Automated Determination of PPQ Levels of Thorium in High Purity Copper

Similar documents
Elemental Scientific. Spike Recovery. Elemental Scientific

Elemental Scientific. Stability. Elemental Scientific

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

hydrideicp Hydride Generation System

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Elemental Scientific. Spike Recovery. Elemental Scientific

Elemental Scientific

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Trace Analyses in Metal Matrices Using the ELAN DRC II

Automating Sample Purification MC-ICPMS. Elemental Scientific ICP ICPMS AA

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Contamination Monitoring of Semiconductor Processes by VPD HR-ICPMS

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Analysis of domestic sludge using the Agilent 4200 MP-AES

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Automation of the radiochemical procedures for the sequential separation of radionuclides

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

DISCLAIMER: This method:

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

ICP-OES Application Note Number 35

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Savillex Technical Note Performance of the DST-1000 Acid Purification System

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

Determination of urea in ultrapure water by IC-MS/MS

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Low level Os isotopic measurements using multiple ion counting

TECHNETIUM-99 IN SOIL

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

ION CHROMATOGRAPHY SYSTEM S 150

ICP-MS Analysis of Bulk & Process Chemicals for Semiconductor Processes. Ann O Connell Chemical Analysis Eng Intel Corporation

Analysis Repeatability of Trace and Major Elements in a Water Sample

JOURNAL OF. Flow Injection. Analysis FIA 研究懇談会会誌

Determination of trace anions in concentrated hydrofluoric acid

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

OPTIMIZING METALS SAMPLE PREP

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

SUPPLEMENTARY INFORMATION

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

enable measurement. This method separates these isotopes effectively.

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Atomic Spectroscopy AA/ICP/ICPMS:

Fergus Keenan Shona McSheehy Julian Wills

URANIUM IN SOIL. Analytical Procedure (2 GRAM SAMPLE) 1. SCOPE

Analytical Method Determination of airborne metal concentrations by ICP-MS Sampling using Solu-Sert filter capsules

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

TECHNETIUM-99 IN WATER

Contact Person(s) : Anna Berne APPLICATION

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Determination of Trace Elements of Ge and P in a Gold Bonding Wire by Inductively Coupled Plasma Atomic Emission Spectrometry

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Accurate Sizing and Precise Counting of 10 nm Gold Nanoparticles using the Enhanced Sensitivity of the NexION 2000 ICP-MS

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

ACCURATE QUANTIFICATION OF RADIOACTIVE MATERIALS BY X-RAY FLUORESCENCE: GALLIUM IN PLUTONIUM METAL

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Mechanistic Insights into Chloride Ion Detection from the Atmospheric-Pressure Afterglow of an Argon Inductively Coupled Plasma

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Dionex IonPac AS28-Fast-4µm column

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Procedures for the determination of stable elements in construction materials from the nuclear reactors at Risø National Laboratory

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Transcription:

PPQ Levels of Th in High Purity Cu TRUFAST ICPMS Automated Determination of PPQ Levels of Thorium in High Purity Copper Using the ESI TRUFAST System and ICPMS Detection By Nathan Saetveit, PhD and Dan Wiederin, PhD. Introduction Alpha decay from semiconductor device package materials is a root cause of soft errors in microprocessors and storage devices. Reducing the concentration of Th (t 1/2 = 1.4 x 10 10 years) and other alpha emitters in copper package material will yield a corresponding reduction in alpha decay events and soft errors. Improved analytical capability allowing the detection of single digit ppq concentrations of Th in Cu would facilitate process improvements and ensure product quality. This note describes an automated in-line system which uses a PFA column packed with TEVA Resin to retain Th from a solution of 1% Cu and 8% HNO 3 while the Cu SC-2 DX TRUFAST Preconcentration and Matrix Removal System matrix passes through to waste. The preconcentrated Th is eluted from the column using high purity HCl and detected by ICPMS. Thorium calibrations are linear over more than 4 orders of magnitude with washout factors (sample-to-sample) better than 1000x. Long-term stability (60 samples) was better than 2% RSD at 100 ppt Th. The Th detection limit (3s) is less than 10 ppq and could be improved further by reducing the procedural blank. Cu samples were prepared by dissolution in high purity HNO 3 and brought to a final concentration of 1% Cu in 8% HNO 3 in a PFA Teflon vessel. ICPMS standards are prepared in 8% HNO 3 at concentrations ranging from 50 ppq to 1 ppb Cu. Matrix effects of high levels of Cu (1% solutions) cause significant signal suppression in ICPMS, so matrix removal is essential to achieve the best detection limits. Simple dilution of the sample matrix is not possible as Th is already close to the detection limit of ICPMS systems. It is therefore necessary to remove the Cu matrix while maintaining or increasing the Th signal. The TRUFAST uses the PFA column packed with TEVA Resin to concentrate Th from a solution of Cu dissolved in HNO 3, allowing for accurate measurement at low ppq levels. The TRUFAST system diagram is shown below. Elemental Scientific Phone: 1.402.991.7800 Email: esi@icpms.com 1500 North 24th Street Fax: 1.402.991.7799 Web: www.icpms.com Omaha, NE 68110 USA 1

Experimental Instrumentation, Sample Introduction, and ICPMS Parameters TRUFAST Sample Introduction System Thermo Scientific ELEMENT2 ICPMS CF-TEVA 50μL resin column Eluent Flow Rate: 200 µl/min Sample volume: 4 ml Table 1. ICPMS Conditions Parameter Value Rf Power 1400 Watts Nebulizer Gas Flow 0.7 L/min Plasma Gas Flow 16 L/min Auxiliary Gas Flow 0.9 L/min Additional Gas Flow 0.15 L/min Argon Sample Loop ICP Torch Sapphire Injector SC DX FAST Autosampler 12% HNO 3 Vacuum 1 FAST Valve Chelation Valve Preconcentration Column PFA-ST Nebulizer Chilled Cyclonic Spray Chamber PC3 @ +2 C ICPMS 2 3% HCl 3 Copyright 2010 Elemental Scientific, Inc. ESI Micro Peripump Figure 1. System diagram for determination of Th and Cu by preconcentration and matrix removal Procedure The TRUFAST system uses two high purity valves to take up an aliquot of sample, concentrate it on a PFA column packed with TEVA Resin, and elute the Th into a PFA nebulizer attached to an ICPMS spray chamber. Standards (8% HNO 3 ) and samples (1% Cu and 8% HNO 3 ) are vacuum loaded into a 4 ml sample loop on the FAST valve. The FAST valve switches to inject the sample into the column. The TEVA Resin retains the Th in the sample while the Cu matrix, which is not retained by the resin, is washed out of the column for 3 minutes with 12% HNO 3. After the Cu matrix has been removed from the column, the chelation valve is switched allowing 3% HCl to pass through the column to elute the Th as a transient peak. The sensitivity of this peak and the concentration of Th in the samples can be measured and calculated via the ICPMS software. Once the Th has been eluted from the column, the chelation valve is switched to recondition the column prior to the next analysis. The total run time is 11 minutes sample-to-sample. 2

Elution An elution curve can be obtained in less than 60 seconds. Th levels at sub-ppt show strong elution peaks, as shown in Figure 2. 1 ppt Th Blank Figure 2. Overlayed elution profiles of the procedural blank and 1 ppt Th 232. Calibration Figure 3 shows Th 232 calibration generated in external standards from 0 ppt to 1000 ppt, demonstrating excellent linearity across four orders of magnitude. However, when generating calibration curves it is best to restrict data point to within an order of magnitude of expected results. This prevents error in the high standards, such as 1000 ppt, from affecting the measurement of lower traces, such as 1 ppt. Figure 3. Calibration curve for Th in Cu up to 1000 ppt. 3

Figure 4 shows a separate calibration of Th 232 from 0 ppt to 1 ppt, with measurements taken at 0.05 ppt, 0.1 ppt, 0.5 ppt, and 1 ppt. The average 10 ppt Th spike recovery for 3 injections in a 1% Cu matrix was 10.1 ppt or 101%. The LOD (3s) for Th in 1% Cu was 8 ppq. Figure 4. Calibration curve for Th in Cu up to 1 ppt. FAST Washout After a 1 ppb sample is analyzed, the thorium washout is better than 1000x in the first blank. The second blank shows 4000x washout, while the third blank shows 8000x washout. The TRUFAST system may be configured to perform additional wash steps if triggered from the instrument QC function. If sample concentration is more than 1000x higher than the desired detection limit, then an extra rinsing step should be employed. Figure 5. Washout analysis of 3 blanks after a 1ppb sample 4

Long-Term Stability Long-term stability was evaluated repeatedly by measuring 100 ppt Th. Three bottles spiked with 100 ppt Th were each analyzed 20 times preceded and followed by blanks. The retention of Th on the column remains constant even after running many samples over a long period of time. Figure 6. 60 replicate determinations of 100 ppt Th in 1% Cu from three 100 ml sample bottles. Conclusion With its automated and versatile sample uptake and introduction capabilities, the TRUFAST system offers easy and efficient determination of low ppq levels of Th in high purity Cu. The TRUFAST system reduces sample prep time while achieving low ppq level detection limits. The long-term stability and washout characteristics of the column provide economical and time-saving advantages over systems with single use columns. References 1) Eichrom Technologies, Inc., Determination of thorium in high purity aluminum by ICP-MS. Application Note 502. 2) C. Cabral Jr., K.P. Rodbell and M.S. Gordon, Alpha particle mitigation strategies to reduce chip soft error upsets. J. Appl. Phys. 101, 014902 (2007). 3) M. Fukuda, Y. Hayashibe and Y. Sayama, Determination of nickel, cobalt, copper, thorium and uranium in high-purity zinc metal by ICP-MS with online matrix separation. Anal. Sci. Vol. 11, 13-16, (1995) Elemental Scientific Phone: 1.402.991.7800 Email: esi@icpms.com 1500 North 24th Street Fax: 1.402.991.7799 Web: www.icpms.com Omaha, NE 68110 USA 5