Practical Quadrupole Theory: Peak Shapes at Various Ion Energies

Similar documents
Extrel Application Note

Design considerations for linear Paul trap mass spectrometer under development

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

Most commercial quadrupole mass spectrometers

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Asymmetrical features of mass spectral peaks produced by quadrupole mass filters

Principles of Quadrupole and Ion Trap Mass Analyzers

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

FOCUS: QUADRUPOLE ION TRAPS ACCOUNT AND PERSPECTIVE

First described by W. Paul and H. Steinwedel in

Quadrupole Miniaturization: Revisited

Field-Free Atmospheric Ionization: Atmospheric Pressure Chemical Ionization & Electrospray

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

Introduction to GC/MS

Types of Analyzers: Quadrupole: mass filter -part1

Quadrupole mass filter operation under the influence of magnetic field

ICPMS Doherty Lecture 1

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

RESEARCH ARTICLE. QMS in the Third Stability Zone with a Transverse Magnetic Field Applied

The Analyzer. We will deal mainly with magnetic sector instruments here although we will talk briefly about other separation methods later.

Introduction to the Q Trap LC/MS/MS System

Quadrupole Storage Mass Spectrometry

Physics of and in Ion Traps

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap

Mass Spectrometry. What is Mass Spectrometry?

Quattro Micro - How does it work?

Analysis of Polar Metabolites using Mass Spectrometry

The development of algebraic methods to compute

Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT)

TANDEM MASS SPECTROSCOPY

Ion traps. Quadrupole (3D) traps. Linear traps

Three dimensional Paul traps may benefit from. Linear Quadrupoles with Added Hexapole Fields

Lecture 8: Mass Spectrometry

Chemistry Instrumental Analysis Lecture 35. Chem 4631

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results

Lecture 8: Mass Spectrometry

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)

Estimating a pseudounitary operator for velocity-stack inversion

Ion Trajectories in the Injection Quadrupole for SIFT-MS

Secondary Ion Mass Spectroscopy (SIMS)

Jumbo Cross Beam Ionizer

Insertion Devices Lecture 2 Wigglers and Undulators. Jim Clarke ASTeC Daresbury Laboratory

Mass Spectrometry in MCAL

Ions trapped within a linear quadrupole ion trap can

Understanding the Quadrupole Mass Filter through Computer Simulation

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading:

5. LIGHT MICROSCOPY Abbe s theory of imaging

Chip-Scale Mass Spectrometers for Portable Gas Analyzers Luis Fernando Velásquez-García. A. I. Akinwande, K. Cheung, and L.-Y Chen.

SUPPLEMENTARY INFORMATION

Harris: Quantitative Chemical Analysis, Eight Edition

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Filters and Tuned Amplifiers

Space Charge Effects in Linear. Quadrupole Ion Traps

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

THE HEAVY ION BEAM PROBE

How an Induction Motor Works by Equations (and Physics)

MS/MS .LQGVRI0606([SHULPHQWV

MS Goals and Applications. MS Goals and Applications

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

Magnetic Measurements

Spatial Frequency and Transfer Function. columns of atoms, where the electrostatic potential is higher than in vacuum

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

1. The range of frequencies that a measurement is sensitive to is called the frequency

Measurement of S-Parameters. Transfer of the Reference Plane. Power Waves. Graphic Representation of Waves in Circuits

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

CHAPTER D3 TOF ION OPTICS

Status of the SHIPTRAP Project: A Capture and Storage Facility for Heavy Radionuclides from SHIP

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Notes on the point spread function and resolution for projection lens/corf. 22 April 2009 Dr. Raymond Browning

Photoelectric Effect

PIC/MCC Simulation of Radio Frequency Hollow Cathode Discharge in Nitrogen

High-Resolution. Transmission. Electron Microscopy

Two-electron systems

All of us engaged in science face exciting opportunities

Beam Diagnostics Lecture 3. Measuring Complex Accelerator Parameters Uli Raich CERN AB-BI

Tendenze nell innovazione della strumentazione in spettrometria di massa:

Beam Diagnostics. Measuring Complex Accelerator Parameters Uli Raich CERN BE-BI

MS Goals and Applications. MS Goals and Applications

The Franck-Hertz Experiment Physics 2150 Experiment No. 9 University of Colorado

ICP-MS. High Resolution ICP-MS.

A Model of Evolutionary Dynamics with Quasiperiodic Forcing

FAIMS Technology at a Glance

Using a Microwave Interferometer to Measure Plasma Density Mentor: Prof. W. Gekelman. P. Pribyl (UCLA)

Quadrupole Mass Spectrometry Concepts. Mass spectrometers for residual gas analysis: Intermediate Level Users Guide

Speculations About a Fourier Series Kicker for the TESLA Damping Ring

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer

Section 1 Simple Harmonic Motion. Chapter 11. Preview. Objectives Hooke s Law Sample Problem Simple Harmonic Motion The Simple Pendulum

Hiden EQP Systems. High Sensistivity Mass and Energy Analyzers for Monitoring, Control and Characterization of Ions, Neutrals and Radicals in Plasma.

arxiv:physics/ v1 [physics.plasm-ph] 5 Nov 2004

MAGNETIC FLUX LEAKAGE INVESTIGATION OF INTERACTING DEFECTS: COMPETITIVE EFFECTS OF STRESS CONCENTRATION AND MAGNETIC SHIELDING

Chapter 11: WinTDR Algorithms

1. Types of Waves. There are three main types of waves:

Born simulation report

Time of Flight Mass Spectrometry of Ions Generated by Molecules in Intense Laser Fields

Transcription:

Practical Quadrupole Theory: Peak Shapes at Various Ion Energies Randall E. Pedder ABB Inc., Analytical-QMS Extrel Quadrupole Mass Spectrometry, 575 Epsilon Drive, Pittsburgh, PA 15238 (Poster presented at the 50th ASMS Conference on Mass Spectrometry and Allied Topics, Orlando, Florida, June 2 to 6, 2002) At the 2001 ASMS conference, a graphical introduction to quadrupole theory was presented. (1) The solutions of the equations of motion for ions traversing through a quadrupole were illustrated, with simple X-Y plots used to illuminate the practical implications of the complex motion described by solving the Mathieu equation, specifically on resolution and mass calibration. This presentation builds upon that graphical approach, using graphical depictions of calculated ion trajectories to illustrate the interaction of ion energy and frequency of ion motion, as well as the effect of a restricted exit aperture on experimentally measured peak shapes. Practical examples of peak shape defects and their source are considered graphically, along with a simple theoretical explanation. The goal of this work is to help build an intuitive understanding of how quadrupoles work, and more importantly, how to optimize performance in experiments involving quadrupoles. Calculated ion trajectories are used to illustrate stability and instability (i.e. what happens if the ion going through the quadrupole is too light, too heavy or just right, based on the applied quadrupole voltages), as well as the consequences of too much ion energy (ions scream through the quadrupole even though they have theoretically unstable trajectories, resulting in poor abundance sensitivity, i.e. poor resolution at the baseline). Peak shape defects due to excessive ion energy are well correlated to simulated ion trajectories. Quasi-stable trajectories of fast moving ions, which should be theoretically unstable, illustrate well the cause of poor abundance sensitivity in such cases. I. INTRODUCTION This presentation was created to address the following questions: What are some of the parameters that impact quadrupole mass peak shape? Why do quadrupole mass peaks tend to lift off the baseline on the low mass side of the peak more so than the high mass side of the peak? Why do mass peaks tend to have notches or splits when ion energy is increased, especially if the quadrupole is equipped with a restricted exit aperture? An ion moving in an RF-only or RF/DC quadrupole field will exhibit periodic motion, with the frequencies and their relative magnitude easily calculated based on operating conditions. The equations, which are used to describe this periodic motion, appear deceptively simple, but in practice, the correlation between the equations of motion and their calculated solutions is rich in its interpretation, if not obscure. Used by permission from Extrel CMS, L.P. Page 1 of 8

The purpose of this presentation is to de-mystify the theory associated with how quadrupoles operate, and to provide some practical application of theoretical quadrupole performance to experimentally measured peak shapes, specifically related to effects of ion energy on peak shapes, and the effect of a restricted exit aperture on peak shapes. More thorough treatments of quadrupole theory can be found in Peter Dawson s book Quadrupole Mass Spectrometry and its Applications. (2), and March and Hughes book Quadrupole Storage Mass Spectrometry. (3) Analytical solutions to the Mathieu equation were calculated via successive approximation of continued fractions to yield a graph of the Mathieu stability diagram. Runge-Kutta numerical integration of the Mathieu equation was used to generate simulated ion trajectories under various conditions. Quadrupole mass peak shapes of Nitrogen (m/z 28) were measured with various ion energies and were correlated graphically to the analytical and numerical integration results, using a quadrupole constructed from 9.5 mm diameter rods operated at an RF frequency of 880 khz, with a smaller than normal exit aperture (~3 mm diameter instead of the standard Extrel aperture of 7.5 mm). This presentation approach is unique to any yet found in the literature, with its focus on practical implications of quadrupole theory, de-emphasizing the complex abstract equations typically utilized in traditional summaries of quadrupole theory. II. THE EXPERIMENT In a quadrupole mass spectrometer system, ions are somehow generated, focused into a quadrupole as close as possible to the centerline of the quadrupole axis, and are filtered while traveling through the quadrupole, with the resulting filtered ions traveling through an exit aperture to a detector or another analyzer. See Figure 1. 'y' Direction 'x' Direction Inscribed Diameter Ions 'z' Direction Filtered Ions Restricted Aperture: ~1/3 Inscribed Diameter Note that in this work, the exit aperture is restricted compared to the ideal in order to exaggerate the effects of ion energy noding. Peak profiles are measured by scanning the RF and DC voltages along a scan line of appropriate ratio through the Mathieu stability diagram (i.e with approximately 6:1 RF to DC ratio). Ion energy through the quadrupole ( z direction) is calculated as the potential difference between the centerline offset potential of the quadrupole (pole bias) and the potential at which the ion was born (or had its last collision in the case of a high pressure source like a chemical ionization or electrospray source). In this presentation, the x and y directions are identified as along the axes that connect the respective quadrupole pairs. Note that the assignment of one direction as x and the other as y is in practice somewhat arbitrary, and is really determined by the resolving DC polarity applied to the rod pairs. Often, an aperture is placed at the exit of a quadrupole, either to shield a detector, or to serve as a conductance limit for a collision cell. Ion trajectories were calculated via Runge-Kutta numerical integration of the Mathieu equation using a Turbo Basic program originally developed by the author at the University of Florida. Iso-beta lines were calculated via successive approximation of the continued fraction solutions of the Mathieu equation. (4) In this presentation, the system modeled and used experimentally had the following characteristics: Rod diameter: ~9.5 mm Inscribed diameter (distance between opposite rods): ~ 8.3 mm Aperture of Lens at exit of quadrupole: ~3 mm (Standard Extrel exit aperture for 9.5 mm quadrupole: 7.5 mm) Quadrupole length: ~20 cm RF frequency: 880 khz Nitrogen ions at m/z 28 III. SOLVING THE EQUATIONS The traditional treatment of quadrupole theory starts with a derivation of the Mathieu equation from F=ma all the way through to the final parameterized form, with the following parametric substitutions: Figure 1. Schematic of quadrupole assembly identifying x, y, and z directions, inscribed diameter, and ion flow. 2 d u + ( a 2q cos 2 ) u = 0 2 u u ξ dξ 8eU = mr 4eV = mr a u q 2 2 u 2 2 0 Ω 0 Ω Used by permission from Extrel CMS, L.P. Page 2 of 8

The u in the above equations represents position along the coordinate axes (x or y), is a parameter representing t/2, t is time, e is the charge on an electron, U is applied DC voltage, V is the applied zeroto-peak RF voltage, m is the mass of the ion, r is the effective radius between electrodes, and is the applied RF frequency. The rigorous analytical solution to this second order linear differential equation is: ( ) = Γ C2n exp(2n + β ) iξ + Γ' c2n n= n= u ξ exp (2n + β ) iξ Which, intuitively obvious to any person skilled in the art, reduces to a similar infinite sum of sine and cosine functions: u( ξ ) = n= ω n Where A n is the amplitude and ω n can be calculated as (n- β/2)*ω. In practice, the amplitudes for the center four frequencies of the infinite series (n=0, +1, -1, and +2) are large compared with the others, allowing us to truncate the solution as an approximation to: u ω 0 For example, consider the motion in the x and y directions for the position a=0.2306 and q=0.7001 in the first region of the Mathieu stability diagram. One could calculate β via successive approximation of the continued fraction solution as described by McLachlen (4) based on a and q. See Figure 2. In this case, β x = 0.918, yielding a trajectory in the x direction that represents a beat pattern with large amplitudes at frequencies of 404 khz and 476 khz, with smaller amplitudes at frequencies of 1.284 MHz and 1.356 MHz. Similarly, β y = 0.0573, yielding a trajectory in the y direction that features a large amplitude fundamental frequency of ion motion at 25 khz, with slower amplitude harmonics at 855kHz and 905 khz, with a tiny harmonic at 1.735 MHz (not shown in figure 2). IV. MATHIEU STABILITY DIAGRAM The traditional way to view the solutions of the Mathieu equation is to plot the calculated β values on A n sin + A cos ω + n n= ω n ω = A A A A ( ξ ) cos + cos + cos + cos 0 + 1 1 1 1 + 2 + 2 ω 'x' Direction 'y' Direction 0 10 20 30 40 Time (microseconds) BetaX = 0.918.404.476 1.284 1.356 BetaY = 0.0573.025.855.905 1 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 Fast Fourier Transform (MHz) Figure 2. Ion trajectories with Fast Fourier Transform Spectra for a point on the stability diagram where a=0.2306 and q=0.7001 for m/z 28 with 2.5 ev ion energy. (Point C on Figure 5) one plot, connecting the points in a,q space that have the same β values (iso-β lines). See figures 3, 4 and 5 The most important of these iso-β lines are the ones that define the stability boundaries themselves, namely β x = 0 and β x = 1, and β y = 0 and β y = 1. It is interesting to note that the βx = 0 and βy = 0 boundaries which define the low mass side of the mass peak (left hand side of the stability diagram) are boundaries associated with low frequencies of ion motion (low frequency boundaries), while the β x = 1 and β y = 1 boundaries which define the high mass side of a mass peak (right hand side of the stability diagram) are high 0.2 0.1 0.0-0.1-0.2 Low Frequency Boundary in the 'y' Direction (BetaY = 0) Low Frequency Boundary in the 'x' Direction. (BetaX = 0) High Frequency Boundary in the 'x' Direction (BetaX = 1) High Frequency Boundary in the 'y' Direction (BetaY = 1) 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 Figure 3. Mathieu Stability Diagram showing the first region of stability, which is two overlapping regions, one for the x direction, and one for the y direction. Used by permission from Extrel CMS, L.P. Page 3 of 8

0.240 0.235 0.230 Iso-Beta Lines for 'X' Direction 0.90 0.92 0.94 0.96 1.0 0.98 Scan Lines for Various Resolutions R =1000 0.0 0.02 R=200 0.04 0.06 Iso-Beta Lines for 'Y' Direction 0.08 R=44 0.10 0.12 0.240 0.235 0.230 Iso-Beta Lines for 'X' Direction 0.90 A B 0.92 0.94 C 0.96 1.0 0.98 0.0 0.02 Iso-Beta Lines 0.04 for 'Y' Direction 0.06 D E 0.08 R=44 0.10 0.12 0.2 0.1 0.0-0.1-0.2 0.69 0.70 0.71 0.1 0.1 Iso-Beta Lines for 'X' Direction 0.2 0.2 0.3 0.3 0.4 0.4 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 frequency boundaries. By simply selecting an operating point in RF-DC space that is contained within this area bound by integer iso-beta lines (or one of the other regions of stability) will result in a theoretically stable trajectory, while operation outside these regions of stability will yield theoretically unstable trajectories. Scanning the quadrupole voltages along a scan line through the stability diagram generates a mass peak. Figure 5 shows such a scan line, while figure 6 illustrates what the peak shape would look like for such a scan line. V. SCANNING ACROSS MASS PEAK 0.5 Iso-Beta Lines for 'Y' Direction When you scan through the stability diagram to generate a mass peak, you are sweeping the amplitude of the RF and DC voltages through the stability region for a given ion. This, in turn, changes the frequencies 0.5 0.6 0.6 0.7 0.7 0.9 1.0 0.8 0.8 0.91.0 Zoom in on Tip of Stability Diagram! Scan Line R = 44 Figure 4. Mathieu Stability Diagram showing Iso-Beta lines, with zoom in view of tip, showing scan lines of various resolutions. 0.69 0.70 0.71 Figure 5. Scan line through the tip of the stability diagram that would yield slightly better than unit mass resolution (R=44) on mass 28. of ion motion for ions of that mass through the quadrupole as can be calculated by solution of the Mathieu equation. Figure 5 shows a stability diagram with a Resolution = 44 scan line, with five points identified as A, B, C, D, and E. Point A corresponds to voltages in which the ion is stable in the x direction, but unstable in the y direction and gets ejected in the y direction. This corresponds to the low frequency of ion motion boundary on the low mass side of the peak. Note that the fundamental frequency trends to zero hertz and that the first and second harmonics converge toward the RF drive frequency as this boundary is approached! Point E corresponds to voltages in which the ion is stable in the y direction, but unstable in the x x200 x1 B A C D E 27.0 28.0 29.0 m/z Figure 6. Experimental mass scan across mass 28 illustrating the relative intensities at the vabrious points along the scan line through the stability diagram shown in Figure 5.b Used by permission from Extrel CMS, L.P. Page 4 of 8

A) 'y' Unstable A) 'y' Unstable in 'y' direction! Beta = 0 Beta = 0.883 B) 0.880 B).389.491 1.269 1.371 Beta = 0.0197 Beta = 0.887 C).0087.871.889 1.751 C).390.490 1.270 1.370 Beta = 0.0573 Beta = 0.918 D).025.855.905 1.735 D).404.476 1.284 1.356 Beta = 0.0804 Beta = 0.978 E) 'x' Unstable.035.845.915 1.725 E) 'x' Unstable.430.450 1.310 1.330 Beta = 0.0823 Beta = 1 0 10 20 30 40 Time (microseconds).036.844.916 1.724 Fast Fourier Transform (MHz) 0 10 20 30 40 Time (microseconds) SPLAT! Ion hit Electrode in 'x' Direction!.440 1.320 Fast Fourier Transform (MHz) Figure 7. Trajectories of ions in the Y direction for various points along the scan line shown in Figure 5, with Fast Fourier Transforms showing frequencies of ion motion. Point A corresponds to the ion crossing the Low Frequency Boundary in the Y direction (low mass side of peak). Figure 8. Trajectories of ions in the X direction for various points along the scan line shown in Figure 5, with Fast Fourier Transforms showing frequencies of ion motion. Point E corresponds to the ion crossing the High Frequency Boundary in the X direction (high mass side of peak). direction and gets ejected in the x direction. This corresponds to the high brequency of ion motion boundary on the high mass side of the peak. Note that the Fundamental frequency of ion motion converges with the first harmonic at one half the RF drive frequency as this boundary is approached. From Figure 5, it can be seen that the different points along the scan line correspond to different sets of βx and βy values. It follows that an ion would have different frequencies of ion motion at different points along the scan line, as shown in figures 7 and 8. When ion energy is increased, the transit time through the quadrupole is decreased, effectively reducing the number of RF cycles (and hence the number of cycles of ion motion) that the ion sees as it transits through the quadrupole. Figure 9 shows this phenomenon, where the trajectory is stretched or compressed, depending on whether the ion energy is increased or decreased. 0.5 ev 1 ev 2.5 ev 10 ev Distance (cm) Figure 9. Calculated ion trajectories for an ion traveling through a quadrupole in the Y direction at various ion energies (point C on figure 5.) The red dotted lines correspond to the quadrupole boundaries. The x axis is distance along the quadrupole, not time. Used by permission from Extrel CMS, L.P. Page 5 of 8

0.5 ev in 'y' Direction 0.5 ev in 'x' Direction 1 ev in 'y' Direction 1 ev in 'x' Direction 2.5 ev in 'y' Direction 2.5 ev in 'x' Direction 10 ev 10 ev Unstable Ion Gets Through Quadrupole!!! in 'x' Direction Distance Through Quadrupole (cm) Figure 10. Calculated ion trajectories through the quadrupole in the y direction for various ion energies, corresponding to point A from Figure 5, the low mass side boundary of the mass peak. Dotted red lines correspond to quadrupole boundaries. VI. ION ENERGY EFFECTS ON ABUNDANCE SENSITIVITY Abundance Sensitivity is a figure of merit that describes the ability of a mass spectrometer to discriminate between adjacent masses of widely different intensity. This is effectively a description of the resolution measured at the baseline. How much of the signal that is measured at mass M is really attributable to M+1 or M-1? How much do the masses leak into their neighbor s intensity measurement? From figure 6, the m/z 28 peak looks well resolved in the x1 trace, but when magnified by 200, it can be seen that there is barely a separation between m/z 27 and m/z 28, while there is almost baseline separation between m/z 28 and m/z 29. This resolution at the baseline is indeed a measure of the abundance sensitivity, which is typically identified as a fraction of the height of a given peak that appears one amu away from that mass peak. In this case the abundance sensitivity would be estimated at a few thousand. Quadrupoles operated under normal conditions generally exhibit poorer abundance sensitivity on the Distance Through Quadrupole (cm) Figure 11. Calculated ion trajectories through the quadrupole in the x direction for various ion energies, corresponding to point E from Figure 5, the high mass side boundary of the mass peak. Dotted red lines correspond to quadrupole boundaries. low mass side of the peak than the high mass side of the peak (typically one order of magnitude difference). This difference in performance can be attributed to the fact that the low mass side of the peak corresponds to a fuzzy low frequency boundary, while the high mass side of the peak corresponds to a sharper high frequency boundary. Figures 10 and 11 illustrate the effects of varying ion energy at either of the low frequency boundary (Point A in Figure 5, illustrated in Figure 10), or the high frequency boundary (Point E in Figure 5, illustrated in Figure 11). It is striking that on the low frequency boundary, ions can have quasi-stable trajectories and get through the quadrupole, if the energy is sufficiently large, that they do not get a full cycle of ion motion, whereas we see more efficient rejection on the high mass side of the peak. This leads to the baseline liftoff that is more pronounced on the low mass side of a quadrupole mass peak. VII. ION ENERGY PEAK SPLITTING When a quadrupole has either an off-axis detector or a restricted exit aperture, it is possible to see nodes Used by permission from Extrel CMS, L.P. Page 6 of 8

1) 6) 0.240 0.235 0.230 10) 0.0 0.02 0.04 0.06 0.08 Iso-Beta Lines for 'Y' Direction 0.10 0.12 0.14 R=44 R=29 2) 7) 0.225 1) 0.16 3) 8) 0.220 0.68 0.69 0.70 0.71 0.72 Figure 12. Stability diagram with only y iso-beta lines identified. Points along R=29 scan line correspond with the peaks and valleys in Figure 13, as well as the trajectories in Figure 14. 4) 9) 8 9 3 4 7 10 2 5 6 5) 10) 5 ev Ion Energy 1 15 ev Ion Energy Points on R=29 Scan Line Distance (cm) Distance (cm) Figure 14. Ion trajectories in the y direction for points along the scan line in Figure 12. Local maxima in the 15 ev spectrum shown in Figure 13 correspond to trajectories where the ion ended up in the middle of the quadrupole. The local minimum corresponds to trajectories in which the ion ends up far from the center of the quadrupole. 27.0 28.0 29.0 m/z Figure 13. Unit mass resolution peak at two different ion energies. appear in the quadrupole peak profile. Such peak splitting can be directly attributed to the periodic nature of ion motion through a quadrupole, and is generally sensitive to variation in ion energy. (i.e. if you change the ion energy, the relative location of the splits in the peak profile will change.) A smaller than normal exit aperture (~3 mm instead of 7.5 mm) was inserted at the end of a quadrupole to exaggerate this effect for this presentation. The quadrupole was tuned to nominally unit mass resolution (resolution = 29 for m/z 28), and ion energy was varied between 5 ev and 15 ev. The scan line through the stability diagram is shown in Figure 12, with various points marked with large red dots. Only iso-β y lines are shown, since peak splitting is attributed to low frequency y motion. An experimentally collected mass peak profile is shown in Figure 13, with red dots marking the points correlating to the points on the scan line shown in Figure 12. Note that the β y values change dramatically along the left hand side of the stability diagram, but more gradually as you move to the right hand side of the stability diagram. (the lines get further apart). This means that the frequency of ion motion changes much more dramatically on the low mass side of the peak, leading to more nodes on the low mass side of the peak, than on the high. Figure 14 shows the trajectories that correspond to the ten red dots identified in Figures 12 and 13. Note the correlation between relative maxima and having the ions end up at the center of the exit of the quadrupole, and the correlation between the relative minimum of the mass peak corresponding to ions that end up nearer to the poles of the quad. The noding is a direct result of efficiency of transferring ions through the quadrupole exit lens. With a restrictive exit aperture, some of the stable ions do not reach the detector. Used by permission from Extrel CMS, L.P. Page 7 of 8

Abundance sensitivity is compromised with increased ion energy. Ions of the wrong mass can indeed travel through the quadrupole to the detector, with worse effects visible for too-heavy ions contributing to the ion signal of the target mass on the low mass side of the mass peak. This effect is evidenced by the exhaggerated liftoff of the low mass side of the 15 ev ion energy peak profile in figure 13, as compared to the 5 ev peak profile in the same figure. VIII. CONCLUSIONS The purpose of this presentation is to help demystify the theory associated with how quadrupoles operate, specifically with regards to the effects of variation of ion energy on peak shape. Using this graphical approach, quadrupole operation can be understood intuitively without extensive study of the equations of motion. Ion Motion in a quadrupole is periodic in nature, with abundance sensitivity on the low frequency ion motion / low mass side of the mass peak not quite as good as the abundance sensitivity on the high frequency ion motion / high mass side of the mass peak. The ion motion at the stability boundary limits correspond to the convergence of the first and second harmonic to the RF drive frequency in the case of the low mass side of the peak / low frequency boundary, and the convergence of the fundamental and first harmonic to one half the drive frequency in the case of the high mass side of the peak / high frequency boundary. Abundance sensitivity is the ultimate determinant of quadrupole resolution, with resolution at the baseline dictating the limit of resolution at half height. While resolution at half height remains virtually unchanged as ion energy is adjusted, the resolution at the baseline (abundance sensitivity) is very strongly dependent upon ion energy. Resolving power is ultimately dictated by the number of cycles of ion motion that the ion sees as it traverses the quadrupole. Increasing ion energy reduces the number of cycles of ion motion, and hence reduces resolution and abundance sensitivity. Increasing rod length or RF frequency increases the number of cycles of ion motion and hence increases resolution and abundance sensitivity. For the case of either restricted exit aperture or offaxis detector, too high an ion energy (or too narrow a range of ion energy) results in peak splitting or nodes in the mass peak, especially noticeable on the low-mass side of the mass peak. IX. REFERENCES 1. Pedder R.E. Practical Quadrupole Theory: Graphical Theory, Extrel Application Note RA_2010, Poster presented at the 49th ASMS Conference on Mass Spectrometry and Allied Topics, Chicago, IL, 2001. 2. March, R.E., and Hughes, R.J. Quadrupole Storage Mass Spectrometry, Wiley Interscience, New York, 1989. Chapter 2: Theory of Quadrupole Mass Spectrometry, Pages 31-110. 3. Dawson, P.H. Quadrupole Mass Spectrometry and its Applications, Elsevier, Amsterdam, 1976. 4. McLachlan, N.W. Theory and Applications of Mathieu Functions, Clarendon: Oxford, 1947. Used by permission from Extrel CMS, L.P. Page 8 of 8