A TES Bolometer for THz FT-Spectroscopy

Similar documents
Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

Temperature and Synchrotron Radiation

M. Kenyon P.K. Day C.M. Bradford J.J. Bock H.G. Leduc

Characterization of high temperature solar thermal selective absorber coatings at operation temperature

Emissivity, Reflectivity and Transmissivity of Semitransparent Fibre Reinforced Plastic Composites

Ir TES electron-phonon thermal conductance and single photon detection

TES bolometer with non-andreev contacts

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

Superconducting Quantum Interference Devices for Current Sensing and Thermometry in the Millikelvin Range

Expressions for f r (T ) and Q i (T ) from Mattis-Bardeen theory

29 May 2015 Carbon Nanotube Based Detectors for THz Radiometry

Spectroscopic Measurements of Optical Elements For Submillimeter Receivers

THz direct detector with 2D electron gas periodic structure absorber.

Low Temperature Thermal Transport in Partially Perforated Silicon Nitride. Membranes

Coherent Synchrotron Radiation and Short Bunches in Electron Storage Rings. G. Wüstefeld, BESSY, Berlin (Germany)

Instrumentation for sub-mm astronomy. Adam Woodcraft SUPA, University of Edinburgh

The in-flight calibration system for the airborne imager GLORIA

Development of a Bolometer Detector System for the NIST High Accuracy Infrared Spectrophotometer

Zero-biased YBCO detectors for the real-time observation of coherent synchrotron radiation

SUPPLEMENTARY INFORMATION

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy

FIRST HIGH-RESOLUTION ANALYSIS OF PHOSGENE 35 Cl 2. CO AND 35 Cl 37 ClCO FUNDAMENTALS IN THE CM -1 SPECTRAL REGION

Comparison of photoexcited p-inas THz radiation source with conventional thermal radiation sources

Performance Testing of the Astro-H Flight Model 3-stage ADR

Output intensity measurement on a diagnostic ultrasound machine using a calibrated thermoacoustic sensor

Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1

SUPPLEMENTARY INFORMATION

Microspectroscopy and imaging in the THz range using coherent CW radiation

Measurement method for the proficiency testing program

1 THE CONCEPT OF TEMPERATURE

Analytical Chemistry II

arxiv: v1 [physics.ins-det] 6 Nov 2007

Total radiation measurements of thermodynamic temperature

Temperature Measurement

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

Performance of horn-coupled transition edge sensors for L- and S-band optical detection on the SAFARI instrument

Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR

File name: Supplementary Information Description: Supplementary Figures, Supplementary Notes and Supplementary References

phase retardance THz intensity ratio THz filling factor in air : 0.2 filling factor in si : 0.8 length of air : 4um length of si : 16um depth : 27.

transmission reflection absorption

Cosmic Microwave. 08/07/2017 Benson Bolometers and the CMB

Bilateral NIST-PTB Comparison of Spectral Responsivity in

High speed photon number resolving detector with titanium transition edge sensor

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

High Precision Dimensional Metrology of Periodic Nanostructures using Laser Scatterometry

CRYOGENIC CONDUCTION COOLING TEST OF REMOVABLE PANEL MOCK-UP FOR ITER CRYOSTAT THERMAL SHIELD

Position calibration of cryogenic dark matter detectors

A Broadband Flexible Metamaterial Absorber Based on Double Resonance

Simple strategy for enhancing terahertz emission from coherent longitudinal optical phonons using undoped GaAs/n-type GaAs epitaxial layer structures

Pixellated Planar Bolometer Imaging Arrays

Characterization of a high-performance Ti/Au TES microcalorimeter with a central Cu absorber

Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements

Optical Properties of Thin Semiconductor Films

Transition Edge Sensor Thermometry for On-chip Materials Characterization

ECE 695 Numerical Simulations Lecture 35: Solar Hybrid Energy Conversion Systems. Prof. Peter Bermel April 12, 2017

Resonator Fabrication for Cavity Enhanced, Tunable Si/Ge Quantum Cascade Detectors

Millimetre wave attenuation of prototype diagnostic components for the ITER bolometers

arxiv: v1 [physics.ins-det] 9 Apr 2016

FIRST carrier spacecraft

Development of Polarization Interferometer Based on Fourier Transform Spectroscopy for Thomson Scattering Diagnostics

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

Large Scale Polarization Explorer

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

Evaluating Labsphere s new UV-2000

Terahertz sensing and imaging based on carbon nanotubes:

Operator's Manual. optris BR 400. Calibration source

Calibrating the Thermal Camera

Supplementary material for High responsivity mid-infrared graphene detectors with antenna-enhanced photo-carrier generation and collection

Eugenio Monticone. Nanoscience and Materials Division Istituto Nazionale di Ricerca Metrologica

The Low Temperature Physics of Thin Films Superconducting Tin and Monolayer Graphene

Refractive indices of water and ice in the to 2.5-gm spectral range

Principles and Applications of Superconducting Quantum Interference Devices (SQUIDs)

Burner Tubing Specification for the Turbulent Ethylene Non-Premixed Jet Flame

Enhanced performance of microbolometer. using coupled feed horn antenna

Antiproton Decelerator at CERN, another advanced CCC for the FAIR project at GSI is under construction and will be installed in CRYRING. This

Gesellschaft für Schwerionenforschung mbh (GSI), Planckstrasse 1, D Darmstadt, Germany

EXPERIMENTAL DETERMINATION OF SPECTRAL AND ANGULAR DEPENDENT OPTICAL PROPERTIES OF INSULATING GLASSES

Low Vibration Cryogenic Equipment

Simplified Collector Performance Model

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

A Multipass Optics for Quantum-Well-Pumped Semiconductor Disk Lasers

DRIFT EFFECTS IN HgCdTe DETECTORS

Vibrational Spectroscopy of Molecules on Surfaces

08 - Miscellaneous and historical detectors

RASE 2009 Development of Sunshield Panels for Passive Radiant Cooler On Board Meteorological Instruments of ISRO.

Future Light Sources March 5-9, 2012 Low- alpha mode at SOLEIL 1

XI. Bolometers Principle XII. Bolometers Response. This lecture course follows the textbook Detection of

Terahertz imaging using the Jefferson Lab - FEL high power broadband terahertz source

Ultra-low noise HEMTs for deep cryogenic lowfrequency and high-impedance readout electronics

NIRSpec Performance Report NPR / ESA-JWST-RP Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.

Characterization of the 300 K and 700 K Calibration Sources for Space Application with the Bepicolombo Mission to Mercury

Temperature Dependence of a Macrobending Edge Filter Based on a High-bend Loss Fiber

TA/TI survey. Phy Phy

Voltage-biased superconducting transition-edge bolometer with strong electrothermal feedback operated at 370 mk

From Last Time. Partially full bands = metal Bands completely full or empty = insulator / seminconductor

Supplementary Figure 1 Schematics of an optical pulse in a nonlinear medium. A Gaussian optical pulse propagates along z-axis in a nonlinear medium

Bolometric Measurement of Beam Intensity at Cryogenic Windows

Force/Torque measuring facility for friction coefficient and multicomponent

Which Optistat is right for you?

Charge readout and double phase

Transcription:

A TES Bolometer for THz FT-Spectroscopy M. Kehrt, J. Beyer, C. Monte, J. Hollandt Physikalisch-Technische Bundesanstalt Abbestraße 2-12, Berlin, Germany E-Mail: Mathias.Kehrt@PTB.de Abstract - We recently reported on a Transition Edge Sensor (TES) composite bolometer for Fourier Transform spectroscopy which was designed, built and characterized. The design is based on a superconducting thermistor with SQUID readout, SiN membrane technology and a thin film metal mesh absorber which is optimized for the spectral range from 0.1 THz to 3 THz. A noise equivalent power (NEP) of 3.8 10-13 W/Hz 0.5 was achieved with a linear dynamic range of at least four orders of magnitude. Keywords (Index Terms) Bolometer, absorber, TES, teraherz, spectrometer, Fourier transform Received: June 25, 2016; Accepted: July 12, 2016. Reference No. ST522; Category 4. I. INTRODUCTION AND BACKGROUND We recently reported on a Transition Edge Sensor (TES) composite bolometer for Fourier Transform spectroscopy which was designed, built and characterized [1]. Since 2009, the Physikalisch-Technische Bundesanstalt (PTB) provides calibration services for THz power as the only national metrology institute worldwide [2]. The knowledge of the optical properties of the used components is of major interest to achieve lowest uncertainties. These are obtained by the Fourier Transform (FT) spectroscopy using cooled detectors. The presented TES-bolometer is designed to replace conventional semiconductor-based bolometers in FT spectrometers operating in the frequency range from 0.1 THz to 3 THz. In the design, special emphasis was given to attaining a high and spectrally uniform absorption in this spectral range as well as on a highly linear response, in order to avoid artifacts in processing the FT spectroscopy data. The composite design allows a separate optimization of all relevant components, especially of the absorber and the thermistor. II. ABSORBER AND THERMISTOR DESIGN One approach realizing the desired spectrally flat absorption spectrum is matching the sheet resistance of the absorber to half the vacuum impedance. By using a continuous metal layer with a thickness smaller than the skin depth and a sheet resistance of R = 188 Ω, a theoretical maximum absorption of 0.5 can be obtained [3]. Page 1 of 5

A sufficiently thin continuous metal layer cannot be fabricated in high quality using materials accessible in standard thin film deposition. Therefore, the absorbers here are based on microstructured metal meshes to match the vacuum impedance. Fig. 1. Microscopic image of a microstructured metal mesh absorber. Here the design with square loops insight each mesh is shown. Fig. 2. Spectrally flat absorption of the two selected absorber designs. Both designs are close to the theoretical absorption limit of 0.5. In order to achieve a spectrally flat absorption the structural dimensions are much smaller than the relevant wavelengths. Two different designs were investigated systematically. One design consists of a simple mesh, the other design has additional square loops inside each mesh cell as shown in Fig. 1. Two structures, one of each design, were identified as optimum absorbers. Both structures have a strip width of 3 µm. The structure without loops in the grid has a pitch of 10 µm, the structure with loops a pitch of 35 µm. The absorbers were lithographically fabricated on 1 µm thin SiN membranes. The metal has a thickness of about 100 nm with a corresponding sheet resistance of approx. 20 Ω. The absorption α determined for both variants is spectrally flat and near its theoretical maximum of 0.5 for this approach, as illustrated in Fig. 2. The results were obtained by reflectance ρ and transmittance τ measurements by α = 1 ρ τ. (1) A further requirement of the bolometer is a high dynamic range in combination with a highly Page 2 of 5

IEEE/CSC & ESAS SUPERCONDUCTIVITY NEWS FORUM (global edition), July 2016. linear response. To achieve this, the thermistor is realized as a superconducting transition edge sensor (TES) which utilizes a steep resistive transition [4]. Our TESs are fabricated from bilayers of a superconducting (niobium) and a normal conducting (aluminium) metal with a thickness of approx. 20 nm for each layer. The critical temperature of the bilayer is reduced compared to a single layer of niobium. This ensures a superconducting operation of the leads (niobium) while the TES is in its transition range (Fig. 3.). The TES shows a superconducting transition at Tc 7.8 K with a width of approx. 20 mk. The achieved Tc is suitable for heliumcooled bolometers as intended here. Operating the TES in strong voltage biased condition a negative electrothermal feedback (netf) extends the dynamic range [4] by ensuring that a constant sum of electrical heating power and incident radiative power stabilizes the TES at a fixed temperature inside the superconducting transition. The heating current, which is measured using an array of 16 Superconducting Quantum Interference Device (SQUID) current sensors, is directly related to the incident radiation. Fig. 3. Resistive transition of the TES and, at higher temperature, of its leads. Fig. 4. The TES-Bolometer, its mounting and its final setup. Left: Bolometer element with a TES in the centre, surrounded by the absorber mesh on a SiN membrane. Middle: Mounted bolometer with SQUID and wiring. Right: Assembly of filter wheel, Winston Cone, bolometer mount and radiation shield. Page 3 of 5

More thermal design details can be found in [1]. Several composite bolometers consisting of TES thermistors and both types of absorber structures were fabricated in a multilayer lithographic process on SiN membranes. On the left of Fig. 4 a bolometer element with a TES in the center is shown. The thermistor in the center is surrounded by the absorber with a physical dimension of 3.3 mm in square. The mounted bolometer with the SQUID readout and wiring is shown in the middle of Fig. 4. III. CHARACTERIZATION The complete detector was built up in a liquid helium cooled cryostat. The bolometer is shielded from the warmer surrounding inside of the cryostat by a gold plated housing. To reduce reflected FIR radiation the inner side of the housing is coated by grained SiC. The THz radiation to be detected passes a wedged diamond window and one of several cooled filters in a filter wheel and is concentrated by a Winston cone onto the bolometer (Fig. 4, right). For the electrical characterization the bolometer was operated in this environment without a radiant load by replacing one of the filters by a copper plate. The dark signal was recorded by a spectrum analyzer. The measured noise spectral density divided by the responsivity [5] leads to a noise equivalent power (NEP) of 3.8 10-13 W/Hz 0.5 for typical operating conditions. Additionally, the linear response range was determined with tunable THz radiation at the Metrology Light Source (MLS), the electron storage ring of PTB [6]. To investigate a power range over several orders of magnitude two operation modes of the MLS were used: normal operating mode providing incoherent THz radiation as well as low-α mode providing coherent THz radiation up to the mw range. In normal mode of operation four different ring currents were used. In low-α mode the MLS was operated at one ring current. For a finer attenuation of the radiant power two wire grid polarizers at different angle settings were utilized. In low-α mode the bolometer was compared with a calibrated broadband THz detector [7]. Here the bolometer shows a linear response up to 330 µw as shown in Fig. 5 in lightblue squares. In normal mode of operation the calibrated detector was not sensitive enough for a comparison. But here the radiant power is known to be strictly proportional to the ring current and the TES bolometer shows a linear response according to the calculated attenuation due to the angle settings of the polarizers. To determine the missing proportionality factor the results of the highest ring current of 140 ma were extrapolated and scaled to the results obtained in low-α mode. The compiled consistent results of all measurements at both operating modes of the MLS are shown in Fig. 5. The TES bolometer provides a linear response over a range of at least four orders of magnitude. Page 4 of 5

Fig. 5. The linear response range of the TES bolometer was determined to be four orders of magnitude by measurements at an electron storage ring (Metrology Light Source of PTB). Two modes of operation of the storage ring (normal mode and THz coherent synchrotron radiation CSR) with different ring currents were used to systematically vary the radiant power on the bolometer. REFERENCES [1] M. Kehrt, C. Monte, J. Beyer, and J. Hollandt, A highly linear superconducting bolometer for quantitative THz Fourier transform spectroscopy, Opt. Express, 23, 11170-11182, (2015). [2] A. Steiger,M. Kehrt; C. Monte, R. Müller, Traceable terahertz power measurement from 1 THz to 5 THz, Opt. Express, 21, 14466-14473, (2013). [3] W. Woltersdorff, Über die optischen Konstanten dünner Metallschichten im langwelligen Ultrarot, Z. Phys., 91, 230 252, (1934). [4] P. L. Richards, Bolometers for infrared and millimeter waves, Appl. Phys., 76, 1 24, (1994). [5] K. Irwin and G. Hilton. Transition-edge sensors. In C. Enss, editor, Cryogenic Particle Detection, vol. 99 of Topics. Appl. Phys., pp. 63 150. Springer, Berlin, Heidelberg, (2005). [6] J. Feikes, M. von Hartrott, M. Ries, et al., Metrology Light Source: The first electron storage ring optimized for generating coherent THz radiation, Phys. Rev. ST Accel. Beams, 14, 030705, (2011). [7] R. Müller, W. Bohmeyer, M. Kehrt, K. Lange, C. Monte, and A. Steiger, Novel detectors for traceable THz power measurements, J. Infrared Millim. Terahertz Waves, 35, 659-670, (2014). Page 5 of 5