Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

Similar documents
A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph

Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit

The Role of the Carbon Cartridge in the One-Pass System

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Application Note. Abstract. Authors. Environmental

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Author. Abstract. Introduction

ENVIRONMENTAL analysis

Appendix 2 Data Distributions for Contaminants in Water and Sediment Samples in Response to the Deepwater Horizon Oil Spill, 2010

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

POTENTIAL VAPOR INTRUSION ENVIRONMENTAL

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Amec Foster Wheeler Environment & Infrastructure, Inc BIG SHANTY ROAD, NW, SUITE 100 KENNESAW, GEORGIA (770)

Ultra-Inert chemistry for Trace Level Analysis

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

STANDARD OPERATING PROCEDURES SOP: 1821 PAGE: 1 of 33 REV: 0.0 DATE: 11/25/94

Application Note. Abstract. Authors. Introduction. Food Safety

2. September 2015 Sub-Concrete Slab Soil Investigation

Non-Potable Water A2LA Certificate

October 26, Dear Mr. White:

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

STANDARD OPERATING PROCEDURES

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

DUAL CARTRIDGE SYSTEM FOR THE EXTRACTION OF ACIDS, BASES, AND NEUTRALS IN WATER

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Certificate of Accreditation

Certificate of Accreditation

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Certificate of Accreditation

Challenging Pesticide Analysis Using an Agilent J&W DB-35ms Ultra Inert GC Column

Semivolatile Organic Reference Materials

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Selection of a Capillary

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

The Suite for Environmental GC Analysis

NEW Agilent J&W Ultra Inert Capillary GC Columns. Proven performance that raises the bar on. consistent. column inertness.

Determination of 24 PAHs in Drinking Water

Selection of a Capillary GC Column

Table 1 Soil and Groundwater Sampling Summary Table

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Methods and Analytes Requiring Laboratory Certification Attachment 3

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Evaluation of Capillary Columns for General Performance Parameters

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Certificate of Accreditation

Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood Using GC/MS

The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Semivolatile Organic Compounds By Gas Chromatography / Mass Spectrometry (GC/MS)

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

317 Elm Street Milford, NH (603) Fax (603)

Certificate of Accreditation

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

MULTIRESIDUE ANALYSIS OF PRIORITY POLLUTANTS IN SURFACE WATERS USING EXACT MASS GC-TOF MS

Table of Contents ~ 1 ~

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Industrial Applications with a New Polyethylene Glycol-Based GC Column

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Facility: Mansfield Revision 3. Title: PAHs by SPME Page 1 of 11

METHOD 8410 GAS CHROMATOGRAPHY/FOURIER TRANSFORM INFRARED (GC/FT-IR) SPECTROMETRY FOR SEMIVOLATILE ORGANICS: CAPILLARY COLUMN

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Understanding Gas Chromatography

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

GC Analysis of Polybrominated Flame Retardants Application

Appendix C. Validation Standard Operating Procedures

Solid Phase Micro Extraction of Flavor Compounds in Beer

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

STANDARD OPERATING PROCEDURES

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

SUBSURFACE (PHASE II) INVESTIGATION REPORT

Low Bleed Stationary Phases for Gas Chromatography

Application Note. Authors. Abstract. Xin Sun Shandong Agriculture University, Taian, , China

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Analysis of Organic Pollutants by Micro Scale Liquid-Liquid Extraction. and On-column Large Volume Injection Gas Chromatography. Mark S.

Rely on Rxi -PAH Columns

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Analytical Trap Comparison for USEPA Method 8260C

Transcription:

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column Application Note Environmental Author Doris Smith and Kenneth Lynam Agilent Technologies, Inc. 285 Centerville Road Wilmington, DE 1988 USA Abstract Trace-level semivolatile organics analyses using methods such as USEPA Method 827 are important tools for assessing environmental contaminants worldwide. The wide-ranging chemical diversity of target semivolatiles can prove chromatographically challenging. This application note demonstrates the benefits of using an Agilent J&W HP-5ms Ultra Inert Capillary GC column with electron impact single quadrupole scanning mass spectrometry for trace-level semivolatiles analysis. Agilent Technologies has implemented new testing procedures to more effectively evaluate GC column inertness performance. These new testing procedures employ deliberately aggressive probes to thoroughly investigate column inertness and quality. These aggressive probes, including 1-propionic acid, 4-picoline, and trimethyl phosphate, are used to verify each column's inertness performance.

Introduction USEPA Method 827 [1] is a commonly used method for detecting semivolatile organic compounds in environmental samples by GC/MS. This method encompasses several classes of analytes, including amines, alcohols, polycyclic aromatic hydrocarbons, and phenols. The acidic and basic nature of many of the analytes makes minimizing any column or instrument activity critical to good chromatography and reliable results. Minimizing activity in the GC column is essential in maximizing an analyte's response. Nitrophenols are among the most active compounds in semivolatiles series. 2,4-Dinitrophenol in particular is notorious for showing low response through adsorption onto active sites in the flow path during analysis. At low concentrations, the response factor (RF) for 2,4-dinitrophenol can fall below the minimum average RF of.5 required by USEPA 827 due to interaction between the analyte and sample flow path. Capillary GC column activity as a potential source of result uncertainty has been effectively eliminated with the Ultra Inert series of columns. A custom standard containing an abbreviated list of analytes specific to USEPA Method 827 was analyzed to evaluate column performance. This semivolatiles short mix contained n- nitrosodimethylamine, aniline, benzoic acid, 2,4-dinitrophenol, 4-nitrophenol, 2-methyl-4,6-dinitrophenol, pentachlorophenol, 4-aminobiphenyl, benzidine, 3,3'dichlorobenzidine, benzo[b]fluoranthene, and benzo[k]fluoranthene, along with the recommended internal standards. These target analytes were chosen based on their chemical activity, as well as their poor chromatographic behavior. The short mix is particularly useful for rapid evaluation of system performance for semivolatiles analysis. Challenging analytes from early-eluting nitrosoamines through late-eluting PAHs are represented in this mix and chromatographic performance can be assessed quickly. A second large mix standard containing a broader selection of semivolatiles was also evaluated to show the Ultra Inert's performance when analyzing a more complex sample. This standard contained a variety of acidic, basic, and neutral groups, which ranged from very low-boiling components to high-boiling polycyclic aromatic hydrocarbons. Experimental An Agilent 689N GC/5975B MSD equipped with a 7683B autosampler was used for this series of experiments. Table 1 lists the chromatographic conditions used for these analyses. Table 2 lists flow path consumable supplies used in these experiments. Table 1. Chromatographic Conditions for EPA Method 827 Calibration Standards GC: Agilent 689N/5975B MSD Sampler: Agilent 7683B, 5.-µL syringe (Agilent p/n 5181-1273) 1. µl splitless injection Carrier: Helium 3 cm/s, constant flow Inlet: Splitless; 26 C, purge flow 5 ml/min at.5 min Gas saver 8 ml/min at 3 min Inlet liner: Deactivated dual taper direct connect (Agilent p/n G1544-87) Column: Agilent HP-5ms Ultra Inert 3 m.25 mm.25 µm (Agilent p/n 1991S-433UI) Oven: 4 C (1 min) to 1 C (15 C/min), 1 C/min to 21 C (1 min), 5 C/min to 31 C, hold 8 min Detection: MSD source at 3 C, quadrupole at 18 C, transfer line at 29 C, scan range 45 to 45 amu Table 2. Flow Path Supplies Vials: Amber screw top glass vials (Agilent p/n 5183-272) Vial caps: Blue screw caps (Agilent p/n 5182-723) Vial inserts: 1 µl glass/polymer feet (Agilent p/n 5181-8872) Syringe: 5 µl (Agilent p/n 5181-1273) Septum: Advanced Green (Agilent p/n 5183-4759) Inlet liners: Deactivated dual taper direct connect (Agilent p/n G1544-87) Ferrules:.4 mm id short; 85/15 Vespel/graphite (Agilent p/n 5181-3323) 2x magnifier: 2x magnifier loupe (Agilent p/n 43-12) Sample Preparation A 12-component custom semivolatiles mix was purchased from Ultra Scientific (Kingston, RI) and used to prepare a seven-level calibration standard set. The stock semivolatiles solution as delivered had a nominal concentration of 2, µg/ml. An internal standard mix as recommended by USEPA Method 827 was purchased from AccuStandard (New Haven, CT). The internal/surrogate solution as delivered had a 2

nominal concentration of 4, µg/ml. The calibration standards were prepared with component and internal standard concentrations of 8, 4, 2, 1, 5, 2, and 1 µg/ml. All solutions were prepared in dichloromethane using class A volumetric pipettes and flasks. The dichloromethane used was Burdick and Jackson spectral grade purchased thorough VWR International (West Chester, PA). Dichloromethane was used as a reagent blank and syringe wash solvent. The EPA 827 Calibration Level 2 standard set was purchased from AccuStandard containing 83 semivolatile components and internal standards. The large mix calibration standard was prepared at an analyte concentration of 5 µg/ml. Results and Discussion Baseline Inertness Profile for Ultra Inert Columns The basic approach for inertness verification for the Agilent J&W Ultra Inert series of capillary GC columns is testing with aggressive active probes at low concentration and low temperature. This is a rigorous approach that establishes consistent baseline inertness profiles for each column in the Agilent J&W Ultra Inert GC column series. The baseline inertness profile then serves as a predictor for successful analysis of chemically active species that tend to adsorb onto active sites, particularly at trace level like the semivolatiles in this application example. A more detailed description of the test mix and additional application examples can be found in references 2 through 7. Semivolatiles Analysis (USEPA 827) In this application note a seven-level semivolatile calibration curve set was evaluated over the concentration range of 1 to 8 µg/ml on an Agilent J&W Ultra Inert HP-5ms 3 m.25 mm.25 µm (p/n 1991S-433UI). An example chromatogram of a 1-µL injection of the 1 µg/ml short mix calibration standard is shown in Figure 1. Scanning mode was used exclusively for this analysis. Pentachlorophenol and benzidine are two components that are used to verify inlet and column inertness. Excessive peak tailing of these components would indicate column activity. Analysis of the short mix standard yielded sharp, symmetrical peak shapes for the problematic analytes as shown in Figure 2. Good separation was obtained in the analysis of the 5-ng on-column 827 large mix standard for each of the semivolatiles, which is shown in Figure 3. Semivolatile analysis by USEPA Method 827 requires a minimum average RF of.5 for a system performance check compound such as 2,4-dinitrophenol. 2,4-Dinitrophenol is a highly active analyte that has proven to be one of the most challenging compounds, often yielding lower than expected 2 3 5 6 11 1 12 1. n-nitrosodimethylamine 2. Aniline 3. 1,4-Dichlorobenzene-d4 (IS) 4. Benzoic acid 5. Naphthalene-d8 (IS) 6. Acenaphthene-d1 (IS) 7. 2,4-Dinitrophenol 8. 4-Nitrophenol 9. 2-Me-4,6-dinitrophenol 13 1. 4-Aminobiphenyl 11. Pentachlorophenol 12. Phenanthrene-d1 (IS) 13. Benzidine 14. Chrysene-d12 (IS) 15. 3,3 -Dichlorobenzidine 16. Benzo[b]fluoranthene 17. Benzo[k]fluoranthene 18. Perylene-d12 (IS) 14 15 16 17 18 1 4 7 8 9 Abundance 12 11 1 9 8 7 6 5 4 3 2 1 4. 6. 8. 1. 12. 14. 16. 18. 2. 22. 24. 26. 28. 3. Time Figure 1. Total ion chromatogram (SCAN mode) of the 1-ng on-column EPA827 short mix standard solution loading on an Agilent J&W HP-5ms Ultra Inert 3 m.25 mm.25 µm capillary GC column (p/n 1991S-433UI). Chromatographic conditions are listed in Table 1. 3

Abundance 36 32 28 1 2 1. Pentachlorophenol 2. Benzidine 24 2 16 12 8 4 15. 16. 17. 18. 19. 2. 21. 22. 23. Time Figure 2. Enlarged section of the total ion chromatogram for a 1-µL injection of 1. µg/ml EPA 827 short mix standard. The peaks of interest noted in the figure are two semivolatiles that are prone to peak tailing. Chromatographic conditions are listed in Table 1. 1 2 1. n-nitrosodimethylamine 2. Pyridine 3. 2-Fluorophenol 4. Phenol-d5 5. Phenol 6. Aniline 7. Bis(2-chloroethyl)ether 8. 2-Chlorophenol 9. 1,3-Dichlorobenzene 1. 1,4-Dichlorobenzene-D4 11. 1,4-Dichlorobenzene 12. Benzyl alcohol 13. 1,2-Dichlorobenzene 14. o-cresol 15. Bis(2-chloroisopropyl)ether 16. p-cresol 17. n-nitroso-di-n-propylamine 18. Hexachloroethane 19. Nitrobenzene-d5 2. Nitrobenzene 21. Isophorone 22. 2-Nitrophenol 23. 2,4-Dimethylphenol 24. Benzoic acid 25. Bis(2-chloroethoxy)methane 26. 2,4-Dichlorophenol 27. 1,2,4-Trichlorobenzene 28. Naphthalene-d8 29. Naphthalene 3. 4-Chloroaniline 31. Hexachlorobutadiene 3 4-2 32. 4-Chloro-3-methylphenol 33. 2-Methylnaphthalene 34. Hexchlorocyclopentadiene 35. 2,4,6-Trichlorophenol 36. 2,4,5-Trichlorophenol 37. 2-Fluorobiphenyl 38. 2-Chloronaphthalene 39. 2-Nitroaniline 4. Dimethyl phthalate 41. 2,6-Dinitrotoluene 42. Acenaphthylene 43. 3-Nitroaniline 44. Acenaphthene-d1 45. Acenaphthene 46. 2,4-Dinitrophenol 47. 4-Nitrophenol 48. Dibenzofuran 49. 2,4-Dinitrotoluene 5. Diethyl phthalate 51. Fluorene 52. 4-Chlorophenyl phenyl ether 53. 4-Nitroaniline 54. 2-Me-4,6-dinitrophenol 55. n-nitrosodiphenylamine 56. Azobenzene 57. 2,4,6-Tribromophenol 58. 4-Bromophenyl phenyl ether 59. Hexachlorobenzene 6. Pentachlorophenol 61. Phenanthrene-d1 62. Phenanthrene 21-31 32 33 34 35 36 37 38 39 42 44 4 41 45 43 48 49 46 47 63. Anthracene 64. Carbazole 65. Dibutylphthalate 66. Fluoranthene 67. Benzidine 68. Pyrene 69. p-terphenyl-d14 7. Benzyl butyl phthalate 71. 3,3-Dichlorobenzidine 72. Benzo[a]anthracene 73. Chrysene-D12 74. Chrysene 75. Bis(2-ethylhexyl)phthalate 76. Di-n-octyl phthalate 77. Benzo[b]fluoranthene 78. Benzo[k]fluoranthene 79. Benzo[a]pyrene 8. Perylene-d12 81. Indeno[1,2,3,cd]pyrene 82. Dibenzo[a,h]anthracene 83. Benzo[g,h,i]perylene 5-54 62 59 61 55 56 58 6 57 63 64 65 68 66 69 7 67 71-75 77 78 79 76 8 82 81 83 5. 1. 15. 2. 25. 3. 35. Figure 3. Total ion chromatogram (SCAN mode) of 5-ng on-column loading of EPA 827 calibration (large mix) standard solution on an Agilent J&W HP-5ms Ultra Inert 3 m.25 mm.25 µm capillary GC column (p/n 1991S-433UI). Chromatographic conditions are listed in Table 1. 4

response factors at lower concentrations. In the analysis of the short mix calibration standard, the response for 2,4-dinitrophenol was greater than.1 at the 1-ng level. The average response was.15 over the concentration range studied. An example chromatogram for the signal-to-noise ratio for a 1-ng on-column loading of 2,4-dinitrophenol is shown in Figure 4. The signal-to-noise ratio for this difficult analyte was greater than 16 to 1. This demonstrates the excellent performance of the HP-5ms Ultra Inert GC column. Linearity was excellent across the range studied, giving R 2 values of.99 or greater for even the more difficult phenols. Figure 5 indicates the correlation coefficients for several of the more active analytes. Abundance 2 19 18 17 16 15 14 13 12 11 1 Figure 4. 9 8 7 6 5 4 3 2 1 2,4-Dinitrophenol S/N = 16.86 Signal = 118 abundance units Noise = 7 abundance units 11. 11.5 12. 12.5 13. 13.5 14. 14.5 15. 15.5 Time Enlarged section of the total ion chromatogram (scan mode) for a 1-µL injection of 1 µg/ml EPA Method 827 short mix standard on an Agilent J&W HP-5ms Ultra Inert 3 m.25 mm.25 µm capillary GC column (p/n 1991S-433UI). The peak in the figure is 2,4-dinitrophenol, one of the more demanding semivolatiles. This injection represents an on-column loading of 1 ng per component. Chromatographic conditions are listed in Table 1. R 2 n-nitrosodimethylamine.995 2,4-Dinitrophenol.99 4-Nitrophenol.995 Pentachlorophenol.995 Benzidine.995 Response 1.e+7 5.e+6 Benzidine Figure 5. 5 Concentration µg/ml Correlation coefficients for some of the more challenging analytes in the EPA Method 827 short mix standard over the 1 to 8 µg/ml range of this study and an example linear regression plot for benzidine. 5

Conclusions This application successfully demonstrates the use of an Agilent J&W HP-5ms Ultra Inert capillary GC column for low-level semivolatile organics. Linearity was excellent for all semivolatiles studied, yielding.99 or greater R 2 values down to a 1-ng column loading of each component. One of the reasons for excellent linearity and high R 2 values is the highly inert surface of the column. The lack of chemically active sites makes these columns an excellent choice for semivolatiles analyses. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. This study was done using SCAN mode on an Agilent 689N/5975B GC/MSD equipped with an inert electron impact source. The signal-to-noise ratio for a 1-ng on-column loading of 2,4-dinitrophenol was greater than 16 to 1 with this system. This result clearly shows the power of using an Agilent J&W HP-5ms Ultra Inert column for low-level semivolatile organics analysis. Lower limits of quantification are expected when using one of Agilent's latest GC/MS offerings, such as the 789A/5975C GC/MSD Triple-Axis Detector coupled with an Agilent J&W HP-5ms Ultra Inert GC capillary column. References 1. US EPA Method 827D, Revision 4, February 27, "Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS)" 2. Mitch Hastings, Allen K. Vickers, and Cameron George, "Inertness Comparison of Sample of 5% Phenyldimethylpolysiloxane Columns," Poster Presentation, 54th Annual Pittsburg Conference, Orlando, FL, March 23 3. "Agilent J&W Ultra Inert GC Columns: A New Tool to Battle Challenging Active Analytes," Agilent Technologies publication 5989-8685EN, May 29, 28 4. Mike Szelewski and Bill Wilson, "Improvements in the Agilent 689/5973 GC/MSD System for Use with USEPA Method 827," Agilent Technologies publication 5988-372EN, November 7, 21 5. Kenneth Lynam, "Semivolatile Analysis Using an Inertness Performance Tested Agilent J&W Ultra Inert DB-5ms Column," Agilent Technologies publication 5989-8616EN, May 13, 28 6. Kenneth Lynam and Doris Smith, "Polycyclic Aromatic Hydrocarbon (PAH) Analysis Using an Agilent J&W DB-5ms Ultra Inert Capillary GC Column," Agilent Technologies publication 5989-9181EN, July 28 7. Kenneth Lynam and Doris Smith, "Polybrominated Diphenyl Ether (PBDE) Analysis Using an Agilent J&W DB-5ms Ultra Inert Capillary GC Column," Agilent Technologies publication 5989-9571EN, August 28 www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 29 Published in the USA January 2, 29 599-3416EN