OPTIMIZING METALS SAMPLE PREP

Size: px
Start display at page:

Download "OPTIMIZING METALS SAMPLE PREP"

Transcription

1 APPLICATION NOTE OPTIMIZING METALS SAMPLE PREP Utlilizing Single Reaction Chamber (SRC)Technology for Trace Metals Analysis for contract laboratories. Contract analytical laboratories use a variety of instrumentation for trace metals analysis and must take advantage of method standardization whenever there is an opportunity. To ensure high quality data and a standard operation for a variety of industrial samples, these laboratories use microwave digestion for sample preparation before inductively coupled plasma-optical emission spectrometry (ICPOES) or ICP-mass spectrometry analysis. This tech note describes optimized single-reaction-chamber microwave digestion methods, including vessel selection, sample size, and temperature pro les, along with ICP-OES data. The on-demand priorities and round-the-clock nature of commercial testing laboratories often requires innovative solutions to meet client demand while maintaining high data quality. Laboratory capability has become even more critical with the increased complexity of sample ranges from multiple industries, including consumer products, clinical samples, pharmaceuticals and ne chemicals. A combination of capabilities and diverse instrumentation will allow a contract laboratory to include development and validation of methods for material components such as polymers, metals, foils, inks, adhesives, labels, liners, seals, coatings, toys, food, and active pharmaceutical ingredients. Without question, the goal is to nd a exible, easy-to-use sample preparation approach that is coupled with the capability of determining any element over wide concentration ranges. Standardizing multi-element sample preparation to include little or no method development, higher temperature and pressure exibility, the capability of processing mixed sample batches and a variety of acid combinations simultaneously, and the option of disposable vials offers increases in laboratory productivity to match fast analysis turnaround times. Conventional closed-vessel microwave digestion has been recognized for the last two decades as the most effective technique for the digestion of the widest range of sample types in metals analysis, largely because of its ability to operate at high temperatures and pressures. Because the Page 1

2 an intense amount of labor dedicated to vessel cleaning and vessel assembly and disassembly (1). The most signi cant limitation comes from the conventional closed-vessel approach s inability to process incoming samples simultaneously; samples must be batched by sample type and acid combination. Given the high demand of contract laboratory environments, there has been a heavy emphasis on improving the technology for sample preparation. Figure 1. SRC Microwave Digestion digestions are in a closed vessel, cross-contamination and the loss of volatile elements are eliminated (1). Additionally, microwave sample preparation techniques have overcome challenges of poor digestion quality by providing better digests with low residual acidity, lower carbon content and lower concentrations of dissolved solids, a critical feature necessary for back-end inductively coupled plasma-mass spectrometry (ICP-MS) and optical emission spectrometry (ICP-OES) analysis (2). It has been recently suggested that an ideal sample preparation procedure would be capable of digesting large sample amounts, provide high throughput, be compatible with multi-element analysis, and be safe, green and easy to use (3). Unfortunately, in the context of commercial high-throughput laboratories, conventional closed-vessel microwave digestion has limitations, such as METHODS AND MATERIALS SRC microwave digestion is a relatively new closed-vessel technology that is different from traditional designs that include multiple closed-vessels inside a cavity (4). The single reaction chamber uses a large, pressurized reaction chamber where all samples are digested simultaneously. Through this approach the chamber itself is the microwave cavity, enabling high microwave energy and distribution. It also provides an opportunity to place multiple vials in the closed system and eliminates the need to rotate samples during a digestion program. The chamber is pre-pressured with inert gas, providing a virtual cap and raising the boiling point of any acids used during a digestion run. The most critical result of pre-pressurization is that it enables different sample types and different acid combinations to be processed in a single operation by placing samples in PTFE, quartz, or disposable glass vials and loading them into an autosample rack set to t the chamber. In addition to the bene ts of mixed-sample batches, no loss of volatile elements, the elimination of vessel assembly and an option of using disposable glass vials, this SRC design can operate at very high temperature and pressure up to 300 o C and 199 bar, respectively. Labs can now digest more dif cult samples with a standardized approach consisting of Figure 2. Schematic ow diagram of SRC microwave operation. Page 2

3 digestion of the full range of incoming sample types, with adjustments to the acid combination to account for jewelry or samples of a siliceous nature. Figure 3. Mixed batch SRC microwave diigestion run. the following steps: with any incoming sample (mix of easy and dif cult), determine sample weight; choose the type of vial, acid combination, temperature, and time of digestion; and analyze. This tech note details these key criteria for routine use of SRC microwave digestion in a commercial contract laboratory. Detailed review of the capabilities and use can be found in a number of recent accounts (1,2,4). All digestions were developed using an UltraWAVE microwave oven based on SRC design (Figure 1). The UltraWAVE delivers 1500 W of microwave power to a 1L chamber that can hold sample racks with 5, 15 or 22 vessels depending on the sample and volume. Direct temperature and pressure control of the chamber provides direct control of each sample (all samples reach the same temperature and pressure). A schematic ow diagram of the system s operation is shown in Figure 2. The experiments were performed with a 15-position rack and 15 (15 ml volume) disposable glass vials. An icap 6300 duo ICP-OES system composed of an EMT duo quartz torch, cyclonic glass spray chamber, and Burgener nebulizer was used for the analysis of Cd, Cr, and Pb in all packaging sample matrices (for samples digestion with HF, a ceramic torch, alumina injector, and PTFE nebulizer and spray chamber is used). The operating conditions for the ICP-OES system were as follows: RF applied power to the torch, 1.15 kw; auxiliary ow rate, 0.5 L/min; nebulizer ow rate, 0.65 L/min maximum integration time, 30 s, with dual radial-axial view. MICROWAVE SAMPLE PREPARATION Incoming samples for this study included packaging components containing plastics, polymers, printed-paper and cardboard. Routine contract testing also includes food items and enamels, paint and epoxy for commercial jewelry. A microwave method was developed that would provide a complete Typical packaging samples (0.2 g) were placed in disposable glass test tubes and placed in a 15-position rack for the microwave. The samples were treated with a 5 ml nitric acid, 0.5 ml hydrochloric acid combination and the rack placed in the microwave chamber (note: the PTFE liner contains 110 ml of water and 10 ml of hydrogen peroxide as the microwave load for parameter control). Following prepressur ization of the chamber with nitrogen to 40 bar, the samples were digested using the following time-temperature Table I: ICP-OES analysis of a typical mixed packaging sample set; plastics, polymers, printed paper following SRC digestion (0.2 g, nitric acid: 25 min to 250 o C with 15 min at 250 o C.) Spike result Spike recovery % Spike duplicate result Spike duplicate recovery % Product box m) < <1.29 Container seal <1.63 <1.63 <1.63 Outer plastic cap <1.48 <1.48 <1.48 Inner plastic cap <1.39 <1.39 <1.39 Inner cap cardboard < <1.48 Label < <1.35 Bottle <1.36 <1.36 <1.36 Pet toy bristles (handle) <1.44 Bristles duplicate <1.39 Entree container <1.36 Cardboard < <1.46 Plastic <1.46 <1.46 <1.46 Page 3

4 microwave method: Ramp to 250 o C over 20 min, hold at 250 o C for 15 min, cool to 60 o C and depressurize the chamber for analysis for total time of 50 minutes. To achieve complete digestions for all sample types, two method development digestion runs were performed at 225 o C and 240 o C before the optimized temperature was determined. An example of a typical pro le for the microwave method is shown in Figure 3. All parameters are displayed in real time to ensure high quality, safety, and control. For mixed batches, it is critical that the method account for all sample types. As an example for highly organic samples that are encountered in packaging materials, high temperatures are needed for complete digestion; however, signi cant pressures are generated during the decomposition. During the course of method development for plastic packaging, pressures often reached 100 bar or greater, indicating the importance of choosing conditions that will digest all sample types simultaneously. RESULTS Three separate digestion runs were performed on a variety of mixed packaging samples from commercial sources. Each of the methods described above formed a clear digest, which was diluted with deionized water to 25 ml. These samples were analyzed for Cd, Cr, and Pb using the ICP-OES system. For quality control (QC), laboratory blank,control spikes and duplicates were analyzed as part of ongoing instrument continuing calibration veri cation to go along with periodic sample duplicates. Quantitative data, detection limits, and QC data are shown in Tables I, II, and III. By examining the results, it was easy to determine packaging that required subsequent testing and agging for the customer. Given the high throughput nature of client demand, the standardization of the SRC microwave method aligned well with the use of ICP-OES analysis for a majority of incoming samples. The use of disposable glass vessels, ACS-grade acid and small sample sizes provided a cost effective approach for microwave digestion of ICP-OES analysis. A switch to PTFE or quartz vessels and trace-level acid provides for a path for coupling microwave digestion and ICP- MS analysis for other element panels and lower concentration ranges. CONCLUSION SRC digestion in commercial contract environments for trace metals provided an optimized sample prep method. To achieve complete digestion for all samples ranges simultaneously, an optimized temperature of 250 o C was used following method development iterations at 225 o C and 240 o C. The packaging sample digestions were performed in disposable glass vials to eliminate the need for vessel cleaning in subsequent digestion runs. With a 15 position rack and 50 min digestion time, 3 sets of 12 customer samples, blank, and control spike samples and duplicates were digested and analyzed in a normal working day. Table II: ICP-OES analysis of mixed sample packaging, containers, and lids follwing SRC microwave digestion (0.2 g, 5 ml nitric acid and 0.5 ml hydrochloric acid: 25 min to 250 o C with 15 min at 250 o C) Spike result Spike recovery % Spike duplicate result Spike duplicate recovery % Clear plastic <1.28 Frosted plastic Clear plastic duplicate Frosted plastic duplicate Outer plastic wrapper <1.44 <1.44 <1.44 Large white backing <1.36 <1.36 <1.36 Inner plastic wrapper <1.37 <1.37 <1.37 Small clear backing <1.28 <1.28 <1.28 Outside package <1.41 <1.41 <1.41 Inside Package <1.42 <1.42 <1.42 Paper back < <1.46 Side dish with lid <1.43 Page 4

5 Standardizing the SRC method with an optimized temperature and choice of vessel material and acid combination enables commercial laboratories to process any combination of customer samples without extensive method development, batching or vessel cleaning and assembly. REFERENCES 1. T. Michel, Breaking the Sample Preparation Bottleneck with a New Approach to Microwave Digestion, American Laboratory 42, (2010). 2. J.A. Nobrega, C. Pirola, L.L. Fialho, G. Rota, C.E.K.M.A de Campos Jordao, F. Pollo, Talanta 98, (2012). 3. J.A. Norbrega, G.L. Donati, in: R.A. Meyers (Ed.), Encyclopedia of Analytical Chemistry, John Wiley, Chichester (2011). 4. T. Michel and S. Hussain, Spectroscopy supplement: Applications of ICP & ICP-MS Techniques for Today s Spectroscopists, (November 2012). Table III: ICP-OES analysis of mixed sample plastic cap, bottle, label packaging follwing SRC microwave digestion (0.2 g, 5 ml nitric acid and 0.5 ml hydrochloric acid: 25 min to 250 o C with 15 min at 250 o C) Spike result Spike recovery % Spike duplicate result Spike duplicate recovery % Plastic <1.39 <1.39 <1.39 Cap <1.37 <1.37 <1.37 Bottle < <1.34 Label <1.37 <1.37 <1.37 Paper < <1.44 Plastic (2) <1.45 <1.45 <1.45 Cap (2) <1.42 <1.42 <1.42 Bottle (2) < <1.34 Label (2) <1.38 <1.38 <1.38 Outer Package <1.92 <1.92 <1.92 Wrapper <1.50 <1.50 <1.50 Wrapper <1.36 <1.36 <1.36 Page 5

Test Method: CPSC-CH-E

Test Method: CPSC-CH-E UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 10901 DARNESTOWN RD GAITHERSBURG, MD 20878 Test Method: CPSC-CH-E1001-08 Standard Operating Procedure

More information

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES APPLICATION NOTE ICP-Optical Emission Spectroscopy Author: Nick Spivey PerkinElmer, Inc. Shelton, CT Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES Introduction Micronutrients contained

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

Testing and Calculating. Cannabis. Uncertainty. Analytical Chemistry and Cannabis

Testing and Calculating. Cannabis. Uncertainty. Analytical Chemistry and Cannabis Microwave Digestion Technology for Measuring Heavy Metals Analytical Chemistry and Regulatory and Analytical Challenges for Microbial Testing in VOL 1 NO 3 SEPTEMBER/OCTOBER 2018 Testing and Calculating

More information

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Application Note - AN1508 Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES Introduction Page 1 The combination of hydride generation with

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

H E L P I N G C H E M I S T S. UltraCLAVE. The Next Generation Microwave Digestion System

H E L P I N G C H E M I S T S. UltraCLAVE. The Next Generation Microwave Digestion System H E L P I N G C H E M I S T S UltraCLAVE The Next Generation Microwave Digestion System UltraCLAVE The Next Generation Microwave Digestion System Improving upon an established technology is always a challenge.

More information

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES APPLICATION NOTE 43145 Analysis of trace elements in polymer s to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES Authors Introduction Sanja Asendorf, Application

More information

Basic Digestion Principles

Basic Digestion Principles Basic Digestion Principles 1 From Samples to Solutions Direct Analytical Method Solid Sample Problems: Mech. Sample Preparation (Grinding, Sieving, Weighing, Pressing, Polishing,...) Solid Sample Autosampler

More information

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Prodigy ICP Application Note: 1035 U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Introduction This Application Note describes the capability of the Teledyne Leeman Labs Prodigy High

More information

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy

More information

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Authors Christine M. Rivera Doug Shrader

More information

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Application Note Food Testing Author John Cauduro Agilent Technologies, Australia Introduction Trace

More information

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission 0 chapter Sodium and Potassium Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission Spectroscopy 67 S. S. Nielsen (ed.), Food Analysis Laboratory Manual Springer

More information

Product Descriptions and Offerings

Product Descriptions and Offerings Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental analysis. Laboratories in industries ranging from environmental science, oil and

More information

Laboratory 101: A Guide to Understanding your Testing Laboratory

Laboratory 101: A Guide to Understanding your Testing Laboratory Laboratory 101: A Guide to Understanding your Testing Laboratory Jim Cronin Product Ecology Manager Environmental Monitoring and Technologies, Inc. (EMT) Morton Grove, Illinois Abstract: Laboratory 101:

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

110 Lowell Rd Hudson, NH USA Tel:

110 Lowell Rd Hudson, NH USA Tel: 110 Lowell Rd Hudson, NH 03051 USA Tel: 603-886-8400 www.teledyneleemanlabs.com Product Descriptions and Offerings Teledyne Leeman Labs is a leading innovator of analytical instrumentation for elemental

More information

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009] 2 Vol. 36(1) [Jan. Feb. 2009] BRIEFING h233i Elemental Impurities Procedures. This proposed new general test chapter is the second of two being developed to replace the general test chapter Heavy Metals

More information

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid Application Semiconductor Authors Abe G. Gutiérrez Elemental Scientific 2440 Cuming St Omaha, NE 68131 USA abe@icpms.com

More information

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES TECHNICAL NOTE 43252 Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES Keywords Auto-dilution, Intelligent dilution,

More information

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY C1_C0011 1 INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY Janya Buanuam,* Thareerut Woratanmanon, Temporn Sookawee Regional Failure Analysis and Reliability Laboratory,

More information

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle

More information

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction

More information

Preparation of Food and Feed Samples for Metals Analysis

Preparation of Food and Feed Samples for Metals Analysis White Paper Preparation of Food and Feed Samples for Metals Analysis Alan Cross Technical Specialist, Metals Laboratory Preparation of Food and Feed Samples for Metals Analysis The relative virtues of

More information

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1 Vol. 36(1) [Jan. Feb. 2010] 1 Page 1 of 23 Time:10:03 Date:10/14/09 Instance: g:/pf/production/final PF36(1)/m5193.xml Template:s:/Pf/Template/PFRedesign/Pf353/PFR-pf-server-2009.3f BRIEFING h233i Elemental

More information

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction

More information

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Thermo Scientific icap TQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively

More information

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS Determination of Total Bromine and Iodine Page 2 of 5 1. SCOPE AND FIELD OF APPLICATION The present EuSalt Analytical Standard describes an inductively coupled plasma optical emission spectrometry method

More information

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7 Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes

More information

TECHNETIUM-99 IN WATER

TECHNETIUM-99 IN WATER Analytical Procedure TECHNETIUM-99 IN WATER (WITH VACUUM BOX SYSTEM) 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in water. 1.2. This method does not address all

More information

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental Application Note - AN1305 This application note describes the capability of the Teledyne Leeman Lab s Prodigy7 High-Dispersion ICP for performing analysis according to SW-846 Method 6010C. This method

More information

á233ñ ELEMENTAL IMPURITIES PROCEDURES

á233ñ ELEMENTAL IMPURITIES PROCEDURES Second Supplement to USP 38 NF 33 Chemical Tests / á233ñ Elemental Impurities Procedures 1 á233ñ ELEMENTAL IMPURITIES PROCEDURES INTRODUCTION This chapter describes two analytical procedures (Procedures

More information

Concept of Elemental Impurities (EI) assessment in finished dosage forms by total extraction testing

Concept of Elemental Impurities (EI) assessment in finished dosage forms by total extraction testing Concept of Elemental Impurities (EI) assessment in finished dosage forms by total extraction testing W. Keurentjes, F. Maris, T. Wang, M. van Tiel Merck & Co. Content Introduction How did we start Straight

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically

More information

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry

More information

DISCLAIMER: This method:

DISCLAIMER: This method: Inorganic arsenic determination in fresh mussels using water bath extraction and anion exchange chromatography-inductively coupled plasma mass spectrometry DISCLAIMER: This method: - has to be considered

More information

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM)

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM) Analytical Procedure RADIUM IN WATER (CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM) 1. SCOPE 1.1. This is a method for separation and measurement of radium-226 and radium-228 in water. This method

More information

Determination the elemental composition of soil samples

Determination the elemental composition of soil samples 4. Experiment Determination the elemental composition of soil samples Objectives On this practice you will determine the elemental composition of soil samples by Inductively Coupled Plasma Optical Emission

More information

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Automated Determination of PPQ Levels of Thorium in High Purity Copper PPQ Levels of Th in High Purity Cu TRUFAST ICPMS Automated Determination of PPQ Levels of Thorium in High Purity Copper Using the ESI TRUFAST System and ICPMS Detection By Nathan Saetveit, PhD and Dan

More information

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos

More information

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords

More information

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Innovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer

Innovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer Innovative 0.01 0.1 1 10 Intuitive Concentration (ppb) R = 0.99999 Versatile Hydra II C Mercury Analyzer At Teledyne Leeman Labs; atomic spectroscopy is our business our only business. We are industry

More information

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation

More information

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Organics Revision Date: July 19, 2017 Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM Parameter Perfluorinated Alkyl Acids (Perfluorobutane Sulphonate (PFBS), Perflourooctane Sulphonate (PFOS),

More information

Product Information. Silicon Carbide Well Plates

Product Information. Silicon Carbide Well Plates Silicon Carbide Well Plates Silicon carbide (SiC) is the perfect material to ensure temperature homogeneity in rectangular multiwell reactors in a microwave field. The high microwave absorbance and excellent

More information

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,

More information

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS Application Note Environmental China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS Fast, accurate quantitative analysis of 16 elements using an Agilent 7800 ICP-MS with

More information

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction application Note Atomic Absorption Authors David Bass Senior Product Specialist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT 06484 USA The Determination of Toxic, Trace, and

More information

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full

More information

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are

More information

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent

More information

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

Quality Control Procedures for Graphite Furnace AAS using Avanta Software Application Note Quality Control Procedures for Graphite Furnace AAS using Avanta Software Introduction With the arrival of quality standards in the analytical laboratory, an ever increasing emphasis is

More information

OES - Optical Emission Spectrometer 2000

OES - Optical Emission Spectrometer 2000 OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the

More information

1.1. This is a method for the separation and measurement of 228 Ra in water via its beta emitting 228 Ac daughter.

1.1. This is a method for the separation and measurement of 228 Ra in water via its beta emitting 228 Ac daughter. Analytical Procedure RADIUM-228 IN WATER (WITH VACUUM BOX SYSTEM) 1. SCOPE 1.1. This is a method for the separation and measurement of 228 Ra in water via its beta emitting 228 Ac daughter. 1.2. This method

More information

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Petrochemical Authors Ed McCurdy & Don Potter Agilent Technologies Ltd. Lakeside Cheadle Royal

More information

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction Determination of Nutrients Determination of total phosphorus Extraktion with aqua regia, reflux method Introduction This document is developed in the project Horizontal. It is the result of desk studies

More information

NICKEL-63/59 IN WATER

NICKEL-63/59 IN WATER Analytical Procedure NICKEL-63/59 IN WATER 1. SCOPE 1.1. This is a method for the separation and measurement of nickel- 63/59 in water samples. 1.2. This method does not address all aspects of safety,

More information

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,

More information

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe: February 20, 2012 Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Subject: ICP-MS Report Job Number: S0CHG688 Dear Joe: Please find enclosed the procedure report for the analysis

More information

TECHNETIUM-99 IN SOIL

TECHNETIUM-99 IN SOIL Analytical Procedure TECHNETIUM-99 IN SOIL 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in soil. 1.2. This method does not address all aspects of safety, quality

More information

Not Just for Paint Chips Anymore. High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions

Not Just for Paint Chips Anymore. High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions Not Just for Paint Chips Anymore High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions Objective Develop a sample preparation procedure for analysis of a range of sample

More information

BRIEFING 1730 THEORY AND PRACTICE OUTLINE

BRIEFING 1730 THEORY AND PRACTICE OUTLINE 40(6) In-Process Revision: PLASMA SPECTROCHEMISTRY-THEORY AND... Page 1 of 15 BRIEFING 1730 Plasma Spectrochemistry Theory and Practice. The Chemical Analysis Expert Committee proposes this new

More information

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

Atomic Spectroscopy AA/ICP/ICPMS:

Atomic Spectroscopy AA/ICP/ICPMS: Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics

More information

High purity acids and bases

High purity acids and bases High purity acids and bases acids and bases Ultrapur acids The life science business of Merck KGaA, Darmstadt, Germany operates as MilliporeSigma in the U.S. and Canada. High purity acids and bases, and

More information

Leaching of Heavy Metals Contaminants in Recycled Polyethylene Terephthalate Used for Food Packaging

Leaching of Heavy Metals Contaminants in Recycled Polyethylene Terephthalate Used for Food Packaging Leaching of Heavy Metals Contaminants in Recycled Polyethylene Terephthalate Used for Food Packaging Michael Whitt, Wyatt Brown and Keith Vorst California Polytechnic State University Summary Twenty-two

More information

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer Application note Agriculture Authors Annie Guerin INRA, Laboratoire d Analyses

More information

Graphite Block Acid Digestion System ODLAB

Graphite Block Acid Digestion System ODLAB Graphite Block Acid Digestion System Table of contents 1. company introduction and product development background 2. Hot Plate and Graphite Block Acid Digestion and Microwave Digestion comparison 1 2 3

More information

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),

More information

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Cadion Method Method to 0.30 mg/l Cd TNTplus 852 Cadmium DOC316.53.01013 Cadion Method Method 10217 0.02 to 0.30 mg/l Cd TNTplus 852 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 1 to 60 mg/l COD (ULR) TNTplus 820 Scope and application: For wastewater, process water, surface water, and cooling water.

More information

Trace elemental analysis solutions for your application. June 6, 2018

Trace elemental analysis solutions for your application. June 6, 2018 Trace elemental analysis solutions for your application June 6, 2018 Understanding how each technique works Components of instrument Selection Criteria Application Fields OUTLINE All these techniques can

More information

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs Application note Authors Food safety Kazuhiro Sakai 1, Junichi

More information

Elemental Scientific. Spike Recovery. Elemental Scientific

Elemental Scientific. Spike Recovery. Elemental Scientific Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dissolution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist Innovative Thermal Desorption From Agilent Technologies Vivek Dhyani Application Chemist What is thermal desorption (TD)? How does thermal desorption compare to Head Space? Thermal desorption (TD) is the

More information

TECHNETIUM-99 IN WATER

TECHNETIUM-99 IN WATER Analytical Procedure TECHNETIUM-99 IN WATER (TEVA DISC METHOD) 1. SCOPE 1.1. This procedure describes a method to separate and measure technetium-99 in water. 1.2. This method does not address all aspects

More information

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017

Sample Preparation of Electronic Device Components for Hexavalent Chromium Analysis by IEC Method :2017 Page 1 of 3 Abstract The increasing use of consumer and electronic device components worldwide has drawn increased attention to their impact on the environment. The correct disposal of these materials

More information

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Authors Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Application Note Food safety Craig Jones and Jenny Nelson Agilent Technologies, USA Introduction In the U.S., marijuana remains

More information

Agenda. Recent Advances in Detection of Heavy Metals in Foods. NQAC Organization

Agenda. Recent Advances in Detection of Heavy Metals in Foods. NQAC Organization Agenda Recent Advances in Detection of Heavy Metals in Foods NQAC Nestlé Quality Assurance Centre Supporting factors for reliable results? Daniel Hammer Nestlé Quality Assurance Centre Singapore 2...by

More information

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction. Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

Even wall thickness to with stand maximum G force.

Even wall thickness to with stand maximum G force. 0 3 0 CENTRIFUGE TUBES EASY TO OPEN & EASY TO CLOSE THE UNIQUE GRIP SCREW CAP PROVIDES A SECURE GRIP IN WET CONDITIONS CORROSIVE CHEMICAL PROOF PRINTING NABL APPROVED ACCURATE GRADUATION CENTRIFUGATION

More information

PAR Method Method to 2.0 mg/l Pb TNTplus 850

PAR Method Method to 2.0 mg/l Pb TNTplus 850 Lead DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows all

More information

The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES)

The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES) The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES) Part I: The 1970s Since its commercial inception in 1974, ICP-OES has seen significant technological advancements over

More information

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP Application Note: 40986 Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP Karen Harper, Applications Group Leader, Thermo Scientific, Cambridge,

More information

Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer

Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer Application Note - AN1502 Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer Page 1 Jeff Forsberg, Product Manager, John McQuatters, Mercury Product Specialist,

More information

TITLE: Surface Decontamination of Iodine Compounds (Potassium Iodide (KI)) by DeconGel TM

TITLE: Surface Decontamination of Iodine Compounds (Potassium Iodide (KI)) by DeconGel TM TITLE: Surface Decontamination of Iodine Compounds (Potassium Iodide (KI)) by DeconGel TM ABSTRACT Surface decontamination efficacy determination of DeconGel TM on stainless steel, carbon steel, floor

More information

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document: SPECIFICATION SPECIFICATION Inductively Coupled Plasma Mass Spectrometry System 1. Scope This specification describes the requirements for an Inductively Coupled Plasma Mass Spectrometry System ( System

More information

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS FIELD APPLICATION REPORT ICP MASS SPECTROMETRY Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS Authors: Kenneth R. Neubauer, Ph.D. Perkin Elmer Life and Analytical Sciences 710

More information

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application Environmental Author Neal Julien Midwest Research Institute

More information