Relative Sensitivity RS Measurements of Gases

Similar documents
Hiden CATLAB Systems Microreactor for Catalysis Studies & Thermal Analysis

Quadrupole Mass Spectrometry Concepts. Mass spectrometers for residual gas analysis: Intermediate Level Users Guide

Ashwani Gupta. Mb: Class IX-X: X: Math & Science Class XI-XII: XII: Accts., Eco. & B. Stds. Carbon and its compounds.

Mass Spectrometer A Comparison of Positive and Negative Ion RGA Methods

vacuum analysis plasma diagnostics surface science gas analysis

4.6 Evaluating spectra

Hiden HPR60 Molecular Beam Mass Spectrometer System

The Claviature of Gas Analysis

Pressure & temperature sampling options for Hiden gas analysis systems

Pellistor Relative Responses

Polymer plants continue to seek ways to increase production and efficiency without compromising safety.

Elements, Mixtures and Compounds

Hiden SIMS Secondary Ion Mass Spectrometers. Analysers for surface, elemental and molecular analysis

CHEMISTRY MOLES PACKET 2017 NAME: PER:

All organic compounds contain carbon, however, not all carbon containing compounds are classified as organic. Organic compounds covalently bonded

Practice Packet Unit 11: Organic Chemistry

Slide 1 / A gas at a pressure of 10.0 Pa exerts a force of N on an area of 5.5 m 2 A 55 B 0.55 C 5.5 D 1.8 E 18

Chapter 13: Partial Pressure Analysis

The stoichiometry of burning hydrocarbon fuels

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

Appendix A Physical and Critical Properties

PROPERTY TABLES AND CHARTS (SI UNITS)

CHEMISTRY & YOU What is the difference between the oxygen you breathe and the oxygen in ozone in the atmosphere?

Methane contains atoms of two elements, combined chemically. Methane is a mixture of two different elements.

FIB - SIMS. Focussed Ion Beam Secondary Ion Mass Spectrometry.

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns

8. Application Programs

Carbon and its Compounds

Candidate Name Centre Number Candidate Number. The Language of Chemistry, Structure of Matter and Simple Reactions

The Spectroscopy of Stars

Charles D Winters/Science Photo Library. How can you tell that a chemical reaction is taking place? ...

Cracking. 191 minutes. 186 marks. Page 1 of 27

Extrel Application Note

Chapter 8 Covalent Bonding

2 Examiner SECTION A. Answer all the questions in the spaces provided.

Which of the following is an element? A Water B Methane C Fluorine D Ammonia

2. Hydrocarbons. 2.1 Composition of Petroleum

Review of investigation of influence of operational parameters on metrological characteristics of QMS within EMRP IND12 project

WP3: In-situ chemical, physical and optical properties of aerosols. WP4: Trace gases networking: Volatile organic carbon and nitrogen oxides

ammonia carbon dioxide hydrogen nitrogen electrical heat solar sound (a) In air, the two most common gases are oxygen and...

Changes in Matter. Introduction to Chemistry

An ion source performs the following two functions:

ChemActivity L2: Mass Spectrometry

Nomenclature. 133 minutes. 130 marks. Page 1 of 22

Chemistry: A Molecular Approach, 2e (Tro) Chapter 2 Atoms and Elements. Multiple Choice Questions

Combustible Gas Catalytic Bead (0-100 %LEL) Part No FM Performance Certified 1,4 FM Performance Certified 1

1 Which of the compounds shown are in the same homologous series? 1 CH 3 OH 2 CH 3 CH 2 OH 3 CH 3 COOH C 3 CH 2 CH 2 OH

PhysicsAndMathsTutor.com. Advanced Subsidiary Unit 1: The Core Principles of Chemistry

CHAPTER 24 Organic Chemistry

Application of the EQP and EQS mass spectrometers to time resolved studies in ion beam and plasma processes

Name: Date: Period: #: BONDING & INTERMOLECULAR FORCES

Balancing chemical reaction equations (stoichiometry)

GCSE SCIENCE A CHEMISTRY

MASS and INFRA RED SPECTROSCOPY

3 - Atomic Absorption Spectroscopy

Level 1 Chemistry, 2014

Hiden Isochema. Gravimetric Gas & Vapor Sorption Analyzers. Hiden Isochema IGA Series. Advancing Sorption Analysis

5.4 Molecular Compounds

CHAPTER 2 Building Blocks of Materials

Gases CHAPTER. Section 10.1 Properties of Gases

Determination of Minimum Detectable Partial Pressure (MDPP) of QMS and its Uncertainty

Chemistry 11. Unit 8 Atoms and the Periodic Table Part II Electronic Structure of Atoms

Monomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

CHAPTER 5 GASES AND THE KINETIC- MOLECULAR THEORY

Thermo Scientific Prima PRO Process Mass Spectrometer

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

Unit 2, Lesson 01: Introduction to Organic Chemistry and Hydrocarbons

Organic Nomenclature

Chapter 8 Covalent Bonding

AppNote 7/1994. Friedhelm Rogies, Andreas Hoffmann Gerstel GmbH & Co.KG, Aktienstrasse , D Mülheim an der Ruhr, Germany

Principles of Chemistry: A Molecular Approach (Tro) Chapter 2 Atoms and Elements

Activity # 2. Name. Date due. Assignment on Atomic Structure

Chapter 20. Mass Spectroscopy

2011, Robert Ayton. All rights reserved.

Periodic Classification of Elements

National 5 Chemistry. Unit 2: Nature s Chemistry. Topic 1 Hydrocarbons

a) ion-ion attractions b) London dispersion forces c) hydrogen bonding forces d) dipole-dipole attractions

Practice questions for Ch. 3

Far UV Absorbance Detector

An extension of the group contribution method for estimating thermodynamic and transport properties. Part IV. Noble gas mixtures with polyatomic gases

Organic Chemistry. REACTIONS Grade 12 Physical Science Mrs KL Faling

Introduction to GC/MS

Prima PRO Process Mass Spectrometer

Chemistry of Life 9/16/15. Chemistry s Building Block: The Atom. Ch 2 BIOL 160

The School For Excellence 2018 Unit 3 & 4 Chemistry Topic Notes Page 1

Organic Chemistry. A brief introduction

Satellites, Weather and Climate Module 33: Atmospheric sciences and the mathematics common core standards. Dr. Janel Hanrahan

Elements and Reactivity Revision Notes

Science For Class IX Atoms and Molecules

2 Copyright Pearson Education, Inc., or its affiliates. All Rights Reserved.

INTRODUCTORY CHEMISTRY Concepts and Critical Thinking Seventh Edition by Charles H. Corwin

THE DETERMINATION OF CRITICAL FLOW FACTORS FOR NATURAL GAS MIXTURES

Q1. Which one of the following is least likely to occur in the reaction between methane and chlorine?

Principles of Chemistry: A Molecular Approach 2e (Tro) Chapter 2 Atoms and Elements

Name: Class: Date: 10. How many neutrons does europium-151 have? What is the isotope s mass number? Number of neutrons = mass number atomic number

Supported molybdenum oxides as effective catalysts for the catalytic fast pyrolysis of lignocellulosic biomass

Organic Chemistry Worksheets

5.3 ORGANIC COMPOUNDS

CHAPTER 2: Atoms, Molecules and Stoichiometry

Chapter 2 Atoms and the Periodic Table

Transcription:

Gas Analysis Relative Sensitivity RS Measurements of Gases Introduction This note describes the main factors that influence the relative sensitivity factor (RSF) of a mass spectrometer and describes how to estimate the RSF for a given molecular species. Mass Spectrometer Figure 1 shows the major components of a quadrupole mass spectrometer. Typically the gas sample is admitted via an inlet system to the ion source, where it is converted to ions by electron-impact. The ions pass through a mass analyser, and then to a detector. A general introduction may be found in [1]. Manufactured in England by: HIDEN ANALYTICAL LTD 420 Europa Boulevard, Warrington, WA5 7UN, England t: +44 (0) 1925 445225 f: +44 (0) 1925 416518 e: info@hiden.co.uk w: www.hidenanalytical.com

S ample Inlet Ion Source Mass Analyser Detector Data system Figure 1 Mass Spectrometer Schematic The sample is typically a gas. It will probably be at a pressure much higher than the operating pressure of the mass spectrometer, so the inlet system will incorporate pressure-reducing means, such as an orifice, a capillary or a leak valve. Fractionation in the inlet system can affect the performance of the instrument; this can be minimised by correct inlet design [2]. In the ion source, energetic electrons (70 ev) bombard the sample, forming positive ions. Often the sample will break into smaller fragment ions. This gives rise to the cracking pattern (or fragmentation pattern ) of ions formed from a particular molecule. The cracking pattern is influenced quite strongly by the kinetic energy of the ionising electrons, but there will also be variation from one design of ion source to another, and also variation due to factors such as mechanical tolerances and surface condition. Nevertheless, standardised cracking patterns will usually provide a good guide to performance. Another factor is the transmission of the mass analyser. For a quadrupole, the transmission tends to reduce at higher masses. Usually the transmission can be regarded as effectively constant in the mid-mass region, say from 20 to 50 amu. It is likely to be significantly higher for the lowest masses - perhaps by a factor of two for H 2 at 2 amu and He at 4 amu. At higher mass a power law such as M -1 or M -2 is sometimes quoted, but this should be applied with caution because there is considerable variation from one instrument to another. 2 For a Faraday detector the sensitivity does not vary significantly with mass, but the sensitivity of an electron multiplier reduces as the mass increases, typically as M -1. Relative sensitivity factor Evidently the instrument sensitivity depends on many factors. We can express this as: RSF = R I. R S. R F. R Q. R D (1) Where, RSF is the overall relative sensitivity factor, assumed to be 1.0 for nitrogen R I is the inlet sensitivity factor R S is the source sensitivity factor R F is the fragmentation factor R Q is the quadrupole transmission is the detection efficiency R D The inlet sensitivity factor R I is impossible to predict with any confidence on theoretical grounds. It is perhaps best regarded as constant when attempting to predict the RSF unless specific information is available, based on measurements on the individual instrument or on the known generic performance of the design. The source sensitivity factor R S and the fragmentation factor R F can be estimated from generic data with a fair degree of confidence. These factors are discussed in more detail below. The quadrupole transmission R Q and the detection efficiency R D are strongly influenced by the instrument design so any assumptions made should be used with caution. In the absence of more

Relative sensitivity factor R R Quadrupole transmisison (typical) 10.0 1.0 specific information, Figure 2 shows the variation in relative transmission to be expected for a typical small quadrupole. ionisation gauges rather than mass spectrometers, and therefore will not necessarily incorporate all the factors mentioned above. Nevertheless, there are useful references, and the spread of results reported from several different sources is not so great as to invalidate the use of generic data. It seems reasonable to assume that typical RSF values can be used in the absence of more definitive calibration checks. Figure 3 and Table 1 show averaged RSF values from Leck [3], where data is tabulated from several studies published in the literature [4-6]. These values may be taken to represent RS in equation 1 above. 0.1 0 10 20 30 40 50 60 70 80 90 100 Mass (amu) Figure 2: Typical transmission for a small quadrupole mass filter. 8 7 6 5 4 3 2 Noble gases Hydrocarbons Other RSF in the literature There is little definitive data in the academic literature relating to RSF for mass spectrometers as such, but there are publications describing RSF for ionisation gauges. There is much in common between the two, because the electron-impact ion source resembles a small extractor ionisation gauge and the basic physical processes that influence ion formation are very similar. Over the years, many manufacturers have listed RSF data for mass spectrometers in their technical literature, and today one can readily find information on the Internet. It is often not clear whether the data is based on actual measurements on the instruments in question or whether (as is perhaps more likely) the data has been taken from earlier work. Furthermore, it seems likely that the data often originates from measurements on 3 1 0 0 20 40 60 80 100 120 Number of electrons per molecule N e Figure 3 :Relative Sensitivity Factor vs No of Electrons.

Helium 0.18 Neon 0.32 Argon 1.36 Krypton 1.92 Xenon 2.82 Hydrogen 0.43 Deuterium 0.40 Ammonia 1.18 Water 1.11 Carbon monoxide 1.01 Nitrogen 1.00 Oxygen 0.85 Hydrogen sulphide 1.82 Carbon dioxide 1.38 Nitrous oxide 1.20 Sulphur hexafluoride 2.35 Mercury 3.30 Methane 1.53 Ethane 2.65 Propane 3.92 Butane 4.42 Pentane 5.06 Hexane 5.90 Heptane 6.94 Iso-octane 7.21 Acetylene 1.66 Ethylene 2.11 Propene 3.04 Butene 3.60 Hexene 5.81 Benzene 4.74 Table 1: Relative sensitivity factors R S for ionisation gauge Note that along the x-axis of Figure 3 is plotted the number of electrons per molecule. This has been found to give a reasonable correlation with measured data, and suggests plausibly that the probability of ionisation taking place is proportional to the number of electrons in the sample molecule. Clearly there are other factors at work as well, but the correlation seems quite convincing for the noble gases and for hydrocarbons. The Other points fall above the general trend are for ammonia and water (at N e = 10) and for hydrogen sulphide (N e 4 = 20). Many other common gases fall close to the yellow trend line. These values are a good guide to the variation in sensitivity that can be expected from the ionisation process, and thus can be used as an estimate of the source sensitivity factor R S. For many common gases the fragmentation factor can be ignored because the base peak corresponds to most of the ion current. For larger molecules with a richer mass spectrum (eg hydrocarbons) this is no longer true, and one needs to make allowance for fragmentation. For example, propane has a base peak at mass 29, but it corresponds to only about 30.2% of the total ion current. Therefore when estimating the RSF we would take the figure for ion gauges (3.92) and multiply it by 0.302, giving us an overall RSF of 1.27. To some extent, the increased sensitivity from the electron configuration is compensated by the reduction due to fragmentation. Even in this case where there is significant fragmentation we would introduce relatively little error if we simply assumed the RSF to be 1.0. In practice, it is often appropriate to ignore the variation is RSF altogether. For the best precision there is no substitute for performing routine calibration check, preferably using a gas mixture whose composition is close to the composition expected from the sample. Measured RSF Table 2 shows RSF values from measurements made at on a few common gases on a QIC system. Data is shown for a mixture of 2% He, Ne, Kr in N 2. Ar was not included because of the interference between Ne + and Ar ++ at mass 20

RSF Measured Conc. Literature average and range H 2 /N 2 1.01 2.5% 0.46 (0.33 0.70) He/N 2 0.27 2% 0.16 (0.13-0.25) Ne/N 2 0.25 2% 0.26 (0.22-0.30) Ar/N 2 1.39 50% 1.24 (1.11 1.42) CO 2 /N 2 1.00 50% 1.36 (0.90 1.54) Kr/N 2 1.47 2% 1.76 (1.00-1.95) Table 2 Measured RSF values For Ne and Kr the measured RSF is quite close to the literature average. The result for He is high, but is only just outside the range of the literature, and is probably influenced by increased quadrupole transmission at low mass. Also tabulated are measurements of H 2 /N 2, Ar/N 2 and CO 2 /N 2 using mixtures set up using a gas blender. The measured RSF for H 2 is somewhat higher than the spread of values found in the literature, again probably because of the quadrupole transmission. Estimated RSF It is preferable to measure the RSF, but this may not be feasible. In such a case we can estimate the RSF for a particular species as follows. We find the value for R S from Table 1, and multiply by factors R F to allow for the fragmentation and R Q to allow for quadrupole transmission. (Figure 2). The values for R F are derived from standard tabulations such as [7] or (He, Kr, Xe) from isotopic abundances. The values are normalised to the value for N 2. Data for common gases are listed in Table 3. 5

Base peak R S R F R Q RSF Helium 4 0.18 1.00 2.00 0.39 Neon 20 0.32 0.90 1.00 0.31 Argon 40 1.36 0.88 1.00 1.27 Krypton 84 1.92 0.57 0.52 0.61 Xenon 132 2.82 0.27 0.33 0.27 Hydrogen 2 0.43 0.98 2.00 0.84 Deuterium 4 0.40 0.99 2.00 0.79 Ammonia 17 1.18 0.53 1.00 0.65 Water 18 1.11 0.81 1.00 0.95 Carbon monoxide 28 1.01 0.92 1.00 0.98 Nitrogen 28 1.00 0.94 1.00 1.00 Oxygen 32 0.85 0.96 1.00 0.86 Hydrogen sulphide 34 1.82 0.50 1.00 0.97 Carbon dioxide 44 1.38 0.80 1.00 1.17 Nitrous oxide 30 1.20 0.89 1.00 1.13 Sulphur hexafluoride 127 2.35 0.58 0.35 0.50 Mercury 202 3.30 0.30 0.22 0.23 Methane 16 1.53 0.46 1.00 0.75 Ethane 28 2.65 0.48 1.00 1.35 Propane 29 3.64 0.32 1.00 1.23 Butane 43 4.42 0.40 1.02 1.90 Pentane 43 5.06 0.33 1.02 1.82 Hexane 57 5.90 0.20 0.77 0.99 Heptane 43 6.94 0.24 1.02 1.82 Iso-octane 57 7.21 0.43 0.77 2.54 Acetylene 26 1.66 0.73 1.00 1.28 Ethylene 28 2.11 0.46 1.00 1.02 Propene 41 3.04 0.32 1.00 1.03 Butene 41 3.60 0.36 1.07 1.49 Pentene 55 0.00 0.34 0.80 0.00 Hexene 41 5.81 0.17 1.07 1.15 Benzene 78 4.74 0.49 0.56 1.38 Table 3: Estimated RSF In Figure 4 we show the measured values from Table 2 together with estimated values from Table 3. There is reasonable agreement except for Kr, where the measured value is approximately 50% higher than the estimate. 6

RSF Measured RSF Estimated RSF 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0 0 10 20 30 40 50 60 70 80 90 100 Mass (amu) Figure 4: Measured and estimated RSF Recommendations Undoubtedly if good precision is required, a regular calibration should be carried out, preferably using a gas mixture whose major constituents are representative of to the expected sample composition. Failing this, an estimate can be made, as described above. There are many possible sources of variation, but nevertheless such an estimate can is usually good to within a factor of two of the actual value determined by in-situ measurement. References 1. R Schubert, Partial pressure analysers, analysis and applications, AVS Monograph M-19 (2000). 2. JH Batey, Vacuum, 44, 639-642 (1993). 3. JH Leck, Total and Partial Pressure Measurement in Vacuum Systems, Blackie (1989). 4. R Holanda, J. Vac. Sci. Technol., 10, 1133-1139 (1973). 5. F Nakao, Vacuum, 25, 431-435 (1975). 6. JE Bartmess & SM Georgiadis, Vacuum, 33, 149-153 (1989). 7. Eight Peak Index of Mass Spectra, Royal Society of Chemistry (1991). 7