SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Similar documents
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ALLOMETRICS, INC Bayport Blvd. Seabrook, TX Terry Baldwin Phone: CALIBRATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. GREENSLADE & CO., INC Wenneca Avenue Fort Worth, TX Larry Borowski Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. J. A. KING & COMPANY, LLC 8813 Old Craft Road Olive Branch, MS CONNIE FOSTER PHONE:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ADVANCED TECHNOLOGY SERVICES HMA 1800 Honda Drive Lincoln, AL Lasheila Hayes Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z ALPHAGAGE th Avenue Rockford, IL James Hoard Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

ACCREDITED LABORATORY. INSPECTION MEASUREMENT COMPANY Wyoming, MI

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

Scope of Accreditation For ISOLAB Inc. Calibration

Certificate of Accreditation to ISO/IEC 17025:2005

Certificate of Accreditation

Certificate of Accreditation

Certificate of Accreditation

CERTIFICATE OF ACCREDITATION

For the National Voluntary Laboratory Accreditation Program

For the National Voluntary Laboratory Accreditation Program

A2LA. G118: Guidance for Defining the Scope of Accreditation for Calibration Laboratories

Perry Johnson Laboratory Accreditation, Inc. Calibration Scopes of Accreditation

For the National Voluntary Laboratory Accreditation Program

CERTIFICATE OF ACCREDITATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Ackred. nr 1678 Kalibrering ISO/IEC 17025

National Voluntary Laboratory Accreditation Program

SCOPE OF ACCREDITATION

Mason Technology Ltd

Document No: TR 12 Issue No: 1

CERTIFICATE OF CALIBRATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Mason Technology Ltd

ESTIMATION OF UNCERTAINTY IN HARDNESS MEASUREMENT OF RUBBER AND OTHER ELASTOPLASTIC MATERIALS

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

An Introduction to the Differences Between the Two Most Recognized Force Standards

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

CALIBRATION CERTIFICATE # 503

R SANAS Page 1 of 7

PERFORMANCE TEST REPORT. Rendered to: VELUX AMERICA, INC. PRODUCT: SUN TUNNEL Domes TYPES: Acrylic and Polycarbonate

Physical Properties Testing Technical Bulletin

CALIBRATION REPORT FOR THERMOHYDROMETER

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

CEEN 3320 Behavior & Properties of Engineering Materials Laboratory Experiment No. 1 Measurement Techniques

ACCREDITED CALIBRATION AND MEASUREMENT CAPABILITY SCOPE. Ring gaugeф(50~100)mm U=1.1 μm

Uncertainty Propagation for Force Calibration Systems

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

CERTIFICATE OF CALIBRATION

Calibration Traceability Guidelines

APEX DYNAMICS, INC. Staianless

MEASUREMENT TRACEABILITY AND ERRORS RELATED TO FORCE MEASUREMENT

Maha Ireland Ltd. 629 Jordanstown Avenue, Greenogue Business Park, Rathcoole, Co. Dublin. Calibration Laboratory Registration number: 287C

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Vocabulary of Metrology

บร การสอบเท ยบ และทดสอบเคร องม อว ด

COMMON MISTAKES IN DIMENSIONAL CALIBRATION METHODS

Reliable Test Results

New Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018

CERTIFICATE OF CALIBRATION

GB/T / ISO 527-1:1993

A Case for Periodic Calibration or Verification of RTDs

Hiroshi Sawai TKG Laboratory Center International Standardisation Office

Transcription:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 POLYMER TESTING INSTRUMENTS, INC. 10 National Drive Fletcher, NC 28732 Larry Eisner Phone: 828 252 1326 CALIBRATION Valid To: June 30, 2019 Certificate Number: 1504.01 In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this laboratory to perform the following calibrations 1 : I. Dimensional Cutting Dies Up to 8 in 310 μin Caliper, optical Dial and Electronic Up to 1 in 120 μin Gage blocks Indicators 3 Calipers 3 Up to 8 in 310 μin Caliper master Micrometers 3 Up to 1 in 140 μin Gage blocks (A2LA Cert. No. 1504.01) Revised 09/28/2017 Page 1 of 6

II. Electrical DC/Low Frequency Electrical Calibration of Thermocouple Indicators 3 Generate and Measure Type J Type K Type T (-210 to 1200) ºC (-200 to 1372) ºC (-250 to 400) ºC 0.22 ºC 0.23 ºC 0.23 ºC Omega / Martel CL3001 Electrical Calibration of RTD Indicators 3 Generate and Measure Pt 385, 100 Ω Pt 385, 1 kω Pt 3902, 100 Ω Pt 3926, 100 Ω Cu 427, 10 Ω (-200 to 800) ºC (-190 to 630) ºC (-200 to 630) ºC (-200 to 630) ºC (-100 to 260) ºC Omega / Martel CL3001 III. Mechanical Force Measure 3 (0 to 500) lbf 0.062 % of full scale ASTM E4 w/ load cells Force Gages and Tensile Testers (200 to 2000) lbf (2000 to 5000) lbf 0.14 % of full scale 0.14 % of full scale Load cell comparison Crosshead Speed (0 to 20) in/min 0.037 in/min Calibrated scale and stop watch Torque Wrenches (0 to 177) in lbf 0.76 in lbf Torque tester Pressure 3 (0 to 2000) psi (2000 to 10 000) psi 0.11 % of full scale 0.11 % of full scale Deadweight tester Vacuum 3 (0 to 30) inhg 0.71 inhg Master gage lab transducer field (A2LA Cert. No. 1504.01) Revised 09/28/2017 Page 2 of 6

Rotational Speed 3, Non-Contact (0 to 2000) rpm 2.4 rpm Tachometer Verification of Scales and Balances 3 Class 1 weights Resolution (R): R = 0.0005 g R = 0.0005 g Up to 1 g Up to 1 g Up to 1 g Up to 200 g Up to 200 g Up to 200 g Up to 1000 g Up to 1000 g Up to 4000 g Up to 4000 g 58 mg 5.8 mg 0.3 mg 58 mg 5.9 mg 0.89 mg 58 mg 6.7 mg 62 mg 23 mg IV. Plastic Industry Specific Measurements Extrusion Plastometer Melt Flow Index 3 Temperature (0 to 400) ºC 0.35 ºC ASTM D1238 Bore Shaft-Inside Diameter Up to 0.4 in 300 μin Piston Foot Diameter Up to 1 in 300 μin Mass (0 to 5) kg 0.07 % (A2LA Cert. No. 1504.01) Revised 09/28/2017 Page 3 of 6

V. Rubber Industry-Specific Measurements Parameter/Equipment Range CMC 2, 4 (±) Comments Extensometer 3 Gage Length Class D 0.007 in ASTM E83 w/ calibrated scale Rubber Hardness Block Standardization Shore A Shore C Shore D Shore M IRHD M 0.71 A units 0.71 C units 0.71 D units 0.71 M units 0.61 units Direct Verification of Durometers ASTM D2240 Indentor Extension and Shape Diameter Radius (0 to 25) mm (0 to 25) mm 100 μin 100 μin Optical inspection under magnification or video comparator Angle (0 to 90) 0.04 Extension (0 to 25) mm 100 μin Indentor Display 0.58R + 12 μduro Gage blocks Spring Calibration Force Types Shore A Type Shore C Types Shore D Types JIS A Types IRHD Types Shore M Types Shore 0,00 0.6 Shore A units 0.61 Shore C Units 0.61 Shore D units 0.58 JIS A units 0.61 IRHD units 0.85 Shore M units 0.59 Shore 0,00 units (A2LA Cert. No. 1504.01) Revised 09/28/2017 Page 4 of 6

Rheometer 3 Torque @ (0.5, 1, 3) degrees arc (0 to 200) in lbf 0.14 in lbf Torque standard ASTM D2084 ASTM D5289 Mooney Viscometer 3 (0 to 200) Mooney 0.12 Mooney ASTM D1646 VI. Thermodynamics Temperature Measure 3 (-200 to 350) ºC 0.35 ºC RTD/Thermocouple; calibration of various temperature devices such as oil baths and ovens Relative Humidity (10 to 75) % RH 1.3 % RH Vaisala HMI41 & Measure 3 HMP46 VII. Time and Frequency Time Interval 3 (0 to 100) hr 3 s/day Digital stop watch 1 This laboratory offers commercial calibration service and field calibration service. 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer s device and to influences from the circumstances of the specific calibration. (A2LA Cert. No. 1504.01) Revised 09/28/2017 Page 5 of 6

3 Field calibration service is available for this calibration and this laboratory meets A2LA R104 General Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer s site being larger than the CMC. 4 In the statement of CMC, R is defined as the resolution of the device measured in Duros. (A2LA Cert. No. 1504.01) Revised 09/28/2017 Page 6 of 6

Accredited Laboratory A2LA has accredited POLYMER TESTING INSTRUMENTS, INC. Fletcher, NC for technical competence in the field of Calibration This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005 General requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of ANSI/NCSL Z540-1-1994 and R205 Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated 8 January 2009). Presented this 16 th day of May 2017. President and CEO For the Accreditation Council Certificate Number 1504.01 Valid to June 30, 2019 Revised on August 3, 2017 For the calibrations to which this accreditation applies, please refer to the laboratory s Calibration Scope of Accreditation.