Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Similar documents
Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

ALLOWAY METHOD OUTLINE

Chloramine (Mono) and Nitrogen, Free Ammonia

Chlorine, Free and Total, High Range

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Chlorine, Free and Total, High Range

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows

LOW-RANGE TOTAL CHLORINE

Nitrogen, Total Inorganic

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Chlorine, Free and Total, Low Range

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Standard Operating Procedure VUSP M

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

CHLORINE, TOTAL (0 to 4.00 mg/l)

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

DR/4000 PROCEDURE. IRON, Total

DPD Test N Tube Method *

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Standard Operating Procedure VUSP N

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Alkalinity, USEPA by Flow Injection Analysis (FIA)

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

SECTION D.2 AMMONIA NITROGEN

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Trihalomethane Formation Potential (THMFP)

PAR Method Method to 2.0 mg/l Pb TNTplus 850

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Chloride by Flow Injection Analysis (FIA)

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

Oxygen Demand, Chemical

NITROGEN, AMMONIA, High Range, Test N Tube

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

Oxygen Demand, Chemical

Hexavalent Chromium by Flow Injection Analysis (FIA)

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

NITROGEN, TOTAL, HR, Test N Tube

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Glossary of Common Laboratory Terms

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority:

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

METHOD DETERMINATION OF PHOSPHORUS BY SEMI-AUTOMATED COLORIMETRY

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

DETERMINATION OF NITRATE IN 2M KCl SOIL EXTRACTS BY FLOW INJECTION ANALYSIS NITRATE REDUCTASE

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

DETERMINATION OF NITRATE IN BRACKISH WATERS BY FLOW INJECTION ANALYSIS NITRATE REDUCTASE

NITROGEN, TOTAL, Test N Tube * Method 10071

Appendix A. Glossary of Definitions, Acronyms, Abbreviations and Symbols. March 1997

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA)

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

Laboratory Chemical Safety. Pathology Training

Results of the EPA Method 1631 Validation Study

Percent Solids Determination

NITRITE, Low Range (0 to mg/l NO 2- -N)

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

HAZARD COMMUNICATION SAFETY PROGRAM

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

University of Tennessee Safety Procedure

Transcription:

Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water

Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished Drinking Water 1.0 Scope and Application 1.1 This method is for the determination of free chlorine (Cl) in finished drinking water. 1. The method is applicable in the range from 0.1 to 4.5 mg/l Cl. 1.3 This method is equally effective in performance and use to SM 4500-Cl G for the purposes of regulatory compliance reporting of Cl..0 Summary of Method An ammonia solution at a ph of 8.3 is added to a sample that contains free chlorine. The free chlorine is immediately converted into monochloramine (NHCl). In the presence of a cyanoferrate catalyst, the monochloramine reacts with a substituted phenol to form an intermediate monoimine compound. The intermediate couples with excess substituted phenol to form a green indophenol compound, which is proportional to the amount of free chlorine in the sample. A sample blank that contains Monochlor F Reagent corrects for background color from the reagent and sample. The measurement wavelength is 655 nm for spectrophotometers or 610 nm for colorimeters. 3.0 Interferences 3.1 The items listed in the Interfering Substances table have been individually checked up to the given concentrations and do not cause interference. The cumulative effects and influence of other ions have not been determined. Measurement results can be verified using sample dilutions or standard additions. Interfering substance Ozone > 1 Sulfide > 0.5 Interference level (mg/l) 4.0 Safety 4.1 The toxicity or carcinogenicity of each reagent used in this method has not been precisely determined; however, each chemical should be treated as a potential health hazard. Exposure to these chemicals should be reduced to the lowest possible level. It is suggested that the laboratory perform personal hygiene monitoring of each analyst using this method and that the results of this monitoring be made available to the analyst. 4. Unknown samples may contain high concentrations of volatile toxic compounds. Sample containers should be opened in a hood and handled with gloves to prevent exposure. 4.3 This method does not address all safety issues associated with its use. The laboratory is responsible for maintaining a safe work environment and a current awareness file of OSHA regulations regarding the safe handling of any chemicals specified in this method. A reference file of material safety data sheets (MSDSs) should be available to all 1

personnel involved in these analyses. Additional information on laboratory safety can be found in Sections 16.3 and 16.4. 5.0 Equipment Note: Brand names, suppliers, and part numbers are for illustrative purposes only. No endorsement is implied. Equivalent performance may be achieved using apparatus and materials other than those specified here, but demonstration of equivalent performance that meets the requirements of this method is the responsibility of the laboratory. 5.1 Sampling equipment 5.1.1 Sample collection bottles Collect samples in chlorine demand-free glass bottles. 6.0 Equipment for sample analysis 6.1 Hach Company DR 6000, DR 3900, DR 1900 spectrophotometer, or equivalent 6. Hach Company sample cells 6.3 Equipment for standard preparation 6.3.1 500 ml bottle or flask Chlorine demand-free glass 6.3. Analytical balance. 7.0 Reagents and Standards 7.1 Deionized water Water in which Cl concentration is below the detection limit of this method. Water prepared by passage of tap water through reverse osmosis and carbon filtration has been shown to be an acceptable source of reagent water. 7. Hach Company Monochlor F Reagent Pillows, Cat. No. 8099, or equivalent 7.3 Hach Company Freechlor F Reagent Solution, Cat. No. 96496, or equivalent 7.4 Hach Company Cl Ampule: 10 ml, 50-75 mg/l as Cl (Cat. No. 146810) or equivalent 7.5 Initial precision and recovery (IPR) solution 7.5.1 Prepare 4 or more replicate 1.0-1.5 mg/l Cl IPR solutions by weight. Tare an empty 500 ml bottle. Transfer the contents of a 10 ml standard ampule to the bottle (Sect. 7.4). Record the weight. Identify the concentration of the standard from the label. Using the concentration and the transferred weight of the standard, calculate the final dilution weight as (W1)(C1)/(C)=(W). Dilute to the final weight with DI water. Assuming the concentration in the Standard ampule is 75 mg/l, the calculated final concentration = 1.5 mg/l Cl.

7.6 Method detection limit (MDL) solution 7.6.1 Prepare 7 or more replicate Cl MDL solutions by weight (approximately 0.0 mg/l). Tare an empty 500 ml bottle. Transfer 60.00 g of a prepared IPR solution to the bottle. Dilute to 450 g with DI water. Final concentration = 0.0 mg/l Cl. 8.0 Sample Collection, Preservation and Storage 8.1 Samples should be collected in clean chlorine demand-free glass bottles. 8.1.1 Pretreat sample containers to remove chlorine demand. Soak containers in a weak bleach solution for at least 1 hr. Rinse fully with DI water. 8. Analyze samples immediately. 9.0 Quality Control 9.1 Each laboratory that uses this method is expected to operate a formal quality assurance program (16.1). The minimum requirements of this program consist of an initial demonstration of laboratory capability and ongoing analyses of laboratory prepared water standards as a test of continued performance to assess accuracy and precision. Laboratory performance is compared to established performance criteria to determine if the results of analyses meet the performance characteristics of the method. 9.1.1 The analyst shall make an initial demonstration of the ability to generate acceptable accuracy and precision with this method. This ability is established as described in Sections 9. and 9.3. The laboratory shall, on an ongoing basis, demonstrate through analysis of the ongoing precision and recovery sample that the analysis system is in control. 9.1. Accompanying QC for the determination of Cl is required per analytical batch. An analytical batch is a set of samples processed during a contiguous 8-hour period. Each analytical batch must be accompanied by an ongoing precision and recovery sample (OPR), matrix spike sample (MS), and matrix spike duplicate sample (MSD) resulting in a minimum of four analyses (1 OPR, 1 sample, MS, and MSD). 9. Initial demonstration of laboratory capability. 9..1 To establish the ability to detect Cl the analyst shall determine the MDL using the apparatus, reagents, and standards that will be used in the practice of this method. An achieved MDL less than or equal to the MDL in Section 13.0 is recommended prior to the practice of this method. 9.. Prepare and measure seven replicates of the MDL standard (Sect. 7.6) according to the procedure in Section 11 9..3 Using the results of the set of seven analyses, compute the MDL using the following equation: 3

where: MDL = x x n n 1 3.14 n = Number of samples (7) x = measured concentration of each sample 9.3 Initial precision and recovery (IPR) - To establish the ability to generate acceptable precision and accuracy, the analyst shall perform the following operations: 9.3.1 Prepare and measure four samples of the IPR standard (Sect. 7.5.1) according to the procedure in Section 11. 9.3. Using the results of the set of four analyses, compute the average percent recovery (x) and the standard deviation of the percent recovery (s) for Cl. Use the following equation for calculation of the standard deviation of the percent recovery: s x x n n 1 where: n = Number of samples (4) x = % recovery in each sample 9.3..1 Compare s and x with the corresponding limits for initial precision and recovery in Table 1 (Sect. 17). If s and x meet the acceptance criteria, system performance is acceptable and analysis of samples may begin. If, however, s exceeds the precision limit or x falls outside the range for recovery, system performance is unacceptable. In this event, correct the problem, and repeat the test. 9.4 Ongoing precision and recovery (OPR) - To demonstrate that the analysis system is in control, and acceptable precision and accuracy is being maintained with each analytical batch, the analyst shall perform the following operations: 9.4.1 Prepare a 1.0-1.5 mg/l recovery standard with each analytical batch as described in Sect. 7.5.1 and measure according to the procedure in Section 11. Calculate the percent recovery and compare this value with the limits for ongoing recovery in Table (Sect. 17). If the percent recovery meets the acceptance criteria, system performance is acceptable. If the percent recovery falls outside the acceptance criteria, system performance is unacceptable. In this event, correct the problem, and repeat the test. 4

9.4.1.1 Measure a field sample. After measuring the background concentration, spike the sample with a known concentration of Cl. The spike concentration should be 1-5 times the background concentration, but still within the reporting range of the method. Prepare a duplicate of this spiked sample. 9.4.1. Measure the spike duplicates and calculate the spike recovery for each sample and the relative percent difference (RPD) between the two results. Use the following equation to calculate the spike recovery: [Conc] [Bkgd] Spike Recovery = 100 [Sp] where: [Conc] = the measured concentration of the spiked sample [Bkgd] = the measured concentration of the un-spiked sample [Sp] = the concentration of the spike where: RPD = Conc 1 Conc ( Conc 100 1 + Conc ) Conc1 = the concentration of the first spiked sample Conc = the concentration of the second spiked sample 9.4.1.3 Compare the spike recoveries and RPD with the corresponding limits in Table (Sect. 17). If recoveries and RPD meet the acceptance criteria, system performance is acceptable and analysis of samples may begin. If recoveries or RPD fall outside the limits, system performance is unacceptable. In this event, correct the problem, and repeat the test. 9.4.1.4 The laboratory should add results that pass to IPR and previous OPR data and update QC charts to form a graphic representation of continued laboratory performance. The laboratory should also develop a statement of laboratory data quality for each analyte by calculating the average percent recovery (R) and the standard deviation of the percent recovery (sr). Express the accuracy as a recovery interval from R - sr to R + sr. For example, if R = 95% and sr = 5%, the accuracy is 85% to 105%. Control charts are acceptable for evaluating process control, but under no circumstances can the control limits be widened beyond those established in the acceptance criteria defined in Section 13. 10.0 Calibration and Standardization 10.1 The Hach DR series spectrophotometers have a built-in calibration that is automatically initiated when the Indophenol Free Cl procedure is selected through the instrument interface. No further initial calibration is required. However, the instruments have the capability of developing a user-calibration. See manufacturer s manual for instructions. 5

10. Calibration Verification 10..1 To verify that the instrument is measuring Cl properly, analyze 0.0 mg/l (Sect. 7.6.1) and 1.5 mg/l (Sect. 7.5.1) Cl standards. Results should be within 15 percent of the actual value. Perform this calibration verification daily while instrument is in use. If the calibration verification standard result is outside the limit, it is unacceptable. In this event, correct the problem, and repeat the test. 11.0 Procedure 11.1 Instrument Setup follow the instrument manufacturer s instructions for instrument setup. 11. Fill two sample cells with 10 ml of sample, one cell will be the sample and one will be the blank. 11.3 Add 5 drops of Freechlor F Reagent to the sample cell. 11.3.1 Stopper the sample cell and invert to mix. 11.4 Add the contents of one Monochlor F Reagent Powder Pillow to both cells. 11.4.1 Stopper the cells and shake for ~0 s to dissolve the reagent. 11.5 Allow the samples to react for 5 min. 11.6 Invert the cells to mix. 11.7 Insert the blank cell into the spectrophotometer and zero. 11.8 Insert the sample cell into the spectrophotometer and read. Results will display in mg/l Cl. 1.0 Data Analysis and Calculations 1.1 Cl concentration is calculated automatically against internal instrument calibration. 13.0 Method Performance Performance of the method was demonstrated in multi-lab studies comparing the method against EPA Reference Method SM 4500-Cl G. The method was evaluated in a low ionic strength reference matrix as well as multiple geographically diverse finished drinking water samples obtained from both surface water and ground water sources. Validation Results Section Limit Method Detection Limit 9. 0.05 mg/l Cl Initial Recovery Range 9.3 89.5% - 101% 6

Initial Precision 95% 9.3 0.0 Matrix Recovery Range 9.4 89.9 110% Matrix Recovery Precision 95% 9.4 0.06 14.0 Pollution Prevention 14.1 Follow guidelines in Section 15. 15.0 Waste Management 15.1 It is the laboratory s responsibility to comply with all federal, state, and local regulations governing waste management, particularly the hazardous waste identification rules and land disposal restrictions, and to protect air, water, and land by minimizing and control all releases from fume hoods and bench operations. Compliance with all sewage discharge permits and regulations is also required. 15. For further information on waste management, consult The Waste Management manual for Laboratory Personnel, and Less is Better: Laboratory Chemical Management for Waste Reduction, both available from the American Society s Department of Government Relations and Science Policy, 1155 16 th Street N.W., Washington, D.C. 0036. 16.0 References 16.1 Protocol for the Evaluation of Alternate Test Procedures for Organic and Inorganic Analytes in Drinking Water, USEPA, EPA-815-R-15-007, February 015. 16. 40 CFR 136, Appendix B. 16.3 OSHA Safety and Health Standards, General Industry, (9 CFR 1910), Occupational Safety and Health Administration, OSHA 06 (Revised, January 1976) 16.4 Safety in Academic Chemistry Laboratories, American Chemical Society, Committee on Chemical Safety, 3 rd Edition, 1979. 16.5 Water Analysis Handbook, Hach Company, 8 th Edition, 013. 7

17.0 Tables 17.1 Acceptance Criteria for Performance tests The QC performance criteria for this method were performed with a Hach Company DR6000 Spectrophotometer and Indophenol Free Cl Reagents. Table 1. Initial Precision and Recovery Acceptance Criteria Parameter Acceptance Criteria Relative Standard Deviation 10% Percent Recovery Range 100 ± 15% Table. Ongoing Precision and Recovery Acceptance Criteria Parameter Acceptance Criteria Lab Fortified Blank Recovery 100 ± 15% Sample Matrix Spike Recovery 100 ± 15% Sample Matrix Spike RPD 10% 8

18.0 Glossary of Definitions and Purposes The definitions and purposes are specified to this method but have been conformed to common usage as much as possible. 18.1 Units of weight and measure and their abbreviations 18.1.1 Symbols o C: degrees Celsius 18.1. Alphabetical characters mg/l: milligram per liter 18. Definitions, acronyms, and abbreviations 18..1 MDL: Method detection limit 18.. IPR: Initial precision and recovery 18..3 OPR: On-going precision and recovery 18..4 MS: Matrix spike 18..5 MSD: Matrix spike duplicate 18..6 LIS: Low ionic strength, deionized water