of alkanes (m/z 43, 57, 71) in un-weathered (bottom) and 99% weathered (top) gasoline. Figure 2. The comparison of the extracted ion chromatogram of alkylbenzenes (m/z 91,106,120) in un-weathered (bottom) and 99% weathered (top) gasoline. Depending on the heat and duration of the fire and the volume of accelerant used, the fingerprint of an arson sample with a particular accelerant will vary across the range of weathering. Each of the compound classes which make up a particular accelerant should be monitored in both reference material and sample analysis to ensure correct identification. This study includes the analysis of samples from a test fire burned with and without gasoline as the accelerant. Chromatograms from this analysis are presented in Figure 5 in this case, presence of an accelerant is quite obvious. Figure 3. The comparison of the extracted ion chromatogram of naphthalenes (m/z 128, 142, 156) in un-weathered (bottom) and 99% weathered (top) gasoline. Figure 5. The comparison of the analysis of a test sample of wood burned without gasoline (top) with a low split (center) and high split (bottom) run of a wood sample burned with gasoline. 4

Throughout the study, the ATD was operated in sample re-collection mode sample re-collection, as implemented in the TurboMatrix 650 ATD system, is activated during the second stage of thermal desorption. The effluent desorbing from the ATD cold trap is split before entering the GC transfer line, in this case 50:1. This split results in 49 ml/min flowing back onto the sample tube (re-collected), and down the analytical column; approximately 98% of the sample is available for reanalysis and archival. In the test sample, the amount of gasoline was so high that it was necessary to re-analyze the sample tube at a significantly higher split ratio to obtain a chromatogram with useful fingerprint data (Figure 5). Once adequate split-ratio conditions were achieved, the sample was re-analyzed 15 consecutive times under identical conditions, each time splitting 50:1 this is demonstrated in Figure 6. The fingerprint of the arson sample can be reproduced time after time, for confirmatory analyses. Conclusion As with all forensic analyses, the data generated in arson investigation must be legally defensible, creating the need for duplicate analyses and sample archival. Sample re-collection allows laboratories to perform multiple analyses of the same sample in addition to preserving both the instrument sample (thermal desorption tube) as well as the bulk debris sample. Thermal extraction combined with dynamic headspace and automated thermal desorption is clearly a clean and very sensitive sampling technique in arson investigation. Additionally, this sample preparation technique eliminates the need for solvent extraction, reducing the use of carbon and eliminating solvent-disposal costs, analyst s preparation time and solvent exposure hazard. References 1. American Society of Testing and Materials. ASTM Method E1387-01 Standard test method for ignitable liquid residues in extracts from fire debris samples by Gas Chromatography. In: Annual Book of ASTM Standards. Philadelphia, PA: ASTM, 2002:543-53. 2. Bertsch W., Zhang Q. Sample preparation for the chemical analysis of debris in suspect arson cases. Anal Chim Acta 1990; 236:183-95. Figure 6. The analysis, re-collection and re-analysis of a single ATD sample tube 5 times, under sample re-collection mode of operation. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA Phone: (800) 762-4000 or (+1) 203-925-4602 For a complete listing of our global offices, visit /lasoffices 2008 PerkinElmer, Inc. All rights reserved. The PerkinElmer logo and design are registered trademarks of PerkinElmer, Inc. TurboMass and TurboMatrix are trademarks and Clarus and PerkinElmer are registered trademarks of PerkinElmer, Inc. or its subsidiaries, in the United States and other countries. All other trademarks not owned by PerkinElmer, Inc. or its subsidiaries that are depicted herein are the property of their respective owners. PerkinElmer reserves the right to change this document at any time without notice and disclaims liability for editorial, pictorial or typographical errors. 008317_02