Maximizing Sample Throughput In Purge And Trap Analysis

Similar documents
System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Solid Phase Micro Extraction of Flavor Compounds in Beer

Optimizing. Abstract: is standardd. procedures. altered to

Maximizing Production While Minimizing Costs

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Solid Phase Micro Extraction of Flavor Compounds in Beer

Optimizing Standard Preparation

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Methanol Extraction of high level soil samples by USEPA Method 8260C

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Determination of Volatile Organic Compounds in Air

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

EPA TO-17 Volatile Organic Compounds

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Ed George and Anaïs Viven Varian, Inc.

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

2017 EPA Method Update Rule and EPA Method 624.1

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Analytical Trap Comparison for USEPA Method 8260C

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

CALA Directory of Laboratories

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

STANDARD OPERATING PROCEDURES

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Target Compound Results Summary

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Thermal Desorption Technical Support

In-Tube Extraction Sample Preparation for Gas Chromatography

Associates. Project No.: October 25, 2017

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Volatile Organic Compounds in Water PBM

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Copies: Dave Favero, RACER File

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Study of Residual Solvents in Various Matrices by Static Headspace

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Headspace Technology for GC and GC/MS: Features, benefits & applications

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

Table 1 Soil and Groundwater Sampling Summary Table

Readiness thru Health

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

USP<467> residual solvents

CIA Advantage-xr. Cryogen-free automated canister and whole-air sampling system

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

January 19, Dear Mr. Nightingale:

317 Elm Street Milford, NH (603) Fax (603)

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Transcription:

Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in today s environmental laboratory. The first and most important is that the analysis must be performed in compliance with EPA methodologies. Next, there is the continued demand to improve productivity. The key to productivity is to analyze as many samples as possible which meet the required quality assurance criteria in the 12 hours after a successful BFB tune check. The need to re-run samples due to moisture interference and carry-over certainly limits productivity. However, the most limiting factor of all is the length of the purge and trap cycle time. This paper will present the optimum purge and trap system conditions used to overcome these productivity limitations. Analytical results including calibration, accuracy and precision data will be presented for both water and soil matrices. DISCUSSION 1. The total timed needed for the auto-sampler to process, transfer, purge and dry purge the sample (total time from purge-ready to desorb-ready). 2. The total desorb, bake and cool-down time of the concentrator (total time from desorb pre-heat to purgeready). 3. The total GC cycle time (total time from GC start to GCready). In addition to the unique dual mode capabilities, the Centurion WS dramatically reduces the sample process and rinse time due to the absence of a syringe mechanism. The Centurion WS uses only pressure and gas flow to transfer samples and rinse water by using a pre-measured sample loop. Table 1 illustrates the GC cycle time of 15 minutes as the limiting factor in sample throughput in the dual concentrator configuration. PURGE AND TRAP CYCLE TIME As previously mentioned the most limiting factor to improving sample throughput is the length of the purge and trap cycle time. The typical purge and desorb parameters recommended or required by EPA methodologies are 12 to 15 minutes. In addition, a bake-out cycle is commonly used to clean the system before the next sample is analyzed, thus making the complete purge-and-trap cycle time nearly 25 minutes. Injections can only be made in 25 minute intervals which equal 28 injections per 12 hours. It is now common for GC cycle times to be as fast as 15 minutes when performing VOC analysis. How can the analyst take advantage of fast GC conditions to boost production without changing the required conditions governed by the EPA and without sacrificing data quality? The answer is to employ the use of a second purge and trap concentrator to the same GC/MSD system and continue to use the traditional conditions required by EPA methods. The EST Centurion WS Autosampler incorporates a mode referred to as the Dual Concentrator Option which easily connects two concentrators to the same GC and the same autosampler. This configuration allows the concentrator to alternate injections to the GC. As the first concentrator begins desorbing a sample, the Centurion WS sends the next sample to the second concentrator to begin purging. The rate-determining step for maximum sample throughput is now one of three factors. Table 1. Rate-Determining Steps PURGE AND DESORB, BAKE PROCESS DRY PURGE AND COOL GC TIME TIMES DOWN TIMES CYCLE <0.5 min 11 min + 1 min + 8 min + TIME 1 min 2 min 12.5 min 11 min 15 min MOISTURE INTERFERENCE As part of the purging process, water vapor is carried along with the purge gas. It is well known that moisture can cause interferences in the GC/MS chromatogram causing elevated detection levels and poor vacuum readings which lead to inconsistent analytical results. This interference can cause sample re-runs and greatly reduce the lab s productivity. The management of this moisture prior to GC introduction is crucial to maximizing sample throughput. 503 Commercial Drive, Fairfield, Ohio 45014 (513) 642.0100 www.estanalytical.com

The EST Encon Evolution Purge and Trap Concentrator utilizes a unique Moisture Reduction Trap (MoRT) to decrease the amount of moisture introduced to the GC. Unlike other concentrators the Evolution positions the moisture management device before the analytical trap to remove the water from the purge gas as illustrated in Figure 1. Through the use of an 8-port valve, the Evolution s MoRT is excluded from the desorb pathway during the transfer of analytes to the GC as illustrated in Figure 2. By positioning the water management device before the trap, dead volume and cold spots are eliminated from the desorb pathway to deliver superb peak symmetry and sensitivity over the entire chromatogram as shown in Figure 3. To Vent Transfer Line to GC Column GC Column Inject Multiport Valve Figure 3. Total Ion Chromatogram of the 10 ppb VOC standard Purge Gas CARRY-OVER REDUCTION Sparging Vessel MoRT Figure 1. Purge Pathway Purge Gas Sparging Vessel Multiport Valve MoRT To Vent Analytical Trap Transfer Line to GC Column Analytical Trap GC Column Inject As detection limits are pushed lower, carryover from check standards and contaminated samples has caused more frequent re-runs of samples. In the past, efforts have been made to reduce carryover with inert sample pathways and programmable flow rates during the bake cycle. Little has been done to deal with the primary cause of the carryover the sparge vessel itself. The Encon Evolution employs a patented mode which heats the sparge vessel during the rinsing process as shown in Figure 4. This takes place during the normal Evolution bake cycle therefore no additional time is added to the overall cycle time. The Figure 4. Sparge Tube Carry-Over Reduction System result is the lowest level of carry-over of any concentrator available today. The bake-out of the glassware delivers superior analytical results and reduces, if not eliminates completely, the number of blanks needed after high standards and contaminated samples. Figure 2. Desorb Pathway 2

EXPERIMENTAL RESULTS A single EST Centurion WS autosampler and two EST Encon Evolution purge and trap concentrators were interfaced to a single GC/MSD to demonstrate the ability of the Dual Mode option to provide quality data with maximum sample throughput. The system conditions used to obtain the results are listed in Tables 2 and 3. The linear range of the system was first established by analyzing a nine-point calibration curve (1ppb-200ppb) for both water and soil samples in accordance to the procedure in EPA method 8260B. The concentration of the internal standards was constant at 50ppb in all analyses. The percent relative standard deviation (%RSD) of the initial calibration (ICAL), the mean percent recovery of ten 1ppb measurements and the percent relative standard deviation of ten 1ppb measurements for each concentrator and matrix are shown in Table 4. Table 2. Purge and Trap Conditions Table 3. GC/MSD Conditions PURGE & TRAP CONCENTRATOR Trap Type EST ENCON EVOLUTION EV1 Proprietary Valve Oven 130 C Transfer Line 130 C Moisture Reduction Trap 39 C (MoRT) Temp Purge Time Purge Flow Purge Gas Dry Purge Dry Purge Flow Desorption Pressure Desorb Bake Temperature Bake Flow 11 minutes 40 ml/min Helium Ambient 1 minute 40 ml/min 12 psi 260 C for 2 minutes 265 C for 8 minutes 120 ml/min Sparge Heater Purge Ambient Carry-Over Reduction System Bake 110 C GC/MS AGILENT 7890/5975* Inlet Inlet Head-Pressure EPC S/SS 17.3 psi Inlet Temperature 200 C Mode Split Split Ratio 40:1 Total Flow Column Oven Temp. Program Column Flow Rate Mode *Agilent Technologies (Wilmington, DE) **Restek Corporation (Bellefonte, PA) 38.8 ml/min RTX-624** 20 m X 0.18mm I.D. 1.0 µm Film Thickness 45 C for 1 minute 45 C - 220 C at 18 C/min 220 C for 0.5 minute 0.8 ml/min Constant Flow PURGE & TRAP AUTOSAMPLER Water Sample Size Soil Sample Size Internal Standard Volume Minimizer Carry-Over Reduction Time EST CENTURION W/S 5 ml 5 grams/10 ml H20 5 ul 2 minutes 3

Table 4. Quality Assurance Data TARGET COMPOUND ICAL %RSD Accuracy at 1ppb Precision at 1ppb ICAL %RSD ICAL %RSD Accuracy at 1ppb Precision at 1ppb ICAL %RSD 1ppb-200ppb Water Matrix Water Matrix 1ppb-200ppb 1ppb-200ppb Water Matrix Water Matrix 1ppb-200ppb Water Matrix (%Recovery) n=10 (%RSD) n=10 Soil Matrix Water Matrix (%Recovery) n=10 (%RSD) n = 10 Soil Matrix CONCENTRATOR #1 CONCENTRATOR #2 Dichlorodifluoromethane 11.8 81.0 11.7 11.8 13.1 80.6 7.1 10.9 Chloromethane 12.3 105.1 8.6 13.1 14.1 99.8 4.4 10.6 Vinyl Chloride 6.5 95.6 7.9 11.3 9.3 90.4 9.6 5.2 Bromomethane 11.6 118.4 3.9 13.4 5.8 106.4 4.8 12.9 Chloroethane 10.7 96.6 4.0 11.0 7.2 96.4 6.1 10.2 Trichlorofluoromethane 3.7 89.8 5.4 11.2 4.5 93.9 4.4 6.5 1,1-Dichloroethene 4.9 98.0 5.6 9.0 6.3 99.1 6.7 8.8 Acetone* 0.999 92.3 5.9 0.997 0.997 114.8 7.3 0.998 Carbon disulfide 4.4 103.8 5.0 11.3 5.3 109.4 5.1 10.2 Methylene Chloride 4.0 103.0 5.0 10.5 3.7 109.5 7.8 11.5 MTBE 3.9 95.4 5.4 9.3 4.4 98.5 4.2 6.3 trans-1,2-dichloroethene 3.5 98.9 7.0 10.9 7.5 109.9 7.0 11.0 1,1-Dichloroethane 4.7 98.0 4.6 8.4 3.9 101.1 5.9 5.5 cis-1,2-dichloroethene 3.3 100.6 3.8 8.1 5.5 103.9 8.3 5.6 2-Butanone 6.5 105.1 9.0 14.8 5.5 104.3 6.6 13.5 2,2-Dichloropropane 3.6 81.8 5.1 8.5 8.4 86.8 6.2 9.7 Bromochloromethane 5.4 105.4 6.7 9.5 4.9 107.4 6.0 7.7 Chloroform 6.3 100.0 6.2 7.8 4.1 101.0 3.6 6.3 1,1,1-Trichloroethane 3.1 102.8 8.7 8.6 4.7 103.6 7.1 10.5 2-Chloroethylvinylether 7.5 96.8 8.0 11.1 9.3 99.8 4.9 7.8 Carbon Tetrachloride 2.7 89.8 7.0 10.0 5.0 96.8 4.7 4.7 1,1-Dichloropropene 5.0 91.1 5.6 11.3 4.2 100.0 7.0 8.0 Benzene 2.6 99.5 4.2 9.3 3.6 103.6 1.9 5.8 1,2-Dichloroethane 4.4 99.0 4.1 10.3 4.6 104.3 3.0 9.3 Trichloroethene 4.4 99.0 7.3 8.3 6.2 104.5 4.4 5.2 1,2-Dichloropropane 5.8 101.8 6.9 7.5 4.6 103.1 4.9 6.5 Dibromomethane 3.7 93.0 6.4 7.7 9.7 102.9 5.3 10.2 Bromodichloromethane 6.0 98.6 6.4 7.7 3.4 98.5 4.0 7.3 cis-1,3-dichloropropene 4.5 89.9 6.8 11.0 4.5 97.0 7.5 7.6 Toluene 4.5 96.5 4.5 9.9 4.8 100.0 7.3 10.1 trans-1,3-dichloropropene 6.0 93.0 8.8 7.5 4.5 95.3 7.5 6.2 1,1,2-Trichloroethane 3.2 101.3 7.9 8.7 4.5 104.4 5.8 10.8 Tetrachloroethene 1.9 98.0 5.5 8.6 5.9 94.4 4.8 8.3 1,3-Dichloropropane 4.1 99.5 4.5 12.7 4.2 101.6 3.2 8.0 Dibromochloromethane 3.7 91.5 5.0 6.7 6.3 92.3 5.7 8.5 2-Hexanone 7.6 85.9 7.0 10.1 7.7 91.3 5.7 9.8 1,2-Dibromoethane 5.2 100.3 6.1 12.2 4.8 97.7 7.7 6.8 Chlorobenzene 3.1 99.8 4.1 7.6 5.2 104.9 6.5 5.0 1,1,1,2-Tetrachloroethane 6.4 91.4 5.6 11.1 4.7 96.1 6.3 6.4 Ethylbenzene 4.4 96.3 5.6 7.1 3.6 96.3 4.8 5.3 Xylene (M&P) 3.9 94.9 5.4 6.2 3.5 94.3 3.9 4.6 Styrene 6.3 92.3 4.0 5.6 5.0 95.1 6.9 5.1 Xylene (o) 4.9 95.1 4.1 6.5 4.8 93.9 3.8 6.3 Bromoform 11.5 90.5 5.4 10.9 9.1 89.1 6.9 7.8 Isopropylbenzene 5.4 91.9 5.2 8.1 4.2 91.5 4.7 7.3 Bromobenzene 3.5 101.6 5.9 5.1 5.7 102.9 5.4 6.6 1,2,3-Trichloropropane 3.6 92.5 7.4 11.5 4.6 97.8 8.0 6.9 1,1,2,2-Tetrachloroethane 3.2 101.0 6.7 11.3 4.3 98.1 5.9 13.7 n-propylbenzene 4.9 94.5 5.0 5.8 3.9 94.4 4.7 4.3 2-Chlorotoluene 6.1 102.1 6.3 7.1 4.4 97.6 5.6 4.4 4-Chlorotoluene 3.7 98.3 7.0 5.8 5.0 102.0 4.3 3.5 1,3,5-Trimethylbenzene 5.4 89.5 5.1 5.0 4.7 90.4 4.3 6.6 tert-butylbenzene 2.9 91.8 7.1 6.9 3.4 92.4 5.5 5.2 sec-butylbenzene 5.6 100.3 7.0 6.1 3.2 96.4 9.5 6.8 1,2,4-Trimethylbenzene 5.5 92.1 3.3 5.4 2.6 94.6 5.8 6.1 1,3-Dichlorobenzene 3.0 100.6 6.4 5.6 3.9 103.3 5.4 6.7 1,4-Dichlorobenzene 2.6 101.8 3.7 5.4 8.4 105.5 4.7 5.7 Isopropyltoluene 7.7 90.8 5.6 6.4 3.4 88.5 5.6 7.2 1,2,-Dichlorobenzene 2.6 103.1 4.7 5.7 4.8 102.1 5.0 5.5 n-butylbenzene 6.3 89.6 6.4 4.9 5.2 95.1 3.9 7.2 1,2-Dibromo-3-chloropropane 10.1 91.6 7.5 8.3 3.8 93.9 9.8 11.9 1,2,4-Trichlorobenzene 5.8 100.4 5.1 5.4 3.8 96.5 4.9 9.4 Naphthalene 11.5 87.0 4.7 5.8 4.3 93.3 6.6 7.1 Hexachlorobutadiene 7.8 87.9 9.6 7.5 5.8 95.8 4.7 6.2 1,2,3-Trichlorobenzene 7.1 95.3 4.8 7.1 4.7 101.4 7.2 9.2 *Compound was linear regressed 4

CONCLUSION From the data presented the instrument configuration and operating conditions described above produce outstanding performance for EPA Method 8260B. In addition to the improved quality of the data, the configuration also maximizes sample throughput by an additional 40 injections per day over conventional configurations by enabling analysts to be able to inject sample every 15 minutes! This translates to an improvement of an additional 34 revenue generating samples each day per instrument. The Centurion WS and Encon Evolution also offer a number of other unique user features and benefits. These include a programmable internal standard delivery mechanism and a sampling routine which only transports soil vials. Centurion WS Autosampler Encon Evolution 503 Commercial Drive, Fairfield, Ohio 45014 (513) 642.0100 www.estanalytical.com 5