1 Calibration Traceability Guidelines Check the website of accrediting bodies for the accredited laboratories individual scope of accreditation ts.nist.gov/standards/accreditation/index.cfm The 1 st page of any ISO is useless after it has been determined that the laboratory is accredited. 1

2 Calibration Traceability Guidelines It is the scope of accreditation that determines the laboratory s capability. The cert should state the best measurement uncertainty for different ranges. On this cert the best measurement uncertainty is % for torque calibrations from N-m and % for force calibrations up to 120,000 LBF. 2

3 7 Essential Elements Chain of Measurement Traceability Measurement Uncertainty Measurement Assurance Realization of SI Units Periodic Recalibration Technical Competence 3

4 Measurement Calibration Hierarchy Reference Standard used in the calibration of equipment BIPM/SI National Metrology Institute (NMI) Primary Reference Laboratory Morehouse Instrument Company Accredited Calibration Service Supplier Instrument/Equipment Measurement Process: A. Reference Standard B. Measurand C. Measurement Result D. Measurement Traceability: 1. Unbroken Chain of Comparisons 2. Measurement Uncertainty 3. Documented Procedure 4. Technical Competence 5. Realization of SI Units 6. Periodic Recalibration 7. Measurement Assurance 4

5 Measurement Traceability Requirements ISO/IEC 17025, Clause Specific requirements Calibration laboratories When using external calibration services, traceability of measurement shall be assured by the use of calibration services from laboratories that can demonstrate competence, measurement capability and traceability... 5

6 Measurement Calibration Hierarchy Reference Standard used in the calibration of equipment Accredited Calibration Service Supplier Laboratory Instrument/Equipment Supplier Evaluation Measurement Process: A. Reference Standard B. Measurand C. Measurement Result D. Measurement Traceability: 1. Unbroken Chain of Comparisons 2. Measurement Uncertainty 3. Documented Procedure 4. Technical Competence 5. Realization of SI Units 6. Periodic Recalibration 7. Measurement Assurance Laboratory Measurement 6

7 Measurement Calibration Hierarchy Reference Standard Accredited Calibration Service Supplier Supplier Evaluation Measurement Process: A. Reference Standard B. Measurand C. Measurement Result D. Measurement Traceability: 1. Unbroken Chain of Comparisons 2. Measurement Uncertainty 3. Documented Procedure 4. Technical Competence 5. Realization of SI Units 6. Periodic Recalibration 7. Measurement Assurance Laboratory use of a calibrated Reference Standard 7

8 Measurement Uncertainty & the Measurement Hierarchy BIPM/SI Uncertainty National Metrology Institute (NMI) Primary Reference Laboratory Morehouse Instrument Company Accredited Calibration Service Supplier Working Standards Instrument/Equipment Measurement Traceability: 1. Unbroken Chain of Comparisons 2. Measurement Uncertainty 3. Documented Procedure 4. Technical Competence 5. Realization of SI Units 6. Periodic Recalibration 7. Measurement Assurance 8

9 Measurement Uncertainty BIPM/SI NMI Primary Standards Accredited Cal. Lab Working Standards Field Measurement All measurements are subject to uncertainty and a measured value is only complete if it is accompanied by a statement of the associated uncertainty. Uncertainty is a measure of how close a particular test result, the product of one laboratory, is to the true value. 9

10 Measurement Uncertainty 10,000 LBF LOAD CELL TYPE A UNCERTAINTY.5 LBF NMI Primary Standards Accredited Cal. Lab Working Standards Field Measurement NIST ( %) TOTAL TEST UNCERTAINTY LBF MOREHOUSE (0.001 %) TOTAL TEST UNCERTAINTY LBF SECONDARY STANDARDS (0.02 %) TOTAL TEST UNCERTAINTY LBF WORKING STANDARDS (0.1 %) TOTAL TEST UNCERTAINTY LBF FIELD MEASUREMENT (1 %) TOTAL TEST UNCERTAINTY LBF 10

11 Measurement Uncertainty BIPM/SI NMI Primary Standards Accredited Cal. Lab Working Standards Field Measurement The further away from calibration by primary standards the larger the Overall Uncertainty will become 11

12 Documented Procedure Documented Procedure Written description of a prescribed course of action or process. Per ANSI/ISO/IEC The laboratory shall use the appropriate methods and procedures for all tests and/or calibrations within its scope. 12

13 Measurement Assurance Measurement Assurance -Practices put in place to monitor a testing or calibration process and to ensure the calibration status of equipment, reference standards or reference materials used in a measurement process. 13

14 Measurement Assurance Measurement Assurance is essential to holding together the entire chain of traceability as its purpose is to prevent incorrect results from being reported 14

15 Measurement Assurance Monitoring the measurement process Monitoring the: Calibration status of calibrated equipment Calibration status of calibrated Reference Standard(s) Calibration status of Reference Material(s) A combination of monitoring both the measurement process and/or the equipment or the reference(s) A means of strengthening measurement traceability A means of verifying the estimated measurement uncertainty 15

16 Measurement Assurance ISO/IEC 17025, Section 5 Technical Requirements When intermediate checks are needed to maintain the confidence in the calibration status of the equipment, these checks shall be carried out according to a defined procedure. 16

17 Measurement Assurance ISO/IEC 17025, Section 5 Technical Requirements Intermediate checks Checks needed to maintain confidence in the calibration status of reference, primary, transfer, or working standards and reference materials shall be carried out according to defined procedures and schedules. 17

18 Measurement Assurance ISO/IEC 17025, 5 Technical The laboratory shall have quality control procedures for monitoring the validity of tests and calibrations undertaken. The resulting data shall be recorded in such a way that trends are detectable and, where practicable, statistical techniques shall be applied to the reviewing of the results. This monitoring shall be planned and reviewed and may include, but not be limited to, the following: 18

19 Measurement Assurance ISO/IEC 17025, 5 Technical Quality control data shall be analyzed and, where they are found to be outside pre-defined criteria, planned action shall be taken to correct the problem and to prevent incorrect results from being reported. 19

20 Technical Competence Technical Competence - is the ability of a person to do their job comparatively and on the same level as other professionals in their field. Per ANSI/ISO/IEC The laboratory shall ensure the competence of all who operate specific equipment, perform tests and or calibrations, evaluate results, and sign test reports and calibration certificates. 20

21 Demonstrating Technical Competence 21 Proficiency testing determines the performance of individual laboratories for specific tests or measurements and is used to monitor laboratories continuing performance

22 Realization of SI Units Realization of SI Units Force is derived from SI units of Mass; length; and time. CIPM/BIPM defines 1N as the force required to accelerate one kg to one meter per second per second in a vacuum 22

23 Periodic Recalibration Per ANSI/NSCL Z540.3 Calibration intervals and/or process controls shall be established for measuring and test equipment that are included in the calibration system to monitor and maintain equipment performance to the stated application requirements. 23

24 Calibration Traceability The Torque Machine at Morehouse belonged to a NMI (National Physical Laboratory) Morehouse uses dead weight primary standards up to 120,000 LBF accurate to.002 % of applied force. 24

25 NIST 8-Step Process for Estimating and Reporting Measurement Uncertainty 1: Specify the measurement process 2: Identify uncertainty components 3: Quantify uncertainty components 4: Convert quantities to standard uncertainties 5: Calculate combined standard uncertainty 6: Expand the combined standard uncertainty by coverage factor (k) 7: Evaluate the expanded uncertainty 8: Report the uncertainty 25

26 Calibration Defined Calibration is the comparison of an unknown (typically referred to as the Unit Under Test or UUT) to a device known within a certain error(typically referred to as the Calibration Standard or Reference Standard) for the purpose of characterizing the unknown Operation that, under specified conditions, in a first step, establishes a relation between the quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication. Calibration Standards in regards to ASTM E74 are typically characterized as either Primary or Secondary Standards 26

27 Primary Force Standard (as defined by ASTM E74-13) Primary Force Standard a deadweight force applied directly without intervening mechanisms such as levers, hydraulic multipliers, or the like, whose mass has been determined by comparison with reference standards traceable to national standards of mass To be a classified as a primary standard the masses of the weights shall be determined within % of their values by comparison with reference standards traceable to national standards of mass 27

28 Primary Force Standard (as defined by ASTM E74-13) Require correction for the effects of Local Gravity Air Buoyancy Must be adjusted to within % or better (N.I.S.T weights are adjusted to within U = %, Morehouse U= %) Made from stable materials with good surface finish (Stainless Steel preferred material) 28

29 Secondary Force Standard (as defined by ASTM E74-13) Secondary Force Standard an instrument or mechanism, the calibration of which has been established by comparison with primary force standards. In order to perform calibrations in accordance with ASTM E74 your force standard must be calibrated with primary standards 29

30 Secondary Force Standard (as defined by ASTM E74-13) Secondary Force Standard Range of use limited by loading ranges established by the standard ASTM E74 Class AA Load Range for calibration of secondary standard load cells. This is found by multiplying the lower limit factor by 2000 (.05 %) 5:1 ratio ASTM E74 Class A Load Range for calibration of testing machine. This is found by multiplying the lower limit factor by 400 (.25 %) 4:1 ratio. Range of use cannot be less than the lowest applied force. Loading range cannot be less than 400 for Class A or 2000 for Class AA times the resolution. 30

31 Calibration Preparation - Stabilization Temperature Stabilization It is recommended that a device be kept in the area or lab where it is to be calibrated for the device to stabilize in the environment. A good rule of thumb is to allow 24 hours for temperature stabilization. Recommended Temperature is 23 degrees C Electrical Stabilization Depending on the equipment common practice is to allow minutes to warm up. Exercise the instrument to be calibrated. The instrument should be set up in the machine and exercised to the maximum force that is to be applied during the actual calibration. Typically we recommend 3-4 exercise cycles; most standards require a minimum of 2 exercise cycles. 31

32 Calibration In Accordance with ASTM E74 At least 30 force applications are required (we typically recommend 3 runs of 11 or 33 force applications) At least 10 must be different forces and each force must be applied at least twice. Either 15 forces applied twice for 30 force applications or 11 forces applied 3 times for 33 force applications. There should be at least one calibration force for each 10% interval throughout the loading range and if the instrument is to be used below 10% of its capacity a low force should be applied. This low force must be greater than the resolution of the device multiplied by 400 for Class A or 2000 for Class AA devices 32

33 Calibration Temperature ASTM E74 requires that the temperature be monitored during calibration as close to the device as possible and that the temperature change not exceed +/- 1 degree C during calibration. Temperature corrections must be applied to non-compensated devices. Deflection generally increases by % for each 1 degree C increase in temperature. If the calibration laboratory is not operating at 23 degrees C they should make corrections by correcting the applied force accordingly. 33

34 Calibration In Accordance with ASTM E74 Randomization of Loading Conditions Shift or rotate the UUT in the calibration machine before repeating any series of forces (suggestion is to rotate 0, 120 and 240 degrees) For Tension and Compression calibration, intersperse the loadings. Be sure to re-exercise the UUT prior to any change in setup. Zero Return during calibration - This is lab-dependent and it is recommended that no more than 5 forces be applied before return to zero. 34

35 Calibration In Accordance with ASTM E74 Deflection calculation Methods Method B Deflection readings should be calculated as the difference between readings at the applied force and the average or interpolated zero force readings before and after the applied force readings. Method A Deflection readings are calculated as the difference between the deflection at the applied force and the initial deflection at zero force. 35

36 Number of Calibration Values 30 +points reduces standard measurement error T distribution = sample vs mean, Z = population 36

37 Calibration In Accordance with ASTM E74 LOAD REVERSAL OR DESCENDING LOADING If a force measuring device is to be used to measure forces during decreasing load sequences, then it must be calibrated in this manner. Separate calibration curves can be used for Ascending values and Descending Values A combined curve may also be used though the STD DEV of the combined curve will be much higher than using separate curves. 37

38 Calibration In Accordance with ASTM E74 Criteria for Use of Higher Degree Curve Fits Resolution must exceed 50,000 counts An F distribution test is used to determine the appropriate best degree of fit (instructions for this test can be found in the Annex A1 of the ASTM E74 Standard) The Standard deviation for the established curve fit is calculated as before using all the individual deflection values 38

39 Calibration In Accordance with ASTM E74 Criteria for Lower Load Limit Uncertainty = 2.4 * STD DEV This corresponds to a 99 % Coverage Factor Based on Uncertainty or Resolution whichever is higher Class A 400 times the uncertainty or resolution Class AA 2000 times the uncertainty or resolution NOTE: Any instrument that is either modified or repaired should be recalibrated Recalibration is required for a permanent zero shift exceeding 1.0 % of full scale 39

40 Calibration In Accordance with ASTM E74 Recalibration Interval Secondary Standards should be calibrated or verified annually to ensure that they do not change more than.032 % over the loading range Instruments used as Class A devices (Typically used to calibrate testing machines) should be calibrated or verified annually to ensure that they do not change more than.16 % over the loading range. If the Calibration device is stable to within.16 % over the loading range then the calibration interval can be 2 years as long as the UUT continues to meet the stability criteria 40

41 ASTM E74 Calibration Data Analysis Deviations from the fitted curve These are the differences between the fitted curve and the observed values Standard Deviation is the square root of the sum of all the deviations squared/n-m-1 N = sample size, m = the degree of polynomial fit Calibration equation Deflection or Response = A0+A1(load)+A2(load)^2+ A5(load)^5 LLF is 2.4 times the standard deviation Class A range is 400 times the LLF. Class AA range is 2000 times the LLF. 41

42 Calibration In Accordance with ASTM E74 Substitution of Electronic Instruments The indicating device used in the original calibration and the device to be substituted shall have been calibrated and the measurement uncertainty determined The uncertainty of each device shall be less than 1/3 of the uncertainty for the force measurement system. Excitation amplitude, wave form, and frequency shall be maintained Cable substitutions should be verified with a transducer simulator 42

43 Calibration In Accordance with ASTM E74 Allow UUT to come to room temperature Warm up Instrumentation Select Test points Fixture UUT in Test Frame Exercise UUT 2-4 times Apply 1 st series of forces (Run1) Rotate the UUT 120 degrees if possible for run 2 Apply 2 nd series of forces (Run2) IF UUT IS COMPRESSION AND TENSION SWITCH TO OTHER MODE AFTER FINISHING RUN 2 AND EXERCISE AND REPEAT ABOVE STEPS Rotate the UUT another 120 degrees if possible for run 3 Apply 3rd series of forces (Run3) 43

44 Accuracy and Precision It is a common mistake to assume that an accurate device is precise or that a precise device is accurate. 44

45 Accuracy Accuracy is the function of bias and precision (estimate of a systematic measurement error) The Qualitative term for the extent of the approximation of the measurement result to a true value (DIN55350) Accuracy does not tell you about the quality of the instrument. 45

46 Bias Measurement bias (bias): estimate of a systematic measurement error (VIM 2.18) Systematic Measurement Error: component of measurement error that in replicate measurements remains constant or varies in a predictable manner (VIM 2.17) Measurement Error: measured quantity value minus a reference quantity value (VIM 2.16) 46

47 Bias 47

48 Bias one part of Accuracy Accuracy - Influenced by both the precision and bias in a measurement process Accuracy: Closeness of agreement between a measured quantity value and a true quantity value of a measurand. (VIM 2.13) Between an individual measurement and the true value Defined with reference to a true value which is unknown Qualitative term If using a conventional true value, the accuracy can be known 48

49 Precision Precision is a measure of spread Precision refers to the repeatability of measurement. 49

50 Precision Example Example: 500 LBF was applied 3 times using a dead weight primary standard to a load cell and the load cell indicator s recorded output was LB, LB, LB. Conclusion: The instrument is precise to +/-.01 LB when 500 LBF is applied. 50

51 Bias and Precision A load cell may not give you the same measurement result at different orientations. A load cell may repeat when a force is applied, but may not read close to the desired engineering unit s value. 51

52 Accuracy Example Example: A Force Gauge was calibrated against a dead weight primary standard and at LBF the recorded output on the indicator was LBF. This measurement was repeated and LBF was observed a second time. The instrument was re positioned and LBF was observed a third time. This force gauge was determined to be accurate to +.1 % of full scale or +.5 LBF. Note: From this example the resolution of the instrument may be.1 LBF or.5 LBF and although the resolution may be the same as the accuracy in this example they are different. 52

53 Resolution Resolution - the smallest amount of input signal change that the instrument can detect reliably Resolution - is the ability of the measurement system to detect and faithfully indicate small changes in the characteristic of the measurement result. 53

54 Uncertainty of Force Standards Regardless of the force facility to be used, it is important to evaluate the uncertainty of the system. This should include contributions from all influencing parameters (e.g. mass, alignment, and environmental factors). The factors or influences to be reflected in calculation of the uncertainty differ between standards as well as processes. 54

55 Accuracy vs. Uncertainty Accuracy determined via a calibration is not the same as uncertainty! an accurate measurement with a large uncertainty is possible. 55

56 A2LA Policy R205 A2LA POLICY R205: SPECIFIC REQUIREMENTS: CALIBRATION LABORATORY ACCREDITATION A2LA Policy R205 states Every measurement uncertainty shall take into consideration the following standard contributors, even in the cases where they are determined to be insignificant, and documentation of the consideration shall be made: 56

57 These uncertainty contributors are : Repeatability (Type A) Resolution Reproducibility Reference Standard Uncertainty Reference Standard Stability Environmental Factors 57

58 Uncertainty a measurement is not very meaningful unless there is some way of estimating the associated uncertainty good analyses should include uncertainty estimates 58

59 Uncertainty Per section of the GUM The word uncertainty means doubt, and thus in the broadest sense Uncertainty of measurement means doubt the validity of the result of a measurement. 59

60 Uncertainty Uncertainty is a measure of how close a particular test result, the product of one laboratory, is to the true value. There is no assurance that any laboratory using the same test method will have the same accuracy; some will be better and some worse. Specifications of apparatus and materials in test methods attempt to control uncertainty but cannot guarantee a value. 60

61 Uncertainty In Metrology, two common types of uncertainty evaluations are Type A and Type B. 61

62 Uncertainty Type A Uncertainty - The method of evaluation of uncertainty by the statistical analysis of a series of observations. Repeatability condition of measurement, Intermediate precision condition of measurement, Reproducibility condition of measurement 62

63 Uncertainty ASTM E2428 (TORQUE) and ASTME74 (FORCE) calibration test for the reproducibility and repeatability condition of measurement and is an example of Type A Uncertainties. The term used in these standards is Lower Limit Factor which applies a coverage factor of 2.4 If the equipment used to perform the test has a relatively low overall uncertainty then a large percentage of the TTU (Total Test Uncertainty) will be quantified with reproducibility and repeatability 63

64 Uncertainty Type A Example A series of measurements are taken to determine the Type A uncertainty of the measurement

65 Uncertainty Type B Uncertainty - method of evaluation of uncertainty by means other than the statistical analysis of a series of observations. Type B uncertainty - Evaluation based on information associated with a quantity value of a certified reference material, - obtained from a calibration certificate or manufacture s specifications, obtained from the accuracy class of a verified measuring instrument, obtained from limits deduced from a test or experiment. Examples include torque or load cell temperature effect, drift, resolution, etc. 65

66 Uncertainty Type B evaluation method: The method of evaluation of uncertainty of measurement by means other than statistical analysis of a series of observations. Examples: Based on specification History of parameter Other knowledge or test of the process parameter 66

67 Uncertainty Type B Examples The temperature effect on force or torque cell output is +/-.004 % per degrees Celsius The specification of the torque arm is +/ inches Bending, cross-force, cosine error, etc. 67

68 Uncertainty So if we are dealing with A component of Uncertainty it is either Type A or Type B, If we have both Type A and Type B uncertainties and RSS (root sum square) then this becomes a Combined Standard Uncertainty and if we apply a coverage factor to the Combined Uncertainty this becomes our Overall Uncertainty also known as Expanded Uncertainty. We will go over an expanded uncertainty analysis in the next section. 68

69 Uncertainty Determining the Uncertainty of a Measurement (UOM) is different from the practice of Determining the Expected Performance of a Device. Determining the Expected Performance of a Device (which includes the De-Rating Specifications associated with its performance and which are NOT part of the Measurement Uncertainty Analysis performed/calculated to determine the UOM any more than the Base EP (Expected Performance) of the device is included in the UOM). In other words the manufacturer s specification is not to be used in place of the uncertainty. 69

70 Uncertainty Example The next example will deal with an example derived from an uncertainty analysis that is in the ASTM E74-13a standard as an appendix. Repeatability, Reproducibility, and Resolution are all accounted for in the ASTM E74 uncertainty or LLF (Lower Limit Factor). 70

71 Uncertainty Example We will gather some necessary information and run through a sample expanded uncertainty calibration. We will need the following: 1. Calibration Report for the Device 2. The uncertainty of the instrument(s) that were used to perform the calibration 3. Calibration History (if available) 4. Manufacturer s Specification Sheet 5. Error Sources, if known 6. Dissemination Error, if known 71

72 Performance & Uncertainty for Calibration performed by Morehouse Type A Uncertainty To do a type A uncertainty analysis, information will be needed from the calibration report. The ASTM E74 LLF or Uncertainty from the report should be entered into the spreadsheet. In ASTM E74-13a, Uncertainty has been changed to LLF (Lower Limit Factor). 72

73 Uncertainty Example Type A and B uncertainty analysis COMPANY LOAD CELL MANUFACTURER SAMPLE MOREHOUSE ENTER LOAD CELL S/N P-7768 YOUR CAPACITY LBF CALIBRATION ASTM E74 Uncertainty for K= LBF INFORMATION THE LOWEST FORCE AT WHICH THE SECONDARY STANDARD WILL BE USED 1000 LBF IN PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K= % HIGHLIGHTED PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) COLUMNS CAL DATE 10/27/2010 Using the Excel sheet available at content/uploads/2012/05/type-a-and-b-uncertainty-analysis- ASTM-E74-1.xls We will enter the load cell S/N, Capacity, ASTM Uncertainty which was.237 LBF, Lowest force the standard will be used, and the Uncertainty of the standard used to perform the calibration at k=1. 73

74 Uncertainty Example Type A and B uncertainty analysis COMPANY LOAD CELL MANUFACTURER SAMPLE MOREHOUSE ENTER LOAD CELL S/N P-7768 YOUR CAPACITY LBF CALIBRATION ASTM E74 Uncertainty for K= LBF INFORMATION THE LOWEST FORCE AT WHICH THE SECONDARY STANDARD WILL BE USED 1000 LBF IN PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K= % HIGHLIGHTED PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) COLUMNS CAL DATE 10/27/2010 The uncertainty of the standard or standards that were used to perform the calibration should be found somewhere on the certificate of calibration with a coverage factor. This coverage factor is typically 2, so you will need to reduce this to k=1. 74

75 Uncertainty Example Type A Uncertainty % Uncertainty Description Uncertainty Distribution Divisor Standard Uncertainty Squared ASTM E74 Uncertainty % at the lowest calibration force to be used % normal E E-09 Combined Type A Uncertainty 9.88E E-09 The information from the ASTM E74 report is your Type A uncertainty. To get this, we are dividing the uncertainty or LLF by the lowest force this instrument is going to be used at. Then we divide by 2.4 the ASTM E74 coverage factor to reduce this to the uncertainty or LLF at k=1 (above). 75

76 Uncertainty Example Type B Uncertainty % Uncertainty Description Uncertainty Distribution Divisor Standard Uncertainty Squared PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY 0.001% rectangular E E-11 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) 0.000% rectangular E E+00 STABILITY OF THE SECONDARY FORCE STANDARD OVER TIME 0.005% rectangular E E-10 CREEP ERROR FOUND ON LOAD CELL SPEC SHEET 0.002% rectangular E E-11 MISALIGNMENT ERROR (SEE ASTM E1012) AND/OR SIDE LOAD SENSITIVITY FROM LOAD CELL SPEC SHEET 0.005% rectangular E E-10 DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) 0.005% rectangular E E-10 TEMPERATURE ERROR +/- FROM LOAD CELL SPEC SHEET % rectangular E E-11 Combined Type B Uncertainty 5.18E E-09 Next, we will be looking at the type B uncertainty. If the system was calibrated in accordance with ASTM E74 and consisted of a meter and load cell, then we will treat the ASTM E74 uncertainty or LLF as a system uncertainty, and there would not be a need to look at the Electrical Calibration Standard Uncertainty. 76

77 Uncertainty Example PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY 0.001% rectangular E E-11 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) 0.000% rectangular E E+00 The first type B uncertainty component to examine is the uncertainty of the standards used to perform the calibration. The can usually be found on the calibration laboratory s scope of accreditation. In this example, the load cell was sent in with an indicator, so we will only consider the Primary Force Standard Uncertainty, which was dead weight with an uncertainty of 0.001% for K=1. 77

78 Uncertainty Example STABILITY OF THE SECONDARY FORCE STANDARD OVER TIME 0.005% rectangular E E-10 This can be determined by comparing the previous calibration with the current calibration. The # to be used should be the number of the lowest calibration force that will be used for calibration. If you do not have any previous calibration data, then the suggestion would be to contact the manufacturer or use a conservative number based on similar systems. 78

79 Uncertainty Example Calibration History can be found from taking the difference from one calibration to the next, or looking at several calibrations, if available. In this example, I would opt to use the % change of %. This is the highest % change throughout the loading range. 79

80 Uncertainty Example A less conservative way to approach change from previous would be to take the Standard Deviation of all of the change from previous numbers. 80

81 UNCERTAINTY EXAMPLE CREEP ERROR FOUND ON LOAD CELL SPEC SHEET 0.002% rectangular E E-11 Creep error This can usually be found on the manufacturer s spec sheet, and is usually % reading for 20 minutes. Since we typically hold the force for around 30 seconds when performing the calibration, the creep error is much lower. If the end user replicates holding the force for 30 seconds, then the creep error of the system should be better than %. A creep test can be performed and is included in the new ASTM revision for those using method A. 81

82 Uncertainty Example MISALIGNMENT ERROR (SEE ASTM E1012) AND/OR SIDE LOAD SENSITIVITY FROM LOAD CELL SPEC SHEET 0.005% rectangular E E-10 A well aligned calibration machine may demonstrate bending less than 2 %. The % can usually be found on the load cell spec sheet under Side Load Sensitivity. Note: If using a Morehouse UCM and Morehouse Ultra Precision Load Cell the Morehouse press will transfer the force applied to the load cell at an angle of no more than 1/16 th inch measured off centerline of the load cell. (This number is usually 0.05 % *.0625 =.003%) 82

83 Uncertainty Example DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) 0.005% rectangular E E-10 Dissemination Error Assuming we have compared results with primary standards accurate to % of applied force, and we achieved actual measurement results comparing 2 standards, each calibrated with primary standards against one another that suggested our measurements to be within %, we will use this number for Dissemination Error. 83

84 Uncertainty Example TEMPERATURE ERROR +/- FROM LOAD CELL SPEC SHEET % rectangular E E-11 Temperature Error - This is found from the load cell spec sheet. It is usually in terms of % of reading/100 per degree F or C. This number should then be multiplied by the maximum temperature difference from the temperature at which the calibration was performed. If the manufacturer s spec sheet suggests.0015 % per degree C and you are operating within +/- 1 degree, then use this number. If you are +/- 2 Degrees C, then use.003 %. 84

85 Type A and B uncertainty analysis COMPANY LOAD CELL MANUFACTURER SAMPLE MOREHOUSE ENTER LOAD CELL S/N P-7768 YOUR CAPACITY LBF CALIBRATION ASTM E74 Uncertainty for K= LBF INFORMATION THE LOWEST FORCE AT WHICH THE SECONDARY STANDARD WILL BE USED 1000 LBF IN PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K= % HIGHLIGHTED PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) COLUMNS CAL DATE 10/27/2010 CALCULATED VALUES LOAD CELL UNCERTAINTY IN % FOR FULL SCALE K= % LOAD CELL UNCERTAINTY IN % FOR FULL SCALE K= % LOAD CELL UNCERTAINTY IN % FOR LOWEST FORCE APPLIED K = % LOAD CELL UNCERTAINTY IN % FOR LOWEST FORCE APPLIED K = % Type A Uncertainty % Uncertainty Description Uncertainty Distribution Divisor Standard Uncertainty Squared ASTM E74 Uncertainty % at the lowest calibration force to be used % normal E E-09 Combined Type A Uncertainty 9.88E E-09 Type B Uncertainty % Uncertainty Description Uncertainty Distribution Divisor Standard Uncertainty Squared PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY 0.001% rectangular E E-11 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) 0.000% rectangular E E+00 STABILITY OF THE SECONDARY FORCE STANDARD OVER TIME 0.005% rectangular E E-10 CREEP ERROR FOUND ON LOAD CELL SPEC SHEET 0.002% rectangular E E-11 MISALIGNMENT ERROR (SEE ASTM E1012) AND/OR SIDE LOAD SENSITIVITY FROM LOAD CELL SPEC SHEET 0.005% rectangular E E-10 DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) 0.005% rectangular E E-10 TEMPERATURE ERROR +/- FROM LOAD CELL SPEC SHEET % rectangular E E-11 Combined Type B Uncertainty 5.18E E-09 Review of everything we entered 85

86 Uncertainty Example FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B % FOR K=2 U = K * UC (UNCERTAINTY % AT THE LOWEST FORCE TO BE APPLIED) 0.022% LBF Summary UC AT CAPACITY * 2 ( (UNCERTAINTY % AT INSTRUMENT CAPACITY) 0.011% LBF After all the data has been entered, the Big U or Expanded Uncertainty for this 10K load cell that had an ASTM E74 uncertainty or LLF of.237 LBF (K=2.4) is now LBF for K=2 at full scale capacity. 86

87 Uncertainty Example This example is just a guideline for calculating expanded uncertainty. The actual uncertainty components in your system may vary. In addition to this example, there is also an Uncertainty example for torque in the next section. 87

88 Uncertainty Example Using handouts and the material provided in class we will fill out an excel spreadsheet. 88

89 Uncertainty -SINGLE RUN CALIBRATION 89

90 Uncertainty -SINGLE RUN CALIBRATION APPLIED LBF FORCE MEASURED LBF INSTRUMENT ERROR in LBF Calculated Residual % Error % % % % % % % % % % Uncertainty %

91 Types of Load Cells Column Load Cell (Single-Column or High- Stress Load Cells) Multi-Column Load Cell S-Beam or S-Type Shear Web 91

92 Column Load Cell The spring element is intended for axial loading, and typically has a minimum of four strain gauges, two in the longitudinal direction, and two oriented transversally to sense the Poisson strain. 92

93 Column Load Cell Column load cells have a reputation for inherent non-linearity. This deviation from linear behavior is commonly ascribed to the change in the crosssectional area of the column (due to Poisson s ratio), which occurs with deformation under load. Column cells also exhibit a larger creep characteristic than other cells and often do not return to zero as well as other cells. The advantage of a column cell is that at higher capacities, its physical size and weight will make the cell more portable. It is not uncommon to have a 1,000,000 LBF column cell weigh less than 100 lbs. 93

94 Multi - Column Load Cell In this type of design, the load is carried by four or more small columns, each with its own complement of strain gauges. The corresponding gauges from all of the columns are connected in a series in the appropriate bridge arms

95 Multi - Column Load Cell 95

96 Multi - Column Load Cell Multi-Column load cells can be more compact than high-stress column cells and offer improved discrimination against the effects of off-axis load components. These cells typically have less creep and have better zero returns than single-column cells. In many cases, a properly designed shear-web spring element can offer greater output, better linearity, lower hysteresis, and faster response. 96

97 S-beam Load Cell This type of design is often used in weighing applications. There are four gauges placed inside the beam. In general, linearity will be enhanced by minimizing the ratio of deflection (at rated load) to the length of the sensing beam, thus minimizing the change in shape of the element. These cells are very sensitive to off-axis loading and ideally suited for scales or tension applications. Ideal for measuring small forces (under 50 LBF) when physical weights cannot be used. 97

98 Shear Web Load Cell This type of load cell is typically the most accurate when installed on a tapered base with an integral threaded rod installed. These cells typically have very low creep and are not as sensitive to offaxis loading as the other cells discussed. These cells would be the recommended choice for force applications from 100 LBF through 100,000 LBF. After 100,000 LBF, the weight of the cell makes it very difficult to use as a field standard. A 100,000 LBF Shear Web cell weighs approximately 57 lbs and a 200,000 LBF shear web cell weighs over 140 lbs. 98

99 Shear Web Load Cell If these cells are used without a base or without an integral top adapter, there may be significant errors associated with various loading conditions. 99

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web:  sales: Morehouse 1 Morehouse Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com sales: edlane@mhforce.com 2 This presentation will cover the calibration

More information

MEASUREMENT TRACEABILITY AND ERRORS RELATED TO FORCE MEASUREMENT

MEASUREMENT TRACEABILITY AND ERRORS RELATED TO FORCE MEASUREMENT MEASUREMENT TRACEABILITY AND ERRORS RELATED TO FORCE MEASUREMENT Henry Zumbrun II, President Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com info: hzumbrun@mhforce.com

More information

Uncertainty Propagation for Force Calibration Systems

Uncertainty Propagation for Force Calibration Systems Uncertainty Propagation for Force Calibration Systems Morehouse Instruments Figure 1. Measurement Uncertainty Pyramid Introduction There are several labs operating throughout the world, whom do not follow

More information

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012 Revision 1.2 October 15, 2012 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability

More information

An Introduction to the Differences Between the Two Most Recognized Force Standards

An Introduction to the Differences Between the Two Most Recognized Force Standards An Introduction to the Differences Between the Two Most Recognized Force Standards Morehouse Instrument Company Introduction Morehouse has been performing both ASTM E74 and ISO 376 calibrations for more

More information

Document No: TR 12 Issue No: 1

Document No: TR 12 Issue No: 1 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Prepared by: SADCAS Technical Manage Approved

More information

ISO 376 Calibration Uncertainty C. Ferrero

ISO 376 Calibration Uncertainty C. Ferrero ISO 376 Calibration Uncertainty C. Ferrero For instruments classified for interpolation, the calibration uncertainty is the uncertainty associated with using the interpolation equation to calculate a single

More information

R SANAS Page 1 of 7

R SANAS Page 1 of 7 ESTIMATION OF THE UNCERTAINTY OF MEASUREMENT BY CALIBRATION LABORATORIES AND SPECIFICATION OF CALIBRATION AND MEASUREMENT CAPABILITY ON SCHEDULES OF ACCREDITATION Approved By: Chief Executive Officer:

More information

Vocabulary of Metrology

Vocabulary of Metrology Vocabulary of Metrology For Understating Uncertainty and Traceability Prepared by: Kim, Sang Ho Engineering Team Leader Modal Shop, Inc A PCB Group Company SI National Metrology Institute Secondary Calibration

More information

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. EVALUATION OF MEASUREMENT UNCERTAINTY POLICY... 5

More information

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE Publication Reference EA-10/14 EA Guidelines on the Calibration of Static Torque Measuring Devices PURPOSE This document has been produced by EA to improve harmonisation in determining the calibration

More information

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05 P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05 This document provides guidance on the estimation of uncertainty

More information

PRELIMINARY DRAFT INDIAN STANDARD METHOD FOR CALIBRATION AND CLASSIFICATION OF TORQUE MEASURING DEVICES INCLUDING TORQUE WRENCH TESTER

PRELIMINARY DRAFT INDIAN STANDARD METHOD FOR CALIBRATION AND CLASSIFICATION OF TORQUE MEASURING DEVICES INCLUDING TORQUE WRENCH TESTER For BIS Use Only Doc: PGD 34(1203)P PRELIMINARY DRAFT INDIAN STANDARD METHOD FOR CALIBRATION AND CLASSIFICATION OF TORQUE MEASURING DEVICES INCLUDING TORQUE WRENCH TESTER Not to be reproduced without the

More information

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 4 5. ESTIMATION OF UNCERTAINTY OF MEASUREMENT POLICY...

More information

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments NABL 122-06 NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments Reviewed by MASTER

More information

CALIBRATION REPORT FOR THERMOHYDROMETER

CALIBRATION REPORT FOR THERMOHYDROMETER CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188 ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited

More information

Conversion Formulas Potential Measurement Errors What is involved in building a Primary Torque Standard Proper Handling techniques

Conversion Formulas Potential Measurement Errors What is involved in building a Primary Torque Standard Proper Handling techniques FUNDEMENTALS OF TORQUE CALIBRATION 1 Lesson Objectives Introduction Definition and Examples of Torque Conversion Formulas Torque Calibration Traceability Standard Types of Torque Standards ASTM E2428 BS7882

More information

Guidelines on the Calibration of Static Torque Measuring Devices

Guidelines on the Calibration of Static Torque Measuring Devices European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET/cg-14/v.01 Previously EA-10/14 July 2007 Calibration Guide EURAMET/cg-14/v.01

More information

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014) SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February

More information

Glossary Innovative Measurement Solutions

Glossary Innovative Measurement Solutions Glossary GLOSSARY OF TERMS FOR TRANSDUCERS, LOAD CELLS AND WEIGH MODULES This purpose of this document is to provide a comprehensive, alphabetical list of terms and definitions commonly employed in the

More information

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT Scott Crone North American Sales Manager AMETEK Calibration Instruments 8600 Somerset Drive Largo, Florida 33773 USA INTRODUCTION Like any other piece of equipment,

More information

Process Accuracy Ratio vs Process Uncertainty Ratio (Risk Mitigation: Calibration and the Customer s Process)

Process Accuracy Ratio vs Process Uncertainty Ratio (Risk Mitigation: Calibration and the Customer s Process) 1. Abstract The requirements for the process accuracy may be dictated by the requirements of the expected output of the process or product. The ratio between the process accuracy (i.e., the acceptance

More information

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant SWGDRUG GLOSSARY These definitions were developed and adopted by the SWGDRUG Core Committee from a variety of sources including The United Nations Glossary of Terms for Quality Assurance and Good Laboratory

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares

More information

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0 076-0-067-ACC-USL-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer Accredited by the American Association for Laboratory Accreditation (ALA) CALIBRATION CERT #788.0 SAMPLE

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ARCONIC POWER AND PROPULSION RESEARCH CENTER 1500 S. Warner St. Whitehall, MI 49461-1895 Michelle McDonald Phone: 231 894 7586 CALIBRATION

More information

Uncertainty of Calibration. Results in Force Measurements

Uncertainty of Calibration. Results in Force Measurements Publication Reference EAL-G Uncertainty of Calibration Results in Force Measurements PURPOSE This document has been produced to improve the harmonization in determination of uncertainties in force measurements.

More information

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty October 21, 2013 Summary The Working Group for Fluid Flow (WGFF) defines Calibration and Measurement Capability (CMC) uncertainty

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 MTS FIELD SERVICE 14000 Technology Drive Eden Prairie, MN 55344 Hugh Casper Phone: 518 852 3977 CALIBRATION Valid To: September 30,

More information

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC 17025 Table of contents 1 GENERAL... 2 2 THE STANDARD SIST EN ISO/IEC 17025... 2 3 SA'S POLICY IN IMPLEMENTING

More information

Review of Anemometer Calibration Standards

Review of Anemometer Calibration Standards Review of Anemometer Calibration Standards Rachael V. Coquilla rvcoquilla@otechwind.com Otech Engineering, Inc., Davis, CA Anemometer calibration defines a relationship between the measured signals from

More information

Uncertainty of Measurement A Concept

Uncertainty of Measurement A Concept Uncertainty of Measurement A Concept Shweta V. Matey 1, Dr. Nitin K. Mandavgade 2, and Dr. R.R. Lakhe 3 1 (Asst. Professor, Dept. of Mechanical Engg, Lokmanya Tilak College of Engineering, Mumbai University,

More information

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Torque Measuring Devices

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Torque Measuring Devices NABL 122-10 NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Torque Measuring Devices Reviewed by MASTER COPY

More information

CMM Uncertainty Budget

CMM Uncertainty Budget Table of Contents Purpose...3 Scope...3 Measurement Example...3 Uncertainty Groups...4 Probe Group...5 Calibration Artifact Group...6 Environment Group...7 CMM Group...8 Part Group...10 Conversion of Potential

More information

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE Michael Bair and Pierre Delajoud 2002 DEC 10 Revised 2004 MAR 26 FORWARD FPG8601 is a pressure standard designed

More information

Extending Transducer Calibration Range by Extrapolation

Extending Transducer Calibration Range by Extrapolation Introduction Extending Transducer Calibration by olation by LaVar Clegg Force and torque transducers must be calibrated in a laboratory in order to be useful in their intended application. Applications

More information

Standard Practices for Air Speed Calibration Testing

Standard Practices for Air Speed Calibration Testing Standard Practices for Air Speed Calibration Testing Rachael V. Coquilla Bryza Wind Lab, Fairfield, California Air speed calibration is a test process where the output from a wind measuring instrument

More information

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE Publication Reference EA-10/13 EA Guidelines on the Calibration of Temperature Block Calibrators PURPOSE This document been produced by EA to improve the harmonisation in the calibration of temperature

More information

M3003 EDITION 2 JANUARY 2007

M3003 EDITION 2 JANUARY 2007 M3003 EDITION 2 JANUARY 2007 The Expression of Uncertainty and Confidence in Measurement JH 2007 CONTENTS SECTION PAGE 1 Introduction 2 2 Overview 4 3 More detail 10 4 Type A evaluation of standard uncertainty

More information

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60 Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60 Bernd Meißner Physikalisch-Technische Bundesanstalt Weighing Instruments Laboratory Braunschweig Abstract

More information

CALIBRATION CERTIFICATE # 503

CALIBRATION CERTIFICATE # 503 CALIBRATION CERTIFICATE # 503 Calibration date : 2011-02-23 Certificate issued : 2012-04-18 Company name Company address City, Province, Canada CLAS 2009-02 Calibration of Mass flow meter Micro Motion

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 7500-1 Third edition 2004-08-15 Metallic materials Verification of static uniaxial testing machines Part 1: Tension/compression testing machines Verification and calibration

More information

Guide to the Uncertainty of Force Measurement

Guide to the Uncertainty of Force Measurement April 1999 Guide to the Uncertainty of Force Measurement Liam Maybank, Andy Knott, and David Elkington Centre for Mechanical and Acoustical Metrology National Physical Laboratory Queens Road Teddington

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 JEM PRECISION 9615 56 Avenue NW Edmonton, Alberta T6E 0B2 Canada Robert Korthuis Phone: 780-440 1933 CALIBRATION

More information

Introduction to the evaluation of uncertainty

Introduction to the evaluation of uncertainty Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials SECTION 1 Introduction to the evaluation of uncertainty F A Kandil National Physical Laboratory

More information

The Determination of Uncertainties in Bend Tests on Metallic Materials

The Determination of Uncertainties in Bend Tests on Metallic Materials Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials Code of Practice No. 09 The Determination of Uncertainties in end Tests on Metallic Materials

More information

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY International Conference on Ceramics, Bikaner, India International Journal of Modern Physics: Conference Series Vol. (01) 8 5 World Scientific Publishing Company DOI: 10.11/S010195101019 ROLE AND SIGNIFICANCE

More information

A Unified Approach to Uncertainty for Quality Improvement

A Unified Approach to Uncertainty for Quality Improvement A Unified Approach to Uncertainty for Quality Improvement J E Muelaner 1, M Chappell 2, P S Keogh 1 1 Department of Mechanical Engineering, University of Bath, UK 2 MCS, Cam, Gloucester, UK Abstract To

More information

Measurement Uncertainty Knowing the Unknown

Measurement Uncertainty Knowing the Unknown Measurement Uncertainty Knowing the Unknown Anish Shah FOUNDER and Chief Metrology Officer at Metrologized, LLC, a company founded in 2014 to further the metrology knowledge-base within the manufacturing

More information

THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY

THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY D. Peschel 1, D. Mauersberger 1, D. Schwind 2, U. Kolwinski 2 1 Solid mechanics department, PTB, Germany 2 Gassmann Theiss Messtechnik

More information

Certificate of Accreditation to ISO/IEC 17025:2005

Certificate of Accreditation to ISO/IEC 17025:2005 United States Department of Commerce National Institute of Standards and Technology Certificate of Accreditation to ISO/IEC 17025:2005 NVLAP LAB CODE: 200508-0 Insco Metrology, Inc. Miami, FL is accredited

More information

ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS

ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS 1 Introduction... 3 2 Objective... 3 3 Supplies... 3 4 Theory...

More information

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY 15th EFLM Continuing Postgraduate Course in Clinical Chemistry and Laboratory Medicine How to assess the quality of your method? Zagreb, 24-25 October 2015 HOW TO ASSESS THE MEASUREMENT UNCERTAINTY Centre

More information

Accuracy Calibration Certificate

Accuracy Calibration Certificate gl06-0-067-acc Mettler Toledo Service Business Unit Industrial 900 Polaris Parkway Columbus, OH 0-800-- Services Provided By: Customer Accuracy Calibration Certificate Company: Sample Company Test Address:

More information

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator Introduction Humidity generators are mechanical devices that provide stable temperature and humidity

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ADVANCED TECHNOLOGY SERVICES HMA 1800 Honda Drive Lincoln, AL Lasheila Hayes Phone:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ADVANCED TECHNOLOGY SERVICES HMA 1800 Honda Drive Lincoln, AL Lasheila Hayes Phone: SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ADVANCED TECHNOLOGY SERVICES HMA 1800 Honda Drive Lincoln, AL 35096 Lasheila Hayes Phone: 205 505 0296 CALIBRATION Valid to: April 30, 2019 Certificate Number:

More information

Loadcell Calibration - Evaluation of Uncertainties

Loadcell Calibration - Evaluation of Uncertainties Loadcell Calibration - Evaluation of Uncertainties This document is a revision of the evaluation issued in June 2015. Novatech s calibration laboratory is not UKAS accredited due to the high number of

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V (Hereinafter called the Organization)

More information

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1. 076-00-067-ACC-S Mettler Toledo, LLC 900 Polaris Parkway Columbus, OH 0.800.METTLER Customer SAMPLE Accuracy Calibration Certificate Company: Address: City: Zip / Postal: State / Province: Mettler Toledo

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares

More information

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION IMEKO 22 nd TC3, 15 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION R. Kumme

More information

UNCERTAINTY SCOPE OF THE FORCE CALIBRATION MACHINES. A. Sawla Physikalisch-Technische Bundesanstalt Bundesallee 100, D Braunschweig, Germany

UNCERTAINTY SCOPE OF THE FORCE CALIBRATION MACHINES. A. Sawla Physikalisch-Technische Bundesanstalt Bundesallee 100, D Braunschweig, Germany Measurement - Supports Science - Improves Technology - Protects Environment... and Provides Employment - No and in the Future Vienna, AUSTRIA, 000, September 5-8 UNCERTAINTY SCOPE OF THE FORCE CALIBRATION

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:. Av. Rio Consulado # 2614, Col. San Juan de Aragón

More information

< Final report > Report on the APMP.M.F-S1 supplementary comparison for 2 MN Force

< Final report > Report on the APMP.M.F-S1 supplementary comparison for 2 MN Force < Final report > Report on the APMP.M.F-S1 supplementary comparison for 2 MN Force Kazunaga Ueda*, Toshiyuki Hayashi*, Hiroshi Maejima*, Rolf Kumme**, Dirk Röske**, Mark Seidel** * National Metrology of

More information

AERO 214. Lab II. Measurement of elastic moduli using bending of beams and torsion of bars

AERO 214. Lab II. Measurement of elastic moduli using bending of beams and torsion of bars AERO 214 Lab II. Measurement of elastic moduli using bending of beams and torsion of bars BENDING EXPERIMENT Introduction Flexural properties of materials are of interest to engineers in many different

More information

Measurement & Uncertainty - Basic concept and its application

Measurement & Uncertainty - Basic concept and its application Measurement & Uncertainty - Basic concept and its application Feb. 7. 2012 JongOh CHOI (choijongoh@kriss.re.kr) Knowledge and measurement When you can measure what you are speaking about, and express it

More information

T e c h n i c a l L e x i c o n

T e c h n i c a l L e x i c o n T e c h n i c a l L e x i c o n T e r m s f r o m F o r c e a n d t o r q u e m e a s u r e m e n t t e c h n o l o g y a n d G T M d a t a s h e e t s Technisches Lexikon (eng.) 16/10/14 Introduction

More information

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES Proceedings of the ASME 2014 Power Conference POWER2014 July 28-31, 2014, Baltimore, Maryland, USA POWER2014-32205 UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES Michael S. Hering Rosemount DP

More information

The Uncertainty of a Waterdraw Calibration vs. Gravimetric Calibration on Small Volume Provers

The Uncertainty of a Waterdraw Calibration vs. Gravimetric Calibration on Small Volume Provers The Uncertainty of a Waterdraw Calibration vs. Gravimetric Written by: Gary Cohrs Introduction The focus of the API Petroleum Measurement Manuals is to reduce the uncertainty and improve the accuracy of

More information

What is measurement uncertainty?

What is measurement uncertainty? What is measurement uncertainty? What is measurement uncertainty? Introduction Whenever a measurement is made, the result obtained is only an estimate of the true value of the property being measured.

More information

POWER QUALITY MEASUREMENT PROCEDURE. Version 4 October Power-Quality-Oct-2009-Version-4.doc Page 1 / 12

POWER QUALITY MEASUREMENT PROCEDURE. Version 4 October Power-Quality-Oct-2009-Version-4.doc Page 1 / 12 POWER QUALITY MEASUREMENT PROCEDURE Version 4 October 2009 Power-Quality-Oct-2009-Version-4.doc Page 1 / 12 MEASNET 2009 Copyright all rights reserved This publication may not be reproduced or utilized

More information

A basic introduction to reference materials. POPs Strategy

A basic introduction to reference materials. POPs Strategy A basic introduction to reference materials POPs Strategy 2009-2010+ A tutorial 16 September 2009 Angelique Botha R&D metrologist Contents Why do we need reference materials? comparability of results metrological

More information

Measurement method for the proficiency testing program

Measurement method for the proficiency testing program APLAC T088 Appendix Measurement method for the proficiency testing program Introductions This measurement method is prepared for use by the APLAC Proficiency Testing Program Photometric measurement of

More information

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS W. Beyer and W. Pahl Physikalisch-Technische Bundesanstalt (PTB) 38116 Braunschweig, Germany Abstract: In accordance with ISO 17025,

More information

1.11 Measurement Uncertainty

1.11 Measurement Uncertainty Document #: 4968 Page 1 of 10 1.11 Measurement Uncertainty In an effort to comply with accreditation requirements, and because scientific measurements in general are subject to variability, a budget estimating

More information

Calibration of Temperature Block Calibrators

Calibration of Temperature Block Calibrators European Association of National Metrology Institutes Calibration of Temperature Block Calibrators EURAMET cg-13 Version 2.0 (03/2011) Previously EA-10/13 Calibration Guide EURAMET cg-13 Version 2.0 (03/2011)

More information

Procedure for Uncertainty Determination for Calibration Consoles

Procedure for Uncertainty Determination for Calibration Consoles Category: ELECTRICITY Procedure: EL-ENG-09-01 Page: 1 of 19 Procedure for Uncertainty Determination for Calibration Consoles Adnan Rashid Senior Electrical Engineer Engineering and Laboratory Services

More information

Specific Accreditation Guidance. Infrastructure and Asset Integrity. Measurement Uncertainty in Geotechnical Testing

Specific Accreditation Guidance. Infrastructure and Asset Integrity. Measurement Uncertainty in Geotechnical Testing Specific Accreditation Guidance Infrastructure and Asset Integrity Measurement Uncertainty in Geotechnical Testing January 2018 Copyright National Association of Testing Authorities, Australia 2018 This

More information

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities T. J. Wiandt 1,2 1 Fluke Corporation, Hart Scientific Division, American Fork, Utah United States. 2 E-mail: tom.wiandt@hartscientific.com

More information

Training Benefits Adding Values to Your Systems/ Organization through: Potential Attendees These Courses have been designed for;

Training Benefits Adding Values to Your Systems/ Organization through: Potential Attendees These Courses have been designed for; Potential Attendees These Courses have been designed for; Calibration engineers and Technicians. Training Benefits Adding Values to Your Systems/ Organization through: Planning Engineers and decision makers.

More information

The Fundamentals of Moisture Calibration

The Fundamentals of Moisture Calibration The Fundamentals of Moisture The following guide will provide you with a basic knowledge of humidity calibration, help you to understand your requirements and select an appropriate solution. 1 Why Humidity

More information

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 Technical Protocol of the CIPM Key Comparison CCAUV.V-K5 2017-03-06 revised 2018-02-13 (changed schedule) Task and Purpose of the Comparison According to the rules set up by the CIPM MRA the consultative

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 QAL-TEK ASSOCIATES 3998 Commerce Circle Idaho Falls, ID 83401 Travis Snowder Phone: 208 523 5557 CALIBRATION Valid To: January 31, 2019

More information

BioMechanics and BioMaterials Lab (BME 541) Experiment #5 Mechanical Prosperities of Biomaterials Tensile Test

BioMechanics and BioMaterials Lab (BME 541) Experiment #5 Mechanical Prosperities of Biomaterials Tensile Test BioMechanics and BioMaterials Lab (BME 541) Experiment #5 Mechanical Prosperities of Biomaterials Tensile Test Objectives 1. To be familiar with the material testing machine(810le4) and provide a practical

More information

AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS

AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS WHITE PAPER AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS The Energy Conservatory Minneapolis, MN Introduction: The purpose of this document is to explain the details of the accuracy specifications

More information

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis TECHNICAL SPECIFICATION ISO/TS 13530 First edition 2009-03-15 Water quality Guidance on analytical quality control for chemical and physicochemical water analysis Qualité de l'eau Lignes directrices pour

More information

Technical Procedure for Measurement Assurance

Technical Procedure for Measurement Assurance Technical Procedure for Measurement Assurance 1.0 Purpose - This procedure specifies the required elements for measurement assurance in the Drug Chemistry Sections of the State Crime Laboratory. 2.0 Scope

More information

Essentials of expressing measurement uncertainty

Essentials of expressing measurement uncertainty Essentials of expressing measurement uncertainty This is a brief summary of the method of evaluating and expressing uncertainty in measurement adopted widely by U.S. industry, companies in other countries,

More information

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks INTERNATIONAL STANDARD ISO 6506-3 Third edition 2014-10-01 Metallic materials Brinell hardness test Part 3: Calibration of reference blocks Matériaux métalliques Essai de dureté Brinell Partie 3: Étalonnage

More information

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator Technical Note The Fluke Calibration 5128A RHapid-Cal Humidity Generator provides a portable, stable test environment

More information

MEASUREMENT UNCERTAINTY OF ROTATING TORQUE TRANSDUCERS WHEN USED IN PARTIAL LOAD RANGES

MEASUREMENT UNCERTAINTY OF ROTATING TORQUE TRANSDUCERS WHEN USED IN PARTIAL LOAD RANGES XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17, 006, Rio de Janeiro, Brazil MEASUREMENT UNCERTAINTY OF ROTATING TORQUE TRANSDUCERS WHEN USED IN PARTIAL LOAD RANGES Georg

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ALABAMA SCALE & INSTRUMENT MARSHALL, INC. d.b.a. ASI Calibration Labs - Texas 1006 East End Blvd North Marshall, TX 75672 Lawson Hill

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. J. A. KING & COMPANY, LLC 8813 Old Craft Road Olive Branch, MS CONNIE FOSTER PHONE:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. J. A. KING & COMPANY, LLC 8813 Old Craft Road Olive Branch, MS CONNIE FOSTER PHONE: SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 J. A. KING & COMPANY, LLC 8813 Old Craft Road Olive Branch, MS 38654 CONNIE FOSTER PHONE: 800 327 7727 CALIBRATION Valid To: August 31, 2019 Certificate Number:

More information

Appendix A: Laboratory Report Format

Appendix A: Laboratory Report Format Appendix A: Laboratory Report Format This appendix contains details on the format for formal written laboratory reports in the form of author's instructions. Author's Instructions for ME354 Formal Lababoratory

More information

Mason Technology Ltd

Mason Technology Ltd Mason Technology Ltd 228 South Circular Road, Dublin 8 Calibration Laboratory Registration number: 043C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in the Schedule

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 LAKE SHORE INDUSTRIAL SERVICES, INC. 2230 Caughey Road Erie, PA 16506 Suzanne R. Zuba / James B. Cross Phone: 814 838 3539 CALIBRATION

More information

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s):

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s): SOP No: ICLN-Validation Levels Between Networks.003.01 Page 1 of 19 1. Purpose: This document establishes guidelines for assigning validation levels to methods based on national and international standards.

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: AMC LABS S.A de C.V Av. México # 329 Col. Valle De

More information

CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu

CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS. A.Gnanavelu CALCULATION OF UNCERTAINTY IN CHEMICAL ANALYSIS A.Gnanavelu UNCERTAINTY (U) Every measurement has error which is unknown and unknowable. This unknown error is called measurement uncertainty. Types of Error

More information