Model 791 Calibration Plate for Tiltmeters and Inclinometers
|
|
- Angelina Parsons
- 5 years ago
- Views:
Transcription
1 User s Manual Model 791 Calibration Plate for Tiltmeters and Inclinometers Serial No. Baseline, L Date measured By mm Introduction The Model 791 Calibration Plate is designed for calibrating any tiltmeter or inclinometer (clinometer) that has a gravity-referenced output. The top surface of the plate contains threaded holes for attaching most Applied Geomechanics tiltmeters and clinometers. These holes also can be used to hold attachment fixtures for tiltmeters from other suppliers. The bottom surface of the plate contains three spherical tooling balls made of hardened steel. These balls, which are separated by a fixed distance L, comprise the legs of the calibration plate. To perform a calibration, NIST-traceable gage blocks are placed under one end of the plate to rotate the tiltmeter through a series of known angles. The output of the tiltmeter at each angle establishes the calibration. Figure 1 shows the calibration plate in use with a Model 701 Biaxial Platform Tiltmeter. Specifications are listed in Table 1. Figure 1. Calibrating a tiltmeter with the Model 791 Calibration Plate
2 page 2 of 6 Table 1. Model 791 Calibration Plate Specifications and Ordering Information MATERIALS BASELINE, L GAGE BLOCK SIZES SIZE & WEIGHT ORDERING INFORMATION Calibration Plate: 6061T6 aluminum Contact Points: 3 hardened tool steel spherical balls, diameter = inch ± inch ( ± mm) Gage Blocks: Hardened tool steel ± 0.03 mm Part No : 10 blocks ranging from 1mm to 25mm for angles between 0.2 and 6 degrees. Other sizes available on request. Calibration Plate: 13 x 6 x 1 in (330 x 152 x 25 mm), 3.6 lb (1.6 kg) Part No Gage Block Set: 2 lb (0.86 kg) Part No : Model 791 Calibration Plate Part No : Rectangular Gage Block Set Part No : Vertical Bracket Kit for Wall Mount Tiltmeters Part No : Tilt Plate Kit for Model 800P Portable Tiltmeter Using Your Model 791 Calibration Plate The Model 791 Calibration Plate is easy to use. Simply follow the steps below: 1. Locate the hole pattern that fits your Applied Geomechanics tiltmeter. Screw the tiltmeter to the calibration plate using the fasteners that were supplied with it or with your own fasteners. The following American thread sizes are used in the holes in the calibration plate: 4-40, 6-32, 8-32, 10-32, 1/4-20 and 5/ Place the calibration plate on a hard, flat and level surface. A granite table or an optical table are excellent choices. 3. If you plan to calibrate the tiltmeter over its full range (on both the positive and negative sides of level), elevate the end of the calibration plate with the two legs. Both legs should be elevated by the same amount. Gage blocks can be used for this purpose. 4. Connect your tiltmeter to a power supply and a readout device. Wait for the output to stabilize and then take the first reading. Enter the reading into a calibration table beside the height of the gage block. Table 2 is an example of a table in which the gage block heights and corresponding angles already have been entered. Table 3 is a blank table that you can copy and use. 5. Place the first gage block under the end of the calibration plate with the single tooling ball (place the block under the tooling ball)(figure 2). Record the output in your calibration table beside the height of the gage block. 6. Repeat step 5 until you have recorded the tiltmeter output for several steps over the desired calibration range. The number of steps that you should record is a function of your unique calibration requirements. 7. Compute the tilt angle θ at each step using the formula: θ = arc sin (h/l) where h is the height of the gage block and L is the baseline length (Figure 2), written on the first page of this document. Enter these values into your calibration table (Table 2 or 3). After completing the calibration procedure described above, your results can be used in a variety of ways. For example: The angular rotation between two gage block steps can be compared to the value indicated by the tiltmeter.
3 page 3 of 6 A linear regression line that best fits all of the calibration data can be computed (Figure 3) and the slope of this line taken as the scale factor of the tiltmeter. This is the method that Applied Geomechanics uses to obtain the scale factors of analog output tiltmeters in its factory calibrations. A polynomial approximation of the calibration data can be obtained using a spreadsheet application such as Excel. Maintaining Your Model 791 Calibration Plate Your Model 791 Calibration Plate and gage blocks are uncomplicated devices that will provide many years of trouble-free service if the following maintenance procedures are followed: Keep this equipment clean and dry. Prolonged exposure to moisture or to corrosive liquids will mar the finish and can damage the threads in the tiltmeter attachment holes. Keep the tooling balls (legs) coated with a thin coating of low-viscosity oil at all times to prevent rust. Avoid impacting the tooling balls against hard objects, as this may dent or flatten their surfaces, reducing the accuracy of the calibration plate.
4 page 4 of 6 Figure 2
5 page 5 of 6 Table 2 Calibration Table Model 791 Tiltmeter Calibration Plate. Baseline Length, L = 300mm Tilt Angle Gage block height, mm degrees arc minutes arc seconds microradians tiltmeter reading , , , , , , , , , , , , , , ,448 6 Output (volts) Degrees=1.021 x (Volts) Angle (degrees) Figure 3. Calibration data for a tiltmeter with analog voltage output. Scale factor = 1.021degrees/volt.
6 page 6 of 6 Table 3 Calibration Table Model 791 Tiltmeter Calibration Plate. Baseline Length, L = mm Tilt Angle Gage block height, mm degrees arc minutes arc seconds microradians tiltmeter reading Should you have questions about tiltmeter calibration, please contact us at the address below.
User s Manual: MODEL 904-T CLINOMETER PAK
User s Manual: MODEL 904-T CLINOMETER PAK Model 904-T Serial No. Bipolar Output (±2.5 volts typical): Unipolar Output (0-5 volts typical): Model 870 Readout Module Serial No. (optional) 2007 Applied Geomechanics
More informationElectrolytic Tiltmeter Temperature Coefficients: Source, Definition and Use to Improve Accuracy
Electrolytic Tiltmeter Temperature Coefficients: Source, Definition and Use to Improve Accuracy Introduction Environmental temperature changes alter the mechanical and electrical characteristics of all
More informationPositive Signal The most technologically advanced. double coated dichroic filters are used, which transmit
The PAPI (Precision Approach Path Indicator) system is a simple and efficient, positive visual aid to the pilot on final approach. The PAPI system consists of a line of four units at 90 to the runway center
More informationMorehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:
Morehouse 1 Morehouse Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com sales: edlane@mhforce.com 2 This presentation will cover the calibration
More informationThermal Cube. Custom-built Heat Flux Sensor
Thermal Cube Custom-built Heat Flux Sensor Specifications and Construction MR. RAZIM REFAI MR. SHAMMAWI ANDERSON DR. ANDRÉ MCDONALD Table of Contents 1. Introduction... 2 1.1 Specifications [1]... 2 1.2
More informationLUDLUM MODEL 42-5 NEUTRON BALL CART. August 2015 Serial Number PR and Succeeding Serial Numbers
LUDLUM MODEL 42-5 NEUTRON BALL CART August 2015 Serial Number PR149512 and Succeeding Serial Numbers LUDLUM MODEL 42-5 NEUTRON BALL CART August 2015 Serial Number PR149512 and Succeeding Serial Numbers
More informationCEEN 3320 Behavior & Properties of Engineering Materials Laboratory Experiment No. 1 Measurement Techniques
Laboratory Experiment No. 1 Measurement Techniques Engineers rely on data from a wide variety of sources to design the things that make up our physical world and to ensure compliance with established specifications.
More informationLP PYRA Installation and Mounting of the Pyranometer for the Measurement of Global Radiation:
CENTRO DI TARATURA SIT N 124 TEL. +39.049.8977150 r.a. FAX +39.049.635596 1 Introduction LP PYRA 03 The LP PYRA 03 pyranometer measures the irradiance on a plane surface (Watt/ m 2 ). Measured irradiance
More informationTHE LOSMANDY G-11 MOUNT
Checking the parts THE LOSMANDY G-11 MOUNT Depending on which accessories you ordered, your G-11 mount was shipped in four or more boxes. The contents of each box are as follows: Equatorial Mount Adjustable
More informationDam Monitoring With Tiltmeters
Dam Monitoring With Tiltmeters INTRODUCTION What is a Tiltmeter? Applied Geomechanics tiltmeters are simple yet precise instruments that work on the concept of a trapped air bubble inside a glass or ceramic
More informationPLOW MOUNT KIT FOR POLARIS RANGER P/N ASSEMBLY / OWNERS MANUAL
PLOW MOUNT KIT FOR POLARIS RANGER P/N 34-3010 ASSEMBLY / OWNERS MANUAL Application PLOW PUSH FRAME NO. 34-0000 or 34-0070 Before you begin, please read these instructions and check to be sure all parts
More informationLaboratory 1: Calibration of a High-Volume Sampler
Laboratory 1 Laboratory 1: Calibration of a High-Volume Sampler 1.1 To the User of this Manual This manual has been prepared to guide the students through a series of laboratory exercises developed to
More information22X Builder s Transit Level Model No Instruction Manual
2595H 7/29/09 10:15 AM Page 1 22X Builder s Transit Level Model No. 40-6910 Instruction Manual Congratulations on your choice of this 22X Builder s Transit Level. We suggest you read this instruction manual
More informationproduct manual H-4204 Proving Ring Penetrometer
05.09 product manual H-4204 Proving Ring Penetrometer nnmm Handle Proving Ring Extension Rod Coupler Penetration Rod Cone Figure #1 General Information The Proving Ring Penetrometer is a cone type of penetrometer
More information22X Builder s Level Model No Instruction Manual
2594H 7/29/09 10:12 AM Page 1 22X Builder s Level Model No. 40-6900 Instruction Manual Congratulations on your choice of this 22X Builder s Level. We suggest you read this instruction manual thoroughly
More informationUncertainty Analysis of Experimental Data and Dimensional Measurements
Uncertainty Analysis of Experimental Data and Dimensional Measurements Introduction The primary objective of this experiment is to introduce analysis of measurement uncertainty and experimental error.
More informationSENSORS CLIMATRONICS CORPORATION (631) MANUAL TACMET II WEATHER STATION, EMI 2.0 INSTALLATION 1.0 INTRODUCTION
SENSORS MANUAL TACMET II WEATHER STATION, EMI P/N M102304 Rev H 1.0 INTRODUCTION Climatronics TACMET II weather station is designed as a stand-alone weather station to provide accurate measurements of
More informationExtending Transducer Calibration Range by Extrapolation
Introduction Extending Transducer Calibration by olation by LaVar Clegg Force and torque transducers must be calibrated in a laboratory in order to be useful in their intended application. Applications
More informationINSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7
INSTRUCTION MANUAL Model 902A Laser Diode Sarcomere Length May 6, 2013, Revision 3 Copyright 2008-2013 Aurora Scientific Inc. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario,
More informationproduct manual H-3220A Benkelman Beam
05.12 product manual H-3220A Benkelman Beam General The H-3220A Benkelman Beam Apparatus is a convenient and accurate device used for measuring the deflection of flexible pavements under moving wheel
More informationCommercial Generator Parts Catalog. PROTEC Series
Commercial Generator Parts Catalog PROTEC Series YD ÁÁ ÁÁ To avoid errors or delay in filling your parts order, always give the MODEL, SPEC NO. and SERIAL NO. from the Onan nameplate. For handy reference,
More informationBINOCULAR STEREOSCOPIC MICROSCOPE MBS-10. User Manual
BINOCULAR STEREOSCOPIC MICROSCOPE MBS-10 User Manual ATTENTION! Мicroscope design may differ somewhat from description in this manual. 2 1 - PURPOSE The MBS-10 microscope is intended for observing both
More informationAltazimuth Mount. Altazimuth Mount
instruction Manual Orion VersaGo II HD Altazimuth Mount #10104 Orion VersaGo II Altazimuth Mount #10105 #10104 #10105 Providing Exceptional Consumer Optical Products Since 1975 OrionTelescopes.com Customer
More informationJumbo Cross Beam Ionizer
Jumbo Cross Beam Ionizer Version 1.0 September 30, 2008 Page 1 of 19 Table of Contents Cross Beam Ionizer... 1 Table of Contents... 2 1.0 Packing List... 3 1.1 Packing List for Cross Beam Ionizer... 3
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. GREENSLADE & CO., INC Wenneca Avenue Fort Worth, TX Larry Borowski Phone:
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 GREENSLADE & CO., INC. 2234 Wenneca Avenue Fort Worth, TX 76102 Larry Borowski Phone: 800 435 2657 CALIBRATION Valid To: October 31, 2018 Certificate Number:
More informationOrder No. G Date: September 8, 2010 REPORT NO CRT-001 TEST OF SMF/PAPI SYSTEM RENDERED TO
REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Order No. G100147381 Date: September 8, 2010 INTRODUCTION REPORT NO. 100147381CRT-001 TEST OF SMF/PAPI SYSTEM RENDERED TO ARGOS INGEGNERIA SpA VIA TIBURTINA,
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 TRI-STATE INSTRUMENT SERVICE, INC. 3024 Wayne Trace Ft. Wayne, IN 46806 Lynn A. Stroble Phone: 260 456 4545 CALIBRATION Valid To: September
More informationThe FLUX F 430 with mechanical seal
The FLUX F 0 with mechanical seal Flexible coupling between motor and pump The pump must be used with many varying liquids, consequently the smallest residual left within the pump is unacceptable. You
More informationMANUAL TT-220 TT-220
MANUAL TT-220 TT-220 INDEX 1. GENERAL 2 1.1 Scope of applications 2 1.2 Basic working principle 2 1.3 Basic configuration TT220 2 1.4 Technical Parameters 3 1.5 Main functions 3 2. OPERATION OF GAUGE 4
More informationIn Association With. Introduces. TPT Contractor
In Association With Introduces TPT Contractor Software for Contractors with TENSIONMETERS Table of Contents Capabilities 4 Setup and Use 7 The Dashboard 12 Tower Setup Form 18 Field Calibration Verification
More informationRoute and retain the clinometer cable so that it does not interfere with moving parts where the unit is mounted.
SYSTEM DESCRIPTION The AngleStar Protractor Systems are economical angle measurement devices. A system consists of a compact gravity-referenced clinometer connected to an associated liquid crystal, (LCD)
More informationMotion in Two Dimensions: Centripetal Acceleration
Motion in Two Dimensions: Centripetal Acceleration Name: Group Members: Date: TA s Name: Apparatus: Rotating platform, long string, liquid accelerometer, meter stick, masking tape, stopwatch Objectives:
More informationNuclear Medicine Physics Course Laboratory Number 3: Dose Calibrator Quality Control
Nuclear Medicine Physics Course Laboratory Number 3: Dose Calibrator Quality Control Purpose 1. To understand quality control tests for accuracy, linearity, and daily constancy that are necessary in the
More informationPC1141 Physics I Compound Pendulum
PC1141 Physics I Compound Pendulum 1 Purpose Determination of the acceleration due to gravity 2 Equipment Kater pendulum Photogate timer Vernier caliper Meter stick 3 Theory One of the most important physical
More informationрн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL
рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically Complete Set ph PRO Controller - 1 pcs. ph electrode - 1 pcs. Calibration
More informationOperating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2
Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2 P/N VALTOL-KT2 Revision 2.0 October 7, 2014 Prepared by: Quality Lab Accessories, LLC 100 Emlen Way, Suite 108 Telford,
More informationFAST CHECK THICKNESS GAGE
FAST CHECK THICKNESS GAGE Recommended for quick thickness checks of flat sheet materials and wide variety of small parts. Supplied with hardened steel gaging contacts, lifting levers, and clockwise continuous
More informationRefer to page E-6 for details. Refer to page E-31 for details. Digital Height Master. Refer to page E-35 for details. High Precision Square
NewProducts Ultra Low xpansion Ceramic Gauge (ZRO CRA Block) Refer to page -6 for details. Gauge Block Comparator GBCD-00A Refer to page -3 for details. Digital Height Master Refer to page -35 for details.
More informationPARTS MANUAL 2 INCH CHIPPER/SHREDDER. SC cc SUBARU
www.bearcatproducts.com 2 INCH CHIPPER/SHREDDER SC2170 170cc SUBARU PARTS MANUAL PN: 1641-00 Rev. 0011 Companion to 1509-00 Range: D01688 - Current Before You Begin DEAR ECHO BEAR CAT CUSTOMER Thank you
More informationconfront fury
confront WINTER S fury WITH A COTECH SNOW PLOW www.cotech.ca table of content mission...2 Snow Removal Equipment...4 Snow plows with a trip system... 4 Extendable Reversible Snow Plows... 6 Inverted Snow
More informationPHYS 1111L - Introductory Physics Laboratory I
PHYS 1111L - Introductory Physics Laboratory I Laboratory Advanced Sheet Acceleration Due to Gravity 1. Objectives. The objectives of this laboratory are a. To measure the local value of the acceleration
More informationFlexitest Switch Assemblies Type FT-19R and FT-19RX Supersedes DB March 2008
Flexitest Switch Assemblies Type FT-19R and FT-19RX Supersedes DB41-078 March 2008 FT-19R and FT-19RX FT-19R 3RU with Individual Clear Covers APPLICATION FEATURES The type FT-19R (standard length) and
More informationUsing Calibrated Specular Reflectance Standards for Absolute and Relative Reflectance Measurements
Using Calibrated Specular Reflectance Standards for Absolute and Relative Reflectance Measurements Applications Overview here are two fundamental techniques for measuring specular reflectance with a UV/VIS/NIR
More informationUser Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration
User Manual Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio with NIST-Traceable Calibration Model 20250-09 THE STANDARD IN PRECISION MEASUREMENT Introduction The Digi-Sense
More informationUpgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory
Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory Bülent ÖZGÜR 1,*, Okhan GANİOĞLU 1, Nasser Al-Qahtani 2, Faisal Al-Qahtani 2 1 TÜBİTAK, National
More informationsaxon Instruction Manual saxon Grandeur Brass Telescope High quality optics
saxon High quality optics Instruction Manual saxon Grandeur Brass Telescope WARNING! Do not use the telescope to look at the sun without an appropriate solar filter. Looking at or near the sun can result
More informationSECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:
SECTION 1 - WHAT IS A BTU METER? ISTEC BTU Meters measure energy usage by multiplying flow rate and temperature difference. As the water (or other liquid) passes through these lines, the multi-wing turbine
More informationBall screw. Linear. Coupling cover. 1Positioning table of a structure with enhanced sealing property. 3 High corrosion resistance
Cleanroom Precision Positioning Table TC Ball screw TC EB Slide table inear Way Side cover Stainless sheet Motor bracket inear Coupling cover Pipe threads for suction connector Points ight weight, low
More informationAmerican Revolution Deluxe Full Cutoff LED Series ARDCL
P R O D U C T O V E R V I E W SERIES ARDCL Applications: Streetscapes Walkways Pathways Parks Features: Colonial LED lantern, replaces HID models up to 150W HPS for street and area lighting applications
More informationBall Reverser. Ball Reverser Actuator. FLENNOR Power Transmission Products. Patented
Ball Reverser FLENNOR Power Transmission Products Ball Reverser Actuator Patented Ball Reverser Principle General Information NORCO'S Ball Reverser Actuator is a unique actuator which provides automatic
More informationValiant Full Cutoff LED Series AVPCL2
PRODUCT OVERVIEW SERIES AVPCL2 Applications: Streetscapes Walkways Pathways Parks Features: Colonial LED lantern, replaces HID models up to 150W HPS for street and area lighting applications Eleven (11)
More informationLUDLUM MODEL NEUTRON DETECTOR. December 2017
LUDLUM MODEL 42-49 NEUTRON DETECTOR December 2017 LUDLUM MODEL 42-49 NEUTRON DETECTOR December 2017 Model 42-49 Neutron Detector Technical Manual Table of Contents Introduction 1 Unpacking and Repacking
More informationPHYS 2211L - Principles of Physics Laboratory I
PHYS 2211L - Principles of Physics Laboratory I Laboratory Advanced Sheet Acceleration Due to Gravity 1. Objectives. The objectives of this laboratory are a. To measure the local value of the acceleration
More informationProduct Data. CNPHP, CNRHP Evaporator Coil N Coil --- Cased Horizontal CNPHP / CNRHP
CNPHP, CNRHP Evaporator Coil N Coil --- Cased Horizontal Product Data 06005 This horizontal design N -coil is an evaporator coil designed to provide the highest standards of reliability and durability.
More informationD501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect
Microscanner D-Series The Only Certified Accurate Surface Instruments in the World A MUST FOR ISO 9001 ISO 9002 ISO 9003 TRACEABILITY PROGRAMS Common Surface Measurement Errors D501 and D501-RS Microscanner
More informationPoints. Precision Positioning Table TU. Variation. Major product specifications With flange
9 0 Precision Positioning Table Ball screw inear Points lide table with high accuracy and high rigidity in a single structure lide table Compact and slim type positioning table with an original U-shaped
More informationKalinka Optics Warehouse User Manual Kobra AK Side Mount Red Dot Sight Manual
Kalinka Optics Warehouse User Manual www.kalinkaoptics.com Kobra AK Side Mount Red Dot Sight Manual CONTENTS 1. Introduction... 4 2. Purpose... 4 3. Specifications... 4 4. Components and Equipment Provided...
More informationCanal Velocity Indexing at Colorado River Indian Tribes (CRIT) Irrigation Project in Parker, Arizona using the SonTek Argonaut SL
Canal Velocity Indexing at Colorado River Indian Tribes (CRIT) Irrigation Project in Parker, Arizona using the SonTek Argonaut SL Authors: Dr. Stuart Styles P.E., Mark Niblack, Beau Freeman Abstract An
More informationATV WP2 FRONT MOUNT PUSH TUBE
ATV WP2 FRONT MOUNT PUSH TUBE ASSEMBLY / OWNER S MANUAL Part No: 16-0000 MAN0352 Rev. 00 OPERATING INSTRUCTIONS Congratulations! You ve just purchased the most durable plow component in the industry. Cycle
More informationTEMPERATURE SENSOR. Model TH-T. Roctest Limited, All rights reserved.
INSTRUCTION MANUAL TEMPERATURE SENSOR Model TH-T Roctest Limited, 2005. All rights reserved. This product should be installed and operated only by qualified personnel. Its misuse is potentially dangerous.
More informationTEST METHOD FOR DETERMINING BREAKAWAY FORCE OF A MAGNET
STANDARD MDFA 101 95 TEST METHOD FOR DETERMINING BREAKAWAY FORCE OF A MAGNET DISCLAIMER: This test method was developed by representative members of the Magnet Distributor and Fabricators Association (MDFA)
More informationSUGAR/BRIX REFRACTOMETER INSTRUCTION MANUAL
SUGAR/BRIX REFRACTOMETER 300003 INSTRUCTION MANUAL 1 TABLE OF CONTENTS INTRODUCTION...3 PANEL DESCRIPTION...3 OPERATING PROCEDURES...4-5 TEMPERATURE CORRECTION...5 TEMPERATURE CORRECTION TABLE...6 PRECAUTIONS...7
More informationMET ONE RAIN GAGE MODELS 380, 385, 380M, 385M
MODELS 380, 385, 380M, 385M REVISION: 2/96 COPYRIGHT (c) 1993-1996 CAMPBELL SCIENTIFIC, INC. WARRANTY AND ASSISTANCE The MET ONE RAIN GAGE--MODELS 380, 385, 380M, 385M is warranted by CAMPBELL SCIENTIFIC,
More informationLever-type Dial Test Indicators
Lever-type Dial Test Indicators G-1 Lever-type Dial Test Indicators TESA IP65 Electronic Dial Test Indicators Provides the advantages of a mechanical test indicator with a digital reading. Inductive patented
More informationRotational Motion. Figure 1: Torsional harmonic oscillator. The locations of the rotor and fiber are indicated.
Rotational Motion 1 Purpose The main purpose of this laboratory is to familiarize you with the use of the Torsional Harmonic Oscillator (THO) that will be the subject of the final lab of the course on
More informationPUSH TUBE STANDARD DUTY
PUSH TUBE STANDARD DUTY ASSEMBLY / OWNER S MANUAL Part No: 15-0070 PUR223 Rev. 00 OPERATING INSTRUCTIONS Congratulations! You ve just purchased the most durable plow component in the industry. Cycle Country
More informationFOR POLARIS ATV s P/N ASSEMBLY / OWNERS MANUAL
MID-MOUNT KIT FOR POLARIS ATV s P/N 15-3591 ASSEMBLY / OWNERS MANUAL Application PLOW PUSH FRAME NO. 15-0070, 33-0000 or 33-0070 Before you begin, please read these instructions and check to be sure all
More informationElcometer 116C. Sling Hygrometer. Operating Instructions
Elcometer 116C English Sling Hygrometer Operating Instructions English is a registered trademark of Elcometer Limited. All other trademarks acknowledged. Copyright Elcometer Limited. 2008. All rights reserved.
More informationNAIL SECTION TABLE OF CONTENTS
NAIL SECTION TABLE OF CONTENTS ANCHOR TRUSS BOX NAILS-1 NAILS-1 COMMMON NAILS-1 & 2 DECK NAILS NAILS-2 & 3 DRYWALL DUPLEX FINISH FLOOR - HARD SCREW NAILS-4 NAILS-4 NAILS-5 NAILS-5 MASONRY - STUB NAILS-6
More informationModel 556X User s Manual. Optical Isolator
Model 556X User s Manual Optical Isolator 550031 Rev. A 2 Is a registered trademark of New Focus Inc. Warranty New Focus, Inc. guarantees its products to be free of defects for one year from the date of
More informationElizabethtown College Department of Physics and Engineering PHY104
Elizabethtown College Department of Physics and Engineering PHY104 Lab #7- Capacitors 1. Introduction The capacitor is one of the essential elements of analog circuitry. It is highly useful for its energy
More informationC-Lube Linear Way ML Linear Way L ML LWL
-Lube Linear ay ML Linear ay L MLLL -Lube Maintenance ree Series -Lube Linear ay ML ML long term maintenance free supported! The aquamarine end plate is the symbol of maintenance free. Identification umber
More informationLM35 Precision Centigrade Temperature Sensors
LM35 Precision Centigrade Temperature Sensors General Description The LM35 series are precision integrated-circuit temperature sensors, whose output voltage is linearly proportional to the Celsius (Centigrade)
More informationExperiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado
Experiment 6 1 Introduction The Compton Effect Physics 2150 Experiment No. 6 University of Colorado In some situations, electromagnetic waves can act like particles, carrying energy and momentum, which
More informationPLOW MOUNT KIT FOR YAMAHA VIKING P/N ASSEMBLY / OWNERS MANUAL. Application PLOW PUSH FRAME NO or
PLOW MOUNT KIT FOR YAMAHA VIKING P/N 34-5020 ASSEMBLY / OWNERS MANUAL Application PLOW PUSH FRAME NO. 34-0000 or 34-0070 Before you begin, please read these instructions and check to be sure all parts
More informationMODEL NAME : LLDMWW0-15K*0*A
Eagle Eye (Outdoor LED Module) Application Note MODEL NAME : LLDMWW0-15K*0*A RoHS Compliant HALOGEN FREE (30)-4022 1 / 16 CONTENTS 1. Product Description 3/16 2. Specifications 2.1. Chromaticity Bins for
More informationElectric Field Mapping
Electric Field Mapping Equipment: mapping board, U-probe, 5 resistive boards, templates, 4 long leads, Phywe 07035.00 voltmeter, DC wall voltage output, 3 pieces of paper Precautions 1. Before turning
More informationDetection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR
Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR Ensuring ultimate cleanliness for maximum adhesion Application Note Author John Seelenbinder Agilent Technologies,
More informationLEAPERS, INC. RED/GREEN DOTS
LEAPERS, INC. 1 RED/GREEN DOTS range estimating scopes UTG reticle intensified scopes TOTAL SOLUTION TO YOUR NEEDS -COMMITMENT TO BEST QUALITY, BEST VALUE AND BEST SERVICEwww.LEAPERS.com 32700 Capitol
More informationVariable Speed 20" Single and Multi-spindle, Belt Drive Industrial Drill Presses
www.clausing-industrial.com Variable Speed 20" Single and Multi-spindle, Belt Drive Industrial Drill Presses Two Speed Variable Speeds 150-2000rpm Variable Speeds 200-1300rpm Variable Speeds 300-2000rpm
More informationMN-ISO-2000FN-E. ISO-2000FN COATING THICKNESS GAGE OPERATION MANUAL
MN-ISO-2000FN-E www.insize.com ISO-2000FN COATING THICKNESS GAGE OPERATION MANUAL V0 Place the probe on the zero plate and wait until beep. Lift the probe for minimal 5 cm off the zero plate and wait
More informationInstruction Manual. Kite-22 Camera Crane Starter Package (KITE-22-STARTER) Exclusive of Weights
Instruction Manual Kite-22 Camera Crane Starter Package (KITE-22-STARTER) Exclusive of Weights All rights reserved No part of this document may be reproduced, stored in a retrieval system, or transmitted
More informationwww. ElectricalPartManuals. com Flexitest Switch Assemblies Type FT-19R and FT-19RX INTL APPLICATION
Effective: January 2004 Supersedes, dated June 1998 APPLICATION QUALITY FEATURES Descriptive Bulletin 41-078INTL Flexitest Switch Assemblies Type The type FT-19R (standard length) and FT-19RX (extended
More informationRestoration of Eureka Clock Serial Number 4793
Page 1 Restored Eureka 4793 Restoration of Eureka Clock Serial Number 4793 Page 2 Restored Eureka 4793 Photos by kind permission of the owner. A short movement with ceramic skeletonised dial. The dial
More informationMOY/SCMI/36 SPECIFICATION OF ACCURACY FOR A PRECISION CLINOMETER
Centre for Basic, Thermal and Length Metrology National Physical Laboratory MOY/SCMI/36 SPECIFICATION OF ACCURACY FOR A PRECISION CLINOMETER A Watts Precision Clinometer fitted with a circular glass scale
More informationEngineering Metrology
Albaha University Faculty of Engineering Mechanical Engineering Department Engineering Metrology Lecture 04: Angular Measurements Ossama Abouelatta o_abouelatta@yahoo.com Mechanical Engineering Department
More informationBallistic pendulum Operating Instructions Fig. 1: Ballistic pendulum SAFETY PRECAUTIONS
R Ballistic pendulum 11229.00 PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 D-37079 Göttingen 6 3.8 3.7 3.6 3.5 Phone +49 (0) 551 604-0 Fax +49 (0) 551 604-107 E-mail info@phywe.de Internet www.phywe.de
More informationIRTS-P PRECISION INFRARED TEMPERATURE SENSOR INSTRUCTION MANUAL
IRTS-P PRECISION INFRARED TEMPERATURE SENSOR INSTRUCTION MANUAL 4/02 COPYRIGHT (c) 2002 CAMPBELL SCIENTIFIC, INC., APOGEE INSTRUMENTS, INC. This is a blank page. Warranty and Assistance The IRTS-P PRECISION
More informationDielectric Analysis of Solid Insulations
using Dielectric Test Fixture 16451B from Keysight By Britta Pfeiffer 2018 by OMICRON Lab V2.0 Visit www.omicron-lab.com for more information. Contact support@omicron-lab.com for technical support. Page
More informationC-Lube Linear Way ML ML MLF
Maintenance ree & U.S. PTTD C-Lube Linear ay Variation of C-Lube Linear ay Shape Length of slide unit Model code C-Lube Linear ay is a linear motion rolling guide, incorporating the C-Lube as a components
More informationFSUE PA Novosibirsk. PO3-9x24, PO3-9x24-1. Instruction manual
FSUE PA Novosibirsk Instrument-making Plant Optical Sights PO3-9x24, PO3-9x24-1 Instruction manual AL3.812.232 RE Because of continuous modernization of Sights, their construction may have some corrections,
More informationElectric Field Mapping
Electric Field Mapping Equipment: mapping board, U-probe, 5 resistive boards, templates, knob adjustable DC voltmeter, 4 long leads, 16 V DC for wall strip, 8 1/2 X 11 sheets of paper Reading: Topics of
More informationiweathar EDGE V Installation Guide Copyright 2012 iweathar. All rights reserved.
iweathar EDGE V Installation Guide Copyright 2012 iweathar. All rights reserved. Table of contents: 1 Introduction...3 2 Station Components...4 3 STATION INSTALLATION PROCEDURE...5 4 Specifications...11
More informationInfrared Thermometer Calibration A Complete Guide
Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate
More informationThe Coupled Pendulum Experiment
The Coupled Pendulum Experiment In this lab you will briefly study the motion of a simple pendulum, after which you will couple two pendulums and study the properties of this system. 1. Introduction to
More information405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook
Reference Manual 405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Book and Flow Handbook www.rosemount.com Reference Manual 405 and 1595 405 Compact Orifice Series and 1595 Conditioning
More informationProduct Data. CNPVP, CNRVP, CNPVT, CNRVT, CNPVU, CNRVU Evaporator Coil N Coil Cased and Uncased Upflow, Downflow
CNPVP, CNRVP, CNPVT, CNRVT, CNPVU, CNRVU Evaporator Coil N Coil Cased and Uncased Upflow, Downflow Product Data CN(P,R)VP CN(P,R)VT (Transition) A06008 A06007 This vertical design N -coil is a furnace
More informationTemperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer
Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer Temperature Control without Compromise: rapid changes, precise temperatures, powerful complex functionality.
More information1 M62 M62.1 CONSERVATION OF ANGULAR MOMENTUM FOR AN INELASTIC COLLISION
1 M62 M62.1 CONSERVATION OF ANGULAR MOMENTUM FOR AN INELASTIC COLLISION PRELAB: Before coming to the lab, you must write the Object and Theory sections of your lab report and include the Data Tables. You
More informationCM3 PYRANOMETER INSTRUCTION MANUAL
CM3 PYRANOMETER INSTRUCTION MANUAL REVISION: 10/02 COPYRIGHT (c) 1999-2002 CAMPBELL SCIENTIFIC, INC. This is a blank page. WARRANTY AND ASSISTANCE This equipment is warranted by CAMPBELL SCIENTIFIC (CANADA)
More information