Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Similar documents
Test of the rem-counter WENDI-II from Eberline in different energy-dispersed neutron fields

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI

Neutron Metrology Activities at CIAE (2009~2010)

NEUTRON H*(10) INSIDE A PROTON THERAPY FACILITY: COMPARISON BETWEEN MONTE CARLO SIMULATIONS AND WENDI-2 MEASUREMENTS

A new neutron monitor for pulsed fields at high-energy accelerators

Radiation Protection At Synchrotron Radiation Facilities

Shielding calculations for the design of new Beamlines at ALBA Synchrotron

Occupational Radiation Protection at Accelerator Facilities: Challenges

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Reactions of neutron-rich Sn isotopes investigated at relativistic energies at R 3 B

Feasibility Studies for the EXL Project at FAIR *

EXPERIMENTAL STUDY OF NEUTRON FIELDS PRODUCED IN PROTON REACTIONS WITH HEAVY TARGETS. Nuclear Physics Institute AS CR, Rez Czech Republic

Measurements of Radiation Doses Induced by High Intensity Laser between and W/cm 2 onto Solid Targets at LCLS MEC Instrument

Fission Fragment characterization with FALSTAFF at NFS

1. RADIOACTIVITY AND RADIATION PROTECTION

Introduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose

Coulomb Breakup as a novel spectroscopic tool to probe directly the quantum numbers of valence nucleon of the exotic nuclei

Radionuclide Imaging MII Detection of Nuclear Emission

Neutron dosimetry and microdosimetry with track etch based LET spectrometer

Tracking at the LAND/R B setup on 17

Validation of the UFS Bonner Sphere Spectrometer and Monte Carlo Methods at the CERN-EU high energy Reference Field (CERF)

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515

Calibration of the GNU and HSREM neutron survey instruments

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Development of a neutron time-of-flight source at the ELBE accelerator

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Neutron multiplicity. Total radioactivity (Ci) Time after beam cut. Target A s 1min 1h 1d 1m 1y 10y 100y

Radioactivity & Radiation Protection *

PoS(NIC XI)248. Cross section measurements of 103 Rh(p, γ) 104 Pd with the Karlsruhe 4πBaF 2 detector

Bonner Sphere Spectrometer. Cruzate, J.A.; Carelli, J.L. and Gregori, B.N.

PERSONNEL DOSE EQUIVALENT MONITORING AT SLAC USING LITHIUM-FLUORIDE TLD S*

Research Physicist Field of Nuclear physics and Detector physics. Developing detector for radiation fields around particle accelerators using:

Part E. Radiation monitors. Radiation Safety - JUAS 2014, X. Queralt. Part E. Radiation monitors

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

PECULIARITIES OF FORMING THE RADIATION SITUATION AT AN AREA OF NSC KIPT ACCELERATORS LOCATION

Fragmentation and space radioprotection

Neutrons for Science (NFS) at SPIRAL-2

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

MONTE-CARLO SIMULATIONS OF TIME- RESOLVED, OPTICAL READOUT DETECTOR for PULSED, FAST-NEUTRON TRANSMISSION SPECTROSCOPY (PFNTS)

Neutron Transport Calculations Using Monte-Carlo Methods. Sean Lourette Fairport High School Advisor: Christian Stoeckl

Improving neutron detection efficiency by using passive converters

Neutron Users Club Meeting

Portable, Low-cost Proportional Counters for Space, Atmospheric and Ground based Applications

Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility

Atoms, Radiation, and Radiation Protection

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL

The Possibility to Use Energy plus Transmutation Setup for Neutron Production and Transport Benchmark Studies

H4IRRAD generic simulation results

Secondary Particles Produced by Hadron Therapy

Compton suppression spectrometry

Lab #13: MEASUREMENT OF RADIATION DOSES

MEASUREMENT AND DETECTION OF RADIATION

Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

Thursday, April 23, 15. Nuclear Physics

Radiation Dose, Biology & Risk

PoS(Baldin ISHEPP XXII)042

Measurements of liquid xenon s response to low-energy particle interactions

PoS(ICHEP2016)474. SoLid: Search for Oscillations with a Lithium-6 Detector at the SCK CEN BR2 reactor

New irradiation zones at the CERN-PS

Jeancarlo Torres, CHP. Naval Surface Warfare Center Carderock Division

Light ion recoil detector

Calibration of the Modular Neutron Array (MoNA)

Radiation and Radioactivity. PHYS 0219 Radiation and Radioactivity

L-35 Modern Physics-3 Nuclear Physics 29:006 FINAL EXAM. Structure of the nucleus. The atom and the nucleus. Nuclear Terminology

IAC-08-A MONTE CARLO SIMULATIONS OF ENERGY LOSSES BY SPACE PROTONS IN THE CRATER DETECTOR

COMET muon conversion experiment in J-PARC

Cross section measurements of the elastic electron - deuteron scattering

JRPR. Measurement of Neutron Production Doubledifferential Cross-sections on Carbon Bombarded with 430 MeV/Nucleon Carbon Ions.

THE GSI FUTURE PROJECT: AN INTERNATIONAL ACCELERATOR FACILITY FOR BEAMS OF IONS AND ANTIPROTONS

Chapter 11: Neutrons detectors

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Recent activities in neutron standardization at NMIJ/AIST

Examples for experiments that can be done at the T9 beam line

PoS(FNDA2006)033. Radiography Using Fission Neutrons. Thomas Bücherl. Christoph Lierse von Gostomski

Secondary Radiation and Shielding Design for Particle Therapy Facilities

SECTION 8 Part I Typical Questions

Recent developments in neutron metrology at the Institute for Radiological Protection and Nuclear Safety (IRSN)

Journal of Radiation Protection and Research

Unfolding of neutron spectra with an experimentally determined diamond detector response function

Neutron flux measurement using fast-neutron activation of 12 B and 12 N isotopes in hydrocarbonate scintillators

DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE

Investigation of nuclear Equation of State (EoS) is an attractive subject not only for nuclear

electrons out of, or ionize, material in their paths as they pass. Such radiation is known as

EEE4106Z Radiation Interactions & Detection

The Physics of Cosmic Rays

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements

This paper should be understood as an extended version of a talk given at the

Altitude Variation of cosmic-ray neutron energy spectrum and ambient dose equivalent at Mt.Fuji in Japan

Transcription:

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry F. Gutermuth *, S. Beceiro, H. Emling, G. Fehrenbacher, E. Kozlova, T. Radon, T. Aumann, T. Le Bleis, K. Boretzky, H. Johansson, O. Kiselev, H. Simon Gesellschaft für Schwerionenforschung Planckstraße 1, 64291 Darmstadt, Germany E-mail: f.gutermuth@gsi.de A neutron field is produced by bombarding a lead brick with 500 and 800 AMeV deuterons. The incoming deuterons break up into protons and neutrons mainly due to Coulomb interaction with the high-z target nuclei. The resulting charged products are deflected by a strong dipole magnet behind the target. The spectrum of the remaining neutrons in forward direction is measured from 0.1 MeV to 1.5 GeV with the time-of-flight method. Two neutron dosimeters are tested in the neutron field described above: The extended-range rem-counter WENDI-2 from Thermo-Electron Corporation reveals a fluence response of 778 psv cm 2 and 947 psv cm 2 for neutrons of energy 482 MeV and 789 MeV, respectively. A thermo-luminescence based extended range neutron dosimeter developed at GSI exhibits a fluence response of 623 psv cm 2 and 634 psv cm 2, respectively. A conventional version of the passive neutron dosimeter yields dose values which are lower by a factor of 5 demonstrating the advantage of the extended-range neutron dose equivalent meters. International Workshop on Fast Neutron Detectors University of Cape Town, South Africa April 3 6, 2006 * Speaker Copyright owned by the author(s) under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike Licence. http://pos.sissa.it

1. Introduction Neutrons constitute a major component in the radiation field outside the shielding of particle accelerators. While the neutrons percolate through shielding walls the spectrum is shifted towards lower energies. Nevertheless, when the energy of the primary beam particles approaches several tens of GeV per atomic mass unit, one has to expect a certain leakage of GeV neutrons. The Gesellschaft für Schwerionenforschung in Darmstadt (GSI) is preparing for a new Facility for Antiproton and Ion Research (FAIR). A large synchrotron for ions from protons up to Uranium-238 will allow beam intensities up to 10 12 particles per second at energies of several ten GeV per atomic mass unit. Thus, appropriate instrumentation will be needed in the future to survey the radiation levels over a substantial number of beam lines and experimental laboratory buildings. Recently neutron dose equivalent meters have been developed which show an enhanced response to neutrons above 20 MeV. The response function of these instruments is determined and optimized mainly with the help of Monte-Carlo simulations. Experimental data is very limited, especially for neutron energies above 200 MeV. Here we report on the production and characterization of a quasi-mono-energetic neutron field via break-up of deuterons with energy of 500 and 800 AMeV *. In these neutron fields two different neutron dosimeters are tested, both optimized for high energetic neutrons: First, the active counter WENDI-2 from Thermo- Electron Corporation and second, a passive neutron dosimeter based on the thermoluminescent technique built at GSI. deuteron beam 2 H + T - O - F START neutrons T - O - F STOP, X, Y lead target ionization chamber ALADIN (dipole magnet) 1 H + 2 H + T - O - F VETO flight path 15m Fig. 1: Scheme of the experimental set-up for time-of-flight neutron spectrometry. * AMeV: kinetic particle energy given per number of particle nucleons 2

2. Neutron spectrometry Deuterons with a kinetic energy of 500 and 800 AMeV are delivered from the GSI synchrotron in the experimental area Cave C where the Large Area Neutron Detector (LAND) is situated. Deuterons hit a 5 cm thick lead target and break up mainly via Coulomb interaction producing a field of mainly neutrons and protons. The protons and other charged particles are separated from the neutrons by a large dipole magnet. The neutrons with energies higher than 100 MeV are detected by LAND. The neutron yield below 100 MeV has been assessed to be smaller than 20 % in terms of fluence and smaller than 30 % in terms of dose. Fig. 1 shows a scheme of the experimental set-up LAND is a large time-of-flight detector wall built at GSI by a collaboration of different nuclear physics institutes. A detailed description of LAND can be found in [1]. Briefly, LAND is a multi-layer array of passive iron converters and active BC 408 plastic scintillators. The signals from 200 different scintillator paddles are processed (anti-coincidence to a veto detector for charged particles) and stored for later evaluation. The energy of the detected neutrons is derived from their time of flight over a 15 m long path. The spatial resolution perpendicular to the beam axis is 10 cm times 10 cm. The detection efficiency is higher than 90 % for neutrons above 500 MeV. The differential neutron yield as measured with LAND is shown in Fig. 2. The neutron spectrum resulting from 500 and 800 AMeV deuterons can be approximated by a Gaussian function with maximum at 482 and 789 MeV and full width at half maximum (FHWM) of 119 MeV and 203 MeV. The total neutron yield is 5.06 and 6.42 neutrons per 100 primary deuterons, respectively. 30 25 MAX at 482 MeV neutron spectrum (0-4 ) 500 MeV D 800 MeV D neutron yield per GeV and steradian 20 15 10 5 FWHM 119MeV MAX at 789 MeV FWHM 203MeV 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 energy / GeV Fig. 2: Neutron yield per primary deuteron as measure with LAND covering a polar angle from 0 to 4 as seen from the production target.. 3

3. Fast-neutron dosimetry The instrumentation used to survey radiation controlled areas is normally calibrated to the operational quantity H*(10) named ambient dose equivalent. The ambient dose equivalent from neutrons is highly depending on neutron energy. The official function to convert neutron fluence in neutron dose equivalent is given in [2] covering neutron energies from thermal to 200 MeV (see blue line in Fig. 3). For higher energies conversion coefficients have been calculated ([3],[4], see red and green lines in Fig. 3) but not officially approved, yet. As a result it is not clear which should be the desired value for the dose response of a neutron dose meter. Sannikov and Savitskaya [3] report a conversion factor of 420 and 580 psv cm 2 for 500 and 800 MeV neutrons, whereas Ferrari and Pellicioni account smaller values of 290 and 377 psv cm 2 for 500 MeV and 1 GeV neutrons, respectively. Apart from that conventional rem-counters are not suitable for neutron energies above 100 MeV as their response function declines steeply above 10 MeV. ambient dose equivalent / (psv cm 2 ) 1000 900 800 700 600 500 400 300 200 100 ICRP74 rem-counter Sannikov & Savitskaya [3] Ferrari & Pellicioni [4] 0 1.0E-04 1.0E-03 1.0E-02 1.0E-01 1.0E+00 1.0E+01 energy / GeV Fig. 3: Fluence to dose conversion function for neutron ambient dose equivalent (blue, red, green lines) and response function of a standard rem-counter (orange line) from 100 kev to 10 GeV. 3.1 Neutron dosimeter WENDI-2 According to the suggestions of Birattari et al. [5] the response of a standard rem-counter can be extended to higher energies by introducing a layer of heavy metal in the moderator. Fast neutrons are converted to neutrons of lower energies mainly through spallation reactions. Olsher et al. [6] developed an optimized rem-counter featuring a tungsten layer which is commercially available from Thermo-Electron Corporation. The instrument specified for neutron energies from thermal to 5 GeV was named WENDI-2, an abbreviation for Wide Area 4

Neutron Detection Instrument, second version. A cylindrically shaped Helium-3 proportional counter is situated in the centre of the moderating (polyethylene) and converting (tungsten) cylindrical structure. The instrument is calibrated using a 37 MBq Americium-241/Beryllium source, exhibiting a calibration factor of 3.2 10 9 counts per Sv. 3.2 Passive thermo-luminescent neutron dosimeter The passive neutron dosimeter consists of a 12 inch polyethylene sphere and a card carrying two crystals of each TLD600 and TLD700 thermo-luminescent material. TLD600 contains lithium-fluoride enriched in Li-6 thus being sensitive to neutrons via the (n,t) reaction. On the contrary, TLD700 contains practically no Li-6 and consequently is not sensitive to low energetic neutrons. The neutron dose is calculated from the difference of the signals in both crystal types. The advantages of the passive neutron dosimeter compared to rem-counters are the following: It does not show pile-up effects in highly pulsed fields and can operate for several months without the need of electricity or maintenance. Furthermore, the passive system is not sensitive to gamma rays and charged particles due to the subtraction of the TLD700 signal from the TLD600 one. The draw-backs are: no information on the time profile of the dose rate is provided and the detection limit is in the order of 100 µsv. Applying the concept of a heavy metal converter from the extended range rem-counters a modified version of the passive dosimeter has been developed. The response function of the device has been calculated using Monte-Carlo methods [7]. Both passive dosimeters are calibrated with a Americium-241/Beryllium source before using them in the fast-neutron field. The standard passive dosimeter will be referred to as STLD and the modified version as STLD(X) from here on. 3.3 Results The response of the neutron dosimeters described above is tested in the fast neutron field characterized in Sec. 2. The detectors are exposed in a distance of 5.36 m to the production target, where a neutron fluence of 21.9 cm -2 and 34.8 cm -2 is produced by 10 6 deuterons of 500 and 800 AMeV, respectively. For the passive dosimeter between 3 10 11 and 1.5 10 12 deuterons are applied to each card. The response of the active detector WENDI-2 can be derived from correlating the instrument counts with the single deuteron counts of the time-of-flight start detector. The resulting fluence responses for the three devices are shown in Tab. 1. The highest fluence responses are observed for the WENDI-2 instrument. The values are between a factor of 1.6 and 2.7 higher than the conversion factors found in [3] and [4]. For the modified passive dosimeter STLD(X) one obtains values which are higher by a factor of 1.1 to 2.1. The standard passive dosimeter STLD, which is mainly sensitive to neutrons with energy smaller than 20 MeV, shows response values which are by a factor of 2.4 to 5.2 lower than the conversion factors found in [3] and [4]. Furthermore, the response of the STLD is smaller by more than a factor of 5 than the response of the STLD(X), which is an indication that the dose contribution of low energy neutrons is comparatively small for both neutron fields under investigation. 5

Tab. 1: Fluence responses of the neutron dose equivalent meters. The devices have been calibrated to an Am-241/Be neutron source (h*(10)=391 psv cm 2 ). All values are given in psv cm 2 (standard deviation in parentheses). Neutron Energy WENDI-2 STLD(X) STLD 482 (119) MeV 778 (39) 623 (32) 121 (15) 789 (203) MeV 947 (47) 634 (56) 112 (32) 4. Conclusion A quasi-mono-energetic field of high energetic neutrons can efficiently be produced by break-up of deuterons and subsequent deflection of the residual charged particles. From 500 and 800 AMeV deuterons breaking up in a 5 cm thick lead brick, one obtains neutron yields in the range of 5 to 6 % (energy > 100 MeV). The resulting neutron spectra show distinct peaks close to the energy per nucleon of the primary deuterons with a full width at half maximum of approximately 25 % of the energy in the maximum. The neutron dose equivalent meter WENDI-2 overestimates the doses resulting from the neutron field described here by a factor of 1.6 to 2.7. The response of the passive dosimeter STLD(X) is lower than the one of the WENDI-2, but still higher than the conversion factors published in [3] and [4]. The observed over-response might originate from neutrons with energies below 100 MeV. However, their dose fraction was assessed to be smaller than 30 % and therefore can not account for the large varieties between the observed responses and the published conversion factors. With rising energy (> 1 GeV) one may expect an even more pronounced difference between the active and the passive dosimeter types. While active instruments count an increasing number of events caused by charged particles, the passive dosimeters STLD and STLD(X) are not affected because of the TLD600/700 subtraction process. References [1] Th. Blaich et al., A large area detector for high-energy neutrons, Nucl.Instr.Meth. A314 (1992) 136-154 [2] ICRP 74, ICRP Publication 74: Conversion coefficients for use in radiological protection against external radiation, International Commission on Radiation Protection (1997) [3] A. V. Sannikov and E. N. Savitskaya, Ambient dose equivalent conversion factors for high energy neutrons based on the ICRP 60 recommendations, Rad.Prot.Dosim. 70 (1997) 383-386 [4] A. Ferrari and M. Pellicioni, Fluence to dose equivalent conversion data and effective quality factors for high energy neutronsl, Rad.Prot.Dosim. 76 (1998) 215-224 [5] Birattari et al., An extended range rem-counter, Nucl.Inst.Meth. A297 (1990) 250-257 [6] R. H. Olsher et al., WENDI: An improved neutron rem meter, Health Phys. 79 (2000) 170-181 [7] F. Gutermuth et al., The response of various neutron dose meters considering the application at a high energy particle accelerator, Kerntechnik 68 (2003) 172-178 6