Lasentec Product Group
|
|
- George Tyler
- 5 years ago
- Views:
Transcription
1 Prepared by: Gregor Hsiao Approved on: ECN# 3506 Effective Date: 1/14/2008 Title: Window Reference Procedure for Version and Higher Purpose: Lasentec D600 Probes and FBRM/Raman Probes using version 6.0 and higher FBRM CI Software The focal point position of the FBRM probe must be established to ensure that the measurements are consistent and repeatable for each instrument. The window reference procedures must be performed to calculate the Lasentec instrument focal point position. The measured statistic, #/Sec, No Wt 0-5 (Statistic #5) reaches a maximum value, also called the peak value, when the focal point is located at the window reference position. The procedure described below is the standard calibration procedure used to reference the Lasentec instrument focal point to the window surface. It is for both a D600 Lasentec probe and the Lasentec FBRM/Raman probe that is being used with version 6.0 and higher and higher of the Lasentec Control Interface (CI) software. Use Instruction Sheet for calibration if version 5.4 of the CI software is used. The Window Reference Procedure should be performed: Before running the PVC Reference Procedure If the PVC Reference Procedure indicates the Lasentec FBRM instrument or the FBRM/Raman instrument is out of calibration For non-standard probe orientations (see Orienting the Probe below) Orienting the Probe At the factory, the probe is calibrated in a vertical position with the window pointing down, as it is in the fixed beaker stand. The vertical (window down) window reference position must established before the PVC Reference Procedure can be run. Non-Standard Orientation If the orientation of the process installation is greater than ±30 from a vertical (window down) position, this Window Reference Procedure should be performed with the Lasentec FBRM probe or FBRM/Raman probe in the same orientation as in the process installation (i.e., horizontal, vertical, or at a given angle). Adjusting/Reading the Micrometer that the probe is installed in a Fixed Beaker Stand with the window facing down. The micrometer is used to adjust the focal point position. The scale on the micrometer knob indicates the relative position of the focal point. Each marked increment on the micrometer equals 10 microns (0.01mm). Half-increment positions can be used to adjust the position to the nearest 5 microns (5 µm). The micrometer can make no more than one full turn. The zero reference mark engraved on the probe head (see Figure 1) indicates where to read the current micrometer position Rev K Page 1 of 9
2 On the Lasentec D600 L, P, S, T, or X probes and the FBRM/Raman probe, the micrometer is a knob protruding from the top (probe window facing down) of the probe head. On the D600R probes, the micrometer is a ring. D600L P, S, T, or X and FBRM/RAMAN PROBES D600R PROBE Figure 1 Micrometer Location The window reference procedure finds the zero point of the window. When the instrument is measuring samples or in a process the focus should be set to -20 µm relative to the zero mark. Turning the micrometer knob (ring) clockwise (top view) moves the focal point further into the slurry (positive distance). Turning the micrometer knob (ring) counterclockwise (top view) moves the focal point back into the probe (negative distance). Coating the Window Surface Using the Lasentec Marker Coat the Lasentec probe window with erasable ink before locating the window reference position. 1. Using the Lasentec Dry Erase Marker provided with the Lasentec FBRM hardware, coat the surface of the clean, dry Lasentec probe window with a uniform, thick layer of ink. While most black, dry-erasable markers will do, it is best to use the marker provided by Lasentec to ensure measurement-to-measurement conformity. 2. Allow two minutes for the ink to dry. Configuring version 6.0 and higher of the FBRM CI software to Locate the Window Reference Position Before locating the window reference position, verify that the Acquisition Module of version 6.0 and higher of the FBRM CI software is running and configured per the directions below. With the CI software, the micrometer position can be adjusted while the Lasentec instrument output is monitored. To configure the software for locating the window reference position: Rev K Page 2 of 9
3 1. From the Setup Menu, select Meas. Config. 2. For Measurement Duration, select 5 sec (or 10 sec if using a low-speed optical line driver). 3. For Time Averaging, select No Time Averaging. 4. For Channel Grouping, select µm 200 Linear Ch. 5. Click the Open File for Save button. 6. Open file C:\Fbrm6?\Backup\ SN-X-mmddyy-WRP 1 and click Save. 7. For When Saving to File, save every v Measurements, set it to save every 1 measurement. 8. Press the NOT measuring, Press to Start button. 9. From the Setup Menu, select Stat. Config. 10. Press the Load Stats. Config button. 11. From the C:\Fbrm6?\Statistics Setup Files folder, select the default0-100-lin.sta statistics setup (the measured statistic #/sec, No Wt 0-5 will reach a maximum value when the focal point is located at the window reference position) and click Open. 12. From the Display Menu, select Stats. 13. In the Show column, make sure there is an X next to the #/sec, No Wt 0-5 statistic (Statistic #5). This will create a trend plot in the trend display. 14. De-select all other statistics by clicking on any existing Xs in the Show column. De-selecting a statistic will remove the trend from the trend display. 15. Under X-axis Scale (located below the trend display), select Record. 16. Under Buffer Length, type in To autoscale the X- and Y-axes, push both toggle switches (under the Trend Display graph) to the right. 18. From the Setup Menu, select Annotation. 19. Make sure the box next to Save with All Records contains an X. 20. Click the Y-axis applies to: box (located below the trend display) and select the #5 statistic. 21. Press the Measuring: Press to Stop button. 1 SN = the instrument s serial number X = electronics type (F or C) mmddyy = current date WRP = window reference procedure (A file name of 550-F WRP means the file is the calibration procedure for the unit with serial number 550, using F electronics, taken on January 1, 2001.) Rev K Page 3 of 9
4 Locating the Window Reference Position The window reference position must be carried out continuously without interruption. It takes less than 25 minutes to perform the procedure. If the procedure is interrupted you must start over from step 1 of the Coarse Adjustment procedure as described below. Important! To eliminate the effects of hysteresis, always reach the micrometer position on a Lasentec D600 or FBRM/Raman probe turning to the left. If you pass the desired micrometer position, move the rotating dial at least 50 µm (5 increments on the rotating dial) to the right and then find the correct point moving to the left. Coarse Adjustment During the coarse adjustment, increase the micrometer position in 20-micron increments (2 marked increments on the micrometer knob) in an attempt to pass the peak value for Statistic #5. You will likely make several coarse adjustments before the peak value is passed. To perform coarse adjustment of the micrometer: 1. Set the initial micrometer position by turning the micrometer knob to the right until it reaches the stop. Do not force it past this point. This point should be approximately -70 µm on the micrometer. 2. Type the micrometer position in the Annotation field (e.g., micrometer = -70 µm). 3. Press the NOT Saving, Press to AutoSave button to begin saving to the data file. 4. Press the NOT Measuring, Press to Start button to begin measuring. 5. Wait until the # Meas Since Start box displays 5 (i.e., 5 measurements have been taken), then click the Saving Data: Press to Stop button to stop saving and the Measuring: Press to Stop button to terminate the measurement. 6. Increase the micrometer position by 20 microns (2 marked increments) by rotating the micrometer knob to the left. 7. Type over the previous entry to record this micrometer position (+20 µm greater than the last recorded measurement) in the Annotation field (e.g., micrometer = -50 µm). 8. Repeat steps 3 through 7 (coarse adjustment) while watching the trend display graph. It may take several coarse adjustments before data will display in this graph and counts will begin to register in the Current Value box. Continue to watch the trend display graph, performing another coarse adjustment after every five measurements. Once counts begin to register in the trend display graph you ll notice that they fluctuate up and down by small increments with each measurement and increase dramatically with each adjustment of the micrometer position. If the trend line disappears off the graph at any point, click the Scale All Trends box to reset the Y-axis range. Once the peak value is reached, counts displayed in the Current Value box will begin to decrease and a definite peak will be seen in the trend display graph. Once you see the peak, perform three additional coarse adjustments to verify there is no higher peak, then proceed to step Rev K Page 4 of 9
5 9. To determine the final micrometer position of the coarse adjustment, minimize the Acquisition Module, open the Data Review Module and from the Setup Menu, select Meas Config. 10. Click the Open File button. 11. Select the data file created while the window reference position was being located (C:\Fbrm6?\Backup\ SN-X-mmddyy-WRP) and click Open. 12. For Time Averaging, select No Time Averaging. 13. Click the Apply Time Averaging box. 14. For Channel Grouping, select µm 200 Linear Ch. 15. From the Setup Menu, select Stat. Config. 16. Press the Load Stats. Config button. 17. From the C:\Fbrm6?\Statistics Setup Files folder, select the default0-100-lin.sta statistics setup (the measured statistic #/sec, No Wt 0-5 will reach a maximum value when the focal point is located at the window reference position) and click Open. 18. From the Display Menu, select Stats/Trend. 19. In the Show column, make sure there is an X next to the #/sec, No Wt 0-5 statistic (Statistic #5). This will create a trend plot in the trend display. 20. De-select all other statistics by clicking on any existing Xs in the Show column. De-selecting a statistic will remove the trend from the trend display. 21. From the Y Axis applies to: box, select the #5 statistic. 22. Under X-axis Scale (located below the trend display), select Record. 23. Click the Scale All Trends box to reset the Y-axis. 24. Make sure the Lock Cursor box is not checked (i.e., an X is not displayed), select the + from the trend toolbar (located below the Trend Display), and move the Current Record crosshair on the trend display graph to the peak value measurement of the trended statistic as determined by the coarse adjustment. 25. From the Setup Menu, select Annotation. 26. Look at the Annotation field for the micrometer reading at this maximum value. This is the peak micrometer position of the coarse adjustment. In some cases, the maximum value of the trended statistic may be observed at more than one consecutive micrometer position. If a plateau is observed, the first micrometer position encountered at the maximum value of the statistic (i.e., the left-hand side of the plateau in the trend display) should be used as the final micrometer position of the coarse adjustment Rev K Page 5 of 9
6 Fine Adjustment The micrometer position is increased in one-half micron increments to confirm the peak value for Statistic #5. (One-half an increment on the micrometer know.) Usually, several fine adjustments must be made before the peak value is confirmed. To eliminate the effects of hysteresis, always reach the micrometer position going to the left. If you pass the desired micrometer position, move the micrometer to the right at least 50 microns (5 marked increments) and then find the correct point by moving the micrometer to the left. To perform the fine adjustment of the micrometer: 1. With the software set to the Acquisition Module, turn the micrometer knob to the right to the stop (approximately 70 µm), then carefully turn the knob to the left to set the micrometer position 30 microns (3 marked increments) before the peak micrometer position found in the coarse adjustment (e.g., if the peak micrometer position of the coarse adjustment is +50 µm, set the micrometer position at +20 µm). 2. Type the new micrometer position in the Annotation field (e.g., micrometer = +20 µm). 3. Press the NOT Saving, Press to AutoSave button to begin saving to the data file. 4. Press the NOT Measuring, Press to Start button to begin measuring. 5. Wait until the # Meas Since Start box displays 5 (i.e., 5 measurements have been taken), then click the Saving Data: Press to Stop button to stop saving and the Measuring: Press to Stop button to terminate the measurement. 6. Increase the micrometer position by 5 microns (one-half of one marked increment) by rotating the micrometer knob to the left. 7. Type over the previous entry to record this new micrometer position in the Annotation field (e.g., micrometer = +25 µm). 8. Look at the trend display graph. The counts should be higher than those achieved in the coarse adjustment (i.e., the trend line going up). If the trend line has disappeared off the top of the graph, click the Scale All Trends button to reset the Y-axis. 9. Repeat steps 3 through 8 until the peak value established in the coarse adjustment is passed by 20 microns (20 µm). You will see gradual stair-step increases before the line in the trend display graph gradually decreases. After the peak, the counts will likely fluctuate up and down again, but will not reach as high as they did at the peak. If counts fluctuate so multiple peaks are observed, re-coat the window with the Dry Erase Marker and repeat the fine adjustment from step Once the peak value from the coarse adjustment has been passed by 20 microns (20 µm), open the Data Review Module to determine the exact micrometer position of the maximum peak value using the trend display graph. 11. From the Setup Menu, select Meas Config. 12. Click the Open File button Rev K Page 6 of 9
7 13. Select the data file created while the window reference position was being located (C:\Fbrm6?\Backup\ SN-X-mmddyy-WRP) and click Open. 14. For Time Averaging, select No Time Averaging. 15. Click the Apply Time Averaging box. 16. For Channel Grouping, select µm 200 Linear Ch. 17. From the Setup Menu, select Stat. Config. 18. Press the Load Stats. Config button. 19. From the C:\Fbrm6?\Statistics Setup Files folder, select the default0-100-lin.sta statistics setup (the measured statistic #/sec, No Wt 0-5 will reach a maximum value when the focal point is located at the window reference position) and click Open. 20. From the Display Menu, select Stats/Trend. 21. In the Show column, make sure there is an X next to the #/sec, No Wt 0-5 statistic (Statistic #5). This will create a trend plot in the trend display. 22. De-select all other statistics by clicking on any existing Xs in the Show column. De-selecting a statistic will remove the trend from the trend display. 23. From the Y Axis applies to: box, select the #5 statistic. 24. Under X-axis Scale (located below the trend display), select Record. 25. Click the Scale All Trends box to reset the Y-axis. 26. The trend display should look similar to Figure 2 below, with two distinct peaks displayed. The left peak is from the coarse adjustment and the right peak is from the fine adjustment. 27. Make sure the Lock Cursor box is not checked (i.e., an X is not displayed), select the + from the trend toolbar (located below the Trend Display), and move the Current Record crosshair on the trend display graph to the peak value measurement of the trended statistic as determined by the fine adjustment. 28. From the Setup Menu, select Annotation. 29. Look at the Annotation field for the micrometer reading at this maximum value. The micrometer position corresponding to this value is the window reference position (e.g., +45 µm). In some cases, the maximum value of the trended statistic may be observed at more than one consecutive micrometer position. If a plateau is observed, the first micrometer position encountered at the maximum value of the statistic (i.e., the left-hand side of the plateau in the trend display) should be used as the window reference position (WRP) Rev K Page 7 of 9
8 Figure 2 Example of Trend Display in the Data Review Module Window Reference Procedure Repeatability While all measurements of the window reference position should be within a range of ±5 microns (i.e., ±0.005 mm or one-half of an increment on the rotating dial), due to inconsistencies in ink application, you should not expect actual counts achieved to be the same each time. Contact a Mettler-Toledo Technology and Application Consultant if there is a greater than ±5-micron discrepancy in the measurement of the window reference position. Setting the Micrometer at the Standard Focal Point Position The focal point position defines the location of the focal point relative to the external surface of the Lasentec FBRM probe window and the window of the FBRM/Raman probe. The standard focal point position for both probes is 20 microns (2 marked increments) inside the sapphire window surface. Unless there is a specific, application-based need for an alternate position, use the standard focal point position for all applications. To set the micrometer at the required focal point position for a Lasentec D600 FBRM probe or FBRM/Raman probe: 1. Turn the micrometer knob at least 50 microns to the right from the window reference position (e.g., if the WRP was +45 µm, turn the micrometer to the right to 5µm). 2. Next, turn the micrometer knob to the left to the WRP minus 20 microns (e.g., turn the micrometer to the right from 5 µm to +25 µm). This is the standard focal point position Rev K Page 8 of 9
9 To eliminate the effects of hysteresis, you must always reach the final micrometer position going to the left. If you pass the required focal point position, repeat steps 1 and 2 above. Determining a Clean Probe Window For the purpose of this procedure, the probe window is considered clean if counts are less than 150 per second and not repeated in a single channel when the probe tip is immersed in de-ionized (DI) water. If acetone or other solvents are required to clean the probe window, the solvent should be thoroughly rinsed away with DI water. For other applications, see the section called Cleaning the Probe Window in the hardware manual Rev K Page 9 of 9
Lasentec Product Group
Document Number: 004-0043 Window Reference Procedure - M500/600 P Probes, version 6.0 and higher FBRM CI Software The Window Reference Procedure is the standard calibration procedure used to reference
More informationDetermination of the Equivalent Weight and the K a or K b for a Weak Acid or Base
INTRODUCTION Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base Chemists frequently make use of the equivalent weight (eq. wt.) as the basis for volumetric calculations.
More informationiworx Sample Lab Experiment GB-2: Membrane Permeability
Experiment GB-2: Membrane Permeability Exercise 1: Movement of Small Positive Ions Across a Membrane Aim: To determine if small, positively charged, hydrogen ions can move across a membrane from a region
More informationpka AND MOLAR MASS OF A WEAK ACID
Experiment 10 pka AND MOLAR MASS OF A WEAK ACID Adapted by the Chemistry Faculty of Eastern Michigan University from EQUL 305,written by Richard C. Bell, Lebanon Valley College, published by Chemical Education
More informationPhysics Labs with Computers, Vol. 1 P14: Simple Harmonic Motion - Mass on a Spring A
Activity P14: Simple Harmonic Motion - Mass on a Spring (Force Sensor, Motion Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Harmonic motion P14 SHM.DS P19 SHM Mass on a Spring
More informationMicroscale Acid-Base Titration
Microscale Acid-Base Titration Experiment 36 A titration is a process used to determine the volume of a solution needed to react with a given amount of another substance. In this experiment, you will titrate
More informationFog Monitor 100 (FM 100) Extinction Module. Operator Manual
Particle Analysis and Display System (PADS): Fog Monitor 100 (FM 100) Extinction Module Operator Manual DOC-0217 Rev A-1 PADS 2.7.3, FM 100 Extinction Module 2.7.0 5710 Flatiron Parkway, Unit B Boulder,
More informationProtocol Particle size distribution by centrifugal sedimentation (CPS)
Protocol Particle size distribution by centrifugal sedimentation (CPS) 1. Method The CPS Disc Centrifuge separates particles by size using centrifugal sedimentation in a liquid medium. The sedimentation
More informationTitration of Acids and Bases
Exercise 3 Page 1 Illinois Central College CHEMISTRY 132 Titration of Acids and Bases Name: Equipment 1-25 ml burette 1-pH electrode 1-50 and 1-150 ml beaker 1-stir plate and stir bar 1-Vernier Interface
More informationMagnetic Fields. Experiment 1. Magnetic Field of a Straight Current-Carrying Conductor
General Physics Lab Department of PHYSICS YONSEI University Lab Manual (Lite) Magnetic Fields Ver.20181029 NOTICE This LITE version of manual includes only experimental procedures for easier reading on
More informationKinetics of Crystal Violet Bleaching
Kinetics of Crystal Violet Bleaching Authors: V. C. Dew and J. M. McCormick* From Update March 12, 2013 with revisions Nov. 29, 2016 Introduction Chemists are always interested in whether a chemical reaction
More informationUsing Conductivity to Find an Equivalence Point
Experiment 25 PRE LAB DISCUSSION In this experiment, you will monitor conductivity during the reaction between sulfuric acid, and barium hydroxide in order to determine the equivalence point. From this
More informationShown below is a sample titration curve for a diprotic acid. Note the two equivalence points.
EXPERIMENT 9 Titration Curve for a Polyprotic Acid INTRODUCTION Other than by strength and concentration, another way of classifying acids involves the number of H + ions an acid can donate. A monoprotic
More informationTITRATION CURVES INTRODUCTION. Read and/or review Sections 4.10 and 16.7 in your textbook.
1 TITRATION CURVES Copyright: Department of Chemistry, University of Idaho, Moscow, ID 83844-2343. 2013. INTRODUCTION Read and/or review Sections 4.10 and 16.7 in your textbook. In an acid - base titration,
More informationAcid-Base Titration. Evaluation copy
Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to
More informationDetermining the Concentration of a Solution: Beer s Law. Evaluation copy. Figure 1
Determining the Concentration of a Solution: Beer s Law Computer 17 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You will use a
More informationExperiment P05: Position, Velocity, & Acceleration (Motion Sensor)
PASCO scientific Physics Lab Manual: P05-1 Experiment P05: Position, Velocity, & Acceleration (Motion Sensor) Concept Time SW Interface Macintosh file Windows file linear motion 30 m 500 or 700 P05 Position,
More informationAcid-Base ph Titration Introduction
Electronic Supplementary Material (ESI) for Chemistry Education Research and Practice. This journal is The Royal Society of Chemistry 2016 Appendix B: Example of Traditional Investigation Acid-Base ph
More informationExperiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE
Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE Concepts: Density Equipment Calibration Approximate time required: 90 minutes for density 90 minutes for two thermometers
More informationHooke s Law. Equipment. Introduction and Theory
Hooke s Law Objective to test Hooke s Law by measuring the spring constants of different springs and spring systems to test whether all elastic objects obey Hooke s Law Equipment two nearly identical springs,
More informationEXPERIMENT 15. USING CONDUCTIVITY TO LOOK AT SOLUTIONS: DO WE HAVE CHARGED IONS OR NEUTRAL MOLECULES? rev 7/09
EXPERIMENT 15 USING CONDUCTIVITY TO LOOK AT SOLUTIONS: DO WE AVE CARGED IONS OR NEUTRAL MOLECULES? rev 7/09 GOAL After you complete this experiment, you should have a better understanding of aqueous solutions
More information1iI1E. The Determination of 0 an Equilibrium Constant [LU. Computer
Computer The Determination of 0 an Equilibrium Constant Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant,
More informationfor a Weak Acid 1-ring stand stir plate and stir bar ph Probe ml burette 1-50 ml and ml beaker Drop counter ph 7.00 buffer 0.
Exercise 4 Page 1 Illinois Central College CHEMISTRY 132 Name: Determination of for a Weak Acid Equipment Objectives 1-ring stand stir plate and stir bar ph Probe 1-25.00 ml burette 1-50 ml and 1-250 ml
More informationph electrode Instruction Manual For use with the RAH-210 & RPH-250
ph electrode Instruction Manual For use with the RAH-210 & RPH-250 The information in this manual was current at the time of printing. The most current versions of all Hydro Instruments manuals can be
More informationActivity P20: Conservation of Mechanical Energy (Force Sensor, Photogate)
Name Class Date Activity P20: Conservation of Mechanical Energy (Force Sensor, Photogate) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Energy P20 Mechanical Energy.DS P23 Cons. Mechanical
More informationEvaluation copy. Ground Temperature Measurements. computer OBJECTIVES MATERIALS PRE-LAB QUESTIONS
Ground Temperature Measurements Computer P3 Throughout the daylight hours the sun heats up the air and the soil. Project 1 in this manual investigates the changes in air temperature during this process.
More informationThe Phase Change Lab: Freezing and Melting of Water
The Phase Change Lab: Freezing and Melting of Water Experiment 3 Freezing temperature is the temperature at which a substance turns from a liquid to a solid. Melting temperature is the temperature at which
More informationUniversity of Minnesota Nano Center Standard Operating Procedure
University of Minnesota Nano Center Standard Operating Procedure Equipment Name: Zeta Potential Analyzer Model: Stabino Location: PAN 185 Badger Name: Not on Badger Revision Number: 0-Inital release Revisionist:
More informationLab 12 - Conservation of Momentum And Energy in Collisions
Lab 12 - Conservation of Momentum And Energy in Collisions Name Partner s Name I. Introduction/Theory Momentum is conserved during collisions. The momentum of an object is the product of its mass and its
More informationLAB 2 - ONE DIMENSIONAL MOTION
Name Date Partners L02-1 LAB 2 - ONE DIMENSIONAL MOTION OBJECTIVES Slow and steady wins the race. Aesop s fable: The Hare and the Tortoise To learn how to use a motion detector and gain more familiarity
More informationSample Alignment (2D detector) Part
Sample Alignment (2D detector) Part Contents Contents 1 How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Customizing scan conditions and slit conditions... 6 2 Sample alignment sequence...13
More informationExperiment C-10 Titration of a Strong Acid and a Strong Base
1 Experiment C-10 Titration of a Strong Acid and a Strong Base Objectives To study the titration process. To follow changes in the ph during the titration process while adding a strong base to a strong
More informationBioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)
Liquid Chromatography Mass Spectrometry (LC/MS) Prelab Questions: Questions to be answered before doing the experiment. The answers are due at the beginning of each experiment without exception (the questions
More informationInvestigation #2 TEMPERATURE VS. HEAT. Part I
Name: Investigation #2 Partner(s): TEMPERATURE VS. HEAT These investigations are designed to help you distinguish between two commonly confused concepts in introductory physics. These two concepts, temperature
More informationTi-Pure TITANIUM DIOXIDE. DETERMINATION OF PARTICLE SIZE DISTRIBUTION OF RPS ( Horiba (LA-900) Procedure) METHOD: T WP
Ti-Pure TITANIUM DIOXIDE DETERMINATION OF PARTICLE SIZE DISTRIBUTION OF RPS ( Horiba (LA-900) Procedure) METHOD: T4400.274.03WP 1 TITANIUM DIOXIDE (Ti-Pure ) Determination of Particle Size Distribution
More informationACID-BASE TITRATION (MICROSCALE)
ACID-BASE TITRATION (MICROSCALE) LAB PH 4.PALM From Science with Handhelds, Vernier Software & Technology, 2002. INTRODUCTION Acids and bases represent a major class of chemical substances. We encounter
More informationThis sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.
ph sensor with electrode ENPH-A016 ph sensor with flat electrode ENPH-F052 The ph sensor replaces the traditional ph meter adding powerful capabilities such as collecting the ph data and displaying the
More informationLab 1 Uniform Motion - Graphing and Analyzing Motion
Lab 1 Uniform Motion - Graphing and Analyzing Motion Objectives: < To observe the distance-time relation for motion at constant velocity. < To make a straight line fit to the distance-time data. < To interpret
More informationKinematics Lab. 1 Introduction. 2 Equipment. 3 Procedures
Kinematics Lab 1 Introduction An object moving in one dimension and undergoing constant or uniform acceleration has a position given by: x(t) =x 0 +v o t +1/2at 2 where x o is its initial position (its
More informationRate Law Determination of the Crystal Violet Reaction. Evaluation copy
Rate Law Determination of the Crystal Violet Reaction Computer 30 In this experiment, you will observe the reaction between crystal violet and sodium hydroxide. One objective is to study the relationship
More informationLab 5 Enthalpy of Solution Formation
Chemistry 3202 Lab 5 Enthalpy of Solution Formation Page 1 of 9 Lab 5 Enthalpy of Solution Formation Introduction This lab activity will introduce you to the measurement of energy change associated with
More informationCHEMISTRY 206 Experiment 4: A KINETIC STUDY
CHEMISTRY 206 Experiment 4: A KINETIC STUDY Instructor s Informal Preamble Chemists are interested in figuring out how reactions happen (i.e., mechanisms), and how quickly they occur (i.e., rates). Both
More informationAcid-Base Titration Curves Using a ph Meter
Acid-Base Titration Curves Using a ph Meter Introduction: In this experiment you will use a ph sensor to collect volume and ph data as you titrate two acids with sodium hydroxide. You will obtain titration
More informationAcid-Base Titration Curves Using a ph Meter
Acid-Base Titration Curves Using a ph Meter Introduction: In this experiment you will use a ph sensor to collect volume and ph data as you titrate two acids with sodium hydroxide. You will obtain titration
More informationConductometric Titration & Gravimetric Determination of a Precipitate
Conductometric Titration & Gravimetric Determination of a Precipitate Experiment 9 In this experiment, you will monitor conductivity during the reaction between sulfuric acid, H2SO4, and barium hydroxide,
More information2 (aq) [FeSCN [Fe 3JSCN] Figure 1
The Determination of an Equilibrium Constant Computer Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant,
More informationAlphaLab Gaussmeter Model VGM -- Instructions
AlphaLab Gaussmeter Model VGM -- Instructions Power: As shipped, a standard 9 volt battery is installed. Battery life is about 10 hours (25 ma drain; 6.5 volt dropout). Low battery appears when about one
More informationDetermining the Concentration of a Solution: Beer s Law
Determining the Concentration of a Solution: Beer s Law LabQuest 11 The primary objective of this experiment is to determine the concentration of an unknown nickel (II) sulfate solution. You will be using
More informationExercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES
Exercise 2-4 Titration of a Buffer Solution EXERCISE OBJECTIVES To define the terms buffer solution and buffer capacity; To titrate a buffer solution with a weak acid solution; To plot a graph using the
More informationHeat of Combustion: Magnesium
Heat of Combustion: Magnesium Experiment 21 In Experiment 17, you learned about the additivity of reaction heats as you confirmed Hess s Law. In this experiment, you will use this principle as you determine
More informationLab Activity H4 It s Snow Big Deal
Lab Activity H4 It s Snow Big Deal OUTCOMES After completing this lab activity, the student should be able to use computer-based data acquisition techniques to measure temperatures. draw appropriate conclusions
More informationActivity P27: Speed of Sound in Air (Sound Sensor)
Activity P27: Speed of Sound in Air (Sound Sensor) Concept Speed of sound DataStudio P27 Speed of Sound 1.DS Equipment Needed Qty Other Qty Sound Sensor (CI-6506B) 1 Tape, duct 1 roll Base and Support
More informationExploring the Poles (Without Leaving Your Classroom!)
Exploring the Poles (Without Leaving Your Classroom!) Computer 37 Magnets have north and south poles. Do you think that the poles of differently shaped magnets are in different places? In this activity,
More informationIV. EQUIPMENT a. Horiba LA 950 main unit b. Dry sample unit that has a silencer behind it c. Cell slide unit that holds
STANDARD OPERATING PROCEDURE No of pages 7 No: METH 006 Version: 1 Date: 05-06- 2011 Title: Particle Size Distribution Analysis by Laser Diffraction method with HORIBA LA 950 in wet mode I. SCOPE AND APPLICATION
More informationAdditivity of Heats of Reaction: Hess s Law
Additivity of Heats of Reaction: Hess s Law Experiment 21 In this experiment, you will use a Styrofoam-cup calorimeter to measure the heat released by three reactions. One of the reactions is the same
More informationStandardizing a Solution of Sodium Hydroxide. Evaluation copy
Standardizing a Solution of Sodium Hydroxide Computer 6 It is often necessary to test a solution of unknown concentration with a solution of a known, precise concentration. The process of determining the
More informationChemical Kinetics: Integrated Rate Laws. ** updated Procedure for Spec 200 use **
Chemical Kinetics: Integrated Rate Laws ** updated Procedure for Spec 200 use ** *DISCLAIMER: It is highly recommended that students bring in their own computers to lab this week to use excel. There may
More informationExperiment 2: The Speed of Light
Experiment 2: The Speed of Light Modern Physics Laboratory Department of Physics and Astronomy Austin Peay State University September 12, 2006 Abstract In this experiment you will determine the value of
More informationMeasurements of a Table
Measurements of a Table OBJECTIVES to practice the concepts of significant figures, the mean value, the standard deviation of the mean and the normal distribution by making multiple measurements of length
More informationCart on a Ramp. Evaluation Copy. Figure 1. Vernier Dynamics Track. Motion Detector Bracket
Cart on a Ramp Computer 3 This experiment uses an incline and a low-friction cart. If you give the cart a gentle push up the incline, the cart will roll upward, slow and stop, and then roll back down,
More informationConnect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer.
Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer. The screen shown in Fig. 1 may be displayed. If status line displays
More informationDensity of Aqueous Sodium Chloride Solutions
Experiment 3 Density of Aqueous Sodium Chloride Solutions Prepared by Ross S. Nord and Stephen E. Schullery, Eastern Michigan University PURPOSE Determine the concentration of an unknown sodium chloride
More informationExperiment 7 : Newton's Third Law
Experiment 7 : Newton's Third Law To every action there is always opposed an equal reaction, or the mutual actions of two bodies upon each other are always equal, and directed to contrary parts. If you
More informationExperimental Procedure
Experimental Procedure Overview The ph meter is used in conjunction with a titration apparatus and a standardized sodium hydroxide solution to determine the molar concentration of a weak acid solution
More informationExperiment 2: Reaction Stoichiometry by Thermometric Titration
Experiment 2: Reaction Stoichiometry by Thermometric Titration Introduction The net result of a reaction (a chemical change) is summarized by a chemical equation. In order to write a chemical equation,
More informationChemistry MultiMeasure Sensor PS-2170
ä 5 Instruction Manual 012-09523A Chemistry MultiMeasure Sensor PS-2170 ph/ise/orp PRESSURE CHEMISTRY TEMPERATURE SENSOR VOLTAGE 1 2 3 4 PS-2170 5 6 7 cc 8 10 15 20 25 30 35 40 45 50 55 60 Included Equipment
More informationAssignment #0 Using Stellarium
Name: Class: Date: Assignment #0 Using Stellarium The purpose of this exercise is to familiarize yourself with the Stellarium program and its many capabilities and features. Stellarium is a visually beautiful
More informationAcid-Base Titration. Sample
Acid-Base Titration Computer 7 A titration is a process used to determine the volume of a solution that is needed to react with a given amount of another substance. In this experiment, your goal is to
More informationEXPERIMENT 9 SALTWATER CONDUCTANCE: The Effect of Concentration
EXPERIMENT 9 SALTWATER CONDUCTANCE: The Effect of Concentration Introduction According to the Theory of Ionization proposed by S. Arrhenius, about 1880, ionic compounds dissolve in water forming cations
More informationNE01 - Centripetal Force. Laboratory Manual Experiment NE01 - Centripetal Force Department of Physics The University of Hong Kong
Background Introduction Laboratory Manual Experiment Department of Physics The University of Hong Kong Circular Motion is one of the simplest forms of 2-dimensional motion in which the locus of the object
More informationConservation of Momentum Using PASCO TM Carts and Track to Study Collisions in One Dimension
14 Conservation of Conservation of Using PASCO TM Carts and Track to Study s in One Dimension When two objects collide momentum is transferred between them. p is defined as the product of mass and velocity
More informationActivity P60: Inverse Square Law Nuclear (Nuclear Sensor, Rotary Motion Sensor)
Name Class Date Activity P60: Inverse Square Law Nuclear (Nuclear Sensor, Rotary Motion Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Radioactivity P60 Nuclear Inv Sqr Law.DS P60
More informationPolar alignment in 5 steps based on the Sánchez Valente method
1 Polar alignment in 5 steps based on the Sánchez Valente method Compared to the drift alignment method, this one, allows you to easily achieve a perfect polar alignment in just one step. By "perfect polar
More informationREVIEW OF LAB TECHNIQUES
Experiment 1 REVIEW OF LAB TECHNIQUES Prepared by Masanobu M. Yamauchi and Ross S. Nord, Eastern Michigan University PURPOSE To review density calculations, Beer s Law and the use of electronic balances,
More informationHESS S LAW: ADDITIVITY OF HEATS OF REACTION
HESS S LAW: ADDITIVITY OF HEATS OF REACTION From Chemistry with Calculators, Vernier Software & Technology In this experiment, you will use a Styrofoam-cup calorimeter to measure the heat released by three
More informationName Date Period. 1. If drops of ACID are added to a ph buffer, then the ph of the buffer will [increase / decrease / stay the same].
Name Date Period ACIDS AND BASES Organisms are often very sensitive to the effect of s and s in their environment. They need to maintain a stable internal ph in order to survive even in the event of environmental
More informationExperiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013
Absorbance Experiment 13 Dilutions and Data Handling in a Spreadsheet rev 1/2013 GOAL: This lab experiment will provide practice in making dilutions using pipets and introduce basic spreadsheet skills
More informationInverted Pendulum System
Introduction Inverted Pendulum System This lab experiment consists of two experimental procedures, each with sub parts. Experiment 1 is used to determine the system parameters needed to implement a controller.
More informationKinematics. Become comfortable with the data aquisition hardware and software used in the physics lab.
Kinematics Objective Upon completing this experiment you should Become comfortable with the data aquisition hardware and software used in the physics lab. Have a better understanding of the graphical analysis
More informationForce vs time. IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report
IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report Newton s second law may be written r r F dt = p where F is the force and p is the change in momentum. The area under the force vs. time
More informationZetasizer Nano-ZS User Instructions
Zetasizer Nano-ZS User Instructions 1. Activate the instrument computer by logging in to CORAL. If needed, log in to the local instrument computer Username: zetasizer. Password: zetasizer. 2. Instrument
More informationExperiment P30: Centripetal Force on a Pendulum (Force Sensor, Photogate)
PASCO scientific Physics Lab Manual: P30-1 Experiment P30: (Force Sensor, Photogate) Concept Time SW Interface Macintosh File Windows File centripetal force 30 m 500 or 700 P30 Centripetal Force P30_CENT.SWS
More informationJasco V-670 absorption spectrometer
Laser Spectroscopy Labs Jasco V-670 absorption spectrometer Operation instructions 1. Turn ON the power switch on the right side of the spectrophotometer. It takes about 5 minutes for the light source
More informationBecome a Microprobe Power User Part 2: Qualitative & Quantitative Analysis
Become a Microprobe Power User Part 2: Qualitative & Quantitative Analysis Mike Spilde Spring IOM Seminar February 5, 2008 Qualitative Analysis Why use qualitative scans? Elemental ID (especially trace
More informationTranspiration. Evaluation copy
Transpiration Computer 9 Water is transported in plants, from the roots to the leaves, following a decreasing water potential gradient. Transpiration, or loss of water from the leaves, helps to create
More informationThe Synthesis and Analysis of Aspirin
The Synthesis and Analysis of Aspirin Computer 22 Aspirin, the ubiquitous pain reliever, goes by the chemical name acetylsalicylic acid. One of the compounds used in the synthesis of aspirin is salicylic
More informationAmmonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A
imagine explore learn Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A Overview Ammonium ions (NH 4+ ) can be found in fertilizer, cleaning products, processed meats and in the waste of
More informationExperiment 11 Beer s Law
Experiment 11 Beer s Law OUTCOMES After completing this experiment, the student should be able to: determine the wavelength (color) of maximum absorbance for a solution. examine the relationship between
More informationDetermining the K sp of Calcium Hydroxide
Determining the K sp of Calcium Hydroxide (Titration Method) Computer 23 Calcium hydroxide is an ionic solid that is sparingly soluble in water. A saturated, aqueous, solution of Ca(OH) 2 is represented
More informationPHY222 Lab 2 - Electric Fields Mapping the Potential Curves and Field Lines of an Electric Dipole
Print Your Name PHY222 Lab 2 - Electric Fields Mapping the Potential Curves and Field Lines of an Electric Dipole Print Your Partners' Names Instructions January 23, 2015 Before lab, read the Introduction,
More informationExperiment 11 Beer s Law
Experiment 11 Beer s Law OUTCOMES After completing this experiment, the student should be able to: determine the wavelength (color) of maximum absorbance for a solution. examine the relationship between
More informationThe Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
More informationGeneral Physics I Lab. M1 The Atwood Machine
Purpose General Physics I Lab In this experiment, you will learn the basic operation of computer interfacing and use it in an experimental study of Newton s second law. Equipment and components Science
More informationOcean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)
Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS) The UV-VIS spectrometer is a portable ultraviolet light and visible light spectrophotometer, combining a spectrometer and a light source/cuvette
More informationHow to perform 2D NMR Experiments on the Varian/Agilent VNMRS 500 when using the Chempack interface
How to perform 2D NMR Experiments on the Varian/Agilent VNMRS 500 when using the Chempack interface 1 June 3, 2014 To start: 1. Insert your sample 2. If running any 2D spectrum or even just a 1D 13 C spectrum
More informationAdditivity of Heats of Reaction: Hess s Law
Additivity of Heats of Reaction: Hess s Law Experiment In this experiment, you will use a Styrofoam-cup calorimeter to measure the heat released by three reactions. One of the reactions is the same as
More informationKa of Unknown Acid In this experiment you will determine the Ka of an unknown acid by titration with the sodium hydroxide.
Ka of Unknown Acid In this experiment you will determine the Ka of an unknown acid by titration with the sodium hydroxide. Because you will be titrating an unknown acid again, you will be using many of
More informationActivity P11: Collision Impulse and Momentum (Force Sensor, Motion Sensor)
Name Class Date Activity P11: Collision Impulse and Momentum (Force Sensor, Motion Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Newton s Laws P11 Impulse.DS P14 Collision P14_COLL.SWS
More informationDustMon A Standard Operating Procedure From the Microtrac Particle Analysis Laboratory (PAL) SL-AN-44 Revision C
DustMon A Standard Operating Procedure From the Microtrac Particle Analysis Laboratory (PAL) SL-AN-44 Revision C Introduction This operating procedure is provided to assist the user of the Dustmon in obtaining
More informationHarmonic Motion. Mass on a Spring. Physics 231: General Physics I Lab 6 Mar. 11, Goals:
Physics 231: General Physics I Lab 6 Mar. 11, 2004 Names: Harmonic Motion Goals: 1. To learn about the basic characteristics of periodic motion period, frequency, and amplitude 2. To study what affects
More information