Z -LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar Reader

Similar documents
Predictor Assay Setup Guide on the BMG LABTECH CLARIOstar Microplate Readers

Predictor Assay Setup Guide on the Tecan Safire 2 Microplate Reader

Optimization of an Adapta Kinase Assay for CAMK1

LanthaScreen Eu Kinase Binding Assay Validation Packet. Optimization of a LanthaScreen Eu Kinase Binding Assay for AURKB

LanthaScreen Eu Kinase Binding Assay for GAK

LanthaScreen Eu Kinase Binding Assay Validation Packet. Optimization of a LanthaScreen Eu Kinase Binding Assay for RPS6KA1

ab Alkaline Phosphatase Assay Kit (Luminometric)

Caspase 3 Inhibitor Drug Detection Kit

ATP Fluorometric/Colorimetric Assay Kit

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology

LanthaScreen Eu Kinase Binding Assay Validation Packet

D-Lactate Fluorometric/Colorimetric Assay Kit

Catalase Fluorometric/Colorimetric Assay Kit

Instructor s Advance Preparation

RayBio MMP-3 Activity Assay Kit

Phospholipase C Fluorometric/Colorimetric Assay Kit

Alkaline Phosphatase Colorimetric Assay Kit

Firefly Luciferase 1. ATP + Luciferin AMP + Oxyluciferin + Light (565 nm)

P450-Glo Screening Systems All technical literature is available on the Internet at: Please visit the web site to verify that you

Color Compensation Set 1

Protein Quantification Kit (Bradford Assay)

RayBio Human Phospho-eIF- 2alpha (Ser52) ELISA Kit

LanthaScreen Eu Kinase Binding Assay Validation Packet. Optimization of a LanthaScreen Eu Kinase Binding Assay for STK33

ab Caspase 3, Caspase 8 and Caspase 9 Multiplex Activity Assay Kit (Fluorometric)

For the rapid, sensitive and accurate detection of Caspase 5 inhibition by various compounds

Caspase 7 Inhibitor Drug Detection Kit

Monensin ELISA Kit. Catalog Number KA assays Version: 11. Intended for research use only.

L-malic acid measurement in wines using the SpectraMax Plus 384 Microplate Reader

Nitric Oxide Synthase Ultrasensitive Colorimetric Assay

Formaldehyde Fluorescent Detection Kit

QuickZyme Total Protein Assay (to be used with acid hydrolyzates)

Hydrogen Peroxide Colorimetric Detection Kit

Mouse Anti-OVA-IgE ELISA KIT

Miniaturized HTS Assays Using the Aurora Discovery BioRAPTR FRD Workstation Automated Protocol #EP029

RayBio Rat TNF-alpha ELISA Kit (For Lysates)

This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power.

For the rapid, sensitive and accurate measurement of Caspase 3 Activity in cell and tissue lysates

RayBio Human IDUA ELISA Kit

Zearalenone ELISA Kit

Aflatoxin M1 (AFM1) ELISA Kit

Tetracyclines (TCs) ELISA Test Kit

QuickZyme Hydroxyproline Assay

Mouse Insulin ELISA KIT S-Type

Predictor herg Fluorescence Polarization Assay Frequently Asked Questions General

Acetylcholinesterase Assay Kit

Application Note TABLE OF CONTENTS

MyBioSource.com. This package insert must be read in its entirety before using this product.

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats

Malachite Green Phosphate Detection Kit Catalog Number: DY996

RayBio Glucose Dehydrogenase Activity Assay Kit

RayBio Human Thyroid Peroxidase ELISA Kit

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit

CERTIFICATE OF ANALYSIS Nitric Oxide Assay Kit

Myeloperoxidase Colorimetric Assay Kit

Optimization of a LanthaScreen Eu Kinase Binding Assay for PIK3CA/PIK3R1

Acetylcholinesterase Assay Kit (Fluorometric - Red)

For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates

IncuCyte ZOOM NeuroTrack Fluorescent Processing

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA

Human 25-hydroxy vitamin D3 (25HVD3) ELISA Kit

Certificate of Analysis

Applications Note 215 September 2009

ADH Activity Assay Kit

Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer.

Xanthine Oxidase Assay Kit

Human anti-ige receptor antibody ELISA Kit

3D-Printed Fluidic Devices Enable Quantitative Evaluation of Blood Components in Modified Storage Solutions for Use in Transfusion Medicine

Human hepatitis B virus surface antigen(hbsag) ELISA Kit

Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit

Human anti-gliadin antibody (IgA)ELISA Kit

Human Syndecan-3 ELISA Kit

For the rapid, sensitive and accurate measurement of Urea levels in various samples.

Human Creatinine Serum Detection Kit

GGT Activity Assay Kit (Colorimetric)

Peroxidase Assay Kit. Catalog Number KA assays Version: 02. Intended for research use only.

Human vitamin A (VA) ELISA

Measurement of L-Malic Acid in Wines Using the SpectraMax Plus 384 Microplate Reader

Kinase-Glo Luminescent Kinase Assay Platform

RayBio Protease Activity Assay Kit

Characterization of Reversible Kinase Inhibitors using Microfluidic Mobility-Shift Assays

Human Coagulation Factor X Total Antigen ELISA Kit

Mouse TNF-alpha ELISAKit

Caspase 3 (active) Red Staining Kit

Nexcelom ViaStain Live Caspase 3/7 Detection for 2D/3D Culture

Human Anti-Ovary Antibody (IgG)ELISA Kit

Additional Materials Required (NOT PROVIDED):

For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates.

For the rapid, sensitive and accurate measurement of Pyruvate Kinase (PK) activity in various samples.

Porcine Immunoglobulin E (IgE)ELISA Kit

Pig immunoglobulin E (IgE) ELISA Kit

Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit

Ammonia Assay Kit. Catalog Number KA assays Version: 02. Intended for research use only.

ab83360 Ammonia Assay Kit

Acetylcholinesterase Assay Kit (Fluorometric - Green)

Testosterone ELISA, EIA

Rat Prolactin ELISA Kit

Alcohol Dehydrogenase Detection Kit

Golimumab Free Anti-Drug Antibody ELISA

Human plasma protease C1 inhibitor (SERPING1) ELISA Kit

Transcription:

08 Jan 15 Page 1 of 18 Z -LYTE Assay Setup Guide on the BMG LABTECH CLARIOstar Reader The BMG LABTECH CLARIOstar Microplate Readers were tested for compatibility with the Life Technologies Z -LYTE Assay using the Z'-LYTE Ser/Thr 11 Kit (PV3670) and FRAP1 (mtor) kinase (PV4753). The following document is intended to demonstrate setup of this instrument and provide representative data. For more detailed information and technical support of Life Technologies assays please call 1-800-955-6288 extension 40266. For more detailed information and technical support of BMG LABTECH instruments or software, please contact BMG LABTECH at 1-877-264-5227 or www.bmglabtech.com. A. Recommended Optics wavelength (nm) Wavelength Selection Excitation 400-12 (or similar) Emission 1 445-12 (or similar) Emission 2 520-12 (or similar) LVF Monochromator LVF Monochromator LVF Monochromator Emission 1 Dichroic 422.5 Linear Variable Dichroic Emission 2 Dichroic 460 Linear Variable Dichroic

08 Jan 15 Page 2 of 18 B. Instrument Setup 1. Make certain plate reader is turned on, and open up CLARIOstar Control software on computer. Insert plate into plate reader. 2. Setup protocol for Z'-LYTE, select "Manage Protocols" from the Protocols tab in the ribbon style menu at the top of the window.

08 Jan 15 Page 3 of 18 3. At this point, a new screen will open (below). Click on the Show all test protocols or "Fluorescence Intensity" button on the left side of the screen, then select New from the tabs at the bottom.

08 Jan 15 Page 4 of 18 4. A new window will pop up. Select Fluorescence Intensity and Endpoint and then select OK.

08 Jan 15 Page 5 of 18 5. A new Protocol window will open automatically. Enter a test name, select plate type, type 2 into box beside No. of multichromatics and in Optic select Top optic. In Speed and Precision select Rapid or Precise or click on.

08 Jan 15 Page 6 of 18 6. Additional selections are now made available to change the speed at which the plate is read. When finished, select the "Layout" tab at the top of the Protocol window.

08 Jan 15 Page 7 of 18 7. Select the wells you wish to read. Note in this step you can select to designate blanks, positive controls, etc. but for this case we marked all wells "Sample" and calculations were performed manually. When finished, select the Multichromatic tab.

08 Jan 15 Page 8 of 18 8. Adjust the monochromator settings by typing in the desired wavelength bandpass. For both chromatics, type in 400-12 for the Excitation. For chromatic 1, type in an Emission of 445-12 and for chromatic 2, type in an Emission 520-12. The Dichroic settings will be adjusted automatically. At this time you can select Start measurement to run this protocol immediately or OK to save this protocol.

08 Jan 15 Page 9 of 18 9. If you selected OK in the previous step you will return to the initial settings window. Select the Start measurement button in the menu bar at the top of the screen.

08 Jan 15 Page 10 of 18 8. A new window will appear allowing you to select which of your test protocols to be run. Select the protocol created for Z'-LYTE, and then press OK.

08 Jan 15 Page 11 of 18 9. A new window will appear. Place the plate into the reader, and select a well to use for adjusting gain and focus by highlighting the well of your choice. The gain or sensitivity can be adjusted for the highlighted channel at this point; in this case a positive control (high FRET) well was used. When finished, click on the "Start Adjustment" tab. This will procedure will need to be followed for both channels.

08 Jan 15 Page 12 of 18 10. In a moment, the instrument will have calculated its optimal focal height and the gain adjustments necessary. When finished with adjusting both channels, click on the "Start Measurement" tab to read.

08 Jan 15 Page 13 of 18 11. When CLARIOstar has finished reading, collect the data by clicking the MARS icon in the Results portion of the ribbon on the toolbar at the top of the window. This will automatically redirect you to the MARS data evaluation file which collects run data. Select the run of interest from the list to open, and view data in a plate layout format.

08 Jan 15 Page 14 of 18 C. Z'-LYTE Kinase Assay using FRAP1 (mtor) Purpose The following is a sample assay performed for demonstration purposes. The section below describes how the data was obtained, and is not intended for use as an assay protocol. We recommend all first-time users follow the appropriate protocols and/or validation packets provided with their specific assay kits, and include all proper controls. The instrument settings above would be sufficient for any Z'-LYTE assay and were optimized for high throughput. The information below is provided as representative data. Assay was run at ATP Km apparent and a kinase concentration producing approximately 30-40% of maximal phosphorylation, as discussed in the Z'-LYTE protocols. ATP and kinase concentrations should be optimized for each kinase by the actual user. Specific Z -LYTE assay protocols and setup information from Life Technologies own in-house SelectScreen Custom Profiling Z -LYTE -based kinase assay service can be located at the following link: http://www.lifetechnologies.com/selectscreen and select the SelectScreen Kinase Profiling Service and select Technology Overview. At a Glance Step 1: Step 2: Step 3: Step 4: Setup the instrument and plate layout. Dilution series of inhibitor. Prepare assay plate with controls. Read. Materials Required Component Part Number Storage FRAP1 (mtpor) kinase PV4753-80 C Z'-LYTE -LYTE Assay Kit Ser/Thr 11 containing PV3670 Z'-LYTE Kinase Assay Kit - Ser/Thr 11 Peptide PV3671-20 C Z'-LYTE Ser/Thr 11 Phospho-Peptide PV3672-20 C 5X Kinase Buffer A PV3189 20-30 C ATP PV3227-20 C Development Reagent B PV4126-80 C Development Buffer P3127 20-30 C FRAP1 (mtor) Kinase Buffer User prepared see http://www.lifetechnologies.com/selectscreen

08 Jan 15 Page 15 of 18 Materials Required But Not Supplied 384-well assay plates. We recommend black Corning 384-well, low-volume, round-bottom (non-binding surface) assay plates (Corning #4514). Other blackwalled, low-binding assay plates, while not tested, may be suitable. Multi-channel pipettor (12 channels) or any pipetting device that can accurately deliver repeated volumes of 2.5 µl and 5 µl. 96-well microplates that can accommodate 300 µl per well PI-103 (CAS 371935-74-9), MW 348.4 DMSO Procedure 1. Prepare initial 100X serial dilution curves in Row A of a 96-well plate: Dilute PI-103 to a 100X initial concentration in 100% DMSO (500 µm). Prepare a set of 3-fold serial dilutions from the initial concentration in a 96-well plate: Add 30 µl DMSO in wells A2 - A10. Add 45 µl of 500 µm of PI-103 in A1. Add 15 µl from well A1 to well A2, and then mix well A2, and take 15 µl from well A2 and add to well A3, mix, and so on. Figure 1: Schematic of initial compound dilution. PI-103 was titrated from a 500 µm starting concentration in the initial dilution series by preparing a 3-fold dilution in 100% DMSO. A secondary dilution to 4X was then prepared in the rows below the initial dilution curve (lighter gray) using kinase buffer. 2. Prepare a 4X Intermediate Dilution of PI-103 in Row B: Add 48 µl of 1X Kinase Buffer A into all wells B1 B10. Add 2 µl of diluted inhibitor from the wells in

08 Jan 15 Page 16 of 18 Row A to wells in Row B. This will produce a final serial dilution starting at 20 µm (B1), which will then produce a final assay concentration starting at 5 µm. 3. Prepare an Assay Plate: Prepare duplicate wells. Add 2.5 μl of the 4X compound dilutions per well into a low volume NBS, 384-well plate (Corning Cat. # 4514)., Using row 1, Columns 1 and 2 are replicates of the highest concentration, Columns 3 and 4 are replicates of the second highest concentration, etc. to column 20. 4. Add 2.5 μl of 1X Kinase Buffer A alone to Columns 21 and 22 (0% inhibition, no compound control), 23 (0% phosphorylation control, no kinase added) and 24 (Phosphopeptide 100% phosphorylation positive control) 5. Add 5 μl of the 2X Peptide/Kinase Mixture (prepared in mtor kinase buffer) to Row 1 Columns 1-22. DO NOT ADD TO COLUMN 23 OR 24. 0% phosphorylation control: Add 5 μl of 2 μm substrate alone without kinase to column 23, rows A-L. 100% phosphorylation control: Add 5 µl of 2 µm phosphopeptide control substrate to column 24, rows A-L. Buffer-only reference: Add 5 µl Kinase Buffer A alone to the remaining 8 wells (columns 23 and 24, rows M-P). 6. Add 2.5 μl of 4X ATP Solution (200 μm) per well to all columns to start reaction. 7. Shake assay plate on a plate shaker for 30 seconds. 8. Incubate assay plate for 60 minutes at room temperature. 9. Add 5 μl of the Development Reagent Solution to each well. Use the lot-specific dilutions indicated on your Certificate of Analysis as dilutions may vary based upon Z'-LYTE peptide and Development Reagent A lot. 10. Shake plate again on a plate shaker for 30 seconds. 11. Incubate for 60 minutes at room temperature. 12. Read and analyze as directed in the protocol.

08 Jan 15 Page 17 of 18 A B C D E F G H I J K L M N O P 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Figure 2: Assay Plate Schematic. Compound titrations, duplicate 10-point titration shown in row A columns 1-20 with known inhibitor (e.g. PI-103). Columns 21 and 22 prepared without any inhibitor as kinase activity controls (purple), column 23 prepared with no kinase, 0% phosphorylation, (blue) and column 24 prepared using phosphopeptide control, 100% phosphorylation, (green). Note 8 wells in bottom right (yellow M23 P24), were prepared without any inhibitor or substrates, as buffer controls. White wells can be used for the assay test compounds.

Emission Ratio Z -LYTE Compatible Microplate Reader 08 Jan 15 Page 18 of 18 D. Results 7.5 7.0 6.5 6.0 5.5 5.0 4.5 4.0 3.5 PI-103 IC50 = 49 nm 3.0 0.1 1 10 100 1000 10000 [PI-103] (nm) Figure 1: Z'-LYTE Kinase Assay. Z -LYTE Assay using Ser/Thr 11 (PV3670) with FRAP1 (mtor) (PV4753) performed using the BMG LABTECH CLARIOstar.