ANTERIOR SEGMENT OPTICAL

Similar documents
CLINICAL SCIENCES. reported significant differences between thickness estimates from healthy and glaucomatous eyes Although promising as

Optical coherence tomography (OCT) is a ranging technique

Obstruction of trabecular orifices in primary open-angle glaucoma

Multiple patterns of angle closure mechanisms in primary angle closure glaucoma in Chinese

Thickness and Birefringence of Healthy Retinal Nerve Fiber Layer Tissue Measured with Polarization-Sensitive Optical Coherence Tomography

Scattering properties of the retina and the choroids determined from OCT-A-scans

MODELING THE FLOW OF AQUEOUS HUMOR IN POSTERIOR CHAMBER. Ram Avtar, Swati Srivastava 1

In Vivo Total Retinal Blood Flow Measurement by Fourier. domain Doppler Optical Coherence Tomography

Descemet stripping endothelial keratoplasty (DSEK) 1 3

3D FEM temperature distribution analysis of the human eye exposed to laser radiation

Clinical Study Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster

Quantitative retinal blood flow measurement with. three-dimensional vessel geometry determination. using ultrahigh-resolution Doppler optical

Blood flow measurement and slow flow detection in retinal vessels with Joint Spectral and Time domain method in ultrahigh speed OCT

Abstract. 1 Introduction

Thermal modelling of the human eye exposed to infrared radiation of 1064 nm Nd:YAG and 2090 nm Ho:YAG lasers

Quantitative Analysis of the Effect of Optical Probe Positioning on the Measurements of Flattened Diameter of Ocular Cornea for Applanation Tonometry

Numerical solution of ocular fluid dynamics

Iris Biomechanics in Health and Disease

Truncated logistic regression for matched case-control studies using data from vision screening for school children.

COMMUNICATIONS. OPEN-ANGLE GLAUCOMA*t PATHOLOGY OF TRABECULAR NERVES IN. (Wolter, 1959). As long as not even the normal functions of the trabecular

Identify the structure labelled 1.

A theoretical model of the human eye based on ultrasound and corneal data

Comparison of keratometric and topographic cylinder and axis measurements on normal corneas with low astigmatism

University of Southampton Research Repository eprints Soton

-I (PH 6151) UNIT-V PHOTONICS AND FIBRE OPTICS

Long Path Industrial OCT High-precision Measurement and Refractive Index Estimation

Assignment 3 Due September 27, 2010

Fluid and structure coupling analysis of the interaction between aqueous humor and iris

For more information, please contact

Thermal Modelling of the Human Eye Exposed to Laser Radiation

Eschenbach Low Vision Training Program

Michelson Interferometer. crucial role in Einstein s development of the Special Theory of Relativity.

Computer-based analysis of rhegmatogenous retinal detachment

Measurement of Corneal Thickness by Loser Doppler Interferometry

Biology 224 Human Anatomy and Physiology - II Week 1; Lecture 1; Monday Dr. Stuart S. Sumida. Review of Early Development of Humans.

Supplementary Figure 1. Characterization of the single-photon quantum light source based on spontaneous parametric down-conversion (SPDC).

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 FRAUNHOFER DIFFRACTION

Experimental research on intraocular aqueous flow by PIV method

The Integration of the Corneal and Limbal Fibrils in the Human Eye

I n previous communications 1 ' 2 a model

NUMERICAL SIMULATION OF AQUEOUS HUMOR FLOW: FROM HEALTHY TO PATHOLOGIC SITUATIONS J.A. FERREIRA, P. DE OLIVEIRA, P. M. DA SILVA AND J.N.

Comparison of noncontact infrared and re m o t e sensor thermometry in normal and dry eye p a t i e n t s

Coherence and width of spectral lines with Michelson interferometer

Short-Pulsed Neodymium-YAG Laser Trabeculotomy

The Excimer Laser and Applications to Corneal Ablative Surgery

THE USE OF ULTRASOUND IN OCULAR BIOMETRY

Experiment 3 The Simple Magnifier, Microscope, and Telescope

Michael K. Smolek, Ph.D. Associate Professor of Ophthalmology LSUHSC School of Medicine New Orleans, LA

ACCURATE PREDICTION OF

Deviations of Lambert-Beer s law affect corneal refractive parameters after refractive surgery

Research Article Clinical Validation of Adjusted Corneal Power in Patients with Previous Myopic Lasik Surgery

ENDOTHELIAL CELL VACUOLATION IN THE CORNEA*

Theory and calibration of the Schi0tz tonometer

Implant Design for Ophthalmic Delivery: The place of materials and fabrication methods

An electrical model of the human eye

Resolution improvement in optical coherence tomography with segmented spectrum management

Trabecular Metal Material Spine Overview. Page 1 of 6

Human Anatomy & Physiology

Comparison of ocular response analyzer, dynamic contour tonometer and Goldmann applanation tonometer

Measurements, applications, and impact of the refractive index of extracellular vesicles

Flow in the anterior chamber of the eye with an implanted iris-fixated artificial lens

Computational model for heat transfer in the human eye using the finite element method

Steady State Heat Analysis of the Eye Using Finite Element Method

X-Ray Diagnostics Commissioning at the LCLS

Flexible A-scan rate MHz OCT: Computational downscaling by coherent averaging

Experiment 3 The Simple Magnifier, Microscope, and Telescope

Advanced Laser Surface Analyzer. Francis V. Krantz Ashland Inc Blazer Parkway Dublin, Ohio Abstract

STA January 11, Object Data is Big Data of Complex Type. In this class, the focus is on the complexity of data,

Being a Physicist Unit 5. Summary Sheets. Gleniffer High School

Use the correct answer from the box to complete the sentence. diffraction reflection refraction

JOURNAL OPHTHALMOLOGY THE BRITISH NOVEMBER, 1922 COMMUNICATIONS ACCOMMODATION AND OTHER OPTICAL SECTION I.

P3 Revision Pack (Please keep this pack with you)

A system of two lenses is achromatic when the separation between them is

Multi-cycle THz pulse generation in poled lithium niobate crystals

Human Eye Response to Thermal Disturbances

Degradation in the degree of polarization in human retinal nerve fiber layer

Performance of a differential contrast sensitivity method to measure intraocular scattering

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY Translated English of Chinese Standard: YY

University of Genoa. PhD program in Civil, Chemical and Environmental Engineering Fluid Dynamics and Environmental Engineering. Cycle XXX.

ABB temperature measurement Radiation thermometry. Measurement made easy. Process temperature measurement practice--non-contacting

Unit P3: Applications of Physics

A Question. Simple Magnifier. Magnification by a Lens 11/29/2011. The last lecture

The Accuracy of Tower' Maps to Display Curvature Data in Corneal Topography Systems

FORMULA FOR THE DEPTH OF THE FEMTOTUNNEL IN THE IMPLANTATION OF THE KERARING SI 5, SI 6 BY MEANS OF INTRALASE ifs 150

Human Trabecular Meshwork Cell Survival Is Dependent on Perfusion Rate

Relation Between Superficial Capillaries and Foveal Structures in the Human Retina

Illuminated Reticle Technologies for Rifle Scopes. Illuminated Reticle Technologies for Riflescopes

Optical model for light distribution during transscleral cyclophotocoagulation

Lens internal curvature effects on age-related eye model and lens paradox

HUMAN EYE AND THE COLOURFUL WORLD

WHAT IS THE MOORFIELDS MDT? MOORFIELDS MDT TEST TASK EASILY UNDERSTOOD. New advances in Moorfields Motion Displacement Perimetry

DIAGNOSIS OF FULL THICKNESS ROTATOR CUFF TEARS

Dominant optic atrophy in a Japanese family with OPA1 frameshift mutation (V942fsX966)

Operating Instructions Spectro-Goniometer Student. 1 Functional Elements. 2 Safety Precautions. Figure 1: Spectro-Goniometer Student

Propagation and Phase Reconstruction of ocular wavefronts with SAR techniques

Optical Instruments. Chapter 25. Simple Magnifier. Clicker 1. The Size of a Magnified Image. Angular Magnification 4/12/2011

MAPPING FRACTURE APERTURES USING MICRO COMPUTED TOMOGRAPHY

OPTO 5320 VISION SCIENCE I

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego

Retinal Nerve Fiber Layer Thickness in Dominant Optic Atrophy

Transcription:

CLINICAL SCIENCES High-Definition Optical Coherence Tomography Imaging of the Iridocorneal Angle of the Eye Hon-Tym Wong, FRCS(Ed); Marcus C. Lim, MRCS(Ed); Lisandro M. Sakata, MD, PhD; Han T. Aung, MBBS; Nishani Amerasinghe, MRCOphth; David S. Friedman, MD, MPH, PhD; Tin Aung, MBBS, FRCS(Ed), PhD Objective: To assess the ability of high-definition optical coherence tomography (HD-OCT) to image the anterior chamber angle. Methods: Forty-five consecutive subjects with phakic eyes underwent gonioscopy and anterior chamber angle imaging with HD-OCT adapted with a 60-diopter aspheric lens mounted over the imaging aperture. The patients fixation was directed to the side using an external fixation light, and scans were taken of the temporal and nasal quadrants. The visibility of angle structures was assessed and the diagnosis of angle closure using HD-OCT was compared with that of gonioscopy. Results: The majority of subjects were Chinese (91.1%) and female (62.2%). Cross-sectional HD-OCT allowed in vivo visualization of the scleral spur in 71 of 90 quadrants (78.9%) and the termination of the Descemet membrane (Schwalbe line) in 84 of 90 quadrants (93.3%). It was possible to image the trabecular meshwork in 56 quadrants (62.2%). Angle closure was observed in 17 eyes with gonioscopy and 12 eyes with HD-OCT (P=.12, McNemar test). The 2 modalities showed good agreement for angle closure diagnosis by quadrant ( =0.65). Conclusion: The adapted HD-OCT provided magnified views of the anterior chamber angle and allowed visualization of the Schwalbe line and trabecular meshwork in most eyes. Arch Ophthalmol. 2009;127(3):256-260 Author Affiliations: Tan Tock Seng Hospital (Dr Wong), Singapore Eye Research Institute and Singapore National Eye Center (Drs Lim, Sakata, H. T. Aung, Amerasinghe, and T. Aung), and Yong Loo Lin School of Medicine, National University of Singapore (Dr T. Aung), Singapore; and Wilmer Eye Institute and Johns Hopkins Bloomberg School of Public Health, Baltimore, Maryland (Dr Friedman). ANTERIOR SEGMENT OPTICAL coherence tomography (AS-OCT) uses a 1310-nm diode laser to obtain realtime images of the anterior chamber angle. 1-4 Recent studies have shown that AS-OCT tends to confirm gonioscopic angle closure in most patients but also detects more closed angles than gonioscopy. 5,6 This discrepancy in findings between gonioscopy and AS-OCT may be due to the different anatomical landmarks and levels of irido-angle contact used to define a closed angle. On AS-OCT images, it is not possible to determine the location of the trabecular meshwork, and the presence of any contact between the iris and the angle wall anterior to the scleral spur is graded as angle closure. 5 However, if this apposition did not reach the level of the posterior trabecular meshwork, the quadrant would be considered open on gonioscopy. l spur determination is impossible in approximately 20% to 30% of AS-OCT images. This could be because of low lateral resolution of AS-OCT images, image artifacts from eyelids or other structures, anatomical variations in eyes, and technical difficulties in imaging some quadrants. 7 This inability to detect the scleral spur may limit the accuracy and usefulness of angle imaging, especially for quantitative analysis of the angle that uses the scleral spur as an anatomical landmark for measurements. 8,9 High-definition OCT (HD-OCT) has been recently introduced for imaging the retina. 10-13 High-definition OCT uses a scanning rate 50 to 60 times faster than current time-domain OCT devices, 12,13 and with an axial resolution of 3 to 5 µm, the higher resolution of HD-OCT allows the identification of retinal layers and enhanced visualization of the macula and optic nerve head. 10,11 A prototype HD-OCT device was also recently used as a noninvasive corneal imaging modality that was capable of in vivo differentiation of corneal layers and demonstration of pathologic abnormalities in the cornea. 14 The aim of this study was to evaluate a modification of a commercially available HD-OCT device (Cirrus-OCT; Carl Zeiss Meditec, Dublin, California) adapted for imaging the anterior chamber angle. 256

METHODS In this hospital-based study, consecutive consenting subjects were recruited from a glaucoma clinic at a Singapore hospital. Informed consent was obtained from all participants, and the study had the approval of the hospital s institutional review board and adhered to the tenets of the Declaration of Helsinki. After an interview about medical and ophthalmic history, each subject underwent the following examinations on the same day: slitlamp biomicroscopy, gonioscopy, and anterior segment imaging (see later). Eyes were excluded if there was a history of previous intraocular surgery or penetrating trauma or any cornea opacities or abnormalities that precluded angle imaging; however, those who had previously undergone laser iridotomy were not excluded. GONIOSCOPY Gonioscopy was performed in the dark in all cases by a single examiner with glaucoma fellowship training (H.T.W.) who was masked to OCT findings. Static and dynamic gonioscopy was performed using a Sussman 4-mirror lens, at high magnification ( 16), with the eye in the primary position of gaze. A 1-mm light beam was reduced to a very narrow slit, and care was taken to avoid light falling on the pupil and accidental indentation during examination. Slight tilting of the gonioscopy lens was permitted in an attempt to gain a view over the convexity of the iris. The angle in each quadrant was graded using the Scheie grading system, which is based on the anatomical structures observed during gonioscopy (grade I=visible ciliary body, grade II=visible scleral spur, grade III=visible anterior trabecular meshwork, grade IV=angle structures not visible). 15 On gonioscopy, an angle quadrant was considered closed if the posterior trabecular meshwork could not be seen in the primary position without indentation (Scheie grade III or IV). HD-OCT IMAGE ACQUISITION Seated subjects were examined by a single examiner (H.T.A.) masked to other test results. A 60-diopter (D) aspheric lens (Volk Optical, Mentor, Ohio) was mounted over the HD-OCT (Cirrus- OCT) imaging aperture, with the same distance from the Volk lens to the instrument s entrance pupil as one would normally use between this lens and the patient s entrance pupil during fundus examination. The magnification of the 60-D lens increases the transverse scan range of the HD-OCT from 6 mm SL Figure 1. A cross-sectional high-definition optical coherence tomography image of the angle showing the scleral spur () and the termination of the Descemet membrane (Schwalbe line [SL]). The trabecular meshwork () is represented by a triangular dark area between the and SL. to 7.2 mm. The HD-OCT instrument was set for full myopic correction, retracting the HD-OCT imaging aperture and Volk lens so as to allow room for the patient at the same working distance as in the usual configuration for retinal scanning. For angle imaging, the patient s fixation was directed to the side of the instrument, using an external fixation light so that the iridocorneal angle was centered in the instrument s field of view. Two scans of the angle of each eye were obtained with HD-OCT in the dark: one image scanning the angle at the 3-o clock position and the other at the 9-o clock position, representing the temporal and nasal quadrants of each eye. The superior and inferior quadrants were not scanned because of technical difficulties in scan acquisition caused by the eyelids. In a subset of the last 31 consecutive patients, anterior segment imaging of the nasal and temporal quadrants was also performed under the same illumination conditions using a commercially available AS-OCT device (Visante; Carl Zeiss Meditec). IMAGE ANALYSIS All HD-OCT and AS-OCT scans were analyzed separately by an examiner with glaucoma subspecialty training (T.A.) masked to other test results. For HD-OCT images in which the trabecular meshwork was visible, angle closure in a quadrant was diagnosed if there was contact between the iris and the trabecular meshwork. For images in which trabecular meshwork was not delineated, angle closure was diagnosed if there was any contact between the iris and angle wall anterior to the scleral spur. For the comparison between HD-OCT and AS-OCT, the angle was considered closed on AS-OCT and HD-OCT imaging if any contact between the iris and angle wall anterior to the scleral spur was noted. The intraobserver reproducibility for the assessment of angle closure in HD-OCT images was assessed in a random subset of 20 subjects (20 eyes, 40 quadrants). Images of the quadrants of the eye were graded for angle closure in 2 sessions separated by an interval of 1 week by a single examiner (T.A.) masked to other results. STATISTICAL ANALYSIS One eye per patient was selected for analysis; this was the right eye if both eyes fulfilled the inclusion criteria. Parametric and nonparametric tests were used to compare continuous variables according to data distribution. The 2 or Fisher exact test was used to compare categorical data. The McNemar test was used to compare differences in distribution of a categorical variable between 2 dependent samples. The statistic was used to assess the agreement between categorical variables. Statistical significance was set at P.05 and statistical analysis was performed using SP (version 11.5; SP Inc, Chicago, Illinois). RESULTS The mean (SD) age of the 45 participants was 62.5 (9.1) years and the majority of subjects were female (62.2%) and Chinese (91.1%). Cross-sectional HD-OCT allowed in vivo visualization of the scleral spur in 71 of 90 quadrants (78.9%) and the termination of the Descemet membrane (Schwalbe line) in 84 of 90 quadrants (93.3%) (Figure 1). There was no difference in the visibility of the scleral spur (P.99, McNemar test) and Schwalbe line (P.99, McNemar test) between nasal and temporal quadrants. It was possible to image the trabecular meshwork in 56 257

Cirrus 1 1 0 6 2 2 2 AC Gonioscopy Visante Figure 2. A high-definition optical coherence tomography scan showing angle closure (AC) where there is iris-angle apposition including the region of the trabecular meshwork (), which is anterior to the scleral spur (). Figure 3. Venn diagram showing the agreement between gonioscopy, Cirrus HD-OCT (Carl Zeiss Meditec, Dublin, California), and Visante AS-OCT (Carl Zeiss Meditec) in detecting angle closure. quadrants (62.2%; 23 nasal and 33 temporal quadrants) (Figure 1). This structure was represented as a triangular dark area between the scleral spur and the Schwalbe line (Figure 1). The angle could not be graded as open or closed in 3 quadrants (3.3%) because of limited visualization of the angle structures. A closed angle in either quadrant was observed in 17 eyes (37.8%) with gonioscopy and 12 eyes (26.7%) with HD-OCT (P=.12, McNemar test) (Figure 2). The 2 modalities showed good agreement ( =0.65) for angle closure diagnosis by quadrant. Identical results were obtained when considering all 4 quadrants for gonioscopy, as all eyes with a closed angle in any quadrant had either the nasal or temporal quadrant also closed. One eye was identified as closed by HD-OCT in the 28 eyes that were open both nasally and temporally by gonioscopy. The intraobserver reproducibility for the assessment of angle closure in HD-OCT images was very good ( =0.81; 95% confidence interval [CI], 0.60-1.00). Thirty-one subjects underwent imaging with both HD-OCT and AS-OCT. The Venn diagram (Figure 3) shows the agreement between gonioscopy, HD-OCT, and AS-OCT in detecting eyes with angle closure. Angle closure in a nasal or temporal quadrant was observed in 11, 10, and 8 eyes with gonioscopy, AS-OCT, and HD-OCT, respectively (gonioscopy vs AS-OCT, =0.64; McNemar, P.99; gonioscopy vs HD-OCT, =0.63; McNemar, P =.38; AS-OCT vs HD-OCT, =0.53; McNemar, P=.69). Anterior segment OCT identified 8 of 11 eyes while HD-OCT identified 7 of 11 eyes diagnosed with angle closure on gonioscopy (P.99, Fisher exact test). Once again, identical results were obtained when considering all 4 quadrants for gonioscopy, as all eyes with a closed angle in any quadrant had either the nasal or temporal quadrant also closed. An open angle in all quadrants was found in 20 eyes on gonioscopy. Of these, 2 eyes and 1 eye were identified as closed by AS-OCT and HD-SD-OCT, respectively. Using gonioscopy as the reference standard, the specificity of AS-OCT was 88.9% (95% CI, 80.1%-97.7%) and HD-OCT, 96.4% (95% CI, 91.0%-100.0%), while the sensitivity of AS-OCT and HD-OCT were 71.4% (95% CI, 58.8%-84.1%) and 64.7% (95% CI, 50.7%-78.7%), respectively. COMMENT Modification of a commercially available HD-OCT device with the use of a 60-D lens enabled high-resolution imaging of the anterior chamber angle. In addition to the scleral spur, we were able to identify a new imaging landmark for the angle, namely the termination of the endothelium/descemet membrane at the Schwalbe line. Grading an angle as open or closed based on apposition of the iris to the area between these landmarks (where the trabecular meshwork lies) would thus be possible. Furthermore, the association of partial trabecular meshwork closure and intraocular pressure could be investigated. This is a distinct advantage over AS-OCT imaging of the angle, in which only the scleral spur is visible (though only in about 70%-80% of images) and angle closure is diagnosed based on the presence of any contact between the iris and the angle wall anterior to the scleral spur. It was possible in some eyes to identify what seems to be the trabecular meshwork (Figure 1). This is an exciting observation as it would allow even more accurate phenotyping of the angle status by the presence of direct iris apposition to the trabecular meshwork in real time and in a noncontact fashion. In 1 case, we were also able to identify peripheral anterior synechiae in the angle (Figure 4). Although the resolution of the HD-OCT was sufficiently high for identification of these structures, penetration of HD-OCT was limited and it was not possible to image the ciliary body. Thus, it would not be possible to evaluate underlying mechanisms for angle closure such as plateau iris, where a large ciliary body and/or anteriorly rotated ciliary processes have been shown to hold the peripheral iris in apposition to the trabecular meshwork. Also, the depth and width of view provided by HD-OCT precluded imaging of the entire angle up to the iris root, and the images obtained may thus be unfamiliar to most ophthalmologists who expect to view the angle in its entirety, such as in images produced by AS-OCT. We believe, however, that imaging iris trabecular meshwork contact is the most crucial feature in diagnosing angle closure and so we may not actually need to see the entire angle. Overall, the agreement in detecting a closed angle quadrant using HD-OCT and gonioscopy was good, with a 258

A PAS Figure 4. A high-definition optical coherence tomography scan showing peripheral anterior synechiae (PAS) in the angle (arrow). B SL Gap Figure 5. A high-definition optical coherence tomography scan showing what appears to be a gap between the iris and angle wall (arrow); however, the quadrant was graded as closed on gonioscopy. Other structures visible include the scleral spur (), the termination of the Descemet membrane (Schwalbe line [SL]), and trabecular meshwork (). of 0.65. However, when we compared the gonioscopy findings with those of HD-OCT, we found that the rate of angle closure diagnosis was lower using HD-OCT, with some eyes graded as open on HD-OCT but closed on gonioscopy. This difference could in part be explained by the presence of a gap between the cornea and sclera such that the iris does not seem to be in apposition with the angle. Another possibility is that HD-OCT images using our jury-rigged device are likely distorted since they are not dewarped or corrected for the index of refraction transition at the air-tear interface and the different group indexes in air, cornea, and aqueous. 9 An example of an eye in which there was disagreement between HD-OCT and gonioscopy is seen in Figure 5, where there appears to be a gap between the iris and the angle wall on the HD-OCT scan when this quadrant was graded as closed on gonioscopy. When HD-OCT was compared with AS-OCT and gonioscopy, we found that AS-OCT detected 2 eyes as having closed angles that were found to be open on gonioscopy and HD-OCT. The higher rate of angle closure diagnosis by AS-OCT compared with gonioscopy has been found in previous studies. 5,6 Figure 6 illustrates an eye in which AS-OCT showed a closed angle in the nasal quadrant (left quadrant of Figure 6A) but HD-OCT found that Figure 6. Anterior segment optical coherence tomography (AS-OCT) and high-definition OCT (HD-OCT) of the nasal quadrant. A, An AS-OCT scan showing a closed angle in the nasal quadrant. B, An HD-OCT scan showing the same nasal quadrant. There is no contact between the iris and angle wall as there appears to be a gap (large arrow). Other structures visible include the scleral spur () and trabecular meshwork (). there was no contact between the iris and the angle wall (Figure 6B). In this case, we wonder if there was a shadow created by the cornea such that the iris apposition to the angle wall could not be seen clearly with the HD device. Compared with gonioscopy, the use of imaging devices such as HD-OCT and AS-OCT to evaluate the angle is also associated with several distinct disadvantages. While the angle grading by gonioscopy is typically for a whole quadrant, the OCT image is only for a section in that quadrant. Thus, the patient who has a closed angle for the most part in a quadrant by gonioscopy may have a slice in that region that is open. In addition, one is unable to perform a dynamic analysis of the angle or to image pigment or other deposits in the angle. Our study had some limitations. For angle imaging using the HD-OCT, the patient s fixation was directed outside the instrument using an external fixation light so that the iridocorneal angle was centered in the instrument s field of view. This external fixation could not be completely standardized and thus the images may not have been centered alike in all cases. As noted earlier, the images obtained with this adapted device are not dewarped, limiting the accuracy of angle findings somewhat. For AS- OCT, we used the normal imaging mode and not the highresolution mode, which may have better visualization of angle structures. Finally, the use of a single observer could result in a systematic bias of the gonioscopy findings (and the OCT analysis). 259

In summary, we found that HD-OCT provided images of the anterior chamber angle with higher resolution than the current AS-OCT device. Despite the fact that the anterior chamber angle images were acquired by an adapted HD-OCT device, it was possible to identify a new anatomical landmark for anterior chamber angle evaluation, the Schwalbe line, which is the anterior boundary of the trabecular meshwork, as well as the trabecular meshwork itself. There are still limitations with this method, including the inability to clearly identify all the angle structures in every eye. With further optimization, HD-OCT has the potential to produce more precise determination of iris posterior trabecular meshwork contact and thus represents a promising imaging method for the detection of eyes at risk of angle closure. More studies are under way to evaluate the use of HD-OCT in angle imaging, including ways to obtain a 3-dimensional image of the angle. Submitted for Publication: May 21, 2008; final revision received July 10, 2008; accepted July 17, 2008. Correspondence: Tin Aung, MBBS, FRCS(Ed), PhD, Singapore National Eye Center, 11 Third Hospital Ave, Singapore 168751 (tin11@pacific.net.sg). Author Contributions: Drs Wong and Lim contributed equally to the work and are joint first authors. Financial Disclosures: Dr Wong has received financial support and honoraria for travel to conferences from Carl Zeiss Meditec and Heidelberg Engineering. Dr Friedman has received an instrument loan and has been a consultant for Carl Zeiss Meditec. Dr T. Aung has received grant funding as well as financial support and honoraria for travel to conferences from Carl Zeiss Meditec. REFERENCES 1. Izatt JA, Hee MR, Swanson EA, et al. Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol. 1994; 112(12):1584-1589. 2. Radhakrishnan S, Rollins AM, Roth JE, et al. Real-time optical coherence tomography of the anterior segment at 1310 nm. Arch Ophthalmol. 2001;119 (8):1179-1185. 3. Radhakrishnan S, Goldsmith J, Huang D, et al. Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol. 2005;123(8):1053-1059. 4. Radhakrishnan S, Huang D, Smith SD. Optical coherence tomography imaging of the anterior chamber angle. Ophthalmol Clin North Am. 2005;18(3):375-381. 5. Nolan WP, See JL, Chew PT, et al. Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes. Ophthalmology. 2007; 114(1):33-39. 6. Sakata LM, Lavanya R, Friedman DS, et al. Comparison of gonioscopy and anteriorsegment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle. Ophthalmology. 2008;115(5):769-774. 7. Sakata LM, Lavanya R, Friedman DS, et al. Assessment of the scleral spur in anterior-segment ocular coherence tomography images. Arch Ophthalmol. 2008; 126(2):181-185. 8. Ishikawa H, Liebmann JM, Ritch R. Quantitative assessment of the anterior segment using ultrasound biomicroscopy. Curr Opin Ophthalmol. 2000;11(2):133-139. 9. Goldsmith JA, Li Y, Chalita MR, et al. Anterior chamber width measurement by highspeed optical coherence tomography. Ophthalmology. 2005;112(2):238-244. 10. Drexler W, Morgner U, Ghanta RK, Kärtner FX, Schuman JS, Fujimoto JG. Ultrahigh resolution ophthalmic optical coherence tomography. Nat Med. 2001; 7(4):502-507. 11. Ko TH, Fujimoto JG, Duker JS, et al. Comparison of ultrahigh- and standardresolution optical coherence tomography for imaging macula hole pathology and repair. Ophthalmology. 2004;111(11):2033-2043. 12. Wojtkowski M, Bajraszewski T, Gorczynska I, et al. Ophthalmic imaging by spectral optical coherence tomography. Am J Ophthalmol. 2004;138(3):412-419. 13. Mumcuoglu T, Wollstein G, Wojtkowski M, et al. Improved visualization of glaucomatous retinal damage using high-speed ultrahigh-resolution optical coherence tomography. Ophthalmology. 2008;115(5):784-791. 14. Christopoulos V, Kagemann L, Wollstein G, et al. In vivo corneal high-speed, ultra high-resolution optical coherence tomography. Arch Ophthalmol. 2007; 125(8):1027-1035. 15. Scheie HG. Width and pigmentation of the angle of the anterior chamber: a system of grading by gonioscopy. Arch Ophthalmol. 1957;58(4):510-512. From the Archives of the Archives 140 Years Ago T he seizing of the iris presents difficulties in many cases of iridectomy. Although the incisions may have been properly made, and the forceps introduced in the usual manner, either they do not seize the iris at all, or only tear out a small piece of it.... For such cases I have now had the situation of the teeth of the forceps altered, in such a manner that the surface in which they grasp is turned to a right angle. Reference: Liebreich R. On a modification of iridectomy forceps. Arch Ophthalmol. 1869;1:22-24. 260