COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

Similar documents
COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

GENERAL INSTRUCTIONS GENERAL PREPARATION

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

GENERAL INSTRUCTIONS GENERAL PREPARATION

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

1) A ph/mv meter or an ion meter, either line operated or portable.

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

POTASSIUM ION SELECTIVE ELECTRODE

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions.

Operating Instructions for Ammonium ISE Specifications

CALCIUM SELECTIVE ELECTRODE

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable.

The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in aqueous solutions.

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable.

NITRATE ION SELECTIVE ELECTRODE

GENERAL INSTRUCTIONS GENERAL PREPARATION

phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement

Ion Selective Electrode Probe

Rapid RH Portable ph Meter. User's Manual

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth

perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

LEAD ION SELECTIVE ELECTRODE

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Understanding ph Troubleshooting and Diagnostic information

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement

User Guide. Perchlorate Ion Selective Electrode

Guide to ph Analysis.

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement

STANDARD OPERATING PROCEDURE

User Guide. Fluoroborate Ion Selective Electrode

User Guide. High Performance Ammonia Ion Selective Electrode

OPERATING INSTRUCTIONS

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

User Guide. Calcium Ion Selective Electrode

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

PRACTICAL 3 ph AND BUFFERS

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

[Agilent Technologies] [Electrochemistry Products] Quick Reference

NO 2. Orion Nitrite Electrode INSTRUCTION MANUAL. Orion 93-46, Orion 97-46, ionplus. Analyze Detect Measure Control

QAM-I-111 Operation and Calibration of the Conductivity Meter

User Guide. ROSS Sodium Ion Selective Electrode

Ammonia Gas Sensing Combination Electrode

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

NH 3 -selective electrodes

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

Thermo Scientific Orion Fluoride Ion Selective Electrodes The standard in fluoride ion analysis EPA compliant

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

User Guide. Cyanide Ion Selective Electrode

F- Cl- Br- I- CN- S2- NO3-

User Guide. Bromide Ion Selective Electrode

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Complex equilibria and calculations of formation constants from potentiometric data

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC

symphony Reference Electrodes

User Guide. High Performance Ammonia Ion Selective Electrode

READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

ph Probes Lab vs. Process Care, Calibrations and Comparison Presented by: Michael Blackburn Technical Sales Specialist at Hach Company

STANDARD OPERATING PROCEDURES

Chemistry MultiMeasure Sensor PS-2170

ORP Probe ORP. Reads. +/-2000mV. Range. 95% in 1s. Response time 100 PSI. Max pressure. 60m (197 ft) Max depth 1 80 C. Temperature range C.

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

ph Probe ReallyEasyData com

User Guide. Iodide Ion Selective Electrode

Buffer Preparation. Learning Objectives:

symphony Multisensor Probes

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Ammonia Gas Sensing Combination Electrode

ORP Probe. Mini ORP. Reads. +/-2000mV. Range SMA. Connector. 95% in 1s. Response time 100 PSI. Max pressure. 60m (197 ft) Max depth 1 80 C

Chemistry Determination of Mixed Acids

Bante210 Benchtop ph Meter Instruction Manual

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual

Transcription:

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com Email: techinfo@coleparmer.com 1 Table of Contents

TABLE OF CONTENTS General Instructions...3 introduction...3 required equipment...3 required solutions...3 Electrode Preparation...4 Calibration Procedure...4 Measurement...5 measuring hints...5 sample requirements...6 units of measurement...6 Electrode Characteristics...6 reproducibility...6 interferences...6 temperature influences...8 electrode response...8 limits of detection...9 ph effects...9 electrode life...10 electrode storage...10 Electrode Theory...10 electrode operation...10 Common Measurement Problems, Simple Causes...11 Troubleshooting Difficult Measurement Problems...12 Some Questions Asked About Ion-Selective Electrodes...12 Calcium Electrode Specifications...14 Ordering Information...14 2 Table of Contents

GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Calcium Ion Electrode is used to quickly, simply, accurately, and economically measure calcium in in-line batch processing, control processing, or water treatment applications. Required Equipment 1. Cole-Parmer Ion Concentration Controller, Cat. No. 56105-00. Cord with U.S. standard plug, Cat. No. 50001-68 or cord with European plug, Cat. No. 50001-70 is necessary. 2. Cole-Parmer Industrial Calcium Ion Electrode, Cat. No. 27077-02. It is a sealed, double junction electrode which is easily connected to the Ion Concentration Controller with stripped wire ends. 3. Cole-Parmer Industrial Housings to protect the Calcium Ion Electrode. Contact Cole-Parmer to choose an Industrial Housing. Required Solutions 1. Deionized or distilled water for solution and standard preparation. 2. Cole-Parmer Ionic Strength Adjuster, 4 M KCl, Cat. No. 27503-52. To prepare this ISA, add 298 grams of reagent-grade potassium chloride (KCl) to a 1000 ml volumetric flask about half full of distilled water. Swirl the flask to dissolve the solid and fill to the mark with distilled water. Cap the flask and invert several times to mix the solution. 3. Cole-Parmer Calcium Standard, 1000 ppm Ca +2, Cat. No. 27503-05. To prepare this standard, add 3.67 grams of reagent-grade calcium chloride (CaCl 2.2H 2 O) to a 1 liter volumetric flask about half full of distilled water. Swirl the flask to dissolve the solid and fill to the mark with distilled water. Cap the flask and invert several times to mix the solution. 3 Table of Contents

ELECTRODE PREPARATION Remove the rubber cap covering the electrode tip. Gently shake the electrode downward in the same manner as a clinical thermometer to remove any air bubbles which may be trapped behind the calcium membrane. Prior to first usage, or after long-term storage, immerse the calcium membrane in calcium standard for thirty minutes. The electrode is now ready for use. CALIBRATION PROCEDURE This quick start procedure is designed to get the unit operational. To set alarms and relays, consult the Operator s manual. Standards must be made in the range of 0.01 to 1000 ppm, as the controller will not accept standards outside this range. 1. Connect the electrode with the stripped wire ends to the controller by inserting the center wire into slot #22 and inserting the outer wire braid into slot #21. 2. Press the [Mode] key 9 times until the display reads [Configuration] 3. Press the [Cal] key. 4. The display should read [KB UNLOCKED]. Press [ENT]. 5. The display should now read: [LCD Contrast: 4]. Press [ENT]. 6. The display should now read: [Access Nr.: 0]. Press [ENT]. 7. The display should now read [Valence: #]. Pressing the [/\] and [\/] keys, choose the proper valence of the ion you are testing for. Press [ENT] (If the factory setting is already the correct #, press [ENT]). E.G. For Silver/sulfide, choose 1; For Sodium, choose +1; for Calcium, choose +2. 8. If you chose a valence which is different from the factory setting, the display will read [Reset ppm Cal?] Press [ENT]. 9. The display will now read [Scale: 100.0 ppm]. Press [Mode]. 10. The display will now read [Configuration]. 11. Press [Mode] key 6 times. Make sure the display reads: [Termoc.: OFF]. If so, go to step 14. If the display reads [Termoc.: ON], go to step 12. 12. Press the [CAL] key. Press the [\/] key so the display reads [CAL TC: OFF]. The display will flash [UPDATE] and then read: [CAL TC:0.198%/oC]. 13. Press the [Mode] key. The display will now read: [Termoc.: OFF]. 14. Press the [Mode] Key 8 times. The display should read [ppm 5 point cal].(note: It is not necessary to do a 5 point calibration. You can calibrate from 1 to 5 points). Prepare 1 to 5 standard ISE solutions whose concentrations vary by tenfold. Use the 1000 ppm ISE solution as the stock solution. Use the serial dilution method for this preparation and add the ISA when required to each standard. 15. To start the calibration sequence, press the [cal] button. The display will now read [Cal Point Nr 1]. 16. Insert the ISE electrode in the standard with the lowest ppm value.(note: for best results, ppm standards should be used in the following order: {0.1, 1, 10, 100, 1000} 4 Table of Contents

17. Press the [cal] button. 18. When a stable millivolt reading is obtained, press the [ENT] button. 19. The display should now read [P1 Decade: 100.0]. 20. Choose the ppm range within which the standard lies using the [/\] or [\/] buttons. - select 10.00 if the standard solution value is < 10.00 ppm. - select 100.0 if the standard solution value is > 10.00 ppm. - and < 100.0 ppm select 1000 if the standard solution is > 100.0 ppm. 21. Press the [ENT] button. 22. The display should now read [P1 Value: (# selected in step 20)]. 23. Choose the actual ppm value of the standard solution by pressing the [/\] or [\/] buttons until the exact ppm value is displayed. Press the [ENT] key to confirm the ppm value. 24. The screen should read [CAL POINT Nr. 2]. If no further calibration points are required, go to step 25. If you wish to enter more calibration points, go to step 26. 25. If no further calibration points are desired, press the [ENT] button. The display will now read [End Point Cal?]. Press the [ENT] button once more. The screen will now display [Zr Pt updated]. Press [ENT]. The display will now read [ppm 5 point cal]. To display Concentration value, press the [Mode] key 10 times. To display millivolt signal supplied by electrode, press the [Mode] key once. 26. If further calibration points are required, repeat steps 16 through 25. After the 5th calibration point is confirmed at step 11, the meter will flash [chk cal pt] and then read [VALUE UPDATED]. Press the [ENT] key. The display will now read [ppm 5 point cal]. To display Concentration value, press the [Mode] key 10 times. To display millivolt signal supplied by electrode, press the [Mode] key once. MEASUREMENT Measuring Hints Check electrode slope weekly if not daily. Always use fresh standards for calibration. All samples and standards should be at the same temperature for precise measurement and below 40 C. A difference of 1 C in temperature will result in a 4% measurement error. All samples and standards should be at the same pressure for precise measurement. The sensing membrane is normally subject to water uptake and might appear milky. This has no effect on performance. For samples with high ionic strength, prepare standards whose composition is similar to the sample. Always check to see that the membrane is free from air bubbles after immersion into standard or sample. A slow responding electrode may be caused by interferences to the electrode. To restore proper performance, soak the electrode in distilled water for about 5 minutes to clean the 5 Table of Contents

membrane, rinse, and soak in standard solution for about 5 minutes. Sample Requirements All samples must be aqueous and not contain organics which can dissolve in the membrane or extract out the liquid ion exchanger. Interferences should be absent. If they are present, use the procedures found in the Interferences section to remove them. The ph range for the calcium ion electrode is 3-10. Units of Measurement Calcium concentrations are measured in units of parts per million, moles per liter, or any other convenient concentration unit. Table 1 indicates some of the concentration units. TABLE 1: Concentration Unit Conversion Factors ppm Ca +2 ppm CaCO 3 moles/liter Ca +2 4.01 10.00 40.10 400.80 ELECTRODE CHARACTERISTICS Reproducibility 10.0 24.9 100.1 1000.9 1.0 x 10-4 2.5 x 10-4 1.0 x 10-3 1.0 x 10-2 Electrode measurements reproducible to ±10% of reading can be obtained if the electrode is calibrated every hour. Factors such as temperature fluctuations, drift, and noise limit reproducibility. Interferences Table 2 lists some common cations that, if present in high enough levels, will cause electrode interferences and measurement errors or electrode drift when using the calcium electrodes. Electrode drift and slow response could indicate the presence of high interferences from the ions listed. Soak the electrode in distilled water for five minutes, then for five minutes in calcium standard solution to restore proper response. 6 Table of Contents

TABLE 2: Concentration of Possible Interferences Causing a 10% Error at Various Levels of Calcium. Interferences (moles/liter) 10-2 M 10-3 M 10-4 M Mg +2 1.0 x 10 +1 1.0 x 10 0 1.0 x 10-1 Zn +2 1.0 x 10 +1 1.0 x 10 0 1.0 x 10-1 Ba +2 7.0 x 10 0 7.0 x 10-1 7.0 x 10-2 K +1 4.0 x 10 0 4.0 x 10-1 4.0 x 10-2 Na +1 2.0 x 10 0 2.0 x 10-1 2.0 x 10-2 Ni +2 5.0 x 10-1 5.0 x 10-2 5.0 x 10-3 Cu +2 4.0 x 10-1 4.0 x 10-2 4.0 x 10-3 Fe +2 2.0 x 10-2 2.0 x 10-3 2.0 x 10-4 Sr +2 6.0 x 10-2 6.0 x 10-3 6.0 x 10-4 H +1 4.0 x 10-2 4.0 x 10-3 4.0 x 10-4 Hg +2 4.0 x 10-2 4.0 x 10-3 4.0 x 10-4 Pb +2 1.0 x 10-4 1.0 x 10-5 1.0 x 10-6 Interferences (ppm) 1000 ppm CaCO 3 100 ppm CaCO 3 10 ppm CaCO 3 Mg +2 2.43 x 10 5 2.43 x 10 4 2.43 x 10 3 Zn +2 6.53 x 10 5 6.53 x 10 4 6.53 x 10 3 Ba +2 9.60 x 10 5 9.60 x 10 4 9.60 x 10 3 K +1 1.56 x 10 5 1.56 x 10 4 1.56 x 10 3 Na +1 4.60 x 10 4 4.60 x 10 3 4.60 x 10 2 Ni +2 2.94 x 10 4 2.94 x 10 3 2.94 x 10 2 Cu +2 2.54 x 10 4 2.54 x 10 3 2.54 x 10 2 Fe +2 1.11 x 10 4 1.11 x 10 3 1.11 x 10 2 Sr +2 5.20 x 10 3 5.20 x 10 2 5.20 x 10 1 H +1 1.4 ph 2.4 ph 3.4 ph Hg +2 8.0 x 10 3 8.0 x 10 2 8.0 x 10 1 Pb +2 2.0 x 10 1 2.0 2.0 x 10-1 7 Table of Contents

Temperature Influences Samples and standards should be at the same temperature, since electrode potentials are influenced by changes in temperature. A 1 C difference in temperature results in a 4% error at the 1.0 x 10-3 M level. Provided that temperature equilibria has occurred, the calcium ion electrodes can be used at temperatures from 0-40 C continuously. Room temperature measurements are recommended, since measurements at temperatures markedly different from room temperature may require equilibrium times up to one hour. Electrode Response Plotting the mv potential against the calcium concentration on semi-logarithmic paper results in a straight line with a slope of about 25 mv per decade. (Refer to Figure 1.) Figure 1 Typical Calcium Electrode Calibration Curve + 100 electrode potential (mv) + 60 + 20-20 ~ 27 mv 10-fold change - 60 1 10 (ppm as CaCO 3 ) 100 1000 0.1 1 10 (ppm as Ca +2 ) 100 1000 10-5 10-4 10-3 10-2 10-1 Ca +2 Concentration (M) The time needed to reach 99% of the stable electrode potential reading, the electrode response time, varies from one minute or less for calcium concentration above 1.0 x10-5 M to several minutes near the detection limit. Refer to Figure 2. 8 Table of Contents

Figure 2 Typical Electrode Time Response to Step Changes in CaCl 2 + 50 + 40 10-3 M to 10-2 M CaCl 2 electrode potential (mv) + 30 + 20 + 10 0-10 - 20 10-3 M to 10-4 M CaCl 2 10-3 M to 10-5 M CaCl 2 1 2 Time (minutes) 3 4 Limits of Detection The upper limit of detection is 1M in pure calcium chloride solutions. The upper limit of detection is above 1.0 x 10-1 M when other ions are present, but the possibility of a liquid junction potential developing at the reference electrode and the salt extraction effect are two limiting factors. Some salts may be extracted into the electrode membrane at high salt concentrations causing deviation from theoretical response. Calibrate the electrode at four or five intermediate points, or dilute the sample, to measure samples between 1.0 x 10-1 M and 1M. The slight water solubility of the ion exchanger in the sensing module, which causes deviation from theoretical response, determines the lower limit of detection. The theoretical response at low levels of calcium chloride compared to actual response is shown in Figure 1. To avoid formation of CaSO 4, the sulfate concentrations must be less than 5 x 10-4 M (50 ppm). To avoid formation of CaCO 3 or formation of the CaHCO 3 + complex, the ph of the solution should be less than 7, and the total carbonate/bicarbonate concentration should be less than 3 x 10-3 M (280 ppm carbonate). ph Effects The calcium electrode has an operating ph range of 3-10. Use at other ph values can adversely affect the membrane. Hydrogen ion interferes with measurements of very low levels of calcium. Hydroxide ion will complex calcium ions. 9 Table of Contents

Electrode Life The calcium electrode will last six months in normal laboratory use. On-line measurement will shorten operational lifetime to several months. In time, the response time will increase and the calibration slope will decrease to the point calibration is difficult and electrode replacement is required. Electrode Storage The Cole-Parmer Calcium Ion Electrode may be stored in-line if the membrane is kept wet. For storage off-line, rinse and dry the calcium membrane electrode and cover the tip with any protective cap shipped with the electrode. ELECTRODE THEORY Electrode Operation The Cole-Parmer calcium electrode consists of an electrode body containing an ion exchanger in a sensing membrane. This sensing membrane contains a liquid internal filling solution in contact with a gelled organophilic membrane containing an calcium selective ion exchanger. An electrode potential develops across the membrane when the membrane is in contact with an calcium solution. Measurement of this potential against a constant reference potential with a digital ph/mv meter or with a specific ion meter depends on the level of free calcium ion in solution. The level of calcium ions, corresponding to the measured potential, is described by the Nernst equation: E = E o + S logx where: E = measured electrode potential E o = reference potential (a constant) S = electrode slope (~25 mv/decade) X = level of calcium ions in solution The activity, X, represents the effective concentration of the ions in solution. Total calcium concentration, C t, includes free calcium ions, C f, plus bound or complexed calcium ions, C b. Since the calcium electrodes only responds to free ion, the free ion concentration is: C f = C t - C b The activity is related to the free ion concentration, C f, by the activity coefficient, γ, by: X = γ C f 10 Table of Contents

Activity coefficients vary, depending on total ionic strength, I, defined as: 2 I = ½ ΣC x Z x where: C x = concentration of ion X Z x = charge of ion X Σ = sum of all of the types of ions in the solution. In the case of high and constant ionic strength relative to the sensed ion concentration, the activity coefficient, γ, is constant and the activity, X, is directly proportional to the concentration. To adjust the background ionic strength to a high and constant value, ionic strength adjuster (ISA) is added to samples and standards. The recommended ISA for calcium is potassium chloride, KCl. COMMON MEASUREMENT PROBLEMS...SIMPLE CAUSES? - NOT REPRODUCIBLE sample carryover? sample interferences or complexing agents present? contaminated reference electrode junction - SLOW RESPONSE (READINGS SLOWLY CHANGING) electrode poisoned by sample? - OUT OF RANGE READING electrode(s) not plugged into controller properly? not enough fill solution left in reference? electrode not in sample solution? air bubble on electrode surface? - LOW SLOPE OR NO SLOPE standards are old? contaminated? made wrong? sample ph has not been adjusted properly to operating range of electrode? air bubble on electrode surface? controller okay? not enough fill solution left in reference? failed electrode? - NOISY RESPONSE (READINGS RANDOMLY CHANGING) controller not grounded? air bubble on electrode surface? controller okay? not enough fill solution left in reference? 11 Table of Contents

- DRIFTY RESPONSE (READINGS CONTINUOUSLY CHANGING) excessive leaking at reference electrode junction? clogged reference electrode junction? sensing membrane poisoned by sample? temperature problems? sample too concentrated? sensing membrane needs conditioning? - INACCURATE (BUT CALIBRATION IS OK) standards are incorrect? sample is not ph adjusted properly? sample carryover? sample interferences or complexing agents present? TROUBLESHOOTING DIFFICULT MEASUREMENT PROBLEMS COMPONENTS OF THE ELECTRODE SYSTEM - CONTROLLER perform checkout procedure in instruction manual. - ION-SELECTIVE ELECTRODE perform electrode slope check, inspect for physical damage. Monitor leak rate on reference side. PROPER CALIBRATION - STANDARD(S) prepare fresh standards by serial dilution. - ph ADJUSTMENT adjust ph to operational range of electrode. SAMPLE VARIABLES - CONCENTRATION RANGE sample out of range? - PRESSURE wrong sample pressure? - TEMPERATURE wrong sample temperature? - ph wrong sample ph? - INTERFERENCES interferences in the sample? - COMPLEXATION complexing agents in sample? SOME QUESTIONS ASKED ABOUT ION-SELECTIVE ELECTRODES 1. How often do you need to calibrate? Recalibrate at least once a week, or daily if in doubt of your results being accurate or reproducible. 2. How long after being opened or made are standards good? The stock standard will last at least six months before discarding, whereas diluted standards treated with ISA/pH buffer should be prepared weekly. 12 Table of Contents

3. How do you store the electrodes? It is best to store them dry when they are not to be used in the next week or so. Empty out the filling solution in the gas-sensing electrode. For shorter periods, store in dilute standard approximating the sample concentration, and ISA/pH buffer added when required. 4. Can you do temperature compensation with ISE? Yes, it is possible, but somewhat difficult. First, you have to know the isopotential point for a given electrode system. Second, the concentration of the sample has to be similar in concentration to the isopotential point or else the temperature correction will be very inaccurate. Third, the temperature of the sample can not exceed the operational temperature range of the ISE. Fourth, very little isopotential point data for ISE s is available at this time. It is best to standardize and measure samples at the same temperature without using temperature compensation. 5. Can you do in-line continuous ISE measurement without treating the sample? Yes, direct measurement is possible in many cases without ISA/buffer addition to the sample stream. However, Fluoride, Sulfide, Ammonia, and Sodium Electrodes do require ph adjustment and must have ISA/ buffer added to the sample stream. 6. Which standards should be used with the ISE? The most obvious choice will be determined by what concentration units are desired (e.g. ppm as what?). Also, if an electrode is being used to measure another ion (e.g. sulfate with a lead electrode), use a stock standard of the ion to be measured (e.g. sulfate). 7. Why buy a combination electrode instead of separate ones? Advantages: no external reference electrode needed. more economical than price of both one less electrode to fit in the process 13 Table of Contents

CALCIUM ELECTRODE SPECIFICATIONS (MODEL # 27077-02) SLOPE 25 +/ 2 mv/decade REPRODUCIBILITY ± 4% INTERFERENCES Pb +2, Hg +2, Cu +2, Ni +2 TEMPERATURE RANGE 0 to 40 degrees C PRESSURE RANGE 0 to 30 psi RESPONSE TIME 95% response in 30 seconds STORAGE for in-line storage keep the membrane wet, for off-line storage rinse and dry the membrane, and cover with a protective cap CONCENTRATION RANGE 5 x 10 6 M to 1M 0.2 ppm to 40,000 ppm ph RANGE 3 to 10 ph RESISTANCE 100 megaohms CARE, MAINTENANCE, treat with distilled water & CLEANING followed by calcium standard TEMPERATURE COMPENSATION Yes ISOPOTENTIAL POINT 1 ppm calcium ON LINE CAPABILITY Yes APPLICATIONS: Water Softening Systems, Drinking and Mineral Waters, Blood Electrolytes/Clinical Analysis ORDERING INFORMATION P/N DESCRIPTION 56105-00 Ion Concentration Controller 50001-68 Cord with U.S. standard plug 50001-70 Cord with European plug 27077-02 Industrial Calcium Ion Electrode, sealed, double junction electrode which is easily connected to the Ion Concentration Controller with stripped wire ends. Cole-Parmer Industrial Housings to protect the Calcium Ion Electrode. Contact Cole-Parmer at 1-800-323-4340 to choose an Industrial Housing. 27503-05 Calcium Standard, 1000 ppm Ca +2 27503-52 Calcium ISA (Ionic Strength Adjustor),4 M KCl 14 Table of Contents