Luminescence free counting of 3 H facilitated by Hidex 300 SL/600 SL TDCR triple coincidence counters. Risto Juvonen

Similar documents
Hidex 300 SL Automatic TDCR Liquid Scintillation Counter

Hidex LSCs. LSC 2013 Barcelona

The measurement of tritium in environmental water samples

1/5 Specifications Hidex 300 SLL automatic TDCR liquid scintillation counter

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS

LIQUID SCINTILLATION COUNTERS. {Beta Counters}

Hidex Liquid Scintillation Counters

Technical Basis for Use of a Correlated Neutron Source in the Uranium Neutron Coincidence Collar

Chapter 14. The Concept of Equilibrium and the Equilibrium Constant. We have for the most part depicted reactions as going one way.

Implementation of ISO/IEC in a low level liquid s cintillation tritium laboratory

QCLAS Sensor for Purity Monitoring in Medical Gas Supply Lines

Geochemical dating of a Swiss freshwater limestone cave using 230 Th/ 234 U ingrow and 226 Ra-excess decay chronometry

Comparison of Different Methods of Environmental Radioactivity Measurements at the Zagreb Radiocarbon and Tritium Laboratory

EYAL'C- \TION OF TELEMETRY RECORDS RECEIVED

Hefei, Anhui, , P.R. CHINA b Institute of Management, University of Science and Technology of China, Beijing,

Heat exchangers: Heat exchanger types:

9 Geophysics and Radio-Astronomy: VLBI VeryLongBaseInterferometry

An ultrasonic air temperature measurement system with self-correction function for humidity

Centroid Detection by Gaussian Pattern Matching In Adaptive Optics

Liquid Scintillation Counter

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion

Chapter 2 Lecture 5 Longitudinal stick fixed static stability and control 2 Topics

Atomic and Nuclear Physics

Atomic and Nuclear Physics

IvanaJakonić 1, Nataša Todorović 2, JovanaNikolov 2, Ines KrajcarBronić 3, BranislavaTenjović 4, MiroslavVesković 2

Determination of Neutron Beam Diameter in 3 rd Horizontal Channel of Dalat Nuclear Reactor Nguyen AS 1*, Dang L 1 and Ho HT 2 1

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

An Adaptive Optimization Approach to Active Cancellation of Repeated Transient Vibration Disturbances

Centroid Detection by Gaussian Pattern Matching in Adaptive Optics

RESOLUTION OF DIFFERENCES IN CONCENTRATION OF NATURALLY OCCURRING TRITIUM IN GROUNDWATER TRACER STUDIES

Power-speed Trade-off in Parallel Prefix Circuits

IMPEDANCE EFFECTS OF LEFT TURNERS FROM THE MAJOR STREET AT A TWSC INTERSECTION

V. A. TSAREV Lebedev Physical Institute of the Academy of Sciences of the USSR Moscow, USSR

TRITIUM DETERMINATION IN WATER

arxiv:hep-ph/ v1 13 Dec 2001

E γ. Electromagnetic Radiation -- Photons. 2. Mechanisms. a. Photoelectric Effect: photon disappears. b. Compton Scattering: photon scatters

MODELING STUDY OF PROPOSED FIELD CALIBRATION SOURCE USING K-40 SOURCE AND HIGH-Z TARGETS FOR SODIUM IODIDE DETECTOR. A Thesis JEREMY LAYNE ROGERS

TECHNETIUM-99 IN WATER

Comparison of Deadtime Correction Factors for Passive Neutron Multiplicity Counting of Correlated and Non-Correlated Neutron Sources

Supplementary Material Comparative assessment of the ELBA coarse-grained model for water

Practical Approaches using TDCR Measurements and Alpha/Beta Separation

A model for measurement of the states in a coupled-dot qubit

D* meson production in diffractive DIS

Čerenkov counting and liquid scintillation counting of 36 Cl

METU Chem. Eng. Dept. Ch.E. 320 Chem. Eng. Lab I EXPERIMENT 33 MOLECULAR WEIGHT DETERMINATION OF POLYMERS BY CAPILLARY FLOW METHOD

Procedure for determining the activity concentrations of lead-210 and polonium-210 in drinking water and ground water

Discussion: Flux correction for cross-contamination. 1. Before measuring: practical considerations for open paths

EFFECT OF PITCH NUMBER IN OVERALL HEAT TRANSFER RATE IN DOUBLE PIPE HELICAL HEAT EXCHANGER

A Spatiotemporal Approach to Passive Sound Source Localization

TECHNETIUM-99 IN SOIL

Bureau International des Poids et Mesures

PhysicsAndMathsTutor.com 1

RESULTS YOU CAN COUNT ON

Detection of 222 Rn and 226 Ra in environmental samples by scintillation method

arxiv: v2 [quant-ph] 7 Feb 2012

"Research Note" ANALYSIS AND OPTIMIZATION OF A FISSION CHAMBER DETECTOR USING MCNP4C AND SRIM MONTE CARLO CODES *

Speed-feedback Direct-drive Control of a Low-speed Transverse Flux-type Motor with Large Number of Poles for Ship Propulsion

SIMPLEX OPTIMIZATION BY ADVANCE PREDICTION IN INSTRUMENTAL NEUTRON ACTIVATION ANALYSIS.

Controller Design Based on Transient Response Criteria. Chapter 12 1

A Proof-of-Principle Investigation for a Neutron-Gamma Discrimination Technique in a Semiconductor Neutron Detector THESIS

22.01 Fall 2015, Problem Set 6 (Normal Version Solutions)

Chapter 8 Thermodynamic Relations

LIQUID SCINTILLATION

TDCR in a nutshell. P. Cassette, Laboratoire National Henri Becquerel, France

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

A Longitudinal Aerodynamic Data Repeatability Study for a Commercial Transport Model Test in the National Transonic Facility

Effect of Different Types of Promoters on Bed Expansion in a Gas-Solid Fluidized Bed with Varying Distributor Open Areas

Numerical modeling of the thermoelectric cooler with a complementary equation for heat circulation in air gaps

Monte Carlo Simulation of Electron and Radiative Emission from Silicon Diodes

Minimizing new tritium line sensitivity by a quality process implemented at Madagascar-I.N.S.T.N

An Integrated Architecture of Adaptive Neural Network Control for Dynamic Systems

Evaluation of effect of blade internal modes on sensitivity of Advanced LIGO

DETERMINATION AND MITIGATION OF PRECIPITATION EFFECTS ON PORTAL MONITOR GAMMA BACKGROUND LEVELS. A Thesis STEPHEN MICHAEL REVIS

Variation Based Online Travel Time Prediction Using Clustered Neural Networks

Assessing the Performance of a BCI: A Task-Oriented Approach

Lecture 6 Design of ESP

General Equilibrium. What happens to cause a reaction to come to equilibrium?

ABSTRACT. SUN, JING. Optimization of the Neutron Absorption Steady Source Measurement

Effects of Vane Sweep on Fan-Wake/Outlet-Guide-Vane Interaction Broadband Noise

Position sensitive detector technology, data reduction and count rates

Final Exam: know your section, bring your ID!

Comparison of Alternative Equivalent Circuits of Induction Motor with Real Machine Data

DETERMINATION OF RADON BY LIQUID SCINTILLATION a/f PARTICLE SPECTROMETRY: Towards the Resolution of a 14C Dating Problem

SHIELDING MATERIALS FOR HIGH-ENERGY NEUTRONS

TIME-RESOLVED LIQUID SCINTILLATION COUNTING

ABSOLUTE NEUTRON FLUX MEASUREMENTS USING AN NE110 SCINTILLATION COUNTER

Analytical Chemistry Detection Limits and the Evaluation of Dredged Sediment

STATIC SIMULATION PROGRAM OF COPPER SOLVENT EXTRACTION CONFIGURATIONS USING MICROSOFT EXCEL SOLVER

Planning with Uncertainty in Position: an Optimal Planner

MultiPhysics Analysis of Trapped Field in Multi-Layer YBCO Plates

UTC. Engineering 329. Proportional Controller Design. Speed System. John Beverly. Green Team. John Beverly Keith Skiles John Barker.

Supplementary information for: All-optical signal processing using dynamic Brillouin gratings

Copyright 2018 Society of Photo-Optical Instrumentation Engineers (SPIE). One print or electronic copy may be made for personal use only.

CHARACTERIZATION OF WOUND MONITORING SYSTEMS USED TO QUANTIFY AND LOCATE PLUTONIUM CONTAMINATION. A Thesis PAUL JAMES DIMMERLING

Size Exclusion Chromatography and Nuclear Magnetic Resonance Techniques for Predicting Asphalt Yields and Viscosities from Crude Oil Analyses

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua

Making sense of Limits of Detection (with respect to Tritium analysis) James Thomson Meridian Biotechnologies Ltd.

Case I: 2 users In case of 2 users, the probability of error for user 1 was earlier derived to be 2 A1

A Differential Equation for Specific Catchment Area

Transcription:

Luminesene free ounting of 3 H failitated by Hidex 300 SL/600 SL TDCR triple oinidene ounters Risto Juvonen LSC 2017 Copenhagen, 4.5.2017

Hidex 300 SL and 600 SL Automati TDCR Liquid Sintillation Counters Automati vial ounters with triple- PMT detetor failitating: Exeptionally high ounting effiieny, Absolute ativity outing of beta isotopes (H-3, C-14,.. ) and Cherenkov isotopes (Y-90, Sr-89,..) using TDCR method without external radioative standard soure Luminesene free ounting mode a/b separation with 2D graphial alibration & validation tool

Hidex 300 SL and 600 SL Automati TDCR Liquid Sintillation Counters Automati vial ounters with triple- PMT detetor failitating: Exeptionally high ounting effiieny, Absolute ativity outing of beta isotopes (H-3, C-14,.. ) and Cherenkov isotopes (Y-90, Sr-89,..) using TDCR method without external radioative standard soure Luminesene free ounting mode a/b separation with 2D graphial alibration & validation tool

Luminesene interferene Luminesene is one of the main interferenes in sintillation ounting Single photon event inreasing the bakground and inreases results variation

Luminesene interferene Methods to redue and orreting luminesene are (1) Dark adaptation of sample (2) Chemial methods (3) Temperature ontrol (4) Counting region settings (5) Delayed oinidene ounting Many labs are ounting samples also with repeats, and removing outliers - How to find outlier? - Do I have to reount the sample or an I alulate the results using the data of remaining repeats?

Luminesene Free Counting method (LF method) for 3 H in water Method is based on ounting region (ROI) settings With double oinidene ounters the onventional method based on ounting regions (ROIs) works relatively well for higher energy isotopes like 14 C but yields in too low ounting effiieny for 3 H water samples. Improved LF method was tested for 3 H water samples and requires triple oinidene ounter like Hidex 300 SL or 600 SL.

7 Triple detetor array and oinidenes R S T: Triple Coinidene Counts S R Sample T R S: Double Coinidene Counts R T: Double Coinidene Counts S T: Double Coinidene Counts R S T R S R T S T C t = triple ounts C d = double ounts

Luminesene do not interfere with triple ounts 14 C in 3ml 2M NaOH + 17 ml OptiPhase HiSafe 3 3 H bkg spetra, 8 ml H 2 O + 12 ml UG ullt All ounts (blue), with luminesene peak Triple oiniodene ounts (red), no luminesene interferene In Luminesene Free mode (LF mode) only Luminesene Free Counts are reorded: 1) All Triple Coinidene ounts 2) Pure Double Coinidene ounts at above luminesene ROI (> h 120) Improved eff. Allows also 3 H water measurements in LF mode

Performane omparison of Lum Free mode and Normal ount mode for 3 H in water Hidex 300 SL 425-020, 8 + 12 ml sample, 10 x 50 min, ka = kb = 1.645 Normal Mode LF Mode Bkg 4.7 CPM 2.8 CPM Eff. 37 % 24 % Ld 2.6 Bq/L 1.2 DPM 3.1 Bq/L 1.5 DPM Outliers High risk Negligible risk

IAEA TRIC2012 3 H Refene samples in LF mode Hidex SL300 test for low-level tritium. I AEA TRIC2012 referene samples were measured, pre-distilled by ion-exhange, 500 ml eletrolyti enrihments, deuterium and spike method were ompared to determine 3H enrihment fators, full error propagation applied, data was proessed using IAEA TRIMS software.

IAEA TRIC2012 3 H Refene samples in LF mode Sample Hidex 300 SL, LF mode TRIC Ref. labs TRIC Top lab 8 ml + 12 ml UGuLLT, 20 x 50 min 2000 min Deuterium Spike TU ( 2 H) Un. TU (spike) Un. TU (known) Un. TU (known) Un. TRIC-20 0 0 0 0 0 0,10 0,02 0,01 TRIC-21 0,26 0,16 0,28 0,17 0,43 0,10 0,44 0,02 TRIC-22 1,13 0,18 1,17 0,18 1,12 0,10 1,12 0,15 TRIC-23 2,71 0,20 2,81 0,20 2,74 0,07 2,70 0,04 TRIC-24 4,25 0,11 4,45 0,21 4,37 0,13 4,45 0,06 TRIC-25 7,57 0,29 7,61 0,25 7,51 0,23 7,45 0,09 *Data provided by IAEA Hydrogeology lab, Vienna Hidex 300 SL an be used to produe aurate data on LF mode for low level 3 H water samples.

12 Quenh Corretion in LF mode TDCR annot be used diretly in LF mode as luminesene interferes with double oinidene ounts ( d ) TDCR = d t + t = t all Availabel methods: 1. Constant quenh orretion fator (preferred when quenhing is onstant) 2. External std urve method 3. Coretion of TDCR with Chemi CPM TDCR = d t Chemi + t = t Chemi all 4. Triple Coinidene Channel Ratio method - Ratio of Triple Coinidenes in two different region used as quenh parameter => (troi1/troi2) vs. Eff. HIdex Oy / Risto Juvonen

13 Summary Luminesene do not interfere with triple oinidene ounts LF mode = all triples + pure doubles above luminesene ROI Improved eff. allows also 3 H water measurements in LF mode In LF mode the Eff. and Bkg are redued, Ld only slightly higher Total ounting time is redued as in LF mode ounting an be started more or less immediately after sample preparation Negligible risk for outliers In LF mode the preferred quenh orretion method is the use of Constant Quenh Corretion Fator (other methods are appliable) Hidex 300 SL and 600 SL provide aurate data in LF mode for low level 3 H water samples HIdex Oy / Risto Juvonen

Thank You! Risto Juvonen Produt manager risto.juvonen@hidex.om Lemminkäisenkatu 62 FI-20750 Turku info@hidex.om www.hidex.om Tel. +358 10 843 5570