AppNote 7/2013. Comprehensive Automation of the SPE- GC/MS Analysis of Opioids, Cocaine and Metabolites from Serum and Other Matrices KEYWORDS

Similar documents
AppNote 5/2008. Automation of Sample Preparation Steps using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

AppNote 14/2012. Thermal Gravimetric Analysis/Mass Spectrometry Simulation using the GERSTEL Automated Pyrolyzer KEYWORDS ABSTRACT

AppNote 8/2008. A New Purge Tool for Use with Automated Headspace Analysis KEYWORDS ABSTRACT

RESEARCH PAPER. Oliver Lerch & Oliver Temme & Thomas Daldrup. Keywords Opiates. Opioids. Cocaine. Solid phase extraction (SPE). Automation.

AppNote 7/2014. Automating Liquid-Liquid Extractions using a Bench-top Workstation KEYWORDS ABSTRACT. Sample Preparation, Lab Automation, LC/MS/MS

AppNote 3/2014. Analysis of Aroma Compounds in Edible Oils by Direct Thermal Desorption GC/MS using Slitted Microvials KEYWORDS ABSTRACT

AppNote 8/2012. Rapid Automated Extraction and Confi rmation of Buprenorphine and Norbuprenorphine in Urine by DPX-LC/MS/MS KEYWORDS ABSTRACT

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

AppNote 4/1998. New Developments in Multi-Dimensional Capillary Gas Chromatography KEYWORDS ABSTRACT

TechNote 5/2002 KEYWORDS ABSTRACT. SBSE, Twister, Thermal Desorption

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances KEYWORDS

AromaOffi ce: Application of a Novel Linear Retention Indices Database to a Complex Hop Essential Oil

AppNote 1/2007. Automated Dynamic Headspace Sampling using Replaceable Sorbent Traps KEYWORDS ABSTRACT

AppNote 2/2013. A High Throughput Automated Sample Preparation and Analysis Workfl ow for Comprehensive Toxicology Urine Screenings using LC/MS/MS

AppNote 15/2012. Large Volume Injection with Solvent Venting - Application to Trace Detection of Analytes in Water INTRODUCTION

AppNote 1/2013. Determination of Barbiturates and 11-Nor-9-carboxy- 9 -THC in Urine using Automated Disposable Pipette Extraction (DPX) and LC/MS/MS

AppNote 3/2007 KEYWORDS INTRODUCTION. Beverage, Detector, Headspace, Quality Control

AppNote 10/2012. Large Volume Injection with GC/MS Multi Column Switching for Direct Determination of Ethyl Carbamate in Alcoholic Beverages KEYWORDS

AppNote 2/1996. Andreas Hoffmann Gerstel GmbH & Co.KG, Eberhard-Gerstel-Platz 1, D Mülheim an der Ruhr, Germany

AppNote 5/2002. Characterization of a Mass Spectrometer based Electronic Nose for Routine Quality Control Measurements of Flavors KEYWORDS ABSTRACT

AppNote 6/2010. part a. Alternative Procedure for Extraction and Analysis of PAHs in Seafood by QuEChERS-SBSE-GC-MS KEYWORDS ABSTRACT

AppNote 1/2004. Use of a Mass Spectral Based Chemical Sensor to Discriminate Food and Beverage Samples: Olive Oils and Wine as Examples KEYWORDS

Off-gassing of Rubber Particles Used for Athletic Fields using Automated Dynamic Headspace Sampling

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to En vi ron men tal Aqueous Samples INTRODUCTION

AppNote 6/2014. Automated Online Desorption and Analysis of DNPH Derivatives of Airborne Aldehydes and Ketones KEYWORDS ABSTRACT

AppNote 8/2002 KEYWORDS ABSTRACT. Electronic Nose, Fast GC, Headspace Analysis, Chemometrics,

Analysis of Food Samples using Thin Film Solid Phase Microextraction (TF-SPME) and Thermal Desorption GC/MS

AppNote 3/2012. Rapid Automated Screening of Extractable Compounds in Materials for Food Packaging, Medical or Technical Purposes KEYWORDS ABSTRACT

AppNote 2/2002. Discrimination of Soft Drinks using a Chemical Sensor and Principal Component Analysis KEYWORDS ABSTRACT

Thermal Desorption Unit TDU

AppNote 8/2009. Fragrance Profi ling of Consumer Products using a Fully Automated Dynamic Headspace System KEYWORDS ABSTRACT

AppNote 1/2011 KEYWORDS ABSTRACT

AppNote 11/2012. Fully Automated Analysis of Doping Substances in Equine Urine by Disposable Pipette Extraction (DPX) coupled to GC/QqQ-MS KEYWORDS

AppNote 2/2012. Determining Phenolic Compounds in Whisky using Direct Large Volume Injection and Stir Bar Sorptive Extraction KEYWORDS ABSTRACT

AppNote 2/2007. Automated Solid Phase Microextraction using the GERSTEL MPS Prepstation and MAESTRO Software KEYWORDS ABSTRACT

AppNote 2/2004. Analysis of Packaging Materials using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

AppNote 2/2001. Stir Bar Sorptive Extraction: Enhancing Selectivity of the PDMS Phase KEYWORDS ABSTRACT

AppNote 6/2012. Flavor and Fragrance Analysis of Consumer Products - Dynamic Headspace Compared to Some Traditional Analysis Approaches KEYWORDS

GERSTEL MultiPurpose Sampler MPS. Versatile autosampler and sample preparation robot

AppNote 8/2000. Volatile Organic Compounds from Adhesives and their Contribution to Indoor Air Problems KEY WORDS ABSTRACT

AppNote 12/2012. Comparison of EG-Silicone-SBSE and Derivatization-PDMS-SBSE for the Analysis of Phenolic Compounds and Off-fl avors in Water KEYWORDS

AppNote 5/2001. The Analysis of the Bitter and Other Flavour Compounds in Beer and Wort by Stir Bar Sorptive Extraction (SBSE) Followed by HPLC

AppNote 4/2005. A Selectable Single or Multidimensional. Fraction Collection and Dual Detection for Trace Analysis of Complex Samples KEYWORDS

AppNote 2/1998. Direct Injection of Distilled Spirits with PTV Matrix Removal: The Perfect Splitless Injection? KEYWORDS INTRODUCTION

AppNote 8/2011 KEYWORDS ABSTRACT INTRODUCTION

AppNote 7/2002. Classifi cation of Food and Flavor Samples using a Chemical Sensor KEYWORDS ABSTRACT

AppNote 9/2012. Solid Phase Micro-Extraction (SPME) Coupled with Selectable 1 D/ 2 D GC-MS for the Determination of Food Product Flavors KEYWORDS

Benzoylecgonine in Urine by SAMHSA GC/MS

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

AppNote 12/2008 KEYWORDS ABSTRACT

Opiates in Urine by SAMHSA GC/MS

KEYWORDS ABSTRACT AppNote 6/2002

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Abstract. Experimental Sample Preparation

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

AppNote 5/1994. The Use of a Multi Purpose Sampler for Headspace GC-MS Analysis of Volatile Organic Compounds in Human Urine

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Confirmation and Quantitation of THC in Oral Fluid Using an MRM Method on a GC-Triple Quadrupole MS

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Selective Sample Enrichment of Gaseous Samples using a Cooled Injection System: Trace Determination of Sulphur Components in Natural Gas

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Determination of Cannabinoids (THC) in Biological Samples

AppNote 6/2010. part b. High Throughput Method for the Determination of PAHs in Seafood by QuEChERS-SBSE-GC-MS SCOPE PRINCIPLE

AppNote 3/1998 KEYWORDS ABSTRACT. Beverage, Food & Flavor, PTV, Large Volume Injection, Headspace, Thermal Desorption.

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber

Appendix II- Bioanalytical Method Development and Validation

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

EVOLUTE WCX Columns for Solid Phase Extraction of Quaternary Amines and Strongly Basic Compounds from Aqueous Samples

AppNote 7/2005. Detection of Packaging Emissions using a Flexible Headspace Sampler Combined with a Multi Sensor System and a Separation Unit KEYWORDS

Tecan AC Extraction Plate

NORBUPRENORPHINE (Buprenorphine s Metabolite ) BUPRENORPHINE in urine by GC/MS Code GC Method of Confirmation by GC-MS

Detection of 9-tetrahydrocannabinol ( 9-THC) in human urine by Solid Phase Extraction and HPLC.

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Psychoactive Drugs Urine LC-MS/MS Analysis Kit User Manual

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

STANDARD OPERATING PROCEDURES

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Analysis of Drugs in Hair using GC/GC/MS CAT Meeting 3/2/07

Sulfotepp impurities in Chlorpyrifos EC formulations

Don t Work Day and Night!

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Transcription:

AppNote 7/2013 Comprehensive Automation of the SPE- GC/MS Analysis of Opioids, Cocaine and Metabolites from Serum and Other Matrices Oliver Lerch Gerstel GmbH & Co. KG, Eberhard-Gerstel-Platz 1, D-45473 Mülheim an der Ruhr, Germany Oliver Temme, Thomas Daldrup University Hospital Düsseldorf, Institute of Legal Medicine, Department of Forensic Toxicology, Moorenstrasse 5, D-40225 Düsseldorf, Germany KEYWORDS Opiates, Opioids, Cocaine, Solid Phase Extraction (SPE), Automation, GC/MS ABSTRACT Analyzing blood serum for opioids, cocaine and metabolites is a routine task in forensic laboratories. The most commonly used methods involve several manual or partly-automated sample preparation steps such as protein precipitation, solid phase extraction, evaporation and derivatization followed by GC/MS or LC/MS determination. In this study a comprehensively automated method is compared with a validated, partly-automated routine method. Following manual protein precipitation, the automated method relies on a MultiPurpose Sampler (MPS) to perform all remaining sample preparation steps. These include solid phase extraction (SPE), evaporation of the eluate, derivatization and introduction to the GC/MS. Quantitative analysis of close to 170 serum samples, as well as more than 50 samples of other matrices like urine, different tissues and heart blood, was performed using both methods. Cocaine, benzoylecgonine, methadone, morphine, codeine, 6-monoacetylmorphine, dihydrocodeine and

7-aminoflunitrazepam were determined quantitatively and the methods were found to produce equivalent analytical results even near the limits of quantification [1]. INTRODUCTION Toxicological chromatographic analysis of biological fluids or tissues usually requires sample preparation for cleanup and enrichment. Recently in conjunction with very sensitive and selective mass spectrometers protein precipitation alone [2] or dilute and shoot methods [3,4] were also used though these methods may suffer from sample dependent matrix effects that can compromise the accuracy of the results. Although the number of LC-MS/MS methods is rapidly increasing [5,6,7,8] GC-MS/(MS) is still the standard routine analysis technique in many forensic laboratories [5,9,10,11]. Solid Phase Extraction (SPE) is the most widely used extraction technique for toxicological analysis of biological fluids and tissues [10,11,12]. Typically polypropylene cartridges with a fixed sorbent bed (typically mixed mode cation exchange cartridges) are used. A range of methods for the analysis of opioids, cocaine and metabolites in different matrices was published. The compounds were analyzed in matrices like urine [10,13,14], whole blood, serum, plasma [6,8,10], saliva [7], hair [9] or post-mortem samples [15]. In this study a validated, partly-automated SPE- GC/MS analysis method for opioids, cocaine and metabolites was completely automated. Analysis results of both methods were compared. Automation was performed by different modules attached to a GERSTEL MultiPurpose Sampler MPS. This allowed mimicking the manual workflow of sample dilution, SPE, evaporation, derivatization and sample injection. EXPERIMENTAL Instrumentation. The system employed for automation of the sample preparation is based on a MultiPurpose Sampler (MPS, GERSTEL). It was configured with two syringes, a 2.5 ml syringe with gas supply for the sample preparation steps and a 10 μl syringe used for sample injection into a Cooled Injection System CIS 4 (GERSTEL) coupled to a 7890 GC/5975 MSD (Agilent Technologies). The MPS was equipped with modules for solid phase extraction (SPE), for evaporation of solvents under controlled vacuum and temperature (MultiPosition evaporation station, mvap), for shaking under controlled temperature conditions (Agitator) and for supplying large volumes of solvents (Solvent Filling Station 2, SFS 2, all GERSTEL). The complete system is shown in figure 1. Figure 1. Setup used for the automated analysis of blood serum for THC and metabolites. Dual Head MultiPurpose Sampler (MPS) equipped with solvent bottle holder, agitator, standard wash station, trays for eluate vials, SPE cartridges and samples, SPE module, two solvent filling stations (SFS), MultiPosition evaporation station (mvap) and solvent bottle holder (from left to right). AN/2013/07-2

Measurements with the partly-automated, validated analysis method were conducted on a 7890 GC/5975 MSD (Agilent Technologies). A 7683B autosampler was used for injection into a hot Split/Splitless inlet (both Agilent Technologies). A bench-top instrument was used for automation of the solid phase extraction. All other sample preparation steps were done manually. Materials. All analytes and deuterated analogues were certified standards purchased from Lipomed AG or LGC Promochem GmbH. All solvents and salts were of analytical grade and purchased from VWR. N-Methyl- N-trimethylsilyltrifluoroacetamide (MSTFA) for silylation was purchased from MachereyNagel or Sigma-Aldrich. Bond Elut Certify 130 mg, 3 ml format SPE cartridges from Agilent Technologies were used. For automated SPE these cartridges were cut at the top, equipped with a transport adapter and a disposable syringe needle (canula) (figure 2). Blood, urine and tissue samples were taken from authentic forensic cases of the Institute of Legal Medicine in Düsseldorf. Figure 2. Top: Solid phase extraction cartridge configured for automated GERSTEL SPE. Bottom: Standard solid phase extraction cartridge. Preparation of standards and solutions. For calibration multi-compound calibration solutions and one multicompound internal standard solution containing deuterated analogues of every analyte were prepared in methanol. The calibration ranged from 25 to 1500 ng/ml (methadone), from 50 to 1500 ng/ml (benzoylecgonine), from 5 to 150 ng/ml (codeine), from 5 to 300 ng/ml (cocaine, dihydrocodeine, morphine), and from 2.5 to 150 ng/ml (7-aminoflunitrazepam, 6-monoacetylmorphine) respectively and was calculated for 0.6 ml serum sample (nine levels). Each 20 μl of the internal standard solution was added to samples, calibration samples or quality control samples. Manual sample pretreatment. All liquids (urine, blood, serum) were handled in the same way: 1. Protein precipitation by drop-wise addition of a mixture of 0.6 ml sample, 0.1 ml water and 20 μl internal standard solution to a mixture of 1 ml acetonitrile and 0.1 ml isopropanol. 2. Mixing and centrifugation. 3. Transfer of an aliquot (0.75 ml) of the supernatant to individual vials (093640-046-00 with cap 093640-075-00) for both analysis methods. Tissues (brain and kidney, native and lyophilized) were homogenized with an Ultra Turrax. An aliquot of approximately 0.6 g was weighed and handled like the liquid samples as described above, except the acetonitrile/isopropanol solution was added to the sample/standard mixture. These protein precipitation steps could be automated by employing a centrifuge combined with the MPS, but this was not within the scope of this study. Automated sample preparation The following list describes the completely automated method. Differences to the partly-automated method are given in [brackets]. 1. Condition the SPE cartridge with 2 ml methanol and 2 ml phosphate buffer (ph 7.9). 2. Dilute the supernatant of the protein precipitation in the SPE syringe [in a vial] and add the diluted sample to the SPE cartridge. 3. Wash the cartridge with 2 ml water, 2 ml acetic acid, and 2 ml methanol. 4. Dry cartridge briefly using a flow of nitrogen. 5. Elute with 1.9 ml [2 ml] of dichloromethane/ isopropanol/ammonia. The first 0.6 ml are discarded and the following 1.3 ml are collected [2 ml are collected] in a vial (093640-046-00, with cap 093640-102-00). 6. Evaporate the eluate to dryness at 70 C, 8 kpa and 300 rpm in the mvap station [manually at 60 C under nitrogen]. 7. Reconstitute in 200 μl isooctane/pyridine/mstfa 14/5/1 v/v/v [isooctane/mstfa 19/1 v/v]. 8. Shake for 5 min at 90 C [30 min at 90 C] for derivatization. 9. Inject 2 μl into the CIS [2 μl into the split/splitless injector]. Calibration solutions were treated analogous to the eluates. AN/2013/07-3

Analysis Conditions. In the following, parameters for the completely automated method are listed. Whenever these differ from the parameters used in the partlyautomated method, the original parameters are listed in [brackets]. MPS: 2 μl injection volume CIS: 50 C; 12 C/s; 280 C (5 min) [270 C isothermal] Inlet Liner: Quartz wool deactivated [glass wool] Injection Mode: Splitless, 3 min [2 min] Pneumatics: He, constant flow, 1 ml/min Oven: 140 C (1 min); 120 C/min; 225 C (5.29 min) 120 C/min; 275 C (5.2 min) Post Run: 300 C (2.5 min) Column: Rxi-5Sil MS, Restek [HP-5ms, Agilent Technologies] 30 m, d i = 0.25 mm, d f = 0.25 μm MSD Mode: Selected ion monitoring (SIM) SIM Masses: see Table 1 Table 1. Quantifier and qualifier ions for analytes and internal standards. Compound Quantifi er [m/z] Qualifier [m/z] Cocaine 182 303, 198 Cocaine-d 3 185 306, 201 Benzoylecgonine 361 256, 346 Benzoylecgonine-d 3 364 259, 349 Methadone 223 294, 236 Methadone-d 9 303 226 a, 303 a, 318 b, b 242 Morphine 429 220, 401 Morphine-d 3 432 223, 404 Codeine 371 234, 343 Codeine-d 3 374 237, 346 6-Monoacetylmorphine 399 340, 400 6-Monoacetylmorphine-d 3 402 343, 403 Dihydrocodeine 373 315, 358 Dihydrocodeine-d 6 379 318, 364 7-Aminoflunitrazepam 326 326 b, 355 a, 356 a, a 327 b, 354 b 7-Aminofl unitrazepam-d 7 362 333, 363 a Qualifi er ion used in partly-automated analysis method. b Qualifi er ion used in fully automated analysis method. According to guidelines of the Society of Toxicological and Forensic Chemistry (GTFCh, Germany) a blank injection of pure derivatization solution was done after every sample, quality control or calibration sample. RESULTS AND DISCUSSION The validated, partly-automated routine analysis method (table 2) could be successfully automated using the MPS starting with the dilution of the sample after protein precipitation and ending with the injection into the GC/MS. Some modifications were necessary to establish the automated method: Table 2. Limit of detection (LOD), limit of quantification (LOQ) and upper limit of calibration (ULOC) for each compound using the validated, partly-automated reference method. Analyte LOD [ng/ml] LOQ [ng/ml] ULOC [ng/ml] Cocaine 1.1 3.5 300 Benzoylecgonine 9 47 1500 Methadone 4.2 16.7 1500 Morphine 1.2 4.9 300 Codeine 0.4 2.6 150 6-Monoacetylmorphine 0.3 0.8 150 Dihydrocodeine 0.8 4.2 300 7-Aminofl unitrazepam 0.6 2.5 150 The dilution of the supernatant after protein precipitation was partly done in the autosampler syringe. Therefore a 0.75 ml aliquot of the supernatant was diluted with 0.75 ml phosphate buffer and 0.75 ml of this mixture was aspirated. After that another 1.75 ml phosphate buffer was aspirated resulting in the final dilution (same as in the reference method). This solution was added to the SPE cartridge and the process was repeated once to transfer the entire sample. AN/2013/07-4

The elution volume was reduced from 2 ml to 1.9 ml. The first 0.6 ml were discarded and the last 1.3 ml were collected based on the established elution profile (figure 3). 100 90 Eluted proportion [%] 80 70 60 50 40 30 20 10 0 Cocaine Benzoylecgonine 7-Aminoflunitrazepam Methadone Morphine Codeine Dihydrocodeine 6-Monoacetylmorphine Eluted fraction Figure 3. Profile of analyte elution from Bond Elut Certify SPE cartridge. Eventually collected fractions between dashed lines. The derivatization time could be shortened from 30 min to 5 min with shaking at 90 C (see figure 4) by employing a mixture of isooctane/pyridine/mstfa 14/5/1 (v/v/v) instead of isooctane/mstfa 19/1 (v/v) which was used originally. 1000000 Peak area 100000 Cocaine Benzoylecgonine 7-Aminoflunitrazepam Methadone Morphine Codeine Dihydrocodeine 6-Monoacetylmorphine 10000 0 5 10 15 20 25 30 35 Derivatization time [min] Figure 4. Optimization of derivatization time (with shaking) at 90 C with a mixture of isooctane/pyridine/ MSTFA 14/5/1 v/v/v. 5 min with shaking is sufficient for complete derivatization. AN/2013/07-5

Almost 170 serum samples and more than 50 samples of other matrices like urine, different tissues and heart blood were analyzed by both methods. Results are equivalent as can be seen in the double logarithmic line- and Bland-Altman-plots (figures 6 and 7). This is true for serum samples and also for alternative matrix samples. Although results between the limit of quantification and the limit of detection may not be reported routinely, they are included in the line plots (dashed red lines in figures 6). Even in this concentration range the method equivalence is obvious. Since only a couple of samples were positive for dihydrocodeine and 7-aminoflunitrazepam these results are not plotted. Samples and quality control samples were also in good concordance for these compounds. 50000 48000 46000 44000 42000 40000 38000 36000 34000 32000 30000 28000 26000 24000 22000 20000 18000 16000 14000 12000 10000 8000 6000 4000 2000 Methadone 442 ng/ml Benzoylecgonine 172 ng/ml Codeine 23 ng/ml Morphine 70 ng/ml 7-Aminoflunitrazepam 9,4 ng/ml 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00 10.50 11.00 Figure 5. Extracted-ion chromatogram resulting from a real serum sample. Quantified compounds are named. AN/2013/07-6

GERSTEL MPS automation [ng/ml] GERSTEL MPS automation [ng/ml] GERSTEL MPS automation [ng/ml] GERSTEL MPS automation [ng/ml] GERSTEL MPS automation [ng/ml] GERSTEL MPS automation [ng/ml] Figure 6. Correlation of determined analyte concentrations in double logarithmic scale. Line with a slope of one representing complete equivalence of results is shown. ng/ml: Nanogram per milliliter or nanogram per gram for tissue respectively; Other: Other matrices than serum - urine, blood, lyophilized kidney tissue, heart blood, lyophilized and native brain tissue; LOD: Limit of detection; LOQ: Limit of quantification; ULOC: Upper limit of calibration. AN/2013/07-7

Relative deviation (MPS/Reference) Relative deviation (MPS/Reference) Relative deviation (MPS/Reference) Relative deviation (MPS/Reference) Relative deviation (MPS/Reference) Relative deviation (MPS/Reference) Figure 7. Relative deviations of measured concentrations displayed in Bland-Altman-plots. ng/ml: Nanogram per milliliter or nanogram per gram for tissue respectively; Other: Other matrices than serum - urine, blood, lyophilized kidney tissue, heart blood, lyophilized and native brain tissue; LOD: Limit of detection; LOQ: Limit of quantification; ULOC: Upper limit of calibration; SD: Standard deviation of relative deviations. No carryover for any of the compounds could be detected when extracting blank serum after real samples. By overlapping sample preparation steps with the GC/MS run a throughput of around 26 samples per day could be achieved which is comparable with the partly-automated reference method. The analyses were performed in different laboratories by different personnel at different times revealing the ruggedness of the instrumentation and methods and the suitability for routine forensic analysis tasks. AN/2013/07-8

CONCLUSIONS The following achievements were made: Comprehensive automation of a validated, partlyautomated analysis method for opioids, cocaine and metabolites from blood serum and other matrices. Analysis results of both methods are equivalent on the basis of GTFCh recommendations. The automated method proved to be rugged and suitable for routine analysis in forensic laboratories. The automated method saves manual work and reduces the risk of human error. It generates a throughput of 29 samples per day, which is similar to the reference method and corresponds to the GC/ MS analysis time. The analysis system is highly flexible and can mimic manual sample preparation workflows. Therefore it can be employed for easy automation of other validated GC or LC analysis methods or for standalone automation of sample preparation. REFERENCES [1] O. Lerch, O. Temme, T. Daldrup, Anal. Bioanal. Chem. (2014) DOI: 10.1007/s00216-014-7815-7 [2] M. Sergi, E. Bafile, D. Compagnone, R. Curini, G. D Ascenzo, F.S. Romolo, Anal. Bioanal. Chem. 393 (2009) 709. [3] S. Guddat, E. Solymos, A. Orlovius, A. Thomas, G. Sigmund, H. Geyer, M. Thevis, W. Schänzer, Drug Test. Anal. 3 (2011) 836. [4] R.L. Fitzgerald, T.L. Griffin, Y. Yun, R.A. Godfey, R. West, A.J. Pesce, D.A. Herold, J. Anal. Toxicol. 36 (2012) 106. [5] M.R. Moeller, T. Kraemer, Ther. Drug Monit. 24 (2002) 210. [6] M. Kjaergaard Bjork, M.K.K. Nielsen, L.O. Markussen, H.B. Klinke, K. Linnet, Anal. Bioanal. Chem. 396 (2010) 2393. [7] N. Badawi, K. Wiese Simonsen, A. Steentoft, I.M. Bernhoft, K. Linnet, Clin. Chem. 55 (2009) 2004. [8] N. Ferreiros Bouzas, S. Dresen, B. Munz, W. Weinmann, Anal. Bioanal. Chem. 395 (2009) 2499. [9] I. Angeli, M. Minoli, A. Ravelli, F. Gigli, F. Lodi, Forensic Sci. Int. 218 (2012) 15. [10] A.W. Jones, A. Holmgren, F.C. Kugelberg, J. Anal. Toxicol. 32 (2008) 265. [11] S.W. Toennes, S. Steinmeyer, H.J. Maurer, M.R. Moeller, G.F. Kauert, J. Anal. Toxicol. 29 (2005) 22. [12] S.M.R. Wille, W.E.E. Lambert, Anal. Bioanal. Chem. 388 (2007) 1381. [13] E. Jagerdeo, M.A. Montgomery, M. Sibum, T.A. Sasaki, M.A. LeBeau, J. Anal. Toxicol. 32 (2008) 570. [14] R. Dams, T. Benijts, W.E. Lambert, A.P. De Leenheer, J. Chromatogr. B 773 (2002) 53. [15] T. Stimpfl, W. Vycudilik, Forensic Sci. Int. 142 (2004) 115. For research use only. Not for use in diagnostic procedures. The information provided for this product is intended for reference and research purposes only. GERSTEL offers no guarantee as to the quality and suitability of this data for your specific application. Information, descriptions and specifications in this publication are subject to change without notice. AN/2013/07-9

GERSTEL GmbH & Co. KG Eberhard-Gerstel-Platz 1 45473 Mülheim an der Ruhr Germany +49 (0) 208-7 65 03-0 +49 (0) 208-7 65 03 33 gerstel@gerstel.com www.gerstel.com GERSTEL Worldwide GERSTEL, Inc. 701 Digital Drive, Suite J Linthicum, MD 21090 USA +1 (410) 247 5885 +1 (410) 247 5887 sales@gerstelus.com www.gerstelus.com GERSTEL AG Wassergrabe 27 CH-6210 Sursee Switzerland +41 (41) 9 21 97 23 gerstelag@ch.gerstel.com www.gerstel.ch GERSTEL K.K. 1-3-1 Nakane, Meguro-ku Tokyo 152-0031 SMBC Toritsudai Ekimae Bldg 4F Japan +81 3 5731 5321 +81 3 5731 5322 info@gerstel.co.jp www.gerstel.co.jp GERSTEL LLP Level 25, North Tower One Raffles Quay Singapore 048583 +65 6622 5486 +65 6622 5999 SEA@gerstel.com www.gerstel.com GERSTEL Brasil Av. Pascoal da Rocha Falcão, 367 04785-000 São Paulo - SP Brasil +55 (11)5665-8931 +55 (11)5666-9084 gerstel-brasil@gerstel.com www.gerstel.com.br Information, descriptions and specifications in this Publication are subject to change without notice. GERSTEL, GRAPHPACK and TWISTER are registered trademarks of GERSTEL GmbH & Co. KG. Copyright by GERSTEL GmbH & Co. KG Awarded for the active pursuit of environmental sustainability