Neutron Metrology Activities at CIAE (2009~2010)

Similar documents
Recent Activities on Neutron Calibration Fields at FRS of JAERI

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Geant4 simulation of SOI microdosimetry for radiation protection in space and aviation environments

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

PROGRESS OF NUCLEAR DATA MEASUREMENT IN CHINA

Introduction to neutron rem meters

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Bonner Sphere Spectrometer. Cruzate, J.A.; Carelli, J.L. and Gregori, B.N.

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007

Flux and neutron spectrum measurements in fast neutron irradiation experiments

Recent activities in neutron standardization at NMIJ/AIST

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI

Introduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

Validation of the UFS Bonner Sphere Spectrometer and Monte Carlo Methods at the CERN-EU high energy Reference Field (CERF)

A new neutron monitor for pulsed fields at high-energy accelerators

STUDY ON THE ENERGY RESPONSE OF PLASTIC SCINTILLATION DETECTOR TO MEV NEUTRONS ABSTRACT

Influence of Sensitive Volume Dimensions on the Distribution of Energy Transferred by Charged Particles

4. The mass of a proton is approximately equal to the mass of A an alpha particle C a positron. B a beta particle D a neutron

Neutron Users Club Meeting

Calibration of the GNU and HSREM neutron survey instruments

STUDY ON IONIZATION EFFECTS PRODUCED BY NEUTRON INTERACTION PRODUCTS IN BNCT FIELD *

Portable, Low-cost Proportional Counters for Space, Atmospheric and Ground based Applications

Characterization of an 241 AmBe neutron irradiation facility by different spectrometric techniques

Measurements with the new PHE Neutron Survey Instrument

Neutron Source Capabilities at Ohio University

Experimental Activities in China

Characterization of 3D thermal neutron semiconductor detectors

Experimental Activities in China

Division 6 Ionizing Radiation

EEE4106Z Radiation Interactions & Detection

Recent developments in neutron metrology at the Institute for Radiological Protection and Nuclear Safety (IRSN)

Nuclear cross-section measurements at the Manuel Lujan Jr. Neutron Scattering Center. Michal Mocko

Neutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace

Calculation of Bubble Detector Response Using Data from the Matroshka-R Study

Chapter V: Cavity theories

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor

CHARGED PARTICLE INTERACTIONS

Chapter V: Interactions of neutrons with matter

CRaTER Science Requirements

A comprehensive study of a wide energy range proportional counter neutron spectrometer

MEASUREMENT AND DETECTION OF RADIATION

CRaTER Pre-Environmental Review (I-PER) Science Requirements Update

Diamond detectors in Bonner Spheres A Novel Approach for Real-time Neutron Spectroscopy

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

Response characteristics of neutron survey instruments. Rick Tanner and David Bartlett, NRPB Hamid Tagziria and David Thomas, NPL

ERINDA PAC 1/4 Testing the UAB Extended Range Bonner Sphere Spectrometer for high energy neutrons UU-TSL, Uppsala, Sweden UAB Barcelona

Atomic Structure. The nuclear atom Electron Configuration. Ms. Thompson - SL Chemistry Wooster High School. Wednesday, July 15, 15

Techniques to Measure Absolute Neutron Spectrum and Intensity for Accelerator Based Neutron Source for BNCT )

CRaTER Pre-Ship Review (PSR) Instrument Calibration Science Requirements Compliance

Queen s University PHYS 352

Radiation Protection Dosimetry Advance Access published May 4, Radiation Protection Dosimetry (2015), pp. 1 5

PARTICLES REVELATION THROUGH SCINTILLATION COUNTER

Neutron Transport Calculations Using Monte-Carlo Methods. Sean Lourette Fairport High School Advisor: Christian Stoeckl

References and Figures from: - Basdevant, Fundamentals in Nuclear Physics

Gas-filled Detectors

DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

Lali Tchelidze and James Stovall

DDX measurement for charge particle production reaction by gridded ionization chamber

Chapter 11: Neutrons detectors

Measurements of the Deuterium-Tritium Branching Ratio Using ICF Implosions

He-3 Neutron Detectors

Neutron Time-Of-Flight Spectrometer Based on HIRFL for Studies of Spallation Reactions Related to ADS Project

IAC-08-A MONTE CARLO SIMULATIONS OF ENERGY LOSSES BY SPACE PROTONS IN THE CRATER DETECTOR

Studies on 3D Fusion Reactions in TiDx under Ion Beam Implantation

Performance Test of the Electronic Personal Neutron Dosemeter in Neutron Fields Simulating Workplaces of MOX Fuel Fabrication Facilities

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Accelerator Mass Spectroscopy

Neutronic Design on a Small Accelerator based 7 Li (p, n) Neutron Source for Neutron Scattering Experiments

Unit 2: Atomic Structure Additional Practice

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

Application and Validation of Event Generator in the PHITS Code for the Low-Energy Neutron-Induced Reactions

Spectral Correction Factors for Conventional Neutron Dose Meters Used in High-Energy Neutron Environments

Calculations of Neutron Yield and Gamma Rays Intensity by GEANT4

SL Chemistry. Monday September 21st Monday, September 21, 15

PECULIARITIES OF FORMING THE RADIATION SITUATION AT AN AREA OF NSC KIPT ACCELERATORS LOCATION

ICRP Symposium on the International System of Radiological Protection

Teacher: Mr. gerraputa. Name: Base your answer to the question on the information below. Given the electron dot diagram:

8. Application to Fast Neutron Therapy

Introduction to neutron sources

Exercise 1 Atomic line spectra 1/9

MONTE CARLO SIMULATIONS OF SOLID WALLED PROPORTIONAL COUNTERS WITH DIFFERENT SITE SIZE FOR HZE RADIATION

Measurements of Radiation Doses Induced by High Intensity Laser between and W/cm 2 onto Solid Targets at LCLS MEC Instrument

Bubble Detector Characterization for Space Radiation

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

Scintillation efficiency measurement of Na recoils in NaI(Tl) below the DAMA/LIBRA energy threshold

Interactions of Particulate Radiation with Matter. Purpose. Importance of particulate interactions

Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector

INTRODUCTION Strained Silicon Monochromator Magnesium Housing Windows for Monochromator Shutter and Collimator Fission Detector HOPG Monochromator

Radiation Quantities and Units

Gamma-ray emission by proton beam interaction with injected Boron atoms for medical imaging. Giada Petringa - Laboratori Nazionali del Sud -

A Comparison between Channel Selections in Heavy Ion Reactions

Investigations on warm dense plasma with PHELIX facility

Fission Fragment characterization with FALSTAFF at NFS

Shielding Considerations

Transcription:

Neutron Metrology Activities at CIAE (2009~2010) Ionizing Radiation Metrology Division China Institute of Atomic Energy (P.O.Box 275(20), Beijing 102413, China) 1. Neutron calibration fields So far the neutron calibration fields have been developed at the energies recommended in ISO 8529-1. Recently we are establishing the 8 and 24keV neutron calibration fields and the simulated workplace neutron fields for pressure water reactors. The details are described as follows. (1) 8 and 24keV neutron calibration fields The 8 and 24keV neutron fields can be produced by the 45 Sc(p,n) 45 Ti reaction at 5SDH-2 tandem accelerator. A new target pipe of which structure is shown in Fig.1 was designed in order to bombard the scandium target with a larger beam current than 20µA, so as to increase the neutron yields. The whole target pipe is made of oxygen free copper for the sake of good heat conductivity. The scandium target evaporated on a tantalum back can be fixed on the end of the target pipe by a cylindrical copper sleeve. Refrigerant is driven by a compressor to flow around the target pipe. The target pipe can be connected with the pipeline of the accelerator by flanges. At present we are to test the target using the 3MeV proton beam of 30µA. 1

Fig.1 Sketch map of the new target pipe for the 45 Sc(p,n) 45 Ti neutron source Due to the lower neutron energies, the hydrogenous proportional counter used for 144keV 1.2MeV neutron fluence measurement is not appropriate for 8 and 24keV energies. Hence a detector of Si semiconductor with a 6 LiF converter of which structure is shown in Fig.2 was designed for the purpose. The 6 LiF with atomic abundance 85.43% of 6 Li was evaporated on a 10µm thick aluminum foil. The thickness of the 6 LiF converter is approximately 800µ cm -2. The α particles produced the 6 Li(n,t) 4 He reaction can be stopped in the aluminum foil and the tritium can penetrate it although some energies may be lost. The residual energy of tritium entering a Si(Au) detector was calculated to be equal to 2.2MeV. The range of 2.2MeV tritium in silicon was also calculated to be approximately equal to 30µm, therefore, a passivated surface of Si(Au) semiconductor with 100µm thick depletion layer was chosen as the detector. In the near future the excitation curve of neutron yields as a function of incident proton energies is to be measured using a long counter, and the neutron spectra of a thick target ~500µg cm -2 and a thin target ~70µg cm -2 ) are to be measured using a lithium glass detector by time-of-flight method. 2

Fig.2 Sketch map of the 6 LiF-Si(Au) detector (2) Simulated workplace neutron fields The simulated workplace neutron fields for pressure water reactors are developing at 5SDH-2 tandem accelerator. A neutron calibration source facility has been developed as shown in Fig.3 according to a prototype as described in ISO 12789. 3

Fig.3 Sketch map of the neutron calibration source facility at CIAE The 5MeV monoenergetic neutron field produced by the 2 H(d,n) 3 He reaction is to be used as the primary source. The neutron fluence rate spectra at the calibration position had been simulated with Monte Carlo program as shown in Fig.4. Spectrum-1 represents the simulated neutron spectrum when light water substitutes for heavy water (see Fig.3) which was used to simulate spectrum-2. Fig.4 Simulated workplace neutron spectra for pressure water reactor 4

A ROSPEC spectrometer and a SSS spectrometer are to be introduced from the BTI Corporation of Canada to measure the neutron field spectra and the dose equivalent rates. The results will be compared to that measured by a Bonner sphere system. 2. Neutron measuring instruments (1) Bonner sphere system A Bonner sphere system has been developed in recent years. The system is composed of a set of polythene spheres (density 0.909g cm -3 as 3, 3.5, 4, 5, 6, 8, 10 and 12 inch in diameters with a 3 He proportional counter (SP9, Centronic Ltd.) as the central detector. The response functions of the system were calculated preliminarily with Monte Carlo program as shown in Fig.5. The angular responses of bare 3 He counter and 3inch sphere were also analyzed as shown in Fig.6. Here, R(0º) is the response when neutron incidence is from the nose of the 3 He counter, R(90º) is the response when neutron incidence is from the vertical direction to the stem of the 3 He counter, and R(180º) is the response when neutron incidence is from the end of the 3 He counter. For the bare 3 He counter the maximum anisotropy (R(180º)/R(90º)) is 1.22, and for 3inch sphere the maximum anisotropy (R(180º)/R(90º)) is 1.07. In addition, the spheres with different thick boron shells were designed to improve the energy resolution in the epithermal region so as to apply the spectrometer to the BNCT epithermal beam. The response functions of 2, 3, 4inch spheres with different thick boron shells are shown in Fig.7. The reason for the drop of the responses in the vicinity of 30keV should be investigated. 5

Fig.5 Response functions of the Bonner sphere system Fig.6 Ratio of the angular responses of bare 3 He counter and 3inch sphere Fig.7 Response functions of 2,3,4inch spheres with different thick boron shells 6

Subsequently, the atomic density of 3 He in the proportional counter will be determined in a thermal field. And then the responses of the detectors will be calibrated experimentally in the monoenergetic neutron fields. (2) Tissue equivalent proportional counter A TEPC of which structure is shown in Fig.8 was introduced from Farwest Corporation of USA. The detector with a spherical A150 plastic wall can simulate a tissue cell with the diameter of 2µm. Fig.8 Sketch map of the TEPC with a spherical A150 plastic wall The TEPC filled with the tissue equivalent gas comprising 63.99 CH 4 32.9 CO 2 and 3.11 N 2 was preliminarily tested using 252 Cf, 60 Co and 137 Cs sources. The measured linear energy spectra are shown in Fig.9. Furthermore, the linear energy spectra for 60 Co and 137 Cs sources were also simulated by Monte Carlo program as shown in Fig.10. Because only one amplifier was used in the measurement, the amplification gain could not adapt well to the full range of the linear energy spectrum for 252 Cf source. This is the reason why the low linear energy events are rare as shown in Fig.9. Consequently two amplifiers should be used in the 7

electronics of TEPC, and the reason why the measured linear energy spectra for 60 Co and 137 Cs sources are deviated from the calculated results also will be investigated in the next work. Fig.9 Measured linear energy spectra in the fields of 252 Cf, 60 Co and 137 Cs 3. Summary Fig.10 Simulated linear energy spectra in the fields of 60 Co and 137 Cs In recent years neutron metrology has been developed continuously at CIAE. The 5SDH-2 accelerator was pulsed in order to measure neutron the monoenergetic field spectra by time-of-flight method. That can improve the characteristic parameters of the monoenergetic neutron calibration fields. A Bonner sphere system has been developed to make it possible that the neutron spectra at workplaces can be measured for the 8

radiation protection purpose. A TEPC was introduced for developing the microdosimetry. The monoenergetic neutron calibration fields at 8 and 24keV and the simulated workplace neutron fields are also being developed. We believe that the neutron metrology at CIAE can make further progress in the near future. 9