Instrumental Limitations for High Absorbance Measurements in Noise

Size: px
Start display at page:

Download "Instrumental Limitations for High Absorbance Measurements in Noise"

Transcription

1 TECHNICAL NOTE Instrumental Limitations for High Absorbance Measurements in Noise UV/Vis/NIR Spectroscopy Author: Jeffrey L. Taylor, Ph.D. PerkinElmer, Inc. Shelton, CT Spectrophotometer Energy Using a Fixed Slit Width Along with instrumental stray light noise is the other major contributing factor limiting high absorbance measurements. As can be seen in Figure 1, the amount of radiant energy in a spectrophotometer varies greatly as a function of wavelength. The amount of noise in a UV/Vis/NIR spectrum is directly related to these energy levels. The ability to measure high absorbance values is critically dependent on the amount of noise close to the 0 %T axis. High absorbance measurements will be more difficult to obtain in low energy wavelength regions of the instrument. Key instrumental parameters associated with the control of spectrum noise are slit width, detector gain, and integration (response) time. LAMBDA 1050 Figure 1. Radiant energy in a spectrophotometer. Note: Low energy regions below 400 nm, around 860 nm, and above 3000 nm

2 Spectrophotometer Energy Using a Servo Slit at Set Gain One common methodology for improving noise through the use of instrumental parameters is to servo the slit, which varies the resolution of the measurement, as a function of wavelength. This technique has the advantage of maintaining a constant energy and noise level throughout the NIR wavelength range. The Servo Slit mode maintains 70% energy through almost the entire NIR region, with the exception of the long wavelength area above 2900 nm. The energy declines here because the slit is open to its maximum value of 20 nm. Once the slit has opened to this maximum size, the energy will start to decline. The Detector Range Problem and Noise As seen in Figure 2, as a sample absorbs more light, less energy falls on the detector. The higher the sample absorbance, the greater the noise contribution to the spectrum. At some point the signal disappears into the noise background of the detector. This is why many spectrophotometers cite noise specifications at several increasing absorbance levels. How a given spectrophotometer s software, firmware, processes noisy, high absorbance data is very important to understanding the photometric accuracy in samples of high absorbance. In Figure 3, the purple spectrum is around 520 nm, not only does the spectrum noise increase, it would appear that the noise is causing the absorbance to go off scale. The subject of what happens as a function of noise for a highly absorbent sample is in considerable detail to the right. The Problem With High Absorbance Spectrophotometers are unable to measure sample absorbance directly. They can only directly measure transmittance of a sample. Absorbance is a post-data acquisition processing step. This fact is critical in understanding high absorbance spectra. High absorbance samples should always be measured in %T mode, so that one can directly see what is happening to the signal in the instrument. The spectrum in Figure 4, is a 7-plus absorbance blocking filter. Note the %T values for the blocking regions. The values are so close to zero %T that some of the values are negative due to instrumental noise. This causes a slight data handling problem for the absorbance processing step. Beer-Lambert Law and Negative Percent Transmission Values At top left is a graphical representation of the relationship between percent transmission and absorbance values. The logarithmic nature of absorbance is apparent. On the bottom is the Beer-Lambert Law's conversion from percent transmission to absorbance. Remember those negative %T values due to noise for a highly absorbing sample? They represent a real problem. What is the logarithm of a negative number? The Log function Y = Log b (X) is the inverse of the exponential function X = b Y. Since the base b is positive (b>0), the base b raised to the power of Y must be positive (by>0) for any real value of y; therefore, the number x must be positive (x>0) as well. So the real base b logarithm of a negative number must be undefined. Figure 2. Servo Slit. Note: Water noise energy regions around 2600 nm, 1800 nm, and 1400 nm Figure 3. Didymium filters scanned on the PerkinElmer LAMBDA 950 without reference beam attenuation. Figure 4. 7-plus absorbance blocking filter. Note: The 10-6 % Transmission Y-axis range for this sample 2

3 Since spectra can not have holes in the wavelength axis due to undefined absorbance values, something must be done. The only solution is to substitute a real value that is on the same order of magnitude as the noise data. While this data can serve as a placeholder, it is inserting an unmeasured, artifactual value into the spectrum. Figure 5. The relationship between absorbance and percent transmission. A = 2 - log 10 %T Smoothing to Eliminate Negative %T Values One methodology that can be employed to eliminate the negative %T values is to minimally smooth the original percent transmission spectrum. This smoothing averages the data above the 0 %T threshold thereby eliminating the negative values in the native spectrum. This result can be seen in Figure 6, obtained from that 7-plus absorbance blocking filter. The original raw data, red spectrum, is subject to a 5 point smoothing which results in the blue spectrum where all %T values are positive. The resulting absorbance transformation would then be free from any artifactual data due to undefined absorbance from negative %T numbers. Since the correct minimal level of smoothing is basically a trialand-error procedure, this type of methodology is difficult to automate properly in instrument software. Total Beam Block in the UV Region So what happens in a spectrophotometer when the beam is completely blocked? The answer depends on how the data are processed before presentation in a spectrum. Some spectrophotometers smooth the spectrum internally before presentation. PerkinElmer is committed to never altering raw data so that the user is unable to see the exact nature of the original data. A highly absorbing sample or blocked beam spectrum in a PerkinElmer spectrophotometer will always be very noisy and centered around zero percent transmission. Since minimal energy strikes the detector at high sample absorbance, the only signal generated is random detector noise. The intensity of that noise will depend on the detector gain setting. The spectra here are all measured in the UV region. In Figure 7, the top spectrum in blue is from a 7 absorbance neutral density screen; whereas, the red spectrum is the result of a totally blocked beam. Note the abundance of negative percent transmission values in the blocked beam spectra. When the spectra are all converted to absorbance (bottom graph), the blocked beam spectrum displays noise truncation at around 8 absorbance, but the neutral density screen appears normal. This truncation is the direct result of the insertion of static 8 absorbance values for negative %T values. Figure 7. Spectra measured in the UV regions a 7 absorbance neutral density screen (blue) and the result of a totally blocked beam (red). Figure 6. A 7-plus absorbance blocking filter in the UV region. Figure 8. Spectra converted to absorbance. 3

4 Total Beam Block in the Visible Region Next, we will consider the visible region. This region typically has the highest relative energy in a UV/Vis/NIR spectrophotometer. In this case, three samples were measured. Their absorbance spectra can be seen in Figure 9. In Figure 10, samples are a 5.8 absorbance (green) and 7.2 absorbance (blue) neutral density screen, as well as a totally blocked beam (red). Note the increasing noise as the sample absorbance increases. This is due to the abundance of negative percent transmission values. The blocked beam spectrum appears smoothed, but in actuality it is the number of substituted 8 absorbance values for undefined negative percent transmission numbers that causes the spectrum to look noise free. Figure 9 displays the associated raw percent transmission spectra for the above mentioned absorbance spectra. Total Beam Block in the NIR Region The last region, the NIR, employs a different lead sulfide detector rather than the PMT used in the UV/Visible region. This solid state PbS detector has a different noise profile and is not as sensitive as the PMT detector. Consequently, it has a larger noise envelope and smaller absorbance dynamic range than the PMT. This can be seen Figure 11 and Figure 12 for a totally blocked beam (blue) and a 5.4 absorbance screen sample (red). Note that in the top percent transmission spectra the blocked beam noise appears to be slightly higher than the zero axis, thereby resulting in a smaller proportion of negative percent transmission values. This translates into a much nosier absorbance spectrum in Figure 12. The smaller absorbance range and difference in noise profile for the NIR detector are clearly be observed here. Figure 9. Raw percent transmission versus absorbance spectra. Figure 11. Comparison of blocked beam spectrum in blue and a spectrum of a 5.4A screen sample in red. Figure 10. Noise as the sample absorbance increases. Figure 12. Negative percent transmission values result in increased noise in absorbance spectrum. 4

5 Table 1. Statistics on noise envelop in three major spectral regions. Spectrum Mean %T Absorbance P to P Noise Root Mean Square Noise Blocked Beam UV 3.4 x x x 10-8 Blocked Beam Visible 5.6 x x x 10-9 Blocked Beam NIR (PbS) 2.9 x x x 10-9 The Statistics of a Blocked Beam A completely blocked sample beam should result in an envelope of random noise centered around the zero percent transmission axis. The size of this noise envelope will depend on the instrumental detector and the spectral region being measured. Table 1 displays statistics on this noise envelop in three major spectral regions, the ultra-violet, the visible, and the near infrared. As expected, the values in the table reflect the relative energy differences in the three regions. The region able to measure the highest absorbance (lowest noise) is the visible, followed by the UV, and then the NIR. This is most apparent from the root mean square calculation of the noise as well as the peak to peak noise envelope. The mean of the percent transmission yields a rough estimate of the maximum absorbance limit. Scanning Versus Static Wavelength Measurements Noise in a spectrum is usually greater than the noise associated with a static wavelength measurement. While it is true that by increasing the data point response time of a scan the noise level is reduced, there will always be a minimal noise level associated with a moving monochromator resulting in gated data acquisition from the detector. Therefore, the best quality high-absorbance measurement will often result from a static wavelength measurement. The spectrum in Figure 13 is the 7-plus absorbance blocking filter reference previously. Note the three noisy high absorbance blocking regions. Measuring a selected wavelength from each of these regions blocking regions in a static mode should give insight into just how accurate high absorbance measurements are. Static Single Wavelength Measurements Table 2 displays static wavelength measurements obtained from a 7-plus absorbance blocking filter. Wavelength data from the three blocking regions were measured at two different slit settings and the mean absorbance values were calculated from 10 individual measurements of 10-second integration time duration. In general, the measurements performed with the 5 nm slit had lower standard deviations for the three wavelengths than the lower energy 1 nm slit configuration. The reduction in noise level is about one order of magnitude improvement from 1 nm to a 5 nm slit. The standard deviation is approximately 1% to 3% of the absorbance value at 1 nm slit and about 0.1% to 0.3% for a slit of 5 nm. This is excellent precision for a high absorbance measurement. The last two entries in the table represent a stacked sample+attenuator experiment to demonstrate that this filter had not maxed-out the ordinate dynamic range. Figure 13. A 7-plus absorbance plus blocking filter in the UV region. Table 2. Static wavelength measurements obtained from a 7-plus absorbance blocking filter. Wavelength Slit Width Absorbance Mean Standard Deviation 255 nm 1 nm nm 1 nm nm 1 nm nm 5 nm nm 5 nm nm 5 nm nm No Attenuator 350 nm 0.6A Attenuator 5 nm nm

6 High Absorbance Dynamic Range Test Highly absorbent samples always challenge the stray light and noise capabilities of the instrument. Any measurement above 7 absorbance relies on the sample to be its own stray light filter. There is a simple test that can validate the applicability of the instrument in making reasonable measurements at elevated absorbance levels. By stacking a 25 %T (0.6 abs.) neutral density screen with the sample, the spectral absorbance should be uniformly offset by the absorbance of the added screen. As seen in Figure 14, the stacked screen/sample spectrum (blue) is offset from the original sample (red) spectrum. A spectral subtraction of the two spectra yields a fairly flat, uniform spectra with a 0.72 absorbance mean. If the subtraction spectrum is reasonably flat, the spectral detail preserved in the screen stacked spectrum, along with the stacked spectral offset linear with respect to the screen absorbance, the measurement is likely valid. Figure 14. A 7-plus absorbance plus blocking filter in the UV region. PerkinElmer, Inc. 940 Winter Street Waltham, MA USA P: (800) or (+1) For a complete listing of our global offices, visit Copyright , PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners _02 PKI

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer C101-E107B UV-3600 Shimadzu UV-VIS-NIR Spectrophotometer High sensitivity, high resolution, and an ultra-low stray-light level achieved with the latest technology lead the way to new solutions. UV-3600

More information

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Summary The far UV region (generally covering

More information

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N

More information

Whole Tablet Measurements Using the Frontier Tablet Autosampler System

Whole Tablet Measurements Using the Frontier Tablet Autosampler System a p p l i c a t i o n N O T E Whole Tablet Measurements Using the Frontier Tablet Autosampler System FT-NIR Spectroscopy Introduction Recent advances in NIR technology have changed the ways in which both

More information

RamanStation 400: a Versatile Platform for SERS Analysis

RamanStation 400: a Versatile Platform for SERS Analysis FIELD APPLICATION REPORT Raman Spectroscopy Author: Dean H. Brown PerkinElmer, Inc. Shelton, CT USA RamanStation 400 RamanStation 400: a Versatile Platform for SERS Analysis Introduction Surface Enhanced

More information

Traceable UV-Vis Reference Materials

Traceable UV-Vis Reference Materials UV-Vis Spectroscopy Traceable UV-Vis Reference Materials For Regulatory Compliance Introduction Many laboratories, particularly in the pharmaceutical industry, are under increasing pressure to show proof

More information

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer APPLICATION NOTE FT-NIR Spectroscopy Authors: Ian Robertson Jerry Sellors PerkinElmer, Inc. Seer Green, UK Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer Introduction

More information

Using Calibrated Specular Reflectance Standards for Absolute and Relative Reflectance Measurements

Using Calibrated Specular Reflectance Standards for Absolute and Relative Reflectance Measurements Using Calibrated Specular Reflectance Standards for Absolute and Relative Reflectance Measurements Applications Overview here are two fundamental techniques for measuring specular reflectance with a UV/VIS/NIR

More information

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology APPLICATION NOTE Gas Chromatography Author: David Scott PerkinElmer, Inc. Shelton, CT Residual Solvents in Pharmaceuticals by USP Chapter Methodology Introduction The synthesis of active pharmaceutical

More information

HYPHENATED TECHNOLOGY GUIDE

HYPHENATED TECHNOLOGY GUIDE HYPHENATED TECHNOLOGY GUIDE TABLE OF CONTENTS TG-IR...3 Hiden Analytical MS Systems for TG-MS...4 TG-MS...5 TG-GC/MS...6 TG-IR-GC/MS...8 HYPHENATED TECHNOLOGY GUIDE Coupling, or hyphenating, two instruments

More information

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS A P P L I C AT I O N N O T E ICP - Mass Spectrometry Authors: Lee Davidowski Chady Stephan PerkinElmer, Inc. Shelton, CT The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical

More information

MOLEBIO LAB #4: Using a Spectrophotometer

MOLEBIO LAB #4: Using a Spectrophotometer Introduction: Spectrophotometry MOLEBIO LAB #4: Using a Spectrophotometer Many kinds of molecules interact with or absorb specific types of radiant energy in a predictable fashion. For example, when while

More information

FLUORESCENCE APPLICATIONS

FLUORESCENCE APPLICATIONS FLUORESCENCE APPLICATIONS CORRECTION OF EMISSION SPECTRA USING THE PERKINELMER MODEL LS-50 LUMINESCENCE SPECTROMETER Spectra collected on luminescence spectrometers consist of true spectral information

More information

1 WHAT IS SPECTROSCOPY?

1 WHAT IS SPECTROSCOPY? 1 WHAT IS SPECTROSCOPY? 1.1 The Nature Of Electromagnetic Radiation Anyone who has been sunburnt will know that light packs a punch: in scientific terms, it contains considerable amounts of energy. All

More information

HYPHENATED TECHNOLOGY GUIDE

HYPHENATED TECHNOLOGY GUIDE HYPHENATED TECHNOLOGY GUIDE TABLE OF CONTENTS TG-IR...3 Hiden Analytical MS Systems for TG-MS...4 TG-MS...5 TG-GC/MS...6 TG-IR-GC/MS...8 HYPHENATED TECHNOLOGY GUIDE Coupling, or hyphenating, two instruments

More information

Chem 310 rd. 3 Homework Set Answers

Chem 310 rd. 3 Homework Set Answers -1- Chem 310 rd 3 Homework Set Answers 1. A double line labeled S 0 represents the _ground electronic_ state and the _ground vibrational_ state of a molecule in an excitation state diagram. Light absorption

More information

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS APPLICATION NOTE Liquid Chromatography/ Mass Spectrometry Authors: Roberto Troiano Milan, Italy Wilhad M. Reuter Shelton, CT Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS Introduction

More information

Evaluating Labsphere s new UV-2000

Evaluating Labsphere s new UV-2000 Ciba Grenzach GmbH Optical Spectroscopy Sabrina Siegle Dr. Wolfgang Schlenker Tel. +49 76 24 / 12-28 13 E-mail wolfgang.schlenker@ciba.com May 26, 9 Evaluating Labsphere s new UV- Abstract Labsphere s

More information

Beer's Law and Data Analysis *

Beer's Law and Data Analysis * OpenStax-CNX module: m15131 1 Beer's Law and Data Analysis * Mary McHale This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 2.0 1 Beer's Law and Data Analysis

More information

Whole Tablet Measurements Using the Spectrum One NTS Tablet Autosampler System

Whole Tablet Measurements Using the Spectrum One NTS Tablet Autosampler System Whole Tablet Measurements Using the Spectrum One NTS Tablet Autosampler System A P P L I C A T I O N N O T E Introduction Recent advances in NIR technology have changed the ways in which both the pharmaceutical

More information

Accurate Measurement of Transmittance and Reflectance for Engineering Applications

Accurate Measurement of Transmittance and Reflectance for Engineering Applications Accurate Measurement of Transmittance and Reflectance for Engineering Applications Dr. Chen Fangzhi Laboratory Manager Façade & Roof Materials Testing Laboratory OTM Solutions Pte Ltd PerkinElmer INTour

More information

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY UV/Vis Spectroscopy is a technique that is widely used to characterize, identify and quantify chemical compounds in all fields of analytical chemistry. The

More information

SPECTROPHOTOMETERS. Visible and UV-Visible

SPECTROPHOTOMETERS. Visible and UV-Visible Visible and UV-Visible The Manifacturer ONDA Spectrophotometers are manifactured by a company with over than ten years of experience in development of UV/Vis instruments, Single beam and Double beam, with

More information

Integrating Spheres in Molecular Spectrophotometry

Integrating Spheres in Molecular Spectrophotometry Integrating Spheres in Molecular Spectrophotometry Theory and Practice 2012 Perkin Elmer 2012 Perkin Elmer General Sphere Theory 3 Integrating Spheres Types of Sphere Measurements Total Reflectance (Specular

More information

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry 13A Measurement Of Transmittance and Absorbance Absorption measurements based upon ultraviolet and visible radiation

More information

Accurate Sizing and Precise Counting of 10 nm Gold Nanoparticles using the Enhanced Sensitivity of the NexION 2000 ICP-MS

Accurate Sizing and Precise Counting of 10 nm Gold Nanoparticles using the Enhanced Sensitivity of the NexION 2000 ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author: Chady Stephan PerkinElmer, Inc. Shelton, Inc. Accurate Sizing and Precise Counting of 10 nm Gold Nanoparticles using the Enhanced Sensitivity of the NexION

More information

Background Information:

Background Information: Measurement of Light in Radiation Damaged Quartz Bars for Q weak Katie Kinsley Advisors: Dave Mack, Julie Roche Thomas Jefferson National Accelerator Facility Newport News, VA Summer 2007 (final edit Sept

More information

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV 9//22 OBJECTIVES Spectrophotometry of Food Dyes & Beer s Law Last Update: 9//22 9:54 AM What is the quantitative basis for the color of substances? How is the absorption/transmission of light measured?

More information

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications Bedanta Kr. Deka, D. Thakuria, H. Bora and S. Banerjee # Department of Physicis, B. Borooah College, Ulubari,

More information

Ultraviolet-Visible and Infrared Spectrophotometry

Ultraviolet-Visible and Infrared Spectrophotometry Ultraviolet-Visible and Infrared Spectrophotometry Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451

More information

High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader

High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader TECHNICAL APPLICATION NOTE High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader NGS Automation Image or Color Block Area Next generation

More information

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

Spectrophotometry. Introduction

Spectrophotometry. Introduction Spectrophotometry Spectrophotometry is a method to measure how much a chemical substance absorbs light by measuring the intensity of light as a beam of light passes through sample solution. The basic principle

More information

Ultraviolet-Visible and Infrared Spectrophotometry

Ultraviolet-Visible and Infrared Spectrophotometry Ultraviolet-Visible and Infrared Spectrophotometry Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451

More information

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

UV-Vis optical fiber assisted spectroscopy in thin films and solutions UV-Vis optical fiber assisted spectroscopy in thin films and solutions Description UV-Visible absorption and transmission spectra provide fundamental information for all experiments related to the attenuation

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Introduction What is FTIR? FTIR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR

More information

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF application Note Liquid Chromatography/ Mass Spectrometry Author Sharanya Reddy PerkinElmer, Inc. Shelton, CT USA Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

More information

An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13)

An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13) An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13) Beer s Law: A = -log T = -logp 0 / P = e x b x C See Table 13-1 for terms. In measuring absorbance or transmittance, one should

More information

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Lee Marotta Tom Kwoka David Scott Miles Snow Toby Astill PerkinElmer, Inc. Shelton, CT Fast, Quantitative Analysis of Residual Solvents in

More information

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS INTRODUCTION In recent years, researchers involved in many unrelated scientific disciplines have acquired an interest in accurately

More information

INTRODUCTION The fundamental law of spectrophotometry is known as the Beer-Lambert Law or Beer s Law. It may be stated as: log(po/p) = A

INTRODUCTION The fundamental law of spectrophotometry is known as the Beer-Lambert Law or Beer s Law. It may be stated as: log(po/p) = A S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY AIM 1. To become familiar with the operation of a conventional scanning ultra-violet spectrophotometer 2. To determine suitable cells and solvents

More information

A Calibration Procedure Which Accounts for Non-linearity in Singlemonochromator Brewer Ozone Spectrophotometer Measurements

A Calibration Procedure Which Accounts for Non-linearity in Singlemonochromator Brewer Ozone Spectrophotometer Measurements Atmos. Meas. Tech. Discuss., https://doi.org/.194/amt-18-17 A Calibration Procedure Which Accounts for Non-linearity in Singlemonochromator Brewer Ozone Spectrophotometer Measurements Zahra Vaziri Zanjani

More information

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR Ensuring ultimate cleanliness for maximum adhesion Application Note Author John Seelenbinder Agilent Technologies,

More information

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters ) September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum

More information

Basics of a/b Discrimination for Liquid Scintillation Counting

Basics of a/b Discrimination for Liquid Scintillation Counting application Note Liquid Scintillation Basics of a/b Discrimination for Liquid Scintillation Counting Author Ronald Edler, PhD and Chuck Passo PerkinElmer, Inc. 940 Winter Street Waltham, MA USA Introduction

More information

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept. Spectrum of Electromagnetic Radiation Electromagnetic radiation is light. Different energy light interacts with different motions in molecules. CHEM*344 Chemical Instrumentation Topic 7 Spectrometry Radiofrequency

More information

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES APPLICATION NOTE ICP-Optical Emission Spectroscopy Author: Nick Spivey PerkinElmer, Inc. Shelton, CT Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES Introduction Micronutrients contained

More information

FLS980 Series Reference Guide

FLS980 Series Reference Guide FLS980 Series Reference Guide Integrating Sphere for Measurements of Fluorescence Quantum Yields and Spectral Reflectance Revision 1 Copyrights Copyright 2016 Edinburgh Instruments Ltd. All rights reserved.

More information

PD300 PD300-1W PD300-3W PD300-TP

PD300 PD300-1W PD300-3W PD300-TP 1.1.1 Photodiode Power 1.1.1.1 Standard Photodiode 50pW to 3W Very large dynamic range Swivel mount for hard to measure places Comes with filter in / filter out options Patented automatic background subtraction

More information

Analysis of Cocoa Butter Using the SpectraStar 2400 NIR Spectrometer

Analysis of Cocoa Butter Using the SpectraStar 2400 NIR Spectrometer Application Note: F04 Analysis of Cocoa Butter Using the SpectraStar 2400 NIR Spectrometer Introduction Near-infrared (NIR) technology has been used in the food, feed, and agriculture industries for over

More information

Analytical Chemistry II

Analytical Chemistry II Analytical Chemistry II L4: Signal processing (selected slides) Computers in analytical chemistry Data acquisition Printing final results Data processing Data storage Graphical display https://www.creativecontrast.com/formal-revolution-of-computer.html

More information

Using the ALTA Handheld Reflectance Spectrometer

Using the ALTA Handheld Reflectance Spectrometer Using the ALTA Handheld Reflectance Spectrometer The ALTA spectrometer is an active spectrometer, which means that it emits energy and measures how much of that energy is reflected back. The instrument

More information

Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking

Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking A p p l i c a t i o n N o t e Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking Peter J. Brescia, Jr., Chris Wilson, and Peter Banks, BioTek Instruments, Inc., Winooski,

More information

Modern Instrumental Methods of Analysis Prof. J. R. Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore

Modern Instrumental Methods of Analysis Prof. J. R. Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore Modern Instrumental Methods of Analysis Prof J R Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore Module No # 02 Lecture No # 07 Ultraviolet and Visible Spectrophotometry

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

Introduction. The amount of radiation absorbed may be measured in a number of ways: Transmittance, T = P / P 0 % Transmittance, %T = 100 T

Introduction. The amount of radiation absorbed may be measured in a number of ways: Transmittance, T = P / P 0 % Transmittance, %T = 100 T Introduction Many compounds absorb ultraviolet (UV) or visible (Vis.) light. The diagram below shows a beam of monochromatic radiation of radiant power P 0, directed at a sample solution. Absorption takes

More information

S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY

S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY PURPOSE 1. To become familiar with the operation of a conventional scanning ultra-violet spectrophotometer 2. To determine suitable cells and solvents

More information

Questions on Instrumental Methods of Analysis

Questions on Instrumental Methods of Analysis Questions on Instrumental Methods of Analysis 1. Which one of the following techniques can be used for the detection in a liquid chromatograph? a. Ultraviolet absorbance or refractive index measurement.

More information

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision Clinical Chemistry (CHE221) Professor Hicks Week 1 Statistics Made Slightly Less Boring and Introduction to Spectrophotometry 3 Accuracy vs Precision Precision is the consistency of a measurement made

More information

February 8, 2018 Chemistry 328N

February 8, 2018 Chemistry 328N Lecture 7 UV-Vis spectroscopy February 8, 2018 First Midterm Exam When: Wednesday, 2/14 When: 7-9 PM (please do not be late) Where: WEL 2.122 This room!!! What: Covers material through today s lecture

More information

Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics

Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics CHIRASCAN SERIES APPLICATION NOTE Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics Abstract: The selection of buffer conditions that maximise

More information

CHEMICAL KINETICS E + 2B 2C + D (1)

CHEMICAL KINETICS E + 2B 2C + D (1) CHEMICAL KINETICS Chemical kinetics is the branch of chemistry that is concerned with the study of the rates and mechanisms of chemical reactions. The rate of a reaction is a measure of its speed. Consider

More information

Absorption spectrometry summary

Absorption spectrometry summary Absorption spectrometry summary Rehearsal: Properties of light (electromagnetic radiation), dual nature light matter interactions (reflection, transmission, absorption, scattering) Absorption phenomena,

More information

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY Introduction The purpose of this experiment is to determine the concentration of zinc in a sample of hair. You will use both the calibration

More information

Chapter 5 Electrons In Atoms

Chapter 5 Electrons In Atoms Chapter 5 Electrons In Atoms 5.1 Revising the Atomic Model 5.2 Electron Arrangement in Atoms 5.3 Atomic Emission Spectra and the Quantum Mechanical Model 1 Copyright Pearson Education, Inc., or its affiliates.

More information

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber CH217 Fundamentals of Analytical Chemistry Module Leader: Dr. Alison Willows Electromagnetic spectrum Properties of electromagnetic radiation Many properties of electromagnetic radiation can be described

More information

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation. NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation. SCANNING UV VISIBLE SPECTROPHOTOMETER Features SCANNING UV VISIBLE SPECTROPHOTOMETER

More information

Emission of Light & Atomic Models 1

Emission of Light & Atomic Models 1 Emission of Light & Atomic Models 1 Objective At the end of this activity you should be able to: o Explain what photons are, and be able to calculate their energies given either their frequency or wavelength.

More information

Instrumental Analysis: Spectrophotometric Methods

Instrumental Analysis: Spectrophotometric Methods Instrumental Analysis: Spectrophotometric Methods 2007 By the end of this part of the course, you should be able to: Understand interaction between light and matter (absorbance, excitation, emission, luminescence,fluorescence,

More information

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction application Note Atomic Absorption Authors David Bass Senior Product Specialist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Toxic, Trace, and

More information

Spectroelectrochemistry Part 2: Experiments and Data evaluation

Spectroelectrochemistry Part 2: Experiments and Data evaluation Spectroelectrochemistry Part 2: Experiments and Data evaluation Purpose of This Note This application note is part 2 of a series of notes introducing Gamry s spectroelectrochemistry system. Part 1 discusses

More information

Basic Instrumentation. Learning Objectives:

Basic Instrumentation. Learning Objectives: Proteomics Basic Instrumentation Basic Instrumentation Handling instruments plays an important role in laboratory working on a daily basis. The use of most instruments is almost inevitable and hence a

More information

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Phys316 Exploration 2: Verifying Stefan-Boltzmann Relationship Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law Where A is the effective radiating area,

More information

Investigating Transition Metal Complexes

Investigating Transition Metal Complexes Exercise 4 Investigating Transition Metal Complexes 4 Introduction Colour is a well known property of the transition metals. The colour produced as parts of the visible spectrum are due to electron transitions

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 31 August 2014 Print version (Harris, Chapt. 1) (pp.1-20) CEE 772: Instrumental Methods in Environmental Analysis Lecture #3 Statistics: Detection Limits Spectroscopy: Beer s Law & Electronic

More information

Chemical Kinetics I, A Survey:

Chemical Kinetics I, A Survey: Chemistry 212 Lab Fall 2011 Chemical Kinetics I, A Survey: I. Graphical Determination of Reaction Order for Crystal Violet Dye II. Determination of the Complete Rate Law Purpose For a simple reaction system

More information

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology Spectrophotometry Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology Content Introduction Beer-Lambert law Instrument Applications Introduction 3 Body fluids such as blood, csf and urine contain organic and

More information

Singlet. Fluorescence Spectroscopy * LUMO

Singlet. Fluorescence Spectroscopy * LUMO Fluorescence Spectroscopy Light can be absorbed and re-emitted by matter luminescence (photo-luminescence). There are two types of luminescence, in this discussion: fluorescence and phosphorescence. A

More information

Characterization of the VIIRS Blackbody Emittance

Characterization of the VIIRS Blackbody Emittance Characterization of the VIIRS Blackbody Emittance Jeremy Kloepfer, Chris Taylor, and Vijay Murgai jeremiah_kloepfer@raytheon.com Conference on Characterization and Radiometric Calibration for Remote Sensing,

More information

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES Chemistry 524--Final Exam--Keiderling Dec. 12, 2002 --4-8 pm -- 238 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted plus one 8.5 x 11 sheet

More information

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy molecular spectroscopy Introduction to Fourier Transform Infrared Spectroscopy Part of Thermo Fisher Scientific Introduction What is FT-IR? FT-IR stands for Fourier Transform InfraRed, the preferred method

More information

Spectrometer User s Guide

Spectrometer User s Guide Spectrometer User s Guide (Order Codes: V-SPEC, SPRT-VIS, SP-VIS, SP-UV-VIS, ESRT-VIS) The spectrometer is a portable light spectrophotometer, combining a spectrometer and a light source/cuvette holder.

More information

1901 Application of Spectrophotometry

1901 Application of Spectrophotometry 1901 Application of Spectrophotometry Chemical Analysis Problem: 1 Application of Spectroscopy Organic Compounds Organic compounds with single bonds absorb in the UV region because electrons from single

More information

An Investigative Analysis of Soil Quality in the United Kingdom Using the TIBCO Spotfire Software for Inorganic Soil Dashboard

An Investigative Analysis of Soil Quality in the United Kingdom Using the TIBCO Spotfire Software for Inorganic Soil Dashboard A P P L I C AT I O N N O T E Analytics Authors: Kathryn Kuhr Robert Packer PerkinElmer, Inc. Shelton, CT An Investigative Analysis of Soil Quality in the United Kingdom Using the TIBCO Spotfire Software

More information

Chapter 18. Fundamentals of Spectrophotometry. Properties of Light

Chapter 18. Fundamentals of Spectrophotometry. Properties of Light Chapter 18 Fundamentals of Spectrophotometry Properties of Light Electromagnetic Radiation energy radiated in the form of a WAVE caused by an electric field interacting with a magnetic field result of

More information

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at 2001 Spectrometers Instrument Machinery Movies from this presentation can be access at http://www.shsu.edu/~chm_tgc/sounds/sound.html Chp20: 1 Optical Instruments Instrument Components Components of various

More information

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Electromagnetic Spectrum - Molecular transitions Widely used in chemistry. Perhaps the most widely used in Biological Chemistry.

More information

Spectroscopy: Introduction. Required reading Chapter 18 (pages ) Chapter 20 (pages )

Spectroscopy: Introduction. Required reading Chapter 18 (pages ) Chapter 20 (pages ) Spectroscopy: Introduction Required reading Chapter 18 (pages 378-397) Chapter 20 (pages 424-449) Spectrophotometry is any procedure that uses light to measure chemical concentrations Properties of Light

More information

Lab Investigation 4 - How could you make more of this dye?

Lab Investigation 4 - How could you make more of this dye? Lab Investigation 4 - How could you make more of this dye? USING SPECTROSCOPY TO DETERMINE SOLUTION CON- CENTRATION Guiding Question How could you make more of this dye? INTRODUCTION A solution is a homogeneous

More information

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur 603 203 DEPARTMENT OF ELECTRONICS AND INSTRUMENTATION ENGINEERING QUESTION BANK V SEMESTER EI6501 Analytical Instruments Regulation 2013 Academic

More information

Lab VI Light Emitting Diodes ECE 476

Lab VI Light Emitting Diodes ECE 476 Lab VI Light Emitting Diodes ECE 476 I. Purpose This experiment examines the properties of light emitting diodes. The use of the monochromator for studying the spectrum of light sources is also examined.

More information

Basic Calibration of UV/ Visible Spectrophotometer

Basic Calibration of UV/ Visible Spectrophotometer International Journal of Science and Technology Volume 2 No. 3, March, 2013 Basic Calibration of UV/ Visible Spectrophotometer Adeeyinwo, C.E, 1 Okorie, N. N. 2, Idowu, G. O. 3 1,2 Department of Chemistry,

More information

UV-Vis Absorption Experiment 5: Beer- Lambert Law and the Temperature Dependence of the Crystal Violet- Sodium Hydroxide Reaction

UV-Vis Absorption Experiment 5: Beer- Lambert Law and the Temperature Dependence of the Crystal Violet- Sodium Hydroxide Reaction 1 UV-Vis Absorption Experiment 5: Beer- Lambert Law and the Temperature Dependence of the Crystal Violet- Sodium Hydroxide Reaction Overview In Part A of this experiment, the absorption behaviour of crystal

More information

Recommended tools and devices: spectrophotometer Spekol, cuvettes, set of chromatic solutions varying in concentration, distilled water

Recommended tools and devices: spectrophotometer Spekol, cuvettes, set of chromatic solutions varying in concentration, distilled water 17_/Work procedure 1/5 Study group: Year of study: Objective: Based on construction of calibration curves for absorbance and transmittance establish the unknown concentration of a solution using absorption

More information

Characterisation & Use of Array Spectrometers

Characterisation & Use of Array Spectrometers Characterisation & Use of Array Spectrometers Mike Shaw, Optical Technologies & Scientific Computing Team, National Physical Laboratory, Teddington Middlesex, UK 1 Overview Basic design and features of

More information

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Spectroscopy. Page 1 of 8 L.Pillay (2012) Spectroscopy Electromagnetic radiation is widely used in analytical chemistry. The identification and quantification of samples using electromagnetic radiation (light) is called spectroscopy. Light has

More information

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs Application Note Author Harry Prest, Ph.D. Senior Chemist in GC-MS Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara,

More information

Spectral Calibration of Ultra- High Resolution Volume Holographic Spectrometer. Jeff Bourne Majid Badiei (Advisor)

Spectral Calibration of Ultra- High Resolution Volume Holographic Spectrometer. Jeff Bourne Majid Badiei (Advisor) Spectral Calibration of Ultra- High Resolution Volume Holographic Spectrometer Jeff Bourne Majid Badiei (Advisor) What is a spectrometer? Any device that converts different input wavelengths into different

More information

Introduction to FT-IR Spectroscopy

Introduction to FT-IR Spectroscopy Introduction to FT-IR Spectroscopy An FT-IR Spectrometer is an instrument which acquires broadband NIR to FIR spectra. Unlike a dispersive instrument, i.e. grating monochromator or spectrograph, an FT-IR

More information

Chapter 4 Spectroscopy

Chapter 4 Spectroscopy Chapter 4 Spectroscopy The beautiful visible spectrum of the star Procyon is shown here from red to blue, interrupted by hundreds of dark lines caused by the absorption of light in the hot star s cooler

More information