LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Similar documents
Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Application Note. Abstract. Authors. Environmental

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Agilent ESI and APCI sources: for polar to non-polar compounds

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Determination of Hormones in Serum by LC/MS/MS Using Agilent Bond Elut Plexa SPE

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Measuring Drug-to-Antibody Ratio (DAR) for Antibody-Drug Conjugates (ADCs) with UHPLC/Q-TOF

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Application Note. Edgar Naegele. Abstract

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

A New Polymer-based SPE Sorbent to Reduce Matrix Effects in Bioanalytical LC-MS/MS

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Supporting Information

770-9P. Sensitivity and Selectivity - A Case Study of LC/MS Enantioselective Resolution of Bupivacaine Using Vancomycin as a Chiral Stationary Phase

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Quantitation of Cystine and Identification of Related Metabolites in White Blood Cells Using a High Resolution Accurate Mass LC/MS Approach

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

LC/Q-TOF Workflows for Comprehensive Micropollutant Analysis

Separation of Large and Small Peptides by Supercritical Fluid Chromatography and Detection by Mass Spectrometry

Computer-assisted analysis of complex natural product extracts

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Agilent 6400 Series Triple Quadrupole LC/MS/MS Users Session

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Application Note. Abstract. Authors. Food Safety and Environmental

Accurate Mass Analysis of Hydraulic Fracturing Waters

Routine Multiresidue Pesticide Analysis using the Agilent 6470 Triple Quadrupole Mass Spectrometer

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

Transcription:

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode Application Note Authors Siji Joseph Agilent Technologies India Pvt. Ltd. Bangalore India Anabel Fandino Agilent Technologies, Inc. Santa Clara, CA USA Abstract A sensitive, precise, and accurate quantitative liquid chromatography/tandem mass spectrometry (LC/MS/MS) method for the measurement of estradiol acetate in human plasma was developed. The sample preparation method employed solid phase extraction on Agilent SampliQ cartridges. The processed samples were chromatographed on an Agilent ZORBAX Eclipse Plus C18 analytical column and analyzed by triple-quadrupole tandem mass spectrometry in the multiple reaction monitoring (MRM) mode using negative electrospray ionization. The principal advantage of the LC/MS/MS method described in this application note is the simultaneous achievement of high absolute recovery (~ 80%), high sensitivity (LLOQ of 100 fg injected on-column or 28.6 pg/ml plasma), high precision ( 5.9% CV) and accuracy (81.5-118.1%), as well as excellent linearity over the concentration range 0.1 to 10000 pg/μl (R 2 0.998) with a short run time of only 1.2 min.

Introduction β-estradiol 17-acetate is a derivative of the prime sex hormone estradiol. The molecular structure is shown in Figure 1. Highly sensitive analytical methods are necessary to determine these types of analytes in human body fluids. Because of its specificity and precision, LC/MS/MS in negative ion mode offers a better and faster high throughput technique to monitor these types of analytes in body fluids in comparison to radioimmunoassay and gas chromatography - mass spectrometry (GC/MS) methods. In addition to HO the usage of hazardous radioactive isotopes and the necessary extensive manual labor, immunoassays can be highly variable and inaccurate owing to cross reactivity issues. Purifying the analyte from a plasma sample is a critical segment in the whole sample preparation process because of sample loss issues. To overcome this, Agilent provides a wide range of SampliQ solid phase extraction (SPE) cartridges, which give excellent recovery for a wide range of compounds. The use of an Agilent ZORBAX column provided excellent separation of the analyte from biological matrix components. H 3 C O O CH 3 Figure 1: β-estradiol 17-acetate structure (C 20 H 26 O 3, molecular weight: 314.42). Experimental Chemicals ß-Estradiol 17-acetate, human plasma, formic acid and solvents for LC/MS/ MS analysis were purchased from Sigma-Aldrich. HPLC grade water was freshly taken from a Milli-Q water purification system. Super gradient grade acetonitrile (ACN) and methanol were purchased from Lab-Scan (Bangkok, Thailand). Preparation of the stock solution, calibration and quality control standards A 0.2 mg/ml stock solution of ß-estradiol 17-acetate was prepared in acetonitrile (ACN)/water, 50/50 v/v (diluent). To prepare the various calibration levels in aqueous solutions, the stock solution was diluted using the diluent at seven concentration levels; the concentrations used for the linearity study were 0.1 pg/μl, 0.5 pg/ μl, 1 pg/μl, 10 pg/μl, 100 pg/μl, 1000 pg/μl, and 10000 pg/μl. In addition to the above concentration levels, a 0.2 pg/μl standard solution was also used for the analyte recovery study from human plasma to make a total of eight concentration levels. Calibration levels in plasma were prepared by spiking 100 µl of each standard solution to 250 µl of human plasma. A 500 µl aliquot of Milli-Q water was added followed by 150 µl of 0.5% formic acid. The samples were vortexed after each addition. Each sample was loaded onto SPE cartridges as described below. Three quality control (QC) samples were prepared at the following concentration levels to check for precision and accuracy: 20 pg/μl (low QC), 200 pg/μl (mid QC), and 2000 pg/μl (high QC). 2

Human plasma extraction procedure Agilent SampliQ C18 end capped (100 mg) cartridges (part number: 5982-1311) were used for the extraction. The SPE cartridges were pre-conditioned with 2 ml methanol and conditioned with 5 ml acetonitrile. Equilibration was achieved using 2 ml of Milli-Q water. After loading the sample, SPE cartridges were washed with 2 ml of water (first wash) followed by 1 ml of ACN/water 20:80 v/v (second wash). After a three minute vacuum drying, Pre-conditioning -2 ml Methanol Condition -5 ml Acetonitrile Equilibration -2 ml Water Sample Load Wash I -2 ml Water Wash II -1 ml Mobile phase A:B; 8:2 Elution -1.5 ml Acetonitrile the samples were eluted out from SPE cartridges using 1.5 ml acetonitrile (3x 500 µl aliquots). This organic fraction was collected separately and evaporated to dryness in a RapidVap Vacuum Evaporation System at 30 C. The sample was reconstituted with 100 μl of mobile phase (mobile phase A/B; 20:80), and the resulting sample was injected into the Agilent 6460 Triple Quadrupole LC/MS System coupled to an Agilent 1200 Series Rapid Resolution LC (RRLC) system. The SPE protocol is compiled in Figure 2. Vacuum dry for 3 minutes Determination of extraction recovery Extraction recoveries of ß-estradiol 17-acetate from spiked plasma samples were determined by comparing the peak areas obtained by extraction of freshly prepared plasma extracts, at different concentration levels, with those from an extraction of an aqueous standard solution at equivalent concentration (n = 3). LC parameters The Agilent 1200 Series RRLC system comprised a well plate autosampler with sample thermostat, degasser, column compartment, and pump. A 0.12 mm inner diameter stainless steel capillary was used for all tubing connections. The tubing length was minimized to reduce peak dispersion. The analytical column used for this experiment was ZORBAX Eclipse Plus C18 RRHT 2.1 x 100 mm 1.8 micron 600 bar (part number: 959764-902); the temperature was 40 C. Solvent A was 0.05% ammonium hydroxide (NH 4 OH) in water and solvent B was 100% ACN. The pump delivered 80% B isocratic mobile phase at a flow rate of 0.5 ml/ min. The run time was 1.2 min with no post run time. The flush port was activated for 5 sec with ACN/water 80:20 v/v. The autosampler was set for a sample injection volume of 1 μl. Evaporation to dryness Reconstitution Figure 2: Solid Phase Extraction protocol used to extract ß-estradiol 17-acetate from human plasma. 3

Triple-quadrupole parameters The Agilent MassHunter Workstation software suite incorporated LC/MS data acquisition, qualitative analysis, and quantitative data processing. A 100 pg/μl β-estradiol 17-acetate standard solution was used for the optimization of the MS parameters. The molecular ion mass of β-estradiol 17-acetate in negative mode was found to be 313 Da and in selective ion mode (SIM), an intense peak was observed with a fragmentor voltage value of 135 V. A collision energy (CE) study gave intense products ions at 253 Da (used as the quantifier ion) and 58.9 Da (used as the qualifier ion). The drying gas Parameter Set value in 6460 Triple Quad (negative mode) Fragmentor voltage [V] 135 Collision energy [V] 18 Sheath gas temperature [ C] 350 Sheath gas flow [L/min] 11 Nebulizer pressure [psi] 25 Cap voltage [V] 3500 Nozzle voltage [V] 2000 Drying gas temperature [ C] 300 Drying gas flow [L/min] 7 MS1/MS2 resolution Wide/Unit Dwell time per transition [ms] 100 Time filtering [min] 0.02 Delta EMV [V] 500 Table 1: Optimized triple-quadrupole parameters. flow was optimized at 7 L/min at a constant temperature of 300 C. Data acquisition was performed in the negative ionization mode. Capillary voltage was -3500 V. Agilent Jet Stream Thermal Gradient Focusing was operated with a sheath gas flow rate of 11 L/min at a constant temperature of 350 C. A dwell time of 100 ms was used for each of the multiple reaction monitoring (MRM) transitions. Prior to the experiment, the instrument was autotuned in negative mode using default settings. The first quadrupole was operated at wide resolution (1.2 amu at FWHM) and the second one was at unit resolution (0.7 amu at FWHM). EMV = 500 V. The signal-tonoise (S/N) ratio was calculated from the height RMS method found in the Method Editor of the MassHunter software. The parameters were as follows: signal definition: height noise definition: root-mean-square (RMS) multiplication factor: 1 noise region: 0.6-0.8 minutes. Table 1 gives the compiled triplequadrupole parameters. Data acquisition and processing Data acquisition was performed with Agilent MassHunter Workstation Data Acquisition software. Data processing was performed with Agilent MassHunter Quantitative Analysis and Agilent MassHunter Qualitative Analysis software. 4

Results and Discussion Quantitation of ß-estradiol 17-acetate in aqueous solutions Figure 3 represents the precursor and product ions of β-estradiol 17-acetate. For quantitation of β-estradiol 17-acetate, the MRM transition 313 253 was monitored. Figure 4 shows an overlay of a 100 fg ß-estradiol 17-acetate chromatogram in aqueous solution with a blank. 100 fg/μl standard solution yielded a S/N ratio of 53 (N = 4, Signal = Height, N oise = 1 x RMS). x10 3 3 2.8 2.6 2.4 2.2 2 1.8 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0 58.6 Qualifier 144.9 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 360 380 400 Counts vs. Mass-to-Charge (m/z) Figure 3: Fragmentation pattern for β-estradiol 17-acetate showing precursor and product ions at CE = 30 V. 253.0 Quantifier 312.9 Precursor ion x10 1 8 7.8 7.6 7.4 7.2 7 6.8 6.6 6.4 6.2 6 5.8 5.6 5.4 5.2 5 4.8 4.6 100 fg ß-estradiol 17-acetate blank 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 1 1.05 1.1 Counts vs. Acquisition Time (min) Figure 4: Overlay of a 100 fg β-estradiol 17-acetate chromatogram in aqueous solution (blue trace) with a blank (red trace). 5

Linearity tests were performed in the concentration range of 100 fg/μl to 10 ng/μl (7 levels, 4 replicates per level). A calibration curve was constructed by plotting the peak area of β-estradiol 17-acetate versus concentration (0.1 pg/μl, 0.5 pg/μl, 1 pg/μl, 10 pg/μl, 100 pg/μl, 1000 pg/μl, 10000 pg/μl, 5 orders of magnitude). The calibration plot was fitted by 1/x weighting. The linearity of the relationship between peak area and β-estradiol 17-acetate concentration is demonstrated by a correlation coefficient of R 2 > 0.9999. The calibration curve for aqueous linearity is shown in Figure 5 with an excellent correlation coefficient of R 2 > 0.9999. Quantifier - 7 Levels, 7 Levels Used, 28 Points, 28 Points Used, 0 QCs x10 6 y = 365.8076 * x - 12.8495 R^2 = 0.99986343 1 Responses 0.1 0.01 0.001 0.0001 1 10 100 1000 10000 Concentration (pg) Figure 5: Linearity curve of β-estradiol 17-acetate from 0.1 pg to 10000 pg injected on-column (linear fit, weight 1/x). Quantitation of ß-estradiol 17-acetate in human plasma To assess the specificity of the analytical method, blank human plasma samples were extracted and analyzed for potential interferences. The obvious response at the retention time of β-estradiol 17-acetate was compared to the lower limit of quantification. Figure 6 shows an overlay of a 100 fg β-estradiol 17-acetate chromatogram recovered from human plasma with a blank. 100 fg/μl in plasma yielded a S/N ratio > 38 (N = 4, Signal = Height, Noise = 1 x RMS). x10 2 3 2.8 2.6 2.4 2.2 2 1.8 1.6 1.4 1.2 1 0.8 0.6 0.4 100 fg ß-estradiol 17-acetate blank 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 1 1.05 1.1 Counts vs. Acquisition Time (min) Figure 6: Overlay of a 100 fg β-estradiol 17-acetate chromatogram recovered from human plasma (blue trace) overlaid with a blank (red trace). 6

Linearity, precision, accuracy, and recovery Linearity was studied over the concentration range of 100 fg/μl to 10000 pg/μl (eight levels, 4 replicates per level, 5 orders of magnitude). The correlation coefficient (R 2 ) was higher than 0.9985 (Figure 7). The calculation was performed using a linear weighted regression (1/x). The results obtained for linearity, precision, and accuracy are shown in Table 2. The method intra-day precision and accuracy values for low QC, mid QC, and high QC levels (20 pg/μl, 200 pg/μl, 2000 pg/μl, n = 4) are 1.25%, 0.42%, 0.92% and 85.0%, 103.2%, 100.9%, respectively. The injector precision and accuracy can be demonstrated using the RSD value of retention time at various linearity levels. Extraction recoveries of ß-estradiol 17-acetate from spiked plasma samples are found to be ~ 80% at 1 pg, 10 pg, and 100 pg concentration levels. Quantifier - 8 Levels, 8 Levels Used, 32 Points, 32 Points Used, 12 QCs x10 7 y = 1118.6141 * x + 87.0591 R^2 = 0.99848795 Responses 0.1 0.01 0.001 0.0001 1 10 100 1000 10000 Concentration (pg) Figure 7: Linearity curve of β-estradiol 17-acetate recovered from human plasma from 0.1 pg to 10000 pg injected on-column (linear fit, weight 1/x). The blue triangles ( ) in the calibration line represent the QC samples. Linearity level amount on-column Retention time RSD [%) Average response Response RSD [%) Average accuracy Accuracy RSD [%) 0.1 pg 0.18 219 1.52 118.1 2.51 0.2 pg 0.34 325 5.24 106.5 7.14 0.5 pg 0.21 658 5.87 102.2 6.76 1 pg 0.29 1032 2.06 84.5 2.25 10 pg 0.00 9200 2.23 81.5 2.25 20 pg (L-QC) 0.00 19112 1.25 85.0 1.26 100 pg 0.00 107822 0.24 96.3 0.24 200 pg (M-QC) 0.18 230983 0.42 103.2 0.42 1000 pg 0.00 1254870 0.29 112.2 0.29 2000 pg (H-QC) 0.00 2257375 0.92 100.9 0.92 10000 pg 0.00 11056383 0.36 98.8 0.36 Table 2: RSD values for retention time, response, and accuracy from plasma linearity study (n = 4). 7

Conclusions Highly sensitive quantitation of β-estradiol 17-acetate from aqueous solutions and human plasma using an Agilent 1200 Series RRLC coupled to an Agilent 6460 Triple Quadrupole LC/MS in ESI negative ion mode is demonstrated. Extraction of spiked β-estradiol 17-acetate from human plasma down to a concentration of 100 fg injected on-column is demonstrated using Agilent SampliQ SPE cartridges. The linearity of the quantitation of β-estradiol 17-acetate in aqueous solution was demonstrated over five orders of magnitude (from 100 fg to 10000 pg on-column). The observed R 2 value for the calibration curve is > 0.9999. The linearity of the quantitation of β-estradiol 17-acetate in human plasma was demonstrated over five orders of magnitude (from 100 fg to 10000 pg on-column). The observed R 2 value for the calibration curve is > 0.998. The observed average S/N value for 100 fg β-estradiol 17-acetate in plasma (injected on-column) is > 33 (n = 4, Signal = Height, Noise = 1 x RMS). RSD of response for all the levels studied was < 5.9%. Accuracy RSD at LLOQ level in plasma is < 2.5% and for all other levels it is < 7.1%. www.agilent.com/chem/qqq This item is intended for Research Use Only. Not for use in diagnostic procedures. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Agilent Technologies, Inc. 2010 Published in the U.S.A. June 28, 2010 5990-5877EN