AppNote 1/2013. Determination of Barbiturates and 11-Nor-9-carboxy- 9 -THC in Urine using Automated Disposable Pipette Extraction (DPX) and LC/MS/MS

Similar documents
AppNote 8/2012. Rapid Automated Extraction and Confi rmation of Buprenorphine and Norbuprenorphine in Urine by DPX-LC/MS/MS KEYWORDS ABSTRACT

AppNote 2/2013. A High Throughput Automated Sample Preparation and Analysis Workfl ow for Comprehensive Toxicology Urine Screenings using LC/MS/MS

AppNote 7/2014. Automating Liquid-Liquid Extractions using a Bench-top Workstation KEYWORDS ABSTRACT. Sample Preparation, Lab Automation, LC/MS/MS

AppNote 5/2008. Automation of Sample Preparation Steps using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

AppNote 6/2014. Automated Online Desorption and Analysis of DNPH Derivatives of Airborne Aldehydes and Ketones KEYWORDS ABSTRACT

AppNote 8/2008. A New Purge Tool for Use with Automated Headspace Analysis KEYWORDS ABSTRACT

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

AppNote 14/2012. Thermal Gravimetric Analysis/Mass Spectrometry Simulation using the GERSTEL Automated Pyrolyzer KEYWORDS ABSTRACT

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

AppNote 3/2007 KEYWORDS INTRODUCTION. Beverage, Detector, Headspace, Quality Control

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances KEYWORDS

AppNote 5/2002. Characterization of a Mass Spectrometer based Electronic Nose for Routine Quality Control Measurements of Flavors KEYWORDS ABSTRACT

AppNote 4/1998. New Developments in Multi-Dimensional Capillary Gas Chromatography KEYWORDS ABSTRACT

AppNote 2/2002. Discrimination of Soft Drinks using a Chemical Sensor and Principal Component Analysis KEYWORDS ABSTRACT

AppNote 3/2014. Analysis of Aroma Compounds in Edible Oils by Direct Thermal Desorption GC/MS using Slitted Microvials KEYWORDS ABSTRACT

AromaOffi ce: Application of a Novel Linear Retention Indices Database to a Complex Hop Essential Oil

AppNote 15/2012. Large Volume Injection with Solvent Venting - Application to Trace Detection of Analytes in Water INTRODUCTION

TechNote 5/2002 KEYWORDS ABSTRACT. SBSE, Twister, Thermal Desorption

AppNote 6/2010. part a. Alternative Procedure for Extraction and Analysis of PAHs in Seafood by QuEChERS-SBSE-GC-MS KEYWORDS ABSTRACT

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

AppNote 2/2007. Automated Solid Phase Microextraction using the GERSTEL MPS Prepstation and MAESTRO Software KEYWORDS ABSTRACT

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to En vi ron men tal Aqueous Samples INTRODUCTION

AppNote 8/2002 KEYWORDS ABSTRACT. Electronic Nose, Fast GC, Headspace Analysis, Chemometrics,

Off-gassing of Rubber Particles Used for Athletic Fields using Automated Dynamic Headspace Sampling

AppNote 2/1996. Andreas Hoffmann Gerstel GmbH & Co.KG, Eberhard-Gerstel-Platz 1, D Mülheim an der Ruhr, Germany

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

AppNote 1/2007. Automated Dynamic Headspace Sampling using Replaceable Sorbent Traps KEYWORDS ABSTRACT

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

AppNote 10/2012. Large Volume Injection with GC/MS Multi Column Switching for Direct Determination of Ethyl Carbamate in Alcoholic Beverages KEYWORDS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

AppNote 1/2004. Use of a Mass Spectral Based Chemical Sensor to Discriminate Food and Beverage Samples: Olive Oils and Wine as Examples KEYWORDS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

AppNote 7/2002. Classifi cation of Food and Flavor Samples using a Chemical Sensor KEYWORDS ABSTRACT

AppNote 5/2001. The Analysis of the Bitter and Other Flavour Compounds in Beer and Wort by Stir Bar Sorptive Extraction (SBSE) Followed by HPLC

Thermal Desorption Unit TDU

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

AppNote 2/2001. Stir Bar Sorptive Extraction: Enhancing Selectivity of the PDMS Phase KEYWORDS ABSTRACT

AppNote 2/2004. Analysis of Packaging Materials using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

Analysis of Food Samples using Thin Film Solid Phase Microextraction (TF-SPME) and Thermal Desorption GC/MS

AppNote 3/2012. Rapid Automated Screening of Extractable Compounds in Materials for Food Packaging, Medical or Technical Purposes KEYWORDS ABSTRACT

AppNote 8/2009. Fragrance Profi ling of Consumer Products using a Fully Automated Dynamic Headspace System KEYWORDS ABSTRACT

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

AppNote 11/2012. Fully Automated Analysis of Doping Substances in Equine Urine by Disposable Pipette Extraction (DPX) coupled to GC/QqQ-MS KEYWORDS

GERSTEL MultiPurpose Sampler MPS. Versatile autosampler and sample preparation robot

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

AppNote 2/2012. Determining Phenolic Compounds in Whisky using Direct Large Volume Injection and Stir Bar Sorptive Extraction KEYWORDS ABSTRACT

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

AppNote 1/2011 KEYWORDS ABSTRACT

AppNote 4/2005. A Selectable Single or Multidimensional. Fraction Collection and Dual Detection for Trace Analysis of Complex Samples KEYWORDS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

AppNote 6/2012. Flavor and Fragrance Analysis of Consumer Products - Dynamic Headspace Compared to Some Traditional Analysis Approaches KEYWORDS

AppNote 8/2000. Volatile Organic Compounds from Adhesives and their Contribution to Indoor Air Problems KEY WORDS ABSTRACT

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Abstract. Experimental Sample Preparation

AppNote 2/1998. Direct Injection of Distilled Spirits with PTV Matrix Removal: The Perfect Splitless Injection? KEYWORDS INTRODUCTION

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Detection of 9-tetrahydrocannabinol ( 9-THC) in human urine by Solid Phase Extraction and HPLC.

Determination of Cannabinoids (THC) in Biological Samples

AppNote 9/2012. Solid Phase Micro-Extraction (SPME) Coupled with Selectable 1 D/ 2 D GC-MS for the Determination of Food Product Flavors KEYWORDS

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

AppNote 12/2008 KEYWORDS ABSTRACT

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

LC-MS/MS Analysis of Phytocannabinoids and their

Psychoactive Drugs Urine LC-MS/MS Analysis Kit User Manual

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

AppNote 7/2013. Comprehensive Automation of the SPE- GC/MS Analysis of Opioids, Cocaine and Metabolites from Serum and Other Matrices KEYWORDS

KEYWORDS ABSTRACT AppNote 6/2002

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

AppNote 8/2011 KEYWORDS ABSTRACT INTRODUCTION

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

AppNote 12/2012. Comparison of EG-Silicone-SBSE and Derivatization-PDMS-SBSE for the Analysis of Phenolic Compounds and Off-fl avors in Water KEYWORDS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Figure 1. The supported liquid extraction process using the ISOLUTE SLE plate (single well shown)

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

NORBUPRENORPHINE (Buprenorphine s Metabolite ) BUPRENORPHINE in urine by GC/MS Code GC Method of Confirmation by GC-MS

Definitive EtG/EtS LC-MS/MS Analysis:

AppNote 7/2005. Detection of Packaging Emissions using a Flexible Headspace Sampler Combined with a Multi Sensor System and a Separation Unit KEYWORDS

Confirmation and Quantitation of THC in Oral Fluid Using an MRM Method on a GC-Triple Quadrupole MS

Benzodiazepines and Barbiturates

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Transcription:

AppNote 1/2013 Determination of Barbiturates and 11-Nor-9-carboxy- 9 -THC in Urine using Automated Disposable Pipette Extraction (DPX) and LC/MS/MS Fred D. Foster, Oscar G. Cabrices, John R. Stuff, Edward A. Pfannkoch GERSTEL, Inc., 701 Digital Dr. Suite J, Linthicum, MD 21090, USA William E. Brewer Department of Chemistry and Biochemistry, University of South Carolina, 631 Sumter St. Columbia, SC 29208, USA KEYWORDS DPX, LC/MS/MS, Sample Preparation, High Throughput Lab Automation ABSTRACT This work demonstrates the use of disposable pipette extraction (DPX) as a fast and automated sample preparation technique for the determination of barbiturates and 11-nor-9- carboxy- 9 -THC (COOH-THC) in urine. Using a GERSTEL MultiPurpose Sampler (MPS) with DPX option coupled to an Agilent 6460 LC-MS/MS instrument, 8 barbiturates and COOH-THC were extracted and their concentrations determined. The resulting average cycle time of 7 min/ sample, including just-in-time sample preparation, enabled high throughput screening. Validation results show that the automated DPX-LC/ MS/MS screening method provides adequate sensitivity for all analytes and corresponding internal standards that were monitored. Lower limits of quantitation (LLOQ) were found to be 100 ng/ml for the barbiturates and 10 ng/ml for COOH-THC and % CVs were below 10 % in most cases.

INTRODUCTION The continuously growing quantity of pain management drugs used has increased the demand from toxicology laboratories for more reliable solutions to monitor compliance in connection with substance abuse and/or diversion. Marijuana is one of the most commonly used illegal recreational drugs in the world; 11-nor-9- carboxy- 9 -THC (COOH-THC) is the metabolite of tetrahydrocannabinol ( 9 -THC), which is mainly determined in biological samples due to its long halflife in the body. Barbiturates, on the other hand, are a class of antidepressants whose abuse has been reported to be steadily increasing. Both COOH-THC and the barbiturate class of compounds are difficult to analyze in LC-MS/MS comprehensive drug testing panels due to their different ionization properties and a separate method was therefore developed for these. Conventional sample preparation techniques used to determine the concentration levels of these drugs in biological fluids involve liquid-liquid- or solid-phase extraction (SPE). However, these extraction methods often require relatively large volumes of solvent leading to an increase in the time needed for sample processing; increased cost per sample and higher limits of detection. Alternative SPE approaches involve the removal of the sample matrix from the analytes of interest by selectively retaining potential interferences on a chemically modified sorbent. The main advantage of this approach is that it is fast and does not involve separate wash or elution steps, enabling screening for multiple drugs that have different chemical properties. Disposable Pipette Extraction (DPX) is a dispersive solid-phase extraction technique, which relies on a disposable pipette tip with sorbent loosely contained inside it enabling highly efficient mixing with sample solutions. DPX was developed as an alternative to traditional SPE, combining efficient and rapid extraction with significantly reduced solvent consumption. The extractions can be fully automated and the extracts injected directly to the LC-MS/MS analysis system. This study details a rapid sample cleanup method for the determination of 8 barbiturates (Amobarbital, Butabarbital, Butalbital, Hexobarbital, Methohexital, Pentobarbital, Phenobarbital, Secobarbital) and COOH-THC in hydrolyzed urine samples using DPX- LC-MS/MS. The DPX extraction process removes potential matrix interferences leaving a cleaned sample extract for injection [1-2]. The extraction of the Barbiturates and COOH-THC is based on the DPX-RP- S extraction method described in an earlier Application Note detailing monitoring of 49 Pain Management drug compounds (Gerstel AppNote 2012/01). The only difference is that, for the Barbiturates and COOH-THC method, the final extract can be directly injected into the LC-MS/MS system following the DPX step without being diluted. The analysis of urine for Barbiturates and COOH-THC, however, required different LC and MS parameters altogether. Therefore, a second LC- MS/MS method was developed, which is described in this work. The reversed phase sorbent (DPX-RP-S) used in the method enables the removal of salts and other matrix components present in urine, resulting in reduced matrix effects. The sorbent was chosen to extract the matrix without binding or absorbing the analytes of interest providing high recoveries. The schematic for the DPX Cleanup procedure is shown in Figure 2. Since the extraction time (3 min) is less than the total LC-MS/MS acquisition time (6.5 min), the extraction can be performed in PrepAhead mode during the chromatographic analysis of the previous sample without adding to the overall analysis time. This means that high throughput is achieved while processing each sample just in time ensuring that all samples receive uniform treatment ensuring best possible quality of results.. EXPERIMENTAL Materials. All stock solutions for the compounds listed in Table 1 were purchased from Cerilliant. An intermediate stock solution containing all analytes was prepared by combining the analyte stock solutions and adding acetonitrile to reach the concentrations required. Deuterated analogues, -butalbital, -methohexital, -pentobarbital, -phenobarbital, -secobarbital, and d 9-11-nor-9-carboxy- 9 -THC, were purchased from Cerilliant. Table 1 shows which deuterated internal standard was to quantitate individual analytes. The high concentration calibration standard and intermediate QC urine samples were prepared by making appropriate dilutions of the combined intermediate analyte stock solution using analyte free urine to give the high calibration concentrations listed in Table 1. Calibration standards were then prepared as dilutions from the high concentration sample AN/2013/01-2

at concentrations of 2000/200, 1000/100, 800/80, 500/50, 200/20, and 100/10 ng/ml (Barbiturates/ COOH-THC). The high and low QC samples were prepared as dilutions from the intermediate QC urine sample at concentrations of 800/80 and 200/20 ng/ml (Barbiturates/COOH-THC). β-glucuronidase, Type-2, from Helix pomatia, (cat.#g0876-5ml) was purchased from Sigma-Aldrich. Fresh urine was obtained from a male volunteer. All other reagents and solvents used were reagent grade. Instrumentation. All automated DPX PrepSequences were processed using a GERSTEL MultiPurpose Sampler in dual-head configuration (MPS XL) with DPX Option as shown in Figure 1. All analyses were performed using an Agilent 1290 HPLC with a Poroshell 120, EC-C18 column (3.0 x 50 mm, 2.7 μm), an Agilent 6460 Triple Quadrupole Mass Spectrometer with Jet stream electrospray source and GERSTEL MPS XL autosampler configured with an Active Washstation. Sample injections were made using a 6 port (0.25 mm) Cheminert C2V injection valve outfitted with a 5 μl stainless steel sample loop. Figure 2 shows a graphical representation of the general DPX cleanup process. The automated DPX extraction used for this method consisted of the following steps: Automated DPX Prep Sequence - DPX Cleanup procedure: 1. Aspirate 750 μl of 100 % acetonitrile from the fast solvent delivery station using the 2.5 ml DPX syringe. 2. Pick up a new DPX tip (DPX-RP-S) located within the tray. 3. Ad00 μl of 100 % acetonitrile through the DPX tip, into the urine sample located on the MPS sample tray. 4. Wait for 6 seconds to allow acetonitrile to completely wet the DPX sorbent. 5. Aspirate the entire sample followed by 1400 μl of air into DPX tip. 6. After equilibrating for 5 seconds, dispense the contents of the DPX tip, as well as the remaining acetonitrile found within the DPX syringe, back into the original shell vial in the tray. 7. Move the DPX tip to the PipWaste position and dispose of the DPX tip. 8. Transfer 100 μl the upper liquid layer located within the original shell vial, into a clean, empty, capped autosampler vial with septum located on a VT98 tray. 9. Inject 10 μl of the extract into the HPLC injection valve. Figure 1. MultiPurpose Sampler (MPS XL) with GERSTEL DPX Option for high throughput barbiturates/cooh-thc screening. Sample pretreatment. Hydrolysis of urine was performed by combining 2 ml of urine, 150 μl of the working internal standard solution, 100 μl of β-glucuronidase, an00 μl of 0.66 M acetate buffer, ph 4, vortex mixing for 30 seconds, and then incubating at 55 C for 2 hours. Aliquots of 260 μl of hydrolyzed urine samples were added into clean shell vials for automated cleanup and injection. Figure 2. Graphical representation of the automated DPX urine cleanup process. AN/2013/01-3

Analysis conditions LC. Pump: gradient (650 bar), flowrate = 0.5 ml/ min Mobile Phase: A - 5 mm ammonium acetate, with 0.05 % ammonium hydroxide B 0.05 % ammonium hydroxide in methanol Gradient: Initial 10 % B 0.5 min 10 % B 4.0 min 40 % B 5.6 min 100 % B 5.8 min 100 % B 6.0 min 10 % B 6.5 min 10 % B Injection volume: 5 μl (loop over-fill technique) Column temperature: 55 C Analysis conditions MS. Operation: electrospray negative mode Gas temperature: 350 C Gas flow (N 2 ): 5 L/min Nebulizer pressure: 50 psi Sheath Gas Temp: 400 C Sheath Gas Flow: 11 L/min Capillary voltage: -4000 V Nozzle voltage: -500 V Delta EMV: 500 V DeltaRT (s): 60 s A total of 27 MRM transitions were monitored in a 5 minute analytical window followed by a column regeneration time of 1.5 minutes. A retention time window value of 60 seconds was used for each negative ion transition being monitored during the course of the dynamic MRM experiment. Detailed mass spectrometric acquisition parameters are available upon request. RESULTS AND DISCUSSION Figure 3 shows representative dynamic MRM chromatograms for the barbiturates, COOH-THC, and internal standards, from a hydrolyzed urine sample spiked sample at the minimum reporting limit (MRL) concentrations after the automated DPX cleanup procedure. Figure 3. Overlaid chromatograms for all dynamic MRM transitions from an extracted urine sample at the MRL. AN/2013/01-4

Table 1 lists the column retention times, concentrations for the highest calibration standard, MRLs and LLOQs for the barbiturates and COOH-THC analyzed in this screening assay. Table 1. Retention times, high calibration standard concentrations, MRLs and LOQs for all barbiturates and COOH-THC. Analyte Precursor Ion [m/z] Precursor Ion [m/z] Fragmentor Voltage [V] Collison Energy [ev] Retention Time [min] High Std Conc Amobarbital 1 225.2 42.1 182.1 90 90 17 7 4.74 2000 100 100 Butabarbital 2 211.2 42.1 168 90 90 18 7 3.68 2000 100 100 Butalbital 2 223.2 42.1 180 85 85 18 5 3.63 2000 100 100 -Butalbital 228.2 42.1 185 85 85 15 7 3.61 - - - 11-Nor-9-carboxy- 9 -THC 6 343.2 299.2 245.2 135 135 20 30 5.71 200 10.0 10.0 d 9-11-Nor-9-carboxy- 9 -THC 6 352.2 308.2 254.2 135 135 20 30 5.71 - - - Hexobarbital 1 235.2 42.1-80 - 20-4.69 2000 100 100 Methohexital 5 261.2 42.1-90 - 20-5.40 2000 100 100 -Methohexital 266.2 42.1-80 - 10-5.42 - - - Pentobarbital 1 225.2 42.0 182 80 80 20 7 4.74 2000 100 100 -Pentobarbital 230.2 42.1 187 90 90 10 7 4.90 - - - Phenobarbital 3 231.1 42.1 85 85 85 10 5 2.12 2000 100 100 -Phenobarbital 236.2 42.1 193 90 90 10 2 2.10 - - - Secobarbital 4 237.2 42.1 194.2 85 85 20 5 5.12 2000 100 100 -Secobarbital 242.2 42.1 199 85 85 20 5 5.12 - - Internal Standards: 1: -Pentobarbital, 2: -Butalbital, 3: -Phenobarbital, 4: -Secobarbital, 5: -Methohexital, 6: d 9-11-Nor-9-carboxy- 9 -THC MRL LOQ AN/2013/01-5

Representative calibration curves are shown in Figure 4. Regression analysis for all barbiturates and COOH- THC resulted in R 2 values of 0.99 or greater. The DPX automated sample cleanup time was 3 min/sample; the total cycle time per sample for the extraction process and injection was 7 min/sample, fitting with the just in time sample preparation strategy available using the MAESTRO software PrepAhead function for increased throughput. Using this automated procedure for extraction and analysis, more than 200 samples can be processed per day. The accuracy and precision of the method was determined for all barbiturates analyzed and for COOH-THC by extracting replicate (n=5) QC samples at 800/80 ng/ml and 200/20 ng/ml concentrations (Barbiturates/COOH-THC). Table 2 shows the resulting accuracy and precision data for all compounds analyzed. Accuracy data averaged 102 % (range: 93.4 % - 113 %) and precision data (% CV) average.55 % (range: 2.00 % -10.4 %) for all compounds analyzed. Figure 4. Representative Calibration Curves: Butabarbital, Phenobarbital, and COOH-THC. Table 2. Extracted QC sample % Accuracies and % CVs. Analyte QCL Avg Accuracy [n=5] CV [%] QCH Avg Accuracy [n=5] Amobarbital 200 98.0 9.68 800 112 2.80 Butabarbital 200 93.4 4.50 800 104 4.68 Butalbital 200 98.3 5.97 800 104 4.27 Hexobarbital 200 93.5 7.57 800 111 4.04 Methohexital 200 96.5 10.3 800 102 2.26 Pentobarbital 200 99.0 10.4 800 113 4.31 Phenobarbital 200 93.8 3.67 800 107 6.99 Secobarbital 200 99.5 5.87 800 107 5.81 11-Nor-9-carboxy- 9 -THC 20 95.4 4.84 80 106 2.00 CV [%] AN/2013/01-6

CONCLUSIONS As a result of this study, we were able to show: The automated DPX cleanup method performed using the GERSTEL MPS XL in Dual Head configuration provides cycle times for barbiturates and COOH-THC in urine of approximately 7 min/sample enabling a throughput of more than 200 samples per day. Barbiturates and COOH-THC can be rapidly and reproducibly isolated from hydrolyzed urine samples using an automated DPX cleanup procedure and directly transferred to the Agilent 6460 Triple Quadrupole Mass Spectrometer for LC-MS/MS determination. Linear calibration curves with R 2 values 0.99 or greater were achieved with LOQs meeting the minimum reportable limits requirements for the determination of the barbiturates and COOH-THC in urine. The DPX-LC/MS/MS method provided good accuracy and precision averaging 102 % (range: 93.4 % - 113 %) accuracy with 5.55 % CV (range: 2.00 % -10.4 %) for all compounds analyzed. REFERENCES [1] Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS, Gerstel AppNote 2012/01. [2] Rapid Automated Extraction and Confirmation of Buprenorphine and Norbuprenorphine in Urine by DPX-LC/MS/MS, Gerstel AppNote 2012/08 AN/2013/01-7 For research use only. Not for use in diagnostic procedures. The information provided for this product is intended for reference and research purposes only. GERSTEL offers no guarantee as to the quality and suitability of this data for your specific application. Information, descriptions and specifications in this publication are subject to change without notice.

GERSTEL GmbH & Co. KG Eberhard-Gerstel-Platz 1 45473 Mülheim an der Ruhr Germany +49 (0) 208-7 65 03-0 +49 (0) 208-7 65 03 33 gerstel@gerstel.com www.gerstel.com GERSTEL Worldwide GERSTEL, Inc. 701 Digital Drive, Suite J Linthicum, MD 21090 USA +1 (410) 247 5885 +1 (410) 247 5887 sales@gerstelus.com www.gerstelus.com GERSTEL AG Wassergrabe 27 CH-6210 Sursee Switzerland +41 (41) 9 21 97 23 gerstelag@ch.gerstel.com www.gerstel.ch GERSTEL K.K. 1-3-1 Nakane, Meguro-ku Tokyo 152-0031 SMBC Toritsudai Ekimae Bldg 4F Japan +81 3 5731 5321 +81 3 5731 5322 info@gerstel.co.jp www.gerstel.co.jp GERSTEL LLP Level 25, North Tower One Raffles Quay Singapore 048583 +65 6622 5486 +65 6622 5999 SEA@gerstel.com www.gerstel.com GERSTEL Brasil Av. Pascoal da Rocha Falcão, 367 04785-000 São Paulo - SP Brasil +55 (11)5665-8931 +55 (11)5666-9084 gerstel-brasil@gerstel.com www.gerstel.com.br Information, descriptions and specifications in this Publication are subject to change without notice. GERSTEL, GRAPHPACK and TWISTER are registered trademarks of GERSTEL GmbH & Co. KG. Copyright by GERSTEL GmbH & Co. KG Awarded for the active pursuit of environmental sustainability