RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA

Similar documents
ACCURATE QUANTIFICATION OF RADIOACTIVE MATERIALS BY X-RAY FLUORESCENCE: GALLIUM IN PLUTONIUM METAL

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES

Physics 3204 UNIT 3 Test Matter Energy Interface

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

USABILITY OF PORTABLE X-RAY SPECTROMETER FOR DISCRIMINATION OF VALENCE STATES

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

ANALYSIS OF LOW MASS ABSORPTION MATERIALS USING GLANCING INCIDENCE X-RAY DIFFRACTION

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

Overview of X-Ray Fluorescence Analysis

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS

GLANCING INCIDENCE XRF FOR THE ANALYSIS OF EARLY CHINESE BRONZE MIRRORS

Atomic & Nuclear Physics

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

Air Filter Alpha Spectrometry Report

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE

Application Note. Abstract. Introduction. Experimental

SCINTILLATION DETECTORS AND PM TUBES

GAMMA RAY SPECTROSCOPY

THE IMPORTANCE OF THE SPECIMEN DISPLACEMENT CORRECTION IN RIETVELD PATTERN FITTING WITH SYMMETRIC REFLECTION-OPTICS DIFFRACTION DATA

ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES

A COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE

Atomic Notation (or Nuclear Symbol): Shorthand for keeping track of protons and neutrons in the nucleus

DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN THICK MATERIALS

Alpha decay usually occurs in heavy nuclei such as uranium or plutonium, and therefore is a major part of the radioactive fallout from a nuclear

Application of Spectral Summing to Indeterminate Suspect Low-Level Drums at Los Alamos National Laboratory

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U

Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn?

MCSHAPE: A MONTE CARLO CODE FOR SIMULATION OF POLARIZED PHOTON TRANSPORT

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS

Chapter 30 Nuclear Physics and Radioactivity

Radionuclide Imaging MII Positron Emission Tomography (PET)

S8 TIGER Series 2 for ASTM D 6443

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

Nuclear Chemistry. Nuclear Terminology

Radionuclide Imaging MII Detection of Nuclear Emission

Nuclear Physics. AP Physics B

The Use of Self-Induced XRF to Quantify the Pu Content in PWR Spent Nuclear Fuel

Nuclear Physics and Astrophysics

Name Date Class NUCLEAR RADIATION. alpha particle beta particle gamma ray

CHEM*3440. X-Ray Energies. Bremsstrahlung Radiation. X-ray Line Spectra. Chemical Instrumentation. X-Ray Spectroscopy. Topic 13

4 α or 4 2 He. Radioactivity. Exercise 9 Page 1. Illinois Central College CHEMISTRY 132 Laboratory Section:

PRACTICAL GAMMA SPECTROSCOPY ASSAY TECHNIQUES FOR LARGE VOLUME LOW-LEVEL WASTE BOXES

EDS User School. Principles of Electron Beam Microanalysis

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Lab 14. RADIOACTIVITY

Chapter 18: Radioactivity And Nuclear Transformation. Presented by Mingxiong Huang, Ph.D.,

PERFORMANCE OF A BOREHOLE XRF SPECTROMETER FOR PLANETARY EXPLORATION

CHARACTERIZATION OF Pu-CONTAINING PARTICLES BY X-RAY MICROFLUORESCENCE

PHYA5/1R (JUN15PHYA51R01) General Certificate of Education Advanced Level Examination June Unit 5 Nuclear and Thermal Physics Section A

Sample Examination Questions

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

RADIOACTIVITY. Nature of Radioactive Emissions

... (1) The diagram shows how aluminium sheet is rolled to form foil of constant thickness. rollers source of radiation

Radioactivity III: Measurement of Half Life.

X-Ray Emission and Absorption

Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol

Introduction to Ionizing Radiation

QUANTIFICATION OF MERCURY, LEAD, AND CADMIUM IN AQUEOUS SOLUTIONS BY ENERGY K X-RAY FLUORESCENCE SPECTROSCOPY

IMPROVING THE ACCURACY OF RIETVELD-DERIVED LATTICE PARAMETERS BY AN ORDER OF MAGNITUDE

NOTES: 25.2 Nuclear Stability and Radioactive Decay

Basic physics Questions

Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner, Johann Wernisch, and Michael Mantler

Characterization of a Portable Neutron Coincidence Counter Angela Thornton and W. Charlton Texas A&M University College Station, TX

Time-Resolved μ-xrf and Elemental Mapping of Biological Materials

Radioactivity and energy levels

Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods. Abstract

EEE4106Z Radiation Interactions & Detection

BORON IN GLASS DETERMINATION USING WDXRF

COMPARISON OF THREE UNIVERSAL CURVES FOR THE ESCAPE PROBABILITY OF X-RAY EXCITED ELECTRONS - I. THEORY

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

2. With the help of diagram explain the Edison s discovery of thermionic emission. Mention the observations and the conclusion.

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

Radiation Detection and Measurement

Fundamentals of Radionuclide Metrology

Level 2 Physics: Atoms & Radioactivity Types of Radioactivity - Answers

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter

PHYSICS A2 UNIT 2 SECTION 1: RADIOACTIVITY & NUCLEAR ENERGY

Radioactivity. Lecture 6 Detectors and Instrumentation

Application of Spectral Summing to Suspect Low Level Debris Drums at Los Alamos National Laboratory

An Analysis of Secondary Enhancement Effects in Quantitative XRFA

Radioactivity. The Nobel Prize in Physics 1903 for their work on radioactivity. Henri Becquerel Pierre Curie Marie Curie

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

Multi-element process analyzer

X-rays. X-ray Radiography - absorption is a function of Z and density. X-ray crystallography. X-ray spectrometry

Hour Exam 3 Review. Quantum Mechanics. Photoelectric effect summary. Photoelectric effect question. Compton scattering. Compton scattering question

Transcription:

90 RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA Christopher G. Worley Los Alamos National Laboratory, MS G740, Los Alamos, NM 87545 ABSTRACT Energy dispersive X-ray fluorescence (EDXRF) is a rapid, straightforward method to determine sample elemental composition. A spectrum can be collected in a few minutes or less, and elemental content can be determined easily if the detector has adequate energy resolution. Radioactive alpha emitters, however, emit X-rays during the alpha decay process that complicates spectral interpretation. This is particularly noticeable when using a portable instrument where the detector is located in close proximity to the instrument analysis window held against the specimen. A portable EDXRF instrument was used to collect spectra from specimens containing plutonium- 239 (a moderate alpha emitter) and americium-241 (a heavy alpha emitter). These specimens were also analyzed with a wavelength dispersive XRF (WDXRF) instrument to compare spectra collected from instruments with very different instrumental geometries. INTRODUCTION Pulse pile-up, escape peaks, and sample diffraction effects can complicate EDXRF spectra, but these artifacts are recognizable by an experienced analyst. Radioactive alpha emitters, however, emit X-rays due to internal processes that result from alpha decay, which further complicates spectral interpretation. (Gamma rays emitted by radioactive samples are indistinguishable to the detector from X-rays of the same energy, but gamma energies are normally too high to be of concern in EDXRF.) This alpha decay effect can be seen in plutonium-239 spectra which exhibit a prominent uranium L X-ray emission peak series. When a plutonium-239 atom alpha decays, there is some probability that the uranium-235 daughter nucleus will be left in an excited state that can then transfer its excess energy to an electron through internal conversion. 1 If the electron is then ejected upon absorption of this energy, an X-ray photon or Auger electron will be emitted. Hence, the uranium L X-ray series is present due to sample radioactive decay rather than source-induced X-ray fluorescence. Therefore, quantifying the uranium based on observed peak intensities can lead to erroneous results. Portable instruments are particularly susceptible to this effect due to the short distance between the specimen and the detector. In the present study, Pu-239 metal was analyzed with a portable EDXRF instrument. The same specimen was also examined by WDXRF, where the detector was much further from the specimen, and a diffraction crystal was located between the specimen and detector. This comparison demonstrated that using WDXRF is more advantageous for analyzing alpha emitters. With WDXRF a high power X-ray source is used to increase the number of fluorescence X-ray photons and counteract signal attenuation resulting from the relatively large distance between the specimen and detector. Because using higher source power

This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website www.dxcicdd.com ICDD Website - www.icdd.com

91 does not increase the alpha-induced X-ray emission intensity, more source-excited X-ray photons are detected in the WDXRF instrumental geometry than are alpha emission-excited X-rays. In another experiment, dissolution and ion exchange solid phase extraction were used to separate americium-241 from an old Pu-239 metal sample, and the americium eluate solution was analyzed with the portable EDXRF instrument. Am-241 is a strong alpha emitter. Thus, this study was used to compare the differences observed in EDXRF spectra collected from moderate alpha emitters such as Pu-239 versus heavy emitters like Am-241. EXPERIMENTAL Instrumentation. The portable EDXRF instrument was a Keymaster Technologies Tracer III-V containing a silver X-ray anode. The maximum operating voltage was 40 kv, and the maximum power was ~0.5W. A Ti/Al filter was used in front of the source to reduce background counts. The WDXRF instrument was a PANalytical PW2404 with a 4000 W rhodium anode. The data were collected while flushing the analysis chamber with helium. A LiF(220) crystal was used for wavelength separation. A 150 μm collimator was used in front of a scintillation detector, and a 200 μm aluminum tube filter was employed. A 30 mm source collimator mask was used. Plutonium metal. An ~0.5 g chunk of Pu-239 metal containing ~30 ppm uranium was sealed inside a 40 mm plastic XRF specimen cup with 7.5 μm Kapton. A 4 μm Prolene secondary film was placed over the primary film for analysis. The specimen was analyzed qualitatively with the portable EDXRF at high and low power settings (40 kv, 10 μa and 5 kv, 0.1 μa respectively) for 30 sec each. The specimen was also qualitatively scanned for 60 sec over the Pu Lα and U Lα energy range with the WDXRF instrument at high power (60 kv, 66 ma) and low power (20 kv, 10 ma) settings. Americium eluate. Approximately 0.5 g of an ~21 year old Pu-239 metal sample containing ~1820 ppm Am-241 was dissolved in 6 M HCl. The americium was separated from the plutonium using ion exchange solid phase extraction (Bio-Rad AG MP-1 200 to 400-mesh resin used), and the final solution was brought to 10 ml with 0.5 M HNO 3. An ~5 ml aliquot of the americium eluate (~90 ppm Am) was transferred to a 40 mm plastic specimen cup and sealed with 7.5 μm Kapton. A 4 μm Prolene secondary film was placed over the primary film for analysis. Qualitative spectra were collected from the solution for 30 sec each with the portable EDXRF instrument operated at 40 kv, 2.6 μa and then at 5 kv, 0.1 μa. RESULTS AND DISCUSSION Plutonium metal. The detector is located in close proximity to the specimen in portable EDXRF instruments in order to collect ample signal using a very low powered, battery operated X-ray source. This also results in the presence of substantial specimen radiation-induced X-ray peaks. Spectra collected from plutonium metal at low and high power are compared in figure 1. The low power setting was used to simulate collecting a spectrum with the source off. (The instrument will not collect data with the source actually turned off). Hence, only sample radiation-induced X-ray emission was detected in this spectrum.

92 Because Pu-239 alpha particle ejection and the subsequent internal conversion process results in the emission of uranium X-rays, a uranium L peak series was present in the plutonium spectrum. The sample only contained ~30 ppm uranium, but a marked uranium L peak series was present even when operating the source at 5 kv. This voltage was too low to induce uranium L fluorescence. Hence, the uranium peaks were due to the Pu-239 alpha decay and internal conversion process. Pu Lα 1,2 6000 40 kv, 10 μa 5000 5 kv, 0.1 μa (Alpha decay-induced XRF) Intensity (counts) 4000 3000 2000 U Lα 1,2 Pu Lβ 2 + U Lβ 1 Pu Lβ 1 1000 Pu Ll U Lβ 2 0 12 13 14 15 16 17 18 19 Energy (kev) Figure 1. EDXRF spectra from plutonium metal acquired at high and low power. When the same specimen was analyzed by WDXRF (figure 2), no uranium L peaks were observed. A high powered X-ray source is necessary in this type of instrument to collect adequate analyte signal due to attenuation by the large distance and diffraction crystal between the specimen and detector. Thus, the relatively weaker alpha decay-induced fluorescence signal is substantially attenuated in this geometry and is not detected. To simplify spectral analysis, WDXRF is therefore preferable to EDXRF for analyzing alpha emitters.

93 Figure 2. WDXRF spectra from plutonium metal acquired at high and low power. Americium eluate. As plutonium ages, any Pu-241 present beta decays to Am-241, which is a strong alpha emitter, which then decays to neptunium-237. A neptunium L XRF peak series will therefore be present in EDXRF spectra from americium-containing samples. Since the plutonium sample used to generate the americium specimen was ~21 years old, it contained a significant amount of americium (1820 ppm). The EDXRF spectra collected from the americium eluate at low and high power settings are essentially identical (figure 3). The low power spectrum is effectively the same as if the X-ray source were shut off. Therefore, almost all of the signal is due to Am-241 alpha decay to Np-237 with X-ray emission from internal conversion. The analyzed solution contained ~90 ppm americium and only 4 ppm neptunium, and the sourceexcited americium Lα peak (14.6 kev) was barely discernable above background in the spectrum collected at high power. However, the signal-to-background of the neptunium L peak series would lead one unfamiliar with analyzing radioactive samples to erroneously conclude that neptunium was a major component of this sample. Given that the source-excited americium peak is just above background, and marked neptunium peaks are present, the alpha decay and internal conversion XRF process must be very efficient compared with source-excited fluorescence. Thus, these results demonstrate that one must be very cautious in interpreting sample content or performing elemental quantification of alpha emitters using EDXRF.

94 4500 Np Lα Np Lβ 1 Intensity (counts) 4000 3500 3000 2500 2000 1500 Np Lβ 2 40 kv, 2.6 μa 5 kv, 0.1 μa 1000 500 Zn Kα Np Ll Am Lα Np Lγ 1 Am gamma peak 0 8 10 12 14 16 18 20 22 24 26 28 Energy (kev) Figure 3. EDXRF spectra from americium eluate solution acquired at high and low power. CONCLUSION Interpretation of EDXRF spectra is typically straightforward; however, spectra from alpha emitters are complicated by X-ray peaks from the radionuclide daughter element (atomic number Z-2). The stronger the alpha emitter, the more pronounced these peaks will be. Hence, the analyst must be careful when interpreting EDXRF spectra collected from radionuclides; peak intensities do not necessarily correlate with elemental concentration. As demonstrated here, portable EDXRF instruments are particularly affected by this phenomenon due to the close coupling of the specimen and the detector. EDXRF spectra collected from plutonium and americium specimens exhibited prominent X-ray peaks due to alpha decay daughter products. Analysis of an americium solution demonstrated that even when the americium content was relatively low, substantial neptunium daughter X-ray peaks were present. No alpha-induced uranium L peaks were seen in WDXRF spectra from the plutonium specimen. The instrument utilizes a high power X-ray source in order to maximize the number of fluorescence X-ray photons and collect adequate signal after attenuation from traversing the relatively large distance between the specimen and detector. However, use of a high power source does not enhance alpha decay-induced X-ray emission. Therefore, no alpha-induced uranium X-ray signal was detected, while substantial source-excited plutonium peaks were observed. Hence, WDXRF is more suitable than EDXRF for elemental quantification of alpha

95 emitters or trace analysis in the energy range where the alpha decay daughter X-ray emission peaks are located. ACKNOWLEDGMENTS I would like to thank Connie Soderberg for preparing the americium eluate solution. I am also grateful to Lav Tandon and Kevin Kuhn for providing the age of the plutonium sample used to prepare the americium specimen. REFERENCES [1] D. Reilly, N. Ensslin, and H. Smith, Jr. Passive Nondestructive Assay of Nuclear Materials, U.S. Government Printing Office: Washington, D.C., 1991.