Nuclear Medicine: Physics and Imaging Methods (SPECT and PET)

Similar documents
Nuclear Medicine: Physics and Imaging Methods (SPECT and PET)

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links

Radioisotopes and PET

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

Radionuclide Imaging MII Positron Emission Tomography (PET)

A. I, II, and III B. I C. I and II D. II and III E. I and III

Tomography is imaging by sections. 1

Year 12 Notes Radioactivity 1/5

6: Positron Emission Tomography

Nuclear Medicine RADIOPHARMACEUTICAL CHEMISTRY

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017

Chapter 18: Radioactivity And Nuclear Transformation. Presented by Mingxiong Huang, Ph.D.,


Some nuclei are unstable Become stable by ejecting excess energy and often a particle in the process Types of radiation particle - particle

CLINICALLY USEFUL RADIONUCLIDES:

Lecture Presentation. Chapter 21. Nuclear Chemistry. James F. Kirby Quinnipiac University Hamden, CT Pearson Education, Inc.

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Basic physics of nuclear medicine

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI)

69 Ga Ga

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras?

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter

Physics of Radiography

Structure of Biological Materials

Nuclear Radiation. Natural Radioactivity. A person working with radioisotopes wears protective clothing and gloves and stands behind a shield.

Outline Chapter 14 Nuclear Medicine

2007 Fall Nuc Med Physics Lectures

Physics of Radiography

Differentiating Chemical Reactions from Nuclear Reactions

Structure of the course

Properties of the nucleus. 8.2 Nuclear Physics. Isotopes. Stable Nuclei. Size of the nucleus. Size of the nucleus

Units and Definition

CHAPTER 7 TEST REVIEW

Chapter 11 Nuclear Chemistry

Medical Physics. Nuclear Medicine Principles and Applications

Nuclear Physics and Astrophysics

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging

General Physics (PHY 2140)

RADIOCHEMICAL METHODS OF ANALYSIS

Chapter 2 PET Imaging Basics

Radioactivity. The Nobel Prize in Physics 1903 for their work on radioactivity. Henri Becquerel Pierre Curie Marie Curie

AEPHY: Nuclear Physics Practise Test

Properties of the nucleus. 9.1 Nuclear Physics. Isotopes. Stable Nuclei. Size of the nucleus. Size of the nucleus

NOTES: 25.2 Nuclear Stability and Radioactive Decay

Objectives: Atomic Structure: The Basics

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

12/1/17 OUTLINE KEY POINTS ELEMENTS WITH UNSTABLE NUCLEI Radioisotopes and Nuclear Reactions 16.2 Biological Effects of Nuclear Radiation

Nuclear Physics Part 2A: Radioactive Decays

Dana-Farber Cancer Institute, 44 Binney Street, Boston, MA 02115, USA ramsey

Chapter 17. Radioactivity and Nuclear Chemistry

Basic science. Atomic structure. Electrons. The Rutherford-Bohr model of an atom. Electron shells. Types of Electrons. Describing an Atom

III. Proton-therapytherapy. Rome SB - 2/5 1

β and γ decays, Radiation Therapies and Diagnostic, Fusion and Fission Final Exam Surveys New material Example of β-decay Beta decay Y + e # Y'+e +

Nuclear Physics Part 2: Radioactive Decay

Nuclides with excess neutrons need to convert a neutron to a proton to move closer to the line of stability.

PHYSICS OF NUCLEAR MEDICINE

This Week. 3/23/2017 Physics 214 Summer

Chapter. Nuclear Chemistry

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

Name Date Class NUCLEAR RADIATION. alpha particle beta particle gamma ray

Chapter Three (Nuclear Radiation)

Radiochemistry and Radiopharmacy III

Nuclear Reactions A Z. Radioactivity, Spontaneous Decay: Nuclear Reaction, Induced Process: x + X Y + y + Q Q > 0. Exothermic Endothermic

Mitigation of External Radiation Exposures

Chap. 15 Radiation Imaging

Stable isotope. Relative atomic mass. Mole fraction 203 Tl Tl Thallium isotopes in Earth/planetary science

Alpha Decay. Decay alpha particles are monoenergetic. Nuclides with A>150 are unstable against alpha decay. E α = Q (1-4/A)

Nuclear Chemistry. Background Radiation. Three-fourths of all exposure to radiation comes from background radiation.

Chapter 20 Nuclear Chemistry. 1. Nuclear Reactions and Their Characteristics

Chapter 37. Nuclear Chemistry. Copyright (c) 2011 by Michael A. Janusa, PhD. All rights reserved.

Essentials of nuclear medicine

1.1 ALPHA DECAY 1.2 BETA MINUS DECAY 1.3 GAMMA EMISSION 1.4 ELECTRON CAPTURE/BETA PLUS DECAY 1.5 NEUTRON EMISSION 1.6 SPONTANEOUS FISSION

Introduction to Nuclear Reactor Physics

21.3 Radioactive Decay

Key Question: What role did the study of radioactivity play in learning more about atoms?

Physics 111 Homework Solutions Week #10 - Thursday

RADIOACTIVITY & HALF-LIFE Part 2

Introduction to Nuclear Engineering. Ahmad Al Khatibeh

Atomic Notation (or Nuclear Symbol): Shorthand for keeping track of protons and neutrons in the nucleus

Chemistry 52 Chapter 11 ATOMIC STRUCTURE. The general designation for an atom is shown below:

Chapter 30 Nuclear Physics and Radioactivity

Nuclear Chemistry - HW

Lecture 5: Tomographic nuclear systems: SPECT

Bases of radioisotope diagnostic methods

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital.

Nuclear Reactions Homework Unit 13 - Topic 4

Nuclear Chemistry AP Chemistry Lecture Outline

Introduction to Nuclear Physics and Nuclear Decay

The Case of Melting Ice

Physics in Nuclear Medicine

Chapter 3 Radioactivity

and have low penetrating power) Alpha particles are released through alpha decay. Beta Particles: An electron that comes from a nucleus through

Alta Chemistry CHAPTER 25. Nuclear Chemistry: Radiation, Radioactivity & its Applications

PS-21 First Spring Institute say : Teaching Physical Science. Radioactivity

This Week. 7/20/2016 Physics 214 Spring

Chapter IV: Radioactive decay

Nuclear Spectroscopy: Radioactivity and Half Life

ZX or X-A where X is chemical symbol of element. common unit: [unified mass unit = u] also known as [atomic mass unit = amu] or [Dalton = Da]

Radioactivity. Radioactivity

ELG7173 Topics in signal Processing II Computational Techniques in Medical Imaging

Transcription:

EL-GY 6813 / BE-GY 6203 / G16.4426 Medical Imaging Nuclear Medicine: Physics and Imaging Methods (SPECT and PET) Jonathan Mamou and Yao Wang Polytechnic School of Engineering New York University, Brooklyn, NY 11201 Based on Prince and Links, Medical Imaging Signals and Systems and Lecture Notes by Prince. Figures are from the book.

Nuclide Imaging Overview Lecture Outline Physics of Radioactive Decay Single Photon Emission Computed Tomography (SPECT) Positron Emission Tomography (PET) Image Quality consideration Resolution, noise, SNR, blurring Image Reconstruction Problems as Solving Linear Equations Least squares POCS Maximum likelihood Sparse reconstruction F16 Yao Wang @ NYU 2

What is Nuclear Medicine Also known as nuclide imaging Introduce radioactive substance into body Detect regional variations of radioactivity as indication of presence or absence of specific physiologic function Tumors may disrupt the normal decay process Blockage in a blood vessel may affect the distribution Detection by gamma camera or detector array From H. Graber, Lecture Note for BMI1, F05 F16 Yao Wang @ NYU 3

Examples: PET vs. CT X-ray projection and tomography: X-ray transmitted through a body from a outside source to a detector (transmission imaging) Measuring anatomic structure Nuclear medicine: Gamma rays emitted from within a body (emission imaging) Imaging of functional or metabolic contrasts (not anatomic) Brain perfusion Myocardial perfusion Tumor detection (metastases) From H. Graber, Lecture Note, F05 F16 Yao Wang @ NYU 4

Atomic Structure An atom={a nucleus, electrons} nucleons = {protons; neutrons} Nuclide: unique combination of protons and neutrons in a nucleus mass number A = # nucleons atomic number Z = # protons = # electrons An element is denoted by its A and Z Ex: 12 6 C or C -12 F16 Yao Wang @ NYU 5

Stable vs. Unstable Nuclides Stable nuclides: # neutrons ~= # protons (A ~= 2Z) when Z is small # neutrons > # protons when Z is large Unstable nuclides (radionuclides, radioactive atoms) Likely to undergo radioactive decay, which gives off energy and results in a more stable nucleus F16 Yao Wang @ NYU 6

Line of Stability Stability depends on ratio Z:N F16 Yao Wang @ NYU 7

Isotopes, etc Isotopes: atoms with the same Z but different A E.g. C-12 and C-11 Chemically identical Isobars: atoms with the same A but different Z Different elements Eg. Carbon-11 and boron-11 Isotones: atoms with the same number of neutrons but different A Isomers: atoms with the same Z and A but with different energy levels (produced after gamma decay) F16 Yao Wang @ NYU 8

What is Radioactivity? F16 Yao Wang @ NYU 9

Decay Modes F16 Yao Wang @ NYU 10

Alpha Decay Alpha decay: the nucleus emits a Helium-4 particle (alpha particle) Alpha decay occurs most often in massive nuclei that have a proton to neutron ratio too large. Alpha radiation reduces the ratio of protons to neutrons in the parent nucleus, bringing it to a more stable configuration. mostly occurring for parent with Z > 82 From: http://www.lbl.gov/abc/wallchart/chapters/03/1.html F16 Yao Wang @ NYU 11

Beta Decay Beta decay occurs when, in a nucleus with too many protons or too many neutrons, one of the protons or neutrons is transformed into the other. Mass number A does not change after decay, proton number Z increases or decreases. Beta minus decay (or simply Beta decay): A neutron changes into a proton, an electron (beta particle) and a antineutrino From: http://www.lbl.gov/abc/wallchart/chapters/03/2.html F16 Yao Wang @ NYU 12

Positron Decay Also known as Beta Plus decay A proton changes to a neutron, a positron (positive electron), and a neutrino Mass number A does not change, proton number Z is reduced From: http://www.lbl.gov/abc/wallchart/chapters/03/2.html F16 Yao Wang @ NYU 13

Mutual Annihilation after Positron Decay The positron later annihilate a free electron, generate two gamma photons in opposite directions The two photons each have energy 511 KeV, which is the energy equivalent to the rest mass of an electron or positron These gamma rays are used for medical imaging (Positron Emission Tomography), detected using a coincidence detection circuit F16 Yao Wang @ NYU 14

Gamma Decay (Isometric Transition) A nucleus (which is unstable) changes from a higher energy state to a lower energy state through the emission of electromagnetic radiation (photons) (called gamma rays). The daughter and parent atoms are isomers. The gamma photon is used in Single photon emission computed tomography (SPECT) Gamma rays have the same property as X-rays, but are generated differently: X-ray through energetic electron interactions Gamma-ray through isometric transition in nucleus From: http://www.lbl.gov/abc/wallchart/chapters/03/3.html F16 Yao Wang @ NYU 15

Measurement of Radioactivity Bq=Becquerel (1 decay/sec) Ci=Curie: (orig.: activity of 1 g of 226Ra) Naturally occurring radioisotopes discovered 1896 by Becquerel First artificial radioisotopes produced by Curie 1934 (32P) Inverse square law: The intensity of radiation incident on a detector at range r from a radioactive source is I AE = 4πr A: radioactivity of the material; E: energy of each photon 2 F16 Yao Wang @ NYU 16

Radioactive Decay Law N(t): the number of radioactive atoms at a given time A(t): is proportional to N(t) From above, we can derive dn A = = λn dt λ :decay constant N( t) = A( t) = The number of photons generated (=number of disintegrations) during time T is T T N = 0 A( t) dt = 0 λn 0 e N A e 0 λt 0 e dt λt λt = = λn N 0 0 e λt (1 e λt ) F16 Yao Wang @ NYU 17

Half-Life Half-life is the time it takes for the radioactivity to decrease by ½. F16 Yao Wang @ NYU 18

F16 Yao Wang @ NYU 19 Statistics of Decay The exponential decay law only gives the expected number (mean value) of atoms at a certain time t. The number of disintegrated atoms over a short time t <<T 1/2 after time t=0 with N 0 atoms follows Poisson distribution t N a t e t e N a N t N a k e a k N t t a k = = = = = λ λ λ λ λ λ λ 0 0 0 0, 1 is small, When ) (1 Strictly speaking is called the Poisson rate. ; ;! } Pr{

Radiotracers: Desired Property Decay mode: Clean gamma decay: do not emit alpha or beta articles Positron decay: positron will annihilate with electrons to produce gamma rays Energy of photon: Should be high so that photons can leave the body w/ little attenuation Hard to detect if the energy is too high Desired energy range: 70-511 KeV Half-life Should not be too short (before detector can capture) or too long (longer patient scan time) Minutes to hours desired Half-value-layer (HVL) Thickness of tissue that absorbs half of the radioactivity produced Should be around the dimension of the organ to be imaged Monoenergetic Energy sensitive detectors can discriminate the primary photons from scattered ones. F16 Yao Wang @ NYU 20

Decay Process Examples α decay U Th + He, T 4.5 10 y 238 234 4 9 92 90 2 1 2 - β decay Th Pa + e + ν, T = 24.1 d 234 234 90 91 e 1 2 n H + e + ν, T = 10.6 m 1 1 0 1 e 12 + β decay C B + e + ν, T = 20.38 m 11 11 + 6 5 e 12 C B + e + ν, T = 19.2 s 10 10 + 6 5 e 12 O N + e + ν, T = 122 s 15 15 + 8 7 e 12 Most of these naturally occurring processes are not useful for medical imaging applications, with too long Half-time, too short HVL, too high energy. They can be used as radiotherapeutic agents, if they can be targeted to tumors, to destroy diseased tissue and stops the cancer from proliferating. e capture Ca + e K + ν, T 1 10 y 41 41 5 20 19 e 1 2 F16 Yao Wang @ NYU 21

Radionuclides in Clinical Use Most naturally occurring radioactive isotopes not clinically useful (long T 1/2, charged particle emission, alpha or beta decay) Artificial radioactive isotopes produced by bombarding stable isotopes with high-energy photons or charged particles Nuclear reactors (n), charged particle accelerators (Linacs, Cyclotrons) Mo Tc + e + 99 T1/2= 2.5d 99m ν From H. Graber, Lecture Note, F05 F16 Yao Wang @ NYU 22

The Technetium Generator Can be produced from an on-site generator 99^Mo (Molybdenum) 99m^Tc (Technetium) 99^Tc, Decay characteristics of 99m^Tc: half life =6.02h, E=140 KeV, HVL=4.6 cm m Tc T1/2= 6h Tc + γ 99 99 ( 140 kev) Used in more than 90% of nuclear imaging More detail: see [Webb, sec. 2.5] F16 Yao Wang @ NYU 23

Radiopharmaceuticals Radionuclide is bound to pharmaceuticals specific to metabolic activities (cancer, myocardial perfusion, brain perfusion) Gamma emitters 99m Tc-Sestamibi (myocardial perfusion, cancer) 99m Tc-labeled hexamethyl-propyleneamine (brain perfusion) Positron emitters 11 C, T 1/2 = 20 min [ 12 C (p,pn) 11 C; 14 N (p,α) 11 C]: many organic compounds (binding to nerve receptors, metabolic activity) 13 N, T 1/2 = 10 min [ 16 O (p,α) 13 N; 13 C (p,n) 13 N]: NH 3 (blood flow, regional myocardial perf.) 15 O, T 1/2 = 2.1 min [ 15 N (p,n) 15 O; 14 N (d,n) 15 O]: CO 2 (cerebral blood flow), O 2 (myoc. O 2 consumption), H 2 O (myoc. O 2 consumption & blood perfusion) 18 F, T 1/2 = 110 min [ 18 O (p,n) 18 F; 20 Ne (d,α) 18 F]: 2-deoxy-2-[ 18 F]-fluoroglucose (FDG, neurology, cardiology, oncology, metabolic activity) From H. Graber, Lecture Note, F05 F16 Yao Wang @ NYU 24

Common Radiotracers Thyroid function Kidney function Most commonly used Oxygen metabolism F16 Yao Wang @ NYU 25

Summary of Physics Radioactive decay is the process when a unstable nuclide is changed to a more stable one Four modes of decay, generating alpha particles, beta particles, positrons and gamma rays respectively Medical imaging exploits positron decay and gamma rays Radioactivity follows an exponential decay law in time, characterized by the decay constant or the half-life Desired properties for radio tracers Common radiotracers in nuclear medicine F16 Yao Wang @ NYU 26

Overview of Nuclear Imaging Modalities Planar Scintigraphy Use radiotracers that generate gammay decay, which generates one photon in random direction at a time Capture photons in one direction only, similar to X-ray, but uses emitted gamma rays from patient Use an Anger scintillation camera SPECT (single photon emission computed tomography) Use radiotracers that generate gamma decay Capture photons in multiple directions, similar to X-ray CT Uses a rotating Anger camera to obtain projection data from multiple angles PET (Positron emission tomography) Uses radiotracers that generate positron decay Positron decay produces two photons in two opposite directions at a time Use special coincidence detection circuitry to detect two photons in opposite directions simultaneously Capture projections on multiple directions Will focus on SPECT and PET only F16 Yao Wang @ NYU 27

SPECT Instrumentation Similar to CT, uses a rotating Anger camera to detect photons traversing paths with different directions Must use collimators so that each detector only detects photons in a straight line parallel with the collimator wall. Recent advances uses multiple Anger cameras (multiple heads), reducing scanning time (below 30 minutes) Anger cameras in SPECT must have significantly better performances than for planar scintigraphy to avoid reconstruction artifacts F16 Yao Wang @ NYU 28

A typical SPECT system Fig. 9.1 A dual head system F16 Yao Wang @ NYU 29

Anger Scintillation Camera Compare the detected signal to a threshold Compute the location with highest activity Convert light to electrical currents Convert detected photons to lights Absorb scattered photons F16 Yao Wang @ NYU 30

Collimators F16 Yao Wang @ NYU 31

Imaging Equation when θ=0 (z,l) R Replace x by l F16 Yao Wang @ NYU 32

Examples Example 1: Imaging of a slab Example 2: Imaging of a two-layer slab Go through on the board F16 Yao Wang @ NYU 33

General Case: Imaging Geometry R l s Note: correction to Fig. 9.8 F16 Yao Wang @ NYU 34

General Case: Imaging Equation R l s Note: There should be some constant in front of the integral. See textbook F16 Yao Wang @ NYU 35

Example 1 Imaging of a rectangular region, with the following structure. Derive detector readings in 4 positions (A,B,C,D) B w 1 w 2 w 3 w 4 H 1 C Α1, µ1 H 2 H 2 A H 1 Α2,µ2 D Do you expect the reading at B and D be the same? What about at A and C? F16 Yao Wang @ NYU 36

How to Reconstruct A(x,y)? Two unknowns: A(x,y), μ(x,y) Inverse square law effect: A(x,y) is weighted differently in different detector readings (s depends on l,θ) F16 Yao Wang @ NYU 37

Approximation for Reconstruction Under this assumption, the measurements ϕ(l,θ) are the Radon transform of A(x,y)! A(x,y) can be reconstructed using the reconstruction methods for CT (e.g. convolution backprojection or filtered backprojection) The filter cut-off frequency needs to be chosen properly to balance between removing noise and blurring Commonly used filter in frequency domain F16 Yao Wang @ NYU 38

Correction for Attenuation Factor Use CT imaging to generate an estimate of the tissue µ at each location, before radiotracer is introduced or before significant radioactive decay starts Based on the estimate of µ(x,y), derive the correction factor a(x,y) Assuming A(x,y)=1 everywhere. Numerically compute the expected measurement φ(l,θ) based on the estimated µ(x,y) using the forward model, which can be thought of as the Randon transform of a(x,y). Solve a(x,y) by performing inverse Randon transform on φ(l,θ) Recover A0(x,y) from the real detector readings assuming the readings are sum of A0(x,y) along all projection lines ( l,θ). Call Then correct by A(x,y)= A0(x,y)/a(x,y) See textbook for more detail (only in the 2 nd edition, pp 307-309) F16 Yao Wang @ NYU 39

CT/SPECT Imaging Obtaining CT and SPECT measurement successively while the patient is inside the imaging apparatus CT before radionuclide injection SPECT after Reconstruct μ(x,y) from CT measurement Reconstruct A(x,y) with correction based on μ(x,y) Patient must stay as stationary as possible to avoid misregistration Simultaneous display of μ(x,y) and A(x,y) images can provide more useful information! CT: provides high resolution anatomical information SPECT: Low resolution functional imaging F16 Yao Wang @ NYU 40

From [Smith&Webb] F16 Yao Wang @ NYU 41

From [Smith&Webb] F16 Yao Wang @ NYU 42

SPECT applications Heart (most used): Coronary artery disease Myocardial infarcts Brain: Perfusion (stroke, epilepsy, schizophrenia, dementia [Alzheimer]) Tumors Respiratory Liver Kidney From Graber, Lecture Slides for BMI1,F05 See Webb Sec. 2.10 F16 Yao Wang @ NYU 43

PET Principle F16 Yao Wang @ NYU 44

Annihilation Coincidence Detection Detect two events in opposite directions occurring simultaneously Time window is 2-20 ns, typically 12 ns No detector collimation is required Higher sensitivity F16 Yao Wang @ NYU 45

F16 Yao Wang @ NYU 46

Detected PET Events F16 Yao Wang @ NYU 47

Coincidence Timing F16 Yao Wang @ NYU 48

A Typical PET Scanner F16 Yao Wang @ NYU 49

3D Scanner has higher SNR because it does not reject photons through septal collimators. But must have more sophisticated mechanism to deal with false coincidences. Majority of commercial systems are 3D. F16 Yao Wang @ NYU 50

Imaging Equation See derivation in the textbook (2 nd ed, pp 309-312)! F16 Yao Wang @ NYU 51

Example 2 Imaging of a rectangular region, with the following structure. Derive detector readings in 2 paired positions (A-C, B-D) B w 1 w 2 w 3 w 4 H 1 C Α1, µ1 H 2 H 2 A H 1 Α2,µ2 D F16 Yao Wang @ NYU 52

Attenuation Correction One can apply filtered backprojection algorithm to reconstruct A(x,y) from the corrected sinogram ϕ c (l,θ) Difference from SPECT: Attenuation correction much easier! Can we directly use the CT measurement or reconstructed μ(x,y) by CT for correction? F16 Yao Wang @ NYU 53

Attenuation Correction (2) CT uses X-rays at different energy (e.g. E1 roughly 80-140 KeV) than PET scan (e.g. E2=511KeV) One cannot use the CT measurement or reconstructed μ(x,y) for correction, because μ(x,y) is energy dependent! The reconstructed μ(x,y; E1) image is segmented into a number of tissue types (muscle, lipid, bone), to which standard values of μ at E2=511KeV are assigned. Based on the segmented image with assigned μ values, μ(x,y; E2), correction factors can be numerically generated. The CT-based attenuation map is smoothed to match PET resolution before calculating the correction factor. PET scan takes quite long during which the patient may have natural cardiac or breathing motion. The two reconstructed images need to be registered before using CT-based attenuation map for correction of PET measurement. F16 Yao Wang @ NYU 54

Reconstruction from Corrected Sinogram F16 Yao Wang @ NYU 55

From [Smith&Webb] All commercial PET systems sold are now combined CT/SPECT scanners! [Smith&Webb] Overlaying the two reconstructed images can provide more info than each alone! F16 Yao Wang @ NYU 56

F16 Yao Wang @ NYU 57

PET applications Brain: Tumor detection Neurological function (pathologic, neuroscience app.) Perfusion Cardiac Blood flow Metabolism Tumor detection (metastatic cancer) From H. Graber, lecture slides for BMI1,F05 See Webb Sec. 2.11.7 F16 Yao Wang @ NYU 58

Radiotracers for PET (Except 82 Rb) must be synthesized on site using a cyclotron FDG is used in 80% of PET studies Bio-chemical research: Design of radio tracers to reveal different types of medical problems F16 Yao Wang @ NYU 59

PET Application: See and Hear F16 Yao Wang @ NYU 60

PET evolution From H. Graber, lecture slides for BMI1,F05 F16 Yao Wang @ NYU 61

Image Quality Consideration We will consider the following for scintigraphy, SPECT, and PET together Resolution: collimator, detector intrinsic Noise SNR Read: Sec. 8.4 in Textbook F16 Yao Wang @ NYU 62

Limit of PET Resolution The positron will travel a random distance before its annihilated by an electron The reconstructed image shows the radioactivity at the annihilation site, but the site where positron is produced Inherent limit of PET resolution: cannot be less than the mean traveling range! Long range (hence low resolution) with high energy positron and less dense tissue F16 Yao Wang @ NYU 63

Relation between True Image and Reconstructed Image in SPECT/PET F16 Yao Wang @ NYU 64

PET vs. SPECT PET has much higher SNR and resolution because Not using collimator (hence higher signal strength) Reduced attenuation of higher energy photons (511 vs. 140 KeV) by tissue Use of complete ring of detector PET reconstruction is not affected by larger attenuation deep inside the body as is with SPECT =>PET reconstruction has same resolution/accuracy throughout the body, but SPECT has low resolution/accuracy deeper inside the body PET has higher spatial resolution (8mm vs 15 mm) Faster imaging with PET PET is more expensive and bulky (need an on site cyclotron to produce positron emitting radiotracers) F16 Yao Wang @ NYU 65

Formulation of Reconstruction Problem as Solving a Linear Equation Reconstruction algorithms discussed so far assume the measured data are Radon transform (line integral) of the unknown image or some weighted versions of the unknown image pixels. Measurements in practical systems may not follow the assumed relations exactly, especially the assumed weighting factors. More general modeling Matrix representation F16 Yao Wang @ NYU 66

Pop Quiz What are f i and a ik for different imaging systems (CT, SPECT, PET)? F16 Yao Wang @ NYU 67

Least squares solution When the number of measurements>=number of pixels, we want to solve f so that it satisfies all the equations with minimal sum of squared errors Closed form solution for minimizing the above error (obtained by setting the derivative with respect to f to zero): Direct inverse is not feasible if the number of image pixels and the number measurements are very large. Iterative algorithms: Solving the matrix equation iteratively based on a given criterion (not necessarily least squares) F16 Yao Wang @ NYU 68

Gradient Descent Algorithms for Least Squares Solution Least squares problem is an optimization problem: Gradient descent Guaranteed to converge to global minimum because the energy function is a convex function of f F16 Yao Wang @ NYU 69

POCS (Projection onto convex set) General idea: If f is known to lie in some convex sets, one can iteratively project the current f to each convex set successively. The iteration will converge to the intersection of all convex sets! And if the intersection is non-empty, it will go to the point on the intersection that is closest to the initial solution. Common convex constraints: positivity, norm, linear constraint, etc. Youla, Dan C., and Heywood Webb. "Image Restoration by the Method of Convex Projections: Part 1 --- Theory." IEEE transactions on medical imaging 1.2 (1982): 81-94. F16 Yao Wang @ NYU 70

POCS for Linear Equations F16 Yao Wang @ NYU 71

Maximum Likelihood Estimate Why minimize the sum of square errors? F16 Yao Wang @ NYU 72

Using Prior Information Sometimes we have prior knowledge about certain properties of f. We can incorporate such knowledge in the energy function to be minimized, or simply add the knowledge as explicit constraint to solve a constrained optimization problem. Such prior constraints are particularly important if the equation is under determined (i.e. number of unknowns or pixels < number of measurements) Popular approach: Assume the image to be reconstructed is sparse in the wavelet transform domain, and minimize the following energy function F16 Yao Wang @ NYU 73

Summary of Nuclear Imaging Principles Three major imaging modalities: Planar scintigraphy SPECT PET Principle of Anger camera: collimator, scintillation crystal, photomultiplier Imaging principles of planar scintigraphy and SPECT Both based on gamma decay Very similar to X-ray projection and CT, except for the attenuation factor Combined SPECT/CT systems Imaging principle of PET: Coincidence detection: detect two photons reaching two opposite detectors simultaneously (within a short time window) Detected signal is the product of two terms, depending on the radioactivity A(x,y) and attenuation μ(x,y) separately The measurement can be corrected based on known μ(x,y) A(x,y) can be reconstructed using convolution or filtered backprojection on the corrected measurement Combined PET/CT systems F16 Yao Wang @ NYU 74

Reference Prince and Links, Medical Imaging Signals and Systems, Chap 7,8,9. Section on attenuation correction for SPECT and PET reconstruction are revised in 2 nd ed. New section added on Iterative Image Reconstruction A. Webb, Introduction to Biomedical Imaging, Chap. 2 Sec. 2.5 for Technetium generation; Sec. 2.10, Sec. 2.11.7 for Clinical applications of nuclear medicine. Other recommended readings: K. Miles, P. Dawson, and M. Blomley (Eds.), Functional Computed Tomography (Isis Medical Media, Oxford, 1997). R. J. English, SPECT: Single Photon Emission Computed Tomography: A Primer (Society of Nuclear Medicine, Reston, VA, 1995). M. Reivich and A. Alavi (Eds.), Positron Emission Tomography (A. R. Liss, NY, 1985). F16 Yao Wang @ NYU 75

Homework Reading: Prince and Links, Medical Imaging Signals and Systems, Ch. 7, 8,9. Note down all the corrections for Ch. 7,8,9 on your copy of the textbook based on the provided errata. Problems from Chap 7,8,9 of the text book P.7.4 P7.6 P7.7 (assume the energy of the photons is E) P7.9 P9.9 (P9.4 in 1 st edition) P9.3 (not present in 1 st edition) P9.4 (not present in 1 st edition) P.9.10 (not present in 1 st edition) Homework will not be collected, solutions will be available Tuesday October 17. F16 Yao Wang @ NYU 76