Technical Note 103. Response Enhancement for the Loral CCDs. using Ultra-Violet Flooding. A. P. Oates (RGO)

Similar documents
ACS after SM4: RELATIVE GAIN VALUES AMONG THE FOUR WFC AMPLIFIERS

5G50.51 Superconductor Suspension

203-NYC-05: Waves, Optics & Modern Physics

Cree XLamp XR-C LED Data Sheet

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5.

On the Emissivity of Silver Coated Panels, Effect of Long Term Stability and Effect of Coating Thickness

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

The Measurement of e/m

TECHNICAL DESCRIPTION SPECTRAFLOW ON LINE ANALYZER for BELT CONVEYOR APPLICATION

ORTEC. SLP Series Silicon Lithium-Drifted Planar Low-Energy X Ray Detector Product Configuration Guide

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

Supporting Information

Optics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P

Application of the GD-Profiler 2 to the PV domain

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

SOME ASPECTS OF PYROMETRY OF "WHITE" OBJECTS V.E.

Measurement method for the proficiency testing program

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

HMS-5000 Manual. Product Name: HMS-5000 Hall Effect Measurement System with variable temperature from 80K to 350K. - Manual version: ver 5.

LAAPD Performance Measurements in Liquid Xenon

0. Table of contents. Author: Jaap Snijder

Product family data sheet

Fall 2014 Nobby Kobayashi

RoHS. Specification CUD8DF1A. Drawn Approval Approval. 서식 Rev: 00

ONYX -MCE MULTI-CHANNEL OPTICAL FIBER PYROMETERS WITH ACTIVE EMISSIVITY COMPENSATION PRECISION TEMPERATURE MEASUREMENT FOR DEMANDING INDUSTRIAL

Cree XLamp XR-E LED PRODUCT FAMILY DATA SHEET FEATURES PRODUCT DESCRIPTION TABLE OF CONTENTS CLD DS-05 REV11

Features. Typical Applications G9 Candle Light. ProLight PTA2-3LxE-N3Lxx 3W AC LED Technical Datasheet Version: DS-0384

Cree XLamp MP-L EasyWhite LEDs Data Sheet / Binning & Labeling Document

Practical 1P4 Energy Levels and Band Gaps

Practical 1P4 Energy Levels and Band Gaps

Pumping Speed in the Drydown Zone

GLP Series Planar HPGe Low-Energy Detector Product Configuration Guide

Product family data sheet

Emissivity: Understanding the difference between apparent and actual infrared temperatures

OPERATING INSTRUCTIONS

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Speed of Light in Air

CERAIDE FOR USE AS AN AGRICULTURAL PLANT GROWTH REGULATOR FOR PROFESSIONAL USE ONLY

Report Liquid Hydrogen Target for the COMPASS Experiment

Sodium Chloride - Analytical Standard

ATOMIC WORLD P.1. ejected photoelectrons. current amplifier. photomultiplier tube (PMT)

Cryogenics for bolometric interferometry: The BRAIN experiment

COBP Photo Diode for Front monitor

EXPERIMENT NO. 4. Thermal Radiation: the Stefan-Boltzmann Law

Infrared Temperature Calibration 101 Using the right tool means better work and more productivity

The Vacuum Case for KATRIN

HPLC Praktikum Skript

MLCC APPLICATION GUIDE

(PT-A6) Visible Light Detector for Security Lighting

Product Description. Table of Contents. CLD-DS30 Rev 11. Product family data sheet

Cree XLamp XR-C LEDs. CLD-DS10 Rev 12H. Product family data sheet

(PT-A6) Visible Light Detector for Security Lighting

RoHS. Specification CUD8AF1C. 서식 Rev: 00

Lighting Solutions for Horticulture. The Light of Professional Knowledge

Pre-Leaving Certificate Examination, 2014 Triailscrúdú na hardteistiméireachta, 2014

SCI. Scientific Computing International. Scientific Computing International. FilmTek. Raising Thin Film Metrology Performance to a New Level

S-S35B-F

Photoemission Spectroscopy

Conductometric Titration & Gravimetric Determination of a Precipitate

Techniques for Fast Measurements of Low Dewpoint in Portable Hygrometers

Detectors for IR astronomy

Technical Notes. Introduction. PCB (printed circuit board) Design. Issue 1 January 2010

Cree XLamp XP-G LEDs Data Sheet

LED Measurement Instrumentation

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES

PHY3901 PHY3905. Hall effect and semiconductors Laboratory protocol

ORTEC. GLP Series Planar HPGe Low-Energy Detector Product Configuration Guide. Configuration Guidelines

Microscopy Cryostat System

UVTOP. Specification S-T39B-B (Formerly UVTOP250TO39BL) RoHS. Drawn Approval Approval

CCD OPERATION. The BBD was an analog delay line, made up of capacitors such that an analog signal was moving along one step at each clock cycle.

QUANTUM DOTS, ADVANTAGES AND DRAWBACKS FOR LIGHTING APPLICATIONS. David Schmidmayr, Johann Zehetner

PYRIPROXYFEN TECHNICAL

2. To measure the emission lines in the hydrogen, helium and possibly other elemental spectra, and compare these to know values.

Exercise 4-3. Titration of Weak Acids EXERCISE OBJECTIVE DISCUSSION OUTLINE. The 5% rule DISCUSSION

Cree XLamp XP-C LEDs Data Sheet

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

An Introduction to Insulation Resistance Testing

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

Mitigation of H2RG persistence with image illumination Brian A. McLeod* a, Roger Smith b

3mm Photodiode PD204-6B

5G50.52 Energy Storage with Superconductors

Photosynthesis. LabQuest OBJECTIVES

ONE-ELECTRON AND TWO-ELECTRON SPECTRA

5mm Silicon PIN Photodiode, T-1 3/4 PD333-3B/H0/L2

Supplementary Information. Back-Contacted Hybrid Organic-Inorganic Perovskite Solar Cells

STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION. Price: $5.00. Revised March 31, 1997

MLCC APPLICATION GUIDE

After successfully completing this laboratory assignment, including the assigned reading, the lab

PoS(PhotoDet 2012)010

PHYS2627/PHYS2265 Introductory quantum physics LABORATORYMANUAL Experiment 1: Experiments of Thermal Radiation

5mm Phototransistor PT333-3B

Silicon PIN Photodiode

CHIRON efficiency. A. Tokovinin. Version 2. March 28, 2011 file: prj/bme/chiron/commissioning/efficiency.tex

The Photoelectric Effect. One weight

Exercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES

Multialkali photocathodes grown by MBE technique. General Physics Institute, USSR, Academy of Sciences, , Moscow, Vavilov street, 38.

Investigation #9 OBSERVATION OF THE PHOTOELECTRIC EFFECT

Physics Procedia 00 (2011) TIPP Technology and Instrumentation in Particle Physics 2011

Photoelectric Effect Experiment

Measuring Laser Diode Optical Power with an Integrating Sphere

Transcription:

echnical Note 103 Response Enhancement for the oral CCDs using Ultra-Violet Flooding. A. P. Oates (RGO) July 1996 1

Response Enhancement for the oral CCDs using Ultra-Violet Flooding. 1. Introduction. he oral CCDs in their raw state, as delivered from Mike esser, have a spectral response similar to that of a thick, front-sided detector, i.e. around 50% response at the peak (600nm) and no response at wavelengths shorter than about 400nm. his can be compared with say the SIe CCDs which have a peak response of about 70% at 600nm and an Ultra-Violet response of about 30% (350nm). Although the oral devices have been thinned and should therefore exhibit some response in the blue (<400nm) this can only be achieved by a process of illuminating the surface with Ultra-violet light for some minutes in dry air (or Oxygen). his lack of Ultra-Violet response is due to the manufacturing process these devices are subject to. his process, to achieve a thinned response, results in the back-side surface being damaged (in fact by the acids used for the etching process). his damage manifests itself in a potential barrier close to the back-side surface where the UV photons are converted into electrons. Unfortunately because of the presence of this barrier, the electrons are inhibited from moving to the charge collection sites underneath the electrodes at the front-side surface, a distance of 10um or so. o overcome this problem the back-side surface can be treated in a variety of ways and in fact the SIe and EEV CCDs are treated in a manner which results in a permanent fix to this problem. he ultra-violet and peak response of these devices is however rather less spectacular than that obtained by the UV flooding technique. he oral CCDs have to have their surface potential modified by introducing Oxygen atoms into the surface via the use of Ultra-Violet light. his process has to be carried out at room temperature in the presence of dry air. Once the response had been modified by this process and the CCD cooled to operating temperature, the response should remain stable for weeks to months.. Flooding Process. he oral response (for device W7-4) as measured at RGO is given in the accompanying figures for varying conditions; before flooding at room temperature, after flooding at room temperature and at operating temperature. All the measurements were taken in photo-diode mode i.e. with the detector acting like a simple diode and measuring the current flowing due to the generation of photoelectrons at all the wavelengths. he first curve shows the situation before the UV flood at room temperature. Note the small response in the UV due to a prior flood some weeks before this date.

Spectral Response of oral W7-4 Science Array. = R. Before Cd UV Flood. APO, RGO. 10/05/96. 100 90 80 70 60 Q.E.(%) 50 40 30 0 10 0 0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 1 1.05 1.1 Wavelength (um) he next figure shows a typical response curve (at room temperature) after the device has been UV flooded. Spectral Response of oral W7-4 Science Array. APO, RGO. 1/05/96. 100 90 80 70 60 Q.E.(%) 50 40 30 0 10 0 0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 1 1.05 1.1 Wavelength (um) he points to note are- the UV (350nm) response is 70%, 3

the peak response (600nm) is 90% the red response (900nm) is 55% his curve was obtained at room temperature but does not change when the chip is cold, see following data Spectral Response of oral W7-4 Science Array. = -100. After Hg/Cd UV Flood. Std Baking. Average of 3 runs. APO, RGO. 14/03/96. 100 90 80 70 60 Q.E.(%) 50 40 30 0 10 0 0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 1 1.05 1.1 Wavelength (um) he oral UV response is 70% between 30nm and 400nm (as measured at RGO). If it is suspected that this response is not being achieved, then, in the first instance, it is recommended that the chip be re-tested on the portable Jelley Rig to confirm this. If the UV response has indeed started to deteriorate then re-flooding the device should be undertaken. If for some reason water vapour contamination of the chip surface is suspected, then a more rigorous process is required involving baking the CCD and/or cryostat at around 90 C - see echnical Note #104 for more details. he UV flooding process for the chip requires the presence of dry air (or dry Oxygen - this has been found to produce no significant improvement in respose and so dry air should be used in preference); the chip will therefore have to be brought up to room temperature for this operation to take place. he cryostat should be emptied of iquid Nitrogen the day before flooding is to be undertaken to allow the chip and inside of the cryostat to reach room temperature in a controlled fashion. he cryostat temperature should not be raised in an artificial manner as this may introduce water vapour onto the surface of the chip if this becomes cooler than the rest of the inside of the cryostat. Ensure that the chip is at ambient temperature before releasing the vacuum and opening the cryostat. Once the chip is at room temperature the next operation will be UV flooding of the chip surface. he UV lamp should be turned on 10 minutes prior to flooding to ensure adequate, even illumination during the flooding process. he output of the lamp after 10mins has been measured at RGO and found to be flat and even over the major part of its illumination area. Care must be taken to avoid the lamp output as the UV can be harmful to skin and eyes - wear goggles and gloves if you are working in close proximity to the lamp for some reason. Once the chip has reached room temperature the cryostat must be mounted on the urbo pump but the cryostat valve left closed so that it remains under vacuum. he line should be pumped down to a vacuum of, say, 10-4 mbar and then the cryostat valve opened to the line and the pump turned off. his will permit dry air to enter the cryostat through the sorb on the back of the pump. Once the cryostat is at ambient pressure the valve on the cryostat should be closed, this prevents any water vapour or other contamination of the inside of the cryostat. he CCD will need to be placed about 5cms in front of the UV lamp, the centre-line of the chip matching that of the lamp. he even illumination output of the lamp will result in an even flooding 4

response of the chip at this distance. It is not too critical (to +/-1cm). he chip should be exposed for about 30minutes, Mike esser reports that any time beyond 10mins does not result in any improvement of UV response. Once the flooding is complete the cryostat should be placed on the pump and evacuated to a few x 10-5. Once this vacuum is attained the cryostat should be cooled immediately. No more than a few hours should elapse between flooding and cooling otherwise the UV enhancement may start to deteriorate. Once the chip is cold, at the ING, the response may be measured as detailed in the next section. o conclude this section, there follows a simple recipe for flooding the oral CCDs- Allow the chip to naturally warm up to room temperature, When at ambient temperature, connect the cryostat to the turbo pump and evacuate the hose to the crysostat, When the line is at good vacuum, 10-4 mbar, open the cryostat valve to the pump, wait a few minutes to pump the whole system down a liitle further, Switch on the UV lamp in preparation for flooding, Switch off the pump. his will purge the cryostat through the pump, with dry air, Dry air is required for the flooding process, Close the cryostat valve, After the UV lamp has been on for 10 minutes, present the cryostat to the lamp for flooding, the distance from the cryostat window to the flooding lamp needs to be about 3.5cms, about 5cms from the chip, Allow the chip to be flooded for 30 minutes, Switch off the lamp, connect the cryostat to the pump and pump down the head, When a vacuum of around a few x 10-5 or better is attained, close off the cryostat valve, disconnect and immediately cool to operating temperature. 3. ING Response Measurement. If possible, a good indication of any fall in response can be obtained by measuring the direct current flowing between RD (Reset Drain) and SS (Substrate) using a pico-ammeter and a suitable connector to the chip. Filtered light at a wavelength of 400nm can easily indicate if there is a problem. For instance at RGO with this set-up, we have measured the room temperature response at 400 nm to be around 0nA. When the chip is operating correctly, this current is seen to increase slightly when the temperature drops to operating (-110 C). In any instance this current should not decrease. For instance, when the worse contamination occurred with another device at RGO, the Room emperature response at 400 nm was about 0nA. However, lowering the temperature produced a linear decrease in response down to around 8-9nA at operating temperature. his reflecting a loss in QE of about 50%. Warming the chip back to room temperature restored the Q.E. to about the level seen at the start. At the ING, the spectral response measurement can only be undertaken using the Portable Jelley Box which was (18/5/96) calibrated against ek5/ek3. he image mode measurement means that the CCD will have to be cold to obtain the response data. Below is data obtained for ek5 for the wavelength range 400-1000nm and with ek3 for the 350nm measurement (the change in head was due to ek5 head being required on the telescope) from which the response of the oral CCD can be obtained. (NOE: At the time quoted above, the 400-1000nm fluxes appeared to require a correction to the ORIGINA RGO flux calibration, of about x1.7 and the 350nm flux about x0.16 to obtain the Q.E. of the oral. his correction was required due to a change in the position of the internal OPA with respect to the calibration made at the RGO, this resulted in the signal at the output port being higher by the factors quoted above.) he data for ek5/ek3 is shown in the table below along with the known spectral response of these devices as measured at RGO. his enables the flux at the output port to be established and hence used to determine the response of oral CCD. Quick-speed, =-100 C Pixel size = 0.04mm x 0.04mm Gain of CCD = 1. e/adu for ek5 (400-1000nm) Gain of CCD = 1.83 e/adu for ek3 (350nm)* amp current of illumination source = 3.15 A 600x600 window used at 0,0 5

F# Wave time Bkgnd. Signal Net RGO Q.E. Derived ING (RGO) Flux (nm) (s) (adu) (adu) (adu) (%) (ph/s/mm ) 1 350 00 889 914 5 30* 134(8478) 400 100 459 356 797 55 136375(80390) 3 500 30 448 7968 750 67 100390(595036) 4 600 30 453 161 1708 7 643670(1558800) 5 700 0 455 1539 1084 73 395940(334400) 6 800 0 465 7113 6648 6 6005310(354040) 7 900 0 478 4984 4506 35 9496665(5593550) 8 1000 0 459 698 5839 10 858571(4867360) he Flux at each wavelength can be determined from the relation- F S = x G d t x 1 QE where F = the Flux (in photons/sec/mm ) at the given wavelength, S = the measured net signal (adu) d = the pixel size in mm, G = the measured Gain (e/adu), t = integration time in seconds, QE = known QE from measurements at RGO. For the ek3 CCD at 350nm this came out as F = x x ph s mm 5 183. 1 = 134 0. 04 00 0. 3 1 he oral was measured in the same way using the flux values from the table above. Gain=1. e/adu, quick speed, =-110, EEV9 CCD controller Pixel size = 0.015mm x 0.015mm amp current of illumination source=3.15a 600x600 window used at 0,0 for ORA1 F# Wave time Bkgnd. Signal Net ING ek Flux oral Q.E. (nm) (s) (adu) (adu) (adu) (ph/s/mm ) (%) 1 350 00 150 1537 35 134 71 400 100 1097 540 1443 136375 74 3 500 30 109 4744 365 100390 83 6

4 600 30 1109 105 11096 643670 97 5 700 0 1110 11613 10503 395940 93 6 800 0 1118 14056 1938 6005310 75 7 900 0 1114 189 11175 9496665 41 8 1000 0 1101 3056 1955 858571 8 he simplified relation for the QE of the oral (or any other chip) is- where S Q E S x G G x d. = xq. E d QE S S G G d d QE = QE of the oral = Net signal from oral (adu) = Net Signal from ek (adu) = Gain of oral (e/adu) = Gain of ek (e/adu) = pixel area of ek in mm = pixel area of oral in mm = Known QE of ek (RGO measurement) So that Q E x S. = 17. xq. E S For the ek#3 with a gain of 1.83 (e/adu) and oral #1 with a gain of 1. (e/adu) At the RGO these measurements will be made in either Diode Mode (see echnical Note #105) or in Imaging mode on the Jelley box. 4. Concluding Remarks he oral #1 Cryostat and the CCD have both been baked well to remove any water vapour contamination.. Unless there has been a SERIOUS ingress of water vapour, the cryostat does not have to be re-baked. his makes matters much easier as the chip itself can be processed, i.e. baked WIH SOME CARE, on its own. his, in both previous cases, was found to be the root cause of changes in response with temperature. Baking the cryostat alone, with no chip installed, did NO solve this problem. If however the cryostat has become contaminated then this will have to be baked in addition to the chip - see echnical Note #104 for details. 7