HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Similar documents
Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

HR/AM Targeted Peptide Quantitation on a Q Exactive MS: A Unique Combination of High Selectivity, Sensitivity and Throughput

Targeted protein quantification

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Peptide Targeted Quantification By High Resolution Mass Spectrometry A Paradigm Shift? Zhiqi Hao Thermo Fisher Scientific San Jose, CA

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Q Exactive TM : A True Qual-Quan HR/AM Mass Spectrometer for Routine Proteomics Applications. Yi Zhang, Ph.D. ThermoFisher Scientific

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Quan/Qual Analyses. Unmatched Confidence for. Thermo Scientific Q Exactive Orbitrap LC-MS/MS System. Identify Quantify Confirm

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

of mass spectrometry

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

MetWorks Metabolite Identification Software

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Key Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia

High-Field Orbitrap Creating new possibilities

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Increasing the Multiplexing of Protein Quantitation from 6- to 10-Plex with Reporter Ion Isotopologues

HILIC Method Development in a Few Simple Steps

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Improving Intact Antibody Characterization by Orbitrap Mass Spectrometry

An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

A Quadrupole-Orbitrap Hybrid Mass Spectrometer Offers Highest Benchtop Performance for In-Depth Analysis of Complex Proteomes

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library

Fast, Effective XPS Point Analysis of Metal Components

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology

Introduction to Fourier Transform Infrared Spectroscopy

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

Monitoring Protein PEGylation with Ion Exchange Chromatography

for XPS surface analysis

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Self-assembling covalent organic frameworks functionalized. magnetic graphene hydrophilic biocomposite as an ultrasensitive

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Quantitation of TMT-Labeled Peptides Using Higher-Energy Collisional Dissociation on the Velos Pro Ion Trap Mass Spectrometer

Improving Experimental Efficiency for Drug Metabolite Profiling and Identification

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Enhancement of Linearity and Response in Charged Aerosol Detection

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Introduction to Fourier Transform Infrared Spectroscopy

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization

Analysis of a Verapamil Microsomal Incubation using Metabolite ID and Mass Frontier TM

Automated and accurate component detection using reference mass spectra

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Multi-residue analysis of pesticides by GC-HRMS

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Relative Quantitation of TMT-Labeled Proteomes Focus on Sensitivity and Precision

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

6 x 5 Ways to Ensure Your LC-MS/MS is Healthy

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

TOMAHAQ Method Construction

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

The key to your laboratory s success

Measurement of Metabolic Stability Using SIM and Identification of Metabolites by Data Dependent Full-Scan MS/MS and CNL Scanning

Thermo Finnigan LTQ. Specifications

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Transcription:

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer 1 1 Thermo Fisher Scientific, San Jose, CA, USA; 2 Thermo Fisher Scientific, Bremen, Germany

m Overview Purpose: The performance of a novel hybrid quadrupole-orbitrap TM mass spectrometer was evaluated in targeted peptide quantification. Methods: Two high-resolution, accurate-mass (HR/AM) targeted quantification methods were used to analyze 14 peptide standards spiked into a complex E. coli digest background. The data were automatically processed using specialized software. Results: A detection limit of 1 amol and a linear dynamic range of 4 orders were obtained with msx tsim methods in a moderately complex background. A detection limit of 1 5 amol and a linear dynamic range of 3 4 orders were obtained with thcd methods in a highly complex background. Introduction Quantitative proteomics enables the identification of a large number of protein candidates, which display biologically interesting dynamics on a global scale in the early discovery phase. A targeted MS approach, in particular selected reaction monitoring (SRM) on a triple quadrupole MS, has become the standard technique for quantitatively analyzing tens to hundreds of peptide candidates across large numbers of samples, either for understanding of signaling regulation or for verification and selection of potential biomarkers. However, the low resolution of triple quadrupole MS limits its ability to achieve high selectivity on the targets from complex backgrounds. It is also difficult to develop SRM method for peptides with high charge states or high mass or for modified peptides. In this study, a true high-resolution, accurate-mass (HR/AM) mass spectrometer, was evaluated for targeted protein quantification. The dynamic range and LOD/LOQ of two HR/AM methods: spectrum multiplexing targeted selected ion monitoring (msx tsim) and targeted HCD (thcd), were investigated. Methods Sample Preparation Thermo Scientific heavy-isotope labeled peptide retention standards were spiked into either 5 ng/µl or 25 ng/µl E coli. whole cell tryptic digest to reach final concentrations of amol/µl, 5 amol/µl, 25 amol/µl, 5 amol/µl, 5 amol/µl, 5 fmol/µl, and 5 fmol/µl. Liquid Chromatography Peptides from 2 µl of each sample were separated on a Michrom Magic C18 nanolc column (75 μm x 15 cm, 3 μm particle) with a 1 hr gradient at a flow rate of 35 nl/min. LC solvents were.1% formic acid in H 2 O (Solvent A) and.1% formic acid in acetonitrile (Solvent B). The LC gradient was -2 min, 3% B; 2-47 min, 3%-3% B; 47-52 min, 3%-95% B; 52-54 min, 95% B; 54-55 min, 95%-5% B; 55- min, 5% B. Mass Spectrometry The samples with 5 ng/µl E coli digest background were analyzed in triplicate with a Full-msx tsim method on a Thermo Scientific Q Exactive hybrid quadrupole-orbitrap MS (Figure 1). The resolution was set at 1K for both scan types. The AGC target was 1E6 for full scan and 2E5 for SIM scan. The maximum ion injection time was 1 ms for full scan and 5 ms for SIM scan. The isolation width for SIM scan was 4 amu. The peptide retention standards were monitored over a 3 min window. The multiplexing level was set at 4, which allowed isolation and accumulation of up to four peptide targets in the C-trap before they were transferred to the Orbitrap mass analyzer for detection (Figure 2). The samples with 25 ng/µl E coli digest background were analyzed in triplicate with a thcd method. The resolution was set at 17.5K. The AGC target was 5E5. The maximum ion injection time was 5 ms. The quadrupole isolation width was 2. amu. The normalized collision energy (NCE) was 27%.The peptide retention standards were monitored over a 3 min window. Data Analysis The data were automatically processed using Thermo Scientific Pinpoint software. Extracted ion chromatograms for peptide retention standards were obtained with a 5 ppm mass tolerance. Peak areas were calculated with 7-point smoothed complete chromatograms. Linear fitting was performed with a weighing scheme of 1/x. HCD Cell, C-Trap for Spectrum Multiplexing Orbitrap with Enhanced Resolution Q Im 2 HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

FIGURE 1. Schematic and key features of the Q Exactive MS. HCD Cell, C-Trap for Spectrum Multiplexing Quadrupole Mass Filter for Precursor Ion Selection Bent Flatapole S-Lens for Improved Sensitivity Orbitrap with Enhanced Resolution Capillary Key Features of the Q Exactive MS Incorporation of the S-lens at the source enhances sensitivity by up to 5 fold. Quadrupole mass filter enables precursor selection for data-dependent MS 2 and selected ion monitoring (SIM). Advanced signal processing increases resolution by about twofold at the same transient length, which results in a maximum resolution of 1K at m/z and a maximum scan speed of 12 Hz at a resolution of 17.5K. Spectrum multiplexing (msx) and parallel ion injection/detection significantly improves duty cycle. FIGURE 2: Spectrum multiplexing and parallel ion injection/detection provide high-throughput analysis. A G C Orbitrap FTMS acquisition scan 1 A G C Orbitrap FTMS acquisition scan 2 Results Collecting ions for scan 2 Collecting ions for scan 3 Up to 1 target ions can be isolated by the quadrupoles sequentially, stored in the C-trap, then transferred to the Orbitrap mass analyzer and detected with high resolution simultaneously. The isolation and trapping of the target ions is also concurrent with the Orbitrap mass analyzer detection of the previous ion packet. Targeted Quantification with msx tsim Full or SIM spectra were acquired at three different resolutions: 35K, 7K and 1K. Baseline separation of two ions that are only 3 ppm apart was only achievable at the resolution of 1K. The high mass accuracy (< 5 ppm) and high resolution ensured accurate identification of target peptides from complex background with msx tsim (Figure 3). FIGURE 3: High resolution ensures accurate target selection. 3 ppm m/z Thermo Scientific Poster Note PN63586_E 6/12S 3

r Bent Flatapole Target peptides in 1 ng of E coli digest were analyzed with alternating full scans and tsim scans (Figure 4). The peptide GLILVGGYGTR* (558.3259, 2+) is barely visible in the full scan with a amu isolation window (Figure 4A), but is accurately identified with a S/N of 65 and mass deviation of ~.2 ppm in the SIM scan with a 4 amu isolation window (Figure 4B). The +1 isotope peak is also accurately identified with a S/N of 39 and mass deviation of ~1.4 ppm in the SIM scan. FIGURE 4: High sensitivity with quadrupole-based SIM scan. Capillary A. 1 558.332 S/N: 7 558.3 558.5 558.7 Full Scan ( -1amu) Scan #: 399 5 7 9 1 B. 1 SIM Scan (556.3-5.3amu) Scan #: 31 558.3258 (S/N: 65) 558.8282 (S/N: 39) 558.3 558.5 558.7 557. 558. 559. 5. m/z Extracted ion chromatograms (XIC) of target peptides in SIM scans were obtained with 5 ppm mass tolerance. The XIC of GISNEGQNASIK* (613.3167, 2+) at 1 fmol, 1 fmol and 1 amol is displayed in Figure 5. Although GISNEGQNASIK* was monitored with two other peptides in the same time range, more than 1 SIM scans with a resolution of 1K were obtained across a 12 sec LC peak, even at a sample amount of 1 amol. As shown in Figure 6, a 4-order linear dynamic range was obtained for most peptides with the msx tsim method. Table 1 lists the LOD and LOQ of target peptides for the method. The CV% from triplicate analyses is also included for the LOQ. FIGURE 5. Multiplexing improves throughput and results in accurate quantification. 3 ppm 1 18.12 1 fmol GISNEGQNASIK* 5 ppm XIC 1 18.11 1 fmol 1 18.1 1 amol 17. 17.2 17.4 17.6 17.8 18. 18.2 18.4 18.6 Time (min) 4 HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

FIGURE 6. Four-order linear dynamic range with msx tsim method. SSAAPPPPPR* GISNEGQNASIK* IGDYAGIK* Peak Area Peak Area (log 1 ) R 2 >.9998 R 2 >.9997 Sample Amount (fmol) Sample Amount (log 1 ) TABLE 1. LOD and LOQ of target peptides for the msx tsim method. LOD (amol) LOQ (amol) CV% (LOQ) IGDYAGIK* 1 5 3 SSAAPPPPPR* 1 5 3 HVLTSIGEK* 1 5 3 LTILEELR* <1 1 4 GLILVGGYGTR* 1 5 7 NGFILDGFPR* <1 1 14 SAAGAFGPELSR* <1 1 4 GISNEGQNASIK* <1 1 1 ELASGLSFPVGFK* <1 1 1 TASEFDSAIAQDK* 1 5 1 SFANQPLEVVYSK* <1 1 8 ELGQSGVDTYLQTK* 1 5 7 LSSEAPALFQFDLK* <5 5 3 GILFVGSGVSGGEEGAR* 5 1 5 Targeted Quantification with thcd With thcd, high resolution and high mass accuracy allow simultaneous accurate identification of multiple fragment ions from complex mixed MS 2 spectra (Figure 7). As shown in Figure 8, 3-4 orders of linear dynamic range were obtained for most peptides with the thcd method. Table 2 lists the LOD and LOQ of target peptides for the thcd method. The CV% from triplicate analyses is also included for the LOQ. FIGURE 7. HCD spectrum of peptide NGFILDGFPR* (573.35, 2+) at 5 amol in 5 ng E coli tryptic digest background. 1 486.2699 y4 1.2971 y5 714.3813 y6 5 amol 974.53 y8 5 7 9 1 m/z Magic is a trademark of Michrom BioResources, Inc and its subsidiaries. This information is not intended to encourage use o property rights of others. Thermo Scientific Poster Note PN63586_E 6/12S 5

FIGURE 8. Three to four orders of linear dynamic range with the thcd method. SFANQPLEVVYSK* LTILEELR* NGFILDGFPR* Peak Area Peak Area (log 1 ) R 2 >.999 R 2 >.998 Sample Amount (fmol) Sample Amount (log 1 ) TABLE 2. LOD and LOQ of target peptides for the thcd method. LOD (amol) LOQ (amol) CV% (LOQ) IGDYAGIK* >1 5 6 SSAAPPPPPR* 1 5 11 HVLTSIGEK* 5 1 11 LTILEELR* 1 5 3 GLILVGGYGTR* >1 5 8 NGFILDGFPR* >1 5 6 SAAGAFGPELSR* 1 5 4 GISNEGQNASIK* 1 5 5 ELASGLSFPVGFK* >1 5 2 TASEFDSAIAQDK 1 5 1 SFANQPLEVVYSK* <1 1 12 ELGQSGVDTYLQTK* 1 5 6 LSSEAPALFQFDLK* 1 5 6 GILFVGSGVSGGEEGAR* >1 5 8 Conclusion The Q Exactive MS is well suited for targeted quantification providing two powerful HR/AM approaches: msx tsim and thcd. Both approaches demonstrate high selectivity, high sensitivity and high throughput due to several hardware and software innovations. For the msx tsim approach, high mass accuracy (< 5 ppm) and high resolution (1K) ensure accurate identification of target ions in the presence of a complex background, while spectrum multiplexing dramatically increases throughput without sacrificing resolution. A LOD of 1 amol and a 4-order linear dynamic range was obtained for most peptides with the msx tsim method in the presence of a medium complex background. Different from triple-quadrupole-based SRM, all fragment ions are simultaneously detected with high mass accuracy and high resolution using the thcd approach. The HR/AM of fragment ions provides unmatched selectivity. The simultaneous detection of all fragment ions allows more efficient sampling of target peptides and results in improved sensitivity. A scan speed of 12 Hz at a resolution of 17,5 ensures high throughput. A LOD of 5 amol and 3 4 orders of linear dynamic range were obtained for most peptides with the thcd method in the presence of a strong complex background. Magic is a trademark of Michrom BioResources, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. 6 HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

www.thermoscientific.com 12 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, San Jose, CA USA is ISO Certified. Africa-Other +27 11 57 18 Australia +61 3 9757 43 Austria +43 1 333 5 34 Belgium +32 53 73 42 41 Canada +1 53 8447 China +86 1 8419 3588 Denmark +45 7 23 62 Europe-Other +43 1 333 5 34 Finland/Norway/Sweden +46 8 556 468 France +33 1 92 48 Germany +49 613 8 114 India +91 22 6742 9434 Italy +39 2 95 591 Japan +81 45 453 91 Latin America +1 561 688 87 Middle East +43 1 333 5 34 Netherlands +31 76 579 55 55 New Zealand +64 9 9 67 Russia/CIS +43 1 333 5 34 South Africa +27 11 57 18 Spain +34 914 845 965 Switzerland +41 61 716 77 UK +44 1442 233555 USA +1 532 4752 PN63586_E 6/12S