Simultaneous Determination of Aromatic Amines and Pyridines in Soil

Size: px
Start display at page:

Download "Simultaneous Determination of Aromatic Amines and Pyridines in Soil"

Transcription

1 Simultaneous Determination of Aromatic Amines and Pyridines in Soil Huang Xiongfeng, Liu Lvye, Xu Qun, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 4 Key Words Aniline Compounds, Environmental Analysis, Environmental Safety, Acclaim 0 C8 Column, HPLC Goal To develop an efficient high-performance liquid chromatography (HPLC) method for fast and simultaneous determination of -methylpyridine, 3-methylpyridine, m-toluidine, and p-toluidine in soil Introduction Aniline and pyridines are widely used in the polymer, fuel, pesticide, rubber, and pharmaceutical industries. Not only do some of these compounds have a putrid odor, but they are also suspected potential carcinogens and mutagens that can be highly toxic to life. Therefore, the Ministry of Environmental Protection of the People s Republic of China has proposed a provisional guideline concentration of 0. mg/l for pyridine in environmental water. The U.S. Environmental Protection Agency (EPA) lists aniline and -methylaniline in its Health Effects Notebook for Hazardous Air Pollutants. 3 As cities grow larger and become more crowded, some of the factories that use these hazardous compounds are closed, and houses and parks are then built on the vacated land. The soil may be polluted, however, and require treatment before building. Therefore, it is necessary to establish a sensitive, efficient, and simple method for the determination of compounds such as pyridines in soil samples. The most common techniques used for determination of aromatic amine and pyridine compounds in environmental samples are gas chromatography (GC) and HPLC, although capillary zone electrophoresis (CZE) and spectrophotometric methods also have been reported. 0 Because the compounds involved are thermolabile and polar, a derivatization step prior to GC analysis is often required and most such procedures are time consuming and complicated. Therefore, HPLC analysis is a good alternative to GC analysis because derivatization is not needed. Figure shows the structures of four aromatic amine and pyridine compounds that will be simultaneously determined in this study using HPLC. NH NH CH 3 N CH 3 N CH 3 CH3 -Methylpyridine 3-Methylpyridine m-toluidine p-toluidine Figure. Structures of four aromatic amine and pyridine compounds.

2 Equipment, Software, and Consumables Thermo Scientific Dionex UltiMate 3000 Rapid Separation LC (RSLC) system, including: LPG-3400RS Quaternary Pump (P/N ) SRD-3400 Integrated Solvent and Degasser Rack (P/N ) WPS-3000TRS Wellplate Sampler, Thermostatted (P/N ), with 5 μl sample loop (P/N ) TCC-3000RS Thermostatted Column Compartment (P/N ) DAD-3000RS Diode Array Detector (P/N ) with.5 μl flow cell (P/N ) Thermo Scientific Dionex Chromeleon Chromatography Data System (CDS) software, version 7. Thermo Scientific Sorvall ST6 Centrifuge (P/N ) Thermo Scientific Target Nylon Syringe Filters, 0.45 μm, 30 mm (P/N F500-) Reagents and Standards Deionized (DI) water, 8. MΩ-cm resistivity, generated using a Thermo Scientific Barnstead GenPure Pro ultrapure water system (P/N ) Acetonitrile, HPLC Grade (Fisher Scientific P/N AC ) Methanol, 99.8%, HPLC Grade (Fisher Scientific P/N AC ) Ammonium Acetate, Crystalline/HPLC (Fisher Scientific P/N A ) Acetic Acid, Optima LC/MS (Fisher Scientific P/N A3-50) -Methylpyridine, 5GR (Fisher Scientific P/N ) 3-Methylpyridine, >98%, GC (Fisher Scientific P/N ) p-toluidine, 99% (Fisher Scientific P/N AC ) m-toluidine, 99% (Fisher Scientific P/N AC ) Conditions Column: Thermo Scientific Acclaim 0 C8, 3 µm Analytical,. 00 mm (P/N 0599) Eluent: Acetonitrile/ammonium acetate buffer (dissolve 3.85 g of ammonium acetate and 3 g of acetic acid in L of DI water) Gradient: 0 min, 5% acetonitrile; 3.5 min, 80% acetonitrile; min, 5% acetonitrile Flow Rate: 0.4 ml/min Injection Volume: µl (partial-loop injection mode) Temperature: 35 C Detection: UV, 60 nm Preparation of Standard Solutions Stock Standard Mix Dissolve 0. g each of -methylpyridine, 3-methylpyridine, m-toluidine, and p-toluidine in a 00 ml volumetric flask and dilute to volume with methanol. The concentration of each compound will be mg/ml. Working Mixed Standard Solutions For calibration, prepare five working mixed standard solutions with different concentrations by diluting the proper amount of each stock standard solution mix with methanol. The volumes of each solution needed to make the calibration standards are shown in Table. Sample Preparation Soil Sample used in this work was collected from Jinqiao Park in Shanghai, People s Republic of China. Soil Sample, also from Shanghai, was supplied by a real estate developer. Add 0 ml of methanol to g of an air-dried soil sample in a 0 ml centrifuge tube and vortex for min. After ultrasonically extracting the sample at 50 W for 40 min, centrifuge the extract for min at 8000 rpm. Transfer the supernatant to a 0 ml tube and use nitrogen to dry the sample. Add ml methanol to the dried sample. Prior to injection, filter the solution through a 0.45 μm filter. Table. Preparation of working mixed standard solutions. Vol of mg/ml Stock Standard Mix Vol of Methanol Final Vol of Working Mixed Standard Solution Final Concn of Each Calibration Standard

3 To prepare the spiked soil sample, add 5 μl of the mg/ml stock standard mix to g of a soil sample. Prepare the sample as described above. The spike concentrations will be 5 mg/kg for each analyte. Results and Discussion Separation of -Methylpyridine, 3-Methylpyridine, p-toluidine, and m-toluidine Initial experiments showed that when using methonal/ water or acetontrile/water as the mobile phase, it was difficult to separate p- and m-toluidine. Therefore, the addition of a phosphate or acetate buffer to the mobile phase was investigated as a way to achieve the separation. The peak resolution between p- and m-toluidine was.4 when using an acetonitrile/acetate buffer mobile phase. In addition to the baseline separation, the peak asymmetry factors of -methylpyridine, 3-methylpyridine, p-toluidine, and m-toluidine were all in the range of 0.9~.4 better than that yielded when using methanol/acetate, methanol/phosphate, and acetonitrile/phosphate mobile phases. Therefore, an acetonitrile/acetate buffer mobile phase was used in this work. Figure shows a chromatogram of a standard mix of -methylpyridine, 3-methylpyridine, p-toluidine, and m-toluidine under the specified and optimized chromatographic conditions. Method Reproducibility, Linearity, and Detection Limit Method precision was estimated by making nine consecutive µl injections of a calibration standard with a concentration of 0 mg/l for each compound. Retention time and peak area reproducibilities (RSDs) are shown in Table. Retention time RSDs all are <0.% and peak area RSDs all are <.%, demonstrating good short-term precision for this method. 45 mau 0 Figure. Separation of a mixed standard solution. Peaks:. -Methylpyridine 5 mg/l. 3-Methylpyridine 5 3. p-toluidine 5 4. m-toluidine 5 Calibration linearity of -methylpyridine, 3-methylpyridine, p-toluidine, and m-toluidine was investigated by making three consecutive μl injections of the working mixed standard solution prepared at five different concentrations (i.e., 5 total injections). A linearity range of 000 mg/l for these four compounds was observed when plotting concentration versus peak area. Detailed calibration data calculated by Chromeleon CDS software are shown in Table. Five consecutive μl injections of a working mixed standard solution with a concentration of mg/l for each compound were used for estimating the method detection limit (MDL) using a signal-to-noise ratio = 3. The calculated MDLs also are listed in Table Minutes 3 Table. Calibration data and MDLs. Target Analyte Retention Time RSD Peak Area RSD Regression Equation r Range MDL -Methylpyridine A = 0.6c Methylpyridine A = 0.6c p-toluidine A = 0.05c m-toluidine A = 0.037c

4 4 Table 3. Sample analysis results. Soil Sample Analyte Detected Added Spiked Sample Found Recovery (%) -Methylpyridine Methylpyridine Not p-toluidine Detected m-toluidine Analysis of Soil Samples -Methylpyridine, 3-methylpyridine, p-toluidine, and m-toluidine were not found in the soil samples. Figure 3 shows chromatograms of both soil samples and Soil Sample spiked with four aromatic amines and pyridines. To judge method accuracy, three injections of Soil Sample spiked with 5 mg/kg of each of the standards were made. Analysis results are summarized in Table 3. The average recoveries were in the range of 76 0%, demonstrating that this method can achieve good recovery and is suitable for the determination of aromatic amines and pyridines in soil. Conclusion This work uses an ammonium acetate/acetonitrile mobile phase for a rapid, simple, and accurate determination of -methylpyridine, 3-methylpyridine, p-toluidine, and m-toluidine in soil. The separation requires only 6 min and demonstrates good reproducibility and recovery. 45 mau D C B Peaks:. -Methylpyridine. 3-Methylpyridine 3. p-toluidine 4. m-toluidine 0 A Minutes 34 Figure 3. Separation of (A) a blank, (B) Soil Sample, (C) Soil Sample, and (D) Soil Sample spiked with 5 mg/kg of each standard.

5 References. Brede, C.; Skjevrak, I.; Herikstad, H. Determination of Primary Aromatic Amines in Water Food Simulant Using Solid-Phase Analytical Derivatization Followed by Gas Chromatography Coupled with Mass Spectrometry. J. Chromatogr., A 003, 983, GB : Environmental Quality Standards for Surface Water; Ministry of Environmental Protection of the People s Republic of China: Beijing, U.S. Environmental Protection Agency. Health Effects Notebook for Hazardous Air Pollutants. [Online] (accessed Oct 3, 04). 4. Chiang, J.S.; Huang, S.D. Simultaneous Derivatization and Extraction of Anilines in Waste Water with Dispersive Liquid Liquid Microextraction Followed by Gas Chromatography Mass Spectrometric Detection. Talanta 008, 75, Jen, J.F.; Chang, C.T.; Yang, T.C. On-Line Microdialysis High-Performance Liquid Chromatographic Determination of Aniline and -Chloroaniline in Polymer Industrial Wastewater. J. Chromatogr., A 00, 930, Sarafraz-Yazdi, A.; Es haghi, Z. Liquid Liquid Liquid Phase Microextraction of Aromatic Amines in Water Using Crown Ethers by High-Performance Liquid Chromatography with Monolithic Column. Talanta 005, 66, Zhao, L.M.; Zhu, L.Y.; Lee, H.K. Analysis of Aromatic Amines in Water Samples by Liquid Liquid Liquid Microextraction with Hollow Fibers and High-Performance Liquid Chromatography. J. Chromatogr., A 00, 963, Thermo Scientific Application Note 9: Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE. Sunnyvale, CA, 04. [Online] dam/tfs/atg/cmd/cmd%0documents/application %0&%0Technical%0Notes/Chromatography/ Liquid%0Chromatography/Liquid%0Chromatog- raphy%0accessories/an-9-lc-aniline-nitroanilines- Water-SPE-AN703-EN.pdf (accessed Oct 3, 04). 9. Li, J.; Yuan, Z.B. Separation of Aniline Derivatives by Micellar Electrokinetic Capillary Chromatography. Chin. Chem. Lett. 004, 5, Gu, X.X.; Li, C.Y.; Qi, X.; Zhou, T.Z. Determination of Trace Aniline in Water by a Spectrophotometric Method after Preconcentration on an Organic Solvent-Soluble Membrane Filter. Anal. Lett. 997, 30, 59. Application Note Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa Australia Austria Belgium Brazil Canada China (free call domestic) AN743-EN 076S Denmark Europe-Other Finland France Germany India Italy Japan Korea Latin America Middle East Netherlands New Zealand Norway Russia/CIS Singapore Sweden Switzerland Taiwan UK/Ireland USA

Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE

Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE Li Lang, Xu Qun, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo

More information

Determination of Benzo(a)pyrene in Sausage and Preserved Ham

Determination of Benzo(a)pyrene in Sausage and Preserved Ham Determination of Benzo(a)pyrene in Sausage and Preserved Ham Ran Liangji, Jin Yan, Xu Qun, Liang Lina, and Jeffrey Rohrer 2 Thermo Fisher Scientific, Shanghai, People s Republic of China; 2 Thermo Fisher

More information

Determination of Carbonyl Compounds in Workplace Air

Determination of Carbonyl Compounds in Workplace Air Determination of Carbonyl Compounds in Workplace Air Huang Xiongfeng, Xu Qun, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo Fisher Scientific, Sunnyvale, CA,

More information

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water APPLICATION NOTE 70923 Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water Authors Chen Jing, Dai Zhenyu, Xu Qun, and Liang Lina, Thermo Fisher Scientific, Shanghai, People

More information

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA New Stationary Phases for Solid-Phase Extraction Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Purpose: To develop five new polymer-based stationary

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo

More information

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners APPLICATION NOTE 21734 Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners Authors Aaron Lamb, Joanne Jones, Thermo Fisher Scientific, Runcorn, UK Keywords Environment, food

More information

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil Che Jinshui, 1 Deng Guifeng, 1 Liang Lina, 1 and Aaron Kettle, 2 1 Thermo Fisher Scientific (China)

More information

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis A. Caruso, M. Santoro, P. Magni, S. Pelagatti, and R. Facchetti Thermo Fisher Scientific, Milan,

More information

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key

More information

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column Application Note 95 Determination of Hydrochloride Purity Using the Acclaim PA Column INTRODUCTION -based medications are prescribed for several heart and blood pressure indications. The fastacting formulations

More information

Monitoring Protein PEGylation with Ion Exchange Chromatography

Monitoring Protein PEGylation with Ion Exchange Chromatography Monitoring Protein PEGylation with Ion Exchange Chromatography Peter Yu, Deanna Hurum, Jinyuan (Leo) Wang, 2 Terry Zhang, 2 and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA; 2 Thermo Fisher Scientific,

More information

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution Marc Plante, Bruce Bailey, Christopher Crafts, Ian N. Acworth Thermo Fisher Scientific, Chelmsford,

More information

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS Jason Lai, Jia Wang, Brad Hart, Pavel Aronov, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Andy Jacobs, Haibo

More information

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection Mark Tracy, Xiaodong Liu, and Ian Acworth, Thermo Fisher Scientific, Sunnyvale, CA,

More information

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany Generic Method Approach for Pharmaceutical Drug Discovery and Development using Reversed-Phase Hydrophilic Interaction Liquid Chromatography with Universal Charged Aerosol Detection Frank Steiner, Michael

More information

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection APPLICATION NOTE 72686 Determination of underivatized aflatoxins,, G2, and in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection Authors Sylvia Grosse, Mauro De Pra, Frank

More information

Fast methods for the determination of ibuprofen in drug products

Fast methods for the determination of ibuprofen in drug products APPLICATION NOTE 779 Fast s for the determination of ibuprofen in drug products Authors Sylvia Grosse, Mauro De Pra, Frank Steiner, Thermo Fisher Scientific, Germering, Germany Keywords Pharmaceutical,

More information

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS Kate Comstock, Tim Stratton, Hongxia (Jessica) Wang, and Yingying Huang

More information

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry O. Scheibner, 1 P. van Baar, 2 F. Wode, 2 U. Dünnbier, 2 K. Akervik, 3 J. Humphries, 3 M. Bromirski 1

More information

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford,

More information

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection Ian Acworth, Bruce Bailey, Paul Gamache, and John Waraska Thermo Fisher Scientific, Chelmsford, MA, USA Introduction

More information

HILIC Method Development in a Few Simple Steps

HILIC Method Development in a Few Simple Steps HILIC Method Development in a Few Simple Steps Monica Dolci, Luisa Pereira, Dafydd Milton and Tony Edge Thermo Fisher Scientific, Runcorn, Cheshire, UK Overview This poster presents a systematic approach

More information

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection and Select Impurities Analysis Using a ydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection Bruce Bailey, Ph. D. Thermo Fisher Scientific, Chelmsford, MA, USA Application

More information

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS Daniel Kutscher, 1 Jörg Bettmer, 2 Torsten Lindemann, 1 Shona McSheehy-Ducos, 1 Lothar Rottmann 1 1 Thermo Fisher Scientific, Germany 2

More information

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic

More information

APPLICATION NOTE Goal. Introduction. Keywords

APPLICATION NOTE Goal. Introduction. Keywords APPLICATION NOTE 72398 A highly sensitive high-performance liquid chromatographycharged aerosol detection method for the quantitative analysis of polysorbate 8 in protein solution Authors Zhen Long, Guobin

More information

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method Xiang He and Marta Kozak ThermoFisher Scientific, San Jose, CA,

More information

Determination of Monochloroacetic Acid in Carbocisteine

Determination of Monochloroacetic Acid in Carbocisteine Determination of Monochloroacetic Acid in Carbocisteine Zhong Xinlin, Ye Mingli, Liang Lina, Xu Qun, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo Fisher Scientific,

More information

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency Anthony Edge, 1 Luisa Pereira, 1 Dafydd Milton 1 and Andrew Shalliker 2 1 Thermo Fisher Scientific,

More information

Enhancement of Linearity and Response in Charged Aerosol Detection

Enhancement of Linearity and Response in Charged Aerosol Detection Enhancement of Linearity and Response in Charged Aerosol Detection Christopher Crafts, Marc Plante, Bruce Bailey, Ian Acworth, Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: This work

More information

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals Analytical Methods to Qualify and Quantify and ylated Biopharmaceuticals Christopher Crafts, Bruce Bailey, Marc Plante, Ian Acworth Thermo Fisher Scientific, Chelmsford, MA Application Note 1015 Key Words

More information

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents Marco Karsten, Bas, Dolman, Giovanni Maio, Frank Steiner, Holger Franz, Frank Arnold and Remco Swart LC Packings,

More information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific,

More information

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by PLC Deanna C. urum, Brian M. De Borba, and Jeffrey S. Rohrer Dionex Corporation, Sunnyvale, CA, USA ITRDUCTI Peptides perform

More information

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford,

More information

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma Jon Bardsley 1, James Goldberg 2 1 Thermo Fisher Scientific, Runcorn, UK; 2 Thermo Fisher Scientific, West Palm Beach,

More information

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Terri Christison, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview

More information

Determination of Additives in Carbonated Beverages

Determination of Additives in Carbonated Beverages pplication Note 9 Determination of dditives in Carbonated everages Introduction The soft drink industry is one of the largest in the world, with revenue from sales of carbonated soft drinks totaling billions

More information

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance David Thomas, Bruce Bailey, Marc Plante and Ian Acworth Thermo Fisher Scientific,

More information

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications Xiaodong Liu and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA Overview Review mixed-mode column technology

More information

Thermo Scientific. Anion-Exchange Column

Thermo Scientific. Anion-Exchange Column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS9-µm Anion-Exchange Column Product Specifications The Themo Scientific Dionex IonPac AS9-µm high-capacity, hydroxide-selective, anion-exchange column is

More information

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration Zhang Xuebin 1, Yu Chongtian 1, Liang Lina 1, Hans-Joachim Huebschmann 2, Thermo Fisher Scientific Shanghai, 1 China,

More information

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components APPLICATION UPDATE 157 Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components Authors Sumate Pengpumkiat, Weerapong Worawirunwong,

More information

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 000 Elemental Analyzer Dr. Liliana Krotz and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy

More information

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System Application Update 55 Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free System Introduction Hydrogen peroxide is an essential chemical in the fabrication

More information

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use Shih-Tse Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang,

More information

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity Agilent 12 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity A case study with degraded omeprazole drug product Application Note Small Molecule

More information

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview

More information

Determination of Methylamine in Drug Products

Determination of Methylamine in Drug Products Determination of Methylamine in Drug Products Thunyarat Phesatcha, Suparerk Tukkeeree, and Jeff Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application

More information

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20544 Key Words Hydrophilic, HILIC,

More information

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Determination of Inorganic Anions in Drinking Water by Ion Chromatography Determination of Inorganic s in Drinking Water by Ion Chromatography Peter Jackson Thermo Fisher Scientific Inc. Application Note Introduction The determination of common inorganic anions in drinking water

More information

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Determination of Tartaric cid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Suparerk Tukkeeree, Chanita Chantarasukon, and Jeff Rohrer Thermo Fisher Scientific, angkok,

More information

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer Olaf Scheibner and Maciej Bromirski Thermo Fisher Scientific, Bremen, Germany Overview Purpose: Improve the performance

More information

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma LC-MS/MS Method for the Determination of Diclofenac in Human Plasma J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20569 Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core

More information

Determination of trace anions in concentrated hydrofluoric acid

Determination of trace anions in concentrated hydrofluoric acid APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion

More information

[ Care and Use Manual ]

[ Care and Use Manual ] Oasis HLB DISKS I. Introduction Oasis HLB is a Hydrophilic-Lipophilic-Balanced, water-wettable, reversed-phase sorbent for all your SPE needs. It is made from a specific ratio of two monomers, the hydrophilic

More information

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Comprehensive Neurochemical Profiling of Brain Tissue Samples Comprehensive Neurochemical Profiling of Brain Tissue Samples Bruce Bailey, PhD Thermo Fisher Scientific, Chelmsford, MA, USA Application Note 1060 Key Words Biogenic amines, amino acid precursors, acid

More information

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter Luisa Pereira, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20755 Key Words Solid core, fused core, superficially

More information

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

A Platform Method for Pharmaceutical Counterion Analysis by HPLC A Platform Method for Pharmaceutical Counterion Analysis by HPLC Xiaodong Liu, Mark Tracy, and Christopher Pohl; Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Results and Simple, fast, economical

More information

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS Application Note Food Testing and Agriculture Author Chen-Hao (Andy) Zhai and Rong-jie Fu Agilent Technologies (Shanghai)

More information

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography Yan Liu, Zhongqing Lu, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA USA Overview Purpose: In this work, new high-pressure

More information

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches Hydrophilic Interaction Liquid Chromatography: Method Development Approaches Monica Dolci, Luisa Pereira, and Tony Edge Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract This poster summarizes the

More information

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device Determination of Bromate in Bottled Mineral Water Using the CRD 00 Carbonate Removal Device Thunyarat Phesatcha, Weerapong Worawirunwong, and Jeff Rohrer Thermo Fisher Scientific, Bangkok, Thailand; Thermo

More information

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Trace Cations in Power Plant Waters Containing Morpholine Application Note 8 Determination of Trace Cations in Power Plant Waters Containing Morpholine INTRODUCTION Morpholine and ammonium are used as additives in power plant waters. Morpholine acts as a corrosion

More information

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry Olaf Scheibner, 1 Patrizia van Baar, 2 Florian Wode, 2 Uwe Dünnbier, 2 Kristi Akervik, 3 Jamie Humphrie,

More information

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis Tim Stratton 1, Yingying Huang 1, Katianna Pihakari 1, Ian Acworth 2, and Michael Weber 2 1 Thermo Fisher Scientific,

More information

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer Reiko Kiyonami, Mary Blackburn, Andreas FR Hühme: Thermo Fisher

More information

Advances in Sample Preparation for Accelerated Solvent Extraction

Advances in Sample Preparation for Accelerated Solvent Extraction Advances in Sample Preparation for Accelerated Solvent Extraction SM Rahmat Ullah, Kannan Srinivasan, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Purpose: To demonstrate the utility

More information

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000 Evaluation of Various Anion-Exchange Chemistries for the Analysis of Haloacetic Acids in Drinking Water Using -D Matrix Elimination Ion Chromatography and Suppressed Conductivity Detection White Paper

More information

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery Metabolite Stability Screening and Hotspot Metabolite Identification by Combining High-Resolution, Accurate-Mass Nonselective and Selective Fragmentation Tim Stratton, Caroline Ding, Yingying Huang, Dan

More information

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites Robert Heintz, Mark Wall, Jennifer Ramirez, Stephan Woods Thermo Fisher Scientific, Madison WI Overview Purpose:

More information

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS Kelly Munro, 1 Anthony Edge, 2 Claudia Martins, 3 David Cowan 4 and Leon Barron 1 1 Analytical & Environmental

More information

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Application Note: 3 Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery Yoko Yamagishi, Thermo Fisher Scientific, C-2F 3-9

More information

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer Application ote: 54 Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap rbitrap Hybrid Mass Spectrometer Kate Comstock, Yingying Huang; Thermo Fisher Scientific,

More information

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD Massimo Santoro, 1 Sergio Guazzotti, 1 Danilo Pierone, 2 Alexandre Souza, 3 Jaqueline Lorena,

More information

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples Application Note 79 Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples INTRODUCTION Cow s milk is of particular dietary value to infants,

More information

Utility of the Charge Detector in Ion Chromatography Applications

Utility of the Charge Detector in Ion Chromatography Applications Utility of the Charge Detector in Ion Chromatography Applications Mrinal K. Sengupta, Sheetal Bhardwaj, Kannan Srinivasan, Chris Pohl, and Purnendu K. Dasgupta Thermo Fisher Scientific, Sunnyvale, California,

More information

of mass spectrometry

of mass spectrometry Thermo Scientific 253 Ultra High resolution isotope ratio MS Discover a new world of mass spectrometry Paleoclimatology Atmospheric science Biogeochemistry Petrology Discover the isotopic anatomy of molecules

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Determination of urea in ultrapure water by IC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore

More information

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS Richard Jack, Xiaodong Liu, Leo Wang, and Chris Pohl OT70806_E 08/13S 1 The world leader in serving science

More information

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: Examples

More information

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection Barbara Shao, Terri Christison, Fei Pang, Cathy Tanner, and Frank Hoefler, Thermo Fisher Scientific, Sunnyvale,

More information

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS high-capacity, carbonate based anion-exchange column is designed for

More information

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS CH.MOUNIKA*, M.KINNERA Research Article SIR.C.R.REDDY COLLEGE OF PHARMACEUTICAL SCIENCES, ELURU.

More information

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Improved Screening for 2 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer Mary lackburn, Jia Wang, Jonathan eck, Charles Yang, Dipankar Ghosh, Thermo Fisher Scientific, San Jose, C,

More information

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water Thermo Scientific Dionex Ion Chromatography Solutions Global water safety bromate analysis in drinking water Safe drinking water public health assurance As a vital limited resource required for survival,

More information

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent Infinity LC System with Diode Array Detector Application Note Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany [mau]

More information

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging Katerina Bousova, 1 Michal Godula, 2 Michele Suman 3 1 Thermo Fisher Scientific, Special Solutions

More information

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Determination of silicate in high-purity water using ion chromatography and online sample preparation PPLICTION NOTE 70 Determination of silicate in high-purity water using ion chromatography and online sample preparation uthors Weerapong Worawirunwong and Jeffrey Rohrer Thermo Fisher Scientific, Bangkok,

More information

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method Thermo Scientific HyperSep Dispersive SPE Products Efficient sample preparation and clean-up using the QuEChERS Method The QuEChERS method overcomes the problems associated with time consuming, expensive

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF Translated English of Chinese Standard: GB31604.11-2016 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net NATIONAL STANDARD OF GB THE PEOPLE S REPUBLIC OF CHINA GB 31604.11-2016

More information

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS Thermo Scientific ConFlo IV Universal Interface Continuous Flow Interface Isotope Ratio MS 3 ConFlo IV Universal Interface for Continuous Flow Isotope Ratio MS The development of Continuous Flow carrier

More information

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute

More information

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1, Zhiqi Hao 1, Markus Kellmann 2 and Andreas FR. Huhmer

More information

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Customer Application Note: 114 Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Sébastien N. Ronkart, Ph.D.; Société wallonne des eaux, rue de la

More information

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System G. Maio 1, F. Steiner 1, H. Franz 1, M. Tracy 2, X. Liu 2, F. Arnold 1 1 Dionex Softron GmbH, D-21 Germering,

More information