Apparatus to measure high-temperature thermal conductivity and thermoelectric power of small specimens

Similar documents
Measurement of small sample thermal conductivity by parallel thermal conductance technique

University of Colorado at Boulder

Thermal Interface Material Performance Measurement

Experimental Setup for the Measurement of Low Temperature Seebeck Coefficient of Single Crystal and Bulk Materials

Lecture 36: Temperatue Measurements

Laboratory 12: Three Thermodynamics Experiments

Specific Heat Measurement of High Temperature Thermal Insulations by Drop Calorimeter Method

Experimental Analysis of Wire Sandwiched Micro Heat Pipes

Automated Instrumentation for High-Temperature Seebeck. Coefficient Measurements

Temperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale

AE 3051, Lab #16. Investigation of the Ideal Gas State Equation. By: George P. Burdell. Group E3

Sensors and Actuators Sensors Physics

CRYOGENIC CONDUCTION COOLING TEST OF REMOVABLE PANEL MOCK-UP FOR ITER CRYOSTAT THERMAL SHIELD

qxbxg. That is, the heat rate within the object is everywhere constant. From Fourier s

Application of the Multi-current Transient Hot-Wire Technique for Absolute Measurements of the Thermal Conductivity of Glycols

THERMAL PROPERTY MEASUREMENT TECHNIQUES. AND SOME RESULTS FOR CdTe. H. F. Poppendiek D. J. Connelly R. K. Livett

Kapitza resistance and thermal conductivity of Mylar at superfluid helium temperature

Department of Mechanical Engineering ME 96. Free and Forced Convection Experiment. Revised: 25 April Introduction

Supplementary Information for On-chip cooling by superlattice based thin-film thermoelectrics

Nonstationary electrical charge distribution on the fused silica bifilar pendulum and its effect on the mechanical Q-factor

An experimental investigation of the thermal performance of an asymmetrical at plate heat pipe

Pin Fin Lab Report Example. Names. ME331 Lab

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

Enhanced Boiling Heat Transfer by using micropin-finned surfaces for Electronic Cooling

Heating surface thermal stabilization for pool boiling on porous coverings

Investigation of anomalous heat production in Ni-H systems

Thermal characteristic evaluation system to evaluate. heat transfer characteristics of mounted materials

Earlier Lecture. In the earlier lecture, we have seen non metallic sensors like Silicon diode, Cernox and Ruthenium Oxide.

PC based Temperature control system for Energy Spectrometer

Experiment 1. Measurement of Thermal Conductivity of a Metal (Brass) Bar

Corrections for Thermal Expansion in Thermal Conductivity Measurement of Insulations Using the High-Temperature Guarded Hot-Plate Method

Dynamic light scattering studies of some modes with wavevectors perpendicular to the twist axis in cholest erics

Total hemispherical emissivity of glass sheets with different thicknesses measured by a transient calorimetric technique

FACILITY FOR HEAT TRANSFER MEASUREMENT FOR SATELLITE PROGRAMME

Standard Test Method for Determination of the Thermal Conductivity of Anode Carbons by the Guarded Heat Flow Meter Technique 1

Low-temperature thermoelectrical power measurements using analogue subtraction

Supplementary Figure 1 Characterization of the synthesized BP crystal (a) Optical microscopic image of bulk BP (scale bar: 100 μm).

MEASUREMENT OF THE THERMAL DIFFUSIVITY OF BUILDING MATERIALS

Lab E3: The Wheatstone Bridge

PROBLEM 1.3. dt T1 T dx L 0.30 m

SUPPLEMENTARY INFORMATION

Simultaneous measurement of heat capacity, thermal conductivity, and thermal diffusivity

Available online at ScienceDirect. Physics Procedia 67 (2015 ) Calibration of a HTS based LOX 400mm level sensor

Ali S. Sharbuddin, International Journal of Advance Research, Ideas and Innovations in Technology.

COVENANT UNIVERSITY NIGERIA TUTORIAL KIT OMEGA SEMESTER PROGRAMME: MECHANICAL ENGINEERING

We are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1%

Slide 1. Temperatures Light (Optoelectronics) Magnetic Fields Strain Pressure Displacement and Rotation Acceleration Electronic Sensors

VHITAL-160 thermal model building and 3D thermal modeling carrying-out

High Temperature Measurement System Design for Thermoelectric Materials In Power Generation Application

True/False. Circle the correct answer. (1pt each, 7pts total) 3. Radiation doesn t occur in materials that are transparent such as gases.

Conjugate heat transfer from an electronic module package cooled by air in a rectangular duct

Experimental Investigation of Adiabatic Film Cooling Effectiveness and Heat Transfer Coefficients over a Gas Turbine Blade Leading Edge Configuration

A Comparison of Figureof Merit for Some Common ThermocouplesintheHighTemperatureRange

LaserComp, Inc., DESCRIPTION OF THE INSTRUMENT

DEVELOPMENT OF DROP WEIGHT IMPACT TEST MACHINE

Thermal Coatings for In-vacuum Radiation Cooling LIGO-T C R. Abbott, S. Waldman, Caltech 12 March, 2007

International Journal of Advance Engineering and Research Development

SIMULTANEOUS MEASUREMENT OF APPARENT THERMAL DIFFUSIVITY AND DISTORTION OF COMPOSITES AT HIGH TEMPERATURE

Thermal Performance Characterization of Embedded Pulsating Heat Pipe Radiators by Infrared Thermography

( ) PROBLEM C 10 C 1 L m 1 50 C m K W. , the inner surface temperature is. 30 W m K

EM375 MECHANICAL ENGINEERING EXPERIMENTATION THERMOCOUPLE LABORATORY

Thermal Characterization of Packaged RFIC, Modeled vs. Measured Junction to Ambient Thermal Resistance

LFA for Thermal Diffusivity and Conductivity of Metals, Ceramics and Polymers

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

AN INVESTIGATION OF ANOMALOUS THERMAL POWER GENERATION FROM A PROTON CONDUCTING OXIDE

Sensing, Computing, Actuating

Thermal Sensors and Actuators

Ion Beam Sources for Neutral Beam Injectors: studies and design for components active cooling and caesium ovens

Mineral-insulated thermocouples to DIN and EN

Analysis of Thermoelectric Generator Performance by Use of Simulations and Experiments

I. Introduction and Objectives

Observations of Thermo-Electric MHD Driven Flows in the SLiDE Apparatus

Thermal Cube. Custom-built Heat Flux Sensor

Segmented Power Generator Modules of Bi 2 Te 3 and ErAs:InGaAlAs Embedded with ErAs Nanoparticles

Department of Energy Science & Engineering, IIT Bombay, Mumbai, India. *Corresponding author: Tel: ,

Thermoelectric effect

IN-PLANE THERMAL CONDUCTIVITY OF FLAT PLATES UTILIZING THE FOURIER HEAT CONDUCTION LAW

Thermocouple Calibrations and Heat Transfer Coefficients

DIRECT RADIOMETRIC TECHNIQUES

TRANSIENT TECHNIQUES FOR MEASUREMENTS OF THERMAL PROPERTIES OF SOLIDS: DATA EVALUATION WITHIN OPTIMIZED TIME INTERVALS

Pakistan of Information Technology, Islamabad, Pakistan.

Comparison tests of cellular glass insulation for the development of cryogenic insulation standards

Critical parameters of


InterPACKICNMM

Sensitivity of surface resistance measurement of HTS thin films by cavity resonator, dielectric resonator and microstrip line resonator

Boiling Radial Flow in Fractures of Varying Wall Porosity

S.E. (Chemical) (Second Semester) EXAMINATION, 2011 HEAT TRANSFER (2008 PATTERN) Time : Three Hours Maximum Marks : 100

EFFECT OF XENON GAS AND FOILS ON A MULTI-FOIL INSULATION

STUDY OF HEAT TRANSFER MECHANISMS DURING THE LENS TM PROCESS

Experimental investigation on up-flow boiling of R1234yf in aluminum multi-port extruded tubes

Principles of Food and Bioprocess Engineering (FS 231) Problems on Heat Transfer

Introduction to Thermoelectric Materials and Devices

Chapter 1: 20, 23, 35, 41, 68, 71, 76, 77, 80, 85, 90, 101, 103 and 104.

Veerapong Kanchanawongkul*

MERGING OF SHEET PLUMES IN TURBULENT CONVECTION

Technical Notes. Introduction. PCB (printed circuit board) Design. Issue 1 January 2010

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

Thermionic Emission. A. Goode Student and A. Finicky Professor. Abstract. φ 2 kt

a. Fourier s law pertains to conductive heat transfer. A one-dimensional form of this law is below. Units are given in brackets.

Transcription:

Apparatus to measure high-temperature thermal conductivity and thermoelectric power of small specimens T. Dasgupta and A. M. Umarji a Materials Research Centre, Indian Institute of Science, Bangalore-560012, India A thermal conductivity apparatus based on the parallel heat-flow technique has been fabricated to measure the thermal conductivity of small specimens. The steady-state method is used to measure samples of 1 1 10 mm 3 dimension in the temperature range of room temperature 700 K. The details of instrument fabrication, the method of calibration, and typical measurements on test samples are described. The apparatus can also be used to measure the Seebeck coefficient in the same temperature range. As an example we report the thermal properties of CrSi 2, which is a potential candidate for high-temperature thermoelectric applications. I. INTRODUCTION In recent years there has been a considerable increase in the effort to identify new and efficient thermoelectric materials that would supercede the existing materials, especially at higher temperatures. For a material to be considered as a good thermoelectric, it must have a large figure of merit defined as Z=S 2 /K, where S is the Seebeck coefficient, is the electrical conductivity, and K is the thermal conductivity. Amongst these three properties, K is most difficult to measure experimentally. For relatively big samples there are well-established, steady-state and nonequilibrium techniques, 1 3 but small samples pose difficulties associated with heat flow during the measurement. The most widely used commercial method is based on the laser flash technique. 3 But the method has limitations for high K materials and at elevated temperatures. Recently Tritt et al. 4 described a method known as the parallel conductance technique, which is suitable for small samples. They measured the K value for a few standard materials from 10 K to room temperature. The problems associated with heat loss increases with temperature and measurement of K at high temperatures is very challenging. In this paper we describe an apparatus, which can measure the K value of small samples 1 mm 1 mm 10 mm based on the parallel conductance technique from room temperature to 700 K. The same apparatus can also be used to measure the Seebeck coefficient of the material without disturbing the physical mounting. The paper is divided into three sections. In Sec. II, the principle of measurement is explained, followed by a description of the apparatus in Sec. III. Section IV contains results obtained for standard samples. a Author to whom correspondence should be addressed; electronic mail: umarji@mrc.iisc.ernet.in II. PRINCIPLE OF MEASUREMENT A. Thermal conductivity measurement Under steady-state conditions the K of a material is defined as K = Q S /A / T/ L, where Q S is the amount of heat passing per unit time through a cross-section area A, and causing a temperature difference, T, over a distance L. Q S /A is therefore the heat flux that is causing the temperature gradient, T/ L. The measurement of thermal conductivity therefore involves the measurement of the heat flux and temperature difference. The difficulty of the measurement is always associated with heat-flux measurements because of the undefined amount of heat loss accompanying the heat transfer process. A schematic representation of a heater block sink assembly is depicted without and with the sample in Figs. 1 a and 1 b, respectively. It is a steady-state method where temperature differences between the source and sink are measured as a function of time and the sample holder and sample act as two parallel heat conducting paths. First, the characteristics of the sample holder are measured, which determines the baseline or background thermal conduction and losses associated with the sample stage. The second step consists of attaching the sample and measuring the new characteristics of the assembly. By subtraction of the background thermal conduction, the thermal conductivity of the sample is calculated. When heat is supplied to the heat source of Fig. 1 a, a part of it is utilized in raising the temperature of the source block while the remaining is dissipated to the surrounding through the cell assembly. At a constant input power, the temperature initially rises linearly as a function of time as the heat loss is negligible. Subsequently, as the temperature increases, the heat loss also increases and at equilibrium both the input power and the heat loss per unit time become equal. The temperature of the heater block at equilibrium reaches a constant value, which is taken as the equilibrium T. Thus 1

FIG. 1. Schematic diagram of the different heat-flow paths for a empty and b with sample. the total heat input is a summation of the heat gained by the source and the heat dissipated to the surrounding. When a sample is placed between the source sink assembly Fig. 1 b, a part of the heat passes through the sample, which depends on the K of the sample. Thus the heat input is distributed among the heating source block, the heat transfer through the sample, and heat loss through the cell assembly. If the same heat input is maintained in both the cases and assuming that the heat dissipated in both is the same, the K of the sample can be found. This is known as the parallel conductance method. Thus mathematically, it can be written as follows: Without any sample as in Fig 1 a, P E = x T E + H ; 2 III. APPARATUS DESCRIPTION A schematic diagram of the apparatus is shown in Fig. 2. The specimen 1 is placed inside a copper sample holder assembly. The sample holder consists of two copper 10 mm cylindrical blocks 2 and 3 into which type-k thermocouples 4 of 40 Imperial standard wire gauge swg are embedded. The thermocouples are in good thermal contact but electrically insulated from the blocks using silica sheaths. The crucial heat source copper block 2 is a 10-mmdiameter solid cylinder weighing 10 g on which 20 of kanthal 32 swg heater is wound. The heater is electrically insulated from the copper block using high-temperature cement. The bottom heat-sink block 3 is welded to the copper with the sample as in Fig 1 b, P S = x T S + H + Q S, 3 where the subscripts E stand for an empty run i.e., without the sample, S stand for a run with the sample. P E and P S is the power supplied, T E is the equilibrium temperature change in the heater block without the sample, and T S is the equilibrium temperature change with the sample. H is the heat loss per unit time. x is the heater constant that depends on the specific heat, mass, etc. of the heater block. Q S is the heat flux through the sample. In Eq. 2 there are two unknowns, x and H. The value of the heater constant x has been obtained from the plot of T E vs P. This is explained in Sec. IV. The other unknown, H, can be assumed to be the same for a constant input power i.e., P E = P S. If the same power is passed for both and assuming H is the same, Eqs. 2 and 3 can be equated to the following: x T E + H = x T S + H + Q S. 4 Rearranging Eq. 4, Q S = x T E T S. 5 Thus, once the heat flux through the sample Q S is known, the K of the material can be calculated using Eq. 1. FIG. 2. Schematic diagram of the thermal conductivity setup see text for explanations.

FIG. 3. Calibration curve of heater power vs T E. The slope of the curve as T E tends to zero gives the value of x. base plate. The heater block is supported by a thin walled quartz tube 5 of 10 mm diam and 100 mm length. This is provided with a spring loading 6 mounted on a stainlesssteel support frame 7. The quartz tube is used to minimize the conductive heat loss from the heater block to the support frame while the spring loading ensures proper contact between the sample and the holder assembly. The thermoelectric power is measured using two additional copper wires connected to the heater and base, respectively. The electrical insulation of the thermocouple wires prevents any electrical interference with these copper wires. The central sample holder is placed inside a kanthal furnace 8 15 mm inner diameter i. d. and 100 mm length the height of which can be adjusted. Height adjustment of the furnace 9 can be done from the bottom in between a run without disturbing the sample. This has been done to minimize the temperature difference of the top and bottom thermocouple at the beginning of an experiment and also for easy loading and removal of the samples. A chromel-alumel thermocouple is embedded along with the heating element to control the furnace temperature through a multistep programmable temperature controller Toho TTM-304. A vacuum-tight glass chamber 10 80 mm diam and 400 mm height encapsulates the whole arrangement to minimize convective heat exchange. The thermal contact between the heat source block and the sample is made or broken by a lifting mechanism 11. Using this, the central hot part of the sample holder is lifted by 0.5 mm under vacuum conditions. A. Sample preparation It is necessary to make samples of proper dimensions as the calculation of K depends on the sample dimensions. The samples used for measurement were either cylindrical or rectangular, and cut using a diamond cutter. In general, it was necessary to have an aspect ratio of 10 for successful experimentation. Extra care was taken to make the top and bottom surfaces parallel. Also, to ensure proper thermal contact with the heater and base, high-temperature silver paste was used as a contact material. B. Data collection The heater is energized by a Keithley 228 constant current source. The thermocouple temperatures are measured by FIG. 4. Plot of T vs time obtained for the different standard samples at room temperature. a Keithley DMM 2700 scanner as a function of time. The current source and the multimeter outputs are interfaced to a computer using an IEEE-488 interface and the data collected in the hard disk using the commercial software XLINX 2700. The measurement consists of the following steps. Initially, the glass chamber is evacuated until a vacuum of 10 3 Torr is reached. The furnace is then heated to the desired temperature and allowed to stabilize. The furnace height is adjusted such that the top and bottom thermocouples read the same value. The constant current source is then put on and the change in temperature is measured as a function of time. Once the equilibrium T S value is reached, the heater is lifted slightly using the lifting mechanism and the change in temperature T E is again measured. It is important to repeat the empty run at each temperature as the heat loss keeps changing with temperature and also the furnace position. IV. RESULTS AND DISCUSSION A. Measurement of heater constant x To find the value of x, the T E of the heater is measured for different input power as shown in Fig. 3. x is found from the slope of the curve as T E tends to zero. In the present case x is 0.01491 W/ K at room temperature. For hightemperature measurements the value of x has been modified depending on the change in specific heat of the heater block material copper. A correction factor in T E has to be done for each temperature due to a change in the value of x. Inthe actual experiments T was always maintained in the linear region of the calibration curve. TABLE I. Thermal conductivity measurements of test materials at room temperature. Sample Height cm Cross-sectional area cm 2 Thermal conductivity Wcm 1 K 1 ±5% Observed Reported Fused silica 1.8 0.125 0.013 0.014 SS 304 1.21 0.031 0.157 0.160 Sapphire 1.38 0.078 0.328 0.320

FIG. 5. Thermal conductivity of fused silica as a function of temperature. Plots 5, 7 are the experimental values taken from the literature. B. Room-temperature thermal conductivity measurements Test runs were carried out on standard samples and the data compared with literature data. Room-temperature thermal conductivity was measured for three different standard materials namely, sapphire, SS 304 and fused silica, which have wide variation in K values. The increase in temperature as a function of time for the different samples is shown in Fig. 4. The saturation T could be achieved in sample dimensions and the observed and reported thermal conductivity values are given in Table I. It is seen from the data that a good match is observed with the data reported in the literature. 5 Measurements on each sample were carried out three times and the deviation in the K value was within ±5% of the absolute value. C. Measurement of K as a function of temperature Thermal conductivity was also measured as a function of temperature for fused silica as shown in Fig. 5. Fused silica has a very low thermal conductivity and the applicability of this technique for measurement of low K materials is demonstrated. The calculated data is compared with the data reported in the literature. 5 8 Table II shows the scatter in the room-temperature data reported in the literature by various groups. This is because of the very low K value of fused silica. Our result lies well within the scatter region reported in the literature. V. SEEBECK COEFFICIENT MEASUREMENT FIG. 6. Seebeck coefficient of CrSi 2 as a function of temperature. The above-described apparatus can be used easily to measure the Seebeck coefficient at respective temperatures. For measuring the Seebeck coefficient S, the sample is placed between the source-sink assembly as shown in Fig 1 b. A constant input power is applied and the corresponding rise in temperature T S and voltage developed across the sample E is measured. The Seebeck coefficient S can then be calculated using the equation S = E, 6 T S where E is the voltage developed across the sample and T S is the corresponding rise in sample temperature. The absolute S values of the samples were obtained by applying the necessary correction due to the Seebeck coefficient of copper. VI. MEASUREMENT OF S AND K VALUES FOR CrSi 2 CrSi 2 has been used as a test thermoelectric material. The variation of S and K of the hot pressed sample with temperature is shown in Figs. 6 and 7, respectively. The sample was prepared by direct hot uniaxial pressing of elemental chromium and silicon powders. The thermoelectric data for CrSi 2 as a function of temperature matches with the reported data by Nishida et al. 9 Pure CrSi 2 is p type, as seen from the data. The S value increases initially with a maximum between 500 and 600 K. The K was also measured as a function of temperature and is seen to decrease with increasing temperature. A detailed study on the thermoelectric properties of CrSi 2 can be found elsewhere. 10 In summary, we have designed and fabricated a setup to measure the thermal conductivity and thermoelectric power of small samples from room temperature to 700 K. Equations have been developed based on the parallel conductance tech- TABLE II. Thermal conductivity data of fused silica at room temperature showing the acceptable scatter in data. Reference 5 6 7 8 Our data Temperature K 300 317 304 350 300 K Wm 1 K 1 1.38 1.51 1.37 1.19 1.31 FIG. 7. Thermal conductivity of CrSi 2 as a function of temperature.

nique to measure K. The K value measurements on test samples like sapphire, fused quartz, and stainless steel gave satisfactory results. The use of this apparatus for measuring both K and S values up to 700 K on CrSi 2 has been demonstrated. 1 M. J. Laubitz, Thermal Conductivity Academic, London, 1969, Vol. 1, Chap.3,p.111. 2 D. R. Flynn, Mechanical and thermal properties of ceramics, NBS Spec. Publication 303, p. 63 1969. 3 W. J. Parker, R. J. Jenkins, C. P. Butler, and G. L. Abbot, J. Appl. Phys. 32, 1679 1961. 4 B. M. Zawilski, R. T. Littleton, and T. M. Tritt, Rev. Sci. Instrum. 72, 1770 2001. 5 Thermophysical Properties of Matter, Vols. 1&2ofThermal Conductivity IFI/Plenum, New York, 1970. 6 J. H. Koenig, Rutgers Univ. N.J. Ceramic Research Station Progress Report, Vol. 2, 1 1953. 7 I. K. Kamilov, Sov. Phys. Solid State 4, 1693 1963. 8 K. L. Wray and T. J. Connolly, J. Appl. Phys. 30, 1702 1959. 9 I. Nishida and T. Sakata, J. Phys. Chem. Solids 39, 499 1978. 10 T Dasgupta and A. M. Umarji, communicated to Materials Research Bulletin 2005.